    744Documenting Process Calibrator

    Users Manual

    PN 691287September 1998 Rev.1, 2/99

    1998,1999 Fluke Corporation. All rights reserved. Printed in U.S.A.

    All product names are trademarks of their respective companies.

    99 Washington StreetMelrose, MA 02176Phone781-665-1400Toll Free 1-800-517-8431

    Visit us at www.TestEquipmentDepot.com

    Test Equipment Depot99 Washington Street

    Melrose, MA 02176-6024

    www.testequipmentdepot.com800-517-8431

    781-665-0780 FAX

    http://www.testequipmentdepot.com/http://www.testequipmentdepot.com/http://www.testequipmentdepot.com/http://www.testequipmentdepot.com/http://www.testequipmentdepot.com/http://www.testequipmentdepot.com/http://www.testequipmentdepot.com/http://www.testequipmentdepot.com/http://www.testequipmentdepot.com/http://www.testequipmentdepot.com/http://www.testequipmentdepot.com/http://www.testequipmentdepot.com/
  • 8/12/2019 744 manual

    2/135

    LIMITED WARRANTY & LIMITATION OF LIABILITY

    Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is three years and be-gins on the date of shipment. Parts, product repairs and services are warranted for 90 days. This warranty extends only to the original buyer or end-user customerof a Fluke authorized reseller, and does not apply to fuses, disposable batteries or to any product which, in Flukes opinion, has been misused, altered, neglectedor damaged by accident or abnormal conditions of operation or handling. Fluke warrants that software will operate substantially in accordance with its functionalspecifications for 90 days and that it has been properly recorded on non-defective media. Fluke does not warrant that software will be error free or operate withoutinterruption.

    Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or differentwarranty on behalf of Fluke. Warranty support is available if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable interna-tional price. Fluke reserves the right to invoice Buyer for importation costs of repair/replacement parts when product purchased in one country is submitted forrepair in another country.

    Flukes warranty obligation is limited, at Flukes option, to refund of the purchase price, free of charge repair, or replacement of a defective product which is re-turned to a Fluke authorized service center within the warranty period.

    To obtain warranty service, contact your nearest Fluke authorized service center or send the product, with a description of the difficulty, postage and insuranceprepaid (FOB Destination), to the nearest Fluke authorized service center. Fluke assumes no risk for damage in transit. Following warranty repair, the product willbe returned to Buyer, transportation prepaid (FOB Destination). If Fluke determines that the failure was caused by misuse, alteration, accident or abnormal condi-tion of operation or handling, Fluke will provide an estimate of repair costs and obtain authorization before commencing the work. Following repair, the productwill be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges (FOB Shipping Point).

    THIS WARRANTY IS BUYERS SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUTNOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE FORANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER ARISING FROMBREACH OF WARRANTY OR BASED ON CONTRACT, TORT, RELIANCE OR ANY OTHER THEORY.

    Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, thelimitations and exclusions of this warranty may not apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court of competent

    jurisdiction, such holding will not affect the validity or enforceability of any other provision.

    Fluke Corporation Fluke Europe B.V.P.O. Box 9090 P.O. Box 1186Everett WA 98206-9090 5602 B.D.U.S.A Eindhoven, The Netherlands

    5/94

    Test Equipment Depot99 Washington Street

    Melrose, MA 02176-6024

    www.testequipmentdepot.com800-517-8431

    781-665-0780 FAX

  • 8/12/2019 744 manual

    3/135

    i

    Table of Contents

    Title Page

    Introduction.................................................................................................................... 1

    Standard Equipment...................................................................................................... 3

    Safety Information ...................................... ............................................. ...................... 7

    Getting Started Exercise ............................................................................................... 10

    Operating Features ....................................................................................................... 12

    Input and Output Jacks............................................................................................. 12

    Keys ......................................... ............................................. .................................. 14

    Display...................................................................................................................... 17

    Setting Up the Calibrator............................................................................................... 19

    Using the Strap and Bail........................................................................................... 19

    Charging the Battery................................................................................................. 20

    Battery Life ............................................................................................................... 22Preserving Battery Life ......................................... ............................................. ....... 23

    Using the Optional Battery Eliminator....................................................................... 23

    Selecting the Display Language ....................................... ........................................ 24

    Adjusting the Display Contrast.................................................................................. 24

  • 8/12/2019 744 manual

    4/135

    744Users Manual

    ii

    Displaying the Date and Time ............................................. ...................................... 24

    Using the Backlight ................................................................................................... 26

    Personalizing the Calibrator...................................................................................... 26

    Using Measure Mode..................................................................................................... 28

    Measurement Ranges............................................................................................... 28Measuring Electrical Parameters ........................................ ...................................... 28

    Testing Continuity ........................................ ............................................. ................ 30

    Measuring Pressure ....................................... ............................................. .............. 30

    Measuring Temperature............................................................................................ 34

    Using Thermocouples .......................................................................................... 34

    Using Resistance-Temperature Detectors (RTDs) ........................................ ....... 37

    Measurement Scale ......................................... ............................................. ............ 41

    Linear-Output Transmitters .................................................................................. 41

    Square-Law Process Variables ......................................... ................................... 42Measuring or Sourcing with Custom Units ....................................... ......................... 43

    Using the 700-IV Current Shunt........ ............................................. ........................... 44

    Damping Your Measurements................................................................................... 44

    Using Source Mode ....................................................................................................... 46

    Sourcing Electrical Parameters................................................................................. 46

    Simulating a 4 to 20 mA Transmitter......................................................................... 48

    Supplying Loop Power ....................................... ............................................. .......... 50

    Sourcing Pressure..................................................................................................... 52

    Simulating Thermocouples........................................................................................ 55

    Simulating RTDs .......................................... ............................................. ................ 56Source Scale............................................................................................................. 59

    Linear-Responding Transmitters ...................................... .................................... 59

    Square-Law Process Variables ......................................... ................................... 59

    Stepping and Ramping the Output Value.................................................................. 60

  • 8/12/2019 744 manual

    5/135

    Contents (continued)

    iii

    Using Manual Step .............................................................................................. 60

    Using Auto Step................................................................................................... 61

    Ramping the Output............................................................................................. 62

    Simultaneous Measure/Source ..................................................................................... 65

    Calibrating a Process Instrument .................................................................................. 68Generating As Found Test Data............................................................................. 68

    Adjusting the Transmitter.......................................................................................... 73

    As Left Test Run .................................................................................................... 74

    Test Comments .......................................... ............................................. ................. 75

    Calibrating a Delta-Pressure Flow Instrument .......................................... ................ 75

    Calibrating a Limit Switch ........................................ ............................................. .... 76

    Transmitter Mode ............................................. ............................................. ................ 79

    Memory Operations....................................................................................................... 81

    Saving Results.......................................................................................................... 81Reviewing Memory .......................................... ............................................. ............ 83

    Data Logging ........................................... ............................................. .................... 83

    Recording Min and Max Measurements ............................................. ...................... 86

    Running a Preloaded Task ........................................... ............................................ 86

    Clearing Memory ........................................ ............................................. ................. 86

    Using the Built-in Calculator.......................................................................................... 87

    Saving to and Recalling from the Registers.............................................................. 87

    Using the Calculator to Set the Source Value .......................................................... 88

    Quick Guide to Applications ........................................ ............................................. ..... 88

    Communicating with a PC .......................................... ............................................. ...... 98Maintenance.................................................................................................................. 98

    Replacing the Battery Pack ............................................. ......................................... 98

    Internal Lithium Backup Battery................................................................................ 99

    Cleaning the Calibrator............................................................................................. 99

  • 8/12/2019 744 manual

    6/135

  • 8/12/2019 744 manual

    7/135

    v

    List of Tables

    Table Title Page

    1. Summary of Source and Measure Functions..................................................................... 4

    2. Input/Output Jacks and Connectors................................................................................... 12

    3. Key Functions .................................................................................................................... 15

    4. Battery Life......................................................................................................................... 225. Thermocouple Types Accepted.......................................................................................... 35

    6. RTD Types Accepted ......................................................................................................... 37

    7. Simultaneous MEASURE/SOURCE Functions with Loop Power Disabled ....................... 66

    8. Simultaneous MEASURE/SOURCE Functions with Loop Power Enabled ........................ 67

    9. Replacement Parts............................................................................................................. 101

  • 8/12/2019 744 manual

    8/135

    vii

    List of Figures

    Figure Title Page

    1. Standard Equipment .......................................................................................................... 5

    2. Definition of Symbols ......................................................................................................... 7

    3. Jumper Connections for Demonstration............................................................................. 11

    4. Measure/Source Example.................................................................................................. 115. Input/Output Jacks and Connectors................................................................................... 13

    6. Keys ................................................................................................................................... 14

    7. Elements of a Typical Display............................................................................................ 18

    8. Using the Bail and Installing the Strap............................................................................... 19

    9. Removing the Battery and Using the Charger.................................................................... 21

    10. Electrical Measurement Connections................................................................................. 29

    11. Gage and Differential Pressure Modules ........................................................................... 31

    12. Connections for Measuring Pressure................................................................................. 33

    13. Measuring Temperature with a Thermocouple................................................................... 3614. Using a Jumper Correctly................................................................................................... 39

    15. Measuring Temperature with an RTD................................................................................ 40

    16. Electrical Sourcing Connections ........................................................................................ 47

    17. Connections for Simulating a 4 to 20 mA Transmitter........................................................ 49

  • 8/12/2019 744 manual

    9/135

  • 8/12/2019 744 manual

    10/135

    1

    Documenting Process Calibrator

    Introduction

    The Fluke 744 Documenting Process Calibrator(hereafter referred to as the calibrator) is a battery-

    powered, hand-held instrument that measures andsources electrical and physical parameters, andprovides basic HART

    communicator functions

    when used with HART-capable transmitters. Referto the 744 HART

    Mode Users Guidefor instructions

    on how to use the HART communication feature.

