2134-2discussion

  • Upload
    meowy

  • View
    218

  • Download
    0

Embed Size (px)

Citation preview

  • 7/23/2019 2134-2discussion

    1/11

    Comment on the relative advantages and disadvantages of

    Venturimeter, orifice plate meter and rotameter as flow measuring devices.

    From table 3, it is observed that loss factor Kv is the smallest.

    This suggests that the venturimeter may give more accurate measurement as

    compared to that of orifice plate.

    its experimental flow rates that are quite close to the theoretical value (d

    = 1.0178 1).

    This implies that it is also able to measure a greater range of values of flow rate

    as the will be less energy lost at high flow velocities.

    The meter has a smooth internal surface that helps to reduce friction as the water

    flows through it.

    !oreover, the gradual reduction and expansion in the diameter of the tube

    reduces this loss.

    "owever, to cater for the gradual reduction and expansion, a longer tube is

    required and the #enturimeter occupies quite a lot of space($%& ratio ofapproximate ') and is an expensive device as it is more difficult to construct

    compared to the other device.

    *t also cannot be altered for measuring pressure beyond a maximum velocity.

    The orifice plate meter is fairly accurate in its measurement

    *t is also the easiest to construct.

    +rifice meter occupies less space thus it could be useful to measure fluid flows within space

    constraints

    The orifice plate meter has a significant head loss.

  • 7/23/2019 2134-2discussion

    2/11

    *ts level of accuracy falls as the flow rate rises. The sudden contraction of the

    diameter causes high loss factor due to turbulent flow.

    tube with diameter larger then the orifice meter plate is required to reduce this

    energy loss.

    Furthermore, as the diameter of the tube is reduced, the angle between the surface of t

    he tube with the hori-ontal should not be too large, so as to reduce energy loss and improve

    accuracy, but that would demand even more hori-ontal space.

  • 7/23/2019 2134-2discussion

    3/11

    Comment on the head losses associated with all the flow meters

    studied in this experiment, emphasising the relationship between the

    mechanism of loss generation and its magnitude

    For the venturimeter, its long section enables gradual constriction and expansion

    of diameter.

    The head loss is thus caused by the dissipation of energy via the friction betweenthe fluid and the inner surface of the pipe.

    *t has been observed that the head loss in the venturimeter is relatively

    small relative to the orifice plate.

    *n the #enturi meter, the gradual reduction and expansion of the diameter

    reduces the separation of flow as well as reduces the separation in the

    deceleration portion of the meter.

    "ence, the energy loss is mainly due to friction with the wall of the tube rather to

    inefficient mixing and separate flow

    The orifice plate has a significant head loss when compared to the other parts of

    the apparatus. #+is observed to be very high too.

    s fluid velocity across the plate is relatively high, the sudden contraction

    and expansion of the diameter before and after the orifice place may result in a

    sharp and significant change in momentum about the orifice, causing

    turbulence to form on both ends.

    The turbulence increases as the flow increases.

    onsequently, this turbulence dissipates energy, thus resulting in the head loss

  • 7/23/2019 2134-2discussion

    4/11

    "ence the energy loss may be due to the high energy flow across the small

    orifice plate, resulting in a significant loss in energy.

    ith the orifice plate experiencing as much as ' times more head loss

    it can be implied that turbulence contributes to significantly higher energy loss than friction

    with the internal walls.

    Explain with the aid of simple sketches what is the

    vena contractor of an

    orifice meter.

    #ena contracta is the point in a fluid stream where the

    diameter of the stream isthe least. The contraction ta/es place at a section slightl

    y downstream from the

    orifice, where the fluid flow is hori-ontal. 0treaml

    ines will converge 1ust

    downstream of the diameter change, and a region of sep

    arated flow occurs from

    the sharp corner of the diameter change and extends past

    the vena contracta.

    #ena contracta is the narrowest central flow region

    of a 1et that occurs 1ust downstream to the

    orifice plate. *t is characteri-ed by high velocity

    , laminar flow. $aminar flow, sometimes

    /nown as streamline flow, occurs when a fluid flows

    in parallel layers, with no disruption

    between the layers.

    The vena contracta refers to the point in the fluid stre

    am where the diameter of

    the streamlines is the smallest, and it occurs 1ust slightly d

    ownstream of the

    orifice, where the flow is nearly hori-ontal and is

    concentric with the orifice and

    flow channel. fter the vena contracta, the streamlines

    diverge and a region of

    separated flow occurs and extends past it.

  • 7/23/2019 2134-2discussion

    5/11

    cross section of a circular orifice of diameter &o

    is shown. The thic/ness of the wall is

    assumed small compared to the diameter of the orifice.

    &ue to the convergence of the

    streamlines approaching the orifice, the cross section o

    f the 1et decreases slightly until the

    pressure is equali-ed over the cross2section, and the ve

    locity profile is nearly rectangular.

    This point of minimum area is called the

    vena contracta

    . eyond the vena contracta, friction

    with the fluid outside the 1et (air) slows it down, an

    d the cross section increases perforce.

    This divergence is usually quite small, and the 1et is ne

    arly cylindrical with a constant

    45

    velocity. The 1et is held together by surface tension, o

    f course, which has a stronger effect the

    smaller the diameter of the 1et

    . omment on the limitations and ma1or sources of err

    or in this experiment

    $imitations of the experiment6

    2 The experiment enables only a small range of flow measured. The

  • 7/23/2019 2134-2discussion

    6/11

    apparatus does not allow measurement over larger range.

    2 This experiment is conducted using water. pparatus w

    ill not wor/ for fluid

    with higher viscosity as very large energy will be requi

    red.2 The weighing tan/ is of limited si-e, not large enou

    gh to measure

    rotameter reading of value lesser than 7. The water i

    n tan/ will overflow.

