NOFLO FAULT <0800> WAIT A REASONABLE PERIOD FOR FAULT TO CLEAR HAS TIME PASSED? FAULT CLEARED?? NOT ENOUGH YES YES DONE NO VIEW & COMPARE PRE, AMB, & VAC READINGS. PA-1 VA-3 ? PUMP OK YES NO INSUFF VACUUM (V>A-3; VA)? YES NO ISOLATE PUMP. DO HYDROPHOBIC PINCH TEST & CHECK “P” VALUE ISOLATED PUMP DRAW OK (PA-3) ? YES NO BAD PUMP CHECK EXH PORT FOR BLOCKAGE 1ST BEFORE REPLACING PUMP DONE INDICATIONS ARE RESTRICTED FLOW OR HIGH FLOW CONDITION LARGE NEG ΔP (PA-3) ? SMALL/NO ΔP (APV) ? YES NO CALL SUPPORT YES NO CLOGGED HYDROPHOBIC? YES NO REPLACE HYDROPHOBIC FILTER DONE LINE RESTRICTION? YES NO FIX CLOG OR REPLACE EOL FILTER DONE BAD SOL OR MAN MAG TESTER LIGHTS ON SOL IN ZONE HOLD? YES NO SOL PCB OKAY. CLEAN/REPLACE MANIFOLD DONE REPLACE SOLENOID PCB. CHECK MANIFOLD AFTERWARDS. DONE CALL SUPPORT LIKELY PCB POWER DOWN. DISCONNECT SOL CONNECTORS. RESTORE POWER. REALLY LOW P, V, & A VALUES (10, 9, 11) ? YES NO - ABOUT SAME AS BEFORE SOLENOID(S) POSSIBLY STUCK OPEN. CLEAN/REPLACE MANIFOLD DONE 1. PATIENCE ANY PHANTOM MANIFOLDS ? YES NO FIX CONFIGURATION SETTINGS 2. PHANTOM MANIFOLDS 3. PUMP 4. BASIC CLOGS/RESTRICTIONS 7. MANIFOLD AND/OR SOLENOID PCB Rev 0 03NOV016 Multi Zone No Flow Zone Fault Quick Diagnosis Flow Chart 5. CLOGGED HYDROPHOBIC FILTER 6. LINE RESTRICTIONS

1. PATIENCE 2. PHANTOM MANIFOLDS WAIT A …...FIX CONFIGURATION SETTINGS 2. PHANTOM MANIFOLDS 3. PUMP 4. BASIC CLOGS/RESTRICTIONS 7. MANIFOLD AND/OR SOLENOID PCB Rev 0 03NOV016 Multi

  • Upload
    others

  • View
    8

  • Download
    0

Embed Size (px)

Citation preview

Page 1: 1. PATIENCE 2. PHANTOM MANIFOLDS WAIT A …...FIX CONFIGURATION SETTINGS 2. PHANTOM MANIFOLDS 3. PUMP 4. BASIC CLOGS/RESTRICTIONS 7. MANIFOLD AND/OR SOLENOID PCB Rev 0 03NOV016 Multi

NOFLO FAULT<0800>

WAIT A REASONABLE

PERIOD FOR FAULTTO CLEAR

HAS TIMEPASSED?

FAULTCLEARED??

NO

T EN

OU

GH

YES

YESDONE

NO

VIEW & COMPAREPRE, AMB, & VAC

READINGS.

P≈A-1V≈A-3

?

PUMPOK

YES

NO INSUFFVACUUM (V>A-3;

V≈A)?

YES

NO

ISOLATE PUMP. DO HYDROPHOBIC

PINCH TEST &CHECK “P” VALUE

ISOLATEDPUMP DRAW OK

(P≈A-3)?

YES

NO BADPUMP

CHECK EXH PORTFOR BLOCKAGE

1ST BEFOREREPLACING PUMP

DONE

INDICATIONS ARERESTRICTED FLOW

OR HIGH FLOWCONDITION

LARGE NEG ΔP(P≈A-3)

?

SMALL/NO ΔP(A≈P≈V)

?

YES

NO

CALLSUPPORT

YES

NO

CLOGGEDHYDROPHOBIC?

YES

NO

REPLACEHYDROPHOBIC

FILTER

DONE

LINERESTRICTION?

YES

NO

FIX CLOG OR REPLACE

EOL FILTERDONE

BADSOL ORMAN

MAGTESTER LIGHTS

ON SOL IN ZONEHOLD?

YES

NO

SOL PCB OKAY.CLEAN/REPLACE

MANIFOLDDONE

REPLACESOLENOID PCB.

CHECK MANIFOLDAFTERWARDS.

DONE

CALLSUPPORT

LIKELYPCB

POWER DOWN.DISCONNECT SOL

CONNECTORS.RESTORE POWER.

REALLY LOWP, V, & A VALUES

(≈10, 9, 11)?

YES

NO - ABOUTSAME AS BEFORE

SOLENOID(S)POSSIBLY

STUCK OPEN.CLEAN/REPLACE

MANIFOLD

DONE

1. PATIENCE

ANYPHANTOM

MANIFOLDS?

YES

NO

FIXCONFIGURATION

SETTINGS

2. PHANTOM MANIFOLDS

3. PUMP

4. BASIC CLOGS/RESTRICTIONS 7. MANIFOLD AND/OR SOLENOID PCB

Rev 003NOV016

Multi ZoneNo Flow Zone FaultQuick Diagnosis Flow Chart

5. CLOGGED HYDROPHOBIC FILTER

6. LINE RESTRICTIONS