CLIA Waiver Guidance

Carol C. Benson1 and Marina V. Kondratovich2

1 Associate Director, Division of Chemistry and Toxicology Devices,

OIVD, CDRH, FDA 2 Statistician, Division of Biostatistics, OSB, CDRH, FDA

AMDM 33rd Annual Meeting, April 20, 2006

Page 2: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

2

CLIA Waiver Highlights

Number of test systems categorized

History of the path to CLIA waiver Consensus guidance – new

thoughts on waiver approaches Next steps

Page 3: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

3

Number of tests categorized – (2006 not

complete)

0

200

400

600

800

1000

1200

1400

1600

1800

waived mod high

2000200120022003200420052006

Page 4: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

4

History of the Path to Waiver

Sept. 1995 CDC/CMS proposed rule Nov. 1997 FDA modernization act March 2001 FDA Draft Guidance

Document (Not Implemented) Sept. 2005 FDA Draft Guidance

Document

Page 5: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

5

42 U.S.C. Section 263a(d)(3)

“simple laboratory examinations and procedures that have been approved by the FDA for home use or that…are simple laboratory examinations and procedures that have an insignificant risk of an erroneous result”

Page 6: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

6

42 U.S.C. Section 263a(d)(3)

“including those that – (A) employ methodologies that are so simple and accurate as to render the likelihood of erroneous results by the user negligible, or (b) …pose no unreasonable risk of harm to the patient if performed incorrectly”

Page 7: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

7

2005 Consensus Draft Guidance

FDA interpretation law CLIAC, AdvaMed, CDC and CMS More flexible Scientifically based flex studies Emphasis on use of QC procedures Emphasis on intended users and

patient specimens over time

Page 8: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

8

Demonstrating “Simple”

Fully automated instrument or unitized test system

Uses direct unprocessed samples Non technique dependent

specimen or reagent manipulation

Page 9: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

9

Demonstrating “Simple” - 2

No operator intervention No technical or specialized

training Clear labeling – PI with procedure

steps at 7th grade Quick reference instructions

Page 10: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

10

Demonstrating “Insignificant Risk of Erroneous Result”

Tier 1 Hazard Analysis Tier 2 Fail-Safe and Failure Alert

Mechanisms “Accuracy” - traceability

Page 11: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

11

Tier 1: Hazard Analysis

Operator error/human factors Specimen handling and integrity Reagent integrity Hardware, software and electronics

integrity System stability Environmental factors

Page 12: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

12

Tier 2 – Fail Safe and Failure Alert Mechanisms

General recommendations in designing

External quality control Additional points for control

materials – stability and reproducibility

Validating Fail Safe/Failure Alert and ext controls – stress system

Page 13: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

13

Demonstrating “Insignificant Risk of Erroneous Result” - “Accuracy”

The term “accurate” tests refers to those tests that are comparable to traceable methods.

Prospective clinical studies of the device proposed for waiver:

- intended clinical testing sites; - intended operators; - intended sample type and matrix whenever possible; - testing over time, as in typical intended use setting.

Page 14: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

14

Demonstrating “Accuracy”

The clinical studies should compare results obtained with the device proposed for CLIA Waiver (WM) to results obtained by Comparative Method (CM).

The CM for the clinical study should be performed in laboratory setting by laboratory professionals.

Page 15: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

15

Demonstrating “Insignificant Risk of Erroneous Result” “Accuracy”

Clinical study sites Clinical study participants Clinical samples Instructions for use - labeling Questionnaire Financial disclosure

Page 16: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

16

Labeling for Waived Devices

Quick reference instructions for use QC recommendations Educational information – GLP

guidelines Safeguards – MedWatch information

on medical products reporting program

Page 17: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

17

Demonstrating “Insignificant Risk of Erroneous Result” – “Accuracy”

Page 18: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

18

Demonstrating “Accuracy” -Quantitative

Selection of Comparative Method (CM): Type A – Reference Method; Type B – Traceable method (measurement

values with the same degree of trueness as reference method or reference materials;

Type C – Traceable method (measurement values with small systematic bias, which may be clinically tolerable).

