Upload
others
View
9
Download
0
Embed Size (px)
Citation preview
Chapter 3
Bioanalytical Method Development and Validation
21
3. Bioanalytical Method Development and Validation
3.1.Background
A sensitive, specific bioanalytical method is critical for a reliable pharmacokinetic
experiment. Chromatographic techniques, especially, high performance liquid
chromatography (HPLC) coupled with different detection systems is a preferred technique,
routinely employed in bioanalytical laboratories as compared to any other method of analysis
owing to their precision, accuracy, reliability and applicability to large-scale analysis. The
guidance document by USFDA systematically describes method development and validation
procedures required to demonstrate the reliability and reproducibility of the bioanalytical
method(CDER, 2001). According to the guidance document, the fundamental parameters of
validation include accuracy, precision, selectivity, sensitivity, reproducibility and stability.
The process of method validation involves laboratory investigations, documentation and
assessment of performance characteristics and their conformity to the acceptance criteria.
Though many validated methods have been reported, it is not necessary that all of them work
in a given laboratory. This occurs due to different instrument and assay conditions and other
environmental variables. Therefore, for a given laboratory, modification of the reported
method or development of a new method becomes essential. Most of the reported methods
are for human experiments and involve higher plasma volume. Our study was a non-clinical
one, which demanded a simple, economical, sensitive and specific bioanalytical method with
lesser plasma volume. Based on the physico-chemical properties of the drugs and the
available literature, we developed and validated two separate HPLC methods with UV
detection for the estimation of glibenclamide (GLI) and metformin (MET).
GLI is an oral hypoglycemic agent belonging to second-generation sulfonylureas. GLI lowers
the blood glucose by stimulating insulin release from pancreatic beta cells and may also
increase the insulin levels by reducing hepatic clearance of insulin. (Laurence, 2006). MET is
an anti-hyperglycemic drug belonging to the class of biguanides. It acts by reducing hepatic
glucose production and enhancing insulin action on muscle and fat tissue. MET brings down
fasting glucose and insulin levels, improves the lipid profile and causes weight loss in type-2
diabetic patients (Kirpichnikov et al., 2002, Laurence, 2006).
Chapter 3
Bioanalytical Method Development and Validation
22
3.2. Materials and Methods
3.2.1. Reference/Working Standards
Name of the Standard Supplier/Manufacturer
Glibenclamide Micro labs, India
Glimepiride Sigma Chemicals, USA
Metformin Marksans Pharma, India
Atenolol Sigma Chemicals, USA
3.2.2. Reagents, chemicals and Instruments
Chemical/Reagent Grade Manufacturer
Acetonitrile HPLC Labscan,Thailand
Dichloromethane HPLC Labscan,Thailand
Formic acid Suprapur Merck India Ltd, Mumbai
Methanol HPLC Labscan,Thailand
Methyl tertiary butyl ether HPLC Labscan,Thailand
n-Hexane AR S.D Fine Chemicals Ltd, Mumbai.
Potassium dihydrogen
orthophosphate
AR S.D Fine Chemicals Ltd, Mumbai.
Potassium hydroxide AR S.D Fine Chemicals Ltd, Mumbai.
Sodium Hydroxide AR S.D Fine Chemicals Ltd, Mumbai.
Water Milli-Q Millipore, USA
Instrument Brand/model Supplier/manufacturer
HPLC-System Prominence Shimadzu, Japan
Detector (PDA) SPD-M20A Shimadzu, Japan
Analytical column Betasil C18 Thermo Electron corporation
Lichrocart-CN Merck, Germany
Analytical balance Sartorius Sartorius Mechatronics, India
Centrifuge Heraeus Multifuge Thermo Electron corporation
Deep Freezer Carrier Carrier India
Water Purification System Milli-Q Millipore, USA
Multi-Pulse Vortexer Glass Col Terre Haute, USA
Micropipette Eppendorf Eppendorf, Germany
Nitrogen evaporator Turbovap LV Caliper Instruments,USA
Vortexer Spinix Spinix, India
Ultrasonic Cleaner UCB150 Spectral Lab, India
pH meter Orion 3 star Thermo Electron corporation
Hot air oven JRIC 7/A Osworld equipment, Mumbai
Chapter 3
Bioanalytical Method Development and Validation
23
3.2.3. Blank Plasma
The blank plasma for method development and validation was obtained from healthy animals
from Central Animal Research Facility, Manipal University, Manipal.
3.2.4. Method Development
Literature review was done to understand the physico-chemical properties of the analytes and
the validated methods reported so far. This was followed by method optimization procedures
involving effect of pH, stationary phase, solvent, mobile phase ratio, flow rate, internal
standard selection and optimization of extraction procedure.
