7/29/2019 554161 e http://slidepdf.com/reader/full/554161-e 1/2  06/05-W97-Sel Instruction sheet 554 161 Discharge tube, with luminescent screen (554 161) 1 Luminescent screen 2 4-mm hole 3 Electrode 4 Female ground joint 1 Description  At reduced pressure, the characteristic luminescent phenomena of electrical discharge and the cathode and canal rays outside the discharge path can be observed and investigated in the discharge tube. Safety note 2 Technical data Material: Glass Vacuum connection: Female ground joint NS 19/38 Electrical connections: 4-mm holes Required voltage: 3 ... 5 kV Dimensions: 70 cm × 4 cm  Weight: 0.5 kg 3 Accessories 1 Rotary-vane vacuum pump D 2.5 E 378 752 1 Small flange DN 10 KF with male ground joint NS 19/39 378 023 1 Cross 378 015 4 Centering rings for DN 16 KF 378 045 4 Clamping rings 378 050 1 Variable leak valve 378 776 1 Ball valve with 2 flanges DN 16 KF 378 777 1 Vacuum meter THERMOVAC TM 21 378 500 1 Gauge tube TR 211 378 501 1 Gauge head cable, 3 m 378 502 1 High-vacuum grease 378 701 1 High voltage power supply, 10 kV 521 70 2 High voltage cable, 1m 501 05 or 1 Safety connection lead, 1m, red 500 641 1 Safety connection lead, 1m, blue 500 642 2 Adapter for safety connection lead 500 17 008 When the discharge tube is operated at high voltages over 5 kV, X-rays are generated. Do not operate the discharge tube with high voltages over 5 kV. There is danger of implosion when a damaged discharge tube is evacuated: Handle the discharge tube carefully and protect it from mechanical stresses. Before evacuating the tube, examine it for damage such as cracks.

554161 e

Embed Size (px)

Citation preview

Page 1: 554161 e

7/29/2019 554161 e

http://slidepdf.com/reader/full/554161-e 1/2

 

06/05-W97-Sel

Instruction sheet 554 161

Discharge tube, with luminescent screen (554 161)

1 Luminescent screen

2 4-mm hole

3 Electrode

4 Female ground joint

1 Description

 At reduced pressure, the characteristic luminescent

phenomena of electrical discharge and the cathode and canalrays outside the discharge path can be observed andinvestigated in the discharge tube.

Safety note

2 Technical data

Material: Glass

Vacuum connection: Female ground joint NS 19/38Electrical connections: 4-mm holes

Required voltage: 3 ... 5 kV 

Dimensions: 70 cm × 4 cm ∅ 

Weight: 0.5 kg

3 Accessories

1 Rotary-vane vacuum pump D 2.5 E 378 752

1 Small flange DN 10 KFwith male ground joint NS 19/39 378 023

1 Cross 378 0154 Centering rings for DN 16 KF 378 0454 Clamping rings 378 0501 Variable leak valve 378 7761 Ball valve with 2 flanges DN 16 KF 378 777

1 Vacuum meter THERMOVAC TM 21 378 5001 Gauge tube TR 211 378 5011 Gauge head cable, 3 m 378 502

1 High-vacuum grease 378 701

1 High voltage power supply, 10 kV 521 70

2 High voltage cable, 1m 501 05

or 

1 Safety connection lead, 1m, red 500 641

1 Safety connection lead, 1m, blue 500 6422 Adapter for safety connection lead 500 17 008

When the discharge tube is operated at high voltages over 5 kV, X-rays are generated.

• Do not operate the discharge tube with high voltagesover 5 kV.

There is danger of implosion when a damaged dischargetube is evacuated:

• Handle the discharge tube carefully and protect it frommechanical stresses.

• Before evacuating the tube, examine it for damage suchas cracks.

Page 2: 554161 e

7/29/2019 554161 e

http://slidepdf.com/reader/full/554161-e 2/2

Instruction sheet 554 161 Page 2/2

LD Didactic GmbH . Leyboldstrasse 1 . D-50354 Huerth / Germany . Phone (02233) 604-0 . Fax (02233) 604-222 . e-mail: [email protected]

by LD Didactic GmbH Printed in the Federal Republic of GermanyTechnical alterations reserved

4 Operation

- Firmly clamp the ball valve with two flanges (a) onto theintake connector of the rotary-vane pump and connect thecross over it.

- Install variable leak valve (b) and gauge tube (c) onto theside of the cross.

- Evenly spread a thin coating of high-vacuum grease on thefemale ground joint of the discharge tube; slide the

discharge tube straight into the male ground joint (d)without excessive force and connect cross to male ground joint.

- Carefully close clamping ring.

- Connect the gauge tube to the vacuum meter THERMOVACTM 21.

Remark: For optimum observation of the luminescent 

 phenomena, carry out the experiment in a completely darkened 

room.

- Connect the electrode of the discharge tube to the right kVoutput ( I < 2 mA) of the high voltage power supply.

- Switch on the high voltage power supply and set the outputvoltage U = 5 kV.

- Switch on the vacuum meter THERMOVAC TM 21.

- Completely open the variable leak valve, partially open theball valve with two flanges and turn on the rotary-valvevacuum pump.

- Gradually and carefully close the variable leak valve, so thatthe pressure in the discharge tube continuously drops andobserve the discharge phenomena during pressuredecrease.

- Hold the pressure constant for a suitable observation periodby carefully operating the variable leak valve (while the

pump is operating) whenever a characteristic luminosity isvisible.

-  At approximately 0.04 hPa, use a permanent magnet todeflect the visible cathode and canal rays onto the screen.