Page 19: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

19

Demonstrating “Accuracy” – Quantitative Study Design

Example WM –assay measuring some analyte; Measuring range = (5 units to 800 units); Medically important point = 200 units;

Specimen type for WM – fingerstick blood

PAIRED study design

Patient

Fingerstick blood Venous blood

WM CM

Page 20: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

20

Demonstrating “Accuracy” – Quantitative Study Design

3 sites 3 intended users (operators) at each site (total number of operators = 9); 360 samples equally distributed among operators (120 samples per site); Samples span the measuring range of the device and adequately represent all possible values of CM; Most of the samples are patient samples (60 spiked); Patient samples collected not less than 2 weeks

Page 21: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

21

Demonstrating “Accuracy” – Statistical Analysis (Quantitative)

Descriptive Statistics

0

Identity line A=B

0

100

200

300

400

500

600

700

800

0 200 400 600 800

CM

WM

For each site separately and combined:

•Scatter plot; •For CM and WM: number results, mean, standard deviation, minimum, maximum, median, box-and-whiskers plot

Page 22: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

22

Demonstrating “Accuracy” – Statistical Analysis (Quantitative)

Regression Analysis (CLSI EP9)• Appropriate type of regression; example: Deming regression

• Slope with 95% CI; example: slope = 1.02 with

95% CI: (0.983 to 1.054)

• Intercept with 95% CI; example: intercept = -2.0 with 95% CI: -11.7 to 7.7)

Identity lineY = X

y = 1.0185x - 2.0197

0

100

200

300

400

500

600

700

800

0 200 400 600 800

CM

WM

Page 23: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

23

Demonstrating “Accuracy” – Statistical Analysis (Quantitative)

Regression Analysis

Using regression equation, calculate the systematic bias (with 95% CI) at medically important points.Example: Y=1.02 *X -2.0 ; Xm = 200;

systematic bias at Xm = (1.02 * 200 -2.0) – 200 = 2 units or 1%

Some public comments:•No criteria for slope and intercept;•Only bias at medically important point (not slope close to 1 and intercept close to 0);•No need for regression analysis.

Page 24: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

24

Demonstrating “Accuracy” – Statistical Analysis (Quantitative)Total Analytical Error (CLSI EP21)

Error = deviation from truth = Result of WM – Result of CM

Total error is a simple metric for all error sources.

CM may be an average of replicates

Page 25: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

25

Demonstrating “Accuracy” – Statistical Analysis (Quantitative)

Total Analytical Error •Total Error – an interval that contains a specified proportion (usually 95% ) of the distribution of differences between the values of measurement and true value.

Estimation of total error (CLSI EP21-A). •Assumption: at some range of analyte, the errors (absolute or relative) are identically distributed independent variables.• Errors: E1, E2, E3, …., EN

total error – interval with 2.5th and 97.5th percentiles; RECOMMENDED SAMPLE SIZE =120 (for more details about why 120, see CLSI C28)

Page 26: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

26

Demonstrating “Accuracy” – Statistical Analysis (Quantitative)

Total Analytical Error • 9 measurements

•Observed ordered differences between WM and CM

•-5, -2, -2, 0, 0, +1, +1, +3, +4

•60%-total analytical error ????

•20th percentile = -2•80th percentile = +3

• 60%-total analytical error is (-2 to +3)• 2.5th and 97.5th percentiles; RECOMMENDED SAMPLE SIZE =120 (for more details about why 120, see CLSI C28)

Page 27: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

27

Demonstrating “Accuracy” – Statistical Analysis (Quantitative)

Total Analytical Error

Zero bias

-250

-200

-150

-100

-50

0

50

100

150

200

0 200 400 600 800

CM

Dif

fere

nce

bet

wee

n m

eth

od

s 97.5th percentile

2.5th percentile

Page 28: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

28

Demonstrating “Accuracy” – Statistical Analysis (Quantitative)

Total Analytical Error

•Divide the measuring range for three medically relevant intervals;•For each interval calculate the total analytical error.

Some Public Comments:•Sample size of 360 is “too burdensome”;•Sample size is not enough ( it should be at least 6 sites and 720 samples);•Larger number of contrived samples.

Page 29: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

29

Quantitative Performance Criteria

Establish Allowable Total Error (ATE) (for 95% of differences for WM and CM): values of WM that fall within ATE zones are values that can be tolerated without invalidating the medical usefulness of the WM results.

Establish Limits for Erroneous Results (LER) (no observations in LER): when WM values fall within LER zones, potential harm can occur to the patients if these results are utilized in medical decision making.

Page 30: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

30

Allowable Total Error (at least 95% of subjects)

Zones of Limits for Erroneous Results (0% of subjects).

