iCheck Chroma User Manual iCheck Chroma is a test kit to measure Vitamin A in oil, empowering you with instant results to make confident decisions.

140901 Chroma Usermanual FINAL Druck (1)

Embed Size (px)

DESCRIPTION

chroma instruction.how to use chroma

Citation preview

  • iCheckChroma

    User Manual

    iCheck Chroma is a test kit to measure Vitamin A in oil, empowering you with instant results to make confident decisions.

  • 2QualityGuarantee

    Dear customer,Congratulations on your acquisition of iCheck Chroma!

    iCheck Chroma will be your reliable partner for the vitamin A analysis. iCheck is a high-tech portable photometer with precise and reliable results.

    iCheck is produced following strict rules of quality assurance according to ISO 9001:2008. This is accomplished by the use of high-grade components and equipment as well as a stream-lined production process. This process includes quality controls of each component and rigorous calibration of the device by trained technicians.

    Your iCheck Chroma comes with a 2-year warranty.

    If you have any questions, please contact us by calling +49 (0)33 28 35 15 000 or sending an e-mail to [email protected].

    www.bioanalyt.com www.facebook.com/bioanalyt

    www.linkedin.com/company/bioanalyt

    Development, manufacture and sales of all BioAnalyt test kits (devices, reagent vials) are carried out in

    accordance with ISO 9001:2008 and have been certified by TV NORD, Germany.

    TOP 100 INNOVATOR 2014

  • 3Copyright 2014 by BioAnalyt

    Version 1

    Contents

    4

    5

    6 11

    12 13

    14

    15

    16 17

    1819 2122

    23

    Case Content

    Device Description

    Instructions

    Menu Functions

    Software Installation

    Data Transfer

    Technical Data

    Frequently Asked Questions

    Power Supply

    Measurement

    General

    USB Stick

  • 4Check yourCase Content

    Grafik Case Content

    Hello! I am all yours now.Let's get off to a good start.

    To get the best out of me you need to make the very first measurement with the User Manual provided.

    measure for life

    iCheckChroma

    User Manual

    iCheck Chroma is a test kit to measure vitamin A in oil, empowering you with instant results to make confident decisions.

    C

    A

    B

    D

    E

    USB

    Grafik Case Content

    Hello! I am all yours now.Let's get off to a good start.

    To get the best out of me you need to make the very first measurement with the User Manual provided.

    measure for life

    iCheckChroma

    User Manual

    iCheck Chroma is a test kit to measure vitamin A in oil, empowering you with instant results to make confident decisions.

    C

    A

    B

    D

    E

    USB

    H

    A iCheck Chroma with metal cap

    B Battery charger and 4 rechargeable batteries

    C User Manual and software on USB stick

    D USB cable for data transfer to computer

    E Stand for reagent vials

    Your iCheck Chroma is delivered in a portable case. The items included in the case are listed below.

  • 5F Measurement chamber for iCheck Chroma reagent vials with removable metal cap (not shown)

    G Display monitor

    H Measurement key

    I Power key (On/O)

    J Menu navigation keys: left, right, up, down

    K Enter (OK) key

    L Escape key

    M USB cable mini-port for data transfer

    Use the 4 keys marked with triangles to navigate the menu structure of the device. To select an option, press the enter key. To exit an operation or to navi-gate one directory up, press the escape key.

    Grafik Device Description

    M

    F

    I

    J

    K L

    H

    G

    Not shown: Battery compartment on the bottom side.

    DeviceDescription

  • 6InstructionsChroma

    1Insert the batt eries

    The iCheck is equipped with 4 rechargeable batt eries (AA). Please charge them fully before device use.

    Open the batt ery compartment in the back of the iCheck by using a flat object.

    Switch on the device

    Start by placing the iCheck on a fl at and stable surface. Make sure the metal cap is covering the measurement chamber.

    Switch on your iCheck by pressing the power key. You are automatically in the measurement mode

    and you can proceed with the measurement. Before your fi rst measurement the monitor will display Empty fi le.

    Aft er your fi rst measurement your device will display Batch 1, Sample 1.

    In order to warm up the photometric unit, run an empty measurement by pressing the measure-ment key in total 3 times.

    The iCheck has an energy-saving function. 10 min-utes aft er the last keystroke the photometer will switch o automatically.

    Important: iCheck must be used with charged batt er-ies at all times even when connected to the computer via USB cable.

    CHROMA

    Batch 1Sample 1

    Baery

    Insert the batt eries as indicated.

    note:The batt eries can be recharged with the supplied charger. It takes about 2-3 hours to fully charge an empty batt ery. Charging is best performed within the temperature range of +5 C to +45 C.

  • 73Control the device

    Perform calibration control of your iCheck Chroma before each measurement set. Please refer to the section Configuration > OPC Test on page 13 for instructions.

