Upload
ugo-f-amadi
View
214
Download
0
Embed Size (px)
Citation preview
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 1/25
CALIBRATION
Procedures, StandardAnd
Methodologies
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 2/25
Introduction
• In this section, the procedures andmethodologies to carry out a
calibration will be discussed. Inaddition, the approach to calculateerrors in a calibration will be shown.
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 3/25
Objective Of Calibration
• The prime objective of calibration isto ensure that measurements takenare accurate at all times taking intoconsideration wear and tear,degradation and abuse. It is with thisin mind that we need to adopt and
appropriate methodology to do aproper calibration.
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 4/25
Scope of Calibration
In developing a calibration program, the followingshould be taken into consideration :
Involve all testers or those who use measuringand test equipment
Calibration must be conducted on all equipmentused to verify product compliance.
Engineering measuring equipment must all becalibrated.
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 5/25
Hierarchy of Standards• The hierarchy of standards used in a calibration is shown as
follows :
Primary or National standards are the ultimate reference standards for determining ameasurement. As an example, a standard measurement of a meter is set by a steel rodof this length which is the primary standard. All other standards used to conduct acalibration must be checked with reference to this standard from the working to thereference standard
PrimaryStandard
Reference Standard
Working Standards
Measuring & Test Equipment
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 6/25
Standard Adequacy
• Working standard used to calibrate internal measuringequipment must be traceable to either a National or PrimaryStandard. In addition, their accuracy must be 4 to 10times more accurate compared to the equipment to becalibrated. For instance, portable water analyzer such as pH
meter, NTU meter with the standard measurement scale,can be calibrated internally by using the standard solutionthat provided.
• Occasionally, an equipment may be state of the art and
hence, there is no reference standard available. In thiscase, this equipment can be calibrated with anotherequipment of the same type, make and model.
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 7/25
Traceability of Calibration
• In a country, a calibration must be traceable to a National
Standard held by a national body (eg. SIRIM or othergovernment agencies of Malaysia). If this national standardis not available. The calibration must be traceable toanother country. Traceability of a calibration is importantas an incorrect reference standard used will not ensure
that the equipment calibrated has the required accuracy.
• As most companies send out their equipment for calibration,mandatory that they check the capability of the calibrationagent. Those that are certified to the ISO 9000 Guide 25
will be able to provide an appropriate traceability. Forothers, you will need to check the reference standards usedand their respective traceability.
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 8/25
Factors to Consider in a Calibration
• Calibration Environment
• Interval of Calibration
• Accuracy of Standard Master
• Out of Tolerance Equipment
• Subcontractor Controls• Uncertainty of Measurement
• Handling the Storage
• Identification and Status of Calibration
• Training• Documentation and Records
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 9/25
Calibration EnvironmentEnvironment conditions which could affect a calibration are as follows. This willlargely dependent on the sensitivity of an equipment to the various conditions. Consult
the equipment manufacturer to determine the appropriate environmental conditionsrequired to do a calibration.
• Temperature• Humidity• Vibration• Dust
• Noise• Light• Air Movement• Magnetic Field
Some examples where controlling certain environment factors is critical for thedifferent types of measurement are as follows :
Chemical Analyses – pH value and NTU are critical
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 10/25
Interval of Calibration• Interval between one calibration to another is an important aspect to
ensure the continuing accuracy of an equipment. This must be establishedfor all measuring and test equipment, each may be set at differentintervals dependent on the degree of usage and other factors such as:
> Stability
> Frequency of Use
> Accuracy> Purpose of Use
• If you are unsure which is the best interval to choose, always follow theequipment manufacturer’s recommendations. This interval should bereviewed from time and as a rule of thumb, it should be adjusted based on
the calibration history. For example, if an equipment is calibrated on a sixmonthly basis and it is found to be out of calibration frequently, then thecalibration interval should be changed either 2 or 3 monthly.
C LIBR TION INTERV L
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 11/25
CALIBRATION INTERVAL
• What actions and the calibration interval on the equipment would you recommend tothe user for the following :
1. An instrument is found to have failed 2 times in the last 2 calibrations
performed at 6 monthly intervals.
2. A newly purchased pH on-line analyzer which will be used in production daily.
Calibration interval recommended by manufacturer is 3 monthly.
3. A Magnetic Flow Meter used to measure the flow of water with
specification of 0.5 to 1 % accuracy.
Accuracy of Magnetic Flow Meter is 0.5 to 1% with Calibration interval of
yearly using the Flow Meter Verificator.
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 12/25
ACCURACY OF REFERENCE STANDARD
• As discussed in the previous section, thereference standard used to do a calibration mustbe at least 4 to 10 times more accurate than the
equipment. In general, a 95% confidence level ispreferred and this gives a + 2 sigma (standarddeviation) or 4 sigma capability in a measurement.
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 13/25
OUT OF TOLERENCE EQUIPMENT
• When an equipment is found to be out of tolerance or not having the requiredaccuracy, certain steps must be taken. Actions to be taken include decisions on
whether the equipment can be used or not and what needs to be done on the affectedproducts as follows :
> Review all measurements prior to and after adjustment is made to the
affected equipment.
> Decide on whether the equipment can be used or has to be thrown away
or used for lesser accuracy measurements.> Review previous batches of products ( 1 week or more) where
measurements have been taken using this out of tolerance equipment
prior to and after an adjustment has been made to correct the out of
calibration situation. Decide whether they meet specification or not.
> If the equipment can be used but needs correction to their readings, then the
compensation or offset factors must be stated on the equipment.
