97050811 Rev. 2 28.04.2017 ATTACHMENT to the USER MANUAL NewTom 5G XL –Dose declaration and ascceptance test EN

estceflasupport.com/data/media/product_documentation/newtom/... · 2019. 8. 23. · 1 IEC Acceptance Test The description of the software utilities for execution of the acceptance

  • Upload
    others

  • View
    0

  • Download
    0

Embed Size (px)

Citation preview

Page 1: estceflasupport.com/data/media/product_documentation/newtom/... · 2019. 8. 23. · 1 IEC Acceptance Test The description of the software utilities for execution of the acceptance

97050811

Rev. 2

28.04.2017

ATTACHMENT to the USER MANUAL

New

Tom

5G

XL –

Dose d

ecla

ration a

nd a

scce

pta

nce t

est

EN

Page 2: estceflasupport.com/data/media/product_documentation/newtom/... · 2019. 8. 23. · 1 IEC Acceptance Test The description of the software utilities for execution of the acceptance

NOTES

This document is provided as a consultation manual intended for the device users.

CEFLA s.c. follows a policy based on the constant development and update of the product. For this reason, it

reserves the right to change the content of this manual without prior notice.

This document can not be modified, copied, reproduced, distributed, saved on magnetic or optical supports, or

published on websites and other on-line services, in full or in part, without the prior written authorisation of CEFLA

s.c.

The original version of this manual is in Italian.

NEWTOM™ 5G is a trademark of CEFLA s.c.

All other products and trade names mentioned in this document are registered marks of the relevant

manufacturers.

INFORMATIVE NOTE OF THE MANUFACTURER ON THE MEDICAL DEVICES

The medical device referred to in this manual consists of a scanner and a control, display and calculation

unit (Main Workstation). Such device, as delivered and configured by the production and assistance

technical personnel, is an X-ray device compliant with the safety requirements set forth by the Italian

Legislative Decree of 19 September 1994, no. 626 implementing Directives 89/391/EEC, 89/654/EEC,

89/655/EEC, 89/656/EEC, 90/269/EEC, 90/270/EEC, 90/394/EEC and 90/679/EEC concerning the

improvement of the health and safety of workers in the workplace, and with the essential requirements set

forth by the Italian Legislative Decree 24 February 1997, no. 46 implementing Directive 93/42/EEC as

amended, on the medical devices.

The medical device referred to in this manual is an X-ray device compliant with Directive 2011/65/EU on the

restriction of the use of certain hazardous substances in electrical and electronic equipment.

Any tampering with, modification, updating or other change both of hardware1 and software2 of the device as

supplied and installed by the company (and in the conditions specified in the attached documentation) may

partially or totally compromise the device expected operation. This may also alter the safety features with

consequent hazard increase for patients, operators and surrounding environment.

For this reason, should the user need to modify the device, he/she must request a written authorisation by

CEFLA s.c.

Failure to comply with what is specified in this informative note will null and void the device warranty and the

civil and/or penal responsibility for any consequent damage and/or accident and/or worsening of the patient,

operator or other people health (including the surrounding environment) will be borne by the person who

tampered with the device or his/her legal representative.

1 Adding of a new memory expansion, a new hardware on the connection bus, a printer, the replacement of the graphic display interface represents an important modification.

2 Including the operative system and the applications already installed upon medical device delivery.

Automatic updates of the operative system, changes to network connection parameters, modification and/or addition and/or removal of interface software with hardware (device driver) and/or services (e.g. file and printer sharing service) and/or applications represent an important modification.

Page 3: estceflasupport.com/data/media/product_documentation/newtom/... · 2019. 8. 23. · 1 IEC Acceptance Test The description of the software utilities for execution of the acceptance

SUMMARY

1 IEC Acceptance Test .......................................................................................................................................... 1

1.1 Positioning of the patient support [paragraph 5.1] .............................................................................................. 1 1.2 Patient positioning accuracy [paragraph 5.2] ...................................................................................................... 1

1.2.1 Axial patient positioning accuracy [paragraph 5.2.1] ................................................................................... 1 1.2.2 Sagittal and coronal patient positioning light [paragraph 5.2.2] ................................................................. 1

1.3 Tomographic section thickness [paragraph 5.3] ................................................................................................... 1 1.3.1 Tomographic section thickness for axial scanning [paragraph 5.3.1] ........................................................... 1 1.3.2 Tomographic section thickness for helical scanning [paragraph 5.3.2] ........................................................ 1

1.4 Dose [paragraph 5.4] ............................................................................................................................................. 2 1.5 Noise, mean CT number and uniformity [paragraph 5.5] ..................................................................................... 2 1.6 Spatial resolution [paragraph 5.6] ........................................................................................................................ 2 1.7 IEC 61223-3-5: reference values ........................................................................................................................... 3

Page 4: estceflasupport.com/data/media/product_documentation/newtom/... · 2019. 8. 23. · 1 IEC Acceptance Test The description of the software utilities for execution of the acceptance

EN DOSE DECLARATION AND ACCEPTANCE TEST 1

1 IEC Acceptance Test The description of the software utilities for execution of the acceptance tests is given in the "NNT User Manual" document. Below are the specific instructions for conducting these tests on the device (as required by IEC 61223-3-4, IEC 61223-3-5, DIN 6868-161). For each test the relative paragraph of the standard to which it refers is indicated. A table indicating the reference values is given at the end of this chapter.

1.1 Positioning of the patient support [paragraph 5.1] Not applicable.

1.2 Patient positioning accuracy [paragraph 5.2] NewTom 5G XL is equipped with two laser pointing systems to facilitate patient positioning. The laser beam is cross-shaped where the horizontal line identifies the axial plane and the vertical line the central sagittal plane of the volume examined. There is no pointer to identify the coronal plane.

1.2.1 Axial patient positioning accuracy [paragraph 5.2.1] Place an object along the axial plane identified by the horizontal line and do a scan. See “Data analysis software” in the "NNT User Manual" document.

