2D Color Analyzer CA-2500 Series Instruction Manual Before using this instrument, please read this manual.

2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

Embed Size (px)

Citation preview

Page 1: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

2D Color Analyzer CA-2500 Series

Instruction ManualBefore using this instrument,please read this manual.

Page 2: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

Safety SymbolsThe following symbols are used in this manual to prevent accidents which may occur as a result of incorrect use of the instrument.

Denotes a sentence regarding safety warning or precaution.Read the sentence carefully to ensure safe and correct use.

Denotes a prohibited operation.The operation must never be performed.

Denotes an instruction.The instruction must be strictly adhered to.

Denotes an instruction.Disconnect AC adapter from AC outlet.

Denotes a prohibited operation.Never disassemble this instrument.

This symbol indicates A.C.

This symbol indicates D.C.

Trademarks● “Microsoft®”, “Windows®”, “Windows®XP”, “Windows® 7”, “Windows® 8”and “Excel®” are registered trademarksofMicrosoftCorporationintheU.S.A.andothercountries.

● “Intel®”and“Pentium®”areregisteredtrademarksofIntelCorporationintheU.S.A.andothercountries. Other company names and product names mentioned in this manual are the registered trademarks or

trademarks of their respective companies.

Notes on this manual● Nopartof thismanualmaybe reprintedor reproduced inany formorbyanymeanswithout thepermissionofKonicaMinoltaOptics,Inc.

● Thecontentsofthismanualaresubjecttochangewithoutnotice.● Everyefforthasbeenmadetoensuretheaccuracyofthecontentsofthismanual.However,shouldyouhaveanyquestionsorcomments,or findanerrorormissingsection,pleasecontactaKonicaMinolta-authorized service facility.

● KonicaMinoltaacceptsnoresponsibilityforconsequencesresultingfromfailuretofollowtheinstructionsoutlined in this manual, the condition above notwithstanding.

Page 3: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

1

Safety Warnings and CautionsToensurecorrectuseof this instrument,readthefollowingpointscarefullyandadheretothem.Afteryouhave read this manual, keep it in a safe place where it can be referred to anytime a question arises.

Warning(Failuretoadheretothefollowingpointsmayresultindeathorseriousinjury.)

Donotusethisinstrumentinplaceswhereflammableorcombustiblegases(gasolineetc.)arepresent. Doing so may cause fire.

AlwaysusetheACadapterandpowercordsuppliedasastandardaccessoryoroptional(AC-A312),andconnectittoindoorACoutletofratedvoltage100-240V 50/60Hz.FailuretofolloweitherofthesemayresultindamagetoinstrumentorACadapter,fireorelectricshock.

Insertthepowerplugfullyandsecurely.Ifthepowerplugisnotpushedcompleteinsertingmaycausefireorelectricshock.

Ifthisinstrumentisnotusedforalongtime,disconnectACadapterfromACoutlet.AccumulateddirtorwateronprongsofACadapterplugmaycausefire.Besuretoremovebeforeuse.

Do not forcibly pull any part on power cord when unplugging since this may damage power cord, resultinginfireorelectricshock.Gentlydisconnectbyholdingplug.Also,donothandlepowerplug with wet hands. Doing so may cause electric shock.Dust could enter inside this instrument from air duct. Do not leave such dust accumulated while usingforthismaycausefire.ContactwiththenearestKonicaMinolta-authorizedservicefacilityforperiodical checkup.Donotforciblybend,twistorpullpowercord.Also,donotplaceheavyobjectonpowercord,ordamageormodifyone.Anyofthesemaycausefireorelectricshockduetodamagetopowercord.

DonotdisassembleormodifythisinstrumentorACadapter.Doingsomaycausefireorelectricshock.

Donotexposethis instrumentto liquidormetalobjectwhichmaycausefireorelectricshock.Shouldeitherofthesehappen,switchpoweroffimmediately,unplugACadapter,andcontactthenearestKonicaMinolta-authorizedservicefacility.ShouldthisinstrumentorACadapterbedamagedorsmokeoroddsmellbegenerated,donotkeepusingonewithoutcorrection.Doingsomaycause fire. Insuchcases,switchpoweroffimmediately,unplugACadapter fromACoutlet,andthencontact thenearestKonicaMinolta-authorized service facility.

Caution

UsethisinstrumentnearACoutletforeasypluggingorunplugginginusingACadapter.

Do not place this instrument on unstable or sloping surface which may drop or overturn it. Droppingoroverturningmayinjuresomeonearound.Takecarenottodropthisinstrumentwhencarrying.

Whencleaning,disconnectthepowerplug.Connectingmaycauseelectricshock.

(Failuretoadheretofollowingpointsmayresultininjuryordamagetothisinstrumentorotherproperty.)

Page 4: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

2

IntroductionThankyouforpurchasingour2DColorAnalyzerCA-2500series.Thisinstrumentmeasuresunevennessofluminanceandcoloronvariousdisplayunitorprojector,backlightandcarinstrumentpanelin2dimension.Read this manual carefully before use."CA-2500"andCA-2500Sareappliedas representativeexamples for explanationand illustrationrespectively in this instruction manual. Only those with different specification are indicated model by model.

Packaging materialRetainaccompanyingpackagingmaterials(carton,protector,andplasticbag)forfutureuse.This isdelicatemeasurement instrument.Usepackagingmaterialssupplied inpurchasing incasethisinstrumentneeds tobe transferred forsuchpurposeasmaintenance inKonicaMinolta-authorizedservice facility. These packaging materials are useful for minimizing shock or vibration to this instrument insuchsituation.Shouldanyofthesepackagingmaterialsbelostorbroken,contactthenearestKonicaMinolta-authorizedservicefacility.

