Upload
others
View
7
Download
0
Embed Size (px)
Citation preview
1417SGC/WGG14x17 Wireless Flat Panel Detector
Detector Room Sharing
Options
Portable Battery Charger
X-raySource
Wireless Detector
Access Point
X-ray Wireless Access Source Detector Point
lrDA
Contact us 10, Yangnyeong-ro, Yangpyeong-ri, Moga-myeon, Icheon-si, Gyeonggi-do, Republic of KoreaTel : +82-2-449-6906FAX : +82-2-449-6909Website : http://www.sghealthcare.com
Specifications 1417SGC/SCC1417SGC: Csl:TI
ScintillatorType 1417SGG: Gd 022 5:Tb
Dimension 18.1 X 15.1 X0.6 in Weight 6.6 lbs Total Pixel Area 16.7x 14 in Pixel Pitch 127 µm Effective Pixel Matrix 3268X2756 Pixels A/D Conversion 14/ 16 bits Preview time (2x2 binning) sec Energy range 40-150 kVp
Pressure Distributed : 661 lbs Point : 330 Limiting Resolution Min. 2.5 / Max. 3.93 Ip/mm
OperatingTime Typ.4 Hours
(incl. battery)
Superb Image Quality1417SGC/SGG’s high Detector QuantumEfficiency (DQE) achieves superb image qualitywith low patient dose.
Lightweight & Fast417WCC/WGC weighs only 6.61b. Image preview occurs in less than 2 seconds.
High Visibility OLEO Illuminated OLED window brightly indicates flat panel detector status to the user.
Ergonomic Design Curved edges and a non-slip surface makes lifting and handling easier.
Durability
Providing patient throughput in your hospital and beyond
“ ”
Supporting up to 660 lb., the 141 ?WCC / WGC is manufactured with a seamless magnesium, unibody construction and is combined with a shock, vibration, and scratch resistant carbon fiber composition .
Water Resistant (IPX6)1417SGC/SGG is water resistant to most typical water spills in a hospital as well as in outdoor applications.
Product Specifications
1417SGC/SGG (R)
Rev No.
2
Detector
Parameter Spec. Unit
Sensor Type Amorphous Silicon with TFT (Single panel) -
Scintillator Type 1417SGC - CsI:Tl -
1417SGG - Gd2O2S:Tb
Total Pixel Matrix 3328 X 2816 Pixels
Total Pixel Area 422.7 X 357.6 mm
Pixel Pitch 127 μm
Effective Pixel Matrix 3268 X 2756 Pixels
A/D Conversion 14 / 16 bits
Data Transfer 802.11 a/g/n Wireless LAN, up to 300Mbps -
Preview time ≤5 sec
Energy range 40 ~ 150 kVp
MTF (@1lp/mm) 1417SGC - Min. 50 / Typ. 59 %
1417SGG - Min. 50 / Typ. 57
DQE (@0.1lp/mm) 1417SGC - Min. 50 / Typ. 65 %
1417SGG - Min. 36 / Typ. 45
Defective lines ≤10 Lines
Defective pixels ≤19000 Pixels
Dimension 460 X 384 X 15.4 mm
Weight 3.1 (incl. battery) kg
Window material Carbon fiber -
Trigger mode Manual Mode
Auto Trigger Mode (Auto Exposure Detection)
-
Power consumption Typ. 32.2 (Charging & Operating)
Typ. 10.2 (operating only)
Typ. 22 (charging only)
W
Power supply box (Option)
Parameter Spec. Unit
Dimension 188 X 92 X 41.5 mm
Weight 0.5 kg
Input rate 100 ~ 240VAC (50/60Hz) -
Output Typ. 24VDC (Max 1.9A) -
Product Specifications
1417SGC/SGG (R)
Rev No.
2
AGI box (Option)
Parameter Spec. Unit
Dimension 92.5 X 119 X 33.5 mm
Weight 0.3 kg
Cables
Item Spec. Unit
Detector - power link cable (Option) 7 m
P-interface cable (Option) 8 m
Trigger cable (Option) 10 m
Power cord 1.8 m
USB cable (Option) 1.8 m
Ethernet cable (Option) 10 m
Battery
parameter Spec. Unit
Size 236.2 x 127.8 x 6 mm
Weight 0.3 Kg
Input 12.6 VDC
Output 11.1 VDC
Cycle life Max. 500 cycles
Charging time Typ. 3 hours
Capacity Typ. 3400 mAh
Battery Charger
parameter Spec. Unit
Size 200 x 100 x 150 mm
Weight 0.9 kg
Input 20 VDC
Output 12.6 VDC
Product Specifications
1417SGC/SGG (R)
Rev No.
2
Battery Charger Adapter
parameter Spec. Unit
Size 160 x 76 x 43.7 mm
Weight Typ. 0.8 kg
Input 100~240VAC, 47~63Hz, 1.4~0.6A -
Output 20VDC, Max 6.0A -
Environmental Specification
Parameter Spec. Unit
Operating Temperature 5 ~ 40 ℃
Operating Humidity 30 ~ 75 % H.R
Storage Temperature -10 ~ 50 ℃
Storage Humidity 10 ~ 80 % H.R
Partial Pressure Max. 100 (@ 40mm diameter) kg
Overall Pressure Max. 150 kg
Product Specifications
1417SGC/SGG (R)
Rev No.
2
Defect Condition
Item Min. Typ. Max. Unit Note
Number of Normal Lines between
1defective line and 1defective line 4 Lines
Number of Normal Lines between
1defective line and 2defective lines 4 Lines
Number of Normal Lines between
2defective lines and 2defective lines 10 Lines
Number of adjacent defective lines 2 Lines
Total Pixel Defects 19000 Pixels
Total Line Defects 10 Lines
Definition of Defect
“Defect conditions #1”for tested dark frames:
- Pixel value is lesser than LOCAL_HORZ_MEDIAN-[Low]
Pixel value is lesser than LOCAL_VERT_MEDIAN-[Low]
- Pixel value is bigger than LOCAL_HORZ_MEDIAN+[High]
Pixel value is bigger than LOCAL_VERT_MEDIAN+[High]
“Defect conditions #2”for tested bright frames:
- Pixel value is lesser than LOCAL_AVERAGE-Low_Lim
- Pixel value is bigger than LOCAL_AVERAGE+High_Lim
whereLow_Lim= FRAME_MEDIAN*[Low]
High_Lim= FRAME_MEDIAN*[High]
LOCAL_AVERAGE is moving average of pixel’s surround within [Radius]
Defect Inspection
1) Integration time 0.5 second.
2) Inspection by human eye.
Pixel defect indicates pixel which is determined by test software. Total pixel
defects(in effective area) is sum of offset defects and substrate transmission map
defects.
Product Specifications
1417SGC/SGG (R)
Rev No.
2
#1 Drawings