Description and Function The 1096+ XYZ Autosampler is fully compatible for operation with an Aurora 1030 TOC analyzer. The 1096 autosampler aspirates liquid samples from vials and transfers each sample aliquot to an Aurora 1030 TOC analyzer for analysis. The autosampler holds four 24-vial trays and a 12-vial standards tray for fully automated, unattended operation. The 1096 supports a number of special functions to address a range of sample conditions and analysis requirements. System configuration for sample pretreatment allows pre-acidification and purging of samples on the 1096 prior to sampling. This technique removes the TIC content of samples, reducing the analysis time required for TOC measurements. Septum piercing allows closed vial sampling for applications where open vials may compromise the accuracy of measurements. Operating Principle The sampling sequence of the 1096 autosampler is user-programmed through the Aurora 1030 touchscreen display. The autosampler arm moves the sampling needle to the X-Y position above a sample vial. The Z-mechanism lowers the sampling needle into the vial and aspirates the specified sample volume and transfers it to the reaction chamber of Aurora 1030. Following the sample transfer step, the sampling needle returns to a wash station to be rinsed. A user- programmable number of rinses are performed to clean the needle before the next vial is sampled. A World of Solutions 1096+ XYZ Autosampler Autosampler for the Aurora 1030 TOC Analyzer Automates introduction of 96 samples to the Aurora 1030 TOC Analyzer Configurable for pre-acidification and sparging of sample TIC content within the autosampler to reduce TOC analysis time Integral wash station performs programmable rinse steps to clean sampling needle and prevent cross-contamination Supports sampling from open vials and septum piercing of sealed vials Principal Applications Drinking water • Wastewater Ultrapure water Cleaning validation Water for injection • Groundwater Surface water Industrial process water • Seawater Sampling Methods Open vial Septum piercing

1096+ XYZ Autosampler - jas-sg.com · The 1096+ XYZ Autosampler is fully compatible for operation with an Aurora 1030 TOC analyzer. The 1096 autosampler aspirates liquid samples from

  • Upload
    others

  • View
    6

  • Download
    0

Embed Size (px)

Citation preview

Page 1: 1096+ XYZ Autosampler - jas-sg.com · The 1096+ XYZ Autosampler is fully compatible for operation with an Aurora 1030 TOC analyzer. The 1096 autosampler aspirates liquid samples from

Description and FunctionThe 1096+ XYZ Autosampler is fully compatible for operation with an Aurora 1030 TOC analyzer. The 1096 autosampler aspirates liquid samples from vials and transfers each sample aliquot to an Aurora 1030 TOC analyzer for analysis. The autosampler holds four 24-vial trays and a 12-vial standards tray for fully automated, unattended operation.

The 1096 supports a number of special functions to address a range of sample conditions and analysis requirements. System configuration for sample pretreatment allows pre-acidification and purging of samples on the 1096 prior to sampling. This technique removes the TIC content of samples, reducing the analysis time required for TOC measurements. Septum piercing allows closed vial sampling for applications where open vials may compromise the accuracy of measurements.

Operating PrincipleThe sampling sequence of the 1096 autosampler is user-programmed through the Aurora 1030 touchscreen display. The autosampler arm moves the sampling needle to the X-Y position above a sample vial. The Z-mechanism lowers the sampling needle into the vial and aspirates the specified sample volume and transfers it to the reaction chamber of Aurora 1030. Following the sample transfer step, the sampling needle returns to a wash station to be rinsed. A user-programmable number of rinses are performed to clean the needle before the next vial is sampled.

A World of Solutions

1096+ XYZ AutosamplerAutosampler for the Aurora 1030 TOC Analyzer

• Automates introduction of 96 samples to the Aurora 1030 TOC Analyzer

• Configurable for pre-acidification and sparging of sample TIC content within the autosampler to reduce TOC analysis time

• Integral wash station performs programmable rinse steps to clean sampling needle and prevent cross-contamination

• Supports sampling from open vials and septum piercing of sealed vials

Principal Applications• Drinking water

• Wastewater

• Ultrapure water

• Cleaning validation

• Water for injection

• Groundwater

• Surface water

• Industrial process water

• Seawater

Sampling Methods• Open vial

• Septum piercing

Page 2: 1096+ XYZ Autosampler - jas-sg.com · The 1096+ XYZ Autosampler is fully compatible for operation with an Aurora 1030 TOC analyzer. The 1096 autosampler aspirates liquid samples from

PO Box 9010 College Station, Texas 77842 • Tel (800) 653-1711 • FAX (979) 690-0440 • www.oico.com • E-mail [email protected]

Microsoft and Windows are registered trademarks of Microsoft Corporation. Publication 22710806Teflon is a registered trademark of E.I. duPont Nemours.

Specifications

Dimensions 64.8 cm H x 52.7 cm W x 49.5 cm D(25.5” H x 20.8” W x 19.5” D)

Weight 11.8 kg (26 lbs)

Benchspace Footprint 2,506 cm2 (389.5 in2)

Environment 10–30 °C operating temperature0–95% relative humidity

Sample Capacity 96 sample vials

Standards 12-position standard tray

Sampling Needle 8.4”, two-hole, stainless steel

Sample Transfer Tubing 0.045” I.D. x 1/8” O.D. Teflon® tubing

Vials 40-mL VOA vials

Caps Open-hole caps

Septa TFE-faced septa

Sample Transfer Volume Range 10 µL–10 mL

Particulate Handling Up to 500 µm

Special Functions Sample pre-treatment for TIC removalSeptum piercing

Power Supply Universal Input 100–240 VAC (±10%); 50/60 Hz; 80 W

Communication RS-485 (Aurora 1030 analyzer to 1096 autosampler)

Certifications CE

Warranty 12 months on parts and labor

1096+ XYZ Autosampler with Aurora 1030C TOC Analyzer

Programmable Autosampler Parameters