Note on useOperating environmentlDonotusethisinstrumentoutdoorsincestandardaccessoryACadapter(AC-A312)isdesignedforindooruse.lNeverdisassemblethisinstrumentforbeingcomposedofdelicateelectroniccomponents.lUsethisinstrumentatratedvoltageof100-240V .ConnectACpowercordtoACoutletwithratedvoltage100-240V 50/60Hz.Connectedvoltageshouldnotbeoutsidetherangeof±10%ofnominal.

l ThisinstrumentisclassifiedintoaPollutionDegree2asinstrumentusedmainlyinmanufacturingplant,laboratory,warehouseorequivalents.Usethisinstrumentinmetaldustfreeandnoncondensingpotentialenvironment.

lThis instrument iscategorized intoOvervoltageCategory Iasequipmentwhich ispoweredbyanACadaptor connected to commercially available power.

lIfprotectiveearthterminalisequippedwiththePCtocontrolthisinstrument,connectthePCterminaltoaprotectiveearthoutlet.Ifnot,electricshockmayoccurduetoearthleakage.

l Take care not to enter foreign substance like water or metal in this instrument. Operating in such state causes serious danger.

l Do not use this instrument under direct sunlight or near heater. The internal temperature of this instrument becomesmuchhigherthanambienttemperaturewhichmaybreakthisinstrument.Also,besuretokeepgoodventilationinusingthisinstrument.Nevercoverairductofthisinstrument.

lAvoidrapidchangeinambienttemperaturewhichmayformdewcondensation.lAvoidusingthisinstrumentinextremelydustyorhumidplace.lUsethis instrumentatambient temperaturebetween10ºCand30ºCandrelativehumidity70%or less

with no condensation. Operating this instrument outside specified temperature and humidity range may unsatisfy its original performance.

lDonotusethisinstrumentataltitudesabove2000meters(6560ft).lEven if theproduct isusedwithin thespecifiedoperating temperature/humidity range, thedisplayedvaluesmaybeaffectedby long-termconditionsofuse. Ifyouhaveanyquestionaboutspecification,pleasecontactthenearestKonicaMinolta-authorizedservicefacility.

This instrumentlDonotsubjectthisinstrumenttostrongimpactorvibration.lDonotforciblypull,bend,orapplystrongforcetopowercordforattachedACadapterorUSBcable.This

may result in snapping.lConnectthisunittopowersourcewithminimalnoise.lShouldbreakageorabnormalitybe foundduringoperation,switchpoweroff immediatelyandunplug.Then,referto"ErrorCheck"onpages29.

lShouldthisinstrumentbreakdown,donottrytodisassembleandrepairitbyyourself.ContactthenearestKonicaMinolta-authorizedservicefacility.

Page 5: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

3

Lens and filter inside lens mount on bodyl Be sure that there is no dirt on surface of lens and filter inside lens mount on body before measurement. Correctmeasurementresultcannotbeobtainedifdirt,dustorhandsoilisleftoruncleaned.

l Do not touch surface of lens and filter inside lens mount on body with hand.l Lens and filter inside lens mount on body could become fogged up if ambient temperature fluctuates under

high humidity. Be careful not to generate such condition for correct measurement cannot be performed.

StorageBodyl Coverlensmountwithstandardaccessorymountcoverinstoringlensinremovedstate.l Do not store this instrument under direct sunlight or near heater. The internal temperature of this

instrument could become much higher than ambient temperature which may break this instrument.l Storethis instrumentatambient temperaturebetween0ºCand30ºCandrelativehumidity70%or lesswithnocondensationorbetween30ºCand35ºCandrelativehumidity55%orlesswithnocondensation.Storageunderhightemperatureandhumiditymaydeteriorateperformanceofthisinstrument.Foraddedsafety, we recommend storage with a drying agent at room temperature.

l Eveniftheproductisstoredwithinthespecifiedstoragetemperature/humidityrange,thedisplayedvaluesmaybeaffectedbylong-termconditionsofstorage.Ifyouhaveanyquestionaboutspecification,pleasecontactthenearestKonicaMinolta-authorizedservicefacility.

l Takecarenottoformcondensation.Avoidrapidchangeinambienttemperaturewhentransferringbodyfor storage.

l Putbodyinpackagingboxsuppliedwhenpurchasedoroptionalsoftcase(CA-A60)tostoreinsafeplace.

Lensl Coverlenswithlenscap,orlenshoodwithhoodcapforstorage.lWhenyouremovelensfrombodyforstorage,coverlenswithlenscapandmountcap.

Temperature/Humidity conditions under which the instrument is used and storedIt is recommended that the instrumentbeusedandstoredunderstandardconditions (Temperature:23ºC;Relativehumidity:40%),andthatareassubjecttohightemperatureand/orhumiditybeavoided.Inadditioninordertomaintainthemeasurement accuracy of this instrument, it is recommenced that it be inspected regularly about once a year. For details onhavingtheinstrumentinspected,pleasecontactthenearestKonicaMinolta-authorizedservicefacility.

Even if theproduct isusedwithin thespecifiedoperating temperature/humidity rangeorstoredwithin thespecifiedstoragetemperature/humidityrange,thedisplayedvaluesmaybeaffectedbylong-termconditionsofuseorstorage.Iftheinstrumentisleftunderthefollowinghigh-temperatureconditionsforalongperiodoftime,thedisplayedvaluesmaychangeasfollows:

Temperature:30ºC;Relativehumidity:70%;Periodundertheseconditions:720hours(30days)Accuracy:Luminance:±0.4%;Chromaticity:±0.003Inter-pointerror:Luminance:±0.2%;Chromaticity:±0.0003

Temperature:35ºC;Relativehumidity:55%;Periodundertheseconditions:336hours(14days)Accuracy:Luminance:±1%;Chromaticity:±0.006Inter-pointerror:Luminance:±0.5%;Chromaticity:±0.001

Thesedifferences indisplayvaluesaredue to the instrumentmaterialsand/orcomponentsbeingaffectedby thetemperatureandhumidityconditionsof long-termuseorstorage. Inparticular,optical filtersareeasilyaffectedbytemperatureorhumidity.Althoughmeasureshavebeentakentoimproveresistancetotemperature/humiditychanges,theaccumulatedeffectoflong-termuseorstoragemayaffectthedisplayedvalues.

Page 6: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

4

CleaningBodyl Ifthisunitbecomesdirty,wipewithdryandsoftcloth.Donotuseorganicsolventlikebenzineorthinnerandotherchemicalagent forcleaning.Shouldnoneof thesemethodsbehelpful,contact thenearestKonicaMinolta-authorizedservicefacility.

Lens and filter inside lens mount on bodyl Shoulditbegottendirtordust,wipeoffwithdryandsoftclothorlenscleaningpaper.Donotuseorganicsolventlikebenzineorthinnerandotherchemicalagentforcleaning.Shouldnoneofthesemethodsbehelpful,contactthenearestKonicaMinolta-authorizedservicefacility.

