    P/N 120-5020G1

    Installation Instructions

    Document No. 8000-PS240-24

    Revision: A, Apr il 19, 2006

    K84

    3Pow

    erSu

    pply

  • 7/31/2019 8000 PS240 24power Supply

    2/14

    Copyright 2005, Del Medical Systems. All rights reserved.

    This document is the property ofDel Medical Systems and contains confidential and proprietaryinformation owned by Del Medical Systems. Any unauthorized copying, use or disclosure of it

    without the prior written permission ofDel Medical Systemsis strictly prohibited.

    Attention: Consult Accompanying Documents - As Applicable

    Del Medical Systems Group

    50B North Gary Avenue Phone:1-847-288-7000

    Roselle, IL 60172 Fax:1-847-288-7011

    USA Toll Free:1-800-800-6006

    www.delmedical.com

  • 7/31/2019 8000 PS240 24power Supply

    3/14

    1.1 Introduction 1-1

    1Specifications,Installation & Service

    1.1 IntroductionThe K-843 Power Supply is a rugged and reliable power supply used to

    supply Del and Universal equipment with auxiliary power.

    The supply has an input of 230/240 VAC and outputs of 24VDC and

    24VAC.

    1.2 Dimensions

    Figure 1-1. Dimensions

    Cable - 30

    2 3/47 cm

    5 1/214 cm

    6 1/216.5 cm

    7/8 (2.2 cm)

    1 3/8 3.5 cm

    1 3/8 (3.5 cm)

    512.7cm

    9.1 M

  • 7/31/2019 8000 PS240 24power Supply

    4/14

    K-843 Power Supply Installation Instructions

    1-2 1.3 Specifications

    1.3 Specifications

    Table 1-1: Specifications

    Specifications

    Compatibility Eureka MC150 Collimator

    Del & Universal Wallstands, Tubestands and Veterinary Systems

    The K-843 Power Supply is to be used in a stationary diagnostic

    x-ray configuration.

    Voltage 230-240 VAC

    Current Input:

    240VAC (0.4A Continuous) [1.25A Momentary] 1.5 min On,

    13.5 min Off

    Output:

    24 VAC, 0-8 Amps Momentary, 1.5 min On, 13.5 min Off

    24 VDC, 3.8 Amps Continuous

    Frequency 50/60 Hz Single PhaseIncoming Power Line 30, 3-Conductor 16 Ga AWG Copper

    Fuse Type Two 3 AG or 3 AB 1.25 Amp 250V T

    Classification Class 1

    Duty Continuous/Intermittent

    Mode of Operation Momentary

    Temperature Limits Transit/Storage Operating 40 F to +158 F +50 F to +95 F 40 C to +70 C +10 C to +35 C

    Relative Humidity Limits Transit/Storage

    10% to 100%Operating10%-80% Non-Condensing

    Atmospheric Limits 14.5 inHg to 30.74 inHg

    500 hPa to 1060 hPA

    Weight 10 lbs (4.4 Kg)

    Degree of protection against the ingress

    of water:

    Ordinary

    Certifications:

    Classified To UL 60601-1,IEC60601-1, EN60601-1, IEC60601-2-32, EN60601-2-32, IEC60601-1-3, EN60601-1-3,

    EN60601-1-2:2000. Certified To CAN/CSA C22.2 NO. 601.1.

    Equipment not suitable for use in the presence of flammable anesthetic mixtures with air, oxygen or nitrous

    oxide.

