3.1 INTRODUCTION

Scanning Tunneling Microscope was developed first in 1982 by Binning, Rohrer,

Gerber and Weibel at IBM in Zurich Switzerland and they won Noble prize for

this invention in 1986. AFM (atomic force microscope) is an improvement of

STM which was first invented by Gerd Binnig, Quate and Christoph Gerber in

1986. Scanning Probe microscopy (SPM) consists of a family of microscopy

where a sharp probe is scanned across a surface and the tip-sample interactions

are monitored. SPM consists of many forms like scanning tunneling microscope

(STM), atomic force microscope (AFM), magnetic force microscope (MFM),

electric force microscope (EFM) etc.

In present work, AFM has been used to measure surface morphology. The

essential property of AFM is the interaction force between the tip and the

sample, which depends on their distance. At close contact the force is repulsive

while at a larger separation the force is attractive. This results in different

operation modes which should be chosen according to the characteristics of the

sample, since each mode has different advantages.

For microanalysis the scanning electron microscope (SEM) has been used.

Microanalysis is the analysis of “very small” samples—by whatever technique is

available. Historically, however, the term has had a much narrower meaning.

When electrons of appropriate energy impinge on a sample, they cause the

emission of x-rays whose energies and relative abundance depend upon the

composition of the sample. Using this phenomenon to analyze the elemental

content of microvolumes (roughly one to several hundred cubic micrometers) is

what we commonly mean by microanalysis.

By scanning the beam in a television-like raster and displaying the intensity of a

selected X-ray line, element distribution images or 'maps' can be produced. Also,

images produced by electrons collected from the sample reveal surface

topography or mean atomic number differences according to the mode selected.

The scanning electron microscope (SEM), which is closely related to the electron

probe, is designed primarily for producing electron images, but can also be used

for elements mapping, and even point analysis, if an X-ray spectrometer is added.

Page 3: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

101

There is thus a considerable overlap in the functions of these instruments.

Structure, microstructure, and defect geometry, as well as chemical composition

and spatial distribution are important parameters determining the behavior of

materials and practical applications.

3.2 SCANNING ELECTRON MICROSCOPE (SEM)

3.2.1 Introduction

The scanning electron microscope (SEM) is a type of electron microscope

capable of producing high resolution images of a sample surface. Due to the

manner in which the image is created, SEM images have a characteristic three

dimensional appearance and are useful for observing the surface structure of the

sample. In scanning electron microscopy (SEM), a fine probe of electrons with

energies typically up to 40keV is focused on a specimen, and scanned along a

pattern of parallel lines. Various signals are generated as a result of the impact of

the incident electrons, which are collected to form an image or to analyse the

sample surface. These are mainly secondary electrons, with energies of a few

tens of eV, high-energy electrons back-scattered from the primary beam and

characteristic X-rays. Such rich physical interaction in a practical tool, making

the SEM the powerful instrument it is today in materials and life-science.

The history of electron microscopy began with the development of electron

optics. In 1926, Busch studied the trajectories of charged particles in axially

symmetric electric and magnetic fields, and showed that such fields could act as

particle lenses, laying the foundations of geometrical electron optics [1]. Nearly at

the same time, the French physicist de Broglie introduced the concept of

corpuscule waves. A frequency and hence a wavelength was associated with

charged particles: wave electron optics began [2]. Following these two discoveries

in electron optics, the idea of an electron microscope began to take shape.

In 1931, independently of the ‘‘material wave’’ hypothesis put forward by de

Broglie several years earlier (1925), Ruska and his research group in Berlin,

were working on electron microscopy. They were disappointed learning that

Page 4: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

102

even with electrons a wavelength would limit the resolution. But they found

using de Broglie equation that electron wavelengths were almost five orders of

magnitude smaller than the wavelength of light used in optical microscopy. It

was thus considered that electron microscopes still could prove a better

resolution than light instruments, and no reason existed to abandon this aim. In

1932, Knoll and Ruska tried to estimate the resolution limit of the electron

microscope. Assuming the resolution limit formula of the light microscope was

still valid for material waves, they replaced the light wavelength by the electrons

wavelength at an accelerating voltage of 75kV. A theoretical limit of 0.22nm

resulted, a value experimentally reached only 40 years later. Although these

calculations proved it was possible to reach a better-than-light-microscope

resolution when working at high magnifications [3], many technical limits had to

be to overcome. Ruska and Knoll tried to implement Busch’s lens formula

experimentally. Their work resulted in the construction of the first transmission

electron microscope (TEM) in 1931, with a magnification of 16 [4].

Knoll built a first ‘‘scanning microscope’’ in 1935. However, as he was not using

demagnifying lenses to produce a fine probe, the resolution limit was around

100mm because of the diameter of the focused beam on the specimen. In 1938,

von Ardenne clearly expressed the theoretical principles underlying the scanning

microscope, as we know it today. Because it was difficult to compete with TEM in

resolution achieved for thin specimens, the scanning microscopy development

was oriented more toward observing the surface of samples. The first true SEM

was described and developed in 1942 by Zworykin, who showed that secondary

electrons provided topographic contrast by biasing the collector positively

relative to the specimen. One of his main improvements was using an electron

multiplier tube as a preamplifier of the secondary electrons emission current. He

reached a resolution of 50 nm, which was still considered low in comparison

with the performance obtainable in TEM [5].

In 1948, Oatley began to build an SEM based on Zworykin’s microscope.

Following this development, Smith (1956) shown that signal processing could be

used to improve micrographs. He introduced nonlinear signal amplification, and

Page 5: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

103

improved the scanning system. Besides, he was also the first to insert a stigmator

in the SEM to correct lens cylindrical imperfections.

In 1960, Everhart and Thornley greatly improved the secondary electron

detection. A new detector was created with a positively biased grid to collect

electrons, a scintillator to convert them to light, and a light-pipe to transfer the

light directly to a photomultiplier tube [5].

In 1963, Pease and Nixon combined all of these improvements in one

instrument: the SEMV with three magnetic lenses and an Everhart–Thornley

detector (ETD). This was the prototype for the first commercial SEM, developed

in 1965—the Cambridge Scientific Instruments Mark I ‘‘Stereoscan’ [6]. The SEM

we are using today are not very different from this first instrument.

3.2.2 Principle of SEM

Scanning electron microscope (SEM) consists of an electron gun and

electromagnetic lens system to study the surface structure and morphology

of solids. A well-defined electron beam impinges on the specimen and leads

to generation of secondary electrons, back scattered electrons, absorbed

electrons, characteristic X-rays etc (figure 3.1). These electrons can be detected

by suitable detectors and give information about the surface structure and

morphology of the specimens. The characteristic X-rays generated are used for

Figure 3.1 Example of some of the different types of signals produced when

high-energy electron impinge on a material.

Page 6: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

104

the identification and estimation of different elements present in the specimen

by energy dispersive spectrometer (EDS) and wavelength dispersive

spectrometer (WDS). The depth of focus is much larger than the optical

microscope image even at lowest magnifications is one of the major advantages

of SEM.

3.2.3 Working of an SEM instrument

Electrons from a thermionic or field-emission cathode are accelerated by a

voltage of 1-50kV between cathode and anode. The smallest beam cross section

at the gun (the crossover) with a diameter of the order of 10-50µm for

thermionic and 10-100nm for field-emission guns, is demagnified by two or

three stage electromagnetic lens system, so that an electron probe of diameter 1-

10nm carrying current of 10-10 to 10-12 Å is formed at the specimen surface.

A deflection coil system in front of the last lens scans the electron probe in a

raster across the specimen and in synchronism with the electron beam of a

separate cathode ray tube (CRT). The intensity of the CRT is modulated by one of

the signals recorded to form an image. The magnification can be increased

simply by decreasing the scan-coil current and keeping image size on CRT

constant. Figure 3.2 shows the general schematic diagram of SEM.

An advantage of SEM is the wide variety of electron-specimen interactions that

can be used to form an image to give qualitative and quantitative information.

The large depth of focus, excellent contrast and the straight forward preparation

of solid specimen are advantages of SEM.

Here an electron beam scans the object (the specimen) and because of

synchronized scans of electron beam and the CRT screen (nowadays, monitor),

there is one-to-one correspondence between the spot on the specimen and the

spot on the screen. Unlike Optical microscopy, SEM requires the vacuum

environment and specimen surface to be electrically conductive.

The electron beam is produced by hair-pin shaped tungsten (W) filament by

thermionic emission. The acceleration voltage of 5 to 50kV can be applied

between anode and cathode and hence we get the electron beam of such energy.

Page 7: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

105

Figure 3.2 Schematic diagram of Scanning Electron Microscope (SEM).

This diverged electron beam passes through a pair of electro-magnetic lenses

(coils) and finely passes through probe forming lens, which makes the beam in a

form of a fine probe (~10nm diameter). This fine electron probe scans on the

specimen area (marked as lines in figure 3.2) in a linear manner. Another

electron beam is in synchronization with this beam which scans on the CRT (or

Page 8: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

106

monitor) with the help of same scan generator. In SEM, the formation of image

takes place because of electron beam-specimen interaction [7].

3.2.4 Sample preparation

Since the electron probe analyses only to a shallow depth, specimens should be

well polished so that surface roughness does not affect the results. Sample

preparation is essentially as for reflected light microscopy, with the provision

that only vacuum compatible materials must be used. Opaque samples may be

embedded in epoxy resin blocks.

In principle, specimens of any size and shape (within reasonable limits) can be

analyzed. Holders are commonly provided for 25mm (1") diameter round

specimens and for rectangular glass slides. Standards are either mounted

individually in small mounts or in batches in normal-sized mounts.

3.2.5 Electron-specimen interaction

Electron beam-specimen interaction gives various signals which can be used to

form images and also can be used for other important information.