Notes on transferl Usetheoriginalpackagingmaterialstominimizetheshockandvibrationtotheproductwhentransportingit.l Pack the product and all of its accessories together in one package before sending them to us for repair, inspectionand/orcalibration.

Maintenancel Periodical checkup is recommended annually to maintain measurement accuracy of instrument. For detailsoncheckup,contactthenearestKonicaMinolta-authorizedservicefacility.

Disposal Methodl MakesurethattheCA-2500,itsaccessoriesandthepackingmaterialsareeitherdisposedoforrecycled

correctly in accordance with local laws and regulations.

Page 7: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

5

CONTENTSSafety Warnings and Cautions …………………1

Introduction …………………………………………2Note on use ………………………………………… 2

Operating environment …………………………………… 2This instrument …………………………………………… 2Lens and filter inside lens mount on body ……………… 3

Storage ……………………………………………… 3Body ………………………………………………………… 3Lens ………………………………………………………… 3

Cleaning ……………………………………………… 4Body ………………………………………………………… 4Lens and filter inside lens mount on body ……………… 4

Notes on transfer …………………………………… 4Maintenance ………………………………………… 4Disposal Method …………………………………… 4

Standard Accessory ………………………………6

Optional Accessory ………………………………7

System Configuration ……………………………8

Names and Functions of Parts ……………… 10<Body> ……………………………………………………… 10<Lens> ……………………………………………………… 11

Installing

Goods/Environment to Prepare on Customer's Side … 14Controlling PC ………………………………………14Mounting stage ………………………………………14Blackout curtain (quickly prepared darkroom) / Dark room ……………………………………………14

Installing ………………………………………… 15

Placing Lens …………………………………… 16Replacing lens ………………………………………17

Connecting AC Adapter ……………………… 18Connection procedure ……………………………18

ON ( | ) / OFF ( ) of Power Switch …………… 19ToON ……………………………………………………… 19To OFF ……………………………………………………… 19

Connecting with PC …………………………… 19Operation procedure ………………………………19

Measurement

Starting Up and Ending System …………… 22Starting up ……………………………………………22Connectionrecognitionerror …………………………… 22

Ending …………………………………………………22

Setting the measurement distance and the pint ring range index ……………………………… 23

Setting measurement distance …………………23Caution ……………………………………………………… 24

Example of reference position to read distance …25Howtoreaddistanceincaseofstandardlensortelephotolens … 25Howtoreaddistanceincaseofwidelens ……………… 25

Others

Outer Dimensions ……………………………… 28

Error Check ……………………………………… 29

Main Specification ……………………………… 30

Page 8: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

6

Standard Accessory

Lens Capl Place on lens for protection in not using this

instrument.For standard or telephoto lensFor wide lens

Mount Capl Place on lens mount for protection when lens is

removed from body.For standard or telephoto lensFor wide lens

Lens Hood For standard or telephoto lens CA-A68

For wide lens CA-A69 For macro measurement CA-A70

Hood Capl Place on lens for protection in not using this

instrument. For standard lens, telephoto lens, or macro measurement CA-A71For wide lens CA-A75

Mount Cover CA-A61l Place for protection of lens mount of body when lens

is removed.

AC Adapter AC-A312l SuppliespowerfromACoutlet. Input: 100-240V 0.75A 50/60Hz Output:12V 3.0A

CA-A71 CA-A75

For wide lens

For standard ortelephoto lens

CA-A68 CA-A69 CA-A70

Attachedlenscap,mountcap,lenshoodandhoodcapcorrespondingwithlenstypeforeachmodel.

For standard or telephoto lens

For wide lens

Page 9: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

7

USB Cable (3 m) IF-A18 l Usedforcommunicationbetweenthis instrument

andPC.

Calibration Data DVDl Suppliedwitheachlens.

Tripod CS-A3

Data Management Software CA-S25wl FordatamanagementofthisinstrumentfromPC.

Mura Evaluation Software CA-Mural Thisprogramisanadd-onsoftwareforCA-S25w. Itusesthe"muraquantitativeevaluationmethod"

developedbySonytoprovidethe"mura indexes"and "mura index images" ref lect ing humansensitivity.

Soft Case CA-A60l Usedtokeep this instrumentandaccessoriesor

carrythemwithhand.Neverusefortransfer.

Optional Accessory

Pan Head CS-A4

Page 10: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

8

System ConfigurationCA-2500A (With all lenses)

Wide lens CA-A58 Hood cap(For wide)CA-A75

Lens hood(For wide)CA-A69

Standard lens CA-A57 Hood cap(For standard/telephoto, macro)CA-A71

Lens hood(For standard/telephoto)CA-A68

Telephoto lens CA-A59

Hood cap(For standard/telephoto, macro)CA-A71

Lens hood(For standard/telephoto)CA-A68

Lens hood(For macro)CA-A70

Mount cover

Soft caseCA-A60

TripodCS-A3

Pan headCS-A4

AC adapterAC-A312

USB cableIF-A18

CA-2500

Data management softwareCA-S25w

Mura Evaluation softwareCA-MuraPC

(on the market)

Highmagnificationmacro ringMacro 2

Lowmagnificationmacro ringMacro 1

CA-2500S (With standard lens)

Standard lens CA-A57 Hood cap(For standard/telephoto, macro)CA-A71

Lens hood(For standard/telephoto)CA-A68

Mount cover

Soft caseCA-A60

TripodCS-A3

Pan headCS-A4

AC adapterAC-A312

USB cableIF-A18

CA-2500

Data management softwareCA-S25w PC

(on the market)

Mura Evaluation softwareCA-Mura

CA-2500W (With wide lens)

Wide lens CA-A58 Hood cap (For wide)CA-A75 Lens hood

(For wide)CA-A69

Mount cover

Soft caseCA-A60

TripodCS-A3

Pan headCS-A4

AC adapterAC-A312

USB cableIF-A18

CA-2500

Data management softwareCA-S25w PC

(on the market)

Mura Evaluation softwareCA-Mura

CA-2500T (With telephoto lens)

Telephoto lens CA-A59

Hood cap(For standard/telephoto, macro)CA-A71

Lens hood(For standard/telephoto)CA-A68

Lens hood(For macro)CA-A70High

magnificationmacro ringMacro 2

Low magnificationmacro ringMacro 1

Mount cover

Soft caseCA-A60

TripodCS-A3

Pan headCS-A4

AC adapterAC-A312

USB cableIF-A18

CA-2500

Data management softwareCA-S25w PC

(on the market)