    CHROMA

    Insert Blank

    CHROMA

    Prepare for measurement

    4Measure the blank

    Press the measurement key. The device will display please insert blank.

    Take a new iCheck Chroma reagent vial. Clean the vial with a paper tissue before inserting.

    Be sure to only hold the vial by its metal top.

    Take the vial you used for blank measurement and proceed with sample injection.

    Your iCheck Chroma will request during each mea-surement that you perform blank measurement. Always use a new vial for the blank measurement and then use same vial to inject your sample.

    Remove the metal cap and insert the vial into the measurement chamber of the iCheck.

    Press the measurement key again. Wait until the iCheck displays Prepare for measurement.

  • 85Prepare your sample

    iCheck Chroma quantitatively analyzes vitamin A in refined edible oil fortified in the range of 3 - 30 mg retinol equivalents (RE)/kg (10 - 100 interna-tional units (IU)/g).

    Mix your oil sample well. Use a new syringe with-out a needle attached and take up approx. 0.7 mL of your oil sample.

    Clean the end of the syringe with a paper tissue. Place the needle on the syringe. Hold the syringe with needle pointing up and gently tap on the syringe with your fingers to get the air bubbles to move up.

    Adjust the volume of the sample to exactly 0.5 mL (500 L) by ejecting excessive volume into the pa-per tissue. Make sure no air bubbles are left inside.

    4

    0.5 mL exactly

    4

    0.5 mL exactly

    InstructionsChroma

    6Inject your sample

    Slowly inject the sample into the iCheck Chroma reagent vial used for blank measurement.

    Take special care to inject the oil sample slowly along the side of the glass and above the clear liquid in the vial in order to produce two distinct layers.

    If you have injected the sample correctly, the content of your vial should look very similar to the image below. The oil will form a thin layer on top of the clear reagents in the vial. Make sure not to shake the vial yet.

    approx. 0.7 mL

    3

  • 9Start the measurement

    Now that your sample is ready, press the measure-ment key on the iCheck Chroma and follow the instruction as indicated on the display.

    The display will indicate Mix your sample at zero and start a countdown: 3 > 2 > 1 > 0.

    7

    2 x

    CHROMA

    Mix sample at zero

    3

    Invert and insert the vial

    8

    Important: Take care that no other objects, liquid or dust enters the measurement chamber. This would result in damaging the sensor and interfere with ac-curate measurement.

    When you see 0 invert the vial with oil sample 2 times and immediately insert the vial into the iCheck.

    You must insert the vial into the iCheck within 2 seconds while the display shows Insert sample.

    Note:The content of the vial will instantly turn blue, if the sample contains vitamin A.

  • 10

    InstructionsChroma

    9Result display

    iCheck Chroma will now analyze your sample. This will take approx. 30 seconds.

    Do not move the vial or the iCheck until the result is displayed.

    Sample result The result is displayed in mg RE/kg. Multiply your

    result by 3.33 to convert from mg RE/kg to IU/g.

    CHROMA

    Batch 1, Sample 18.55 mg RE/kg

    If your sample has under 3 mg RE/kg the device will display: Out of linear range. You will observe no blue color in the reagent vial.

    If your sample has over 30 mg RE/kg the device will display: Out of linear range. You will observe intense blue color in the reagent vial.

    CHROMA

    Batch 1, Sample 1Out of Linear Range

    (Range: 3 - 30 mg RE/kg)

    In case your sample has over 30 mg RE/kg dilute the sample with unfortified refined edible oil. Multiply the result by the dilution factor (DF).

    DF = (fortified oil, g + unfortified oil, g)/fortified oil, g.

    Note: For conversion of mg RE to IU please refer to the following factors: 1 mg RE = 1 mg retinol = 3333 IU 1000 IU = 0.3 mg RE.For support with dilution and calculation please contact BioAnalyt Support.

  • 11

    10Data storage

    For documentation purposes, iCheck Chroma has an internal memory to store up to 100 individual measurements including such information as batch number, sample number, date, time, and result.

    For detailed description of the data transfer to a computer please refer to the Data Transfer section.

    11Disposal

    Reagent vials contain hazardous chemicals and are disposed of according to national regulations for hazardous material. Collect the vials in a con-tainer and hand them over to a chemical waste company. Material safety data sheet (MSDS) of the reagent vials is provided with each shipment.

    Take extra care when disposing of the used needles to prevent injury: discard used needles into special container.

  • 12

    MenuFunctions

    By pressing the enter key you enter the menu of iCheck Chroma. Using the arrow keys you can scroll through the dierent options of the menu and with the enter key you can choose a function.

