Page 1: 3D Sensor Studies at New Mexico

Sally Seidel 1

3D Sensor Studies at New Mexico

Sally Seidel for Martin Hoeferkamp,

Igor Gorelov, Elena Vataga, and Jessica Metcalfe

University of New Mexico

Page 2: 3D Sensor Studies at New Mexico

Sally Seidel 2

Introduction• We have characterized 3D sensors of pitch 200 µm × 100 µm.• We report probe station based studies of depletion voltage, leakage

current, electrode capacitance, capture time, and signal rise time, supported by simulations

• The devices: non-irradiated and irradiated (1014, 2×1014, 1015 cm-2 55-MeV-p), unannealed, from Sherwood Parker.

Page 3: 3D Sensor Studies at New Mexico

Sally Seidel 3

Equipment• Picoprobe Model 35 (26 GHz bandwidth, 14 pS rise time, 0.05 pF capacitance)

• Picoprobe Model 12 (500 MHz bandwidth, 0.8 nS rise time, 0.1 pF capacitance)

• Kentech APG1 Pulser (300 pS pulse width)

• Tektronix 7254B Oscilloscope (2.5GHz bandwidth)

• 1064 nm, 960 nm, 820 nm IR lasers, 12GHz Photoreceiver

• Cascade REL-6100 semiautomatic probestation

• Micromanipulator HC-1000 Thermal Chuck (-60C)

• Peltier Thermal Chuck (-20C)

• Eichhorn+Hausmann MX203 wafer thickness and flatness gauge

Page 4: 3D Sensor Studies at New Mexico

Sally Seidel 4

3D Sensor Configuration

• Top view layout

• Layout dimensions:200 m x 100 m spacing,17 m electrode diameter,121 m electrode length.

• Configuration of the devices: <100> p-type silicon.– Alternating columns of n- and p-electrodes– Most electrodes are connected together along each column– Some electrodes are left isolated, to be contacted and measured individually

Page 5: 3D Sensor Studies at New Mexico

Sally Seidel 5

Electrode Leakage Current • Measured leakage current versus fluence:

• Prior to irradiation, the n-electrodes are shorted together by a surface electron layer.

Page 6: 3D Sensor Studies at New Mexico

Sally Seidel 6

Electrode Depletion Voltage

• Pixel cell depletion voltage measured via pulse height:

• Pixel cell depletion voltage measured via LCR meter:

Page 7: 3D Sensor Studies at New Mexico

Sally Seidel 7

Array Depletion Voltage • To test the entire device, we completely flood the 3D sensor with a uniform

1064 nm laser spot and scan the bias voltage above full depletion.

• Photo with IR filter of laser illuminating the sensor:

Page 8: 3D Sensor Studies at New Mexico

Sally Seidel 8

Array Depletion Voltage • Array depletion measured from signal efficiency (pulse height relative to the maximum for the non-irradiated device) versus bias:

• Result: very low values of depletion voltage for the entire sensor array,• Vdepletion ~ 15V for non-irradiated sensor• Vdepletion ~ 60V for sensor irradiated to 2x1014

• Vdepletion ~ 130V for sensor irradiated to 1x1015

Page 9: 3D Sensor Studies at New Mexico

Sally Seidel 9

Electrode Capacitance • Electrode capacitance using standard (HP4284A) LCR meter techniques

• Electrode capacitance versus fluence, type, and frequency:

Fluence (cm-2) Electrode Frequency (kHz)(55-MeV-p) type 10 100 10000 p 71 58 460 n 59 38 322×1014 p 96 69 532×1014 n 72 72 621×1015 p 98 70 551×1015 n 91 80 60

Page 10: 3D Sensor Studies at New Mexico

Sally Seidel 10

Electrode Capacitance

• Electrode capacitance versus temperature and frequency:

Page 11: 3D Sensor Studies at New Mexico

Sally Seidel 11

Electrode Capacitance • Direct measurement is checked by indirect measurement through signal decay time :

PICOPROBE 35

R=1.25M

C=.05pF

To Oscilloscope

Pulsed1064 nm and 960 nm Laser+Vbias

Gnd

• Indirect measurement using decay time of IR pulse on an isolated electrode.

• Electrode is grounded through input impedance of a Picoprobe 35.

• The IR laser induced charge is collected.

• When the laser is turned off the signal decay follows an exponential with a time constant = R*(C+C3D) , referred to here as RC time constant.

• C3D is extracted from the decay time constant using values of probe resistance and capacitance.

