2.2 Analysis of Common Ions

Embed Size (px)

Citation preview

  • 8/11/2019 2.2 Analysis of Common Ions

    1/17

    MODULE 2.2

    Analysis of common ions at low

    concentrations in water

    Ultraviolet And Visible Spectrometry 1

    Spectrophotometric instrumentation 2

    Analysis by direct absorpt ion 3

    Nitrate determination 3

    Analysis after formation of derivative 4

    Chloride: Automated method using

    mercuric thiocyanate and ferric nitrate

    4

    Fluoride: Zr-Alizarin lake method 5

    Nitrite 5

    Phosphate 5

    Automatic procedures 5

    Field Techniques 6

    Flame photometry 6

    Ion Chromatography 7

    Examples Of The Use Of Other Techniques 11

    Ammonia 12

    Fluoride 13

    Sulphate 13

    Free chlorine (Residual chlorine) 14

    Sulphide 14

    Sample Collection and preservation 15

    Spectrophotometric determination of H2S 15

    Titrimetric method 15

    Cyanide 16

  • 8/11/2019 2.2 Analysis of Common Ions

    2/17

    MODULE 2.2

    Analysis of common ions at low concentrations inwater:

    In this chapter the application of instrumental techniques to determine the

    concentration of ions which are present in mg l -1 concentration range are

    discussed . One of the major advantages of the use of instrumental techniques is

    that elaborate sample preparation is not necessary. However when it comes to

    the analysis of ions atg l-1 levels, a preconcentration step is needed inorder to

    bring the concentration of the analyte ions within the working range of the

    instruments. In that case the instrumental part becomes just one part of a more

    complex analytical procedure. The analysis of ions present at g l-1

    concentration

    level are discussed in the next chapter.

    Ultraviolet And Visible Spectrometry:

    This technique is based on Beer-Lambert law. That is at sufficiently low

    concentrations of the absorbing species the above law is obeyed which can be

    expressed mathematically as

    A = cl

    Where A = absorbance of radiation at a particular wavelength;

    o

    t

    I

    A log I

    =

    = Intensity of incident radiation;

    = Intensity of transmitted radiation;

    = Molar absorptivity; l mol

    oI

    tI

    -1cm-1

    1

  • 8/11/2019 2.2 Analysis of Common Ions

    3/17

  • 8/11/2019 2.2 Analysis of Common Ions

    4/17

    For absorption in the visible region, a tungsten lamp is used as a source and

    that for absorption in uv region hydrogen lamp is used. The light of the desired

    wavelength is isolated using diffraction monochromator. Normally for all

    absorption measurements 1 cm cells are used and for all uv absorption

    measurements cells made of quartz are used. Different types of detectors such

    as photocell, photomultiplier or photodiode array detectors are used in the

    absorption measurements.

    Analys is by direct absorption:

    Nitrate determination:

    This method is useful only for screening water samples that have low

    organic matter contents, i.e., uncontaminated natural waters and potable water

    supplies.The calibration curve follows Beer's law upto 11

    ppm. Measurement of uv absorption at 220nm enables rapid determination of

    nitrate. Because dissolved organic matter also may absorb at 220 nm and

    does not absorb at 275 nm, a second measurement made at 275 nm, may

    be used to correct the value. The extent of this empirical correction is

    related to the nature and concentration of organic matter and may vary from one

    water to another. This method is therefore not recommended if a significant

    correction for organic matter absorbance is required, although it may be useful in

    monitoring levels within a water body with a constant type of organic

    matter. Sample filtration is intended to remove possible interference from

    suspended particles. Acidification with 1N HCl is designed to prevent interference

    from hydroxide or carbonate concentration upto 1000 ppm as CaCO

    -

    3NO

    -

    3NO

    -

    3NO

    -

    3NO

    3. Chloride

    does not interfere in the determination. However organic dissolved matter,

    surfactant, , and Cr(VI) do interfere.-2NO

    3

  • 8/11/2019 2.2 Analysis of Common Ions

    5/17

    Analysis after formation of derivat ive:

    Chloride: Automated method using mercuric thiocyanate andferric n itrate:

    Thiocyanate ion is liberated from mercuric thiocyanate solution by the

    formation of soluble mercuric chloride. In the presence of ferric ion, free

    thiocyanate ion, forms a highly coloured ferric thiocyanate complex

    (max=470 nm) the absorbance of which is proportional to the chloride

    concentration.