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 14/25
Subcontractor Controls
• In choosing an appropriate subcontractor to perform yourcalibration, it is critical that an accredited calibrating labor agent is used (those certified to ISO 9000 Guide 25).Ask for a copy of their certification before selecting them.You will also need to ensure that the reference standards
used by the calibrating lab or agent is traceable to nationalstandard. The calibration method used should also bechecked to ensure they used an appropriate or recognizedprocedure. Both of these requirement can be verified ontheir Certificate of Calibration, which is issued to you uponcompletion of a calibration for each set of equipment.
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 15/25
Uncertainty Of Measurement
The uncertainty of measurement of a calibration needs to becalculated to determine all the possible sources of errors. It isgiven by the formula :
Uncertainty of Measurement = + t (from a Students t-table) x s(standard deviation of variation).
As a rule of thumb, the uncertainty of measurement must not exceed30% of the tolerance of a measurement. If it does, it’s mean that theequipment is not suitable for use as its required accuracy is not
sufficient. A measurement taken using this equipment will give a falseor inaccurate value.
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 16/25
Handling and Storage• The proper handling and storage of measuring equipment and their
reference standards are an essential part of establishing aneffective Calibration Program. Some of the guidelines to follow areas follows:
Handling – ensure the equipment is appropriately handled during use, storage andtransportation.
Proper Storage – ensure equipment is stored in their respective casing or specialstorage container under appropriate environmental conditions as required.
Proper Packing for Transportation – ensure the equipment is appropriately packed toprotected them from movement and damage during transportation (eg. Use “bubblebags” or impact absorbent materials).
Check that the calibration status is valid before issuing an equipment for use(especially newly purchased equipment).
Ensure that the equipment is appropriately sealed with a tamper-proof sticker ormarking, etc to prevent unauthorized adjustment or repair.
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 17/25
Identification and Status of Calibration
• Upon completion of a calibration, each equipmentmust be appropriately labeled to indicate theirstatus. Depending on the size of the equipment, itmay not be possible to attach a calibration labelon all instruments. For an equipment such asthermometer it may not be practical to stick sucha label. Therefore, other means such as colorcoding can be used instead of a label to indicate
their calibration status. Alternatively, the labelmay be attached to the casing or container.
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 18/25
Example of the various labels or identifications which can be used are asfollows:
• Passed Calibration
• Tamper Proof Label
These labels are used to prevent tampering on the equipment after it is calibrated.
An example is as follows :
ID#: DATE:
BY: DUE:
VOID Calibration Seal
If Broken
PBA
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 19/25
• “DO NOT USE” Label
This label is used to indicate that a specific equipment cannot be usedeither because it is damaged or needs repair or malfunctioned or itscalibration is out of tolerance.
• “No Calibration Required” Label
This label is used to indicate those equipment which do not require anycalibration. These are mainly those used for rough measurement and forwhich accuracy is not important. Some examples are processinstrumentation such as chlorine analyzers and ultrasonic level sensor, etc.
DO NOT USE
PBA Perbadanan Bekalan AirPulau Pinang Sdn. Bhd.
CALIBRATION NOT
REQUIRED
Perbadanan Bekalan AirPulau Pinang Sdn. Bhd.PBA
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 20/25
Training
• Personnel appointed to perform calibration mustbe appropriately trained together with propereducation and experience. They must understandthe characteristic and features of an equipment,and its usage (ie. Required accuracy, stability,
sensitivity, etc). Furthermore, the personnel mustknow the calibration procedures andmethodologies including the calculation ofuncertainty of measurement and sources oferrors. Various training courses are now availableand those responsible for performing calibrationmust attend these courses.
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 21/25
Documentation and Records
• Appropriate documentation and records must be maintained aspart of a calibration program. Essential ones are as follows :
a) Master List of Equipment
A master list of equipment shall be maintained to indicate all the measuring equipment used in acompany. It should state as a minimum, the following details :
Equipment Identification (ID) Location
Required Accuracy
Measurement Range
Calibration Interval
Calibration Lab/Agent (internal or external)
Date Calibrated
Date due for Next Calibration
Pass / Fail Calibration
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 22/25
b) Calibration Plan
For each type of equipment, a calibration plan shall be developed. This plan shalloutline all the steps to conduct a calibration including :
Reference standards to be used ;
Recording methodologies ;
Environmental conditions for calibration ;
Calculations ;
Approach to handle an out of tolerance situation.
The plan should also outline the recalling system to ensure each piece of equipment iscalibrated either on or before their due date. A reminder system or calibrationschedule is necessary.
c) Training Records
Training records shall be maintained for all personnel authorized to performcalibration.
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 23/25
d) Out of Calibration Notice and Records
Notices advising the relevant department and personnel of an equipment found to be
out of calibration and the corresponding actions (s) taken shall be maintained.
e) Manufacturers’ Handbooks
Manufacturers’ handbooks or reference manuals used as part of calibration shall alsobe controlled and kept by the department responsible for calibration.
f) Calibration Certificate
Calibration certificate provide by an external calibrating organization must be kept asproof of calibration and are to be maintained as part of the calibration records.
Internal calibration certificate also required to be kept at the respective cost centre.
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 24/25
g) Records of Usage
Records of usage are to show the persons who have used the equipment and toindicate the degree of usage and what measurements were taken.
h) Measurement Equipment History Record
A history record should be maintained for each piece of equipment. It should containthe following details :
make, model and type or brand
date of purchase
department using this equipment and its location use
date of first calibration
adjustment made during calibration (if any)
repair & service details
action taken on out of tolerance situation (if any)
date of disposal (when it is unusable).
7/29/2019 23871687 Calibration
http://slidepdf.com/reader/full/23871687-calibration 25/25
Thank youPrepared By:
Mohd Suhaimi Shariff