1.2.2 Sagittal and coronal patient positioning light [paragraph 5.2.2] Place an object along the sagittal plane identified by the vertical line and do a scan. See “Data analysis software” in the "NNT User Manual" document.

1.3 Tomographic section thickness [paragraph 5.3]

1.3.1 Tomographic section thickness for axial scanning [paragraph 5.3.1] Refer to the "NNT User Manual" document.

1.3.2 Tomographic section thickness for helical scanning [paragraph 5.3.2] Not applicable.

FWHM value

Page 5: estceflasupport.com/data/media/product_documentation/newtom/... · 2019. 8. 23. · 1 IEC Acceptance Test The description of the software utilities for execution of the acceptance

EN DOSE DECLARATION AND ACCEPTANCE TEST 2

1.4 Dose [paragraph 5.4] Perform the scans for each modality as requested by Standard IEC 60601-2-44. The device automatically calculates the radiologic loading factors based on the scanned object. There is no way to manually select the radiological load for a scan.

1.5 Noise, mean CT number and uniformity [paragraph 5.5] Refer to the "NNT User Manual" document.

1.6 Spatial resolution [paragraph 5.6] Refer to the "NNT User Manual" document.

Page 6: estceflasupport.com/data/media/product_documentation/newtom/... · 2019. 8. 23. · 1 IEC Acceptance Test The description of the software utilities for execution of the acceptance

EN DOSE DECLARATION AND ACCEPTANCE TEST 3

1.7 IEC 61223-3-5: reference values The following table contains the reference values and the operating conditions to be applied in the various tests.

5.1 Positioning of the patient

support

N.A.

5.2 Patient positioning

accuracy

5.2.1 Axial patient positioning

accuracy

10mm from isocenter

5.2.2 Sagittal and coronal patient

positioning light accuracy

10mm from isocenter

5.3 Tomographic section

thickness

5.3.1 Tomographic section

thickness for axial scanning

Loading factors: automatically selected by the

software.

Thickness Recommended criteria

5.0 mm 1.0 mm

1.0 mm 0.5 mm

0.5 mm -0 +0.5 mm

5.3.2 Tomographic section

thickness for helical scanning

N.A.

5.4 Dose

Loading factors: automatically selected by the

software with Head Phantom

See Appendix A.

5.5 Noise, mean CT number

and uniformity

Loading factors: automatically selected by the

software.

Values are calculated as Arbitrary Unit (a.u.)

because of the machine is not designed to give HU.

Measured with Catphan 500.

Page 7: estceflasupport.com/data/media/product_documentation/newtom/... · 2019. 8. 23. · 1 IEC Acceptance Test The description of the software utilities for execution of the acceptance

EN DOSE DECLARATION AND ACCEPTANCE TEST 4

FOV Mode Voxel Dim. Stand. Dev Mean CT N. Uniformity

[21 x 19]

[18 x 16]

[15 x 22e] *

[15 x 12]

[15 x 5]

* Voxel size

only 300um

– Only Eco

Scan and

Regular

Scan

- Eco Scan -

300 um < 65 675 ÷ 1300 < 200

250 um < 100 675 ÷ 1300 < 175

200 um < 110 675 ÷ 1300 < 150

- Regular

Scan

Standard

Dose

300 um < 45 675 ÷ 1300 < 200

250 um < 70 675 ÷ 1300 < 175

200 um < 75 675 ÷ 1300 < 150

- Regular

Scan

Boosted

Dose

300 um < 30 675 ÷ 1300 < 200

250 um < 40 675 ÷ 1300 < 175

200 um < 45 675 ÷ 1300 < 150

- Enhanced

Scan

Standard

Dose

300 um < 35 675 ÷ 1300 < 125

250 um < 70 675 ÷ 1300 < 100

200 um < 80 675 ÷ 1300 < 75

- Enhanced

Scan

Boosted

Dose

300 um < 35 675 ÷ 1300 < 125

250 um < 70 675 ÷ 1300 < 100

200 um < 80 675 ÷ 1300 < 75

[15 x 5]

HiRes Eco Scan -

150 um < 80 675 ÷ 1300 < 125

125 um < 170 675 ÷ 1300 < 125

100 um < 190 675 ÷ 1300 < 125

HiRes Regular

Scan

Standard

Dose

150 um < 70 675 ÷ 1300 < 125

125 um < 150 675 ÷ 1300 < 125

100 um < 160 675 ÷ 1300 < 125

HiRes Regular

Scan

Boosted

Dose

150 um < 60 675 ÷ 1300 < 125

125 um < 115 675 ÷ 1300 < 125

100 um < 130 675 ÷ 1300 < 125

HiRes Enhanced

Scan

Standard

Dose

150 um < 60 675 ÷ 1300 < 125

125 um < 100 675 ÷ 1300 < 100

100 um < 110 675 ÷ 1300 < 75

HiRes Enhanced

Scan

Boosted

Dose

150 um < 60 675 ÷ 1300 < 125

125 um < 100 675 ÷ 1300 < 100

100 um < 110 675 ÷ 1300 < 75

[12 x 8]

[10 x 10]

[10 x 5]

[8 x 8]

[8 x 5]

- Eco Scan -

300 um < 35 675 ÷ 1300 < 75

250 um < 70 675 ÷ 1300 < 75

200 um < 80 675 ÷ 1300 < 75

- Regular

Scan

Standard

Dose

300 um < 30 675 ÷ 1300 < 75

250 um < 60 675 ÷ 1300 < 75

200 um < 65 675 ÷ 1300 < 75

Page 8: estceflasupport.com/data/media/product_documentation/newtom/... · 2019. 8. 23. · 1 IEC Acceptance Test The description of the software utilities for execution of the acceptance