    The interference due to particulate matter can be overcome by filtration or

    centrifugation before analysis.

    The method is applicable to potable, surface, and saline waters, and

    domestic and industrial waste waters. The concentration range of chloride that

    can be measured is 1 to 200 ppm.

    Fluoride:Zr-Alizarin lake method:

    This SPANDS Colorimetric method is based on the reaction

    between fluoride and a coloured zirconium-alizarine lake. Fluoride reacts with the

    Zirconium alizarin lake, dissociating a portion of it into a colourless complex

    anion (ZrF62-). As the amount of fluoride increases, the colour of zirconium dye

    lake becomes progressively lighter.

    The reaction between fluoride and zirconium ions is influenced

    greatly by the acidity of the reaction mixture. If the proportion of acid in the

    reagent is increased, the reaction can be made almost instantaneous. The

    absorbance measurements can be done at 570nm and concentrations can be

    4

  • 8/11/2019 2.2 Analysis of Common Ions

    6/17

    determined with calibration graph. Both the standards and samples should be

    kept for one hour before absorption measurements.

    Nitrite:

    Nitrite is determined through formation of a purple azo dye produced

    at pH 2.0 to 2.5 by coupling diazotised sulfanilamide with

    N-(1-naphthyl)-ethylenediaminedihydrochloride. The absorbance of the resulting

    purple azo dye can be measured at 543 nm using a spectrophotometer. Beer's

    law is obeyed upto 25 g

    -

    2NO

    -1. Higher Concentration of can be determined by

    diluting the sample. Free chorine and the following ions which form precipitatesunder test conditions such as Sb

    -

    2NO

    3+ ,Au3+, Bi3+, Fe3+, Pb2+, Hg2+, Ag+ interfere .

    Cupric ion may cause low results by catalysing decomposition of the diazonium

    salt.

    Phosphate: The procedure for phosphate involves the addition of a mixed

    reagent (sulphuric acid, ammonium molybdate,ascorbic acid, antimony

    potassium tartrate) to a known volume of sample, diluting to volume, shaking and

    leaving for 10 min. A blue phosphomolybdenum complex is produced and the

    absorbance is measured at 880 nm.

    The concentration is calculated using a predetermined calibration graph

    derived from standard solution treated in the same way.

    Automatic procedures:have been developed for most of the ions listed in

    the previous paragraph.

    Instead of mixing reagents for each analysis, streams of each reagent

    (segmented by air bubbles to diminish mixing effects) in narrow-bore tubes are

    5

  • 8/11/2019 2.2 Analysis of Common Ions

    7/17

    mixed by combining the flows at a T-junction or within a mixing cell. A sample is

    introduced from an automatic sampler as a continuous flow into the reaction

    stream. The combined flow is then led into a spectrophotometer and the

    absorption measured. The flows of all the reagents and samples are produced

    from a multi-channel peristaltic pump.

    Field Techniques:

    Field techniques are very important and they give immediate

    measurement of ion concentration. Using the automated procedures described

    above unmanned field stations can be set up. Otherwise portable instruments

    can be used to perform the analysis at the site.

    Flame photometry:

    This technique is used in water analysis for determining the concentration

    of alkali and alkali metal ions such as sodium, potassium and calcium. The

    following diagram (fig 2) shows the basic components of a flame photometer.