3.2.5. Method development: Glibenclamide
3.2.5.1. Preparation of solutions and buffers
3.2.5.1.1. Mobile Phase [25mM Potassium dihydrogen orthophosphate buffer (pH 6.5)]
1701.125mg of Potassium dihydrogen orthophosphate was dissolved in 500 mL of water and
pH was adjusted to 6.50 with potassium hydroxide. The resulting solution was filtered
through 0.22µ millipore membrane filter using vacuum filtration system.
3.2.5.1.2. Rinsing and reconstitution solution [ methanol : water (50:50%,v/v)]
250 mL of methanol was added into a glass bottle containing 250 mL of water and the
resulting mixture was mixed well and sonicated for 10 minutes.
3.2.5.1.3. Diluent [ methanol : water (50:50% v/v)]
100 mL of methanol was transferred into a glass bottle containing 100mL of water, mixed
well and sonicated.
3.2.5.1.4. Extraction Buffer (0.20% v/v formic acid in water )
400µL formic acid was mixed with 200 mL water and sonicated
3.2.5.2. Preparation of Stock solutions
3.2.5.2.1. Glibenclamide main stock (1.000 mg/mL)
10.000 mg of Glibenclamide was taken in 10 mL volumetric flask and 5 mL methanol was
added and sonicated to dissolve. The volume was made up to the mark with methanol.
Chapter 3
Bioanalytical Method Development and Validation
24
3.2.5.2.2. Glibenclamide intermediate stock (200.000 µg/mL)
2 mL of Glibenclamide main stock was taken in a 10 mL volumetric flask and made up to the
volume by diluent.
3.2.5.2.3. Internal standard (Glimepiride) stock solution (200.000µg/mL)
2.000 mg of Glimepride was taken in 10 mL volumetric flask and 5 mL methanol was added
and sonicated to dissolve well. The volume was made up to the mark with methanol.
3.2.5.2.4. ISTD Working standard solution (40.000 µg/mL)
Working standard solution was prepared by diluting 2mL of stock solution to 10 mL
volumetric flask with diluents. Then it was vortexed for 2 minutes to mix well.
3.2.5.2.5. Preparation of Calibration standards and Quality Control Samples
Table 1.1: CC Standards and QCs preparation of GLI
Standards
Intermediate
stock taken
(mL)
Final volume
(mL)
Spiking
stock conc.
(ng/mL)
Plasma conc.
(ng/mL)
STD-1 0.010 10 200.000 10.000
STD-2 0.020 10 400.000 20.000
STD-3 0.060 10 1200.000 60.000
STD-4 0.120 10 2400.000 120.000
STD-5 0.500 10 10000.000 500.000
STD-6 0.800 10 16000.000 800.000
STD-7 1.000 10 20000.000 1000.000
STD-8 1.200 10 24000.000 1200.000
LLOQC 0.010 10 200.000 10.000
LQC 0.030 10 600.000 30.000
MQC 0.600 10 12000.000 600.000
HQC 1.100 10 22000.000 1100.000
3.2.6. Method development: Metformin
3.2.6.1. Preparation of solutions and buffers
3.2.6.1.1. 50mM Sodium hydroxide
Sodium hydroxide (200mg) was weighed accurately, transferred to a glass bottle containing
100 mL of Milli-Q water and sonicated.
Chapter 3
Bioanalytical Method Development and Validation
25
3.2.6.1.2. Rinsing solution: [ 0.1% ammonia in { methanol : water (70:30%,v/v}]
1 mL of ammonia solution was transferred in to a glass bottle containing 999 mL of
methanol: water (70:30 v/v) and sonicated to mix well
3.2.6.1.3. Diluent [ methanol : water (40:60% v/v)]
40 mL of methanol was transferred into a glass bottle containing 60 mL of Milli-Q water,
mixed well and sonicated.
3.2.6.1.4. Reconstitution solution [methanol : water (50:50% v/v)]
100 mL of methanol was transferred into a glass bottle containing 100mL of Milli-Q water,
mixed well and sonicated.
3.2.6.1.5. Buffer for mobile phase (25mM KH2PO4, pH 7.00)
1701.125mg of Potassium dihydrogen orthophosphate was dissolved in 500 mL of Milli-Q
water and pH was adjusted to 7 with potassium hydroxide. The resulting solution was filtered
through 0.22µ millipore membrane filter using vacuum filtration system
3.2.6.2. Preparation of Stock solutions
3.2.6.2.1. Metformin main stock (2.000 mg/mL)
20.000 mg of metformin was taken in 10 mL volumetric flask and 5 mL methanol was added
and sonicated to dissolve. The volume was made up to the mark with methanol.
