P/N 120-5020G1
Installation Instructions
Document No. 8000-PS240-24
Revision: A, Apr il 19, 2006
K84
3Pow
erSu
pply
7/31/2019 8000 PS240 24power Supply
2/14
Copyright 2005, Del Medical Systems. All rights reserved.
This document is the property ofDel Medical Systems and contains confidential and proprietaryinformation owned by Del Medical Systems. Any unauthorized copying, use or disclosure of it
without the prior written permission ofDel Medical Systemsis strictly prohibited.
Attention: Consult Accompanying Documents - As Applicable
Del Medical Systems Group
50B North Gary Avenue Phone:1-847-288-7000
Roselle, IL 60172 Fax:1-847-288-7011
USA Toll Free:1-800-800-6006
www.delmedical.com
7/31/2019 8000 PS240 24power Supply
3/14
1.1 Introduction 1-1
1Specifications,Installation & Service
1.1 IntroductionThe K-843 Power Supply is a rugged and reliable power supply used to
supply Del and Universal equipment with auxiliary power.
The supply has an input of 230/240 VAC and outputs of 24VDC and
24VAC.
1.2 Dimensions
Figure 1-1. Dimensions
Cable - 30
2 3/47 cm
5 1/214 cm
6 1/216.5 cm
7/8 (2.2 cm)
1 3/8 3.5 cm
1 3/8 (3.5 cm)
512.7cm
9.1 M
7/31/2019 8000 PS240 24power Supply
4/14
K-843 Power Supply Installation Instructions
1-2 1.3 Specifications
1.3 Specifications
Table 1-1: Specifications
Specifications
Compatibility Eureka MC150 Collimator
Del & Universal Wallstands, Tubestands and Veterinary Systems
The K-843 Power Supply is to be used in a stationary diagnostic
x-ray configuration.
Voltage 230-240 VAC
Current Input:
240VAC (0.4A Continuous) [1.25A Momentary] 1.5 min On,
13.5 min Off
Output:
24 VAC, 0-8 Amps Momentary, 1.5 min On, 13.5 min Off
24 VDC, 3.8 Amps Continuous
Frequency 50/60 Hz Single PhaseIncoming Power Line 30, 3-Conductor 16 Ga AWG Copper
Fuse Type Two 3 AG or 3 AB 1.25 Amp 250V T
Classification Class 1
Duty Continuous/Intermittent
Mode of Operation Momentary
Temperature Limits Transit/Storage Operating 40 F to +158 F +50 F to +95 F 40 C to +70 C +10 C to +35 C
Relative Humidity Limits Transit/Storage
10% to 100%Operating10%-80% Non-Condensing
Atmospheric Limits 14.5 inHg to 30.74 inHg
500 hPa to 1060 hPA
Weight 10 lbs (4.4 Kg)
Degree of protection against the ingress
of water:
Ordinary
Certifications:
Classified To UL 60601-1,IEC60601-1, EN60601-1, IEC60601-2-32, EN60601-2-32, IEC60601-1-3, EN60601-1-3,
EN60601-1-2:2000. Certified To CAN/CSA C22.2 NO. 601.1.
Equipment not suitable for use in the presence of flammable anesthetic mixtures with air, oxygen or nitrous
oxide.
No user serviceable parts
7/31/2019 8000 PS240 24power Supply
5/14
K-843 Power Supply Installation Instructions
1.4 Abbreviations 1-3
1.4 Abbreviations% Percent
AWG American Wire Gauge
Btu British Thermal Unit C Degree Celsius
CE Communauts Europennes
cm Centimeter
F Degree Fahrenheit
ga Gauge
hPa Hecto Pascal
inHg Inches Mercury
Kg Kilogram
Lb Pound
M Meter
max. Maximum
min. Minimum
mm Millimeter
UL Underwriters Laboratories
7/31/2019 8000 PS240 24power Supply
6/14
7/31/2019 8000 PS240 24power Supply
7/14
K-843 Power Supply Installation Instructions
1.5 Installation 1-5
4 Run the cables from the equipment that will be supplied power
though the holes from which the plugs were removed.
5 Connect the cable wires to the power supply as shown in Figure 1-3.
Figure 1-3. Power Supply Connections
6 Reattach cover.
7 Place a bushing around each cable where the cable enters the cover
and snap the bushing into the cover.
Warning
Be sure that you plug power supply into Hospital Only or
Hospital Grade receptacle only or injury may result.
8 Connect power input cable to local type receptacle or hard wire cable
to circuit breaker box according to local electrical codes.
9 If necessary, plug power supply cord into wall receptacle. The green
LED on the power supply will light up to indicate that it is working
properly.
24VAC
+24VDC
24VDC RET
Ground
7/31/2019 8000 PS240 24power Supply
8/14
K-843 Power Supply Installation Instructions
1-6 1.6 Replacing Fuses
1.6 Replacing FusesTools Required:
Phillips head screwdriver
small flat-tip screwdriver
Warning
Turn off all electrical power to power supply before servicing
power supply. Also, make sure that power sources are locked out
and tagged Equipment Being Serviced before servicing power
supply. You could get seriously injured if you do not.
1 Turn off all power to power supply.
2 Unscrew two cover screws (1 in Figure 1-4) and remove cover (2).
Figure 1-4. Power Supply Cover
2
1 (x2)
7/31/2019 8000 PS240 24power Supply
9/14
K-843 Power Supply Installation Instructions
1.6 Replacing Fuses 1-7
Caution
Replace fuses only with same type and rating of fuse or machine
may get damaged.