EN DOSE DECLARATION AND ACCEPTANCE TEST 5

- Regular

Scan

Boosted

Dose

300 um < 25 675 ÷ 1300 < 75

250 um < 45 675 ÷ 1300 < 75

200 um < 50 675 ÷ 1300 < 75

- Enhanced

Scan

Standard

Dose

300 um < 35 675 ÷ 1300 < 75

250 um < 70 675 ÷ 1300 < 75

200 um < 80 675 ÷ 1300 < 75

- Enhanced

Scan

Boosted

Dose

300 um < 35 675 ÷ 1300 < 75

250 um < 70 675 ÷ 1300 < 75

200 um < 80 675 ÷ 1300 < 75

HiRes Eco Scan -

150 um < 80 675 ÷ 1300 < 45

125 um < 170 675 ÷ 1300 < 45

100 um < 190 675 ÷ 1300 < 45

HiRes Regular

Scan

Standard

Dose

150 um < 70 675 ÷ 1300 < 45

125 um < 150 675 ÷ 1300 < 45

100 um < 160 675 ÷ 1300 < 45

HiRes Regular

Scan

Boosted

Dose

150 um < 60 675 ÷ 1300 < 45

125 um < 115 675 ÷ 1300 < 45

100 um < 130 675 ÷ 1300 < 45

HiRes Enhanced

Scan

Standard

Dose

150 um < 60 675 ÷ 1300 < 45

125 um < 115 675 ÷ 1300 < 45

100 um < 130 675 ÷ 1300 < 45

HiRes Enhanced

Scan

Boosted

Dose

150 um < 60 675 ÷ 1300 < 45

125 um < 115 675 ÷ 1300 < 45

100 um < 130 675 ÷ 1300 < 45

[6 x 6]

- Eco Scan -

300 um < 300 800 ÷ 1300 < 75

250 um < 400 800 ÷ 1300 < 75

200 um < 500 800 ÷ 1300 < 80

- Regular

Scan

Standard

Dose

300 um < 30 800 ÷ 1300 < 75

250 um < 60 800 ÷ 1300 < 75

200 um < 65 800 ÷ 1300 < 75

- Regular

Scan

Boosted

Dose

300 um < 25 800 ÷ 1300 < 75

250 um < 45 800 ÷ 1300 < 75

200 um < 50 800 ÷ 1300 < 75

-

Enhanced

Scan

Standard

Dose

300 um < 35 800 ÷ 1300 < 75

250 um < 70 800 ÷ 1300 < 75

200 um < 80 800 ÷ 1300 < 75

- Enhanced

Scan

Boosted

Dose

300 um < 35 800 ÷ 1300 < 75

250 um < 70 800 ÷ 1300 < 75

200 um < 80 800 ÷ 1300 < 75

Page 9: estceflasupport.com/data/media/product_documentation/newtom/... · 2019. 8. 23. · 1 IEC Acceptance Test The description of the software utilities for execution of the acceptance

EN DOSE DECLARATION AND ACCEPTANCE TEST 6

HiRes Eco Scan -

150 um < 80 800 ÷ 1300 < 45

125 um < 170 800 ÷ 1300 < 45

100 um < 190 800 ÷ 1300 < 45

HiRes Regular

Scan

Standard

Dose

150 um < 70 800 ÷ 1300 < 45

125 um < 150 800 ÷ 1300 < 45

100 um < 160 800 ÷ 1300 < 45

HiRes Regular

Scan

Boosted

Dose

150 um < 60 800 ÷ 1300 < 45

125 um < 115 800 ÷ 1300 < 45

100 um < 130 800 ÷ 1300 < 45

HiRes Enhanced

Scan

Standard

Dose

150 um < 60 800 ÷ 1300 < 45

125 um < 115 800 ÷ 1300 < 45

100 um < 130 800 ÷ 1300 < 45

HiRes Enhanced

Scan

Boosted

Dose

150 um < 60 800 ÷ 1300 < 45

125 um < 115 800 ÷ 1300 < 45

100 um < 130 800 ÷ 1300 < 45

5.6 Spatial resolution

Loading factors: automatically selected by the software.

Scan mode: MTF50 (lp/cm) MTF10 (lp/cm)

Only [6x6] Eco Scan

[all FOVs] Eco / Regular Scan non-HiRes

[all FOVs] Eco / Regular Scan HiRes

[all FOVs] Enhanced Scan non-HiRes

[all FOVs] Enhanced Scan HiRes

> 7

> 8

> 9

> 6

> 8

> 13

> 15

> 18

> 12

> 15

Page 10: estceflasupport.com/data/media/product_documentation/newtom/... · 2019. 8. 23. · 1 IEC Acceptance Test The description of the software utilities for execution of the acceptance

EN DOSE DECLARATION AND ACCEPTANCE TEST 7

APPENDIX A: Dose declaration (5.4)

All measures are calculated at the center of the FOV.