    The calibrator lets you troubleshoot, calibrate, verify,and document your work on process instruments.Calibrator Specifications are at the back of themanual.

    A summary of the measuring and sourcing functionsprovided by the calibrator is shown in Table 1. In

    addition to these functions, the calibrator has thefollowing features:

    General features:

    An analog display to make it easy to readmeasurements when the input is unstable.

    A setup option that lets you set the display toEnglish, French, German, Italian, or Spanish.

    A thermocouple (TC) input/output jack andinternal isothermal block with automaticreference-junction temperature compensation.Or, you can manually enter an external

    temperature reference.

    The ability to store results for later review.

  • 8/12/2019 744 manual

    11/135

    744

    Users Manual

    2

    The ability to automatically log up to 8,000 datapoints.

    A serial computer interface for

    uploading/downloading tasks, lists, and results.Automatic calibration procedures fortransmitters and limit switches using split screenMEASURE/SOURCE mode.

    Transmitter mode in which the calibrator can beconfigured to emulate the functions of a processinstrument.

    Built-in calculator with square-root function, and

    accessible registers containing measure andsource values.

    An optional bar code wand for enteringalphanumeric characters.

    Measuring features:

    Damping (smoothing of the last severalreadings), with display indicator of dampedstatus.

    Display of measurements in engineering units,percent of scale, square-law inputs, or customunits.

    The ability to capture and display minimum andmaximum measured levels.

    Sourcing features:

    The ability to set source values to engineeringunits, percent of scale, square-law outputs, orcustom units.

    Manual and automatic stepping, and an outputramp feature for testing limit switches. Tripdetect is either a 1 V change or a continuitystatus change (Open or Short) from one rampincrement to the next.

    For performance testing and calibration instructionsorder the 74X Series Calibration Manual(PN 602505).

    To contact Fluke, call:

    USA and Canada:1-888-99-FLUKE (1-888-993-5853)

    Europe: +31 402-678-200

    Japan: +81-3-3434-0181

    Singapore: +65-738-5655

    Anywhere in the world: +1-425-356-5500

    Or visit us on the World Wide Web: www.fluke.com

    http://www.fluke.com/http://www.fluke.com/
  • 8/12/2019 744 manual

    12/135

    Documenting Process CalibratorStandard Equipment

    3

    Standard Equipment

    The items listed below and shown in Figure 1 areincluded with your calibrator. If the calibrator isdamaged or something is missing, contact the place

    of purchase immediately. To order replacementparts or spares, see the user-replaceable parts listat the end of this manual.

    TL24 industrial test leads (two sets)

    AC20 test clips (two sets)

    TP20 test probes (one set) HART interface cable

    BP7235 rechargeable nickel-metal hydride pack

    BC7217 battery charger with Instruction Sheet

    Adjustable quick-release strap (PN 946769)

    Jumper for three-wire RTD measurementconnections (two included, PN 944632)

    744 Users Manual

    English (PN 691287)French (PN 691300)

    German (PN 691311)Italian (PN 691318)Spanish (PN 691303)

    744 HARTMode Users Guide

    English (PN 691292)French (PN 691326)German (PN 691334)

    Italian (PN 691337)Spanish (PN 691329)

    DPC/TRACKSoftware utility version withinterface cable (9-pin male-female straight-through, PN 943738).

  • 8/12/2019 744 manual

    13/135

    744

    Users Manual

    4

    Table 1. Summary of Source and Measure Functions

    Function Measure Source

    dc V 0 V to +/-300 V 0 V to 15 V (10 mA max)

    ac V 0 V to 300 V rms, 20 Hz to 5 kHz No sourcingFrequency 1 Hz to 1 kHz (100 mV to 300 V rms)

    1 kHz to 30 kHz (0.5 V to 30 V rms)

    30 kHz to 50 kHz (1 V to 30 V rms)

    0.1 V to 10 V p-p sine wave, or peak square

    wave, 0 Hz to 50 kHz

    Resistance 0to 11 k 0to 11 kdc Current 0 mA to 110 mA 0 to 22 mA (28 V max), sourcing or sinkingContinuity Beep and the word Shortindicates continuity No sourcing

    Thermocouple Types E, N, J, K, T, B, R, S, C, L, or U

    RTD 100Platinum (3926)100Platinum (385)120Nickel (672)

    200Platinum (385)500Platinum (385)1000Platinum (385)

    10Copper (427)100Platinum (3916)

    Pressure 27 modules ranging from 0 to 10 in. H2O

    (2.5 kPa) to 0 to 10,000 psi (69,000 kPa)

    Note

    Loop Power 24 or 28 V (22 mA max)

    Note: Use an external hand pump or other pressure source as a pressure stimulus for the source pressure function.

  • 8/12/2019 744 manual

    14/135

    Documenting Process CalibratorStandard Equipment

    5

    Jumper(2 Black)

    AC20

    Test Clip(2 Red and 2 Black)

    TP20

    Test Probe(1 Red and 1 Black)

    7 8 9

    4 5 6

    1 2 3

    0 .

    V V

    Hz

    TC

    RTD

    CLEAR(ZERO)

    ENTER

    VRTD

    MEASSOURCE

    mA mA

    RTD

    V

    300V

    MAX30VMAX

    TC

    TL24

    Test Leads(2 Red and 2 Black)

    Strap

    ot01f.eps

    Figure 1. Standard Equipment

  • 8/12/2019 744 manual

    15/135

    744Users Manual

    6

    BP7235Nickel-Metal Hydride

    Battery Pack

    HARTInterface

    Cable

    UsersManual

    HART ModeUsers Guide

    BC7217Battery Charger

    ot02f.eps

    Figure 1. Standard Equipment (cont)

  • 8/12/2019 744 manual

    16/135

    Documenting Process CalibratorSafety Information

    7

    Safety Information

    This calibrator is designed and tested in accordancewith IEC1010-1 and CAN/CSA C22.2 No.

    1010.1-92. Use the calibrator only as specified in

    this manual, otherwise the protection provided bythe calibrator may be impaired.

    Symbols used on the calibrator and in this manual

    are explained in Figure 2.

    AC-Alternating Current

    DC-Direct Current

    CAUTION see explanation in manual

    Overvoltage (Installation) Category II, Pollution Degree 2 per

    IEC 1010-1 refers to the level of Impluse Withstand Voltage protection

    provided. Typical locations include; Mains Wall outlets, local appliances

    and PORTABLE EQUIPMENT.

    Fuse

    Common (LO) Input equipotentiality

    Recycling

    Pressure

    ON/OFF

    Equipment protected throughout by DOUBLE INSULATION or

    REINFORCED INSULATION

    Conforms to relevent European Union directives.

    Conforms to relevent Canadian Standards Association directives.

    CAT II

    gj56f.eps

    Figure 2. Definition of Symbols

  • 8/12/2019 744 manual

    17/135

    744Users Manual

    8

    Safety Information (cont)

    A Warning identifies conditions and actions that posehazards to the user; a Caution identifies conditions

    and actions that may damage the calibrator or theequipment under test.

    Warning

    To avoid electric shock or personalinjury, adhere to the following practices:

    Do not use the calibrator if it is damaged.Before you use the calibrator, inspect the

    insulating cover. Look for cracks or missingplastic. Pay particular attention to the

    insulation surrounding the connectors.

    Disconnect the power and discharge all high-voltage capacitors in the equipment undertest before testing resistance or continuity.

    Inspect the test leads for damaged insulation

    or exposed metal. Check test lead continuity.Replace damaged test leads before using thecalibrator.

    Do not use the calibrator if it operatesabnormally. Protection may be impaired.When in doubt, have the calibrator serviced.

    Select the proper function and range for yourmeasurement.

    Use caution when working above 30 V acrms, 42 V ac pk, or 60 V dc. Such voltagespose a shock hazard.

    When using the probes, keep your fingersaway from the probe contacts. Keep your

    fingers behind the finger guards on theprobes.

    Connect the common test lead before youconnect the live test lead. When youdisconnect test leads, disconnect the livetest lead first.

    Replace the battery as soon as there is a lowbattery indication on the display. The

    possibility of false readings can lead toelectric shock and personal injury.

  • 8/12/2019 744 manual

    18/135

    Documenting Process CalibratorSafety Information

    9

    Safety Information (cont)

    Warning (cont)

    Do not apply more than the rated voltage, asmarked on the calibrator, between theterminals, or between any terminal and earthground.

    When using probes, keep your fingersbehind the finger guards.

    Do not use the calibrator with any part of thecase or cover removed.

    Do not operate the calibrator aroundexplosive gas, vapor, or dust.

    When using a pressure module, make surethe process pressure line is shut off anddepressurized before you connect it to ordisconnect it from the pressure module.

    Disconnect test leads before changing toanother measure or source function.

    When servicing the calibrator, use onlyspecified replacement parts.

    Do not use any battery eliminator other than

    the Fluke model BE9005 Battery Eliminator.

    Caution

    To avoid possible damage to thecalibrator or the equipment under test,follow these guidelines:

    Disconnect circuit power and discharge allhigh-voltage capacitors before testing

    resistance, continuity, diodes, orcapacitance.

    Use the proper terminals, function, and rangefor your measurements.

  • 8/12/2019 744 manual

    19/135

    744Users Manual

    10

    Getting Started Exercise

    The following is a brief getting started exercise thatwill make it easier to understand the instructions in

    the rest of the manual.

