1 Applying Electrical Concepts to Pacemakers Module 4

1 Applying Electrical Concepts to Pacemakers Module 4

Embed Size (px)

Citation preview

Page 1: 1 Applying Electrical Concepts to Pacemakers Module 4

1

Applying Electrical Concepts to PacemakersModule 4

Page 2: 1 Applying Electrical Concepts to Pacemakers Module 4

2

Objectives

• Upon completion you will be able to:

– Recognize a high impedance condition

– Recognize a low impedance condition

– Recognize capture threshold

– Determine which sensitivity value is more (or less) sensitive

Page 3: 1 Applying Electrical Concepts to Pacemakers Module 4

3

Electrical Information

• Why is this electrical information relevant?

• A pacemaker is implanted to:

– Provide a heart rate to meet metabolic needs

• In order to pace the heart, it must capture the myocardium

• In order to pace the heart, it must know when to pace, i.e., it must be able to sense

• A pacemaker requires an intact electrical circuit

Page 4: 1 Applying Electrical Concepts to Pacemakers Module 4

Ohm’s LawRelevance to Pacemaker Patients

• High impedance conditions reduce battery current drain

– Can increase pacemaker battery longevity

– Why?

• R = V/I If “R” increases and “V” remains the same, then “I” must decrease

• Low impedance conditions increase battery current drain

– Can decrease pacemaker battery longevity

– Why?

• R = V/I If “R” decreases and “V” remains the same, then “I” must increase

Page 5: 1 Applying Electrical Concepts to Pacemakers Module 4

5

The Effect of Lead Performance on Myocardial Capture

What would you expect to happen if a lead was partially fractured?

• Impedance (or Resistance) would rise

• Current would decrease and battery energy conserved

- but -

Could you guarantee that enough current (I) can flow through this fractured lead so that each time the pacemaker fired the myocardium would beat?

Page 6: 1 Applying Electrical Concepts to Pacemakers Module 4

Lead wire fracture

Increased resistance

High Impedance ConditionsA Fractured Conductor

• A fractured wire can cause Impedance values to rise

– Current flow from the battery may be too low to be effective

• Impedance values may exceed 3,000

Other reason for high impedance: Lead not seated properly in pacemaker header.

Page 7: 1 Applying Electrical Concepts to Pacemakers Module 4

Lead Impedance Values Change as a Result of:

• Wire fractures

• Insulation breaks

Typically, normal impedance reading values range from 300 to 1,000

– Some leads are high impedance by design. These leads will normally show impedance reading values greater than 1,000 ohms

• Medtronic High Impedance leads are:

– CapSure® Z

– CapSure® Z Novus

Page 8: 1 Applying Electrical Concepts to Pacemakers Module 4

Low Impedance Conditions

• Insulation breaks expose the lead wire to the following

– Body fluids, which have a low resistance, or

– Another lead wire (in a bipolar lead)

• Insulation break that exposes a conductor causes the following

– Impedance values to fall

– Current to drain through the insulation break into the body, or into the other wire

– Potential for loss of capture

– More rapid battery depletion

Current will follow the path of LEAST resistance

Page 9: 1 Applying Electrical Concepts to Pacemakers Module 4

Capture Threshold

• The minimum electrical stimulus needed to consistently capture the heart outside of the heart’s own refractory period

Ventricular pacemaker 60 ppm

Capture Non-Capture

Page 10: 1 Applying Electrical Concepts to Pacemakers Module 4

Effect of Lead Design on Capture

• Lead maturation

– Fibrotic “capsule” develops around the electrode following lead implantation

– May gradually raise threshold

– Usually no measurable effect on impedance

Page 11: 1 Applying Electrical Concepts to Pacemakers Module 4

Steroid Eluting Leads

• Steroid eluting leads reduce the inflammatory process

– Exhibit little to no acute stimulation threshold peaking

– Leads maintain low chronic thresholds

Silicone rubber plugcontaining steroid

Tines forstablefixation

Porous, platinized tipfor steroid elution

Page 12: 1 Applying Electrical Concepts to Pacemakers Module 4

Effect of Steroid on Stimulation Thresholds

Pulse Width = 0.5 msec

03 6

Implant Time (Weeks)

Textured Metal Electrode

Smooth Metal Electrode

1

2

3

4

5

Steroid-Eluting Electrode

0 1 2 4 5 7 8 9 10 11 12

Vol

ts

References: Pacing Reference Guide, Bakken Education Center, 1995, UC199601047aEN. Cardiac Pacing, 2nd Edition, Edited by Kenneth A. Ellenbogen. 1996.

Page 13: 1 Applying Electrical Concepts to Pacemakers Module 4

Myocardial Capture

• Capture is a function of:

– Amplitude—the strength of the impulse expressed in volts

• The amplitude of the impulse must be large enough to cause depolarization (i.e., to “capture” the heart)

• The amplitude of the impulse must be sufficient to provide an appropriate pacing safety margin

– Pulse width—the duration of the current flow expressed in milliseconds

• The pulse width must be long enough for depolarization to disperse to the surrounding tissue

Page 14: 1 Applying Electrical Concepts to Pacemakers Module 4

Comparison5.0 Volt Amplitude at Different Pulse Widths

Amplitude5.0 V

0.25 ms 1.0 ms0.5 ms

Page 15: 1 Applying Electrical Concepts to Pacemakers Module 4

The Strength-Duration Curve

• The strength-duration curve illustrates the relationship of amplitude and pulse width

– Any combination of pulse width and voltage, on or above the curve, will result in capture

Pulse Width

.50

1.0

1.5

2.0

.25V

olts

0.5 1.0 1.5

Capture

No Capture

Rheobase

Chronaxie

Page 16: 1 Applying Electrical Concepts to Pacemakers Module 4

DurationPulse Width (ms)

Clinical Utility of the Strength-Duration Curve

• By accurately determining capture threshold, we can assure adequate safety margins because:

– Thresholds may differ in acute or chronic pacing systems

– Thresholds fluctuate slightly daily

– Thresholds can change due to metabolic conditions or medications

0.5 1.0 1.5

.50

1.0

1.5

2.0

.25

Sti

mu

lati

on

Th

resh

old

(V

olt

s)

X Programmed Output

Page 17: 1 Applying Electrical Concepts to Pacemakers Module 4

Programming Outputs

• Primary goal: Ensure patient safety and appropriate device performance

• Secondary goal: Extend the service life of the battery

– Typically program amplitude to < 2.5 V, but always maintain adequate safety margins

• A common output value might be 2.0 V at 0.4 ms

– Amplitude values greater than the cell capacity of the pacemaker battery (usually about 2.8 V) require a voltage multiplier, resulting in markedly decreased battery longevity

Page 18: 1 Applying Electrical Concepts to Pacemakers Module 4

18

Pacemaker Sensing

• Refers to the ability of the pacemaker to “see” signals

– Expressed in millivolts (mV)

• The millivolts (mV) refers to the size of the signal the pacemaker is able to “see”

0.5 mV signal

2.0 mV signal

Page 19: 1 Applying Electrical Concepts to Pacemakers Module 4

SensitivityThe Value Programmed into the IPG

Time

5.0 mV

2.5 mV

1.25 mV

Page 20: 1 Applying Electrical Concepts to Pacemakers Module 4

Time

5.0 mV

2.5 mV

1.25 mV

5 mV sensitivity

At this value the pacemaker will not see the 3.0 mV signal

SensitivityThe Value Programmed into the IPG

Page 21: 1 Applying Electrical Concepts to Pacemakers Module 4

SensitivityThe Value Programmed into the IPG

At this value, the pacemaker can see both the 3.0 mV and the 1.30 mV signal. So, is “more sensitive” better, because the pacemaker sees smaller signals?

1.25 mV Sensitivity

Time

5.0 mV

2.5 mV

1.25 mV

But what about this?

