2134-2discussion (1)

  • Upload
    meowy

  • View
    213

  • Download
    0

Embed Size (px)

Citation preview

  • 7/23/2019 2134-2discussion (1)

    1/10

    Comment on the relative advantages and disadvantages of

    Venturimeter, orifice plate meter and rotameter as flow measuring devices.

    The orifice plate meter is fairly accurate in its measurement

    It is also the easiest to construct.

    Orifice meter occupies less space thus it could be useful to measure fluid flows within space

    constraints

    The orifice plate meter has a significant head loss.

    Its level of accuracy falls as the flow rate rises. The sudden contraction of the

    diameter causes high loss factor due to turbulent flow.

    A tube with diameter larger then the orifice meter plate is required to reduce this

    energy loss.

    Furthermore, as the diameter of the tube is reduced, the angle between the surface of t

    he tube with the horizontal should not be too large, so as to reduce energy loss and improve

    accuracy, but that would demand even more horizontal space.

  • 7/23/2019 2134-2discussion (1)

    2/10

    Comment on the head losses associated with all the flow metersstudied in this experiment, emphasising the relationship between the

    mechanism of loss generation and its magnitude

    For the venturimeter, its long section enables gradual constriction and epansion

    of diameter.

    The head loss is thus caused by the dissipation of energy via the friction between

    the fluid and the inner surface of the pipe.

    It has been observed that the head loss in the venturimeter is relatively

    small relative to the orifice plate.

    In the !enturi meter, the gradual reduction and epansion of the diameter

    reduces the separation of flow as well as reduces the separation in the

    deceleration portion of the meter.

    "ence, the energy loss is mainly due to friction with the wall of the tube rather toinefficient miing and separate flow

  • 7/23/2019 2134-2discussion (1)

    3/10

    The orifice plate has a significant head loss when compared to the other parts of

    the apparatus. !Ois observed to be very high too.

    As fluid velocity across the plate is relatively high, the sudden contraction

    and epansion of the diameter before and after the orifice place may result in a

    sharp and significant change in momentum about the orifice, causing

    turbulence to form on both ends.

    The turbulence increases as the flow increases.

    #onsequently, this turbulence dissipates energy, thus resulting in the head loss

    "ence the energy loss may be due to the high energy flow across the small

    orifice plate, resulting in a significant loss in energy.

    $ith the orifice plate eperiencing as much as % times more head loss

    it can be implied that turbulence contributes to significantly higher energy loss than friction

    with the internal walls.

    Explain with the aid of simple sketches what is the

    vena contractor of an

    orifice meter.

    !ena contracta is the point in a fluid stream where the

    diameter of the stream is

    the least. The contraction ta&es place at a section slightl

    y downstream from theorifice, where the fluid flow is horizontal. 'treaml

    ines will converge (ust

    downstream of the diameter change, and a region of sep

    arated flow occurs from

    the sharp corner of the diameter change and etends past

    the vena contracta.

    !ena contracta is the narrowest central flow region

    of a (et that occurs (ust downstream to the

  • 7/23/2019 2134-2discussion (1)

    4/10

    orifice plate. It is characterized by high velocity

    , laminar flow. )aminar flow, sometimes

    &nown as streamline flow, occurs when a fluid flows

    in parallel layers, with no disruption

    between the layers.

    The vena contracta refers to the point in the fluid stre

    am where the diameter of

    the streamlines is the smallest, and it occurs (ust slightly d

    ownstream of the

    orifice, where the flow is nearly horizontal and is

    concentric with the orifice and

    flow channel. After the vena contracta, the streamlines

    diverge and a region of

    separated flow occurs and etends past it.

    A cross section of a circular orifice of diameter *o

    is shown. The thic&ness of the wall is

    assumed small compared to the diameter of the orifice.

    *ue to the convergence of the

    streamlines approaching the orifice, the cross section o

    f the (et decreases slightly until thepressure is equalized over the cross+section, and the ve

    locity profile is nearly rectangular.

    This point of minimum area is called the

    vena contracta

    . eyond the vena contracta, friction

    with the fluid outside the (et -air slows it down, an

    d the cross section increases perforce.

    This divergence is usually quite small, and the (et is ne

    arly cylindrical with a constant/0

    velocity. The (et is held together by surface tension, o

    f course, which has a stronger effect the

    smaller the diameter of the (et

  • 7/23/2019 2134-2discussion (1)

    5/10

    . #omment on the limitations and ma(or sources of err

    or in this eperiment

    )imitations of the eperiment1

    + The eperiment enables only a small range of flow m

    easured. The

    apparatus does not allow measurement over larger range.

    + This eperiment is conducted using water. Apparatus w

    ill not wor& for fluid

    with higher viscosity as very large energy will be requi

    red.

    + The weighing tan& is of limited size, not large enou

    gh to measure

    rotameter reading of value lesser than 23. The water i

    n tan& will overflow.

    4a(or sources of errors1

    5

    The reading of manometer. The readings may not be acc

    urate as it tends

    to (ump, even after allowing some time before the rea

    ding is ta&en. Thus

    reading ta&en might not be the actual result.

    5

    The manometer is not totally transparent, due to prese

    nce of water vapour

    and bubbles in the tube, thus readings may not be accurat

    e.

    5

    4easuring of a few meters using the same apparatus may

    result in more

    error in the measurements and calculation.

    5

  • 7/23/2019 2134-2discussion (1)

    6/10

    6sage of stopwatch and weight to measure the mass flow

    rate may not be

    accurate due to human errors.