    1. When you first unpack the calibrator, you willneed to charge the battery. See Figure 9 andcharge the battery for 2 hours.

    2. Reinstall the battery in the calibrator.

    3. Connect the calibrators voltage output to itsvoltage input as follows: connect leftmost pair ofjacks (V RTD SOURCE) to the right most pairof jacks (V MEAS). (See Figure 3.)

    4. Pressto turn on the calibrator. Pressandto adjust the display contrast for the bestlooking display. The calibrator powers up in the

    dc voltage measurement function, and is takingreadings on the V MEAS pair of input jacks.

    5. Pressto switch to the SOURCE screen.The calibrator is still measuring dc voltage, andyou can see the active measurements at the top

    of the display.

    6. Pressto select dc voltage sourcing.Press 5 on the keypad andto beginsourcing 5.0000 V dc.

    7. Now pressto go to the split-screen,simultaneous MEASURE/SOURCE mode. The

    calibrator is simultaneously sourcing dc voltsand measuring dc volts. You can see the

    measurement readings in the top window, andthe active source value in the bottom window asshown in Figure 4.

  • 8/12/2019 744 manual

    20/135

    Documenting Process CalibratorGetting Started Exercise

    11

    7 8 9

    4 5 6

    1 2 3

    0 .

    V V

    Hz

    TC

    RTD

    CLEAR(ZERO)

    ENTER

    TC

    mA mAVRTD

    RTD

    SOURCE

    30V

    MAX

    30V

    MAX

    30V

    MAX

    MEAS

    CAT

    V

    300V

    MAX

    ot03f.eps

    Figure 3. Jumper Connections for Demonstration

    gj4s.eps

    Figure 4. Measure/Source Example

  • 8/12/2019 744 manual

    21/135

    744Users Manual

    12

    Operating Features Input and Output Jacks

    Figure 5 shows the calibrator input and output jacks.Table 2 explains their use.

    Table 2. Input/Output Jacks and Connectors

    No. Name Description

    1 Battery Eliminator jack Jack for the Model BE9005 Battery Eliminator. Use the battery eliminator for bench-top

    applications where ac line power is available. This input does not charge the battery.

    2 SERIAL PORT Connects the calibrator to an RS-232 serial port on a personal computer.

    3 Pressure module

    connector

    Connects the calibrator to a pressure module.

    4 TC input/output Jack for measuring or simulating thermocouples. This jack accepts a miniature

    polarized thermocouple plug with flat, in-line blades spaced 7.9 mm (0.312 in) center to

    center.

    5, 6 MEAS V jacks Input jacks for measuring voltage, frequency, or three- or four-wire RTDs (Resistance

    Temperature Detectors).

    7, 8 SOURCE mA,

    MEAS mA RTDjacks

    Jacks for sourcing or measuring current, measuring resistance and RTDs, and

    supplying loop power.

    9,10 SOURCE V RTDjacksOutput jacks for sourcing voltage, resistance, frequency, and for simulating RTDs.

    D ti P C lib t

  • 8/12/2019 744 manual

    22/135

    Documenting Process CalibratorOperating Features

    13

    7 8 9

    4 5 6

    1 2 3

    0 .

    V

    MEAS

    SOURCE SETUP mA

    V

    Hz

    TC

    RTD

    CLEAR(ZERO)

    ENTER

    TC

    10

    6

    4

    1

    2

    3

    79 8

    5

    mA mAVRTD

    RTD

    SOURCE

    30V

    MAX

    30V

    MAX

    30V

    MAX

    MEAS

    CAT

    V

    300V

    MAX

    DOCUMENTING PROCESS CALIBRATOR744

    ot05f.eps

    Figure 5. Input/Output Jacks and Connectors

  • 8/12/2019 744 manual

    23/135

    Documenting Process Calibrator

  • 8/12/2019 744 manual

    24/135

    Documenting Process CalibratorOperating Features

    15

    Table 3. Key Functions

    No. Name Description

    1 key Cycles the calibrator through MEASURE, SOURCE, and MEASURE/SOURCE modes.

    2 key Selects mA (current) measure or source function. For loop power on/off, go to the Setupmode.

    3 key Enters and exits Setup mode to modify operating parameters.

    4 Softkeys Perform the function defined by the label above each key on the display.

    5 key Turns the backlight on and off.

    6 key Selects the pressure measurement or sourcing function.7 key Selects TC (thermocouple) or RTD (resistance temperature detector) measurement or

    sourcing functions.

    8 key Toggles between HART communication mode and analog operation. In calculator mode, thiskey provides the square root function.

    9 key Clears a partial data entry, or zeros the output when in the SOURCE mode. When using apressure module, zeros the pressure module reading.

    744

  • 8/12/2019 744 manual

    25/135

    744Users Manual

    16

    Table 3. Key Functions (cont)

    No. Name Description

    10 ,, , and keys

    Adjust the display contrast. Make choices from lists on the display. Increase or decrease the source level when using the step feature. In calculator mode, provide arithmetic functions (+ - ).

    11 key Terminates a numeric entry when setting a source value, or confirms your choice in alist. In calculator mode, provides the equals arithmetic operator (=).

    12 key Toggles between resistance and continuity functions in MEASURE mode, or selectsthe resistance function in SOURCE mode.

    13 Numeric keypad Used whenever a numeric entry is required.

    14 key Toggles between ac voltage and frequency functions in MEASURE mode, or selectsfrequency output in SOURCE mode.

    15 key Selects the dc voltage function in MEASURE mode, or selects dc voltage in SOURCEmode.

    16 key Turns the power on and off.

    Documenting Process Calibrator

  • 8/12/2019 744 manual

    26/135

    Documenting Process CalibratorOperating Features

    17

    Display

    Figure 7 shows the features of a typical display. Thedisplay shown is MEASURE mode. Near the top of

    the display is Source Off. This is the area of thedisplay that shows what is happening in the othermode (SOURCE or MEASURE). The other parts ofthe display are as follows:

    Status Bar: shows the time and date (if set inSetup mode), and shows the status of LoopPower, Battery Save, and Backlight Timeout; all

    of which are set in Setup mode. The low battery

    and backlight on symbols also appear here.

    Mode Indicator: Shows whether the calibrator isin MEASURE or SOURCE mode. In split screenMEASURE/SOURCE mode, there is a ModeIndicator for each window.

    Measured Value: Shows the measured value inyour choice of engineering units or percent ofscale.

    Range Status: Shows whether Auto Range is on,and what range is currently being used.

    Custom Units Indicator: Shows that the displayedunits are custom. The original engineering unitsof the measure or source function are notdisplayed.

    Secondary Value: Shows the measure or sourcevalue in original engineering units wheneverscaling or custom units are active.

    744

  • 8/12/2019 744 manual

    27/135

    744Users Manual

    18

    Measured Value

    Softkey Labels

    Status Bar

    Time and Date DisplayBattery Gauge

    Loop Power Annunciator

    Mode Indicator

    Custom Units Indicator

    Secondary Value

    Source Status

    Undamped (Unsettled)

    Indicator

    ot07c.eps

    Figure 7. Elements of a Typical Display

    Documenting Process Calibrator

  • 8/12/2019 744 manual

    28/135

    ocu e t g ocess Ca b atoSetting Up the Calibrator

    19

    Setting Up the Calibrator

    Using the Strap and Bail

    After you unpack the calibrator, attach its carryingstrap as shown in Figure 8. You can adjust the strapas necessary to hang the calibrator on any sturdy

    support. Figure 8 also shows you how to open thebail to stand the calibrator at a comfortable viewing

    angle for benchtop use.

    gj8f.eps

    Figure 8. Using the Bail and Installing the Strap

    744

  • 8/12/2019 744 manual

    29/135

    Users Manual

    20

    Charging the Battery

    Before you use the calibrator for the first time,charge its battery pack in the external battery

    charger. The Model BC7217 charger charges boththe NiMH and Ni-Cd battery packs.

    Figure 9 shows how to remove the battery. Removethe battery door and tap the calibrator with yourhand to get the battery out. Place the battery in thecharger and connect the charger to line power. Thecharger automatically senses line voltage and

    adjusts itself accordingly.

    A discharged battery is fully charged in 2 hours orless in fast-charge mode (steady indicator light on

    the charger). Full charge is maintained after thattime in trickle-charge mode (blinking indicator lighton the charger). Switching between charging modes

    is automatic. You can leave the battery pack on

    trickle charge indefinitely without damage.

    Note

    When you remove a charged battery fromthe charger, wait for the blinking indicator togo off before you insert a dischargedbattery. It takes about 2 seconds for thebattery charger to reset.

    Documenting Process Calibrator

  • 8/12/2019 744 manual

    30/135

    gSetting Up the Calibrator

    21

    BATTERYACCESS TTERY CCESS

    BATTERYACCESS TTERY CCESS

    BATTERYACCESS TTERY CCESS

    1 2

    3

    4

    gj9f.eps

    Figure 9. Removing the Battery and Using the

    Charger

    744

  • 8/12/2019 744 manual

    31/135

    Users Manual

    22

    Battery Life

    The calibrator senses which battery is installed(standard NiMH, or optional Ni-Cd model BP 7217).

    If the NiMH battery is installed, a battery gauge bargraphshows on the upper right of the display.If the Ni-Cd battery is installed, there is no batterylevel indication except for a low battery symbol,that appears when it is time to charge the battery.

    Table 4 shows the typical operating time for a new,fully charged NiMH (Nickel-Metal Hydride) battery

    pack. The optional Ni-Cd battery (Model BP7217)

    provides approximately half the battery life of the

    NiMH battery. Calibrator performance is guaranteedto meet specifications until the battery gauge readsempty (or).