    View data

    You can select the following options: View Samples

    To display individual measurement results. View Statistics

    To display the meanstandard deviation of a selected batch.

    iCheck CHROMA

    View DataDelete DataTransfer DataConfiguration

    Delete data

    You can select the following options: Delete Sample

    To delete an individual result. Delete Batch

    To delete a batch with several measurements. Delete File

    To delete the file with all measurements performed on the device.

    Transfer data

    Use this function to transfer the data from the iCheck to your computer. Refer to the section Data transfer of this manual.

    Note: Data menu function is only displayed after a minimum of 1 measurement. Data Transfer menu function is only displayed after a minimum of 2 measurements.

  • 13

    Configuration

    To configure your iCheck you can select the follow-ing options:

    OPC Test Use this function to check the calibration of your device. To do this cover the measurement cham-ber with the metal cap and select OPC Test with the enter key. Control that the value displayed at (i.e. 45) is with in the range indicated in the back of the device (i.e. OPC: 40 - 70). OPC stands for Optical Perfor-mance Control. When the value is within the range proceed with your measurement. When the value is outside the indicated range, repeat the measurement. If the value remains outside contact BioAnalyt Support at [email protected] for assistance.

    Note:The Calibration Data of your iCheck can be provided on request. For this contact BioAnalyt Support at [email protected] and provide the serial num-ber indicated on the back of your device.

    Configuration

    DisplayDate & TimeOPC Test

    Display Set the brightness and contrast of the display.

    Date & Time Set the correct time and date of your time zone.

  • 14

    SoftwareInstallation

    Software installation

    The data stored on your iCheck can be transferred to a computer. To do so, install BioAnalyt Lab soft-ware which is provided on the USB stick .

    Initiate BioAnalyt Lab software set-up by double-clicking on the Set-up icon on the USB stick. Fol-low the instructions on your computer and make sure that BioAnalyt Lab is installed in the Pro-grams directory. Create a shortcut to your desk-top if you wish to. Finish installation by clicking Finish. The driver will automatically be installed.

    Upon accepting the License Agreement, a window will pop up where you can enter your personal information. This information can be viewed and edited by clicking on the Settings window.

    Note:BioAnalyt Lab software only works with Windows operating system (XP and later versions).

    Software update

    BioAnalyt Lab software can be updated by click-ing on the Update window. For the program to detect whether there are new updates available from BioAnalyt computer must be connected to the internet.

    If your current version (e.g. 1.1.0) is dierent from the newest version click on the Update key to proceed with the software update.

  • 15

    DataTransfer

    Start BioAnalyt Lab program by double-clicking the link on your desktop or by going to the Start Menu >> Programs >> BioAnalyt GmbH >> Bio-Analyt Lab.

    Plug in your iCheck to your computer via USB cable. A configuration window will appear after you connect your device to the computer. Here you can enter the information about your device. The serial number of the device can be found on the back side of the iCheck.

    Now the information about iCheck is saved on your computer and will be displayed the next time you connect your iCheck to your computer. This way, information about multiple iCheck devices can be stored on your computer.

    To initiate data transfer click on Start Transfer. Wait for data transfer to proceed and the sign Data Transfer in Progress to disappear.

    Now, all your data is saved and listed under Documents. You can view, save and edit this data by clicking on the Documents window.

    To save your data in CSV or EXCEL format select the file in the Documents, select the format and save the files to the desired location.

  • 16

    TechnicalData

    iCheck Chroma is a state-of-the-art portable photom-eter for measuring vitamin A concentration in edible oil. The iCheck is specifically developed and equipped with innovative technology to enable technically sophisticated measurement of absorption. iCheck Chroma displays vitamin A concentration in mg RE/kg.

    iCheck Chroma is equipped with the latest light emit-ting diode (LED) technology. Using LEDs as the light source eliminates all known problems with halogen lamps used in other readers. Modern LEDs emit particularly stable light and have very low energy consumption. Furthermore, each iCheck Chroma is calibrated at the production site in Germany with vitamin A standards according to quality manage-ment system. This eliminates the need for the user to calibrate the device. Based on the half life of the LED, approximately 10 million measurements can be performed reliably on iCheck. iCheck Chroma reagent vial is a 2-in-1 disposable reaction vial and cuvette. iCheck Chroma reagent vials contain a certified volume of reagents to determine vitamin A in the sample. One iCheck Chroma test kit consists of 100 reagent vials. No additional chemicals or equipment are required.