In inelastic scattering, the trajectory of the incident electron is only slightly

perturbed, but energy is lost through interactions with the orbital electrons of

the atoms in the specimen. ‘Loss’ of kinetic energy (of the primary electron) is

mainly caused by the interaction with the electrons of atoms of the specimen.

The concept of interaction volume of the primary beam electrons and the

sampling volume of the emitted secondary radiation are important both in

interpretation of SEM images and in the proper application of quantitative X-ray

microanalysis. The image details and resolution in the SEM are determined not

by the size of the electron probe by itself but rather by the size and

characteristics of the interaction volume.

When the accelerated beam electrons strike a specimen they penetrate inside it

to depths of about 1µm and interact both elastically and inelastically with the

solid, forming a limiting interaction volume from which various types of

Page 9: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

107

radiation emerge, including BSE, SE, characteristic and bremsstrahlung x-rays,

and cathodoluminescence in some materials (figure 3.3).

The combined effect of elastic and inelastic scattering controls the penetration of

the electron beam into the solid. The resulting region over which the incident

electrons interact with the sample is known as interaction volume. The

interaction volume has several important characteristics, which determine the

nature of imaging in the SEM.

The energy deposition rate varies rapidly throughout the interaction volume,

being greatest near the beam impact point. The interaction volume has a distinct

shape (figure 3.3). For low-atomic-number target it has distinct pear shape. For

intermediate and high-atomic number materials the shape is in the form of hemi-

sphere. The interaction volume increases with increasing incident beam energy

and decreases with increasing average atomic number of the specimen. For

secondary electrons the sampling depth is from 10 to 100nm and diameter

equals the diameter of the area emitting backscattered electrons. BSE are

emitted from much larger depths compared to SE. Ultimately the resolution in

Figure 3.3 The interaction volume.

Page 10: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

108

the SEM is controlled by the size of the interaction volume.

3.2.6 Magnification, brightness and contrast of the SEM images

Electron images are obtained by rastering the electron beam across the

specimen surface using the deflection coils inside the objective lens and

synchronously rastering the output signal of the detector on a cathode-ray tube

(CRT). The ratio of the area rastered on the specimen to that of the CRT gives the

magnification. For example, a rastered area of 200µm2 (200 x 10-6 m2) displayed

on CRT with an area of 20cm (200,000 x 10-6 m2) yields a magnification of 1000x.

This is a very different process than the production of an image by an optical

microscope. Electronic images are sequentially "constructed" during the

rastering of the beam, whereas in optical systems all parts of the sample are

imaged simultaneously. Brightness can be increased/decreased by amplifying

the signal from the sample and it can be affected by topography, composition,

electrical conductivity, and other properties of the sample. Contrast reflects the

variation in the signal from point to point. Contrast can also be enhanced by

electronically increasing the difference between small and large signals. Working

distance (WD), probe current and probe diameter, incorrect electron gun

alignment, astigmatism, diffused scattering of electrons from the edges (edge

effect), electrostatic charging of the sample, external disturbances(magnetic

fields, flooring, improper grounding) are factors affecting the quality of the

image. The SEM image looks three dimensional because of SEM's ability to focus

up to certain depth and hence the images can be easily interpreted.

3.3 IMAGE FORMATION AND ELEMENT ANALYSIS WITH A SEM-EDAX

SYSTEM

On the page about the principle of a scanning electron microscope (SEM), we saw

that the sample in this instrument is bombarded by an electron beam in order to

obtain a detailed topographical image of the surface of the sample from the

ejected electrons (secondary electrons, see figure 3.4 A here below). Besides,

there are scanning electron microscopes which are equipped with EDS (Energy

Page 11: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

109

Dispersed Spectroscopy) or EDAX (Energy-Dispersed Analysis of X-rays)

detectors that capture the emitted X-ray (figure 3.4 C, 3.4 D and 3.4 E). With such

instruments, like the one shown here below (figure 3.18), it is possible to

determine which elements are present in the surface layer of the sample (at a

depth in the micrometer range) and where these elements are present

("mapping technique"). This particular microscope also allows one to capture

directly reflected electrons, the so-called back scattered electrons (figure 3.4 B),

from which one can obtain a global appreciation whether one or several

elements are present in the surface layer of the sample. Also the so-called Auger

electrons, which are emitted just under the surface (figure 3.4 F, 3.4 G, 3.4 H),

provide information about the nature of the atoms in the sample.

Figure 3.4 A: The bombarding electrons (=primary electrons) can penetrate in

Figure 3.4 (A, B, C, D, E, F, G, H) Secondary emission in Scanning Electron

Microscopy.

Page 12: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

110

the electron shells of the atoms composing the surface of the sample. The energy

(negative charge, mass, velocity) of these incident electrons can be converted to

eject local electrons, so-called secondary electrons, from the shells of the atoms

in the surface of the specimen. This information can be utilized to reconstruct a

detailed topographical image of the sample (SEI = Secondary Electrons Imaging).

The final image looks like a shadow-cast photograph of the surface of the sample.

This record of the morphology is the best-known application of a scanning

electron microscope.

Figure 3.4 B: Primary electron can also be reflected by atoms at about 10-100

nanometer depth at the surface. These so-called "back-scatter" conserve their

energy at incidence, but their direction of propagation has been modified upon

interaction. One can obtain a rough representation whether the surface of the

sample is constituted of a single or multiple elements.

Figure 3.4 C, D, E: At the surface of the sample electrons in the deeper electron

shells (shell K in 3.4 C) can be ejected by primary electrons (Pe- indicated in red),

resulting in an electron hole. When this lower-shell position is filled by an

electron from higher shell (green arrow in 3.4 D) energy is released. This can be

as light (photons; the phenomenon is also called Cathodoluminescence) or as X-

ray. Because each element emits an own characteristic energy value, the

elements present in the micrometer range depth of the sample can be

determined.

Figure 3.4 F, G, H: Another phenomenon is that the energy released upon filling a

hole in the K shell by an electron from the L shell is used to expulse an electron

from the external M shell: a so-called Auger electron. The released energy is

characteristic for the type of atom. Auger electrons are produced in the

outermost surface layer (at nanometer depth) of the sample [8].

3.4 RESULTS AND DISCUSSION

Surface morphological studies were carried out using the scanning electron

microscopy. The SEM images of as-deposited and annealed samples of ZnTe films

Page 13: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

111

are shown in figure 3.5 (A, B, C, D & E). From these images it is clear that, the

surface morphology is different.

Figure 3.5 (A, B, C, D & E) SEM photographs of as-grown ZnTe thin film and

annealed at 323K, 373K, 423K and 473K respectively.

68 nm

Page 14: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

112

3.4.1 For different annealing temperature

The SEM of the as-deposited film (68nm) and thermally annealed films at 323K,

373K, 423K, 473K are displayed in figures 3.5 (A, B, C, D & E) respectively. It was

found that as deposited film at 300K (figure 3.5 A) exhibited a rough surface and

there is no smooth, feature surface that can easily be observed.

However when the films are annealed at higher temperatures, we get the grass

root like structures seen in figure 3.5 (B, C). These films exhibit non-uniform

surfaces with voids. On annealing, voids and porosity of the films are found

to decrease with increase in annealing temperature. Annealing resulted in

the transformation of the as deposited films to one with a morphology

consisting of almost uniform surface. The SEM micrographs also show an

improvement in the structural homogeneity with heat treatment.

The surface morphology of the thin films annealed at 423K and 473K is

presented in figure 3.5 (D, E). For annealed films at 423K and 473K, the SEM

images show uniform surfaces with reduced intergranular spacing. The observed

surface is uniform and randomly distributed small rods are observed over the

surface. This is most probably referred to the foreign particles during annealing

process. It is concluded that the film surfaces become dense, uniform and

compact in nature after post-annealing [9, 10].

3.4.2 For different thickness

The surfaces of the films with different thicknesses were examined by SEM.

Figure 3.6 shows the SEM images of ZnTe films of different thicknesses. All films

are deposited as well as analyzed at room temperature. The SEM morphology of

the film deposited with 10 SILAR cycles having thickness 68nm has discussed in

previous section as figure 3.5 (A). Figures 3.5(A) and 3.6(B) show that the film

compactness was high; free of pinholes. Figure 3.6 (B) shows morphology of

ZnTe thin film for thickness 103nm with initial notches. Initial notches observed

for 103nm is now converted into cracks for thickness 156nm shown in figure 3.6

(C). These cracks increase with increase in film thicknesses due to effects of

evaporation of absorbed water and reorganization of the grain as shown in

figure 3.6 (D, E) for 210nm and 278nm thicknesses.

Page 15: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

113

Figure 3.6 (B, C, D & E) SEM photographs of ZnTe thin film having

thicknesses 103nm, 156nm, 210nm & 278nm respectively.

One possible explanation for this unexpected behavior could be that the ‘‘growth

stresses’’ in these films are of opposite sign leading to compressive strains in the

plane of the substrate and hence causing the band gap to shrink (chapter 5,

section 5.6.1.8 (table 5.2)) rather than expand. However, this is contrary to the

microstructural evidence (chapter 4, section 4.8.2.1(table 4.8)), albeit indirect,

that these films are in less tension: thicker films undergo mud cracking,

indicating that the films are indeed under biaxial tension [11-14].

3.5 ENERGY DISPERSIVE SPECTROSCOPY (EDS/EDAX) ON THE SCANNING

ELECTRON MICROSCOPY (SEM)

The brief introduction and working of energy dispersive spectroscopy

(EDS/EDAX) on the scanning electron microscopy (SEM) are described as follow:

Page 16: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

114

3.5.1 Introduction

In this cursory treatment of the subject, we can divide our task into three major

parts. First, we consider the processes that follow the excitation of the sample by

an electron beam. We are most interested in the process by which x-rays are

emitted, but our efforts will be repaid if we also look at some of the other

interactions that occur. Next, we are interested in the means by which the

emitted x-rays are collected, sorted, and counted. That is, we want to know how

the energetic emissions of an electron-excited sample get translated into

analyzable data. Finally, we look at the analysis techniques themselves [15, 16].

