Authors Weijun Yao, Chuanhong Tu Agilent Technologies Co., Ltd. (Shanghai) 412 YingLun Road Waigaoqiao Free Trade Zone Shanghai, 200131 P. R. C. Abstract The Agilent 6820 gas chromatograph/nitrogen phosphorus detector (GC/NPD) was employed to determine carba- mate pesticides, following the China National Standard Method GB/T 5009.104-2003. Seven common carbamates were fully separated using an HP-5ms column, 30 m × 0.32 mm × 0.25 μm. Each compound showed good linearity in the 10–1000-ppb range, with a detection limit lower than 10 ppb, which is 10–1000 times lower than the maximum allowable residue level. Introduction The nitrogen phosphorus detector (NPD) is both sensitive and selective for organic compounds con- taining nitrogen and phosphorus, and is widely used in the trace analysis of organonitrogen or organophosphorus pesticides. The carbamates are types of organic synthetic pesticides, widely used to prevent plant diseases. The application of these pesticides may leave residues in agricultural prod- ucts that may pose potential human health risks via the food chain. The Chinese government pub- lishes maximum residue limits for the pesticides carbofuran, primicarb, and carbaryl in food. GB/T 5009.104-2003 (originally GB14877-1994) is a method used to determine six carbamate residues based on GC/NPD using a packed column [1]. In recent years, capillary-column methods have gradually replaced packed-column methods Analysis of Carbamate Pesticides Using Agilent 6820 Gas Chromatograph/Nitrogen Phosphorus Detector Application because of the excellent separation capability of the capillary column. In this note, the Agilent 6820 GC/NPD is used to determine seven common car- bamates including metolcarb, isoprocarb, baycarb, propoxur, carbofuran, primicarb, and carbaryl. Experimental All experiments were performed on an Agilent 6820 GC with NPD, deactivated splitless inlet liner (p/n 5183-4696), and Agilent advanced green septa (p/n 5183-4759). Agilent Cerity Networked Data System (NDS) is used for instrument control and data acquisition. Instrumental parameters are listed in Table 1. The standard pesticides were dissolved in acetone solution. Samples were injected manually using a 10-µL syringe (p/n 5182-3428). Environmental and Food Analysis Table 1. Instrumental Parameters Instrument Agilent 6820 GC Software Cerity NDS for chemical QA/QC Inlet Split/Splitless inlet Inlet temperature 250 ºC Injection mode Splitless Injection volume 1 μL Purge time 0.75 min Column HP-5ms, 30 m × 0.32 mm × 0.25 μm (p/n 19091S-413) Carrier gas Nitrogen, constant pressure of 5 psi, 1.0 mL/min (50 ºC) Oven temp 50 ºC (1 min) Ramp 1: 20 ºC/min to 100 ºC Ramp 2: 5 ºC/min to 150 ºC (5 min) Ramp 3: 10 ºC/min to 200 ºC (10 min) Detector NPD, 325 ºC, white bead (p/n G1534-60570) H2, 3 mL/min Air, 60 mL/min Detector gas Makeup, N2, 10 mL/min

5989-1334EN

  • Upload
    anhthoo

  • View
    216

  • Download
    1

Embed Size (px)

Citation preview

Page 1: 5989-1334EN

AuthorsWeijun Yao, Chuanhong Tu

Agilent Technologies Co., Ltd. (Shanghai)

412 YingLun Road

Waigaoqiao Free Trade Zone

Shanghai, 200131 P. R. C.

Abstract

The Agilent 6820 gas chromatograph/nitrogen phosphorusdetector (GC/NPD) was employed to determine carba-mate pesticides, following the China National StandardMethod GB/T 5009.104-2003. Seven common carbamateswere fully separated using an HP-5ms column, 30 m × 0.32 mm × 0.25 µm. Each compound showed good linearityin the 10–1000-ppb range, with a detection limit lowerthan 10 ppb, which is 10–1000 times lower than the maximum allowable residue level.

Introduction

The nitrogen phosphorus detector (NPD) is bothsensitive and selective for organic compounds con-taining nitrogen and phosphorus, and is widelyused in the trace analysis of organonitrogen ororganophosphorus pesticides. The carbamates aretypes of organic synthetic pesticides, widely usedto prevent plant diseases. The application of thesepesticides may leave residues in agricultural prod-ucts that may pose potential human health risksvia the food chain. The Chinese government pub-lishes maximum residue limits for the pesticides carbofuran, primicarb, and carbaryl in food. GB/T 5009.104-2003 (originally GB14877-1994) is amethod used to determine six carbamate residuesbased on GC/NPD using a packed column [1]. Inrecent years, capillary-column methods have gradually replaced packed-column methods

Analysis of Carbamate Pesticides UsingAgilent 6820 Gas Chromatograph/NitrogenPhosphorus Detector

Application

because of the excellent separation capability ofthe capillary column. In this note, the Agilent 6820GC/NPD is used to determine seven common car-bamates including metolcarb, isoprocarb, baycarb,propoxur, carbofuran, primicarb, and carbaryl.