    No user serviceable parts

  • 7/31/2019 8000 PS240 24power Supply

    5/14

    K-843 Power Supply Installation Instructions

    1.4 Abbreviations 1-3

    1.4 Abbreviations% Percent

    AWG American Wire Gauge

    Btu British Thermal Unit C Degree Celsius

    CE Communauts Europennes

    cm Centimeter

    F Degree Fahrenheit

    ga Gauge

    hPa Hecto Pascal

    inHg Inches Mercury

    Kg Kilogram

    Lb Pound

    M Meter

    max. Maximum

    min. Minimum

    mm Millimeter

    UL Underwriters Laboratories

  • 7/31/2019 8000 PS240 24power Supply

    6/14

  • 7/31/2019 8000 PS240 24power Supply

    7/14

    K-843 Power Supply Installation Instructions

    1.5 Installation 1-5

    4 Run the cables from the equipment that will be supplied power

    though the holes from which the plugs were removed.

    5 Connect the cable wires to the power supply as shown in Figure 1-3.

    Figure 1-3. Power Supply Connections

    6 Reattach cover.

    7 Place a bushing around each cable where the cable enters the cover

    and snap the bushing into the cover.

    Warning

    Be sure that you plug power supply into Hospital Only or

    Hospital Grade receptacle only or injury may result.

    8 Connect power input cable to local type receptacle or hard wire cable

    to circuit breaker box according to local electrical codes.

    9 If necessary, plug power supply cord into wall receptacle. The green

    LED on the power supply will light up to indicate that it is working

    properly.

    24VAC

    +24VDC

    24VDC RET

    Ground

  • 7/31/2019 8000 PS240 24power Supply

    8/14

    K-843 Power Supply Installation Instructions

    1-6 1.6 Replacing Fuses

    1.6 Replacing FusesTools Required:

    Phillips head screwdriver

    small flat-tip screwdriver

    Warning

    Turn off all electrical power to power supply before servicing

    power supply. Also, make sure that power sources are locked out

    and tagged Equipment Being Serviced before servicing power

    supply. You could get seriously injured if you do not.

    1 Turn off all power to power supply.

    2 Unscrew two cover screws (1 in Figure 1-4) and remove cover (2).

    Figure 1-4. Power Supply Cover

    2

    1 (x2)

  • 7/31/2019 8000 PS240 24power Supply

    9/14

    K-843 Power Supply Installation Instructions

    1.6 Replacing Fuses 1-7

    Caution

    Replace fuses only with same type and rating of fuse or machine

    may get damaged.

    3 Pry fuses (F1 & F2) out of their holders.

    4 Reverse steps to reassemble.

    Figure 1-5. Fuse Location

    F1

    F2

    1.25 Amp 250V T @ 240VFor Both Fuses:

  • 7/31/2019 8000 PS240 24power Supply

    10/14

    K-843 Power Supply Installation Instructions

    1-8 1.8 Cleaning Power Supply

    1.8 Cleaning Power SupplyTools Required:

    cleaning wipes

    non-abrasive, hospital-grade cleaner

    Use cleaning wipes and non-abrasive, hospital-grade cleaner to clean

    external surfaces of the K-843 Power Supply.

    Warning

    Turn off all electrical power to power supply at its disconnect

    switch before servicing unit. Also, make sure that disconnect

    switch is locked out and tagged Equipment Being Serviced

    before servicing unit.

    Warning

    This equipment is NOT classified as anaesthetic-proof and may

    ignite flammable anesthetics. Flammable agents used for skin

    cleaning or disinfecting may also produce an explosion hazard.

    Ensure the power has been disconnected before starting any cleaning

    operation.

    Ensure no liquid gets into the unit.

    Do not immerse the equipment, including any components or

    accessories, in liquid.

    Do not autoclave the equipment, including any component or

    accessories.

    Do not use acid or abrasive products.

    Clean painted parts with a cloth and products appropriate for

    cleaning plastic materials; after cleaning wipe the surfaces with a

    clean, dry cloth.

    Do not spray cleaning or disinfection solution directly on the

    equipment. To disinfect, moisten a cloth with a 70% Isopropyl alcohol

    solution or equivalent and wipe the surface of the equipment.

    When disinfecting the examination room, ensure the unit is covered

    with plastic sheets.