Initial EDS/EDAX analysis usually involves the generation of an X-ray spectrum

from the entire scan area of the SEM. SEM photograph is a secondary electron

image of a specimen and the corresponding X-ray spectra was generated from

the entire scan area. The Y-axis shows the counts (number of X-rays received

and processed by the detector) and the X-axis shows the energy level of those

counts.

3.5.2 Principle of EDAX

The process of x-ray emission is shown schematically in figure 3.7. First, an

electron from, say, a scanning electron microscope, ejects an electron from an

Figure 3.7 X-ray microanalysis is based on electronic transitions between

inner atomic shells. An energetic electron from an electron column

dislodges an orbital electron from a shell of low energy (E1). An electron

from a shell of higher energy subsequently fills the vacancy, losing energy

in the process. The lost energy appears as emitted radiation of energy E2 -

E1.

Page 17: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

115

inner shell of a sample atom. The resulting vacancy is then filled by an electron

from a higher energy shell in the atom. In “dropping” to a state of lower energy,

this vacancy-filling electron must give up some of its energy, which appears in

the form of electromagnetic radiation. The energy of the emitted radiation, then,

is exactly equal to the energy difference between the two electronic levels

involved. Since this energy difference is fairly large for inner shells, the radiation

appears as x-rays.

To complicate matters a bit, there are many energy levels-therefore many

potential vacancy-filling mechanisms-within every atom. As a consequence, even

a sample of pure iron will emit x-rays at many energies.

When excited by electrons of sufficient energy, every element in a sample will

emit a unique and characteristic pattern of x-rays. Furthermore, under given

analysis conditions, the number of x-rays emitted by each element bears a more

or less direct relationship to the concentration of that element.

Converting these x-ray emissions to analyzable data is the job of a series of

electronic components (see figure 3.8), which, in the end, produce a digital

spectrum of the emitted radiation.

Figure 3.8 In energy dispersive microanalysis, each emitted x-ray produces

a charge pulse in a semiconductor detector. This tiny and short-lived

current is converted first into a voltage pulse, then into a digital signal

reflecting the energy of the original x-ray. The digital signal, in turn, adds a

single count to the appropriate channel of a multi-channel analyzer (MCA).

Page 18: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

116

The x-ray photon first creates a charge pulse in a semiconductor detector; the

charge pulse is then converted into a voltage pulse whose amplitude reflects the

signal, which causes one count to be added to the corresponding channel of a

multichannel analyzer. After a time, the accumulated counts from a sample

produce an x-ray spectrum [15, 16].

EDAX makes use of the X-ray spectrum emitted by a solid sample bombarded

with a focused beam of electrons to obtain a localized chemical analysis. All

elements from atomic number 4 (Be) to 92 (U) can be detected in principle, inner

shell of a sample atom. The resulting vacancy is then filled by an electron though

not all instruments are equipped for 'light' elements (Z < 10). Qualitative analysis

involves the identification of the lines in the spectrum and is fairly

straightforward owing to the simplicity of X-ray spectra. Quantitative analysis

(determination of the concentrations of the elements present) entails measuring

line intensities for each element in the sample and for the same elements in

calibration standards of known composition.

3.5.3 Accuracy and sensitivity

X-ray intensities are measured by counting photons and the precision obtainable

is limited by statistical error. For major elements it is usually not difficult to

obtain a precision (defined as 2σ) of better than ± 1% (relative), but the overall

analytical accuracy is commonly nearer ± 2%, owing to other factors such as

uncertainties in the compositions of the standards and errors in the various

corrections which need to be applied to the raw data. As well as producing

characteristic X-ray lines, the bombarding electrons also give rise to a continuous

X-ray spectrum, which limits the detachability of small peaks, owing to the

presence of 'background'.

Using routine procedures, detection limits are typically about 1000 ppm (by

weight) but can be reduced by using long counting times.

3.5.4 Spatial resolution

Spatial resolution is governed by the penetration and spreading of the electron

beam in the specimen (figure 3.9). Since the electrons penetrate an

Page 19: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

117

approximately constant mass, spatial resolution is a function of density. In the

case of silicates (density about 3 g cm-3), the nominal resolution is about 2 µm

under typical conditions, but for quantitative analysis a minimum grain size of

several micrometers is desirable.

Figure 3.9 Simulated trajectories of electrons (energy 20 keV) in Si

(rectangle = 1x2µm).

Better spatial resolution is obtainable with ultra-thin (~100 nm) specimens, in

which the beam does not have the opportunity to spread out so much. Such

specimens can be analyzed in a transmission electron microscope (TEM) with an

X-ray spectrometer attached, also known as an analytical electron microscope, or

AEM.

3.6 X-RAY SPECTROSCOPY

Soon after x-rays were discovered in 1895, it became apparent that x-ray

energies are intimately related to the atomic structure of the substances that

emit them. And since the atomic structure of each chemical element is different,

it follows that each element-when stimulated to do so-emits a different pattern

of x-rays. By the 1920s, these characteristic patterns had been recorded for most

of the elements. Until the late forties, however, analyzing substances by

stimulating and recording their x-ray emissions remained the province of the

research scientist.

Page 20: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

118

3.6.1 Atomic structure

According to Rutherford-Bohr model of the atom, electrons orbit around the

positive nucleus. In the normal state the number of orbital electrons equals the

number of protons in the nucleus (given by the atomic number, Z). Only certain

orbital states with specific energies exist and these are defined by quantum

numbers. With increasing Z, orbits are occupied on the basis of minimum energy,

those nearest the nucleus, and therefore the most tightly bound, being filled first.

Orbital energy is determined mainly by the principal quantum number (n). The

shell closest to the nucleus (n = 1) is known as the K shell; the next is the L shell

(n = 2), then the M shell (n = 3), etc. The L shell is split into three subshells

designated L1, L2 and L3, which have different quantum configurations and

slightly different energies (whereas the K shell is unitary). Similarly, the M shell

has five subshells. This model of the inner structure of the atom is illustrated in

figure 3.10.

The populations of the inner shells are governed by the Pauli Exclusion Principle,

which states that only one electron may possess a given set of quantum numbers.

The maximum population of a shell is thus equal to the number of possible states

possessing the relevant principal quantum number. In the case of the K shell this

is 2, for the L shell 8, and for the M shell 18. Thus for Z ≥ 2 the K shell is full, and

for Z ≥10 the L shell is full.

Electrons occupying outer orbits are usually not directly involved in the

Figure 3.10 Schematic diagram of inner atomic electron shells.

Page 21: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

119

production of X-ray spectra, which are therefore largely unaffected by chemical

bonding etc.

3.6.2 Origin of Characteristic X-rays

'Characteristic' X-rays result from electron transitions between inner orbits,

which are normally full. An electron must first be removed in order to create a

vacancy into which another can 'fall' from an orbit further out. In electron probe

analysis vacancies are produced by electron bombardment, which also applies to

X-ray analysis in the TEM.

X-ray lines are identified by a capital Roman letter indicating the shell containing

the inner vacancy (K, L or M), a Greek letter specifying the group to which the

line belongs in order of decreasing importance α, β, etc.), and a number denoting

the intensity of the line within the group in descending order (1, 2, etc.). Thus the

most intense K line is Kα1 (The less intense Kα2 line is usually not resolved, and

the combined line is designated Kα1, 2 or just Kα). The most intense L line is Lα1.

Because of the splitting of the L shell into three subshells, the L spectrum is more

complicated than the K spectrum and contains at least 12 lines, though many of

these are weak.

Characteristic spectra may be understood by reference to the energy level

diagram (figure 3.11), in which horizontal lines represent the energy of the atom

with an electron removed from the shell (or subshell) concerned. An electron

transition associated with X-ray emission can be considered as the transfer of a

vacancy from one shell to another, the energy of the X-ray photon being equal to

the energy difference between the levels concerned. For example, the Kα line

results from a K-L3 transition (figure 3.11). Energies are measured in electron

volts (eV), 1 eV being the energy corresponding to a change of 1V in the potential

of an electron (= 1.602 x10-19J). This unit is applicable to both X-rays and

electrons. X-ray energies of interest in electron probe analysis are mostly in the

range 1-10keV.

The 'critical excitation energy' (Ec) is the minimum energy which bombarding

electrons (or other particles) must possess in order to create an initial vacancy.

Page 22: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

120

Figure 3.11 Energy level diagram for Ag showing transitions responsible

for main K and L emission lines (arrows show direction of vacancy

movement), energy of emission line indicated in brackets.

Figure 3.12 shows the dependence of Ec on Z for the principal shells. In electron

probe analysis the incident electron energy (E0) must exceed Ec and should

preferably be at least twice Ec to give reasonably high excitation efficiency. For

atomic numbers above about 35 it is usual to change from K to L lines to avoid

the need for excessively high electron beam energy (which has undesirable

implications with respect to the penetration of the electrons in the sample, and in

any case may exceed the maximum available accelerating voltage).

3.6.3 Wavelengths, energies and intensities of X-ray lines

The preceding discussion treated X-rays as photons possessing a specific energy

(E). Sometimes it is more appropriate to describe X-rays by their wavelength (λ),

which is related to energy by the expression:

E.λ=12396

Page 23: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

121

Figure 3.12 Energies of principal characteristic lines (_) and their

excitation energies (----).

where E is in electron volts and λ is in Å, where 1Å = 10-10m.