3.2.6.2.2. Metformin intermediate stock (1000.000 µg/mL)
5 mL of metformin main stock was taken in to 10 mL volumetric flask and diluted up to the
mark with diluent.
3.2.6.2.3. Internal standard (atenolol) main stock solution (1.000 mg/mL)
25.000 mg of atenolol was taken in 25 mL volumetric flask and 10 mL methanol was added
and sonicated to dissolve well. The volume was made upto mark with methanol.
3.2.6.2.4. IS Working standard solution (100.000µg/mL)
Working standard solution was prepared by diluting 1 mL of main stock (IS) to 10 mL
volumetric flask with diluent. Then it was vortexed to mix well.
Chapter 3
Bioanalytical Method Development and Validation
26
3.2.6.2.5. Preparation of Calibration standards and Quality Control Samples
Table 1.2: CC Standards and QCs preparation of MET
Standards
Intermediate
stock taken
(mL)
Final
volume
(mL)
Spiking
stock
conc.(ng/mL)
Plasma conc.
(ng/mL)
STD-1 0.012 10 1200.000 60.000
STD-2 0.024 10 2400.000 120.000
STD-3 0.090 10 9000.000 450.000
STD-4 0.180 10 18000.000 900.000
STD-5 0.360 10 36000.000 1800.000
STD-6 0.720 10 72000.000 3600.000
STD-7 1.080 10 108000.000 5400.000
STD-8 1.440 10 144000.000 7200.000
STD-9 1.800 10 180000.000 9000.000
LLOQC 0.012 10 1200.000 60.000
LQC 0.036 10 3600.000 180.000
MQC 0.900 10 90000.000 4500.000
HQC 1.620 10 162000.000 8100.000
3.2.7. Optimization of extraction method
Different extraction techniques like, protein precipitation, liquid-liquid extraction and solid
phase extraction were tried. Optimization was done to minimize interferences at the retention
times of analyte and ISTD. All extraction trials were done at the MQC concentration. Trials
were taken to achieve consistent recovery.
3.2.7.1. Protein Precipitation Method
Various protein precipitation agents like acetonitrile, methanol, and 5% perchloric acid were
attempted.
3.2.7.2. Solid Phase Extraction Method
Different SPE trials were taken by altering buffering and washing procedures in order to
optimize extraction efficiency.
3.2.7.2.1. Liquid-Liquid Extraction Method
Methyl tertiary-butyl ether, n-Hexane, and ethyl acetate and methyl tertiary butyl ether in
combination with dichloromethane, methyl tertiary butyl ether in combination with n-Hexane
at different concentrations were attempted.
Chapter 3
Bioanalytical Method Development and Validation
27
3.2.8. Method validation
Validation experiments were conducted as per the USFDA guidelines. The parameters
evaluated were, system suitability, specificity, sensitivity, carryover, linearity, precision and
accuracy, recovery, dilution integrity and stability experiments. Following acceptance criteria
was applied to qualify validation tests
Table 1.3: Acceptance criteria for bioanalytical method validation
Validation Test Acceptance Criteria
System Suitability Area ratio: %CV should be ≤ 5.0%
RT: %CV should be ≤ 2.0%
Carryover Check ≤ 20% of response of the mean extracted analyte at LLOQ.
≤5% of response of the mean extracted ISTD
Linearity r2 ≥ 0.9800
Sensitivity Mean % deviation from the nominal concentration should be within
± 20%
% CV of area ratio of LLOQ should be 20%.
Specificity/Selectivity No endogenous peak shall be present within 10% window of the
retention time of analyte and an internal standard
If any peak is present at the retention time of analyte, its response
should be ≤ 20% of response of an extracted Lower calibration
standard i.e. LLOQ standard
Accuracy and Precision The mean calculated concentration of the LLOQC must be within
80-120% of its mean % nominal concentration and for low, medium
and high quality control standards must be within 85-115% of their
nominal concentration
The %CV for LLOQC must be ≤ 20% and for low, medium and
high quality standards must be ≤ 15%
Recovery % CV across QC level must be ≤ 15%
Dilution integrity
% deviation from the nominal concentration should be within ± 15%
% CV of diluted samples must be ≤ 15%
Stock Solution Stability Mean % change must be within ± 10%
Plasma Stability The calculated concentration of the LQC and HQC must be within
85-115% of its nominal concentration
Mean % change must be within ± 15%
Chapter 3
Bioanalytical Method Development and Validation
28
3.3. Results and Discussion
3.3.1. Method Development
3.3.1.1. Optimized Chromatographic Conditions
Table 1.4: Final Chromatographic conditions
Conditions Glibenclamide Metformin
Chromatographic mode Reversed Phase Reversed Phase
Isocratic/gradient mode Isocratic Isocratic
Internal Standard Glimepiride Atenolol
Rinsing solution Methanol : Water
(50:50%, v/v)
Methanol : Water
(50:50%, v/v)
Tray temperature (°C) 5 5
Injection volume (µl) 30 50
Stationary Phase
Column
Length x i.d (mm)
Particle size (µ)
Thermo Betasil C18
150 x 4.6 mm i.d.,
5
Lichrocart-CN
250 x 4.0 mm i.d.