    MonochromatorDetector

    & read out

    Nebuliser & burner

    sample

    solution

    in

    flame

    lens

    Fig2. Outline of a flame photometer

    6

  • 8/11/2019 2.2 Analysis of Common Ions

    8/17

    The liquid sample to be analysed is sprayed under controlled conditions

    into the flame where the water evaporates, leaving behind the inorganic salts as

    minute particles. These salts decompose into constituent atoms or radicals and

    may become vapourised. The vapours containing the metal atoms are excited by

    thermal energy of the flame and this causes electrons of the metallic atoms to be

    raised to higher energy levels. When these excited electrons fall back to their

    original positions, they give off discrete amounts of radiant energy. The emitted

    radiation is passed through the monochromater where the desired region

    isolated. A photocell and an amplifier are then used to measure the intensity of

    isolated radiation. Normally for alkali metals a propane-compressed air mixture is

    used as a fuel.

    A linear concentration range (for sodium and potassium 1 to 10 mg l-1 and

    for calcium 10-50 mg l-1

    ) is within the range expected for environmental water

    samples.The method is simple and sample preparation is not needed. However

    care has to be taken that the calibration of the instrument and analytical

    measurements are performed quickly after each other.

    Ion Chromatography:

    The major application of this instrument is for inorganic anions in

    environmental analysis. This is a kind of ion-exchange chromatography used for

    the separation of inorganic and some organic cations and anions using a

    conductivity detector after suppressor column. The schematic diagram of an ion

    chromatograph is shown in fig 3a.

    7

  • 8/11/2019 2.2 Analysis of Common Ions

    9/17

    Na2SO4

    H2SO4

    H2SO4

    bath solution2Na+

    CO32-

    Eluate containing

    Na+,CO32-,HCO3-,+

    analyte anions

    SO42-

    2H+

    H2CO3

    Sulfonated

    polyethylene

    hollow fiber

    Eluate containing

    CO2,H2O,H++

    analyte anions

    Fig 3b Micro membrane suppression

    Eluent

    reservoir

    Pump

    Sample injection

    Separator column

    micro membrane

    suppressor

    column

    conductivity

    detector

    Fig 3a Schematic diagram of

    ion chromatograph

    8

  • 8/11/2019 2.2 Analysis of Common Ions

    10/17

    The stationary phase is a pellicular material, the particles consisting of an

    impervious central core surrounded by a thin outer layer (about 2m thick)

    incorporating cation-or anion-exchange sites. The thin layer contributes to faster

    rates of exchange that results in higher efficiencies. For the separation of anions,

    mobile phases containing solutions of electrolytes such as Na2CO3 or NaHCO3

    are used. For the separation of cations, HCl solution is used as mobile

    phase.The detection of low concentration of ionic solutes in the presence of high

    concentrations of eluting electrolyte is not possible.

    Unless the latter is removed. This is achieved by using a micro membrane

    suppressor column (fig 3b) immediately after the separator column, which

    converts the electrolyte into unionised water, leaving the solute ions as the only

    ionic species in the mobile phase.

    The reactions for the separation of inorganic anions on an-ion-exchange

    column in the form using sodium hydrogen carbonate as mobile phase

    are summarised in the following equation.

    -

    3HCO

    Separator Column:

    n n3 3 3 n

    n 2 33 4 4

    n (Res N R HCO ) x (Res N R ) x nHCO

    wherex F ,Cl ,NO ,SO ,PO etc.

    + +

    + +

    =

    i i 3

    .............................(1)

    Suppressor reactions:

    9

  • 8/11/2019 2.2 Analysis of Common Ions

    11/17

    The most wide spread method of achieving low back ground conductivity is

    by the use of "micromembrane suppressor" as illustrated in fig.3b. The effluent

    flows between two semi-permeable membranes which separate it from the

    counterflow of sulphuric acid. The migration of each ion is determined by the

    relative concentrations of the ions in the two liquids, the ion moving into the

    solution of lower concentration.

    For the analysis of anions in environmental water samples which are found at mg

    l-1 concentrations by this technique, the sample has to diluted before injection.

    This along with the filtration of the water sample is often the only sample

    preparation necessary for carryingout the analysis. It takes only a few minutes

    for the analysis of common anions that are present in water as shown in fig.4.