Declared typical loading factors1

FOV Scan protocol Dose

protocol

Tube

voltage Scan time

Shot

time

Number

of shots FPS

Radio-logic

load

Anodic

current

cm x cm // // KV sec msec # Hz mAs mA

21x19

Eco Standard 110 18,0 5 270 15,00 10,29 7,62

Regular Standard 110 18,0 9 400 22,22 21,29 5,91

Regular Boosted 110 18,0 9 400 22,22 39,96 11,10

Enhanced Standard 110 25,8 9 600 23,26 37,35 6,92

Enhanced Boosted 110 25,8 9 600 23,26 62,51 11,58

18x16

Eco Standard 110 18,0 5 270 15,00 9,90 7,34

Regular Standard 110 18,0 9 400 22,22 21,06 5,85

Regular Boosted 110 18,0 9 400 22,22 37,50 10,42

Enhanced Standard 110 25,8 9 600 23,26 38,37 7,11

Enhanced Boosted 110 25,8 9 600 23,26 65,48 12,13

15 x 12

Eco Standard 110 18,0 5 270 15,00 11,56 8,57

Regular Standard 110 18,0 9 400 22,22 24,14 6,71

Regular Boosted 110 18,0 9 400 22,22 41,05 11,41

Enhanced Standard 110 25,8 9 600 23,26 44,13 8,17

Enhanced Boosted 110 25,8 9 600 23,26 74,93 13,88

15 x 5

Eco Standard 110 18,0 5 270 15,00 17,08 12,65

Regular Standard 110 18,0 9 400 22,22 34,36 9,54

Regular Boosted 110 18,0 9 400 22,22 56,44 15,68

Enhanced Standard 110 25,8 9 600 23,26 61,59 11,41

Enhanced Boosted 110 25,8 9 600 23,26 100,11 18,54

12 x 8

Eco Standard 110 18,0 5 270 15,00 16,78 12,43

Regular Standard 110 18,0 9 400 22,22 34,37 9,55

Regular Boosted 110 18,0 9 400 22,22 56,85 15,79

Enhanced Standard 110 25,8 9 600 23,26 59,89 11,09

1 Loading factors automatically set by the software using a “Head Phantom” in compliance with IEC 60601-2-44: 2009 Par. 203.108

positioned in the centre of the FOV.

Page 11: estceflasupport.com/data/media/product_documentation/newtom/... · 2019. 8. 23. · 1 IEC Acceptance Test The description of the software utilities for execution of the acceptance

EN DOSE DECLARATION AND ACCEPTANCE TEST 8

Declared typical loading factors1

FOV Scan protocol Dose

protocol

Tube

voltage Scan time

Shot

time

Number

of shots FPS

Radio-logic

load

Anodic

current

cm x cm // // KV sec msec # Hz mAs mA

Enhanced Boosted 110 25,8 9 600 23,26 98,38 18,22

10 x 10

Eco Standard 110 18,0 5 270 15,00 16,55 12,26

Regular Standard 110 18,0 9 400 22,22 34,21 9,50

Regular Boosted 110 18,0 9 400 22,22 56,60 15,75

Enhanced Standard 110 25,8 9 600 23,26 59,82 11,08

Enhanced Boosted 110 25,8 9 600 23,26 96,47 17,87

10 x 5

Eco Standard 110 18,0 5 270 15,00 21,00 15,56

Regular Standard 110 18,0 9 400 22,22 42,36 11,77

Regular Boosted 110 18,0 9 400 22,22 69,82 19,40

Enhanced Standard 110 25,8 9 600 23,26 71,31 13,21

Enhanced Boosted 110 25,8 9 600 23,26 115,17 21,33

8x8

Eco Standard 110 18,0 5 270 15,00 18,77 13,92

Regular Standard 110 18,0 9 400 22,22 37,71 10,48

Regular Boosted 110 18,0 9 400 22,22 61,94 17,22

Enhanced Standard 110 25,8 9 600 23,26 68,83 12,75

Enhanced Boosted 110 25,8 9 600 23,26 107,10 19,84

8x5

Eco Standard 110 18,0 5 270 15,00 20,51 15,20

Regular Standard 110 18,0 9 400 22,22 41,27 11,46

Regular Boosted 110 18,0 9 400 22,22 66,91 18,60

Enhanced Standard 110 25,8 9 600 23,26 73,28 13,57

Enhanced Boosted 110 25,8 9 600 23,26 129,15 23,93

6x6

Eco Standard 110 18,0 5 180 10,00 2,70 3,00

Regular Standard 110 18,0 9 400 22,22 41,40 11,93

Regular Boosted 110 18,0 9 400 22,22 67,71 19,51

Enhanced Standard 110 25,8 9 600 23,26 76,76 14,22

Enhanced Boosted 110 25,8 9 600 23,26 131,88 24,42

15x5 HiRes

Eco Standard 110 18,0 9 360 20,00 52,16 16,10

Regular Standard 110 20,2 15 360 17,82 111,90 20,72

Regular Boosted 110 20,2 15 360 17,82 172,79 32,00

Page 12: estceflasupport.com/data/media/product_documentation/newtom/... · 2019. 8. 23. · 1 IEC Acceptance Test The description of the software utilities for execution of the acceptance

EN DOSE DECLARATION AND ACCEPTANCE TEST 9

Declared typical loading factors1

FOV Scan protocol Dose

protocol

Tube

voltage Scan time

Shot

time

Number

of shots FPS

Radio-logic

load

Anodic

current

cm x cm // // KV sec msec # Hz mAs mA

Enhanced Standard 110 36,0 15 600 16,67 165,64 18,40

Enhanced Boosted 110 36,0 15 600 16,67 269,98 30,00

12x8 HiRes

Eco Standard 110 18,0 9 360 20,00 47,59 14,69

Regular Standard 110 20,2 15 360 17,82 102,01 18,89

Regular Boosted 110 20,2 15 360 17,82 170,29 31,54

Enhanced Standard 110 36,0 15 600 16,67 163,62 18,18

Enhanced Boosted 110 36,0 15 600 16,67 269,98 30,00

10x10

HiRes

Eco Standard 110 18,0 9 360 20,00 45,29 13,97

Regular Standard 110 20,2 15 360 17,82 100,29 18,57

Regular Boosted 110 20,2 15 360 17,82 168,53 31,21

Enhanced Standard 110 36,0 15 600 16,67 148,13 16,46

Enhanced Boosted 110 36,0 15 600 16,67 269,98 30,00

10x5 HiRes

Eco Standard 110 18,0 9 360 20,00 59,17 18,27

Regular Standard 110 20,2 15 360 17,82 125,37 23,23

Regular Boosted 110 20,2 15 360 17,82 172,79 32,00

Enhanced Standard 110 36,0 15 600 16,67 193,89 21,55

Enhanced Boosted 110 36,0 15 600 16,67 269,98 30,00

8x8 HiRes

Eco Standard 110 18,0 9 360 20,00 55,62 17,18

Regular Standard 110 20,2 15 360 17,82 119,24 22,09

Regular Boosted 110 20,2 15 360 17,82 172,79 32,00

Enhanced Standard 110 36,0 15 600 16,67 182,30 20,26

Enhanced Boosted 110 36,0 15 600 16,67 269,98 30,00

8x5 HiRes

Eco Standard 110 18,0 9 360 20,00 63,02 19,47

Regular Standard 110 20,2 15 360 17,82 135,56 25,13

Regular Boosted 110 20,2 15 360 17,82 172,79 32,00

Enhanced Standard 110 36,0 15 600 16,67 194,23 21,59

Enhanced Boosted 110 36,0 15 600 16,67 269,98 30,00

6x6 HiRes Eco Standard 110 18,0 9 360 20,00 64,21 19,82

Regular Standard 110 20,2 15 360 17,82 135,63 25,15

Page 13: estceflasupport.com/data/media/product_documentation/newtom/... · 2019. 8. 23. · 1 IEC Acceptance Test The description of the software utilities for execution of the acceptance

EN DOSE DECLARATION AND ACCEPTANCE TEST 10

Declared typical loading factors1

FOV Scan protocol Dose

protocol

Tube

voltage Scan time

Shot

time

Number

of shots FPS

Radio-logic

load

Anodic

current

cm x cm // // KV sec msec # Hz mAs mA

Regular Boosted 110 20,2 15 360 17,82 172,79 32,00

Enhanced Standard 110 36,0 15 600 16,67 205,50 22,84

Enhanced Boosted 110 36,0 15 600 16,67 269,98 30,00

15x22e

Eco Standard 110 36,0 5 540 15,00 24,11 8,93

Regular Standard 110 36,0 9 800 22,22 50,97 7,08

Regular Boosted 110 36,0 9 800 22,22 86,03 11,95

CineX [18x19] Scan time ≤20s 110 0,2÷4,0 10 20÷400 20,00 12,00 3,00

[18x19] Scan time ≥20s 110 3,15÷5,40 10 315÷540 15,00 16,20 3,00

Ray2D [18x19] 75 0,01÷0,5 10 1÷50 15,63 1,5÷35,0 3÷70

Page 14: estceflasupport.com/data/media/product_documentation/newtom/... · 2019. 8. 23. · 1 IEC Acceptance Test The description of the software utilities for execution of the acceptance

EN DOSE DECLARATION AND ACCEPTANCE TEST 11

Declared Dose: CTDI (CTDI100 2, CTDIperh

3, CTDIw 4, CTDIvol

5, CTDIfreeair 6)

and DLP7 using typical loading factors8

FOV Scan

protocol

Dose

protocol

CTDI 100

Center

CTDI 100

North

CTDI 100

West

CTDI 100

South

CTDI 100

East CTDI perh CTDI w CTDI vol DLP

CTDI

free air

cm x cm // // mGy mGy mGy mGy mGy mGy mGy mGy mGy*cm mGy

21x19

Eco Standard 1,6 2,0 1,8 1,6 1,7 1,8 1,7 1,7 35 2,1

Regular Standard 3,0 3,8 3,5 3,2 3,4 3,5 3,3 3,3 68 4,2

Regular Boosted 4,9 6,1 5,9 5,0 5,6 5,6 5,4 5,4 109 6,7

Enhanced Standard 5,7 7,1 6,7 6,2 6,8 6,7 6,4 6,4 129 7,1

Enhanced Boosted 8,6 10,7 9,9 9,1 10,1 10,0 9,5 9,5 193 10,5

18x16

Eco Standard 1,4 1,9 1,7 1,5 1,7 1,7 1,6 1,6 28 2,1

Regular Standard 2,9 3,7 3,4 3,1 3,4 3,4 3,2 3,2 57 4,1

Regular Boosted 4,5 5,6 5,2 4,8 5,2 5,2 5,0 5,0 87 6,4

Enhanced Standard 5,7 7,1 6,6 6,0 6,6 6,6 6,3 6,3 110 7,2

Enhanced Boosted 8,8 10,8 10,0 9,3 10,0 10,0 9,6 9,6 169 10,9

15 x 12

Eco Standard 1,5 2,0 1,7 1,6 1,8 1,8 1,7 1,7 23 2,4

Regular Standard 3,0 3,9 3,4 3,2 3,5 3,5 3,3 3,3 45 4,3

Regular Boosted 4,5 5,8 5,2 4,8 5,3 5,2 5,0 5,0 67 6,8

Enhanced Standard 5,9 7,5 6,9 6,3 6,9 6,9 6,5 6,5 88 7,8

Enhanced Boosted 9,1 11,6 10,6 9,7 10,6 10,6 10,1 10,1 136 11,8

15 x 5 Eco Standard 2,2 2,8 2,6 2,4 2,6 2,6 2,5 2,5 16 3,4

Regular Standard 4,4 5,6 5,2 4,7 5,2 5,2 4,9 4,9 31 6,8

2 CTDI100 Values expressed in mGy. The tolerance is ± 30%. CTDI100 values measured according to IEC 60601-2-44: 2009 Par.

203.109.1.

3 CTDIperh (CTDI peripheral) Values expressed in mGy. The tolerance is ± 30%. CTDIw and CTDIvol values calculated according to IEC

60601-2-44: 2009 Par. 201.3.211 and 201.3.212 c).

4 CTDIW (CTDI weighted)Values expressed in mGy. The tolerance is ± 30%. CTDIw and CTDIvol values calculated according to IEC

60601-2-44: 2009 Par. 201.3.211 and 201.3.212 c).

5 CTDIvol (CTDI volumetric)Values expressed in mGy. The tolerance is ± 30%. CTDIw and CTDIvol values calculated according to IEC

60601-2-44: 2009 Par. 201.3.211 and 201.3.212 c).

6 CTDIfree air Values expressed in mGy. The tolerance is ± 30%. CTDIfree air values calculated according to IEC 60601-2-44: 2009 Par.

203.109.2.

7 DLP Values expressed in mGy*cm. The tolerance is ± 30%. DLP values as defined in IEC 60601-2-44:2009 par. 201.3.214.

8 Loading factors automatically set by the software using a “Head Phantom” in compliance with IEC 60601-2-44: 2009 Par. 203.108

positioned in the centre of the FOV.

Page 15: estceflasupport.com/data/media/product_documentation/newtom/... · 2019. 8. 23. · 1 IEC Acceptance Test The description of the software utilities for execution of the acceptance

EN DOSE DECLARATION AND ACCEPTANCE TEST 12

Declared Dose: CTDI (CTDI100 2, CTDIperh

3, CTDIw 4, CTDIvol

5, CTDIfreeair 6)

and DLP7 using typical loading factors8

FOV Scan

protocol

Dose

protocol

CTDI 100

Center

CTDI 100

North

CTDI 100

West

CTDI 100

South

CTDI 100

East CTDI perh CTDI w CTDI vol DLP

CTDI

free air

cm x cm // // mGy mGy mGy mGy mGy mGy mGy mGy mGy*cm mGy

Regular Boosted 6,4 8,3 7,7 6,9 7,7 7,7 7,3 7,3 47 10,2

Enhanced Standard 8,6 11,2 10,1 9,2 10,1 10,1 9,6 9,6 62 11,8

Enhanced Boosted 12,7 16,5 14,9 13,6 14,9 15,0 14,2 14,2 91 17,5

12 x 8

Eco Standard 1,7 2,1 1,8 1,6 1,8 1,8 1,8 1,8 16 2,7

Regular Standard 3,6 4,1 3,6 3,2 3,6 3,6 3,6 3,6 33 6,3

Regular Boosted 5,3 6,3 5,4 4,8 5,4 5,5 5,4 5,4 50 12,5

Enhanced Standard 7,0 8,1 7,1 6,1 6,8 7,0 7,0 7,0 65 10,5

Enhanced Boosted 10,5 12,3 10,8 9,2 10,2 10,6 10,6 10,6 98 15,7

10 x 10

Eco Standard 1,5 1,6 1,3 1,2 1,3 1,4 1,4 1,4 16 2,8

Regular Standard 3,0 3,1 2,7 3,5 2,7 3,0 3,0 3,0 34 5,7

Regular Boosted 4,5 4,6 4,1 3,7 4,1 4,1 4,2 4,2 49 8,6

Enhanced Standard 6,4 6,5 5,4 5,4 5,4 5,7 5,9 5,9 68 9,7

Enhanced Boosted 9,5 9,6 8,0 8,0 8,0 8,4 8,7 8,7 100 14,1

10 x 5

Eco Standard 2,0 2,1 1,8 1,6 1,7 1,8 1,9 1,9 12 4,0

Regular Standard 4,1 4,1 3,6 3,2 3,5 3,6 3,8 3,8 24 8,0

Regular Boosted 6,1 6,1 5,4 4,8 5,2 5,4 5,6 5,6 36 12,1

Enhanced Standard 9,2 9,3 7,9 6,9 7,9 8,0 8,4 8,4 54 13,2

Enhanced Boosted 13,6 13,7 11,7 10,2 11,7 11,8 12,4 12,4 80 19,6

8x8

Eco Standard 1,8 1,6 1,3 1,2 1,3 1,4 1,5 1,5 14 3,3

Regular Standard 3,7 3,3 2,7 2,4 2,6 2,7 3,1 3,1 28 6,7

Regular Boosted 5,4 4,8 3,9 3,6 3,8 4,0 4,5 4,5 42 9,9

Enhanced Standard 7,5 6,4 5,2 4,7 5,2 5,4 6,1 6,1 56 11,8

Enhanced Boosted 10,8 9,2 7,5 6,7 7,5 7,7 8,7 8,7 81 16,8

8x5

Eco Standard 2,1 1,9 1,5 1,4 1,5 1,6 1,8 1,8 11 3,9

Regular Standard 4,4 3,8 3,1 2,9 3,1 3,2 3,6 3,6 23 7,9

Regular Boosted 6,3 5,6 4,6 4,2 4,5 4,8 5,3 5,3 34 11,6

Enhanced Standard 8,5 7,5 5,8 5,3 5,8 6,1 6,9 6,9 44 13,4

Enhanced Boosted 13,6 12,1 9,3 8,5 9,3 9,8 11,0 11,0 71 21,3

Page 16: estceflasupport.com/data/media/product_documentation/newtom/... · 2019. 8. 23. · 1 IEC Acceptance Test The description of the software utilities for execution of the acceptance

EN DOSE DECLARATION AND ACCEPTANCE TEST 13

Declared Dose: CTDI (CTDI100 2, CTDIperh

3, CTDIw 4, CTDIvol

5, CTDIfreeair 6)