5
Mobile phase Acetonitrile: 25mM KH2PO4
at pH 6.50
(45:55%, v/v)
Acetonitrile: 25mM KH2PO4
at pH 7.00
(28:72 %v/v)
Column Temperature (°C) 40 40
Flow rate (ml/min) 0.6 0.7
Detection wavelength (nm) 228 236
Retention time (min)
Analyte
IS
8.20 (Glibenclamide)
9.20 (Glimepiride)
6.66 (Metformin)
7.69 (Atenolol)
Run time (min) 11 10
3.3.1.2. Optimized Extraction Procedure
Glibenclamide:
Various methods were tested for glibenclamide extraction. Protein precipitation and solid
phase extraction were ruled out because of poor recovery. Multiple trials were taken with the
organic solvents, Tertiary-butyl methyl ether (TBME), n-Hexane, ethyl acetate, TBME with
dichloromethane and TBME with n-Hexane. TBME+n-Hexane at different strengths were
Chapter 3
Bioanalytical Method Development and Validation
29
attempted. Both analyte and ISTD were extracted well by TBME: n-Hexane (90:10% v/v)
with no significant interferences.
0.100 mL of validation/study sample was taken in a microcentrifuge tube and 0.010 mL of IS
working solution was added except for blank, where 0.010 mL of diluent was added.
Thereafter, 0.100 mL of buffer (0.2% formic acid in water) was added and votexed to mix.
1.00 mL of extraction solvent (TBME:n-Hexane 90:10 v/v) was added and vortexed for five
minutes. The mixture was centrifuged at 3500 rpm for 05 minutes. The supernatant was
separated and dried in nitrogen evaporator at 40oC then reconstituted with 0.300 mL of
methanol:water (50:50 v/v), votexed to mix well and transferred in to auto sampler vial
containing inserts.
Metformin:
After taking trials with different methods of extraction, protein precipitation method was
optimized with 2% v/v acetic acid in acetonitrile as a precipitating agent for the extraction of
metformin from the plasma and following procedure was used.
0.100 mL of validation/study sample was taken in a micro centrifuge tube and 0.010 mL of IS
working solution was added except in blank plasma, where 0.010 mL of diluent was added in
place of IS working solution. 0.010 mL of 50 mM sodium hydroxide was added and mixed
on the vortexer. To that mixture, 0.75 mL of 2% acetic acid in acetonitrile was added and
centrifuged for 10 min at 10,000 r.p.m. at 4 oC. 0.200 mL of supernatant was transferred in to
auto sampler vial and 0.200 mL of reconstitution solution was added and injected in to
HPLC system.