    To replace the battery, refer to Replacing theBattery later in this manual for instructions. In thecase of the Ni-Cd battery, for longest battery life andbest performance, wait forto appear before youcharge the battery.

    Table 4. Typical Battery Life with Standard NiMH Battery Pack

    Operating Modes Backlight Off Backlight On

    Measure, continuous 13 Hours 12 Hours

    Measure and source, with loop power on, continuous 7 Hours 6 Hours

    Typical intermittent operation >16 Hours >16 Hours

    Documenting Process Calibrator

  • 8/12/2019 744 manual

    32/135

    Setting Up the Calibrator

    23

    Preserving Battery Life

    An optional Auto Battery Savefeature turns thecalibrator off after a selected idle time. The default

    setting for Auto Battery Saveis Off. When AutoBattery Saveis On, and you are using the optional Ni-Cd battery pack (Model BP7217), thesymbolshows in the upper right corner of the display. Thesetting is preserved after you turn off the power. AutoBattery Save works the same when using the batteryeliminator. Turn on the Auto Battery Savefeature asfollows:

    1. Press.2. Pressto highlight Offfollowing Auto Battery

    Save.

    3. Pressor the Choicessoftkey.

    4. Press tohighlight On, then press.

    5. To accept the timeout period shown on thedisplay, you can finish here. Press Doneto exitSetup mode and do not go on to step 6.

    6. To change the timeout period, presstohighlight the timeout period following BatterySave Timeout.

    7. Pressor the Choicessoftkey.8. Enter your choice of timeout period in minutes

    (accepted range: 1 to 120 minutes).

    9. Press the Donesoftkey.

    10. Press the Donesoftkey orto exit Setupmode.

    Using the Optional Battery EliminatorCaution

    To avoid damage to the calibrator, useonly Fluke Model BE9005 Series BatteryEliminator, available from your Flukerepresentative.

    744

  • 8/12/2019 744 manual

    33/135

    Users Manual

    24

    Where ac power is available, you can use theoptional Fluke Model BE9005 Battery Eliminator toconserve battery power. When the battery eliminatoris used, the battery is internally disconnected, and

    can be removed from the calibrator. The batteryeliminator does not charge the battery. The batteryeliminator is handy for troubleshooting processinstruments on the workbench, and for long-termdata logging. When you calibrate an instrument, youwill get best results using battery power.

    Selecting the Display Language

    The calibrator displays information in five languages.English is the default. To change the displaylanguage, proceed as follows:

    1. Press.

    2. Press the third softkey from the left twice.

    3. Pressthree times.

    4. Press.

    5. Pressorto highlight your choice oflanguage.

    6. Pressto confirm your choice. Thelanguage you choose is the power-up default.

    7. Pressto exit Setup mode.

    Adjusting the Display Contrast

    Pressor to increase contrast. Pressor todecrease contrast. When theandkeys arebeing used to select an item from a list, for examplein Setup mode, use the or keys. In calculatormode, all four direction keys are used for arithmetic

    functions.

    Displaying the Date and TimeThe date and time can be shown at the top of thedisplay during normal operation. In Setup mode youcan turn this date and time display on or off. You canalso control the format used to display the date andtime. You should set the calendar and clock whetheror not you use the date and time display, since a

    timestamp is applied to all saved results.

    Documenting Process CalibratorS tti U th C lib t

  • 8/12/2019 744 manual

    34/135

    Setting Up the Calibrator

    25

    Proceed as follows to set up the time and datedisplays:

    1. Press.

    2. Press the Next Pagesoftkey. The displayappears as follows:

    gj38s.eps

    3. Use theandkeys to move the cursor tothe parameter you want to change, then press

    or the Choicessoftkey to choose asetting for that parameter. For example, thefollowing display appears after you select DateFormat:

    gj39s.eps

    4. Pressorto move the cursor to the desired

    date format.

    5. Pressto go back to thedisplay.

    6. Make another selection or press the Donesoftkey orto save your settings and exitSetup mode.

    744Users Manual

  • 8/12/2019 744 manual

    35/135

    Users Manual

    26

    Using the Backlight

    Pressto toggle the display backlight on and off.When the backlight is on, the symbol shows at the

    top of the display. You can minimize battery usageby setting the calibrator to turn the display backlightoff automatically. When the backlight is on and AutoBacklight Off is activated, thesymbol shows atthe top of the display. To automatically turn off thebacklight after a set time, proceed as follows:

    1. Press.

    2. Pressto highlight Offfollowing Auto Backlight

    Off.

    3. Pressor the Choicessoftkey.

    4. Pressto highlight On, then press.

    5. To accept the timeout period shown on the

    display, press Doneto exit, and do not go on tostep 6.

    6. To change the timeout period, presstohighlight the timeout period following BacklightTimeout.

    7. Pressor the Choicessoftkey.

    8. Enter your choice of timeout period in minutes

    (accepted range: 1 to 120 minutes).

    9. Press the Donesoftkey.

    10. Press the Donesoftkey orto exit Setupmode.

    Personalizing the Calibrator

    You can load your name or some other alphanumeric

    identifier into the calibrator to be displayed at power-up and in saved results. Proceed as follows to loadan identifier:

    1. Press.

    2. Press Next Pagetwice.

    3. Pressto move the cursor to the same lineas ID.

    Documenting Process CalibratorSetting Up the Calibrator

  • 8/12/2019 744 manual

    36/135

    Setting Up the Calibrator

    27

    4. Pressor the Choicessoftkey. Thedisplay appears as follows:

    gj40s.eps

    5. The IDstring is shown at the bottom of the boxedarea. To erase a character, press the BackSpacesoftkey. To erase the whole string, press.

    6. Press,, , or to select a character, thenpress. Use the numeric keypad if youwant to enter a number.

    7. Repeat step 6 until you are satisfied with the IDstring appearing in the window.

    8. Press the Donesoftkey.

    9. Press the Donesoftkey orto exit Setupmode.

    744Users Manual

  • 8/12/2019 744 manual

    37/135

    Users Manual

    28

    Using Measure Mode

    Note

    To achieve best noise rejection and highestaccuracy performance, do not use thebattery eliminator, and tie all three common

    jacks together.

    The operating mode (i.e., MEASURE, SOURCE) is

    shown in a reverse-video bar on the display. If thecalibrator is not in MEASURE mode, pressuntilMEASURE is shown. You must be in MEASUREmode to change any of the MEASURE parameters.

    Measurement Ranges

    The calibrator normally changes to the appropriatemeasurement range automatically. The lower right

    side of the display shows either Range or AutoRange depending on the range status. Auto Rangeswitch points are shown in the specifications at the

    end of this manual. When you press the Rangesoftkey, the range is locked. Press it again to cycle toand lock on the next higher range. Auto Range is

    reactivated when you select another measurement

    function.

    If the range is locked, overrange inputs produce adisplay of - - - - - -. In Auto Range, out of range inputsproduce a display of ! ! ! ! ! !.

    Measuring Electrical Parameters

    When you turn on the calibrator, it powers up in the

    dc voltage measurement function. Figure 10 shows

    electrical measurement connections. To select anelectrical measurement function from eitherSOURCE or MEASURE/SOURCE mode, first pressfor MEASURE mode, then proceed as follows:

    1. Pressfor current,for dc voltage,once for ac voltage or twice for frequency, or

    for resistance.

    Documenting Process CalibratorUsing Measure Mode

  • 8/12/2019 744 manual

    38/135

    Using Measure Mode

    29

    TC

    mA mA VVRTD

    RTD

    SOURCE 300VMAX30V

    MAX30VMAX

    30VMAX

    MEAS

    CAT

    TC

    mA mA VVRTD

    RTD

    SOURCE 300VMAX30V

    MAX30VMAX

    30VMAX

    MEAS

    CAT

    Red

    Red

    Black

    Black

    +

    gj10f.eps

    Figure 10. Electrical Measurement Connections

    744Users Manual

  • 8/12/2019 744 manual

    39/135

    Users Manual

    30

    Note

    When measuring frequency, you areprompted to select a frequency range. If youexpect the frequency you are measuring to

    be below 20 Hz, pressto select thelower frequency range, then press.

    2. Connect the test leads as shown in Figure 10,depending on the measurement function.

    Testing Continuity

    When testing continuity, the beeper sounds and the

    word Shortappears on the display when theresistance between the MEAS jack and itscommon jack is less than 25. The word Openappears when the resistance is greater than 400.Proceed as follows to test continuity:

    1. Remove power from the circuit to be tested.

    2. If necessary, pressfor MEASURE mode.

    3. Presstwice so that Openappears.

    4. Connect the calibrator to the circuit to be testedas Figure 10 shows.

    Measuring Pressure

    Many ranges and types of pressure modules areavailable from Fluke. See Accessories near the

    back of this manual. Before you use a pressuremodule, read its Instruction Sheet. The modulesvary in how you use them, how you zero them, whattypes of process pressure media are allowed, andaccuracy specification.

    Figure 11 shows gage and differential modules.Differential modules also work in gage mode by

    leaving the low fitting open to atmosphere.

    To measure pressure, attach the appropriatepressure module for the process pressure to betested as described in the modules InstructionSheet.

    Proceed as follows to measure pressure:

    Warning

    To avoid a violent release of pressure in

    a pressurized system, shut off the valveand slowly bleed off the pressure beforeyou attach the pressure module to thepressure line.

    Documenting Process CalibratorUsing Measure Mode

  • 8/12/2019 744 manual

    40/135

    g

    31

    Differential

    HighLow

    15 PSID/G

    BURSTPRESSURE45PSIG

    PRESSUREM ODULE700P04

    1bar100kPaRANGE

    Gage

    RANGE 100 PSIG

    BURSTPRESSURE300 PSIG

    PRESSURE MODULE700P06

    7bar700kPa

    gj11f.eps

    Figure 11. Gage and Differential Pressure Modules

    Caution

    To avoid mechanically damaging the

    pressure module, never apply more than10 ft.-lb. of torque between the pressure

    module fittings, or between the fittingsand the body of the module. Alwaysapply appropriate torque between thepressure module fitting and connectingfittings or adapters.

    To avoid damaging the pressure modulefrom overpressure, never apply pressureabove the rated maximum printed on the

    pressure module.

    To avoid damaging the pressure modulefrom corrosion, use it only with specifiedmaterials. Refer to the printing on thepressure module or the pressure module

    instruction sheet for the acceptablematerial compatibility.

    744Users Manual

  • 8/12/2019 744 manual

    41/135

    32

    1. Connect a pressure module to the calibrator asshown in Figure 12. The threads on thepressure modules accept standard NPT pipefittings. Use the supplied NPT to ISO

    adapter if necessary.

    2. Pressfor MEASURE mode.

    3. Press. The calibrator automatically senseswhich pressure module is attached and sets itsrange accordingly.

    4. Zero the pressure module as described in themodules Instruction Sheet. Modules vary in

    zeroing procedures depending on module type.You MUST perform this step before you executea task that sources or measures pressure.

    5. If desired, you can change pressure displayunits to psi, mHg, inHg, mH

    2O, inH

    2O@,

    inH2O@60F, ftH

    2O, bar, g/cm

    2, or Pa. Metric

    units (kPa, mmHg, etc.) are shown in Setup

    mode in their base units (Pa, mHg, etc.).Change the pressure display units as follows:

    a. Press.

    b. Press Next Pagetwice.

    c. Pressor the Choicessoftkey withthe cursor on Pressure Units.

    d. Select the pressure units withor.e. Press.

    f. Press Done.

    Documenting Process CalibratorUsing Measure Mode

  • 8/12/2019 744 manual

    42/135

    33

    7 8 9

    4 5 6

    1 2 3

    0 .

    V

    MEAS

    SOURCE SETUP mA

    V

    Hz

    TC

    RTD

    CLEAR

    (ZERO)

    ENTER

    TC

    Differential

    Module

    Tank

    HL

    Isolation

    Valve

    Gage

    Module

    mA mA VV

    RTD

    RTD

    SOURCE 300V MAX30V

    MAX

    30V

    MAX

    30V

    MAX

    MEAS

    CAT

    DOCUMENTING PROCESS CALIBRATOR744

    ot12c.eps

    Figure 12. Connections for Measuring Pressure

    744Users Manual

  • 8/12/2019 744 manual

    43/135

    34

    Measuring Temperature

    Using Thermocouples

    The calibrator supports eleven standard

    thermocouples, each identified with an alphacharacter: E, N, J, K, T, B, R, S, C, L, or U. Table 5summarizes the ranges and characteristics of thesupported thermocouples.

    To measure temperature using a thermocouple,proceed as follows:

    1. Attach the thermocouple leads to the appropriateTC miniplug, then to the TC input/output as

    shown in Figure 13. One pin is wider than theother. Do not try to force a miniplug in the wrongpolarization.

    Note

    If the calibrator and the thermocouple plugare at different temperatures, wait oneminute or more for the connectortemperature to stabilize after you plug theminiplug into the TC input/output.

    2. If necessary, pressfor MEASURE mode.

    3. Press. The display prompts you to selectthe thermocouple type.

    4. Select the desired thermocouple type using theorfollowed by.

    5. If necessary, you can change between C or FTemperature Unitsas follows:

    a. Press.

    b. Press the Next Pagesoftkey twice.

    c. Use theandkeys to move the cursorto the desired parameter. Then press eitheror the Choicessoftkey to choose asetting for that parameter.

    d. Pressorto move the cursor to thedesired setting.

    e. Pressto go back to thedisplay.

    f. Press the Done softkey orto exit Setupmode.

    6. If necessary, you can change between ITS-90orIPTS-68 Temperature Scalein Setup mode. The

    procedure is the same as steps a-f above.

    Documenting Process CalibratorUsing Measure Mode

  • 8/12/2019 744 manual

    44/135

    35

    Table 5. Thermocouple Types Accepted

    Type Positive Lead

    Positive Lead (H)

    Color Negative Lead Specified Range

    Material ANSI* IEC** Material (C)

    E Chromel Purple Violet Constantan -250 to 1000

    N Ni-Cr-Si Orange Pink Ni-Si-Mg -200 to 1300

    J Iron White Black Constantan -210 to 1200

    K Chromel Yellow Green Alumel -270 to 1372

    T Copper Blue Brown Constantan -250 to 400

    B Platinum (30% Rhodium) Gray Platinum (6% Rhodium) 600 to 1820

    R Platinum (13% Rhodium) Black Orange Platinum -20 to 1767

    S Platinum (10% Rhodium) Black Orange Platinum -20 to 1767

    C *** Tungsten (5% Rhenium) White Tungsten (26% Rhenium) 0 to 2316

    L (DIN J) Iron Constantan -200 to 900

    U (DIN T) Copper Constantan -200 to 600

    *American National Standards Institute (ANSI) device negative lead (L) is always red.

    **International Electrotechnical Commission (IEC) device negative lead (L) is always white.

    *** Not an ANSI designation but a Hoskins Engineering Company designation.

    744Users Manual

  • 8/12/2019 744 manual

    45/135

    36

    TC

    mA mA VVRTD

    RTD

    SOURCE 300VMAX30V

    MAX

    30V

    MAX

    30VMAX

    MEAS

    CAT Process Temperature

    Warning

    30V maximum toTC Miniplug

    gj12f.eps

    Figure 13. Measuring Temperature with a Thermocouple

    Documenting Process CalibratorUsing Measure Mode

  • 8/12/2019 744 manual

    46/135

    37

    Using Resistance-Temperature Detectors(RTDs)

    The calibrator accepts RTD types shown in Table 6.RTDs are characterized by their resistance at 0 C(32 F), which is called the ice point or R

    0. The

    most common R0is 100. Most RTDs come in a

    three-terminal configuration. The calibrator acceptsRTD measurement inputs in two-, three-, or four-wireconnections as shown in Figure 15. A four-wire

    configuration provides the highest measurement

    precision, and two-wire provides the lowestmeasurement precision.

    Table 6. RTD Types Accepted

    RTD Type Ice Point (R0) Material Range (C)

    Pt100 (3926) 100 Platinum 0.003926 /C -200 to 630

    *Pt100 (385) 100 Platinum 0.00385 /C -200 to 800

    Ni120 (672) 120 Nickel 0.00672 /C -80 to 260

    Pt200 (385) 200 Platinum 0.00385 /C -200 to 630

    Pt500 (385) 500 Platinum 0.00385 /C -200 to 630

    Pt1000 (385) 1000 Platinum 0.00385 /C -200 to 630

    Cu10 (427) 9.035 ** Copper 0.00427 /C -100 to 260

    Pt100 (3916) 100 Platinum 0.003916 /C -200 to 630

    *Per IEC 751-Standard **10 @ 25 C

    744Users Manual

  • 8/12/2019 744 manual

    47/135

    38

    To measure temperature using an RTD input,proceed as follows:

    1. If necessary, pressfor MEASURE mode.

    2. Presstwice so that Select RTD Typeshows.

    3. Pressorto select the desired RTD type.

    4. Press.

    5. Pressorto select a 2-, 3-, or 4- wireconnection.

    6. Attach the RTD to input jacks as the display or

    Figure 15 shows. Use the supplied jumperbetween the mA RTD MEAS low jack and theV MEAS low jack as shown if you are using a 3-wire connection.

    7. Press.

    Caution

    Do not force a dual banana plug between

    any two jacks in the horizontalorientation. Doing so will damage the

    jacks. Use the supplied jumper wire whenneeded for RTD measurements. You canuse a dual banana plug in the verticalorientation. See Figure 14.

    Documenting Process CalibratorUsing Measure Mode

  • 8/12/2019 744 manual

    48/135

    39

    TC

    mA mAV

    RTDRTD

    SOURCE

    30VMAX

    30VMAX

    30V

    MAX

    MEAS

    CAT

    V

    300V

    MAX

    OK

    VRTD

    MEASSOURCE

    mA mA

    RTD

    V

    300V

    MAX30V MAX

    TC

    VRTD

    MEASSOURCE

    mA mA

    RTD

    V

    300V

    MAX30V MAX

    TC

    WRONG

    gj14f.eps

    Figure 14. Using a Jumper Correctly

    8. If necessary, you can change between C or Ftemperature units in Setup mode as follows:

    a. Press.

    b. Press the Next Pagesoftkey twice.

    c. Use theandkeys to move the cursorto the parameter you which to change, thenpressor the Choicessoftkey tochoose a setting for that parameter.

    d. Pressorto move the cursor to thedesired setting.

    e. Pressto go back to thedisplay.

    f. Press the Donesoftkey orto exit Setupmode.

    9. If necessary, you can change between ITS-90 orIPTS-68 Temperature Scalein Setup mode. Theprocedure is the same as steps a through fabove.

    744Users Manual

  • 8/12/2019 744 manual

    49/135

    40

    TC

    mA mA VVRTD

    RTD

    SOURCE 300VMAX30V

    MAX30VMAX

    30VMAX

    MEAS

    CAT

    TC

    mA mA VVRTD

    RTD

    SOURCE 300VMAX30V

    MAX30VMAX

    30VMAX

    MEAS

    CAT

    TC

    mA mA VVRTD

    RTD

    SOURCE 300VMAX30V

    MAX30VMAX

    30VMAX

    MEAS

    CAT

    RTD

    RTD

    RTD

    gj15f.eps

    Figure 15. Measuring Temperature with an RTD

    Documenting Process CalibratorUsing Measure Mode

  • 8/12/2019 744 manual

    50/135

    41

    Measurement Scale

    This feature lets you scale the measurements inaccordance with a particular process instruments

    response. Percent of scale works for linear-output

    transmitters or square-law transmitters such asdifferential pressure transmitters that report flow rate.

    Linear-Output Transmitters

    1. If necessary, pressfor MEASURE mode.

    2. Select a measurement function (,,,,, or) as previously described.

    3. Press the Scalesoftkey.

    4. Select % scalefrom the list.

    5. Use the numeric keypad to enter the 0% of scalevalue (0% Value).

    6. Press.

    7. Use the numeric keypad to enter the 100% of

    scale value (100% Value).

    8. Press.

    9. Press the Donesoftkey.

    Percent of scale remains in effect until you change to

    another measurement function or press the Scalesoftkey and select another scale mode.

    744Users Manual

  • 8/12/2019 744 manual

    51/135

    42

    Square-Law Process Variables

    When you select within scaling, the calibrator takesthe square root of its input and displays the

    measurement in percent. For example, when

    connected to the output of a delta-pressuretransmitter, the calibrator reading is proportional toflow rate.

    1. If necessary, pressfor MEASURE mode.

    2. Select a measurement function (,,,,, or) as previouslydescribed.

    3. Press the Scalesoftkey.

    4. Select scalefrom the list.

    5. Use the numeric keypad to enter the 0% of scalevalue (0% Value).

    6. Press.

    7. Use the numeric keypad to enter the 100% ofscale value (100% Value).

    8. Press.

    9. Set Modeto .

    10. Press.

    11. Press the Donesoftkey.

    Square root percent of scale remains in effect untilyou change to another measurement function orpress the Scalesoftkey and select another scalemode.

    Documenting Process CalibratorUsing Measure Mode

  • 8/12/2019 744 manual

    52/135

    43

    Measuring or Sourcing with CustomUnits

    Warning

    To avoid possible electric shock, whenusing Custom Units for measurement,always refer to the secondary valuedisplayed below and to the right of themain display for the actual value of themeasurement in native engineering units.

    You can set up the measurement or source displayto show your own custom units. To do this, you

    select a function, for example mV dc, scale it as youchoose, then enter an alphanumeric name for yourcustom units, for example, PH.

    Proceed as follows to set up a custom unit:

    1. When measuring or sourcing the function of yourchoice, press the Scalesoftkey, then selectCustom Units from the list.

    2. Enter the 0% and 100% scale points for the inputof your transfer function.

    3. Press the Custom Unitssoftkey.

    4. Enter the 0% and 100% scale points for theoutput of your transfer function.

    5. Enter the name of the custom units (up to fourcharacters), for example PH(for pH), using the

    alphanumeric entry window, then press.

    While Custom Unitsare active, thesymbol showson the display to the right of the custom unit. Onceyou have programmed the custom measurementunit, the custom unit is available for calibrationprocedures in split-screen MEASURE/SOURCE

    mode. To cancel Custom Units, press theCustom Unitssoftkey again.

    744Users Manual

  • 8/12/2019 744 manual

    53/135

    44

    Using the 700-IV Current Shunt

    To source and measure current simultaneously, youneed to use a current shunt and use the volts

    measure function. The Fluke 700-IV current shunt is

    designed specifically for use with the 700 SeriesDocumenting Process Calibrators. To measurecurrent with the current shunt, proceed as follows:

    1. Connect the current shunt to the MEAS V jacks(rightmost).

    2. Connect the current signal to be measured to thecurrent shunt.

    3. Select the dc voltage measure function.

    4. Press the Scalesoftkey.

    5. Select Current Shuntfrom the list.

    6. The calibrator is automatically configured usingcustom units appropriate for the current shunt.

    Damping Your Measurements

    The calibrator normally applies a software filter todampen measurements in all functions except

    continuity. The specifications assume that damping is

    turned on. The damping method is a running averageof the last several measurements. Flukerecommends that you leave damping on. Turningdamping off may be useful when measurementresponse is more important than accuracy or noisereduction. If you want to turn off damping, press the

    More Choicessoftkey twice, then press the Dampensoftkey so that Off appears. Press Dampenagain to

    turn damping back on. The default state is On.

    Documenting Process CalibratorUsing Measure Mode

  • 8/12/2019 744 manual

    54/135

    45

    Note

    If a measurement falls outside a random noisewindow, a new average is started. If damping isturned off, or until measurements are fully

    damped, thesymbol is displayed.

    744Users Manual

  • 8/12/2019 744 manual

    55/135

    46

    Using Source Mode

    The operating mode (i.e., MEASURE, SOURCE) isshown in a reverse-video bar on the display. If the

    calibrator is not in SOURCE mode, pressuntilSOURCE is shown. You must be in SOURCE modeto change any of the SOURCE parameters.

    Sourcing Electrical Parameters

    To select an electrical sourcing function, proceed asfollows:

    1. Connect the test leads as shown in Figure 16,

    depending on the source function.

    2. Pressfor current,for dc voltage,for frequency, orfor resistance.

    3. Enter the desired output value, then press

    . For example, to source 5.0 V dc, press.

    Note

    If you are sourcing frequency, respond to thedisplay prompt to select a zero-symmetricsine or positive square wave. The amplitude

    you specify is p-p amplitude.

    4. To change the output value enter a new valueand press.

    Note

    If you are sourcing current, wait for thesymbol to go out before you use the output.

    5. To set the output value to 0 in the present sourcefunction, press.

    6. To turn off sourcing completely, presstwice.

    Note

    Use the source current function to drive acurrent loop. This is different than the looppower function in which the calibrator is

    powering a process instrument. To sourceloop power, use the Loop Powerfunctionaccessible from Setup mode.

    Documenting Process CalibratorUsing Source Mode

  • 8/12/2019 744 manual

    56/135

    47

    TC

    mA mA VVRTD

    RTD

    SOURCE 300VMAX30V

    MAX30VMAX

    30VMAX

    MEAS

    CAT

    TC

    mA mA VVRTD

    RTD

    SOURCE 300VMAX30V

    MAX30VMAX

    30VMAX

    MEAS

    CAT Red

    Black

    +

    +

    Red

    Black

    Common

    Common

    gj16f.eps

    Figure 16. Electrical Sourcing Connections

    744Users Manual

  • 8/12/2019 744 manual

    57/135

    48

    Simulating a 4 to 20 mA Transmitter

    You can configure the calibrator as a load on acurrent loop through the SOURCE mA function.

    When you press thekey in SOURCE mode, the

    display prompts you to select Source mAor SimulateTransmitter. When you Source mAthe calibrator issourcing current, and when you Simulate Transmitterthe calibrator is sourcing a variable resistance toregulate current to the specified value. Connect anexternal loop supply to the positive (top) mA jack asshown in Figure 17.

    Note

    Also see Transmitter Mode, in which the

    calibrator can be set up to temporarily takethe place of a two-wire process transmitter.

    Documenting Process CalibratorUsing Source Mode

  • 8/12/2019 744 manual

    58/135

    49

    TC

    mA mA VVRTD

    RTD

    SOURCE 300VMAX30V

    MAX30VMAX

    30VMAX

    MEAS

    CAT

    MEASSOURCE

    SETUPmA

    V V

    HzTC

    RTD

    7 8 9

    4 5 6

    1 2 3

    0 .

    Loop

    Power

    Supply

    +24

    UUT

    Black

    Red

    +

    ENTER

    CLEAR(ZERO)

    DOCUMENTING PROCESS CALIBRATOR744

    ot17c.eps

    Figure 17. Connections for Simulating a 4 to 20 mA Transmitter

    744Users Manual

    S C h lib i i i h h i

  • 8/12/2019 744 manual

    59/135

    50

    Supplying Loop Power

    The calibrator supplies loop power at 28 V or 24 V dcthrough an internal series resistance of 250 . The28 V setting supplies enough current for two or three

    4-20 mA devices on the loop in addition to the two-wire transmitter but uses more battery power. Usethe 24 V setting if there are two or fewer devices onthe loop in addition to the two-wire transmitter. (Eachdevice on a typical 4- to 20-mA loop has a resistanceof 250, thus dropping 5 V at 20 mA. A typicaltransmitter must have 11 V minimum in order to

    operate correctly at its top end.)

    When loop power is enabled, the mA (middlecolumn) jacks are dedicated to sourcing andmeasuring the current loop. This means that theSOURCE mA, measure RTD, and measure functions are not available (see Table 8, later in thismanual.)

    Connect the calibrator in series with the instrumentcurrent loop as Figure 18 shows. Proceed as follows

    to source loop power:

    1. Pressfor Setup mode.

    2. Note that following Loop Power, Disabledishighlighted. Press.

    3. Use theorarrow keys to select Enabled 24Vor Enabled 28 V.

    4. Press.

    5. Press the Donesoftkey.

    Documenting Process CalibratorUsing Source Mode

  • 8/12/2019 744 manual

    60/135

    51

    TC

    mA mA VVRTD

    RTD

    SOURCE 300VMAX30V

    MAX30VMAX

    30VMAX

    MEAS

    CAT

    MEASSOURCE

    SETUPmA

    V V

    HzTC

    RTD

    7 8 9

    4 5 6

    1 2 3

    0 .