Mura Evaluation softwareCA-Mura

Page 11: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

9

CA-2500SW (With standard and wide lenses)Wide lens CA-A58 Hood cap (For wide)

CA-A75Lens hood(For wide)CA-A69

Standard lens CA-A57 Hood cap(For standard/telephoto, macro)CA-A71

Lens hood(For standard/telephoto)CA-A68

Mount cover

Soft caseCA-A60

TripodCS-A3

Pan headCS-A4

AC adapterAC-A312

USB cableIF-A18

CA-2500

Data management softwareCA-S25w PC

(on the market)

Mura Evaluation softwareCA-Mura

CA-2500ST (With standard and telephoto lenses)

Wide lens CA-A58 Hood cap (For wide)CA-A75 Lens hood

(For wide)CA-A69

Telephoto lens CA-A59

Hood cap(For standard/telephoto, macro)CA-A71

Lens hood(For standard/telephoto)CA-A68

Lens hood(For macro)CA-A70

Mount cover

Soft caseCA-A60

TripodCS-A3

Pan headCS-A4

AC adapterAC-A312

High magnificationmacro ringMacro 2

Low magnificationmacro ringMacro 1

USB cableIF-A18

CA-2500

Data management softwareCA-S25w PC

(on the market)

Mura Evaluation softwareCA-Mura

CA-2500WT (With wide and telephoto lenses)

Standard lens CA-A57 Hood cap(For standard/telephoto, macro)CA-A71

Lens hood(For standard/telephoto)CA-A68

Telephoto lens CA-A59

Hood cap(For standard/telephoto, macro)CA-A71

Lens hood(For standard/telephoto)CA-A68

Lens hood(For macro)CA-A70

Mount cover

Soft caseCA-A60

TripodCS-A3

Pan headCS-A4

AC adapterAC-A312

High magnificationmacro ringMacro 2

Low magnificationmacro ringMacro 1

USB cableIF-A18

CA-2500

Data management softwareCA-S25w PC

(on the market)

Mura Evaluation softwareCA-Mura

StandardaccessoryOptional accessory

Page 12: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

10

Names and Functions of Parts<Body>

Power switch Switchesthis instrumenton/off. [|]sideforON; [○]sideforOFF.GreenLEDlightsupwhileswitchedon. ………………………………………… (p.19)

AC adapter input terminalTowhichaccessoryACadapterisconnected. ………………………… (p.18)USB connector TowhichUSBcableisconnectedwhenusedwithPC. ……………… (p.19)Air duct Airduct toair-cool interior.Donotcoverhere.Air-cooling fanbuilt in this

instrument operates while switched on. Lens mount To install lens on.Lens attaching portion To which lens hood is attached.Screw hole for fixing Usedtosetthisinstrumentontripodorjig.

Lens attaching portion

USB connector

Lens mount

AC adapter input terminal

Screw hole for fixing (Tripod screw)

Air duct

Screw hole for fixing (ISO screw)

Handle

Power switch

Page 13: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

11

<Lens>

Body and lens arranged in a pair are calibrated. Pair those with identical numbers.CA-2500A, CA-2500S, CA-2500SW, CA-2500ST

CA-2500A, CA-2500W, CA-2500SW, CA-2500WT

CA-2500A, CA-2500T, CA-2500ST, CA-2500WT

Lens hood Protects lens. Remove this once when replacing lens.Focus ring Adjustsfocus.Aperture ring Beingfixed,thisshouldnotbeused.Donotoperatebyremovingorrevolving

screw for this would deteriorate the performance.Macro ring Placed between telephoto lens and body when utilizing telephoto lens for

macro.Youcannotplace2typesofmacroringsimultaneously.

Focus ring

Distance indicator

Lens hood

Lens

(Low magnification) (High magnification)

Macro ringTelephoto lens

Wide lens

Standard lens

Aperture ring

Page 14: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

12

Page 15: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

13

Installing

Page 16: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

14

Installing

Goods/Environment to Prepare on Customer's SideGetpreparedforbelowgoods/environmentbeforeusingthisinstrument.

Controlling PCOS : Windows®XPProfessionalSP3,Windows®XPProfessionalx64EditionSP2 Windows®7Professional32bit(x86),Windows®7Professional64bit(x64) Windows®8Pro32bit(x86),Windows®8Pro64bit(x64) (English,Japanese,orChinese(simplifiedfont)versionoftheaboveOS) ThePCmustmeetthesystemrequirementsspecifiedbytheapplicable

OSorlistedbelow,whicheverishigher.CPU : EquivalenttoPentium®4,2.8GHzorbetterMemory : 1024MBormoreHarddiskdrive : Atleast80MBofavailablediskspaceisrequiredonthesystemdrive

(drivewheretheOSisinstalled). To install the calibration data, the following disk space is required for

eachlens: •Forastandardlens:approximately540MB •Forawideanglelens:approximately470MB •Foratelephotolens:approximately1.3GB

To save measurement data on the hard disk, you need to reserve the followingadditionalspaceontheharddisk:

•Forexample,approximately110MBisrequiredtosave10piecesofmeasurementdatawith980×980resolution.

•About1MBisrequiredforconfigurationfilessuchasthespotsettingfile and measurement condition file.

Display : Displayhardwarecapableofdisplaying1280×768pixels /65,536colors(HighColor:16-bitcolor)orbetter

Other : CD-ROMdriveRequiredforinstallation1DVD-ROMdriveRequiredforinstallation of calibration data

* Inplaceof theabove twodrives,youcanuseasingledrive that iscapableofreadingbothCD-RandDVD-Rmedia.

AUSBportisrequiredforconnectionoftheinstrument.

Inaddition,followingitemsareuseful.

Mounting stageTable on which this instrument is installed. Below are recommended conditions.lCanchangemeasurementdistanceaccordingtomeasurementobject.l Painted with black to prevent the entrance of light diffusely reflecting.

Blackout curtain (quickly prepared darkroom) / Dark room Neededtoblockthisinstrumentandmeasurementobjectfromoutsidelight.Get prepared for blackout curtain for quickly prepared dark room or dark room.

Page 17: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

15

Installing

Installing

Tosetthisinstrumentontripodorjig,usescrewholeforfixingatthebottomofthisinstrument.2typeholesareavailableasfollows;Tripodscrewhole:Tosetontripod.Screwdepthis6.5mm.ISOscrewhole:Tosetonjig.UserISO5mmscrewandscrewdepthis6.5mm.