and DLP7 using typical loading factors8

FOV Scan

protocol

Dose

protocol

CTDI 100

Center

CTDI 100

North

CTDI 100

West

CTDI 100

South

CTDI 100

East CTDI perh CTDI w CTDI vol DLP

CTDI

free air

cm x cm // // mGy mGy mGy mGy mGy mGy mGy mGy mGy*cm mGy

6x6

Eco Standard 0,5 0,4 0,3 0,3 0,3 0,3 0,4 0,4 3 0,9

Regular Standard 3,8 2,9 2,3 2,1 2,2 2,4 2,8 2,8 21 7,4

Regular Boosted 5,6 4,2 3,4 3,1 3,3 3,5 4,2 4,2 31 11,2

Enhanced Standard 7,5 5,7 4,4 4,0 4,3 4,6 5,6 5,6 41 13,2

Enhanced Boosted 11,7 8,9 6,8 6,1 6,7 7,1 8,6 8,6 63 20,2

15x5

HiRes

Eco Standard 5,5 7,2 6,9 6,0 6,7 6,7 6,3 6,3 40 9,6

Regular Standard 11,2 14,7 13,8 12,1 13,6 13,6 12,8 12,8 82 19,1

Regular Boosted 16,1 21,2 20,0 17,6 19,5 19,6 18,4 18,4 118 27,4

Enhanced Standard 21,7 27,5 25,2 23,1 25,8 25,4 24,2 24,2 155 29,1

Enhanced Boosted 32,7 41,5 37,9 34,8 38,9 38,3 36,4 36,4 233 43,2

12x8

HiRes

Eco Standard 4,8 5,9 5,2 4,3 4,9 5,1 5,0 5,0 46 7,9

Regular Standard 9,6 11,8 10,5 8,7 9,8 10,2 10,0 10,0 92 16,1

Regular Boosted 14,9 17,9 16,2 13,3 15,2 15,7 15,4 15,4 142 24,5

Enhanced Standard 17,3 19,8 17,5 14,9 16,7 17,2 17,2 17,2 159 26,0

Enhanced Boosted 26,4 30,3 26,8 22,7 25,5 26,3 26,4 26,4 243 39,3

10x10

HiRes

Eco Standard 4,2 4,8 3,9 3,4 3,8 4,0 4,0 4,0 46 7,1

Regular Standard 9,0 9,5 7,6 6,6 7,5 7,8 8,2 8,2 94 14,4

Regular Boosted 13,9 14,7 12,0 10,2 11,7 12,1 12,7 12,7 146 22,2

Enhanced Standard 14,8 15,2 12,7 11,2 12,7 12,9 13,6 13,6 155 21,7

Enhanced Boosted 24,3 24,8 20,8 18,4 20,8 21,2 22,2 22,2 255 35,7

10x5

HiRes

Eco Standard 6,3 6,8 5,5 4,6 5,0 5,5 5,8 5,8 37 10,5

Regular Standard 12,7 13,7 11,1 9,4 9,9 11,0 11,6 11,6 74 20,8

Regular Boosted 16,6 18,2 14,9 12,1 13,4 14,6 15,3 15,3 98 27,1

Enhanced Standard 23,7 24,8 19,7 17,5 19,7 20,4 21,5 21,5 138 32,6

Enhanced Boosted 30,7 32,1 25,5 22,6 25,5 26,4 27,9 27,9 179 42,6

8x8

HiRes

Eco Standard 5,0 4,6 3,6 3,1 3,5 3,7 4,2 4,2 38 9,1

Regular Standard 10,2 9,4 7,4 6,4 7,1 7,6 8,5 8,5 78 18,3

Regular Boosted 13,8 12,6 10,0 8,7 9,6 10,2 11,4 11,4 105 24,8

Page 17: estceflasupport.com/data/media/product_documentation/newtom/... · 2019. 8. 23. · 1 IEC Acceptance Test The description of the software utilities for execution of the acceptance

EN DOSE DECLARATION AND ACCEPTANCE TEST 14

Declared Dose: CTDI (CTDI100 2, CTDIperh

3, CTDIw 4, CTDIvol

5, CTDIfreeair 6)

and DLP7 using typical loading factors8

FOV Scan

protocol

Dose

protocol

CTDI 100

Center

CTDI 100

North

CTDI 100

West

CTDI 100

South

CTDI 100

East CTDI perh CTDI w CTDI vol DLP

CTDI

free air

cm x cm // // mGy mGy mGy mGy mGy mGy mGy mGy mGy*cm mGy

Enhanced Standard 18,3 15,8 12,6 11,4 12,6 13,1 14,8 14,8 137 28,3

Enhanced Boosted 25,0 21,6 17,2 15,7 17,2 17,9 20,3 20,3 187 39,0

8x5

HiRes

Eco Standard 6,1 5,7 4,1 3,8 4,1 4,4 5,0 5,0 32 10,9

Regular Standard 12,7 12,0 8,5 8,1 8,6 9,3 10,4 10,4 67 22,0

Regular Boosted 15,3 14,8 10,6 9,9 10,6 11,5 12,8 12,8 82 26,8

Enhanced Standard 20,7 17,6 14,7 12,6 14,7 14,9 16,8 16,8 108 32,4

Enhanced Boosted 27,5 23,4 19,5 16,7 19,6 19,8 22,4 22,4 143 42,2

6x6

HiRes

Eco Standard 5,3 4,0 3,1 2,9 3,1 3,3 4,0 4,0 29 11,3

Regular Standard 10,7 8,2 6,2 5,9 6,1 6,6 8,0 8,0 58 22,3

Regular Boosted 13,1 9,9 8,2 7,1 7,9 8,3 9,9 9,9 72 26,8

Enhanced Standard 18,0 13,4 10,9 9,4 10,9 11,1 13,4 13,4 98 34,4

Enhanced Boosted 22,7 16,8 13,7 11,8 13,7 14,0 16,9 16,9 123 43,0

15x22e

Eco Standard 2,0 2,5 2,3 2,1 2,3 2,3 2,2 2,4 65 4,7

Regular Standard 4,0 5,0 4,6 4,2 4,5 4,6 4,4 4,8 129 9,3

Regular Boosted 5,9 7,3 6,9 6,3 7,1 6,9 6,6 7,2 193 14,1

Page 18: estceflasupport.com/data/media/product_documentation/newtom/... · 2019. 8. 23. · 1 IEC Acceptance Test The description of the software utilities for execution of the acceptance

