1
July-05-2013
GEM R&D Activity in USTC
Yi Zhou
University of Science & Technology of China
2
Outline
The NS2 Self-stretching Technique
The 30cm×30cm GEM of USTC
3
The self-stretching technique
Readout PCB
Readout PCB
GEMs
GEMs
Drift
Drift
NS1 is relaying on the elastic properties of the periphery frame
The screw is pushing the frame before the stretching.
The screw is pulling out the console
Detector Base
Detector Base
NS1
NS2
L. Franconi, A. Marinov, R. De Oliveira Status of no-stretch no-spacer GEM assembly
4
The advantages of NS2
The whole construct progress is driving the screws, it is very easy and very fast, so that we can assemble a detector in 2 hours;
All the GEM foils are self-stretched, the detector doesn’t need support frames anymore. It means that there is no dead area inside the detector, no glue aging and gas flow problems anymore;
All the screws can be released and tightened again, so that any part of the detector is replaceable, it means that the cost of the GEM project can be greatly decreased;
5
3D model of NS2 GEM
6
10M
RGEM1
RGEM2
RGEM3
10M
10M
10M
100k Drift
dI=2mm
dT2=2mm
dT1=2mm
dD=3mm
HV 100k100k100k100k100k
2.2nF2.2nF2.2nF2.2nF
Connection of HV divider & filter
RD=1M
RGEM1=0.55M
RT1=1M
RGEM2=0.5M
RT2=1M
RGEM3=0.45M
RI=1M
Low Pass Filter
7
The temperature of the dividers
HV=4500V
8
Inefficiency estimate
LCR Meter
Capacity of 10cm×10cm GEM foil: ~4.5nF
After each discharge, the GEM sector need recharge, if the protected resistor is 10MΩ, the recharge time is about:
T = 4.5nF × 10MΩ ≈ 45ms this 45ms charging time causes
the inefficiency.
If we require the inefficiency < 1%, we need P×Φ×S×T < 1% , we have:
Here: P: the discharge probability; Φ: the flux of the beam; S: the area of a GEM sector; T: the recharge time(~45ms)
%1SP
T
9
Foil assemblage
10
Gas tight test
11
Detector assemblage
12
HV discharge problem & solution
13
GEM system
14
Signal of 30cm×30cm GEM from Rui
15
We need the Cu-target X-ray tube
16
Ultrasonic bath for detector clean
17
Summary & near future work
System calibration for gain measurement; uniformity test for gain and energy resolution;
Setup the clean system;
The NS2 self-stretching technique is a very efficient way for the GEM detector construction; The HV part on the GEM detector should be sealed;
Summary:
Near future work: