® ® www.paclp.com 6200 Series Sample Analysis Data Sheet 6200 series ® ANTEK 6200 SERIES SAMPLE ANALYSIS DATA LINEARITY 2000 ppm 1000 ppm 500 ppm 250 ppm 125 ppm 3.91 ppm 7.81 ppm 31.25 ppm 62.50 ppm 0 20 40 60 80 HOURS 100 90 80 70 60 50 40 30 20 10 0 PPM Gasoline Sample: run as unknown for 79 hours, 15 minutes STABILITY OVERVIEW Average: 77.45 ppm Standard Deviation: 0.47 ppm RSD: 0.612% DETAILED CONCENTRATION DATA Highest Recorded Value: 78.90 ppm Lowest Recorded Value: 75.50 ppm First Hour Average: 78.07 ppm Last Hour Average: 77.96 ppm A dibutyl sulfide in toluene standard, run as an unknown, was used to calibrate the analyzer at 2000 ppm. Half dilutions were made of the standard, and eight of those dilutions were also run as unknowns: EXPECTED AVERAGE SD %RSD 2000.00 1993.57 14.81 0.74 1000.00 998.67 8.41 0.84 500.00 496.92 4.01 0.81 250.00 248.20 1.86 0.75 125.00 123.75 1.43 1.16 62.50 63.03 0.42 0.66 31.25 34.01 0.69 2.01 7.81 7.05 0.25 3.54 3.91 3.72 0.39 10.56 2000 1500 1000 500 0 0 500 1000 1500 2000 0.999 Correlation STABILITY STUDY: 79 HOURS Copyright 2012/1 PAC L.P. All rights reserved 00.00.088 HOURS INJECTIONS AVERAGE SD %RSD %SD FULL SCALE* 14:45 177 70.11 1.27 1.81 0.423 2:25 29 28.33 0.59 2.07 0.195 3:30 42 154.59 1.61 1.04 0.537 1:40 20 28.56 0.34 1.18 0.112 16:15 195 71.19 1.39 1.96 0.465 3:20 40 155.60 1.70 1.09 0.566 5:10 62 28.84 0.66 2.30 0.221 COMBINED RUNS AVERAGE SD %RSD %SD FULL SCALE* MID RANGE : 70.65 0.76 1.08 0.254 LOWER RANGE : 28.58 0.25 0.89 0.843 UPPER RANGE : 155.10 0.72 0.46 0.239 * Full scale: 300 ppm REPEATABILITY

6200 Series Sample Analysis Data Sheet 6200 series · 2016-05-23 · sulfur and nitrogen/sulfur compounds. The Antek line also includes process analyzers for monitoring total nitrogen

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: 6200 Series Sample Analysis Data Sheet 6200 series · 2016-05-23 · sulfur and nitrogen/sulfur compounds. The Antek line also includes process analyzers for monitoring total nitrogen

® ®

www.paclp.com

6200 Series Sample Analysis Data Sheet

6200 series®

ANTEK 6200 SERIES SAMPLE ANALYSIS DATA

LINEARITY

2000 ppm

1000 ppm

500 ppm

250 ppm 125 ppm

3.91 ppm7.81 ppm

31.25 ppm62.50 ppm

0 20 40 60 80 H O U R S

100

90

80

70

60

50

40

30

20

10

0

PP

M

Gasoline Sample: run as unknown for 79 hours, 15 minutesSTAbILITY ovERvIEw Average: 77.45 ppm Standard Deviation: 0.47 ppm RSD: 0.612%

DETAILED CoNCENTRATIoN DATA Highest Recorded Value: 78.90 ppm Lowest Recorded Value: 75.50 ppm First Hour Average: 78.07 ppm Last Hour Average: 77.96 ppm

A dibutyl sulfide in toluene standard, run as an unknown, was used to calibrate the analyzer at 2000 ppm. Half dilutions were made of the standard, and eight of those dilutions were also run as unknowns: ExPECTED AvERAGE SD %RSD 2000.00 1993.57 14.81 0.74 1000.00 998.67 8.41 0.84 500.00 496.92 4.01 0.81 250.00 248.20 1.86 0.75 125.00 123.75 1.43 1.16 62.50 63.03 0.42 0.66 31.25 34.01 0.69 2.01 7.81 7.05 0.25 3.54 3.91 3.72 0.39 10.56

2000

1500

1000

500

00 500 1000 1500 2000

0.999 Correlation

STAbILITY STuDY: 79 houRS

Copyright 2012/1 PAC L.P. All rights reserved 00.00.088

houRS INjECTIoNS AvERAGE SD %RSD %SD fuLL SCALE* 14:45 177 70.11 1.27 1.81 0.423 2:25 29 28.33 0.59 2.07 0.195 3:30 42 154.59 1.61 1.04 0.537 1:40 20 28.56 0.34 1.18 0.112 16:15 195 71.19 1.39 1.96 0.465 3:20 40 155.60 1.70 1.09 0.566 5:10 62 28.84 0.66 2.30 0.221 CoMbINED RuNS AvERAGE SD %RSD %SD fuLL SCALE* mid range : 70.65 0.76 1.08 0.254 lower range : 28.58 0.25 0.89 0.843 upper range : 155.10 0.72 0.46 0.239