Since X-ray lines originate in transitions between inner shells, the energy of a

particular line shows a smooth dependence on atomic number, varying

approximately as Z2 (Moseley's law). The energies of the Kα1, Lα1 and Mα1 lines

are plotted against Z in figure 3.12.

The total X-ray intensity for a particular shell is divided between several lines. In

the case of the K shell, more than 80% of the total intensity is contained in the

combined Kα1, 2 line (figure 3.13). The relative intensity of the Kβ line decreases

with decreasing atomic number, in accordance with the electron occupancies of

the relevant energy levels.

3.6.4 The continuous spectrum

Electron bombardment not only produces characteristic X-ray lines resulting

from electron transitions between inner atomic shells but also a continuous X-

ray spectrum or continuum, covering all energies from zero to E0 (the incident

electron energy). This continuum arises from interactions between incident

electrons and atomic nuclei. The intensity of the continuum decreases

monotonically with increasing X-ray energy, and is approximately proportional

to Z.

Page 24: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

122

Figure 3.13 Typical K spectra.

The main significance of the continuum in the present context is that it

contributes the 'backgrounds' upon which characteristic elemental lines are

superimposed.

3.7 ENERGY-DISPERSIVE SPECTROMETER

Energy-dispersive spectrometers (EDSs) employ pulse height analysis, a detector

giving output pulses proportional in height to the X-ray photon energy is used in

conjunction with a pulse height analyzer (in this case a multichannel type). A

solid state detector is used because of its better energy resolution. Incident X-ray

photons cause ionization in the detector, producing an electrical charge, which is

amplified by a sensitive preamplifier located close to the detector. Both detector

and preamplifier are cooled with liquid nitrogen to minimize electronic noise. Si

(Li) or Si drift detectors (SDD) are commonly in use.

3.7.1 Energy resolution

The ED spectrum is displayed in digitized form with the x-axis representing X-

ray energy (usually in channels 10 or 20 eV wide) and the y-axis representing the

number of counts per channel (figure 3.14). An X-ray line (consisting of

effectively mono-energetic photons) is broadened by the response of the system,

producing a Gaussian profile. Energy resolution is defined as the full width of the

peak at half maximum height (FWHM). Conventionally, this is specified for the

Mn Kα peak at 5.89keV. For Si(Li) and SDD detectors, values of 130-150eV are

Page 25: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

123

typical (Ge detectors can achieve 115eV). The resolution of an EDS is about an

order of magnitude worse than that of a WDS, but is good enough to separate the

K lines of neighboring elements (figure 3.14).

3.7.2 Dead time and throughput

In processing the pulses from a solid state detector prior to pulse-height analysis,

it is necessary to use certain integrating time to minimize noise. The system

consequently has a specific 'dead time', or period after the arrival of an X-ray

photon during which the system is unresponsive to further photons. This limits

the rate at which pulses can be processed and added to the recorded spectrum.

‘Throughput' passes through a maximum above which it decreases with further

increases in input count rate. The maximum throughput rate is a function of the

integration time and the design of the system.

Energy resolution is determined partly by the statistics of the detection process

and partly by noise fluctuations in the baseline upon which the pulses are

superimposed. The longer the integration time, more the noise is smoothed out,

and the better the energy resolution. There is thus a 'tradeoff' between

resolution and throughput. Maximum throughput rates have been typically in the

region of 20 000 counts s-1 for Si (Li) and 100 000 counts s-1 and above for SDD.

Figure 3.14 ED spectrum of jadeite (part), showing K peaks of Na, Al and Si.

Page 26: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

124

3.8 QUALITATIVE ANALYSIS

Qualitative analysis is the process of identifying which elements are present in a

sample. As suggested in our discussion of minimum detection limits, qualitative

analysis has as its goal a statement of the form, “Elements X, Y, and Z are

definitely present in the sample; if other elements are present, they must be

present at concentrations less than the minimum detection limit (MDL).” MDLs

must always be kept in mind during qualitative analysis.

3.8.1 Line identification

The object of qualitative analysis is to find what elements are present in an

'unknown' specimen by identifying the lines in the X-ray spectrum using tables of

energies or wavelengths. Ambiguities are rare and can invariably be resolved by

taking into account additional lines as well as the main one.

3.8.2 Qualitative ED analysis

The ED spectrometer is especially useful for qualitative analysis because a

complete spectrum can be obtained very quickly. Aids to identification are

provided, such as facilities for superimposing the positions of the lines of a given

element for comparison with the recorded spectrum (figure 3.15). Owing to the

relatively poor resolution, there are cases where identification may not be

immediately obvious. An example showing unresolved S K and Pb M lines is

given in figure 3.16.

Figure 3.15 Line markers for S K and Ba L lines in the ED spectrum of barite.

Page 27: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

125

3.9 QUANTITATIVE ANALYSIS – EXPERIMENTAL

In its simplest form, qualitative analysis proceeds by determining the energies of

peaks present in the spectrum and comparing them with a chart listing the

known energies of x-ray emissions. Modern analyzers have automated this

process to varying degrees, and most provide markers that can be called to the

video display by atomic number or symbol.

In highly automated versions, software routines detect the location of spectral

peaks, compare them with tabulated energy values, check for inconsistencies (for

example, an apparent Kβ peak but no corresponding Kα), then print out a list of

the elements present.

In general, however, routines of this type are not intended to make sophisticated

judgments, but rather to limit the number of judgments required of the user.

3.9.1 Counting statistics

X-ray intensities are measured by counting pulses generated in the detector by

X-ray photons, which are emitted randomly from the sample. If the mean number

of counts recorded in a given time is n, then the numbers recorded in a series of

discrete measurements form a Gaussian distribution with a standard deviation

(σ) of n1/2/n. A suitable measure of the statistical error in a single measurement

is ±2σ. It follows that 40 000 counts must be collected to obtain a 2σ precision of

± 1% (relative). Such statistical considerations thus dictate the time required to

measure intensities for quantitative analysis.

Figure 3.16 Unresolved peaks in an ED spectrum (PbS sample).

Page 28: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

126

3.9.2 Choice of conditions

The optimum choice of accelerating voltage is determined by the elements

present in the specimen. The accelerating voltage (in kV) should be not less than

twice the highest excitation energy Ec (in keV) of any element presents, in order

to obtain atomic number is commonly Fe, which also has the highest excitation

energy (7.11 keV), hence the accelerating voltage should be at least 15 kV. Line

intensities increase with accelerating voltage, but so does electron penetration,

making spatial resolution worse and increasing the absorption suffered by the

emerging X-rays.

The other important variable selected by the user is beam current. The higher

the current the higher the X-ray intensity, but there are practical limitations.

Some samples are prone to beam damage, which necessitates the use of a low

current. In the case of ED analysis, the limited throughput capability of the

system has to be considered and a current as low as a few nA may be

appropriate.

3.10 QUANTITATIVE ANALYSIS - DATA REDUCTION

Quantitative analysis seeks to establish not only the identities of the elements

present in a sample, but also their concentrations, together with an indication of

the confidence that can be placed in the computed results. Assuming that a

qualitative analysis has been concluded, the quantitative analysis must proceed

through several phases: background removal, deconvolution of overlapped

peaks, and calculation of elemental concentration.

3.10.1 Castaing's approximation

As shown by Castaing (1951), the relative intensity of an X-ray line is

approximately proportional to the mass concentration of the element concerned.

This relationship is due to the fact that the mass of the sample penetrated by the

incident electrons is approximately constant regardless of composition. (The

electrons are decelerated by interactions with bound electrons and the number

of these per atom is equal to the atomic number, which, in turn, is approximately

Page 29: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

127

proportional to the atomic weight). Given this approximation, an 'apparent

concentration' (C') can be derived using the following relationship:

(

) 3.1

where Isp and Ist are the intensities measured for specimen and standard

respectively, and Cst is the concentration of the element concerned in the

standard. To obtain the true concentration, certain corrections are required.

3.10.2 Matrix corrections

'Apparent concentrations' require various corrections which are dependent on

the composition of the matrix within which the element concerned exists. The

first of these corrections arises because the mass penetrated is not strictly

constant, but is affected by differences in the stopping power of different

elements. This is the 'stopping power correction'. The second allows for the loss

of incident electrons, which escape from the surface of the sample after being

defected by target nuclei and thus can make no further contribution to X-ray

production, this is the 'backscattering correction'. The third takes account of

attenuation of the X-rays as they emerge from a finite depth in the sample

('absorption corrections'), and the fourth allows for enhancement in X-ray

intensity, arising from fluorescence by other X-rays generated in the sample

('fluorescence corrections').

The absorption correction is generally the most important. It is dependent on the

angle between the surface of the specimen and the X-ray path to the

spectrometer, or 'X-ray take-off angle (ψ)' (figure 3.17), which is about 40° for

current models of electron probes, although earlier instruments used different

angles. The angle of incidence of the beam is also relevant, standard correction

methods assume normal incidence and modifications are required for non-

normal incidence (as sometimes applies in SEMs).

3.10.3 ZAF corrections

The acronym 'ZAF' describes a procedure in which corrections for atomic

number effects (Z), absorption (A) and fluorescence (F) are calculated separately

Page 30: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

128

Figure 3.17 X-ray source region, with path of X-rays to the spectrometer ψ=

take-off angle).

from suitable physical models. (The atomic number correction encompasses

both the stopping power and backscattering factors). Certain specific methods of

calculating these corrections, developed in the 1960s, constitute the standard

form of the ZAF correction, which is still used quite successfully. Its main

drawback is that the absorption correction is inadequate when the correction is

large; alternative absorption correction procedures are therefore preferable.