    5

    7aralla error reading from the apparatus may occur due to the colour of

    the fluid used.

    #onclusion

    From the eperiment, we are able to recognize the wor

    &ing mechanism of the

    venturimeter, orifice meter and rotameter used to measu

    re flow.

    $e have also compared the advantages and disadvantages of

    the three meters,as well as calculated the head loss from each type of meter

    . From these, we are

    able to better understand the principles behind the w

    or&ing mechanism of each

    meter, the efficiency of each meter and thus determine

    wisely which meter is

    better suited for various applications.

    In this eperiement, the 8 devices + the venture me

    ter, orifice meter and the rotameter has been close

    ly

    eamined.

    A

    lso, the coefficient of discharge , #d

    , for the!

    enturi meter and orifice meter is determined9 where

    #d

    for

    the

    !

    enturi meter is 3.:;

  • 7/23/2019 2134-2discussion (1)

    7/10

    >A

    ? -3.3/::@ 3.88// /3%,

    Futhermore, the energy losses in the

    !

    enturi meter, orifice meter, rotameter as well as t

    he wide angle

    diffuser and a :3BFelbow is determined.

    A

    ll in all, the ob(ectives of this eperiment are ac

    hieved

    )imitations of the eperiment1

    + The eperiment enables only a small range of flow m

    easured as thepump=s power is limited and some of the apparatus do no

    t allow

    measurement over a larger range of flow rates.

    + This eperiment cannot be conducted with fluids of gre

    ater viscosity as a

    lot of energy will be required to pump the fluid aro

    und the eperiment set

    up. Also corrosive fluids or fluids at higher temperatur

    es cannot beeperimented with as some of the connections and tubing

    s are less

    resistant to corrosion andC or high temperatures.

    + The weighing tan& is of limited size, and is not suitab

    le to be used to time

    greater mass flow rates of fluids and inaccuracies will ari

    se if it fills up too

    fast. Also, the water in tan& will overflow easily if the

    flow is too high.

    + Also, the eperiment may not be suitable for dar& colo

    ured or opaque

    fluids as the rotameter float has to be visible to ta&e

    that reading.

    4a(or sources of errors1

    5

    The pressure of the A# pump is not constant and fluctuat

    es quite a bit.

    This will result in estimation errors while trying to r

    ead the measurement

  • 7/23/2019 2134-2discussion (1)

    8/10

    values off the instruments. The readings will fluctuate,

    even after allowing

    the set up some time to settle. This problem may be solved

    by using a

    stabiliser for the pump pressure or a *# pump.5

    #ondensation and dirt build up occurs inside the manome

    ter tubes,

    ma&ing them not totally transparent. This may cause inaccura

    cies in the

    readings, even if efforts are ta&en to reduce paralla e

    rrors. #onducting

    the eperiment in a dryer environment may reduce the r

    is& ofcondensation.

    5

    6sing a hand stopwatch and manually loading weights to

    measure the

    mass flow rate may not be accurate due to human errors an

    d bouncing of

    the lever armature. It may be better to use an electron

    ic timer attached to

    the contact point between the arm and the frame and using mechanically

    loaded weights or a calibrated spring to counter the

    weight of the water.

    Implementing this will ma&e the eperiment more accura

    te but will also

    increase the overall cost.

    5

    7aralla error reading from the apparatus may occur as

    the manometerdiameter is small. Also, at the thic&ness, surface tension

    effects may

    become significant, resulting in inaccurate readings.

    5

    The tubings and connections may not be totally airtight

    or watertight and

    any lea&age of either fluid will result in inaccuracies i

    n the reading ta&en.

    #onclusionFrom this eperiment, we have become more familiar wit

  • 7/23/2019 2134-2discussion (1)

    9/10

    h the wor&ing

    mechanisms of the various types of flow measuring devices such

    as the !enturi

    meter, orifice meter and rotameter. $e have also dete

    rmined the coefficient ofdischarge of the !enturi meter and calibrated the rota

    meter.

    $e have also compared the advantages and disadvantages of

    the three meters,

    and calculated the head loss from each measuring apparat

    us. From these, we

    are able to better understand the principles behind t

    he wor&ings of each meter,

    and determine which meter is better suited for various applications. Thus, I feel

    that this eperiment has been a satisfactory one.

    "uman reaction time and paralla is one possible error.

    b

    $ater is used in this eperiment. Any other substances th

    at have higher viscosity may not

    be used. There still eist some air bubbles inside the tub

    e that could have led to

    inaccurate readings and affect subsequent calculations.

    c

    The flow rate in the eperiment is control by the pum

    p which may not pump water at a

    constant rate

    .

    d

    Friction at the (oints of the lever system may cause

    inaccuracies when weighing the

    water.

    e

    The readings on the manometer are always fluctuating and

    consistent results cannot be

    obtained.

    #OD#)6'IOD 1

    The usage of the venture meter, orifice meter and the rotameter has been eamined. From the

  • 7/23/2019 2134-2discussion (1)

    10/10

    three, the venture meter gives us the most accurate re

    ading and the rotameter involves the

    most heat loss. $e can also see that the main reaso

    n for any energy loss is due to turbulent

    flow, friction and drag loss. In order to obtain an accurate reading, the pressure and velocity

    difference had to be significant across the meters.

    #oefficient of discharge for venturimeter and orifice

    4eter,

    #d

    ? /.3/;A

    against Eotameter reading, we obtain the calibration curv

    e equation,

    >A

    ? -3.3/