3.3.1.3.Chromatography
Typical chromatograms obtained from blank sample, lower limit of quantification, upper
limit of quantification, low, medium, high quality control samples are represented in the
following figures. The retention times of glibenclamide and ISTD (glimepiride) were
approximately 8.2 and 9.2 respectively. The retention times of metformin and ISTD
(atenolol) were approximately 6.6 and 7.9 respectively. Representative chromatograms for
GLI are presented in Figure 1.1 to 1.8 and for MET in Figure 1.9 to 1.16
Chapter 3
Bioanalytical Method Development and Validation
30
Figure 1.1: Representative Chromatogram of blank solution (GLI study)
Figure 1.2: Representative Chromatogram of aqueous MQC equivalent (GLI study)
Figure 1.3: Representative Chromatogram of blank K2EDTA rat plasma (GLI study)
Chapter 3
Bioanalytical Method Development and Validation
31
Figure 1.4: Representative Chromatogram of LLOQ (GLI study)
Figure 1.5: Representative Chromatogram of ULOQ (GLI study)
Figure 1.6: Representative chromatogram of LQC (GLI study)
Chapter 3
Bioanalytical Method Development and Validation
32
Figure 1.7: Representative Chromatogram of MQC (GLI study)
Figure 1.8: Representative Chromatogram of HQC (GLI study)
Chapter 3
Bioanalytical Method Development and Validation
33
Figure 1.9: Representative Chromatogram of blank solution (MET study)
Figure 1.10: Representative Chromatogram of aqueous MQC equivalent (MET study)
Figure 1.11: Representative Chromatogram of blank K2EDTA rat plasma (MET study)
Chapter 3
Bioanalytical Method Development and Validation
34
Figure 1.12: Representative Chromatogram of LLOQ (MET study)
Figure 1.13: Representative Chromatogram of ULOQ (MET study)
Figure 1.14: Representative Chromatogram of LQC (MET study)
Chapter 3
Bioanalytical Method Development and Validation
35
Figure 1.15: Representative Chromatogram of MQC (MET study)
Figure 1.16: Representative Chromatogram of HQC (MET study)
Chapter 3
Bioanalytical Method Development and Validation
36
3.3.2. Bioanalytical Method validation
3.3.2.1.System suitability
System suitability test was performed to ensure that the complete testing system (including
instrument, reagents, column, analyte) is suitable for the intended application. System
suitability test was done separately before the start of every sequence. Five replicates of
system suitability solution were injected. Percent coefficient of variation (%CV) for the peak
area ratio and retention times of analyte & ISTD were calculated. The observed percentage
CV for peak response ratio was within the acceptance limits with a value of ≤0.74% for GLI
and ≤1.06% for MET. The observed RTs were within the acceptance limits with the values of
≤0.43% and ≤0.58% for GLI and MET respectively.
3.3.2.2.Calibration curve
The calibration curve was constructed based on reported Cmax of the analyte. Calibration
curve consisted of reconstitution solution, blank sample (matrix sample processed without
internal standard and analyte), zero sample (matrix sample processed with internal standard)
and non-zero samples (calibration standards). The calibration curves were linear over the
range of 10 to 1200 ng/mL for glibenclamide and 60 to 9000 ng/mL for metformin. The
coefficient of determination was ≥0.9960. The linearity parameters are presented in Table 1.5
and 1.10 for GLI and MET respectively. Typical calibration curves for GLI and MET are
presented in figure 1.17 and 1.18 respectively.
3.3.2.3.Specificity/Selectivity
Specificity is the ability of an analytical method to differentiate and quantify the analyte in
presence of other components in the sample. For selectivity, blank plasma of six different lots
were taken and analyzed. Selectivity was evaluated by comparing extracted blank plasma
response with extracted LLOQ. No significant interference from the blank plasma was
observed at the retention times of both analyte and internal standards in both glibenclamide
and metformin study.
Chapter 3
Bioanalytical Method Development and Validation
37
Figure 1.17: Representative calibration curve of GLI in the range of 10.000 to 1200 ng/mL
Figure 1.18: Representative calibration curve of MET in the range of 60.000 to 9000 ng/mL
3.3.2.4.Sensitivity
The lower limit of quantification is the lowest concentration that can be determined with
acceptable accuracy and precision (20% variation) using a given method. This was performed
by injecting six different aliquots of extracted LLOQ concentration. Percentage deviation
y = 0.0014x - 0.0204
R² = 0.9966
0
0.6
1.2
1.8
2.4
0 200 400 600 800 1000 1200 1400
Calibrat ion Curve
y = 0.0002x + 0.0264
R² = 0.9979
0
0.2
0.4
0.6
0.8
1
1.2
1.4
1.6
1.8
2
0 2000 4000 6000 8000 10000
Calibrat ion Curve
Chapter 3
Bioanalytical Method Development and Validation
38
from the nominal concentration and percentage CV were calculated. The developed methods
were found to be sensitive with % CV of 15.80, Mean % concentration of 96.68 for
glibenclamide and %CV of 13.16%, Mean % Concentration of 107.04 for metformin.
3.3.2.5.Carryover
Carry over test was performed to confirm the absence carryover of analyte from the previous
injection. The experiment was performed at ULOQ level. The samples were injected in the
following order: blank reconstitution solution; aqueous equivalent ULOQ sample; blank
reconstitution solution; aqueous equivalent ULOQ sample; blank reconstitution solution;
extracted blank matrix sample; extracted ULOQ sample; extracted blank matrix sample;
extracted ULOQ sample; extracted blank matrix sample, followed by three LLOQs. No
significant carryover was observed in both GLI and MET studies.