3.10.4 Accuracy

For major elements, it is usually possible to obtain a statistical precision of ±1%

relative (2σ). However, various factors apply which limit the accuracy obtainable

in the final result. Instrumental instabilities should be less than 1%, and

uncertainty in standard compositions may be similarly small (though not

always). The largest uncertainties are generally in the matrix corrections,

especially the absorption correction when absorption is severe. Generalization is

difficult, but ±1% is attainable in favorable cases.

3.10.5 Detections limits

With decreasing concentration, statistical errors and uncertainties in

background corrections become dominant. For a concentration in the region of

100 ppm the intensity measured on the peak consists mainly of background. The

smallest detectable peak may be defined as three times the standard deviation of

the background count [17-24].

Page 31: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

129

3.11 EXPERIMENTAL SET UP OF ENERGY DISPERSIVE ANALYSIS OF X-RAY

(EDAX)

As growth of ZnTe thin films depends on various growth conditions, it is

essential to know about the chemical content of the grown thin films. For device

applications of these thin films their chemical content is an important parameter

to look for its influence on electrical properties such as resistivity, carrier

concentration etc. Energy Dispersive Analysis of X-ray (EDAX), a most suitable

technique for this purpose is employed in the present work and hence it is

discussed in brief along with the analysis and conclusions of investigations made

for the material in thin form used in the present thesis.

Energy Dispersive Analysis of X-ray spectroscopy (EDAX) is a chemical analysis

technique used in conjunction with scanning electron microscopy (SEM). The

EDAX technique detects X-rays emitted from the sample during bombardment by

an electron beam to characterize the elemental composition of the analyzed

volume. This technique generates a spectrum in which the peaks correspond to

specific X-ray lines and the elements can be easily identified. Quantitative data

can also be obtained by comparing peak heights or areas in the unknown with a

standard material. Data collection and analysis with EDAX is a relatively quick

and simple process because the complete spectrum of energies is acquired

simultaneously.

The essential feature of analysis is the localized excitation of a small area at the

sample surface by a finely focused electron beam, or probe giving a resolution of

about 1 µm. The energy of the beam is typically in the range 10-20keV. This

causes X-ray to be emitted from the point of the material. The energy of the X-

rays emitted depends on the material under examination. The X-rays are

generated in a region of about 2 microns in depth. By moving the electron beam

across the material an image of each element in the sample can be acquired. In

principle, the concentration of an element could be determined by comparison of

the intensity of a particular characteristic line from the sample to that of a known

standard, usually the pure element, under identical experimental conditions.

Analysis at a point could be carried out, or the specimen could be moved

Page 32: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

130

continuously in one direction while the X-ray output is recorded to give the

distribution of the element.

Figure 3.18 shows an experimental arrangement of EDAX attached to a SEM

(Model: JSM-5610LV Make: JEOL, Japan) used in experiment study. An EDAX

system comprises four basic components that must be designed to work together

to achieve optimum results: the beam source, the X-ray detector, the pulse

processor and the analyzer. X-ray detector measures the relative abundance of

emitted X-rays versus their energy. The detector is typically lithium-drifted

silicon, a solid state surface barrier device. When an incident X-ray strikes the

detector, it creates a charge pulse that is proportional to the energy of X-ray. The

charge pulse is converted to a voltage pulse (which remains proportional to the

X-ray energy) by a charge-sensitive preamplifier. The signal is then sent to a

multichannel analyzer where the pulses are sorted by voltage. The energy, as

determined from the voltage measurement, for each incident X-ray is sent to a

computer for display and further data evaluation. The spectrum of X-ray energy

versus counts is evaluated to determine the elemental composition of the

sampled volume.

Elements of low atomic number are difficult to detect by EDAX. The Si(Li)

detector is often protected by a Beryllium window. The absorption of the

soft X-rays by the Be preclude the detection of elements below an atomic number

of 11 (Na). In windowless systems, elements with as low atomic number as 4

(Be) have been detected, but the problems involved get progressively worse

as the atomic number is reduced [25, 26].

Brief Description:

Scanning Electron Microscope XL 30 ESEM with EDAX: Resolution: upto 20A;

Acc. voltage: 30 kV; Magnification: upto 2, 50, 000x.

Page 33: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

131

Figure 3.18 Experimental set up of Energy Dispersive analysis of X-rays

(ESEM).

Specifications:

Emission current:0 to 200 µA

Accelerating Voltage: 0.2 to 30 kV

Resolution: With LaB6 filament 2nm at 30 kV, With W filament 3.5nm at

30 kV

Magnification:10x to 400000x or higher

Automatic scaled micron marker

Eucentric goniometer stage

Specimen movement: X=50mm,Y=50mm

Rotation: n x 360 degrees

Z movement: 25mm internal & external

Secondary and back scattered electron detectors

3.12 RESULTS AND DISCUSSION

Figure 3.19 and 3.20 show EDAX spectra of ZnTe thin films deposited by SILAR

method with different thicknesses and annealing temperatures respectively. The

weight percentage of the constituent elements obtained from the EDAX of grown

Page 34: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

132

thin films with different thicknesses and annealing temperatures is shown in

table 3.1.

3.12.1 For different thickness

The elemental composition of the ZnTe films deposited at different thicknesses

68nm, 117nm, 155nm, 186nm and 212nm at room temperature (300K) were

evaluated by EDAX technique. The percentage of Zn increases with thickness

while the amount of detected Te changes randomly.

The peak corresponding to silicon, sodium and oxygen shows a transition from L

to the K-shell which can be termed as a K-Alpha peak (Kα). The peaks

corresponding to Zn and Te show three types of transitions from L family i.e.; L-

Alpha (Lα), L-beta (Lβ), L-Gamma (Lγ). These all transitions are shown in the

EDAX patterns of ZnTe thin films of different thickness.

3.13.2 For different annealing temperature

Figure 3.20 shows the EDAX patterns of ZnTe thin films of thickness 68nm at

room temperature (300K) and annealed at 323K, 373K, 423K and 473K

temperature. The peak of Zn decrease with annealing temperature 323K and

373K while increase with 423K and 473K though the amount of detected Te

decreases. This observation is also evidenced by the improvement of hexagonal

ZnTe phase which is revealed from x-ray diffraction analysis [27]. This result is

consistent with x-ray diffraction analysis of the samples with phase correspond

to ZnTe. The peak corresponding to silicon, sodium and oxygen shows a

transition from L to the K-shell which can be termed as a K-Alpha peak (Kα). The

peaks corresponding to Zn and Te show three types of transitions – L-Alpha (Lα),

L-beta (Lβ), L-Gamma (Lγ). All these transitions are shown in the EDAX patterns

of ZnTe thin films.

The results obtained in this research for films deposited at different thicknesses

as well as for different annealing temperatures show that the films are non-

stoichiometric. Generally elements of low atomic number are difficult to detect

by EDAX. So, in the cases of ZnTe films hydrogen is difficult to detect. The

presence of Silicon and Oxygen peaks are due to the glass substrates [28] while

Page 35: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

133

Figure 3.19 EDAX spectra of ZnTe thin films with different thickness grown

by SILAR method.

Page 36: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

134

Figure 3.20 EDAX spectra of ZnTe thin films with different annealing

temperature grown by SILAR method.

Page 37: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

135

the presence of Na is due to the starting material Na2Te. The O2 peak detected

from the EDAX spectrum is unavoidable in any chemically deposited thin film.

This has previously been reported [29].

Table 3.1 Chemical compositions (wt. %) of ZnTe thin films grown by SILAR

method.

Elem

ents

Wt% of elements obtained from EDAX Wt%

o

f elem

ents

calculated

Different thickness Different annealing temperature

68nm 117nm 155nm 186nm 212nm 300K 323K 373K 423K 473K

Zn 51.57 52.06 51.64 51.28 52.85 51.57 50.6 50.62 53.11 56.33 33.89

Te 43.03 42.09 43.35 43.36 42.01 43.03 44.26 44.03 42.77 39.29 66.11

O 1.87 1.12 1.08 1.17 1.14 1.87 0.97 1.02 0.67 0.71

Na 1.24 1.67 1.36 1.33 1.29 1.24 1.03 1.07 0.83 1.08

Si 2.29 3.06 2.57 2.86 2.71 2.29 3.14 3.26 2.62 2.59

3.13 ATOMIC FORCE MICROSCOPY

Atomic force microscopy is currently applied to various environments (air,

liquid, vacuum) and types of materials such as metal semiconductors, soft

biological samples, conductive and non-conductive materials. With this

technique size measurements or even manipulations of nano-objects may be

performed.

3.13.1 Introduction

Atomic Force Microscope (AFM) is one type of scanning probe microscope,

which has an ability to create three dimensional micrograph of sample surface

with resolution down to the nano-meter and angstrom scale. The AFM is one of

the foremost tools for imaging, measuring and manipulating matter at the

nanoscale. AFM is also known as scanning force microscope (SFM) because by

using AFM one can image the surface with atomic resolution and at the same

time one can also measure the force at nano-scale. This force is between the tip

and the sample surface like Vander Waal force with resolution in the range of

few nano-newtons.

The AFM is being increasingly used in solving processing and materials problems

Page 38: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

136

in a wide range of technologies being used in electronics, telecommunications,

biological, chemical, automotive, aerospace and energy industries. The materials

being investigated include thin and thick film coatings, ceramics, composites,

glasses, synthetic and biological membranes, metals, polymers and

semiconductors. The AFM is also being applied to the studies of phenomena such

as abrasion, adhesion, cleaning, corrosion, etching, friction, lubrication, plating

and polishing.

3.13.2 Principle of AFM

Atomic Force Microscopy (AFM) measures the interaction force between the tip

and the surface. The tip may be dragged across the surface, or may vibrate as it

moves.