EN DOSE DECLARATION AND ACCEPTANCE TEST 15

Declared Dose: Air Kerma9 and DAP10 using typical loading factors11

FOV Scan protocol Dose protocol Air Kerma DAP

cm x cm // // mGy mGy*cm2

21x19

Eco Standard 2,0 786

Regular Standard 3,9 1.519

Regular Boosted 6,2 2.450

Enhanced Standard 6,6 2.587

Enhanced Boosted 9,6 3.791

18x16

Eco Standard 2,0 604

Regular Standard 3,8 1.184

Regular Boosted 5,9 1.824

Enhanced Standard 6,7 2.064

Enhanced Boosted 10,1 3.125

15 x 12

Eco Standard 2,1 458

Regular Standard 4,2 916

Regular Boosted 6,2 1.358

Enhanced Standard 7,3 1.580

Enhanced Boosted 11,0 2.388

15 x 5

Eco Standard 2,5 258

Regular Standard 5,0 516

Regular Boosted 7,5 767

Enhanced Standard 8,7 896

Enhanced Boosted 12,8 1.318

12 x 8

Eco Standard 2,6 316

Regular Standard 5,2 642

Regular Boosted 7,8 958

Enhanced Standard 8,9 1.090

9 Air Kerma Values expressed in mGy. The tolerance is ± 30%. Air Kerma values as defined in IEC 60601-1-3:2008 par. 3.4 calculated

according to IEC 60601-2-63:2012 par. 203.6.4.5.

10 Values expressed in mGy*cm2. The tolerance is ± 30%. DAP values as defined in IEC 60601-2-63:2012 par. 201.3.204 calculated

according to IEC 60601-2-63:2012 par. 203.6.4.5.

11 Loading factors automatically set by the software using a “Head Phantom” in compliance with IEC 60601-2-44: 2009 Par. 203.108

positioned in the centre of the FOV.

Page 19: estceflasupport.com/data/media/product_documentation/newtom/... · 2019. 8. 23. · 1 IEC Acceptance Test The description of the software utilities for execution of the acceptance

EN DOSE DECLARATION AND ACCEPTANCE TEST 16

Declared Dose: Air Kerma9 and DAP10 using typical loading factors11

FOV Scan protocol Dose protocol Air Kerma DAP

cm x cm // // mGy mGy*cm2

Enhanced Boosted 13,1 1.613

10 x 10

Eco Standard 2,6 343

Regular Standard 5,4 696

Regular Boosted 8,0 1.044

Enhanced Standard 9,1 1.183

Enhanced Boosted 13,3 1.728

10 x 5

Eco Standard 2,9 213

Regular Standard 6,0 438

Regular Boosted 9,0 652

Enhanced Standard 9,9 718

Enhanced Boosted 14,6 1.054

8x8

Eco Standard 2,8 244

Regular Standard 5,7 493

Regular Boosted 8,5 732

Enhanced Standard 10,0 868

Enhanced Boosted 14,3 1.233

8x5

Eco Standard 2,9 173

Regular Standard 5,9 353

Regular Boosted 8,7 519

Enhanced Standard 10,2 606

Enhanced Boosted 15,9 949

6x6

Eco Standard 0,7 39

Regular Standard 6,1 321

Regular Boosted 8,3 435

Enhanced Standard 10,7 560

Enhanced Boosted 16,3 857

15x5 HiRes

Eco Standard 7,0 718

Regular Standard 13,9 1.429

Regular Boosted 19,8 2.029

Enhanced Standard 21,1 2.168

Page 20: estceflasupport.com/data/media/product_documentation/newtom/... · 2019. 8. 23. · 1 IEC Acceptance Test The description of the software utilities for execution of the acceptance

EN DOSE DECLARATION AND ACCEPTANCE TEST 17

Declared Dose: Air Kerma9 and DAP10 using typical loading factors11

FOV Scan protocol Dose protocol Air Kerma DAP

cm x cm // // mGy mGy*cm2

Enhanced Boosted 31,4 3.223

12x8 HiRes

Eco Standard 6,8 829

Regular Standard 13,3 1.631

Regular Boosted 20,4 2.501

Enhanced Standard 21,7 2.664

Enhanced Boosted 32,6 3.997

10x10 HiRes

Eco Standard 6,7 868

Regular Standard 13,6 1.766

Regular Boosted 20,8 2.701

Enhanced Standard 20,5 2.660

Enhanced Boosted 33,4 4.346

10x5 HiRes

Eco Standard 7,8 566

Regular Standard 15,4 1.111

Regular Boosted 19,9 1.442

Enhanced Standard 24,3 1.761

Enhanced Boosted 31,6 2.288

8x8 HiRes

Eco Standard 7,7 665

Regular Standard 15,2 1.318

Regular Boosted 20,7 1.788

Enhanced Standard 23,9 2.068

Enhanced Boosted 32,7 2.829

8x5 HiRes

Eco Standard 8,2 487

Regular Standard 16,4 975

Regular Boosted 20,0 1.191

Enhanced Standard 24,3 1.447

Enhanced Boosted 31,5 1.877

6x6 HiRes

Eco Standard 8,5 444

Regular Standard 16,7 874

Regular Boosted 20,2 1.062

Enhanced Standard 25,8 1.352

Page 21: estceflasupport.com/data/media/product_documentation/newtom/... · 2019. 8. 23. · 1 IEC Acceptance Test The description of the software utilities for execution of the acceptance

EN DOSE DECLARATION AND ACCEPTANCE TEST 18

Declared Dose: Air Kerma9 and DAP10 using typical loading factors11

FOV Scan protocol Dose protocol Air Kerma DAP

cm x cm // // mGy mGy*cm2

Enhanced Boosted 31,9 1.675

15x22e

Eco Standard 4,4 1.901

Regular Standard 8,7 3.779

Regular Boosted 13,0 5.647

FOV Scan Protocol shots Air Kerma DAP

mGy mGy*cm2

CineX

[18x19]

Scan time ≤20 sec 400 shots 2,80 1105

Full FOV

Scan time >20 sec 540 shots 3,79 1495

FOV Scan Protocol mA Air Kerma DAP

mGy mGy*cm2

Ray2D [18x19]

Exposure Time: 500ms

3 mA 0,092 37

6 mA 0,16 63

12 mA 0,29 114

24 mA 0,51 202

36 mA 0,74 290

48 mA 0,93 368

60 mA 1,11 438

70 mA 1,25 492