3.3.2.6.Accuracy and precision
The accuracy of a method is the closeness of mean values obtained under prescribed
conditions. The precision of an analytical method is the closeness of individual measures
when different aliquots of same solution are analysed. Within-batch accuracy and precision
were assessed by analyzing one calibration curve and 4 sets of QC samples (6 replicates each
of the LLOQC, LQC MQC and HQC) in different batches. Between batch accuracy and
precision evaluation were also assessed by analyzing 4 batches on different days. The
accuracy and precision for all the batches at LLQC and LQC, MQC and HQC levels were
calculated. Mean percentage nominal concentration and coefficient of variation for all the
batches individually (intra) and collectively (inter) were found to be within acceptance limits
for both GLI and MET studies.
For glibenclamide, within batch accuracy was 95.46 to 105.6% and precision 15.80 to
19.76% at LLOQC, accuracy of 88.82 to 109.19 % and precision of 2.07 to 13.16% across
LQC, MQC and HQC.
For glibenclamide, between-batch accuracy was 98.44% and precision of 17.03% at LLOQC,
accuracy of 97.23 to 103.78% and precision of 6.18 to 9.46% across LQC, MQC and HQC.
The data is presented in the Table 1.6
For metformin, within batch accuracy was 93.04 to 117.10 % and precision 9.37% to 16.99%
at LLOQC, accuracy of 89.35 to 109.19 % and precision of 2.07 to 13.16% across LQC,
MQC and HQC.
Chapter 3
Bioanalytical Method Development and Validation
39
For metformin, between-batch accuracy was 98.44% and precision of 17.03% at LLOQC,
accuracy of 97.23 to 113.08% and precision of 1.94 to 9.55% across LQC, MQC and HQC.
The data is presented in the Table 1.11
3.3.2.7.Recovery
Recovery of an analyte was determined at low, medium and high quality control samples.
Analyte was spiked to the reconstituted solvent of the above blank samples to obtain the post
spiked LQC, MQC and HQC samples. Post spiked and extracted quality control samples
were analyzed and percentage recovery at each level was calculated by comparing the
response area of low, medium and high quality control levels and an internal standard. Mean
percentage recovery across all QC levels for GLI and MET is 93.45% and 73.26%
respectively. The mean percentage recovery for glimepiride (ISTD in GLI study) and atenolol
(ISTD in MET study) were 93.52% and 76.20% respectively. The summary statistics is
presented in Table 1.7 and 1.12 for GLI and MET respectively.
3.3.2.8.Dilution integrity
Dilution integrity was done in order to handle high sample concentrations exceeding ULOQ
or the possibility of insufficient sample volume in the study samples. A test for sample
dilution with blank matrix was performed. This was carried out by spiking 2xULOQ spiking
stock solutions in biological matrices to 1/2 and 1/4 dilute samples. Six aliquots each of
diluted samples and calibration standards were processed, extracted and analyzed.
Percentage CV and accuracy were calculated. The mean percentage nominal concentrations
and %CV were found to be within acceptance limits.
3.3.2.9.Stability
3.3.2.9.1. Stock solution stability
Short term and long term solution stabilities for the main stock and spiking stock solutions at
room temperature and at 2-8°C respectively were evaluated at MQC level. Five injections
each of comparison (freshly prepared MQC equivalent concentration) and stability samples
were performed. Mean percentage change was calculated by comparing the area of stability
and comparison samples. Both, main stock and spiking stocks of GLI and MET studies were
found to be stable at 2-8 oC for 9 days (long term) and for 8 hours at room temperature. The
results are presented in Table 1.8 and 1.13 for GLI and MET respectively.
3.3.2.9.2. Stability of drug in plasma
Chapter 3
Bioanalytical Method Development and Validation
40
The stability of the analytes in plasma were assessed at different conditions anticipated
during the pharmacokinetic studies like bench top stability, freeze thaw stability, in-injector
stability and long term stability at LQC and HQC levels.
3.3.2.9.3. Bench top stability
Bench top stability was evaluated to confirm that analyte degradation does not occur during
time of exposure to room temperature, which was based on the expected duration of the
samples maintained at room temperature during the study. Samples were prepared at low and
high quality control levels and kept at bench top at room temperature for a minimum of four
hours (stability samples). Later, fresh calibration standards and quality control samples
(comparison samples) were prepared, extracted and analyzed with the stability samples.
Mean percentage change was calculated.
The plasma samples were found to be stable at room temperature for 07 hours with mean%
change of 0.98% at LQC level and 6.67% at HQC levels for glibenclamide and 3.73% at
LQC, 1.76% at HQC for metformin.
