NOFLO FAULT<0800>
WAIT A REASONABLE
PERIOD FOR FAULTTO CLEAR
HAS TIMEPASSED?
FAULTCLEARED??
NO
T EN
OU
GH
YES
YESDONE
NO
VIEW & COMPAREPRE, AMB, & VAC
READINGS.
P≈A-1V≈A-3
?
PUMPOK
YES
NO INSUFFVACUUM (V>A-3;
V≈A)?
YES
NO
ISOLATE PUMP. DO HYDROPHOBIC
PINCH TEST &CHECK “P” VALUE
ISOLATEDPUMP DRAW OK
(P≈A-3)?
YES
NO BADPUMP
CHECK EXH PORTFOR BLOCKAGE
1ST BEFOREREPLACING PUMP
DONE
INDICATIONS ARERESTRICTED FLOW
OR HIGH FLOWCONDITION
LARGE NEG ΔP(P≈A-3)
?
SMALL/NO ΔP(A≈P≈V)
?
YES
NO
CALLSUPPORT
YES
NO
CLOGGEDHYDROPHOBIC?
YES
NO
REPLACEHYDROPHOBIC
FILTER
DONE
LINERESTRICTION?
YES
NO
FIX CLOG OR REPLACE
EOL FILTERDONE
BADSOL ORMAN
MAGTESTER LIGHTS
ON SOL IN ZONEHOLD?
YES
NO
SOL PCB OKAY.CLEAN/REPLACE
MANIFOLDDONE
REPLACESOLENOID PCB.
CHECK MANIFOLDAFTERWARDS.
DONE
CALLSUPPORT
LIKELYPCB
POWER DOWN.DISCONNECT SOL
CONNECTORS.RESTORE POWER.
REALLY LOWP, V, & A VALUES
(≈10, 9, 11)?
YES
NO - ABOUTSAME AS BEFORE
SOLENOID(S)POSSIBLY
STUCK OPEN.CLEAN/REPLACE
MANIFOLD
DONE
1. PATIENCE
ANYPHANTOM
MANIFOLDS?
YES
NO
FIXCONFIGURATION
SETTINGS
2. PHANTOM MANIFOLDS
3. PUMP
4. BASIC CLOGS/RESTRICTIONS 7. MANIFOLD AND/OR SOLENOID PCB
Rev 003NOV016
Multi ZoneNo Flow Zone FaultQuick Diagnosis Flow Chart
5. CLOGGED HYDROPHOBIC FILTER
6. LINE RESTRICTIONS