    10

  • 8/11/2019 2.2 Analysis of Common Ions

    12/17

    0 2

    NO2- 1.2 ppm

    NO3- 1.4 ppm

    SO42- 1.1 ppm

    Cl-

    Br- 1.4 ppm

    46 8 12 1410

    1.0 ppm

    Time (Min)

    Fig 4 Separation of ions in environmental water

    Although ion chromatography is quite commonly used for the analysis of

    anions in water, some common cations (Na+, K+, NH4+, Ca2+, Mg2+) also are

    being analysed using both the ion suppression system and also conventionalchromatographs with conductivity detection.

    Examples Of The Use Of Other Techniques:

    The most widely used methods covered so far are only for the analysis

    common ions. There are however, a few frequently used methods which have

    not been covered. For example the analysis of species such as ammonia,

    fluoride and sulphate, free chlorine, sulphide and cyanide which have not been

    discussed previously are described here.

    11

  • 8/11/2019 2.2 Analysis of Common Ions

    13/17

    Ammonia:

    In environmental water ammonia is the only alkaline gas commonly

    found. The ammonia could be determined by a simple acid-base titration

    provided it is isolated from solution. This may be done by addition of magnesium

    oxide to ensure that the sample is slightly alkaline. Ammonia is then distilled off

    into known excess acid. By tiltrating the excess acid with standard alkali, the

    ammonia concentration can be determined.

    Ion selective electrodes (i.e. an electrode whose potential measured with

    respect to a reference) respond in a linear manner to the logarithm of the activity

    of analyte over a four to six order of magnitude range of activity. Electrodes do

    not consume unknown samples, and they introduce negligible contamination.

    Response time is seconds or minutes. Since the electrodes respond only to the

    activityof the uncomplexed ions, the ligands must be absent or masked.

    In order to know the concentrations, an inert salt is often used to bring all

    the standards and samples to a high constant ionic strength. If the activity

    coefficients remain constant, the electrode potential gives concentrations directly.

    Commercial electrodes are available for the detection of anions (e.g. for halides,

    NO3-, CN- , SCN-, S2-); cations (e.g. for H+,Na+,K+,Ca2+,Cd2+,Cu2+,Pb2+) and

    gases (e.g. NH3,O2,CO2,NO2).

    In the ammonia selective electrode which is a gas sensing type, the

    ammonia diffuses through a permeable membrane and causes a pH change in a

    small volume of internal solution which is sensed by a glass electrode. Before

    taking the measurement, concentrated sodium hydroxide solution is added to

    samples and standards which serves to increase the pH to above 11 so that all

    ammonia is in the unprotonated form and provides a constant ionic strength.

    12

  • 8/11/2019 2.2 Analysis of Common Ions

    14/17

    Fluoride:

    Fluoride ion can be estimated, based on potentiometric measurements

    with a membrane electrode consisting of a single crystal of europium doped

    lanthanum fluoride, LaF3. The purpose of Eu doping is to improve electrical

    conductivity. The membrane is cut as a 1-mm thick disc, a few mm in diameter,

    the disc is sealed into the end of a rigid plastic tube, filled with an equimolar

    solution of KCl and NaF, into which dips an AgCl electrode. A reference

    electrode (saturated calomel electrode) is inserted into the test solution along

    with the fluoride electrode. The potential difference is measured. Fluoride

    concentration down to 10

    -6

    M can be measured. This electrode shows extremely

    high specificity to the analyte ion, the only pretreatment necessary being the

    addition of buffer solution to maintain constant pH and ionic strength.

    Alternatively fluoride can be determined by spectrophotometry or by ion-

    chromatography both of which have been discussed already. Fluoride reacts with

    zirconium-alizarin lake to form colourless ZrF62-and the dye. The colur of the dye

    lake becomes progressively weak with increase in amount of F-.

    To the sample solution is added drop of sodium arsenite solution to

    remove residual chlorine if any. Then enough of zirconyl-alizarin reagent is

    added.The solution is thoroughly mixed and absorbance measurement were

    done at 570 nm after keeping the samples and standards for 1 hr.