Forfurtherdetail,seep.28.

Warning for safety Donotusethisinstrumentinplaceswhereflammableorcombustiblegases(gasolineetc.)arepresent. Doing so may cause fire.

Caution for safety Do not place this instrument on unstable or sloping surface which may drop or overturn it. Dropping or overturning

mayinjuresomeonearound.Takecarenottodropthisinstrumentwhencarrying.

*Whenstandardlensandhoodareplaced

ISO screw hole

Optical axis

Tripod screw hole

UNC 1/4 Depth 6.5

95.2

186.2

68.2

125.2

Page 18: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

16

Installing

Placing LensBelowshows lens typeaccompanying respectiveCA-2500series.Lenshood isservedasstandardcorresponding to lens type.CA-2500A Standard,wideandtelephotolenses,and2macroringsCA-2500S StandardlensCA-2500W WidelensCA-2500T Telephotolensand2macroringsCA-2500SW StandardandwidelensesCA-2500ST Standardandtelephotolenses,and2macroringsCA-2500WT Wideandtelephotolenses,and2macrorings

Operation procedure

1. Remove mount cover from lens mount on body.

2. Place lens on lens mount before rotating clockwise.lBesuretoinsertlensstraightly.Ifforciblyput

aslant, lens mount could be broken. lRevolvelensuntilfixed.

For CA-2500A, CA-2500T, CA-2500ST, or CA-2500WTInusingamacroringforeitherofCA-2500A,CA-2500T,CA-2500ST,orCA-2500WT,sandwich it between lens and lens mount.

3. Align 2 pins of lens hood and 2 holes of body. Then, insert lens hood as if it covers lens. Finally, rotate screw portion of lens hood clockwise.lSelect appropriate lens hooddepending

on lens type. Inappropriate lenshoodmaycause the lens head to be positioned back from the lens hood, which leads to incorrect measurement.

lHoldthecircumferenceoflenshood.lRotatescrewportionoflenshooduntilfixed.

Mount cover

Lens

Macro ring

Lens hood

Lens hood pin hole

Page 19: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

17

Installingl Checkwhetherbodyandlenshaveidenticalserialnumbers. Serialnumbercanbefoundonlabelofbodyandlenssiderespectively. Topursuehigh-accuracymeasurement,calibrationwasdoneonmaker's

side for each unit before shipment combining with each accessory lens forCA-2500series.Thismeansthatattachedcalibrationcoefficientdatatoeachunitisvalidexclusivelyforthepackagedcombinationofunitandlens.

l SetlenstypetouseinsoftwareCA-S25w. (Fordetail,seetheinstructionmanualfordatamanagementsoftwareCA-S25w.) This enables the application of appropriate calibration coefficient data depending on usage condition

which would finally give correct measurement result.

Replacing lensFollow reverse procedure to above to replace lens. First, remove lens hood and then lens.

XXXXXX1

WARNING:WARNUNG:Niemals das gehause offnen.

Do not remove covers.警告:カバーを開けないでください。

12 4

Page 20: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

18

Installing

Connecting AC Adapter UseaccessoryACadapterforpowersource.

Warning (Failuretoadheretothefollowingpointsmayresultindeathorseriousinjury.) AlwaysusetheACadapterandpowercordsuppliedasastandardaccessoryoroptional(AC-A312),and

connect it to indoorACoutletofratedvoltage100-240V 50/60Hz.Failuretofolloweitherof thesemay

result in damage to unit, fire or electric shock.

Ifthisinstrumentisnotusedforalongtime,disconnectACadapterfromACoutlet.AccumulateddirtorwateronprongsofACadapterplugmaycausefireandshouldberemoved.Besuretoremovebeforeuse.

Do not forcibly pull any part on power cord when unplugging since this may cause fire or electric shock. Gentlydisconnectbyholdingplug.Also,donothandlepowerplugwithwethands.Doingsomaycauseelectric shock.

Donotforciblybend,twistorpullpowercord.Also,donotplaceheavyobjectonpowercord,ordamageormodifyone.Anyofthesemaycausefireorelectricshockduetodamagetopowercord.

DonotdisassembleormodifythisinstrumentorACadapter.Doingsomaycausefireorelectricshock.

ShouldthisinstrumentorACadapterbedamagedorsmokeoroddsmellbegenerated,donotkeepusingonewithoutcorrection.Doingsomaycausefire.Switchpoweroffimmediately,unplugACadapterfromACoutletinsuchcase.Ifoperatedonbattery,removebatteryandcontactthenearestKonicaMinolta-authorizedservice facility.

Caution (Failuretoadheretofollowingpointsmayresultininjuryordamagetoinstrumentorotherproperty.)

UsethisinstrumentnearACoutletforeasypluggingorunplugginginusingACadapter.

Connection procedure 1. Be sure that power switch is OFF ( [ ]

mark pressed).

2. Connect AC adapter plug to AC adapter input terminal of body.

3. Plug AC adapter to outlet (AC 100-240 V , 50/60 Hz).l InsertACadapterplugallthewayseatedinACoutlet.

DC・IN

Page 21: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

19

Installing

ON ( | ) / OFF ( ) of Power SwitchTo ON

1. Press [ | ] side of power switch to ON.

To OFF

2. Press [ ] side of power switch to OFF after completion of measurement.

Connecting with PCBeforeconnectingCA-2500bodywithPC,installsoftwareCA-S25winPC.(Fordetail,seetheInstallationGuidefordatamanagementsoftwareCA-S25w.)UseaccessoryUSBcable(3m)IF-A18forthispurpose.Noproblemwouldoccur ifUSBcable ispluggedorunpluggedwhilepower ison,but instructionhereindicates the case when power is off.

Operation procedure1. Switch power off [ ] .

2. Connect USB cable to USB connector on body.

3. Insert all the way to the end. Check whether the connection is secure.

Communication interfaceof this instrument complieswithUSB2.0.AlwaysholdplugportionofUSBcable inunplugging.Neverhold cord portion to unplug. MatchtheUSBcableplugshapewiththeslotofeitherbodyorPCbeforeinserting.AnyUSBport isusableonPCif ithasmultiple,butabnormaloperationcouldbedetectedonPC in case that it isusedtogetherwithotherUSBdevicethanCA-2500.