3.3.2.9.4. Freeze and thaw stability
From the stage of collection to analysis, the samples undergo multiple cycles of freeze and
thaw, at least two times. Therefore, it is necessary to verify the validation samples to multiple
freeze-thaw cycles. The samples were exposed to four freeze thaw cycles. Stability samples
were analyzed against fresh calibration standards and quality control samples (comparison
samples). Mean % change was calculated and verified against acceptance criteria.
The plasma samples were found to be stable after repeating three cycles of freezing at -20oC
and thawing to room temperature with mean% change of 3.11% at LQC level and 5.26% at
HQC levels for glibenclamide and 0.56% at LQC and 8.21% at HQC for metformin
3.3.2.9.5. In-injector stability
Stability of processed samples in the instrument over the anticipated run time was assessed.
Stability samples were analyzed with fresh calibration standards along with low and high
quality control samples (comparison samples) and mean percentage change was calculated.
The plasma samples were found to be stable 4oC for 30 hours with mean% change of 5.14%
at LQC level and 5.83% at HQC levels for glibenclamide and 1.48% at LQC and 4.54% at
HQC for metformin
Chapter 3
Bioanalytical Method Development and Validation
41
3.3.2.9.6. Long term stability
To confirm analyte stability in the matrix during storage period, long-term stability studies
were conducted. Stability samples were stored at -20°C for 70 days. Long-term stability was
assessed by analyzing fresh calibration standards, low and high quality control samples
(comparison samples) with stability samples. Mean percentage change was calculated.
The plasma samples were found to be stable at -20oC for 70 days with mean% change of
3.83% at LQC level and 1.62 % at HQC levels for glibenclamide and 1.55% at LQC and
1.01% at HQC for metformin.
The stability data is presented in Table 1.9 and 1.14 for GLI and MET respectively.
3.4. Conclusion
Sensitive, specific and precise HPLC assays for the determination of glibenclamide and
metformin in rat plasma were developed and validated. The developed methods are suitable
for pharmacokinetic drug interaction studies in rodents.
Chapter 3
Bioanalytical Method Development and Validation
42
Table 1.5: Summary of calibration curve parameters of glibenclamide
Linearity Slope Intercept r2
1 0.001429379 -0.019494894 0.9996
2 0.001428497 -0.026486986 0.9995
3 0.001430286 -0.020494674 0.9966
4 0.001429299 -0.017495085 0.9996
Table 1.6: Between batch accuracy and precision of glibenclamide
QC Levels LLOQC
(10ng/mL)
LQC MQC HQC
(30 ng/mL) (600ng/mL) (1100ng/mL)
Batch-1
8.005 33.343 631.561 933.422
8.502 32.569 672.4 931.014
7.909 27.036 707.629 1053.373
11.935 28.572 659.422 1008.384
12.006 34.326 612.934 947.001
9.234 26.948 646.838 989.017
Batch-2
8.005 28.885 595.038 962.811
8.652 34.192 572.4 1096.594
12.326 33.375 597.629 982.373
11.492 30.572 583.598 1038.384
10.874 29.095 612.934 1226.001
12.034 31.572 606.838 1209.017
Batch-3
8.652 31.45 545.038 1162.811
7.997 28.392 572.475 1096.594
8.212 29.343 647.629 1230.429
11.326 27.569 615.598 1218.384
9.084 34.036 630.934 1056.034
12.002 38.572 628.838 990.017
Batch-4
10.234 30.326 685.543 1098.116
8.095 33.772 672.4 1062.174
9.671 29.343 597.629 1090.429
7.874 28.569 612.257 1069.575
10.134 27.936 610.425 1126.001
12.002 32.092 626.816 1090.017
n 24 24 24 24
Mean 9.844 30.912 622.7 1069.499
% CV 17.03 9.462 6.197 8.648
Mean %
nominal conc. 98.44 103.04 103.78 97.23
Chapter 3
Bioanalytical Method Development and Validation
43
Table 1.7: Recovery across QC levels (GLI study)
QC Levels Mean % Recovery
Glibenclamide
LQC 94.04
MQC 92.06
HQC 94.27
Mean 93.45
SD 1.21
CV 1.30
ISTD Glimepiride
93.52
n=6 for all QCs; n=18 for ISTD
Table 1.8: Stock solution stability (GLI Study)
Stability
Level
Glibenclamide Glimepiride
Mean % Change Mean % Change
Long term
(9 days at 2-8oC) Main Stock -4.539 -1.032
Spiking
stock -2.8603 -2.98
Short term
(08 hours at room
temperature)
Spiking
stock -2.921 -3.258
Table 1.9: Glibenclamide plasma stability
Stability Level
Fresh Sample Conc.
(ng/mL)
Stability sample Conc.
