Upload
meowy
View
213
Download
0
Embed Size (px)
Citation preview
7/23/2019 2134-2discussion (1)
1/10
Comment on the relative advantages and disadvantages of
Venturimeter, orifice plate meter and rotameter as flow measuring devices.
The orifice plate meter is fairly accurate in its measurement
It is also the easiest to construct.
Orifice meter occupies less space thus it could be useful to measure fluid flows within space
constraints
The orifice plate meter has a significant head loss.
Its level of accuracy falls as the flow rate rises. The sudden contraction of the
diameter causes high loss factor due to turbulent flow.
A tube with diameter larger then the orifice meter plate is required to reduce this
energy loss.
Furthermore, as the diameter of the tube is reduced, the angle between the surface of t
he tube with the horizontal should not be too large, so as to reduce energy loss and improve
accuracy, but that would demand even more horizontal space.
7/23/2019 2134-2discussion (1)
2/10
Comment on the head losses associated with all the flow metersstudied in this experiment, emphasising the relationship between the
mechanism of loss generation and its magnitude
For the venturimeter, its long section enables gradual constriction and epansion
of diameter.
The head loss is thus caused by the dissipation of energy via the friction between
the fluid and the inner surface of the pipe.
It has been observed that the head loss in the venturimeter is relatively
small relative to the orifice plate.
In the !enturi meter, the gradual reduction and epansion of the diameter
reduces the separation of flow as well as reduces the separation in the
deceleration portion of the meter.
"ence, the energy loss is mainly due to friction with the wall of the tube rather toinefficient miing and separate flow
7/23/2019 2134-2discussion (1)
3/10
The orifice plate has a significant head loss when compared to the other parts of
the apparatus. !Ois observed to be very high too.
As fluid velocity across the plate is relatively high, the sudden contraction
and epansion of the diameter before and after the orifice place may result in a
sharp and significant change in momentum about the orifice, causing
turbulence to form on both ends.
The turbulence increases as the flow increases.
#onsequently, this turbulence dissipates energy, thus resulting in the head loss
"ence the energy loss may be due to the high energy flow across the small
orifice plate, resulting in a significant loss in energy.
$ith the orifice plate eperiencing as much as % times more head loss
it can be implied that turbulence contributes to significantly higher energy loss than friction
with the internal walls.
Explain with the aid of simple sketches what is the
vena contractor of an
orifice meter.
!ena contracta is the point in a fluid stream where the
diameter of the stream is
the least. The contraction ta&es place at a section slightl
y downstream from theorifice, where the fluid flow is horizontal. 'treaml
ines will converge (ust
downstream of the diameter change, and a region of sep
arated flow occurs from
the sharp corner of the diameter change and etends past
the vena contracta.
!ena contracta is the narrowest central flow region
of a (et that occurs (ust downstream to the
7/23/2019 2134-2discussion (1)
4/10
orifice plate. It is characterized by high velocity
, laminar flow. )aminar flow, sometimes
&nown as streamline flow, occurs when a fluid flows
in parallel layers, with no disruption
between the layers.
The vena contracta refers to the point in the fluid stre
am where the diameter of
the streamlines is the smallest, and it occurs (ust slightly d
ownstream of the
orifice, where the flow is nearly horizontal and is
concentric with the orifice and
flow channel. After the vena contracta, the streamlines
diverge and a region of
separated flow occurs and etends past it.
A cross section of a circular orifice of diameter *o
is shown. The thic&ness of the wall is
assumed small compared to the diameter of the orifice.
*ue to the convergence of the
streamlines approaching the orifice, the cross section o
f the (et decreases slightly until thepressure is equalized over the cross+section, and the ve
locity profile is nearly rectangular.
This point of minimum area is called the
vena contracta
. eyond the vena contracta, friction
with the fluid outside the (et -air slows it down, an
d the cross section increases perforce.
This divergence is usually quite small, and the (et is ne
arly cylindrical with a constant/0
velocity. The (et is held together by surface tension, o
f course, which has a stronger effect the
smaller the diameter of the (et
7/23/2019 2134-2discussion (1)
5/10
. #omment on the limitations and ma(or sources of err
or in this eperiment
)imitations of the eperiment1
+ The eperiment enables only a small range of flow m
easured. The
apparatus does not allow measurement over larger range.
+ This eperiment is conducted using water. Apparatus w
ill not wor& for fluid
with higher viscosity as very large energy will be requi
red.
+ The weighing tan& is of limited size, not large enou
gh to measure
rotameter reading of value lesser than 23. The water i
n tan& will overflow.
4a(or sources of errors1
5
The reading of manometer. The readings may not be acc
urate as it tends
to (ump, even after allowing some time before the rea
ding is ta&en. Thus
reading ta&en might not be the actual result.
5
The manometer is not totally transparent, due to prese
nce of water vapour
and bubbles in the tube, thus readings may not be accurat
e.
5
4easuring of a few meters using the same apparatus may
result in more
error in the measurements and calculation.
5
7/23/2019 2134-2discussion (1)
6/10
6sage of stopwatch and weight to measure the mass flow
rate may not be
accurate due to human errors.