  • 17

    TECHnICAL DATA

    Analyte: Retinyl palmitate

    Sample type: Refined edible oils: palm, sunflower, corn, peanut, rapeseed, coconut

    Sample preparation: Not required

    Analysis method: Photometric

    Sample volume per analysis: 500 L (0.5 mL)

    Linear range: 3 30 mg retinol equivalents (RE)/ kg

    Time per analysis: 2 min

    Optimal temperature for measurement:

    20 30C

    Variation: 10%

    Method comparison: Validated against reference method high-performance liquid chromatography (HPLC)

    Sta qualification: 1 day training

    Weight: 0.45 kg

    Dimensions: 11 x 4 x 20 cm (W x H x L)

    Quality assurance

    iCheck and iCheck Chroma reagent vials are produ-ced according to quality management system (DIN EN ISO 9001:2008) certified by TV Nord in Germany. We guarantee that each vial contains the exact mix-ture of reagents safely sealed and crimped to ensure a shelf-life of 12 months (at 20 - 30 C), and that your iCheck delivers reliable results for at least 2 years.

  • 18

    Frequently AskedQuestions

    Make sure that the batteries are fully charged. In the lower right corner of the display a battery symbol is shown indicating the remaining battery charge. To recharge the batteries, place them in the charger provided in the case, connect it to a power supply and wait until the light turns green, indicating that power is at 100%. Place the batteries back in the device, switch it on. If iCheck is still not turning on, please contact BioAnalyt Support.

    You may use other AA/2100mAh/1.2V batteries. How-ever you cannot recharge those in the supplied charger.

    If OPC value is outside of the range indicated on the back of your device, measure again. If the value is out of range again, please contact BioAnalyt Support.

    iCheck does not turn on.

    May I use other batteries?

    My OPC value is outside the range. What should I do?

    Power Supply

  • 19

    No, there is no need to calibrate iCheck Chroma, because the device is calibrated during the manufac-turing process and calibration is programmed into the software.

    1. iCheck Chroma variation for repeated measure-ments of the same sample is below 10%. Therefore 10% deviation from expected value is acceptable.

    2. The oil sample was not well mixed before taking up with the syringe. Mix the oil sample well before tak-ing up into the syringe.

    3. No device warm-up was carried out.

    4. Incorrect execution of the test: It is very important to mix the vial at exactly 0. Insert the vial immediately after mixing. Maximum waiting time allowed between mixing and inserting into iCheck Chroma is 2 seconds.

    Do I need to calibrate iCheck Chroma?

    The result I received for a sample is higher/lower than expected. What might be the reason for this?

    Measurement

  • 20

    Frequently AskedQuestions

    5. Incorrect volume:Make sure that exactly 500 L (0.5 mL) of sample is injected into iCheck Chroma reagent vials.

    6. Incorrect injection:There has to be a distinct interface between oil sam-ple and reagents in the vial before mixing.

    7. The operator was not well trained. Contact BioAnalyt to obtain training and iCheck certification.

    1. Unclean vial surface: Make sure the reagent vial you are measuring is ab-solutely clean and does not have any fingerprints on it. If not, wipe the vial with a paper tissue (optional: wet the tissue with alcohol to improve the cleaning).

    2. Sunlight: Do not measure in the direct sunlight.

    What might interfere with the measurement procedure?

    Measurement

  • 21

    iCheck reagent vials must be stored upright at room temperature and protected from direct sunlight.

    1. It is recommended to measure in the range of 20 - 30 C (68 - 86 Fahrenheit). Do not use iCheck at tem-peratures above 40 C.

    2. It is further recommended to store the iCheck and the iCheck reagent vials at least two hours before starting the measurement in the room in which the measurement will be performed. This procedure ensures that both, the vials and the device have the same temperature.

    3. Humidity does not aect iCheck measurements.

    For selecting a new batch press the right arrow key. The batch function is used to group samples, e.g. samples from 1 day or 1 region can be measured in batch 1. If you proceed to measure the samples of a dierent day or region, select a new batch (i.e. 2).

    How should I store the reagent vials?

    Do temperature or humidity inuence the iCheck measurements?

    What is a batch and how can I select a new batch?

  • 22

    Frequently AskedQuestions

    General

    Which form of vitamin A can be measured?

    Does the Data Transfer work with other operating systems like Apple OSX etc.?

    How can iCheck Chroma test kits be ordered?

    Where do I get help with other questions that are not mentioned here?

    iCheck Chroma measures retinyl palmitate, a com-mon type of vitamin A added to refined edible oil.

    No, BioAnalyt Lab may only be used with Windows Operating System.

    An order can be placed by visiting the BioAnalyt website www.bioanalyt.com/shop or by sending e-mail to [email protected].

    We would love to hear from you! Please send us an e-mail at [email protected].

    You can also join the discussion by following us on Facebook or LinkedIn.

    www.facebook.com/bioanalyt www.linkedin.com/company/bioanalyt

  • 23

    USB Stick

    Find the BioAnalyt Lab Software and further product information on the USB stick.

  • 24

    measure for lifeBioAnalyt GmbHRheinstrae 1714513 Teltow, GermanyT +49 (0)33 283 51 5 [email protected]