    Sulphate:

    Ion chromatography is the only instrumental method for the direct

    determination of sulphate. Sulphate may be precipitated either with Ba2+

    or

    2-aminoperimidinium salts. The precipitate may be weighed for a direct

    determination of the sulphate as a gravimetric method. Other methods using

    13

  • 8/11/2019 2.2 Analysis of Common Ions

    15/17

    insoluble salt precipitation are indirect, estimating the excess of the cation after

    precipitation of the sulphate. Excess barium may be determined by titration with

    EDTA or by atomic absorption.Excess 2-aminoperimidinium ions may be

    estimated by visible spectrophotometry.

    Free chlor ine (Residual chlor ine):

    The effectiveness of chlorination of raw water for public water supply

    can be checked by estimating free or residual Cl2in water samples. Free

    available Cl2consists of Cl2,HOCl, and HClO2. Free Cl2 reacts instantly with N,

    N-diethyl phenylene diamine (DPD) indicator to form a red colour. This colour is

    discharged by the addition of ammonium iron(II) sulphate solution.

    The water sample is treated with DPD reagent solution and titrated

    with standard ferrous ammonium sulphate solution until it becomes colourless.

    The amount of free chlorine in the water sample can be determined from the

    volume of ammonium iron (II) sulphate consumed.

    Sulphide:

    The major effluents containing sulfide in significant quantities are

    tannery wastes, sulphide dye liquors, oil refinery wastes, viscose rayon wastes,

    septic sewage and ammonical gas liquor.

    Two methods namely colorimetric and titrimetric methods are used.

    The colorimetric procedure is simple to carryout, rapid, sensitive and applicable

    to sulphide concentrations of 0.2 to 20 ppm. Titrimetric procedure, though is

    applicable to samples containing more than 1 ppm sulphide, it is generally used

    for samples of higher concentration and containing more interferences.

    14

  • 8/11/2019 2.2 Analysis of Common Ions

    16/17

    Sample Collection and preservation:

    Sample should be collected with least aeration as sulfide is

    volatilised and oxygen destroys sulphide. Samples have to be preserved by the

    addition of zinc acetate solution.the pH of the solution should be adjusted to > 9

    by the addition of sodium hydroxide. The sample should contain representative

    porportion of suspended solids.

    Spectrophotometric determination of H2S :

    In this method the absorbing solution used contains CdSO4 and

    NaOH. The precipitated cadmium sulphide on acidification releases S2- ions

    which interacts with FeCl3 and -diethylphenylene diamine to give a dye ethylene

    blue which can be measured at 670 nm. Beer's law is obeyed from 0.1 to 10 ppm

    S2-. Gases like SO2, O3, NH3do not interfere.

    Titrimetric method:

    This procedure measures total sulphides excepting acid insoluble

    metallic sulphides.

    To the Sulphide in acetate solution is added a known excess of iodine

    solution which is later acidified. The excess iodine is back titrated with standard

    thiosulphate solution using starch as indicator. Knowing the amount of the

    standardised iodine consumed, and the volume of sample taken, the amount of

    sulphide present in the water sample can be calculated.

    ..(4)2 2H S I 2H 2I S++ + +

    22 2 3 4 6I S O 2I S O

    2 + + .(5)

    1 ml 0.05N I2=0.85mg H2S

    15

  • 8/11/2019 2.2 Analysis of Common Ions

    17/17

    Cyanide:

    This method is applicable to all types of industrial effluents, domestic

    waste waters, polluted waters, raw and treated waters.

    The cyanide in the sample is distilled and absorbed in alkali. CN-in the

    alkaline distillate on treatment with bromine water is converted to cyanogen

    bromide (CNBr) which reacts with p-phenylenediamine reagent to form a red dye.

    The absorbance of the dye is measured at 530nm and the concentration of the

    cyanide can be read from a calibration graph.

    Most of the simple cyanides M(CN)x are readily converted to HCN by

    acid distillation. The complex cyanides such as alkali ferri and ferrocyanides are

    not converted to HCN during distillation. Acid cuprous chloride and /or

    magnesium chloride are added to the sample which converts the complex

    cyanide to simple cyanides which are then converted to HCN by distillation.

    16