5
7aralla error reading from the apparatus may occur due to the colour of
the fluid used.
#onclusion
From the eperiment, we are able to recognize the wor
&ing mechanism of the
venturimeter, orifice meter and rotameter used to measu
re flow.
$e have also compared the advantages and disadvantages of
the three meters,as well as calculated the head loss from each type of meter
. From these, we are
able to better understand the principles behind the w
or&ing mechanism of each
meter, the efficiency of each meter and thus determine
wisely which meter is
better suited for various applications.
In this eperiement, the 8 devices + the venture me
ter, orifice meter and the rotameter has been close
ly
eamined.
A
lso, the coefficient of discharge , #d
, for the!
enturi meter and orifice meter is determined9 where
#d
for
the
!
enturi meter is 3.:;
7/23/2019 2134-2discussion (1)
7/10
>A
? -3.3/::@ 3.88// /3%,
Futhermore, the energy losses in the
!
enturi meter, orifice meter, rotameter as well as t
he wide angle
diffuser and a :3BFelbow is determined.
A
ll in all, the ob(ectives of this eperiment are ac
hieved
)imitations of the eperiment1
+ The eperiment enables only a small range of flow m
easured as thepump=s power is limited and some of the apparatus do no
t allow
measurement over a larger range of flow rates.
+ This eperiment cannot be conducted with fluids of gre
ater viscosity as a
lot of energy will be required to pump the fluid aro
und the eperiment set
up. Also corrosive fluids or fluids at higher temperatur
es cannot beeperimented with as some of the connections and tubing
s are less
resistant to corrosion andC or high temperatures.
+ The weighing tan& is of limited size, and is not suitab
le to be used to time
greater mass flow rates of fluids and inaccuracies will ari
se if it fills up too
fast. Also, the water in tan& will overflow easily if the
flow is too high.
+ Also, the eperiment may not be suitable for dar& colo
ured or opaque
fluids as the rotameter float has to be visible to ta&e
that reading.
4a(or sources of errors1
5
The pressure of the A# pump is not constant and fluctuat
es quite a bit.
This will result in estimation errors while trying to r
ead the measurement
7/23/2019 2134-2discussion (1)
8/10
values off the instruments. The readings will fluctuate,
even after allowing
the set up some time to settle. This problem may be solved
by using a
stabiliser for the pump pressure or a *# pump.5
#ondensation and dirt build up occurs inside the manome
ter tubes,
ma&ing them not totally transparent. This may cause inaccura
cies in the
readings, even if efforts are ta&en to reduce paralla e
rrors. #onducting
the eperiment in a dryer environment may reduce the r
is& ofcondensation.
5
6sing a hand stopwatch and manually loading weights to
measure the
mass flow rate may not be accurate due to human errors an
d bouncing of
the lever armature. It may be better to use an electron
ic timer attached to
the contact point between the arm and the frame and using mechanically
loaded weights or a calibrated spring to counter the
weight of the water.
Implementing this will ma&e the eperiment more accura
te but will also
increase the overall cost.
5
7aralla error reading from the apparatus may occur as
the manometerdiameter is small. Also, at the thic&ness, surface tension
effects may
become significant, resulting in inaccurate readings.
5
The tubings and connections may not be totally airtight
or watertight and
any lea&age of either fluid will result in inaccuracies i
n the reading ta&en.
#onclusionFrom this eperiment, we have become more familiar wit
7/23/2019 2134-2discussion (1)
9/10
h the wor&ing
mechanisms of the various types of flow measuring devices such
as the !enturi
meter, orifice meter and rotameter. $e have also dete
rmined the coefficient ofdischarge of the !enturi meter and calibrated the rota
meter.
$e have also compared the advantages and disadvantages of
the three meters,
and calculated the head loss from each measuring apparat
us. From these, we
are able to better understand the principles behind t
he wor&ings of each meter,
and determine which meter is better suited for various applications. Thus, I feel
that this eperiment has been a satisfactory one.
"uman reaction time and paralla is one possible error.
b
$ater is used in this eperiment. Any other substances th
at have higher viscosity may not
be used. There still eist some air bubbles inside the tub
e that could have led to
inaccurate readings and affect subsequent calculations.
c
The flow rate in the eperiment is control by the pum
p which may not pump water at a
constant rate
.
d
Friction at the (oints of the lever system may cause
inaccuracies when weighing the
water.
e
The readings on the manometer are always fluctuating and
consistent results cannot be
obtained.
#OD#)6'IOD 1
The usage of the venture meter, orifice meter and the rotameter has been eamined. From the
7/23/2019 2134-2discussion (1)
10/10
three, the venture meter gives us the most accurate re
ading and the rotameter involves the
most heat loss. $e can also see that the main reaso
n for any energy loss is due to turbulent
flow, friction and drag loss. In order to obtain an accurate reading, the pressure and velocity
difference had to be significant across the meters.
#oefficient of discharge for venturimeter and orifice
4eter,
#d
? /.3/;A
against Eotameter reading, we obtain the calibration curv
e equation,
>A
? -3.3/