Upload
kale-truslow
View
213
Download
0
Tags:
Embed Size (px)
Citation preview
1
Management of Laboratory Equipment
Management of Laboratory Equipment
2
Purchasing & Inventory
AssessmentOccurrence
Management
Information Management
Process Improvement
Customer Service
Facilities & Safety
The Quality SystemThe Quality System
Organization Personnel
Documents & Records
Equipment
Process Control (QC & EQA) &
Specimen Management
3
Problem Scenario #1Problem Scenario #1
• You’re in the middle of performing ELISAs and the reader fails. You don’t have a documented procedure for troubleshooting, the maintenance log has not been updated for 2 years, and the manufacturer’s instructions are missing.
• What should you do? Why?
• You’re in the middle of performing ELISAs and the reader fails. You don’t have a documented procedure for troubleshooting, the maintenance log has not been updated for 2 years, and the manufacturer’s instructions are missing.
• What should you do? Why?
4
Functioning Equipment is Vital for: Functioning Equipment is Vital for:
• Producing reliable test results• Minimizing instrument breakdown • Lowering repair costs• Preventing delays in reporting test results• Maintaining productivity
• Producing reliable test results• Minimizing instrument breakdown • Lowering repair costs• Preventing delays in reporting test results• Maintaining productivity
5
A Good Equipment Program Achieves A Good Equipment Program Achieves
• Maintaining a high level of performance• Lengthening instrument life• Reducing interruption of services due to
breakdowns and failures• Improving customer satisfaction• Improving the technologist’s confidence and
knowledge
• Maintaining a high level of performance• Lengthening instrument life• Reducing interruption of services due to
breakdowns and failures• Improving customer satisfaction• Improving the technologist’s confidence and
knowledge
6
Basic Laboratory EquipmentBasic Laboratory Equipment
• Microscope• Balance/scale• Centrifuge• Micropipette• Spectrophotometer • Refrigerator• Freezer
• Microscope• Balance/scale• Centrifuge• Micropipette• Spectrophotometer • Refrigerator• Freezer
• Autoclave• Hot air oven• Incubator• pH Meter• Water bath• Washer• Shaker / rotator• Vortex
• Autoclave• Hot air oven• Incubator• pH Meter• Water bath• Washer• Shaker / rotator• Vortex
7
Equipment Management Equipment Management
• Selection• Acquisition• Installation• Calibration / Validation• Maintenance • Troubleshooting • Service and repair• Retiring equipment / disposition
• Selection• Acquisition• Installation• Calibration / Validation• Maintenance • Troubleshooting • Service and repair• Retiring equipment / disposition
8
Equipment Management OversightEquipment Management Oversight• Assign responsibilities for all activities
• Train all personnel on equipment management requirements and responsibilities
• Monitor equipment management activities by: Routinely reviewing all records Ensuring all procedures are followed Updating procedures, if necessary
• Assign responsibilities for all activities
• Train all personnel on equipment management requirements and responsibilities
• Monitor equipment management activities by: Routinely reviewing all records Ensuring all procedures are followed Updating procedures, if necessary
9
Equipment ManagementEquipment Management• Selection• Acquisition• Installation• Calibration / Validation• Maintenance • Troubleshooting • Service and repair• Retiring equipment / disposition
• Selection• Acquisition• Installation• Calibration / Validation• Maintenance • Troubleshooting • Service and repair• Retiring equipment / disposition
10
Selection CriteriaSelection Criteria
• Use Matching equipment with service provided
• Performance characteristics• Facility requirements• Cost• Supply of reagents• Ease of operation• Warranty• Availability of manufacturer technical support• Service Contracts
• Use Matching equipment with service provided
• Performance characteristics• Facility requirements• Cost• Supply of reagents• Ease of operation• Warranty• Availability of manufacturer technical support• Service Contracts
11
Selection CriteriaSelection Criteria
• Location in the laboratory available space, accessibility
• Safety
• Location in the laboratory available space, accessibility
• Safety
12
Equipment ManagementEquipment Management• Selection• Acquisition• Installation• Calibration / Validation• Maintenance • Troubleshooting • Service and repair• Retiring equipment / disposition
• Selection• Acquisition• Installation• Calibration / Validation• Maintenance • Troubleshooting • Service and repair• Retiring equipment / disposition
13
Acquiring EquipmentAcquiring Equipment
• Purchase, Lease, or Rent Central acquisition Bulk procurement
• Donor provided• Conditions of contract
• Purchase, Lease, or Rent Central acquisition Bulk procurement
• Donor provided• Conditions of contract
14
Acquiring EquipmentAcquiring Equipment
• Contract considerations: Parts Manual Installation Operators’ Manual Trial period Contents of service contracts
• Contract considerations: Parts Manual Installation Operators’ Manual Trial period Contents of service contracts
15
Equipment ManagementEquipment Management
• Selection• Acquisition• Installation• Calibration / Validation• Maintenance • Troubleshooting • Service and repair• Retiring equipment / disposition
• Selection• Acquisition• Installation• Calibration / Validation• Maintenance • Troubleshooting • Service and repair• Retiring equipment / disposition
16
Installation ChecklistInstallation Checklist
• Prior to installation: verify physical requirements have been met
Safety checks, electrical, space, ventilation, water supply, ambient temperature, etc.