Page 22: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

20

Page 23: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

21

Measurement

Page 24: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

22

Measurement

Starting Up and Ending SystemStarting up

Switchpoweronasinstructed.Ifnot,erroneousoperationcouldoccur.

1. Switch power on for body. (Press [ | ] side.) l Motorsoundsindicatingcontrollerinsidemachinestartsinitialization.l Fan inside body starts to run. l15-minutewarm-upat least(to leaveinthisstateafterswitchingon.) isrecommendedforaccurate

measurement.

2. Start up Windows by turning on controlling PC.

3. Connect PC and CA-2500 with USB cable.lPCrecognizesthatCA-2500hasbeenconnected.

4. Select "CA-S25w" from Start → [Program].lCA-S25wstartstorun.

Connection recognition error Sometimes,PCfailstorecognizeitsconnectionwithCA-2500ifCA-2500wasswitchedonafterconnectedwithUSBcabletoPCwhichhadalreadystartedup.SuchconditioncouldnotallowcontrolofCA-2500fromsoftware.Startupthisinstrumentaccordingtotheprocedureasaboveagain.l IncasethatPChasalreadystarted,unplugUSBcableonceandthenconnectafterswitchingCA-2500

on. lSwitchCA-2500onbeforePCifUSBcablecannotbeunpluggedforsomereasonconcerningsetting. Shouldconnectionerroroccur,unplugUSBcableonceandconnectagain.Then,bothcanstartup

normally.

Ending

1. Select "End CA-S25w" from "File" of menu bar.lCA-S25wends.lYoucanfinishthroughstandardapplicationoperationofOS.

2. Finish Windows and switch off PC.

3. Switch off body. (Press [ ] side.)

Page 25: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

23

Measurement

Setting the measurement distance and the pint ring range index Setting measurement distanceFivetypesofobjective lensesareavailableforCA-2500dependingonthesizeofmeasuringobjects.Therelationbetweenthesizeofmeasuringobjectsandthemeasurementdistanceforeachlensisasfollows:

CA-2500A, CA-2500S, CA-2500SW, CA-2500ST CA-2500A, CA-2500W, CA-2500SW, CA-2500WTStandard lens Wide lens

Size of measurement object (One-side length of square) Measurement distance Size of measurement object

(One-side length of square) Measurement distance

98mmapprox. 250mm 145mmapprox. 200mm

210mmapprox. 500mm 410mmapprox. 500mm

440mmapprox. 1000mm 850mmapprox. 1000mm

890mmapprox. 2000mm 1770mmapprox. 2000mm

CA-2500A, CA-2500T, CA-2500ST, CA-2500WTTelephoto lens With low-magnification macro ring (Macro 1)

Size of measurement object (One-side length of square) Measurement distance Size of measurement object

(One-side length of square) Measurement distance

115mmapprox. 900mmapprox. 57mmapprox. 500mm

275mmapprox. 2000mmapprox. With high-magnification macro ring (Macro 2)420mmapprox. 3000mmapprox. Size of measurement object

(One-side length of square) Measurement distance

27mmapprox. 300mm

Optical lens features in varying light intensity and illuminance distribution of imaging surface if focus ring is rotated. Therefore, lens calibration has been done lens by lens according to the distance from lens to measurementobject(distanceindicatoroffocusring).Eachhasitsowncalibrationcoefficient.Correctcalibrationcoefficientcanbeappliedbyselectingwhatkindof lens isused forwhich focusingdistance,insoftwareCA-S25w.

CA-2500A, CA-2500S, CA-2500SW, CA-2500STStandard lens

Distance With distance label On indicator

0.25m ○ Decimalpointindicatorposition(SeeFigure1-1.)

Intermediateindicatorbetweenpreviousandnextreferencepositions(Center)(SeeFigure1-2.)

0.3m ○ Decimal point indicator position

Intermediateindicatorbetweenpreviousandnext referencepositions(Center)

0.5m ○ Decimal point indicator position

Intermediateindicatorbetweenpreviousandnextreferencepositions(Center)(SeeFigure1-3.)

1 m ○ Centerof"1"

Intermediateindicatorbetweenpreviousandnext referencepositions(Center)

○ Endoffocus-ringtravel([ ]side)

Page 26: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

24

Measurement

CA-2500A, CA-2500S, CA-2500SW, CA-2500STWide lens

Distance With distance label On indicator

0.2m ○ Indicatorposition(SeeFigure3-1.)

0.24m ○ Indicatorposition

0.3m ○ Indicatorposition(SeeFigure3-2.)

0.5m ○ Indicatorposition

1 m ○ Indicatorposition

○ Indicatorposition

CA-2500A, CA-2500S, CA-2500SW, CA-2500STWide lens

Distance With distance label On indicator

0.9m ○ Decimalpointindicatorposition(SeeFigure2-1.)

Intermediateindicatorbetweenpreviousandnext referencepositions(Center)

1 m ○ Centerof"1"

Intermediateindicatorbetweenpreviousandnextreferencepositionsin1:2(short-distanceside)(SeeFigure2-2.)Intermediateindicatorbetweenpreviousandnextreferencepositionsin2:1(long-distanceside)(SeeFigure2-3.)

1.5 m ○ Decimal point indicator position

Intermediateindicatorbetweenpreviousandnextreferencepositionsin1:2(short-distanceside)Intermediateindicatorbetweenpreviousandnextreferencepositionsin2:1(long-distanceside)

3m ○ Centerof"3"

Intermediateindicatorbetweenpreviousandnextreferencepositionsin1:2(short-distanceside)Intermediateindicatorbetweenpreviousandnextreferencepositionsin2:1(long-distanceside)

○ Endoffocus-ringtravel([ ]side)

With low-magnification macro ring (Macro 1)

Distance With distance label Precise position on focus ring

○ Endoffocus-ringtravel([ ]side)

With high-magnification macro ring (Macro 2)

Distance With distance label Precise position on focus ring

○ Endoffocus-ringtravel([ ]side)

Caution TheCA-2500has28calibrationpointsasshown in the tableabove,atwhich themeasuringaccuracyisensured.Whenusing the instrumentatapointother than thecalibrationpoints,use thecalibrationcoefficient of the closest calibration point. Doing so may result in a decrease in measuring accuracy.

Page 27: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

25

Measurement

Example of reference position to read distanceHow to read distance in case of standard lens or telephoto lensThering(focusring)onwhichnumbervaluesarewrittenrotatesincaseofstandardlensortelephotolens.If the"distanceindicator" includesdecimalpoint,"." isdeterminedasreferenceposition;andif itdoesnotincludedecimalpoint,thecenterofthenumberor[ ]isdeterminedasreferenceposition.

Standard lens

Telephoto lens

How to read distance in case of wide lensThering(focusring)onwhichindicatorposition(thewhitecircleinthefiguresbelow)iswrittenrotatesincaseofwidelens.Settheindicatorpositiontothelineofnumbervalue.

Wide lensFigure 3-1Example when setting to "0.2"

Figure 3-2Example when setting to "0.3"

0.30.5 0.24 0.21 0.30.5 0.24 0.21

Figure 1-1Example when setting to "0.25"

0.30.25 0.30.25 0.5 1

Figure 1-3 Example when setting to "diverting point between 0.5 and 1 in 1:1"

Figure 1-2Example when setting to "diverting point between 0.25 and 0.3 in 1:1"

Setinreferenceofthe1/2position between decimal point of0.5andthecenterof1.

Setinreferenceofthe1/2position between decimal pointsof0.25and0.3.

10.9 1.510.9 1.51

Figure 2-3 Example when setting to "diverting point between 1 and 1.5 in 2:1"

Figure 2-2 Example when setting to "diverting point between 1 and 1.5 in 1:2"

Figure 2-1 Example when setting to "0.9"

Setinreferenceofthe1/3position between the center of 1 and decimal point of 1.5.

Setinreferenceofthe2/3position between the center of 1 and decimal point of 1.5.

Page 28: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

26

Measurement

Page 29: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

27

Others

Page 30: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

28

Others

Outer Dimensions

WARNING:WARNUNG:Niemals das gehause offnen.

Do not remove covers.警告:カバーを開けないでください。

XXXXXX1

192.1

223

Reference position

113160

97

72.7

Tripod socketUNC 1/4, Depth : 6.5

Positioning hole Depth : 4.5

Ø5 positioning hole Depth : 4.5

M5, Depth : Ø6.5

Rubber foot diameter : Ø15

186.2

128.961.2

68.2 6060

81.295.2111.2125.2

160

52.8

91

5

164.

3211.

3

12.7

47

9

Page 31: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

29

Others

Error CheckShoulderrorbe found in this instrument, trycorrectiveactionsshown in the following table. If thisdoesnothelp, this instrumenthaspossiblybeenbroken.Pleasecontact thenearestKonicaMinolta-authorizedservice facility with indicated error number.

ErrorNo.

Symptom ItemtoCheck CorrectiveAction

1 Power does not turn on.Does not normally start up.

HasACadapterbeenpluggedin outlet?

ConnectACadapter.

HasACadapterbeenconnected with body?

ConnectACadapter.

HaveyouconnectedcorrectAC adapter?

UseourspecifiedstandardaccessoryAC adapter and power cord or optionally offeredACadapterAC-A312.*WhenyoudisconnectawrongadapterandreconnectanAC-A312adapter,waitat least 5 minutes before turning the power switch on after turning it off.

IstheACpowerwithinrating? BesuretheACpoweriswithin±10%ofthe nominal voltage.

2 Time for machine initialization after switched on varies.

Itcouldvarydependingonthe machine position when switched on.Itshouldnotberegardedas failure if it takes only a few seconds.

Iflongtimeisneeded,switchpoweroffand contactthenearestKonicaMinolta-authorized service facility.

3 PCcannotcommunicate withCA-2500.

CheckwhetherPCcorrectlyrecognizesCA-2500.

PCcouldnotproperlyrecognizethisinstrumentdependingonPCtypeorhow CA-2500wasstartedup.RestartPC, software,andCA-2500accordingtoeach instruction.

Checkwhetherdevicedrivernormallyoperates with device driver manager.

Ifnot,uninstallsoftwareonceforreinstallation.SeetheinstructionmanualforCA-S25w.

HasUSBcablebeenconnected securely?

ConnectthisinstrumentandPCsecurely.

CheckthestandardforUSB. CannotusewiththosesupportingexceptforUSB2.0.UsePCwithwhichUSB2.0isusable.

Page 32: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

30

Others

Main SpecificationModel CA-2500S CA-2500W CA-2500T

Light receptor CCD imagesensor (monochrome);2/3-inch;Effectivenumberofpixels:1,000x1,000pixels;EquippedwithXYZfilter(closelymatchesCIE1931color-matchingfunction)andNDfilter

Lens InterchangeableStandard,wide,andtelephotolenses;low-magnificationandhigh-magnificationmacrorings(forusewithtelephotolens)

Measurementpoints(Resolution) 980x980(490x490or196x196selectable byusingDataManagementSoftwareCA-S25w)

Colorindicationmodes XYZ,Lvxy,Lvu’v’, T uv,Dominantwavelength,Excitationpurity,Lv contrast

Display modes Pseudocolor,Chromaticitydiagram,Spot,Crosssection,Colordifference

Measurementsizes(lengthpersideofsquare)(*1)

Standardlens Widelens Telephoto lens Withlow-magnificationmacro ring

Withhigh-magnificationmacro ring

Approx.98mm or more (depending on the distance)

Approx.145mm or more (depending on the distance)

Approx.115mm or more (depending on the distance)

Approx.57mm(Fixed)

Approx.27mm(Fixed)

Measurablesizefortypicalmeasurementdistances(size/distance)

98mm/250mm Approx.

145mm/200mmApprox.

115 mm/900mmApprox.

57mm/500mmApprox.(Fixed)

27mm/300mmApprox.(Fixed)

210mm/500mmApprox.

410mm/500mmApprox.

275mm/2,000mm Approx.

440mm/1,000mm Approx.

850mm/1,000mm Approx. 420mm/3,000mm

Approx.890mm/2,000mm Approx.

1,770mm/2,000mm Approx.

Measurementluminancerange(includingNDfilteruse)

0.05-100,000cd/m2 0.05-100,000cd/m2 0.25-100,000cd/m2 0.25-100,000cd/m2 0.5-100,000cd/m2

Measurementtime(*2)Single:Approx.5sec.ormore;4-timeintegration:Approx.6sec.ormore;16-timeintegration:Ap-prox.10sec.ormore;64-timeintegration:Approx.25sec.ormore;256-timeintegration:Approx.80sec.ormore

Accuracy(*3)

Luminance ±3% ±3% ±3% ±3% ±3%

Chromaticity ±0.005 ±0.005 ±0.005 ±0.005 ±0.005

Temperature/humiditydrift(withintheoperatingtemperature/humidityrange)Luminance:±2%ofchangecomparedtoreferencetemperatureandrelativehumidityof23°Cand40%Chromaticity:±0.004ofchangecomparedtoreferencetemperatureandrelativehumidityof23°Cand40%

Repeatability (*4)

Luminance 0.5% 0.5% 0.5% 0.5% 0.5%

Chromaticity 0.001 0.001 0.001 0.001 0.001

Inter-pointerror(*5)

Luminance(*6) ±2% ±2% ±2% ±2% ±2%

Chromaticity(*6) ±0.002 ±0.002 ±0.002 ±0.002 ±0.002

Luminance(*7) ±3% ±3% ±3% ±3% ±3%

Chromaticity(*7) ±0.003 ±0.003 ±0.003 ±0.003 ±0.003

Other functions Measurementsync(Synchronizationfrequencyselectable),UsercalibrationIntegrationfunction

Interface USB2.0orhigherOperating temperature and humidity range(*8) 10-30°C,Relativehumidity70%orless/Nocondensation

Storagetemperatureandhumidityrange(*8)

0-30°C,Relativehumidity70%orless/Nocondensation,30-35°C,Relativehumidity55%orless/Nocondensation

SizeBody only 160(W)×164(H)×192(D)mm(Heightincludinghandle:211mm)Whenlensandlenshoodareattached 224(D)mm 219(D)mm 224(D)mm 230(D)mm 237(D)mm

Weight 3.5kgapprox.(whenstandardlensandlenshoodareattached)

Power source ACadapter100-240V ,0.75A, 50/60Hz

Standardacces-sories

LensHood CA-A68 CA-A69 CA-A68 CA-A70

HoodCap CA-A71 CA-A75 CA-A71

CalibrationdataDVD Includedwitheachlens.

Other MountCover,ACAdapterAC-A312,ACcable,USBCableIF-A18,DataManagementSoftwareCA-S25w,SDKCA2D-SDK

Optional accessories SoftCaseCA-A60,TripodCS-A3,PanHeadCS-A4,Calibrationcertificate,MuraEvaluationSoftwareCA-Mura

Page 33: 2D Color Analyzer CA-2500 Series Instruction Manual · PDF file2D Color Analyzer CA-2500 Series Instruction Manual ... For details on having the instrument inspected ... CA-2500 CA-2500

31

Others

*1: Errorinangleofview:7%

*2: Measurementtimediffersdependingonbrightnessofmeasurementobject,PCperformance,anddataprocessingcontents. ThespecificationsabovewereobtainedunderKonicaMinolta'smeasurementconditionsshownbelow: PCCPU : Intel®Core™i7-37703.40GHz Memory :8GB Dataprocessing :Pseudocolordisplayofluminancedata Resolution :980x980 Shutterspeed :Ymeasurement:1/64sec.,XZmeasurement:1/32sec. Measurementsubjectbrightness: Standard/widelens:Approx.80cd/m2,Telephotolens:Approx.300cd/m2

Low-magnificationmacroringandtelephotolens:Approx.400cd/m2, High-magnificationmacroringandtelephotolens:Approx.600cd/m2

*Themeasurementtimebecomeslongerwhentheobjectisdark.Thelongestmeasurementtimeisapprox.14secondswith1-timeintegration,approx.45secondswith4-timeintegration,approx.3minuteswith16-timeintegration,approx.11minuteswith64-timeintegrationandapprox.44minuteswith256-timeintegration

*3:ThespecificationsabovewereobtainedunderKonicaMinolta'smeasurementconditionsshownbelow: Measurementsubjectbrightness: Standard/widelens:Approx.35cd/m2, Telephotolens:Approx.140cd/m2

Low-magnificationmacroringandtelephotolens:Approx.250cd/m2, High-magnificationmacroringandtelephotolens:Approx.250cd/m2

Distance : Minimumdistanceofeachlens Evaluation : Basedontheaverageobtainedwithin10%rangeatthecenterofthescreen Temperature : 23°C±2°C Relativehumidity : 40%±10% Measuringlight : White,referencelightsource Integration : 64times(Normalmode)

*4:ThespecificationsabovewereobtainedunderKonicaMinolta'smeasurementconditionsshownbelow: Resolution : 196x196 Shutterspeed : Ymeasurement:1/64sec.,XZmeasurement:1/32sec. Gain : Normal(x1), Lightintensitylevel: Midpointofappropriatelightintensityrange Evaluation : Basedonthemaximumrepeatability(2 )ofallpixels Temperature : 23°C±2°C Relativehumidity : 40%±10% Measurementsubject: White,referencelightsource Integration : 64times(Normalmode)

*5:ThespecificationsabovewereobtainedunderKonicaMinolta'smeasurementconditionsshownbelow: Measurementsubjectbrightness: Standard/widelens:Approx.40cd/m2,Telephotolens:Approx.160cd/m2

Low-magnificationmacroringandtelephotolens:Approx.200cd/m2, High-magnificationmacroringandtelephotolens:Approx.350cd/m2

Distance :Calibrationdistanceofeachlens Resolution :196x196 Evaluation(*6) :Basedonthemaximum/minimumvaluesobtainedinasquareatthecenterofthescreenmeasuring60%oftheheight

and width of the entire screen (*7) :Basedonthemaximum/minimumvaluesobtainedintheentirescreen

Temperature :23°C±2°C Relativehumidity :40%±10% Measurementsubject :White,referencelightsource Integration :64times(Normalmode)

*8:Even if the instrument isstoredwithin thespecifiedusage(orstorage) temperaturehumidity range, thedisplayedvaluemaychangedependingonlong-periodusageorstorageconditions.

• Windows®,Excel®andWord®areregisteredtrademarksoratrademarksofMicrosoftCorporationintheUnitedStatesandothercountries.• Intel® andPentium® are a registered trademarks or trademarksofIntelCorporationintheUnitedStatesandothercountries.• Thespecificationsanddrawingsgivenherearesubjecttochangewithoutpriornotice.• Forspecification,pleasecontactyournearestKonicaMinoltasalesofficeordealer.