    Red

    TEST DC PWR

    + +

    +

    Black

    ENTER

    CLEAR(ZERO)

    DOCUMENTING PROCESS CALIBRATOR744

    ot18c.eps

    Figure 18. Connections for Supplying Loop Power

    744Users Manual

    S i P Caution

  • 8/12/2019 744 manual

    61/135

    52

    Sourcing Pressure

    The calibrator provides a source pressure displayfunction that requires the use of an external pressure

    hand pump. Use this function to calibrate instruments

    that require a pressure source or differential pressuremeasurement. See Figures 19 and 34 for informationabout that application.

    Many ranges and types of pressure modules areavailable from Fluke. See Accessories near theback of this manual. Before you use a pressure

    module, read its Instruction Sheet. The modules varyin how you use them, how you zero them, what types

    of process pressure media are allowed, andaccuracy specification.

    To use the source pressure display, see Figure 19and proceed as follows:

    Warning

    To avoid a violent release of pressure ina pressurized system, shut off the valve

    and slowly bleed off the pressure beforeyou attach the pressure module to thepressure line.

    Caution

    To avoid mechanically damaging the

    pressure module, never apply more than10 ft.-lb. of torque between the pressure

    module fittings or between the fittingsand the body of the module. Alwaysapply appropriate torque between thepressure module fitting and connectingfittings or adapters.

    To avoid damaging the pressure modulefrom overpressure, never apply pressureabove the rated maximum printed on thepressure module.

    To avoid damaging the pressure modulefrom corrosion, use it only with specifiedmaterials. Refer to the printing on thepressure module or the pressure module

    instruction sheet for the acceptablematerial compatibility.

    Documenting Process CalibratorUsing Source Mode

    1 Connect a pressure module and pressure source base units (Pa mHg etc ) Change the pressure

  • 8/12/2019 744 manual

    62/135

    53

    1. Connect a pressure module and pressure sourceto the calibrator as Figure 19 shows. The threadson the pressure modules accept NPT fittings.Use the supplied NPT to ISO adapter if

    necessary.

    2. If necessary, pressfor SOURCE mode.

    3. Press. The calibrator automatically senseswhich pressure module is attached and sets itsrange accordingly.

    4. Zero the pressure module as described in themodules Instruction Sheet. Modules vary in

    zeroing procedures depending on module type.You MUST perform this step before you executea task that sources or measures pressure.

    5. Pressurize the pressure line with the pressuresource to the desired level as shown on the

    display.

    6. If desired, you can change pressure display unitsto psi, mHg, inHg, mH

    2O, inH

    2O, inH

    2O@60F,

    ftH2O, bar, g/cm2, or Pa. Metric units (kPa,mmHg, etc.) are shown in Setup mode in their

    base units (Pa, mHg, etc.). Change the pressuredisplay units as follows:

    a. Press.

    b. Press Next Pagetwice.c. Presswith the cursor on Pressure

    Units.

    d. Select the pressure units with theorkeys.

    e. Press.

    f. Press theDone

    softkey.

  • 8/12/2019 744 manual

    63/135

    Documenting Process CalibratorUsing Source Mode

    Simulating Thermocouples 5 Press the or key followed by to

  • 8/12/2019 744 manual

    64/135

    55

    Simulating Thermocouples

    Note

    Refer to Measuring Temperature earlier in themanual for a table of data relating tothermocouple types supported by the calibrator.

    Connect the calibrator TC input/output to theinstrument under test with thermocouple wire and the

    appropriate thermocouple mini-connector (polarizedthermocouple plug with flat, in-line blades spaced 7.9mm [0.312 in] center to center). One pin is wider thanthe other. Do not try to force a miniplug in the wrongpolarization. Figure 20 shows this connection.Proceed as follows to simulate a thermocouple:

    1. Attach the thermocouple leads to the appropriate

    TC miniplug, then to the TC input/output asFigure 13 shows.

    2. If necessary, pressfor SOURCE mode.

    3. Pressfor the display that prompts you to

    enter thermocouple type.4. Press theorkey followed byto

    select the desired thermocouple type.

    5. Press theorkey followed bytoselect Linear T(default), or Linear mV, (forcalibrating a temperature transmitter that

    responds linearly to millivolt inputs).

    6. Enter the temperature you want to simulate asprompted by the display and press.

    Note

    If you use copper wire instead ofthermocouple wire, the reference junction isno longer inside the calibrator. The referencejunction is moved to the instrument(transmitter, indicator, controller, etc.) input

    terminals. You must measure this externalreference temperature accurately and enterit into the calibrator. Do this by pressingand setting Ref. Junc. Compensat and Ref.Junc. Temp.After you enter the externalreference temperature, the calibratorcorrects all voltages to compensate for thisnew reference junction temperature.

    744Users Manual

    Simulating RTDs Proceed as follows to simulate an RTD (Resistance-

  • 8/12/2019 744 manual

    65/135

    56

    Simulating RTDs

    Note

    Refer to Table 6 for information about RTD(Resistance-Temperature Detector) typessupported by the calibrator.

    Connect the calibrator to the instrument under test asshown in Figure 21. The figure shows connections

    for two, three, or four-wire transmitters. For three orfour-wire transmitters, use the 4-inch long stackablejumper cables to connect the third and fourth wires atthe source V RTD jacks.

    Proceed as follows to simulate an RTD (ResistanceTemperature Detector):

    1. If necessary, pressfor SOURCE mode.

    2. Press

    until the select RTD type display isshowing.

    3. Press theorkeys followed bytoselect the desired RTD type.

    4. Enter the temperature you want to simulate as

    prompted by the display, then press.

  • 8/12/2019 744 manual

    66/135

    744Users Manual

  • 8/12/2019 744 manual

    67/135

    58

    1

    2

    4TC

    mA mA VVRTD

    RTD

    SOURCE 300VMAX30V

    MAX30VMAX

    30VMAX

    MEAS

    CAT

    TC

    mA mA VVRTD

    RTD

    SOURCE 300VMAX30V

    MAX30VMAX

    30VMAX

    MEAS

    CAT -PS

    +PS

    M

    -IN

    +IN

    S

    Z

    3

    ot21f.eps

    Figure 21. Connections for Simulating an RTD

    Documenting Process CalibratorUsing Source Mode

    Source Scale Percent of scale remains in effect until you change to

  • 8/12/2019 744 manual

    68/135

    59

    This feature lets you scale the output in accordancewith the input requirements of a particular processinstruments response. Percent of scale works for

    linear-responding transmitters, or square-rootresponding transmitters.

    Linear-Responding Transmitters

    1. If necessary, pressfor SOURCE mode.

    2. Select a source function (,,,,, or) as previously described and entera value.

    3. Press the Scalesoftkey.

    4. Select %from the list.

    5. Use the numeric keypad to enter the 0% of scalevalue (0% Value).

    6. Press.

    7. Use the numeric keypad to enter the 100% ofscale value(100% Value).

    8. Press the Donesoftkey.

    another source function or press the Scalesoftkeyand select another scale mode.

    Square-Law Process Variables

    When you select within scaling, the calibratoroutput value is the percent value entered, squared,and converted to engineering units.

    1. If necessary, pressfor SOURCE mode.

    2. Select a source function (,,,,, or) as previously described.

    3. Press the Scalesoftkey.

    4. Select from the list.

    5. Use the numeric keypad to enter the 0% of scalevalue (0% Value).

    6. Press.

    7. Use the numeric keypad to enter the 100% of

    scale value(100% Value).

    8. Press.

    744

    Users Manual

    9. Press the Donesoftkey. 1. Refer to the appropriate Using Source Mode

  • 8/12/2019 744 manual

    69/135

    60

    Square root percent of scale remains in effect untilyou change to another source function or press the

    Scalesoftkey and select another scale mode.

    Stepping and Ramping the Output Value

    Two features are available for adjusting the value ofsource functions, except pressure, which requiresthat you use an external pressure source:

    Stepping the output manually with theandkeys, or in automatic mode.

    Ramping the output with optional continuity or Vtrip detect.

    Using Manual Step

    The manual Stepfeature allows you to select a stepsize in engineering units (mV, V, mA, C, etc.) or %

    of scale. Stepping the output in % of scale is usefulfor quickly jumping between 0% and 100% (set stepsize = 100%) or 0-50-100% (set step size = 50%).Stepping works in SOURCE and inMEASURE/SOURCE modes. Proceed as follows toselect a step size:

    subheading earlier in this manual (e.g.,Sourcing Electrical Parameters) and connectthe calibrator to the circuit to be tested.

    2. If necessary, pressfor SOURCE mode.

    3. Set the calibrator for the desired source value.

    4. If you want to step the source value in % ofscale, set the % of scale value as describedpreviously under Sourcing in Percent of Scale.

    5. Press the Step softkey.

    6. Use the numeric keypad to enter the step size inthe units shown on the display.

    7. Press the Donesoftkey.

    8. Now you can adjust the output in steps by

    pressing theandkeys.

    Documenting Process CalibratorUsing Source Mode

    Using Auto Step 7. Select values for the following parameters as

  • 8/12/2019 744 manual

    70/135

    61

    g p

    To have the calibrator make a sequence of stepsautomatically, either once through the sequence orrepetitively, proceed as follows:

    1. Refer to the appropriate Using Source Modesubheading earlier in this manual (e.g.,Sourcing Electrical Parameters) and connectthe calibrator to the circuit to be tested.

    2. If necessary, pressfor SOURCE mode.

    3. Set the calibrator for the desired source value.

    4. If you want to step the source value in % ofscale, set the % of scale value as describedpreviously under Sourcing in Percent of Scale.

    5. Press theStepsoftkey.

    6. Press the Auto Stepsoftkey.

    you are prompted by the display:

    Starting point (in units or % of scale) Ending point Number of steps Time per step Single shot or continuous repetition Ramp pattern if continuous Start delay

    8. To start automatic stepping, press the Start Stepsoftkey. The softkey label changes to Stop Step.

    9. To stop automatic stepping, press the Stop Step

    softkey.

    10. Press the Donesoftkey to resume normaloperation.

    744

    Users Manual

    Ramping the Output Proceed as follows to ramp (i.e., sweep the source):

  • 8/12/2019 744 manual

    71/135

    62

    Ramping sweeps the source up or down in value.Use the ramp feature to check a limit switch or alarm,or any time you want a smoothly increasing or

    decreasing output function. You can set thecalibrator to ramp up or down in engineering units(mV, V, mA, C, etc.) or % of scale.

    During ramping, the output is adjusted 4 times persecond. The size of the steps is determined by yourchoices of endpoints and ramp time. For example, ifyou set the calibrator to ramp from 1 mV to 1 V over

    10 seconds, the output is adjusted in approximately

    25 mV steps.

    Ramping continues until the selected limit is reached,or until an optional trip condition is encountered. Theoptional trip detect works as follows: during ramping,the calibrator checks for either a 1 V change in dcvoltage or a change in continuity status (Openor

    Short) from one second interval to the next.

    1. Refer to the appropriate heading earlier in thismanual (e.g., Sourcing Electrical Parameters)and connect the calibrator to the circuit to be

    tested. Figure 22 shows an example.

    2. To automatically stop ramping if a trip conditionis detected, connect a voltage trip circuit to the VMEAS jacks or a continuity trip circuit to the mARTD MEAS jacks. (Continuity detection is notavailable when sourcing current.)

    3. If necessary, pressfor SOURCE mode.

    4. Set the calibrator for the desired source value aspreviously described.

    5. If you want to ramp the output in % of scale, set% of scale as described previously underSourcing in Percent of Scale.

    6. Press the More Choicessoftkey.

    Documenting Process CalibratorUsing Source Mode

    7. Press the Rampsoftkey. The display changes toth f ll i

    10. Press the Donesoftkey. Note the RAMPi t t t SOURCE t th t f th

  • 8/12/2019 744 manual

    72/135

    63

    the following:

    gj41s.eps

    8. Fill in the parameters as prompted. Enter theStart Value, End Value, and Ramp Time.

    9. To automatically stop ramping if a trip conditionis detected, set the Trip Detectto Enabled, andselect Voltageor Continuityas the trip function.

    annunciator next to SOURCEat the top of thedisplay.

    11. Select a low-to-high ramp or a high-to-low ramp

    with the Ramp Up/Downsoftkey.

    12. To start ramping, press the Start Rampsoftkey.

    13. Ramping continues until a trip is detected (ifenabled), the ramp time expires, or you pressthe Stop Rampsoftkey.

    744Users Manual

  • 8/12/2019 744 manual

    73/135

    64

    TC

    mA mA VVRTD

    RTD

    SOURCE 300VMAX30V

    MAX30VMAX

    30VMAX

    MEAS

    CAT

    MEASSOURCE

    SETUPmA

    V V

    HzTC

    RTD

    7 8 9

    4 5 6

    1 2 3

    0 .

    Input

    4 20 mA

    Trip Relay

    Output

    +

    +

    Source mA

    Voltage Trip Detect

    Red

    Black

    Red

    Black

    ENTER

    CLEAR(ZERO)

    DOCUMENTING PROCESS CALIBRATOR744

    ot22c.eps

    Figure 22. Checking a Relay Output Trip Alarm

    Documenting Process CalibratorSimultaneous Measure/Source

    Simultaneous Measure/Source Table 7 shows the functions you can usesimultaneously when Loop Power is disabled Table

  • 8/12/2019 744 manual

    74/135

    65

    Use the MEASURE/SOURCE mode to calibrate oremulate a process instrument. Pressso that asplit screen display appears as shown below.

    gj42s.eps

    simultaneously when Loop Power is disabled. Table8 shows the functions you can use simultaneously

    when Loop Power is enabled.

    You can use the Stepor Auto Stepfeatures to adjustthe output in MEASURE/SOURCE mode, or you canuse the calibration routine provided when you pressthe As Foundsoftkey.

    The following two softkeys displayed inMEASURE/SOURCE mode are for use in calibrating

    a process instrument.

    As Found, which lets you set up a calibrationroutine to obtain and record as found data.

    Auto Step, which lets you set up the calibrator forauto-stepping, as previously described.

    Instructions for calibrating a process instrumentfollow.

    744Users Manual

    Table 7. Simultaneous MEASURE/SOURCE Functions with Loop Power Disabled

  • 8/12/2019 744 manual

    75/135

    66

    Measure Function Source Function

    dc V mA Freq TC RTD Pressure

    dc V mA

    ac V

    Frequency (20 Hz)

    Low Frequency (

  • 8/12/2019 744 manual

    76/135

    67

    Measure Function Source Function

    dc V mA Freq TC RTD Pressure

    dc V mA

    ac V

    Frequency (20 Hz)

    TC

    Pressure

    744Users Manual

    Calibrating a Process Instrument In this case the calibrator is simulating the output of athermocouple and measuring the resulting current

  • 8/12/2019 744 manual

    77/135

    68

    Note

    To calibrate a HART-capable transmitter

    using the built-in HART interface, you use adifferent procedure than the one that follows.

    See the HART Mode Users Guide forinstructions.

    When the calibrator is in simultaneousMEASURE/SOURCE mode, a built-in calibrationroutine is activated when you press the As Foundsoftkey. (As Found data are the test results showingthe condition of a transmitter before it is calibrated.)The calibrator runs preloaded tasks (procedures) thatare developed using a host computer and compatibleapplication software.

    Generating As Found Test Data

    The following example shows how to generate asfounddata for a thermocouple temperaturetransmitter. The way you set up the template for the

    procedure is similar for Delta-P and 1 Pt. and 2 Pt.Switch tests.

    thermocouple and measuring the resulting currentfrom the transmitter. Other transmitters use thissame method. Just go back to MEASUREMENT or

    SOURCE mode and change the operating

    parameters before you press As Found.

    1. Connect the test leads to the instrument undertest as shown in Figure 23. The connectionssimulate a thermocouple and measure thecorresponding output current.

    2. If necessary, pressfor MEASURE mode.

    3. Press.

    4. Pressfor SOURCE mode.

    5. Press.

    6. Use theandkeys to select thethermocouple type, then press.

    7. Enter a source value, for example 100 degrees,then.

    Documenting Process CalibratorCalibrating a Process Instrument

    8. Pressfor MEASURE/SOURCE mode. Thedisplay changes to:

  • 8/12/2019 744 manual

    78/135

    69

    p y g

    gj42s.eps

    9. Press the As Foundsoftkey, followed by theInstrumentsoftkey. The display changes to:

    gj44s.eps

    10. Enter values for 0%and 100%of 4.0 mA and

    20.0 mA, respectively. Set Toleranceto 0.5% of

    span. (Use other values if necessary for yourapplication.)

    744Users Manual

  • 8/12/2019 744 manual

    79/135

    70

    mA mA VVRTD

    RTD

    SOURCE 300VMAX30V

    MAX30VMAX

    30VMAX

    MEAS

    CAT

    MEAS

    SOURCE SETUPmA

    V V

    HzTC

    RTD

    7 8 9

    4 5 6

    1 2 3

    0 .

    Black

    Power

    Supply

    TEST DC PWR

    ++

    +

    +

    Original

    Circuit Wiring

    Red

    Color dependson type of TC

    ENTER

    CLEAR(ZERO)

    TC

    Miniplug

    DOCUMENTING PROCESS CALIBRATOR744

    ot23c.eps

    Figure 23. Calibrating a Thermocouple Temperature Transmitter

    Documenting Process CalibratorCalibrating a Process Instrument

    11. If the process instrument needs more time tosettle than the calibrators normal settling time

    in percent of scale. Our example uses five points(0%, 25%, 50%, 75%, and 100%), rising only.

  • 8/12/2019 744 manual

    80/135

    71

    (about 2 seconds) at each new stimulus level,enter that time in seconds for Delay.

    12. Use the arrow keys to move the cursor down toenter 0%and 100%values for SOURCEtemperature. Our example uses 100C and300C.

    13. If the instrument calibration procedure requiresyou to enter either the measurement value or

    source value by hand at each step, press theUser Value softkey, for user entered values.

    Custom Units lets you define your own units suchas PH. See Creating Custom MeasurementUnits, earlier in this manual for an example.

    When you use custom units, thesymbolappears next to the value on the display and in

    results.

    Press the Donesoftkey after you have

    programmed your custom unit.

    14. TheStrategyis the number of test points andwhich test points are performed rising and falling

    Rising is indicated by the up arrow on thedisplay. Change to another test strategy by

    pressingon this line. A list of strategies

    appears from which to choose. Select one, thenpress Done.

    15. When you finish selecting the calibrationparameters, the display appears as follows:

    gj45s.eps

    744Users Manual

    16. Press the Donesoftkey to accept the calibrationparameters. The display changes to:

    measurement to stop changing, the Auto Testworks correctly for instruments with built-in

  • 8/12/2019 744 manual

    81/135

    72

    gj46s.eps

    17. You now have the choice of starting anautomatic test or stepping through the test pointsmanually. Press the