confirm responsibility for installation
• Prior to installation: verify physical requirements have been met
Safety checks, electrical, space, ventilation, water supply, ambient temperature, etc.
confirm responsibility for installation
17
Installation Installation
• Upon receipt: verify package contents do not attempt to use prior to proper
installation • If required, ensure the equipment is installed
by the manufacturer
• Upon receipt: verify package contents do not attempt to use prior to proper
installation • If required, ensure the equipment is installed
by the manufacturer
18
InstallationInstallation
• After installation Establish inventory record Define conditions Develop and implement protocols for
calibration, performance verification, and operating procedures
Establish maintenance program Provide training for all operators
• After installation Establish inventory record Define conditions Develop and implement protocols for
calibration, performance verification, and operating procedures
Establish maintenance program Provide training for all operators
19
Equipment ManagementEquipment Management• Selection• Acquisition• Installation• Calibration / Validation• Maintenance • Troubleshooting • Service and repair• Retiring equipment / disposition
• Selection• Acquisition• Installation• Calibration / Validation• Maintenance • Troubleshooting • Service and repair• Retiring equipment / disposition
20
Equipment CalibrationEquipment Calibration
• Perform initial calibration Calibrators or standards Follow manufacturer’s instructions
• Determine frequency of routine calibrations
• Perform initial calibration Calibrators or standards Follow manufacturer’s instructions
• Determine frequency of routine calibrations
21
Performance ValidationPerformance Validation
• Validate the performance of new equipment prior to use Test known samples, analyze data Establish stability/uniformity temperature
controlled equipment Check accuracy/precision for pipettors Check centrifuge rpms
• Validate the performance of new equipment prior to use Test known samples, analyze data Establish stability/uniformity temperature
controlled equipment Check accuracy/precision for pipettors Check centrifuge rpms
22
Equipment ManagementEquipment Management• Selection• Acquisition• Installation• Calibration / Validation• Maintenance • Troubleshooting • Service and repair• Retiring equipment / disposition
• Selection• Acquisition• Installation• Calibration / Validation• Maintenance • Troubleshooting • Service and repair• Retiring equipment / disposition
23
Maintenance ProgramMaintenance Program• Systematic and routine cleaning, adjustment, or
replacement of instrument and equipment parts Performed periodically, daily, weekly, monthly
Example:• Cleaning optical lenses• Thermostat adjustments• Changing motor brushes
• Systematic and routine cleaning, adjustment, or replacement of instrument and equipment parts Performed periodically, daily, weekly, monthly
Example:• Cleaning optical lenses• Thermostat adjustments• Changing motor brushes
24
Function ChecksFunction Checks
• Monitoring of instrument parameters to verify that that your equipment is working according to the manufacturer’s specification
• Performed periodically- daily, weekly, monthly• Performed after major instrument repair Example:
Daily monitoring of temperatures Checking wavelength calibration Checking autoclave indicator paper
• Monitoring of instrument parameters to verify that that your equipment is working according to the manufacturer’s specification
• Performed periodically- daily, weekly, monthly• Performed after major instrument repair Example:
Daily monitoring of temperatures Checking wavelength calibration Checking autoclave indicator paper
25
Implementing a Maintenance ProgramImplementing a Maintenance Program
• Assign responsibility• Oversight of all laboratory equipment• Individual responsibilities
• Develop written policies and procedures• Train staff• Keep records
• Assign responsibility• Oversight of all laboratory equipment• Individual responsibilities
• Develop written policies and procedures• Train staff• Keep records
26
Implementing a Maintenance ProgramImplementing a Maintenance Program
• Create a record for equipment inventory• Name, Model #, Serial #• Location in lab• Date purchased• Manufacturer and vendor contact
information• Warranty, expiration date• Spare parts
• Create a record for equipment inventory• Name, Model #, Serial #• Location in lab• Date purchased• Manufacturer and vendor contact
information• Warranty, expiration date• Spare parts
27
Implementing a Maintenance Program Implementing a Maintenance Program
• For each piece of equipment: Establish routine maintenance plan Establish required function checks Develop a list of spare parts