The interaction force will depend on the nature of the sample, the probe tip and

the distance between them. AFM provides a 3D profile of the surface on a

nanoscale, by measuring forces between a sharp probe (<10 nm) and surface at

very short distance (0.2-10 nm probe-sample separation). The probe is

supported on a flexible cantilever. The AFM tip “gently” touches the surface and

records the small force between the probe and the surface.

3.14.3 Working of an AFM instrument

The probe is placed on the end of a cantilever (which one can think of as a spring

as shown in figure 3.21). The amount of force between the probe and sample is

dependent on the spring constant (stiffness of the cantilever and the distance

Figure 3.21 Spring depiction of cantilever and SEM image of cantilever with probe (tip).

Page 39: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

137

between the probe and the sample surface. This force can be described using

Hooke’s Law:

3.2

Where, F = Force; k = spring constant; x = cantilever deflection.

If the spring constant of cantilever (typically ~ 0.1-1 N/m) is less than surface,

the cantilever bends and the deflection is monitored. This typically results in

forces ranging from nN (10-9 ) to µN (10-6) in the open air.

The motion of the probe across the surface is controlled similarly to the STM

using feedback loop and piezoelectronic scanners (figure 3.22(A, B)). The

primary difference in instrumentation design is how the forces between the

probe and sample surface are monitored. The deflection of the probe is typically

measure by a “beam bounce” method. The beam-bounce method is now widely

used as a result of the excellent work by Alexander and colleagues [30]. In this

system an optical beam is reflected from the mirrored surface on the back side of

the cantilever onto a position-sensitive photodetector. In this arrangement a

small deflection of the cantilever will tilt the reflected beam and change the

position of beam on the photodetector. A third optical system introduced by

Figure 3.22 (A, B) Schematic of AFM instrument showing “beam bounce”

method of detection using a laser and position sensitive photodiode

detector.

A

B

Page 40: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

138

Sarid [31] uses the cantilever as one of the mirrors in the cavity of a diode laser.

Motion of the cantilever has a strong effect on the laser output, and this is

exploited as a motion detector. A semiconductor laser diode is bounced off the

back of the cantilever onto a position sensitive photodiode detector. This

detector measures the bending of cantilever during the tip is scanned over the

sample. The measured cantilever deflections are used to generate a map of the

surface topography.

3.14 MEASUREMENT OF FORCES

The dominant interactions at short probe-sample distances in the AFM are Van

der Waals (VdW) interactions. However long-range interactions (i.e. capillary,

electrostatic, magnetic) are significant further away from the surface. These are

important in other SPM methods of analysis.

Several forces typically contribute to the deflection of an AFM cantilever which

are intermolecular microscopic forces. They are described below:

3.14.1 Intermolecular microscopic interaction

In general, the attractive interaction between two microscopic or macroscopic

bodies is

( )

3.3

where r is the distance between two bodies, C is a constant and n is integer.

Some examples of microscopic interactions are:

3.14.1.1 Coulomb interaction

Coulomb interaction exists between two isolated charged bodies. If the two

charges are q1 and q2 and distance between them is r then the interaction energy

is given by the expression

( )

3.4

This interaction is very strong.

Page 41: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

139

3.14.1.2 Ionic interaction

Ionic interaction results from the interaction between ions. If we add or remove

the electron in a atom then what results is a ion. Interaction between ions with

charge +/- is +/-

. This interaction is attractive between ions of opposite

charge and repulsive between ions of same charge.

3.14.2 Van der Waals interaction

Van der Waals interaction includes following:

3.14.2.1 Van der Waals attractive interaction

Consider two identical inert gas atoms at a separation r in comparison with the

radius of the atoms. Then there is no electrostatic potential so we have no

electrostatic interaction, but atoms have some dipole moment and the atoms

induce dipole moments in each other and the induced moments cause an

attractive interaction between the atoms which may be given as

( )

3.5

This is the Van der Waal’s interaction or the induced dipole – dipole interaction.

It is the principle attractive interaction in crystals of inert gases.

3.14.2.2 Van der Waals repulsive interaction

As the two atoms are brought together their charge distributions gradually

overlap, thereby changing the electrostatic energy of the system. At sufficiently

closed separation, the overlap energy is repulsive in large part because of the

Pauli Exclusion Principle. The statement of this principle is that two electrons

can not have their entire quantum numbers same. When the charge distribution

of two atoms overlaps, there is a tendency of electrons from atom B to occupy in

part state of atom A, which is already occupied by electron of atom A and vice

versa. The Pauli’s principle prevents multiple occupancy and electron

distribution of atoms with closed shell can overlap only if accompanied by the

partial promotion of electrons to the unoccupied high-energy states of the atoms.

Thus the electron overlap increases the total energy of the system and gives a

Page 42: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

140

repulsive contribution to the interaction. This repulsive potential is of the form

of

( )

3.6

where b is a constant. This repulsive interaction is called the Van der Waals

repulsive interaction.

The total potential energy of two atoms at separation r is given by

( ) [(

)

(

)

] 3.7

Where ε and σ are new parameters appear which can be expressed as 4εσ6=a,

4εσ12=b.

This potential is called the Lennard – Jones potential.

3.14.2.3 Ion – dipole interaction

The interaction between ion and dipole is given as

( )

3.8

where q is the charge of the ion, µ is the dipole moment, Ѳ is the angle between

the direction of the dipole moment and the ion.

3.14.2.4 Dipole – dipole interaction

The interaction between the dipoles are given by the relation

( ) [ ]

3.9

The force commonly used with AFM is a Van der Waals force. Force verses

distance (tip to sample) curve is shown in the figure 3.23.

3.15 PRIMARY MODES OF IMAGING

Primary modes of imaging (1) contact mode (2) Intermittent mode and (3) non-

contact mode are described below:

Page 43: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

141

Figure 3.23 Force Vs distance (tip to sample) curve.

3.15.1 Contact Mode AFM (repulsive VdW)

When the spring constant of cantilever is less than surface, cantilever bends. The

force on the tip is repulsive. By maintaining a constant cantilever deflection

(using the feedback loops) the force between the probe and the sample remains

constant and an image of the surface is obtained.

Advantages: fast scanning, good for rough samples, used in friction analysis.

Disadvantages: at time forces can damage/deform soft samples (however

imaging in liquids often resolves this issue).

3.15.2 Intermittent mode (Tapping)

The imaging is similar to contact. However, in this mode the cantilever is

oscillated at its resonant frequency. The probe lightly “taps” on the sample

surface during scanning, contacting the surface at the bottom of its swing. By

maintaining constant oscillation amplitude a constant tip-sample interaction is

maintained and an image of the surface is obtained.

Advantages: allows high resolution of samples that are easily damaged and/or

loosely held to a surface; good for biological samples.

Disadvantages: more challenging to image in liquids, slower scan speeds needed.

Page 44: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

142

3.15.3 Non-contact mode (attractive VdW)

The probe does not contact the sample surface, but oscillates above the adsorbed

fluid layer on the surface during scanning (all samples unless in a controlled UHV

or environmental chamber have some liquid adsorbed on the surface). Using a

feedback loop to monitor changes in the amplitude due to attractive VdW forces

the surface topography can be measured.

Advantages: Very low force exerted on the sample (10-12 N), extended probe

lifetime.

Disadvantages: generally lower resolution; contaminant layer on surface can

interfere with oscillation; usually need ultra-high vacuum (UHV) to have best

imaging.

3.16 FORCE CURVES

Force curves measure the amount of force felt by the cantilever as the probe tip

is brought close to - and even indented into - a sample surface and then pulled

away. In a force curve analysis the probe is repeatedly brought towards the

surface and then retracted, figure 3.24. Force curve analyses can be used to

determine chemical and mechanical properties such as adhesion, elasticity,

Figure 3.24 A typical force curve showing the various probe-sample

interactions.

Page 45: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

143

hardness and rupture bond lengths. The slope of the deflection (C) provides

information on the hardness of a sample. The adhesion (D) provides information

on the interaction between the probe and sample surface as the probe is trying

to break free. Direct measurements of the interactions between molecules and

molecular assemblies can be achieved by functionalizing probes with molecules

of interest.

3.17 SURFACE ROUGHNESS PARAMETERS AND STATISTICAL MEASURES

WITH AFM

Surface roughness can be quantified using a variety of different parameters [32]

that can be divided into two groups. (1) Single values to describe the roughness

and (2) statistical measures of the roughness. Two common single value

parameters for roughness are the arithmetic roughness (Ra) and the rms

roughness (Rrms). Unfortunately, these two parameters only describe the vertical

roughness of a surface. Two surfaces with identical vertical roughness can have

very different surface morphology [35, 36], and a measure of the horizontal

roughness is needed to accurately specify their roughness distinctly [33].

Statistical measures of surface roughness provide a more complete description

of the surface than single valued parameters [32]. Two commonly used statistical

measures of surface roughness are the bearing ratio and the power spectral

density (PSD). The bearing ratio is a useful parameter for characterizing abrasive

wear of the surfaces [32] and the PSD is the power spectrum of the Fourier

transform of the surface profile [32] which provides useful information on the

geometrical structure of the surface. In particular, it is useful for detecting

periodic structure in the profile.

The meanings of all such parameters are discussed below.