3 Pry fuses (F1 & F2) out of their holders.
4 Reverse steps to reassemble.
Figure 1-5. Fuse Location
F1
F2
1.25 Amp 250V T @ 240VFor Both Fuses:
7/31/2019 8000 PS240 24power Supply
10/14
K-843 Power Supply Installation Instructions
1-8 1.8 Cleaning Power Supply
1.8 Cleaning Power SupplyTools Required:
cleaning wipes
non-abrasive, hospital-grade cleaner
Use cleaning wipes and non-abrasive, hospital-grade cleaner to clean
external surfaces of the K-843 Power Supply.
Warning
Turn off all electrical power to power supply at its disconnect
switch before servicing unit. Also, make sure that disconnect
switch is locked out and tagged Equipment Being Serviced
before servicing unit.
Warning
This equipment is NOT classified as anaesthetic-proof and may
ignite flammable anesthetics. Flammable agents used for skin
cleaning or disinfecting may also produce an explosion hazard.
Ensure the power has been disconnected before starting any cleaning
operation.
Ensure no liquid gets into the unit.
Do not immerse the equipment, including any components or
accessories, in liquid.
Do not autoclave the equipment, including any component or
accessories.
Do not use acid or abrasive products.
Clean painted parts with a cloth and products appropriate for
cleaning plastic materials; after cleaning wipe the surfaces with a
clean, dry cloth.
Do not spray cleaning or disinfection solution directly on the
equipment. To disinfect, moisten a cloth with a 70% Isopropyl alcohol
solution or equivalent and wipe the surface of the equipment.
When disinfecting the examination room, ensure the unit is covered
with plastic sheets.
7/31/2019 8000 PS240 24power Supply
11/14
K-843 Power Supply Installation Instructions
1.9 Electrical Diagram 1-9
1.9 Electrical Diagram
Figure 1-6. Electrical Diagram
7/31/2019 8000 PS240 24power Supply
12/14
K-843 Power Supply Installation Instructions
1-10 1.10 Parts List
1.10 Parts ListFor your convenience, replacement parts and accessories can be ordered
from Del Medical Systems by fax 24 hours a day. Please have the
following information available to ensure quick, easy, and accurate
service. Your name and telephone number
Your P.O. (Purchase Order) number
Your preferred method of delivery
The part number and quantity of all items required
To Order by Fax
Fax your order to Del Medical Systems at 1-800-288-7011. Fax orders can
be sent 24 hours a day, 7 days a week.
If you need additional assistance, please call Del Medical Systems at 1-800-
800-6006 and speak to one our Customer Service Representatives.
Telephone hours are 8:00 a.m. to 5:00 p.m., Monday through Friday
(Central Standard or Daylight Time).
General Part Numbers
This section contains all part numbers necessary to order K843 power
supply replacement parts.
The parts lists follow the illustration for a particular assembly and
represent components of that assembly. The number listed in the quantity
column is the number of the specific part required to complete the
assembly and may not reflect the quantity needed for the entire system.
The lists are divided into four columns. The item/index numbers refer to
the identification number located on the drawing. The part number is the
Del Medical part number, used to identify the part for ordering. The part
description column lists each part name, and the quantity column lists the
quantity of that part used in that particular assembly.
7/31/2019 8000 PS240 24power Supply
13/14
K-843 Power Supply Installation Instructions
1.10 Parts List 1-11
Figure 1-7. Power Supply Layout
Table 1-2: Parts List
Fig ref. Part number Description Qty
2 120-5037G2 Chassis Assembly (See for break-
down of assembly)
1
3 122-5089G1 Transformer Assembly (230V) 1
6 46-170021P93 Fuse, 1.25A, 250V 2
7 126-5215G1 Power Supply Cable Assembly 1
8 623-5010P1 LED, Panel Mount 1
9 205-5023P1 Top Cover 1
10 784-12-19100011 Nut, 10-32 Hex 3
12 768-20-15002511 Screw, #7 X 1/4 Sheet Metal 2
13 112-5270G2 Spare Fuse Kit (Not Shown) 1
15 46-208697P47 Bushing, Nylon Heyco #1200 116 126-5216G1 Power Cord Assembly 1
17 46-208761P1 Cable Tie, .09W X 3.875L 2
18 407-5031P1 Plug, .625 X .550 2
8
17
109
18 3
6
1612
2
15
7/31/2019 8000 PS240 24power Supply
14/14
K-843 Power Supply Installation Instructions
Figure 1-8. Power Supply Chassis Assembly
Table 1-3: Parts List
Fig ref. Part number Description Qty
2 202-5081P1 Plate, Base 24V Power Supply 1
3 203-5139P1 Bracket, Transformer Mounting 1
4 408-5183P2 Label, Fuse Rating 1
5 412-5006P1 Fishpaper Insulator 1
6 621-5012P1 Rectifier, Bridge, 25A, 400V 1
7 4455-1174 Fuse Block, 12 Pole .166
8 4455-1000 Terminal Strip, 4 Pos 1
9 760-22-19105011 Screw, PHMS SEMS 10-32 X 1/2 4
10 760-22-14205011 Screw, PPNHMS SEMS 6-32 X 1/2 5
11 402-5039P1 Label, Earth Ground 1
11
3
47
10
5
9
10
6
10
2
10