• For each piece of equipment: Establish routine maintenance plan Establish required function checks Develop a list of spare parts
28
Implementing a Maintenance Program: Spare Parts
Implementing a Maintenance Program: Spare Parts
• Establish and maintain an inventory of most frequently used spare parts. Include in record of inventory:
Spare parts per equipment Part number Average use Minimal # of items to be stored Cost and date of ordering Dates of entry and issuance of part from inventory
stock Balance of items remaining in inventory
• Establish and maintain an inventory of most frequently used spare parts. Include in record of inventory:
Spare parts per equipment Part number Average use Minimal # of items to be stored Cost and date of ordering Dates of entry and issuance of part from inventory
stock Balance of items remaining in inventory
29
Implementing a Maintenance Program: Documents
Implementing a Maintenance Program: Documents
• Develop written procedures for all equipment Concise step-by-step instructions for
performing maintenance and function checks Include guide for troubleshooting
• Establish maintenance records to track: function checks and routine maintenance calibration manufacturer’s service
• Develop written procedures for all equipment Concise step-by-step instructions for
performing maintenance and function checks Include guide for troubleshooting
• Establish maintenance records to track: function checks and routine maintenance calibration manufacturer’s service
30
Sample RecordsSample Records
• Charts• Logs• Checklists• Graphs• Service Reports
• Charts• Logs• Checklists• Graphs• Service Reports
31
32
Equipment ManagementEquipment Management• Selection• Acquisition• Installation• Calibration / Validation• Maintenance • Troubleshooting • Service and repair• Retiring equipment / disposition
• Selection• Acquisition• Installation• Calibration / Validation• Maintenance • Troubleshooting • Service and repair• Retiring equipment / disposition
33
TroubleshootingTroubleshooting
• Check manufacturer’s instructions• Determine source of problem
Sample problem Reagent problem Equipment problem Check electrical supply Check water supply
• Make one change at a time
• Check manufacturer’s instructions• Determine source of problem
Sample problem Reagent problem Equipment problem Check electrical supply Check water supply
• Make one change at a time
34
When In-house efforts fail:When In-house efforts fail:
• Do NOT use equipment that does not function properly
• Options for testing: Refer to nearby laboratory Obtain backup instrument from central stores Store samples appropriately
• Seek help from manufacturer or other technical expert
• Place a malfunction notice on equipment
• Do NOT use equipment that does not function properly
• Options for testing: Refer to nearby laboratory Obtain backup instrument from central stores Store samples appropriately
• Seek help from manufacturer or other technical expert
• Place a malfunction notice on equipment
35
Equipment ManagementEquipment Management• Selection• Acquisition• Installation• Calibration / Validation• Maintenance • Troubleshooting • Service and repair• Retiring equipment / disposition
• Selection• Acquisition• Installation• Calibration / Validation• Maintenance • Troubleshooting • Service and repair• Retiring equipment / disposition
36
Service and RepairService and Repair
• Schedule service that must be periodically performed by the manufacturer
• Options Centrally service small equipment, e.g.,
microscopes, washers, pipettes Team of biomedical service technicians
• Schedule service that must be periodically performed by the manufacturer
• Options Centrally service small equipment, e.g.,
microscopes, washers, pipettes Team of biomedical service technicians
37
Equipment DocumentationEquipment Documentation
• Develop a problem log record for each piece of equipment
• Date problem occurred, removed from service
• Reason for breakdown or failure• Corrective action taken• Date returned to use• Change in maintenance or function checks
• Develop a problem log record for each piece of equipment
• Date problem occurred, removed from service
• Reason for breakdown or failure• Corrective action taken• Date returned to use• Change in maintenance or function checks
38
Equipment ManagementEquipment Management• Selection• Acquisition• Installation• Calibration / Validation• Maintenance • Troubleshooting • Service and repair• Retiring equipment / disposition
• Selection• Acquisition• Installation• Calibration / Validation• Maintenance • Troubleshooting • Service and repair• Retiring equipment / disposition
39
40
Retiring Equipment / DispositionRetiring Equipment / Disposition• When?