For 360 samples, upper limit of 95% CI is less than 1%.

Allowable Total Error, Zones of Limits for Erroneous Results

Page 31: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

31

Quantitative Performance Criteria

For analytes that have existing performance limits for professional use, these limits should be used (CLIA, 42CFR 493.929)

Allowable percent is not more than 20%; It can be different rules for defining the ATE

zones for different ranges of CM

Example of ATE: if CM>90 units, CM ± 20%*CM;

if 5≤CM≤90, CM ± 18 units

Page 32: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

32

Quantitative Performance Criteria

For analytes not listed in the CLIA regulations, other criteria may be acceptable (consult with OIVD)

ATE and LER zones could be based on • medical decision making, • consideration of biological variations of analyte, • needs for accuracy of the samples within the

reference intervals or • other scientific approaches.

Page 33: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

33

Demonstrating “Accuracy” – Qualitative Study Design

Selection of comparative method: type A – quantitative reference method; type B – quantitative traceable of type B; type C – quantitative traceable of type C; type D –qualitative reference method; type E – qualitative method which was

tested by reference specimen panels (e.g, panels prepared by WHO, CDC, NIST).

Page 34: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

34

Demonstrating “Accuracy” – Qualitative Study Design 3 or more clinical sites and intended

users (9) 120 samples positive by CM 120 samples negative by CM Prospective patient samples, archival,

contrived matrix-specific. Each sample split (or pair of samples):

one part for test system (WM) and other part for comparator method (CM).

Page 35: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

35

Demonstrating “Accuracy” – Qualitative Statistical Analysis

Positive and negative agreements between WM and CM (95% two-sided CI) – for every site and combined

CM Pos

CM Neg

WM Pos

115 2

WM Neg

5 118

120 120

Pos. Agreement = 95.8% (115/120) with low limit of 95% two-sided CI of 90.5%;Neg. Agreement = 98.3% (118/120) with low limit of 95% two-sided CI of 94.1%.

Page 36: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

36

Qualitative Performance Criteria, I

Positive and negative agreements between WM and CM should be not less than 95% (for some analytes, can be higher)

Some Public Comments :• Too strong requirements for sensitivity and specificity of WM;• 120 samples positive by CM is “too burdensome”.

Page 37: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

37

Demonstrating “Accuracy” – Qualitative Study Design

Near Cutoff Studies

Page 38: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

38

Demonstrating “Accuracy” – Qualitative Study Design

Near Cutoff Studies

Weak Positive concentration is a concentration above the cutoff that the test yields positive results 95% of the time.

Weak Negative concentration is a concentration below the cutoff that the test yields negative results 95% of the time.

The length of 95% interval near the cutoff is very important analytical characteristics of the qualitative method.

Page 39: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

39

Demonstrating “Accuracy” – Qualitative Study Design

Near Cutoff Studies

• Prepare 60 aliquots of one sample with Weak Positive concentration (CM get positive results 95-99%);• Prepare 60 aliquots of one sample with Weak Negative concentration (CM get negative results 95-99%);

• Select 3 intended use sites;• Test 20 samples of each type at each testing site.

Page 40: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

40

Demonstrating “Accuracy” – Qualitative Statistical Analysis

Near Cutoff Studies

Statistical analysis Percent of positive results for weak positive sample (for every site and combined);

Percent of negative results for weak negative sample (for every site and combined).

Page 41: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

41

Qualitative Performance Criteria, II

Near Cutoff Studies

• The percent of positive results for the weak positive sample should be close to 95%;

•The differences in percents of the positive results among 3 sites for the weak positive sample should not be clinically or statistically significant .

Similar for weak negative sample

Page 42: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

42

Some Comments on Entire Guidance

“…This is a tremendous guideline. One simple comment is that it could be envisioned to apply to all diagnostic assays, not just waiver assays.”

“This draft Guidance must be withdrawn… The “accuracy” study for waiver should only be required to demonstrate that the waived user can operate the device as well as a professional user.”

“Agreement studies” in 2001 FDA Guidance.

Page 43: 1 CLIA Waiver Guidance Carol C. Benson 1 and Marina V. Kondratovich 2 1 Associate Director, Division of Chemistry and Toxicology Devices, OIVD, CDRH, FDA

43

Plan Forward Draft guidance - solicit comments Revise draft to final guidance Issue proposed rule Issue final rule