3.17.1 Roughness average /arithmetic roughness

In general, the roughness average can be considered as the arithmetical mean

deviation of all the points or it can be defined as the average deviation of

roughness of all points in the profile from a mean line over the evaluation length

[37]. The arithmetic roughness Ra of a surface can be expressed as

Page 46: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

144

∑ | |

3.10

where N is the number of data points of the profile and

3.11

Where are the data points that describe the relative vertical height of the

surface and is the mean height of the surface given by the equation

∑ | |

3.12

3.17.2 Root mean square roughness

Root mean square roughness is defined as “The average of the measured height

deviations taken within the evaluation length and length measured from the

mean line”[38]. It can be mathematically expressed as

3.13

3.17.3 Maximum Profile Peak Height

It is defined as “the height of the maximum/highest high intensity peaks in the

roughness profile over the evaluation length” [37]. The maximum profile peak

height is represented mathematically as

| | 3.14

3.17.4 Maximum Profile Valley Depth or

It is defined as “the depth of the deepest valley in the roughness profile over the

evaluation length” [37] which may be given mathematically as

| | 3.15

3.17.5 Maximum Height of the Profile or

Maximum height of the profile basically is a measure of the maximum peak to

valley height in the roughness profile. The exact definition of it may be given as

“The absolute value between the highest and lowest peaks” [38]. The

Page 47: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

145

mathematical expression used to calculate the parameter is

| | | | 3.16

3.17.6 The Amplitude Distribution Function (ADF)

The amplitude distribution function is a probability function that gives the

probability that a profile of the surface has a certain height z at any position x [39].

3.17.7 The Bearing Ratio Curve (BRC)

The Bearing Ratio Curve is related to the ADF. It is the corresponding cumulative

probability distribution. The bearing ratio curve is always integral (from the top

down) of the ADF [54]. The bearing ratio ( ) is defined as the length of the profile

above a horizontal line through the distribution [32] and is typically shown as a

graph in which the ordinate is the height below the highest peak in the profile.

This makes the bearing ratio curve more accurate reflection of the vertical

roughness of the surface than arithmetic or RMS roughness. The length of the

surface is a measure of the vertical roughness of the surface (of the film) whereas

the shape of the bearing ratio curve is an indication of the topography of the

surface [40]. It finds much greater use in evaluating surface finish.

3.17.8 Power Spectrum

The most important property of rough surfaces is the surface roughness power

spectrum. It determines the contact area between two solids and can provide

both lateral and longitudinal information. Thus it is more informative

measurement than all the statistical quantities.

Surface roughness is often understood as a departure of the roughness

parameters from the planarity. A convenient way of describing surface

roughness is to represent it in terms of profile height z(x, y). For typical digitized

AFM scans, the values of x and y are quantized. To determine PSD, one needs to

transform 2D AFM images from real space to reciprocal space. Two types of

transformations are used by various authors: one is the Fast Hartley transform

(FHT) [35] and the other is the Fast Fourier transform (FFT) [41-47].

Even though the 2D FFT and FHT give us a transformed version of the reciprocal

Page 48: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

146

space, it is still difficult to make use of this 2D information. The standard method

to solve this problem is to extract a 1D magnitude of the 2D transform; that is, a

1D power spectral density is plotted against the spatial frequency. For a typical

2D AFM Scan, it consists of 250 000 (i.e. 512 × 512) data points, but the 1D PSD

requires 512 pieces of information to describe the surface roughness. This

method was proposed independently by Dumas et al. [42] and Strausser et al. [43,

48].

In AFM, one and two-dimensional power spectral density (PSD) are used to

characterize the structure of the surfaces. In the present work, (1D) PSD is

considered, as the simulations used are line scans, not two dimensional images.

The 1D PSD is given by the relation [48],

(

∫ ( ) ( ) )

3.17

where L is the length of the profile and y(x) is the profile.

The power spectral density is advantageous because it allows comparison of the

roughness data taken over various spatial frequency regions. Such methodology

also offers a convenient representation of the direct space periodicity and

amplitude of the roughness.

3.18 ATOMIC FORCE MICROSCOPE

The instrument used to obtain data/images for this work was the Veeco CP II,

which is shown in figure 3.25. The CP-II Scanning Probe Microscope (SPM)

provides high performance and value for material and physical science research.

Brief Description:

The CP-II head, designed to accept an on-axis microscope, utilizes patented Scan

Master closed-loop can linearization to image areas of up to 100 x 100 µm. The

system is designed for ease-of-use with integrated high-magnification color

premounted tips, and switching samples fast and easy. The system’s complete

library of both standard and advanced image modes, and a host of accessories

and hardware options, guarantee application flexibility and make the CP-II the

best performer of any scientific SPM.

Page 49: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

147

Figure 3.25 Scanning probe microscope (AFM).

Specifications:

Available Modes (Contact & Non-Contact AFM, STM, LFM, EFM, MFM,

Nanolithography and Nano-manipulation)

100 µm Scanner- Z-range: 7.5 µm

DAC XY-Resolution: 0.25 Å & Z- Resolution: 0.025Å

5 µm Scanner- Z-range: 2.5 µm;

DAC XY-Resolution: 0.0013 Å & Z- Resolution: 0.009 Å with associated

image analysis software.

3.19 RESULTS AND DISCUSSION

The as-grown thin film of ZnTe was viewed under the non-contact AFM mode.

Silicon nitride tip was used to obtain topographic images of surface of the thin

film. One can also obtain various line analyses spectrums from real images by the

digital Instruments 2SPM controller using di CP-II ProScan 1.9 software.

Figure 3.26 (A, B) shows the 2D and 3D AFM images covering an area of 1µm ×

1µm of as prepared ZnTe thin film of thickness 68 nm by SILAR method at room

temperature (300 K). The images are computer generated and the original data

can be manipulated so that the surface of interest can be viewed from different

directions. Distance scales in the x, y and z directions are marked on the 3D

Page 50: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

148

images and the z scale can be exaggerated as required.

The 3D images (maximum 256 × 256 points) do facilitate over all visualization of

the film surface. An image in a 3D format gives a rendition of what the surface

topography actually looks like. That is, the data is displayed in the x, y and z axis.

Often the scale between the x, y and z axis are not equal. The surface features are

very small with relationship to the x and y dimensions; however, in 3D image

they look large. This figure reflects that the surface possess hilly (also valley)

regions and is not absolutely flat surface [49]. The two dimensional image shows

the x and y axis and color is used to depict the height of image. An AFM image

displayed in the 2D format looks much like an image obtained from a traditional

microscope such as an optical microscope.

To know the surface morphology of ZnTe thin film, the line analysis is carried out

in detail. The line profile is a two dimensional profile or cross section extracted

from an AFM image. This may be taken horizontally, vertically or at obtuse

angles. The results of line analysis for ZnTe thin film is given in table 3.1. A line

Figure 3.26 AFM images of the as-deposited ZnTe film: (A) 2D and (B) 3D

(C) histogram (D) height profile and (E) power spectrum.

Page 51: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

149

trace of the surface profile along the chosen traverse length can be printed out

together with the height profile, histogram, power spectrum, bearing ratio along

with various statistical parameters are mentioned in table 3.1. The line – 1, line –

2, line – 3, line – 4 and line - 5 (shown in fig. 3.15(A)) had drawn on the surface of

the film show indirectly the typical arrangement of the growth features in those

regions. The height profiles taken along the horizontal line – 1 to line - 5 of the

AFM image of ZnTe thin film is shown in table 3.1. This reveals the fact that the

difference between the peak p and the valley z or v values (Rp-v) are around120

nm (maximum) and 84.79 nm (minimum) of line – 4 and line – 3 respectively

indicating film at location of line 4 has maximum peak to valley difference than

at location of line – 3. The rms (Rq) and average (Ra) roughness values

(maximum) for line – 2 are found to be 20.15 nm and 16.69 nm, whereas for line

– 3 these values (minimum) are 13.93 nm and 10.81 nm respectively, indicating

that there is significant difference in average and rms values of peaks across

these lines while for lines 1, 4 and 5, the rms (Rq) and average (Ra) roughness

values are average around 17.70 nm and 13.96 nm respectively, comparable

with lines 2 and 3. It may be noted that these parameters indicate roughness in

vertical direction. Furthermore, the arrangement of atoms as revealed in 3D

image shows peaks and valleys. This type of arrangement of atoms is uniform

from 84.79 nm to 120 nm and is depicted in 3D image of the same film. However

these two single value parameters (i.e. Rrms and Ra), though simple and reliable,

make no distinction between peaks and valleys and do not account for the lateral

distribution of surface features.

Table 3.1 Line analyses of the AFM profiles of as deposited ZnTe film.

Line 1 Line 2 Line 3 Line 4 Line 5 Rp-v, nm 94.01 95.10 84.79 120.0 107.5 Rms Rough (Rq), nm 17.74 20.15 13.39 17.45 17.92 Ave Rough (Ra), nm 13.79 16.69 10.81 12.97 15.13 Mean Ht, nm 50.42 50.46 50.40 50.42 50.46 Median Ht, nm 46.56 46.98 50.71 46.58 47.93 Arc length, µm 5.575 5.500 5.517 5.567 5.298 Bearing [email protected]%, nm 57.24 59.71 56.71 54.83 60.48 Bearing [email protected]%, nm 36.62 33.39 38.34 38.21 33.01 Peak (Rp), nm 57.04 51.47 55.14 79.99 79.23 Valley (Rv), nm -36.98 -43.63 -29.65 -39.98 -28.24

Page 52: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

150

Bearing ratio for line – 3 is more than line – 4 is more than line – 1 is more than

line – 2 is more than line – 5 suggesting horizontal roughness along for line – 3 is

more than line – 4 is more than line – 1 is more than line – 2 is more than line – 5.

The height distributions (Histograms) obtained from 2D AFM image of ZnTe film

is shown in figure 3.26 (c). The histogram is a continuous bar diagram in which

each column represents a height range. The height of each column represents

the number of image pixels which have a height value in a particular range. The

histogram of height distributions in figure 3.26 (c) shows that most of the

corrugation heights on film surface lie between 9nm to 10 nm above the lowest

point within the given area. The histogram indicates that the height distribution

is asymmetric.

