    Remove the screw-top cap from the vialand transfer the BEARING nsPVAEmbolisation Particlesinto the barrel ofthe syringe. Replace the plunger on the syringe.

    To ensure proper hydration and suspension,gently agitate the saline/contrast mixtureof BEARING nsPVA EmbolisationParticlesand then wait 2-3 minutes,prior to injection.

    Remove the plunger from a standard30 mL syringe.

    Aspirate 10 mLof contrast and 10 mLof0.9% NaCl.

    Attach the 30 mL syringe to one port of aluer-lock 3-way stopcock. Attach a 1 mL or3 mL injection syringe to another port onthe stopcock.

    NOTE:If desired, a catheter may be attached to theremaining port on the stopcock.

    BEARING nsPVA Embolisation Particles

    Preparation Chart

    NOTE:Purge all air from the system prior to injection.

    Draw the BEARING nsPVA EmbolisationParticlessaline/contrast mixture into theinjection syringe slowly and gently to minimisethe potential of introducing air into the system.

    Inject the BEARING nsPVAEmbolisation Particlessaline/contrast mixture into the delivery catheterunder fluoroscopic visualisation using aslow pulsatile injection while observing thecontrast flow rate.

    NOTE

    : If there is no effect on the flow rate, repeatthe delivery process with additional injections ofthe BEARING nsPVA Embolisation Particles saline/contrast mixture. Consider using larger sizedBEARING nsPVA Embolisation Particles if theinitial injections do not alter the contrast flow rate.Exercise conservative judgment in determining theEmbolisation endpoint.

    10 mL 10 mL

    2-3 min.

    Merit Medical Systems, Inc.:

    1600 West Merit Parkway South Jordan, Utah 84095 1-801-253-1600 1-800-35-MERIT

    Merit Medical EUROPE, AFRICA AND THE MIDDLE EAST (EMEA):

    Amerikalaan 42, 6199 AE Maastricht-Airport The Netherlands Tel: +31 43 358 82 22

    Merit Medical Ireland Ltd. Parkmore Business Park West Galway, Ireland +353 (0) 91 703 733

    Free phone for specific country: Austria 0800 295 374 Belgium 0800 72 906 (Dutch) 0800 73 172 (French)Denmark 80 88 00 24 France 0800 91 60 30 Finland 0800 770 586 Germany 0800 182 0871

    Ireland (republic) 1800 553 163 Italy 800 897 005 Luxembourg 8002 25 22

    Netherlands 0800 022 81 84 Norway 800 11629 Sweden 020 792 445 UK 0800 973 115

    2013 Merit Medical Systems, Inc. All rights reserved. 402841101/A ID 031813

    www.merit.com

    NOTE: Upon completion of the treatment, remove the catheter while maintaining gentle suction soas not to dislodge BEARING nsPVA Embolisation Particles still within the catheter lumen.