When experts indicate not repairable Outmoded, will replace with new equipment
• Why? Prevent inaccurate test results Free up valuable space Hazardous
• How? Salvage any useable parts Consider biohazard, follow safety disposal procedures
• When? When experts indicate not repairable Outmoded, will replace with new equipment
• Why? Prevent inaccurate test results Free up valuable space Hazardous
• How? Salvage any useable parts Consider biohazard, follow safety disposal procedures
41
Benefits of a Maintenance ProgramBenefits of a Maintenance Program
• Safety• Fewer interruptions of work• Lower repair costs• Elimination of premature replacement• Less standby equipment• Identification of high maintenance cost• Reduction of variation in test results• Greater confidence in the reliability of results
• Safety• Fewer interruptions of work• Lower repair costs• Elimination of premature replacement• Less standby equipment• Identification of high maintenance cost• Reduction of variation in test results• Greater confidence in the reliability of results
4242
Function Checks Verify that Equipment is Working Properly
Function Checks Verify that Equipment is Working Properly
Performed routinely Daily, weekly, monthly After adjustment or repair
Example: Monitoring refrigerator temperatures Verifying pipette accuracy Checking centrifuge speed
Performed routinely Daily, weekly, monthly After adjustment or repair
Example: Monitoring refrigerator temperatures Verifying pipette accuracy Checking centrifuge speed
4343
Refrigerator and Freezer: Use and Care
Refrigerator and Freezer: Use and Care
• Keep organized• Periodically clean inside and outside• Ensure door is completed sealed when
closing• CAUTION! – DO NOT store food items or
beverages in laboratory refrigerator or freezer
• Keep organized• Periodically clean inside and outside• Ensure door is completed sealed when
closing• CAUTION! – DO NOT store food items or
beverages in laboratory refrigerator or freezer
4444
Refrigerator & Freezer:Temperature Checks
Refrigerator & Freezer:Temperature Checks
Monitor daily• Refrigerator: 2oC to 8oC • Freezer: -20oC, -40oC, or -80oC
Monitor daily• Refrigerator: 2oC to 8oC • Freezer: -20oC, -40oC, or -80oC
4545
Refrigerator & Freezer:Temperature LogRefrigerator & Freezer:Temperature Log
4646
Types of PipettesTypes of Pipettes• Precision pipettes (Not disposable)
Precise and accurate volumes (e.g., 50 µl for Determine) Use disposable, single-use, pipette tips
• Graduated plastic bulb pipettes (Disposable) Dispenses approximate volume Easy to use
• Precision pipettes (Not disposable) Precise and accurate volumes (e.g., 50 µl for Determine) Use disposable, single-use, pipette tips
• Graduated plastic bulb pipettes (Disposable) Dispenses approximate volume Easy to use
4747
Pipette: Use and CarePipette: Use and Care
• Select the appropriate pipette for the volume required
• Ensure that the pipettor, tips, and specimen are at the same temperature
• Firmly attach tip
• Select the appropriate pipette for the volume required
• Ensure that the pipettor, tips, and specimen are at the same temperature
• Firmly attach tip
4848
Pipette: Use and Care (Cont’d)Pipette: Use and Care (Cont’d)
• Hold the pipette vertically when aspirating• Place tip just below the sample• Avoid air bubbles• Discard contaminated tips in appropriate
container after completion of task
• Hold the pipette vertically when aspirating• Place tip just below the sample• Avoid air bubbles• Discard contaminated tips in appropriate
container after completion of taskDO NOT RE-USE- Pipette tips - Graduated plastic bulb pipettes
4949
Precision Pipettes Require Performance Checks
Precision Pipettes Require Performance Checks
• Performed periodically• Required supplies:
Pipette Pipette tips Analytical balance Weigh boats Distilled or deionzed water
• Performed periodically• Required supplies:
Pipette Pipette tips Analytical balance Weigh boats Distilled or deionzed water
5050
1. Pipette a series of 10 samples into a weigh boat on an analytical scale
2. Record weight of each sample to calculate calibration results3. Verify calculated results are within limits
Range Max/Min Accuracy 10 µl + 1.0 µl 10% 100 µl + 10.0 µ l 10% 200 µl + 20.0 µ l 10 %
4. If the results are not within limits, remove from service until appropriate adjustment can be made
5. Decontaminate pipette and scale after use
1. Pipette a series of 10 samples into a weigh boat on an analytical scale