The Mean height indicates the central value in the roughness profile over the

evaluation length and is found to be 50.42 nm, while median height is a midpoint

on the roughness profile over the evaluation length such that half of the data fall

above it and half below it which turns out to be 46.58 nm. The horizontal

measurement of roughness n(0) can be studied from height profile (figure 3.26

(D)), which indicates the number of intersections that the profile makes with the

mean line of the profile over a specified distance.

The power spectral density curves corresponding to line – 4 is shown in figure

3.26 (E). The curve indicates that the overall surface of ZnTe film is non-

homogeneous. At higher spatial frequencies the amplitude of the PSD curve

gradually decreases. This PSD curve also provide information across the entire

frequency range of the scanned surface (1µm × 1µm) including waviness (mid

frequency) and form (lower frequency). The peaks in the PSD indicate the

periodicity of the surface. The frequency of each peak gives the length that

defines this periodic surface and the spread in the peak indicates the magnitude

of the deviation from the average value. The presence of pin holes in studied

films is not observed. The surfaces of the films show crack free areas in the

explored section.

Page 53: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

151

3.20 CONCLUSIONS

SEM micrographs of ZnTe thin films show the film to be dense and composed of

largely distributed petite rods on uniform background that was free of pits or

pinholes after annealing treatment are seen in the SEM micrographs. EDAX

suggests that the elemental compositions of the all samples are non-

stoichiometric and impurities of O, Si and Na have been detected. The peak of Zn

decrease with annealing temperature 323K and 373K while increase with 423K

and 473K though the amount of detected Te decreases. The percentage of Zn

increases with thickness while the amount of detected Te changes randomly.

Atomic force microscopy (AFM) measurements were performed to obtain the

surface roughness of as-deposited ZnTe thin film. It is observed clearly that the

film surface is not homogeneous; the RMS roughness varies between 13.39 nm to

20.15 nm. The surface morphology of the as-deposited ZnTe films appears to

change significantly as a function of the thermal annealing temperature. The

microstructural analysis of the film shows that SEM micrographs illustrate the

arising and broadening in cracks with film thickness indicating that the films are

indeed under biaxial tension. Finally, it is concluded that the overall surface of as

prepared ZnTe film investigated in present case is substantiated.

REFERENCES

1. Oatley, C.W., 1982, the early history of the scanning electron

microscope, J. Appl. Phys. 53 (2).

2. Hawkes, P., 2004, Recent advances in electron optics and electron

microscopy. Ann. Fondation Louis de Broglie 29-1.

3. Ruska, E., 1986, the development of the electron microscope and of

electron microscopy. Nobel Lecture.

4. Haguenau, F., Hawkes, P.W., Hutchison, J.L., Satiat-Jeunemaıˆtre, B., Simon,

G.T., Williams, D.B., 2003, key events in the history of electron

microscopy. Microsc. Microanal. 9, 96–138.

5. Goldstein, J., et al., 2003, Scanning Electron Microscopy and X-ray

Microanalysis, 3rd ed. Plenum Press, New York.

Page 54: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

152

6. Breton, P.J., 1999, from microns to nanometers: early landmarks in the

science of SEM imaging, Scanning Microsc. 13-1, 1–6.

7. J. I. Goldstein, H. Yakowitz, 1977, Practical Scanning Electron Microscopy,

Plenum Press.

8. http://www.vcbio.science.ru.nl/en/fesem/eds/print/

9. Atoyan, D., 2006, Characterization of Thin-Film Zinc Telluride on Glass

Prepared by Low-temperature Nanosecond Pused-Laser Deposition,

Master Thesis.

10. Rouleau, C. M., Lowndes, D. H., McCamy, J. W., Budai, J. D., Poker, D. B.,

Geohegan, D. B., Puretzky, A. A., Zhu, S., 1995, Growth of Highly Doped p-

type ZnTe Films by Pulsed Laser Ablation in Molecular Nitrogen Applied

Physics Letter 67: pp. 2545 – 2547.

11. Z. C. Jin, I. Hamberg, and C. G. Granqvist, 1988, J. Appl. Phys. 64, 5117.

12. V. Srikant, V. Sergo, and D. R. Clarke, 1995, J. Am. Ceram. Soc. 78, 1931and

references there in.

13. V. Srikant, V. Sergo, and D. R. Clarke, 1995, J. Am. Ceram. Soc. 78, 1935.

14. V. Srikant, 1995, Ph.D. thesis, University of California, Santa Barbara.

15. J. I. Goldstein, D. E. Newbury, P. Echlin, D. C. Joy, C. Fiori, and E. Lifshin,

1981, Scanning Electron Microscopy and X-Ray Microanalysis (Plenum,

New York).

16. K.F. J. Heinrich, 1981, Electron Beam X-Ray Microanalysis (Van Nostrand

Reinhold, New York).

17. Agarwal B.K., 1991, X-ray Spectroscopy, 2nd edn, Springer-verlag, Berlin.

18. Goldstein, J. I., et al., 2003, Scanning Electron Microscopy and X-ray

Micronalysis, 3rd ed Plenum Press, New York.

19. Heinrich, K.F.J., 1981, Electron Beam X-ray Microanalysis, Van Nostrand

Rheinhold, New York.

20. Heinrich, K.F.J. and Newbury, D .E. (eds), 1991, Electron Probe

Quantitation, Plenum Press, New York.

21. Reed, S.J.B., 1993, Electron Microprobe Analysis, 2nd edn. Cambridge

University Press, Cambridge.

22. Reimer, L., 1985, Scanning Electron Microscopy, Springer-Verlag, Berlin.

Page 55: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

153

23. Russ, J. C., 1984, Fundamentals of Energy Dispersive X-ray Analysis,

Butterworths. London.

24. Scott, V. D. and Love, G., 1994, Quantitative Electron Probe Microanalysis,

2nd edn. Ellis Horwood, Chichester.

25. J. P. Sibilia, 1998, a Guide to Materials Characterization and Chemical

Analysis, VCH Publishers.

26. D.Brondon and W.D.Kaplan, 1999, Micro structural Characterization of

Materials, John-Wileyand Sons.

27. Q.J.Feng, D.Z.Shen, J.Y.Zhang, C.X.Shan, Y.M.Lu, Y.C.Liu, X.W.Fan, 2004,

Journal of Magnetism and Magnetic Materials Vol. 279, p. 435.

28. Prabahar S. and Dhanam M., 2005, CdS thin films from two different

chemical baths-structural and optical analysis‟, J. Crys. Growth, Vol. 285,

pp. 41- 48.

29. C.D.Lokhande, B.R.Sankapal, R.S.Mane, H.M. Pathan, M.uller, M.Giersig and

V. Ganesan, 2002, Appl. Sur. Sci., 193, 1.

30. Martin A. Hubbe, Joel J. Pawlak and Alexander A. Koukoulas, 2008,

“Paper’s appearance: A review,” BioResources 3(2), 627-665.

31. Dror Sarid, Dorian Kermisch, 1979, a practical integrated optics device,

Proc. SPIE 0176, Guided Wave Optical Systems and Devices II, 41.

32. B. Nowicki, 1985, Multiparameter representation of surface

roughness, Wear 102, pp. 161-176.

33. J. M. Bennett and J. H. Dancy, 1981, Stylus profiling instrument for

measuring statistical properties of smooth optical surfaces, Applied

Optics,Vol. 20, Issue 10, pp. 1785-1802 .

34. J. C. Stover, 1990, Optical Scattering, Measurements and Analysis

(McGraw- Hill, New York.

35. S. D. Halepete, H. C. Lin, S. J. Fang and C. R. Helms, 1995, Mater. Res. Soc.

Symp. Proc.,386, 383.

36. S. J. Fang, W. Chen, T. Yamanaka and C. R. Helms, 1996, Appl. Phys. Lett.,

68, 2837.

37. Standards: ASME B46.1-1995, ASME B46.1-1985, ISO 4287-1997, ISO

4287/1-1997.

38. Standards: ASME B46.1-1995, ISO 4287-1997.

Page 56: 100 - Shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/25230/9/09... · 2018-07-09 · 100 3.1 INTRODUCTION Scanning Tunneling Microscope was developed first in 1982 by Binning,

154

39. Standards: ISO 4287-1997.

40. K. L. Westra, D. J. Thomson, 1995, J. Vac. Sci. Technol. B13(2), 344.

41. C. R. Helms, 1996, In Semiconductors Characterization: Present Status

and Future Needs, edited by W. M. Bullis, D.G. Seiler, A.C. Diebold, p.110.

42. Ph. Dumas, B. Bouffakhreddine, C. Amra, O. Vatel,, E. Andre, R. Galindo and

F. Salvan, 1993, Europhys. Lett., 22, 717.

43. Y. E. Strausser, B. Doris, A. C. Diebold and H.R. Huff, 1994, Extended

Abstract of 185th Electrochem. Soc. Meeting, No., 284, 461.

44. R. M. Feenstra M.A. Lutz, F. stern, K. Ismail, P.M. Mooney, F. K. LeGoues, C.

Stanis, J. O. Chu and B. S. Meyerson, 1995, J. Vac. Sci. Technol. B13, 1608.

45. T. Yoshinobu, A. Iwamoto, Iwasaki, 1994, Jpn. J. Appl. Phys. Part. 1(33).

46. W. M. Bulliks, 1996, Micro., 14, 47.

47. D. Courboin A. Grouillet and E. Andre, 1995, Surf. Sci., 342, L1111.

48. Y. E. Strausser, 1994, Industrial Applications of Scanned Probe

Microscopy, National Institute of Standards and Technology, Gaitherburg,

MD, P.25.

49. S. E. Shaheen, C. J. Brabec, N. Serdar Sariciftci , F Pandinger, T. Fromherz

and J. C. Hummelen, 2001, Appl. Phys. Lett. 78(6), 841.