    !a1or sources of errors6

    8

    The reading of manometer. The readings may not be acc

    urate as it tends

    to 1ump, even after allowing some time before the reading is ta/en. Thus

    reading ta/en might not be the actual result.

    8

    The manometer is not totally transparent, due to prese

    nce of water vapour

    and bubbles in the tube, thus readings may not be accurat

    e.

    8

    !easuring of a few meters using the same apparatus mayresult in more

    error in the measurements and calculation.

    8

    9sage of stopwatch and weight to measure the mass flow

    rate may not be

    accurate due to human errors.

    8

    :arallax error reading from the apparatus may occur du

    e to the colour ofthe fluid used.

    onclusion

    From the experiment, we are able to recogni-e the wor

    /ing mechanism of the

    venturimeter, orifice meter and rotameter used to measu

    re flow.

    e have also compared the advantages and disadvantages of

    the three meters,

    as well as calculated the head loss from each type of meter. From these, we are

  • 7/23/2019 2134-2discussion

    7/11

    able to better understand the principles behind the w

    or/ing mechanism of each

    meter, the efficiency of each meter and thus determine

    wisely which meter is

    better suited for various applications.

    *n this experiement, the 3 devices 2 the venture me

    ter, orifice meter and the rotameter has been close

    ly

    examined.

    lso, the coefficient of discharge , d

    , for the

    #

    enturi meter and orifice meter is determined; where

    d

    for

    the

    #

    enturi meter is .4, while ?d

    for the orifice meter is .3

  • 7/23/2019 2134-2discussion

    8/11

    measurement over a larger range of flow rates.

    2 This experiment cannot be conducted with fluids of gre

    ater viscosity as a

    lot of energy will be required to pump the fluid aro

    und the experiment setup. lso corrosive fluids or fluids at higher temperatur

    es cannot be

    experimented with as some of the connections and tubing

    s are less

    resistant to corrosion and% or high temperatures.

    2 The weighing tan/ is of limited si-e, and is not suitab

    le to be used to time

    greater mass flow rates of fluids and inaccuracies will ari

    se if it fills up toofast. lso, the water in tan/ will overflow easily if the

    flow is too high.

    2 lso, the experiment may not be suitable for dar/ colo

    ured or opaque

    fluids as the rotameter float has to be visible to ta/e

    that reading.

    !a1or sources of errors6

    8

    The pressure of the pump is not constant and fluctuates quite a bit.

    This will result in estimation errors while trying to r

    ead the measurement

    values off the instruments. The readings will fluctuate,

    even after allowing

    the set up some time to settle. This problem may be solved

    by using a

    stabiliser for the pump pressure or a & pump.

    8 ondensation and dirt build up occurs inside the manome

    ter tubes,

    ma/ing them not totally transparent. This may cause inaccura

    cies in the

    readings, even if efforts are ta/en to reduce parallax e

    rrors. onducting

    the experiment in a dryer environment may reduce the r

    is/ of

    condensation.8

  • 7/23/2019 2134-2discussion

    9/11

    9sing a hand stopwatch and manually loading weights to

    measure the

    mass flow rate may not be accurate due to human errors an

    d bouncing of

    the lever armature. *t may be better to use an electronic timer attached to

    the contact point between the arm and the frame and u

    sing mechanically

    loaded weights or a calibrated spring to counter the

    weight of the water.

    *mplementing this will ma/e the experiment more accura

    te but will also

    increase the overall cost.

    8 :arallax error reading from the apparatus may occur as

    the manometer

    diameter is small. lso, at the thic/ness, surface tension

    effects may

    become significant, resulting in inaccurate readings.

    8

    The tubings and connections may not be totally airtight

    or watertight and

    any lea/age of either fluid will result in inaccuracies in the reading ta/en.

    onclusion

    From this experiment, we have become more familiar wit

    h the wor/ing

    mechanisms of the various types of flow measuring devices such

    as the #enturi

    meter, orifice meter and rotameter. e have also dete

    rmined the coefficient of

    discharge of the #enturi meter and calibrated the rotameter.

    e have also compared the advantages and disadvantages of

    the three meters,

    and calculated the head loss from each measuring apparat

    us. From these, we

    are able to better understand the principles behind t

    he wor/ings of each meter,

    and determine which meter is better suited for various a

    pplications. Thus, * feelthat this experiment has been a satisfactory one.

  • 7/23/2019 2134-2discussion

    10/11

    )

    "uman reaction time and parallax is one possible error.

    b)

    ater is used in this experiment. ny other substances th

    at have higher viscosity may not

    be used. There still exist some air bubbles inside the tub

    e that could have led to

    inaccurate readings and affect subsequent calculations.

    c)

    The flow rate in the experiment is control by the pum

    p which may not pump water at a

    constant rate

    .

    d)

    Friction at the 1oints of the lever system may cause

    inaccuracies when weighing the

    water.

    e)

    The readings on the manometer are always fluctuating and

    consistent results cannot be

    obtained.

    +E$90*+E 6The usage of the venture meter, orifice meter and the r

    otameter has been examined. From the

    three, the venture meter gives us the most accurate re

    ading and the rotameter involves the

    most heat loss. e can also see that the main reaso

    n for any energy loss is due to turbulent

    flow, friction and drag loss. *n order to obtain an acc

    urate reading, the pressure and velocitydifference had to be significant across the meters.

    oefficient of discharge for venturimeter and orifice

    !eter,

    d

    A 4.4=>

    A .'

  • 7/23/2019 2134-2discussion

    11/11

    @

    A (.4>