Experimental

All experiments were performed on an Agilent6820 GC with NPD, deactivated splitless inlet liner (p/n 5183-4696), and Agilent advanced green septa(p/n 5183-4759). Agilent Cerity Networked DataSystem (NDS) is used for instrument control anddata acquisition. Instrumental parameters arelisted in Table 1. The standard pesticides were dissolved in acetone solution. Samples wereinjected manually using a 10-µL syringe (p/n 5182-3428).

Environmental and Food Analysis

Table 1. Instrumental Parameters

Instrument Agilent 6820 GCSoftware Cerity NDS for chemical QA/QCInlet Split/Splitless inletInlet temperature 250 ºCInjection mode SplitlessInjection volume 1 µL Purge time 0.75 minColumn HP-5ms, 30 m × 0.32 mm × 0.25 µm

(p/n 19091S-413)Carrier gas Nitrogen, constant pressure of 5 psi,

1.0 mL/min (50 ºC)Oven temp 50 ºC (1 min)

Ramp 1: 20 ºC/min to 100 ºCRamp 2: 5 ºC/min to 150 ºC (5 min)Ramp 3: 10 ºC/min to 200 ºC (10 min)

Detector NPD, 325 ºC, white bead (p/n G1534-60570)H2, 3 mL/min Air, 60 mL/min

Detector gas Makeup, N2, 10 mL/min

Page 2: 5989-1334EN

2

Results and Discussions

All compounds were separated with a HP-5mscolumn. The resolution of baycarb and propoxur is1.0, although their structures are very similar. Thechromatogram of these pesticides is shown inFigure 1. The peak areas are linear with concentra-tions for all carbamates in the range of 10–1000 ppb.Typical calibration curves for carbofuran andprimicarb are shown in Figure 2.

min16 18 20 22 24 26

pA

20

30

40

50

60

70

80

90

Peak identification1. Metolcarb2. Isoprocarb3. Baycarb4. Propoxur

1

23 4 5

6

7

5. Carbofuran6. Primicarb7. Carbaryl

Figure 1. Chromatogram of 1-ppm carbamates using NPD detector.

Primicarby = 0.1277x + 0.5557R2

= 0.9992

Carbofurany = 0.0706x + 0.7961R2

= 0.9969

0

20

40

60

80

100

120

140

pA

0 200 400 600

Concentration, ppb

800 1000 1200

Figure 2. Calibration curves for primicarb and carbofuran.

Page 3: 5989-1334EN

3

The chromatogram for the 10-ppb carbamate standard is shown in Figure 3. The signal to noiseratios are between 6–12 for the 10-ppb carbamates.The limits of detection are 4–6 ppb; these are 10 to1000 times lower than the maximum allowableresidue limits stipulated in China National Standard(GB14928.2-94, GB14928.7-94, and GB14971-94),which is 50-ppb primicarb and 5000-ppb carbarylin cereals [2, 3, 4]. The NPD can easily meet theserequirements for the routine analysis of carbamatepesticide residues.

1

2

34

5

6

7

min16 18 20 22 24 26

pA

30

30.1

30.2

30.3

30.4

30.5

30.6

Figure 3. Chromatogram of 10-ppb carbamates standard with 1-µL injection.

Conclusion

The Agilent 6820 GC with NPD detector can beused for the sensitive and selective measurementof carbamate pesticide residues in food. The detec-tion limit is much lower than the maximumresidue limits. Good linearity was obtained in the range of 10–1000 ppb.

References 1. GB/T 5009.104-2003, Determination of carba-

mate pesticide residues in plant-originatedfoods.

2. GB14928.2-94 Maximun residue limits of primicarb in food.

3 GB14928.7-94 Maximun residue limits of carbofuran in rice grains.

4. GB14971-94 Maximun residue limits of carbarylin food.

Page 4: 5989-1334EN

Agilent shall not be liable for errors contained herein or for incidental or consequentialdamages in connection with the furnishing, performance, or use of this material.

Information, descriptions, and specifications in this publication are subject to changewithout notice.

© Agilent Technologies, Inc. 2004

Printed in the USAAugust 11, 20045989-1334EN

www.agilent.com/chem

For More Information

For more information on our products and services,visit our Web site at www.agilent.com/chem.