  • 7/31/2019 8000 PS240 24power Supply

    11/14

    K-843 Power Supply Installation Instructions

    1.9 Electrical Diagram 1-9

    1.9 Electrical Diagram

    Figure 1-6. Electrical Diagram

  • 7/31/2019 8000 PS240 24power Supply

    12/14

    K-843 Power Supply Installation Instructions

    1-10 1.10 Parts List

    1.10 Parts ListFor your convenience, replacement parts and accessories can be ordered

    from Del Medical Systems by fax 24 hours a day. Please have the

    following information available to ensure quick, easy, and accurate

    service. Your name and telephone number

    Your P.O. (Purchase Order) number

    Your preferred method of delivery

    The part number and quantity of all items required

    To Order by Fax

    Fax your order to Del Medical Systems at 1-800-288-7011. Fax orders can

    be sent 24 hours a day, 7 days a week.

    If you need additional assistance, please call Del Medical Systems at 1-800-

    800-6006 and speak to one our Customer Service Representatives.

    Telephone hours are 8:00 a.m. to 5:00 p.m., Monday through Friday

    (Central Standard or Daylight Time).

    General Part Numbers

    This section contains all part numbers necessary to order K843 power

    supply replacement parts.

    The parts lists follow the illustration for a particular assembly and

    represent components of that assembly. The number listed in the quantity

    column is the number of the specific part required to complete the

    assembly and may not reflect the quantity needed for the entire system.

    The lists are divided into four columns. The item/index numbers refer to

    the identification number located on the drawing. The part number is the

    Del Medical part number, used to identify the part for ordering. The part

    description column lists each part name, and the quantity column lists the

    quantity of that part used in that particular assembly.

  • 7/31/2019 8000 PS240 24power Supply

    13/14

    K-843 Power Supply Installation Instructions

    1.10 Parts List 1-11

    Figure 1-7. Power Supply Layout

    Table 1-2: Parts List

    Fig ref. Part number Description Qty

    2 120-5037G2 Chassis Assembly (See for break-

    down of assembly)

    1

    3 122-5089G1 Transformer Assembly (230V) 1

    6 46-170021P93 Fuse, 1.25A, 250V 2

    7 126-5215G1 Power Supply Cable Assembly 1

    8 623-5010P1 LED, Panel Mount 1

    9 205-5023P1 Top Cover 1

    10 784-12-19100011 Nut, 10-32 Hex 3

    12 768-20-15002511 Screw, #7 X 1/4 Sheet Metal 2

    13 112-5270G2 Spare Fuse Kit (Not Shown) 1

    15 46-208697P47 Bushing, Nylon Heyco #1200 116 126-5216G1 Power Cord Assembly 1

    17 46-208761P1 Cable Tie, .09W X 3.875L 2

    18 407-5031P1 Plug, .625 X .550 2

    8

    17

    109

    18 3

    6

    1612

    2

    15

  • 7/31/2019 8000 PS240 24power Supply

    14/14

    K-843 Power Supply Installation Instructions

    Figure 1-8. Power Supply Chassis Assembly

    Table 1-3: Parts List

    Fig ref. Part number Description Qty

    2 202-5081P1 Plate, Base 24V Power Supply 1

    3 203-5139P1 Bracket, Transformer Mounting 1

    4 408-5183P2 Label, Fuse Rating 1

    5 412-5006P1 Fishpaper Insulator 1

    6 621-5012P1 Rectifier, Bridge, 25A, 400V 1

    7 4455-1174 Fuse Block, 12 Pole .166

    8 4455-1000 Terminal Strip, 4 Pos 1

    9 760-22-19105011 Screw, PHMS SEMS 10-32 X 1/2 4

    10 760-22-14205011 Screw, PPNHMS SEMS 6-32 X 1/2 5

    11 402-5039P1 Label, Earth Ground 1

    11

    3

    47

    10

    5

    9

    10

    6

    10

    2

    10