* Full scale: 300 ppm

REPEATAbILITY

Page 2: 6200 Series Sample Analysis Data Sheet 6200 series · 2016-05-23 · sulfur and nitrogen/sulfur compounds. The Antek line also includes process analyzers for monitoring total nitrogen

www.paclp.com ® ®

solidpartnersproven solutions

U.S.A.PAC, LP | 8824 Fallbrook Drive | Houston, Texas 77064T: +1 800.444.TEST | O: +1 281.940.1803 | F: +1 [email protected] | [email protected]

FRANCEBP 70285 | Verson | 14653 CARPIQUET CedexT: +33 231 264 300 | F: +33 321 266 [email protected] | [email protected]

GERMANYBadstrasse 3-5 | P.O.Box 1241 | D-97912 Lauda-Königshofen, T: +49 9343 6400 | F: +49 9343 640 [email protected]| [email protected]

SINGAPORE61 Science Park Road | #03-09/10 The GalenSingapore Science Park III | Singapore 117525T: +65 6412 0890 | F: +65 6412 [email protected] | [email protected]

NETHERLANDSP.O.Box 10.054 | 3004 AB RotterdamInnsbruckweg 35 | 3047 AG RotterdamT: +31 10 462 4811 | F: +31 10 462 [email protected] | [email protected]

RUSSIAShabolovka Street | 34, Bldg. 2 | 115419 MoscowT: +7 495 617 10 86 | F: +7 495 913 97 [email protected] | [email protected]

CHINARoom 1003, Sunjoy Mansion | No. 6 RiTan Rd.Chao Yang District | Beijing 100020 T: +86 10 650 72236 | F: +86 10 650 [email protected] | [email protected]

INDIA1036 Regus | Trade Center, Level 1Bandra (E) - 400 051 | MumbaiT: +91 22 40 700 447 / 700 | F: +91 22 40 700 [email protected] | [email protected]

MIDDLE EASTA1 Quds Street, A1 Tawar road | LIU#H13 Dubai Airport FreezoneNear Dubai Airport (terminal 2) | P.O.Box #54781 | Dubai, UAE T: +971 04 2947 995 | F: +971 04 2395 [email protected] | [email protected]

SOUTH KOREA#621 World Vision Building | 24-2, Youido-dongSeoul 150-010 T: +82 2785 3900 | F: +82 2785 [email protected] | [email protected]

THAILAND26th Floor, M. Thai Tower | All Seasons Place87 Wireless Road | Lumpini, Phatumwan | Bangkok 10330T: +66 2627 9410 | F: +662627 [email protected] | [email protected]

PAC Authorized Representatives are also located in most countries worldwide. For more information visit www.paclp.com

Copyright 2012/1 PAC L.P. All rights reserved - addition to 00.00.086

ANTEK bY PACLeading Elemental Analysis and Chromatography Since 1967, Antek has been a global leader in lab and on-line elemental analysis instrumentation and custom chromatography instruments that detect sulfur and nitrogen/sulfur compounds. The Antek line also includes process analyzers for monitoring total nitrogen and sulfur in liquids, gases and LPGs and determination of nitrogen, sulfur, total hydrocarbon, and aromatics for the quality assurance for CO2.

• Routine maintenance agreements available

ThE ANTEK ADvANTAGE• Experienced

We are experts at sulfur and nitrogen analysis technology with over 40 years experience in analytical instrumentation.

• Recognized We comply with the most widely accepted methods for sulfur and nitrogen determination—methods of choice by regulatory agencies and method writing institutions worldwide. ASTM D5453, D6667 total sulfur: gasoline, diesel, oil, gas ASTM D4629 total nitrogen: gasoline, diesel, oils Approved and recognized by CARB and US EPA

• Responsive Our experienced staff of engineers and chemists will help you solve your analytical challenge and fit the 6200 Series to your unique application.

• Cost-effective Our fast, reliable technologies further benefit your process with low cost of ownership. We eliminate need for expensive chemicals, column switching, reagents, and tape.

6200 SERIESADDITIoNAL fEATuRES & bENEfITS

• Advanced communications with remote diagnostics and remote operation capabilities

• Fast recovery time when changing between high and low concentrations: 1000 ppm to 50 ppm in 10–15 minutes (2 to 3 injections) compared to several hours

• The Ultraviolet Fluorescence detection system with its inherent linearity gives the 6200 a large dynamic range

• Touch screen control (man/machine interface)

• Easy to maintain; only 1 moving part• Each application reviewed by experienced,

knowledgable technical staff• Worldwide representative/distributor network• Comprehensive in-house or on-site training

available