(ng/mL) Mean%
Change Mean %CV Mean %CV
Bench topa LQC 30.851 6.84 30.550 4.16 -0.98
HQC 1169.624 6.42 1091.585 2.40 -6.67
In-Injectorb LQC 30.851 6.84 29.267 6.73 -5.14
HQC 1169.624 6.42 1101.452 1.69 -5.83
Freeze thawc LQC 30.851 6.84 29.890 5.32 -3.11
HQC 1169.624 6.42 1108.053 8.92 -5.26
Long termd LQC 30.251 5.26 29.092 4.95 -3.83
HQC 1124.826 10.17 1106.653 10.69 -1.62 aStability in plasma at room temperature for 7hours bStability in plasma at 4oC in auto sampler for 30 hours cStability in plasma after 3rd freeze (-20oC) - thaw cycle dStability in plasma for 70 days at -20oC
n=6; Specified concentrations: LQC: 30.000 ng/mL; HQC:1100.000 ng/mL
Chapter 3
Bioanalytical Method Development and Validation
44
Table 1.10: Summary of calibration curve parameters of metformin
Linearity Slope Intercept r2
1 0.00173465 0.0085769 0.9945
2 0.00024236 0.0019262 0.9979
3 0.00175078 0.0267986 0.9976
4 0.00171293 0.00936981 0.9938
Table 1.11: Between batch accuracy and precision of metformin
QC levels LLOQC LQC MQC HQC
(60 ng/mL) (180 ng/mL) (4500 ng/mL) (8100 ng/mL)
Batch-1
54.75 172.271 4420.618 7774.503
58.127 189.724 4342.535 7781.07
55.232 165.217 4437.871 8065.91
54.658 172.747 4268.01 7722.569
46.82 182.975 4246.815 8206.582
65.341 206.873 4425.854 9198.712
Batch-2
71.102 181.395 5111.759 9016.458
48.794 207.44 4967.067 8950.74
66.515 199.876 4847.924 8891.085
68.939 185.483 4764.899 8667.464
53.869 199.96 4867.679 8463.503
49.612 187.454 5063.789 8958.234
Batch-3
64.515 188.24 4167.647 7681.264
50.444 218.204 4550.722 7456.511
59.108 201.134 4687.722 7469.813
66.335 196.096 4372.766 7351.797
73.078 186.171 4545.001 7916.367
71.843 198.349 4851.92 7436.367
Batch-4
72.15 192.362 3962.1 7084.241
58.428 202.969 4129.928 7292.116
78.617 200.941 4034.504 7289.145
71.939 238.1 3895.433 7995.818
69.86 180.067 3979.694 8104.048
70.567 206.855 4123.938 8122.56
n 24 24 24 24
Mean 62.527 194.204 4461.091 8037.370
% CV 14.76 8.16 8.19 7.88
Mean %
nominal conc. 104.21 107.89 99.14 99.23
Chapter 3
Bioanalytical Method Development and Validation
45
Table 1.12: Recovery across QC levels (MET study)
QC Levels Mean % Recovery
Metformin
LQC 71.69
MQC 70.21
HQC 77.88
Mean 73.26
SD 4.069
CV 5.55
ISTD Atenolol
76.20
n=6 for all QCs; n=18 for ISTD
Table 1.13: Stock solution stability (MET Study)
Stability
Level
Metformin Atenolol
Mean % Change Mean % Change
Long term
(9 days at 2-8oC) Main Stock -0.86 -1.032
Spiking
stock 2.15 1.85
Short term
(08 hours at room
temperature)
Spiking
stock 1.15 -4.75
Table 1.14: Metformin plasma stability
Stability Level
Fresh Sample Conc
(ng/mL)
Stability sample Conc
(ng/mL) Mean%
Change Mean %CV Mean %CV
Bench topa LQC 178.300 3.20 171.655 2.54 -3.73
HQC 8503.279 2.03 8353.754 2.69 -1.76
In-Injectorb LQC 178.300 3.20 180.930 1.49 -1.48
HQC 8503.279 2.03 8117.427 1.89 -4.54
Freeze thawc LQC 178.300 3.20 179.299 2.71 0.56
HQC 8503.279 2.03 7804.766 2.85 -8.21
Long termd LQC 183.633 3.01 180.780 1.96 -1.55
HQC 7923.612 2.72 8003.311 1.95 1.01 aStability in plasma at room temperature for 7hours bStability in plasma at 4oC in auto sampler for 30 hours cStability in plasma after 3rd freeze (-20oC) - thaw cycle dStability in plasma for 70 days at -20oC
n=5; Specified concentrations: LQC:180.000 ng/mL; HQC:8100.000 ng/mL