Page 12: 3D Sensor Studies at New Mexico

Sally Seidel 12

Electrode Capacitance • Performed at different bias voltages, using the procedure of Parker et al., Proc. IEEE Trans. Nucl. Sci., Oct 2001, p.

1635: Isolated electrode grounded through the 1.25 MΩ input impedance of the picoprobe. T = 0 when the laser is turned off. After the light emission ends and the charge is collected, the pulse height follows an exponential of time constant 177 ns. Averaging the values for 50 V to 100 V gives a p-electrode capacitance of 91.6 fF.

• Irradiated 2×1014 cm-2 55-MeV-p sensor p-electrode

Decay Curves, 3D sensor irrad 2x1014 55MeVp

0

0.05

0.1

0.15

0.2

0.25

0.E+00 1.E-07 2.E-07 3.E-07 4.E-07 5.E-07

Time (S)

Pu

lse

he

igh

t (A

U)

100V, tau=177nS

90V,tau=176.5nS

80V,tau=177nS

70V,tau=177nS

60V,tau=177.4nS

50V,tau=183.5nS

40V,tau=198nS

30V,tau=198nS

20V,tau=207nS

10V,tau=231nS

0V,tau=252nS

Capacitance, 3d Irrad 2x1014 cm-2 55MeVp

020406080

100120140160180

0 20 40 60 80 100

Bias Voltage (V)

Cap

acit

ance

(fF

)

Page 13: 3D Sensor Studies at New Mexico

Sally Seidel 13

Electrode Capacitance • A summary of the capacitance versus fluence for a p- and an n-electrode using the

direct capacitance measurement technique and for the p-electrode using the indirect measurement technique. The indirect measurement gives about a 50fF higher result.

Page 14: 3D Sensor Studies at New Mexico

Sally Seidel 14

Electrode Capacitance Calculation• 3D electrostatic calculation (IES Coulomb):

– p electrode length = 121 µm

– p electrode diameter = 17 µm nominal

– Center electrode to nearest neighbors

Prediction for p electrode = 28 fF

n n

n n

p p p

Capacitance at 17 m is 28 fF

We are systematically varying the geometrical parameters to understand the impact of each one on capacitance. An example for electrode diameter:

Page 15: 3D Sensor Studies at New Mexico

Sally Seidel 15

Position Scans • Scan the laser across one electrode cell to measure uniformity of signal collection

Y

X

Page 16: 3D Sensor Studies at New Mexico

Sally Seidel 16

Position Scans • Signal collection versus position:• Non-irradiated 3D sensor, p-electrode

X

Y

Page 17: 3D Sensor Studies at New Mexico

Sally Seidel 17

Charge Collection • Pulse the IR Laser as fast as possible and observe the rise time of the signal

PICOPROBE 35

R=1.25M

C=.05pF

Pulsed1064 nm IR Laser+Vbias

Gnd

• Measure the output rise time while reducing the laser pulse duration

Page 18: 3D Sensor Studies at New Mexico

Sally Seidel 18

Charge Collection • Input 0.3 nS laser duration:

• Output irradiated (1015) p electrode,~1.5 nS rise time

• Output non-irradiated p electrode, ~ 2.5 nS rise time

NOTE: The system isolation was improved, and a broken cable shield replaced, after this measurement was recorded. Revised graphs are in preparation.

Page 19: 3D Sensor Studies at New Mexico

Sally Seidel 19

Capture Time • For an irradiated 3D sensor, pulse the laser at a distance of 30 µm from

the electrode and measure the output. Repeat with laser pulse at a distance of 90 µm.

Page 20: 3D Sensor Studies at New Mexico

Sally Seidel 20

Capture Time

• The 60 µm difference in laser position results in a collection time difference of 50.6 nS – 47.4 nS = 3.2 nS

NOTE: The system isolation was improved, and a broken cable shield replaced, after this measurement was recorded. Revised graphs are in preparation.

Page 21: 3D Sensor Studies at New Mexico

Sally Seidel 21

Plans

Plans for 2007-2008:

– Repeat charge collection and capture time measurement with new low-noise system.

– Complete systematic simulation of full scope of geometrical options.

– Implement TCAD device simulation for improved capacitance and charge collection prediction.

– Systematics studies with 820 nm and 960 nm lasers.

– Irradiate ATLAS geometry devices at LANL and Sandia.

– Apply these measurement techniques to the ATLAS geometry devices.

There is a larger range of measurements we would like to do additionally if a

TurboDAQ system becomes available.

We are 5 people available for testbeam staffing as well.

Page 22: 3D Sensor Studies at New Mexico

Sally Seidel 22

Budget for FY 2008

$110,000 for electrical engineer, travel, and materials and supplies.