2. Record weight of each sample to calculate calibration results3. Verify calculated results are within limits
Range Max/Min Accuracy 10 µl + 1.0 µl 10% 100 µl + 10.0 µ l 10% 200 µl + 20.0 µ l 10 %
4. If the results are not within limits, remove from service until appropriate adjustment can be made
5. Decontaminate pipette and scale after use
Pipette: Steps for Checking Reproducibility
Pipette: Steps for Checking Reproducibility
5151
Pipette: TroubleshootingPipette: Troubleshooting
Leakage
Problem Potential Cause Action
• Tip(s) incorrectly attached
• Attach firmly
• Foreign articles between the tip and cone
• O-ring damaged
• Clean tip cones
• Change the O-ring
• Incorrect operation
• Tip incorrectly attached
• Follow manufacturer’s instructions carefully
• Firmly attach tip
Inaccurate dispensing
5252
Centrifuge: Use and CareCentrifuge: Use and Care
• Always operate with the lid closed• Balance contents before turning on• Check for vibration • Do not open the lid until the rotor has come to
a complete stop• Keep lids on tubes when spinning
• Always operate with the lid closed• Balance contents before turning on• Check for vibration • Do not open the lid until the rotor has come to
a complete stop• Keep lids on tubes when spinning
5353
Centrifuges: Function ChecksCentrifuges: Function Checks
• Proper balanceProper balance• Lubrication Lubrication • Rotor functionRotor function
• Proper balanceProper balance• Lubrication Lubrication • Rotor functionRotor function
5454
Centrifuge: Routine MaintenanceCentrifuge: Routine Maintenance• Clean interior, condenser coils, fan, and screensClean interior, condenser coils, fan, and screens• Investigate unusual noises or vibrations Investigate unusual noises or vibrations • Inspect for evidence of wear, cracks in fitting, corrosion, uneven Inspect for evidence of wear, cracks in fitting, corrosion, uneven
wear, or signs of fatigue:wear, or signs of fatigue: Head, shaft head and couplingHead, shaft head and coupling RotorRotor Brushes and bearingsBrushes and bearings Power supplyPower supply Motor and lubricantMotor and lubricant Gaskets, seals, mounts and lubricantsGaskets, seals, mounts and lubricants
• Calibrate speedCalibrate speed
• Clean interior, condenser coils, fan, and screensClean interior, condenser coils, fan, and screens• Investigate unusual noises or vibrations Investigate unusual noises or vibrations • Inspect for evidence of wear, cracks in fitting, corrosion, uneven Inspect for evidence of wear, cracks in fitting, corrosion, uneven
wear, or signs of fatigue:wear, or signs of fatigue: Head, shaft head and couplingHead, shaft head and coupling RotorRotor Brushes and bearingsBrushes and bearings Power supplyPower supply Motor and lubricantMotor and lubricant Gaskets, seals, mounts and lubricantsGaskets, seals, mounts and lubricants
• Calibrate speedCalibrate speed
5555
Centrifuge SafetyCentrifuge Safety
• Increase the speed slowly until optimal speed is reachedIncrease the speed slowly until optimal speed is reached• Disconnect the centrifuge from the electrical source before Disconnect the centrifuge from the electrical source before
preventive maintenance, cleaning or inspectionpreventive maintenance, cleaning or inspection• Take caution when removing spills and broken specimen Take caution when removing spills and broken specimen
tubes after a run tubes after a run • If tubes are broken, keep the door closed and allow to sit If tubes are broken, keep the door closed and allow to sit
undisturbed for 30 minutes before attempting to clean undisturbed for 30 minutes before attempting to clean • Use tweezers to remove broken glassUse tweezers to remove broken glass
• Increase the speed slowly until optimal speed is reachedIncrease the speed slowly until optimal speed is reached• Disconnect the centrifuge from the electrical source before Disconnect the centrifuge from the electrical source before
preventive maintenance, cleaning or inspectionpreventive maintenance, cleaning or inspection• Take caution when removing spills and broken specimen Take caution when removing spills and broken specimen
tubes after a run tubes after a run • If tubes are broken, keep the door closed and allow to sit If tubes are broken, keep the door closed and allow to sit
undisturbed for 30 minutes before attempting to clean undisturbed for 30 minutes before attempting to clean • Use tweezers to remove broken glassUse tweezers to remove broken glass
5656
Keep a Log for All Maintenance ActivitiesKeep a Log for All Maintenance Activities