Page 22: 1 Applying Electrical Concepts to Pacemakers Module 4

Sensing Amplifiers/Filters

• Accurate sensing requires that extraneous signals are filtered out

– Because whatever a pacemaker senses is by definition a P- or an R-wave

– Sensing amplifiers use filters that allow appropriate sensing of P- and R-waves, and reject inappropriate signals

• Unwanted signals most commonly sensed are:

– T-waves (which the pacemaker defines as an R-wave)

– Far-field events (R-waves sensed by the atrial channel, which the pacemaker thinks are P-waves)

– Skeletal muscle myopotentials (e.g., from the pectoral muscle, which the pacemaker may think are either P- or R-waves)

– Signals from the pacemaker (e.g., a ventricular pacing spike sensed on the atrial channel “crosstalk”)

Page 23: 1 Applying Electrical Concepts to Pacemakers Module 4

Sensing Accuracy

• Affected by:

– Pacemaker circuit (lead) integrity

• Insulation break

• Wire fracture

– The characteristics of the electrode

– Electrode placement within the heart

– The sensing amplifiers of the pacemaker

– Lead polarity (unipolar vs. bipolar)

– The electrophysiological properties of the myocardium

– EMI – Electromagnetic Interference

Page 24: 1 Applying Electrical Concepts to Pacemakers Module 4

Lead Conductor Coil IntegrityAffect on Sensing

• Undersensing occurs when the cardiac signal is unable to get back to the pacemaker

– Intrinsic signals cannot cross the wire fracture

• Oversensing occurs when the severed ends of the wire intermittently make contact

– Creates signals interpreted by the pacemaker as P- or R-waves

Page 25: 1 Applying Electrical Concepts to Pacemakers Module 4

Lead Insulation IntegrityAffect on Sensing

• Undersensing occurs when inner and outer conductor coils are in continuous contact

– Signals from intrinsic beats are reduced at the sense amplifier, and amplitude no longer meets the programmed sensing value

• Oversensing occurs when inner and outer conductor coils make intermittent contact

– Signals are incorrectly interpreted as P- or R-waves

Page 26: 1 Applying Electrical Concepts to Pacemakers Module 4

Unipolar Pacemaker

• Where is the sensing circuit?

_Cathode

Anode

Lead tip to canThis can produce a large potential difference (signal) because the cathode and anode are far apart

Click for Answer

Page 27: 1 Applying Electrical Concepts to Pacemakers Module 4

Bipolar Pacemaker

• Where is the sensing circuit?

Anode and Cathode

Click for Answer

Lead tip to ring on the leadThis usually produces a smaller potential difference due to the short inter-electrode distance

• But, electrical signals from outside the heart (such as myopotentials) are less likely to be sensed

Page 28: 1 Applying Electrical Concepts to Pacemakers Module 4

28

Cardiac Conduction and Device Sensing

By now we should be familiar with the surface ECG and its relationship to cardiac conduction. But, how does this relate to pacemaker sensing?

Page 29: 1 Applying Electrical Concepts to Pacemakers Module 4

29

Vectors and Gradients

Sense

The wave of depolarization produced by normal conduction creates a gradient across the cathode and anode. This changing polarity creates the signal.

Once this signal exceeds the programmed sensitivity – it is

sensed by the device.

Click for More

2.5 mV

Page 30: 1 Applying Electrical Concepts to Pacemakers Module 4

30

Changing the Vector

Sense

2.5 mV

A PVC occurs, which is conducted abnormally. Since the vector relative to the lead has changed, what effect

might this have on sensing?

Click for More

In this case, the wave of depolarization strikes the anode and cathode almost simultaneously. This

will create a smaller gradient and thus, a smaller signal.

Page 31: 1 Applying Electrical Concepts to Pacemakers Module 4

31

Putting It All Together

• Appropriate output programming can improve device longevity

– But, do not compromise patient safety!

• Lead design can improve device longevity via

– Steroid eluting leads

• Can help keep chronic pacing thresholds low by reducing inflammation and scarring

– High Impedance leads

• Medtronic CapSure Z and Medtronic CapSure Z Novus

• Designed so electrode is high, butV low so current (I) is low as well, reducing battery drain

• Control of manufacturing

– Batteries, circuit boards, capacitors, etc., specific to needs, can lead to improved efficiencies and lowered static current drain

– Highly reliable lead design

Page 32: 1 Applying Electrical Concepts to Pacemakers Module 4

32

Putting It All Together

• Pacemaker Longevity is:

– A function of programmed parameters (rate, output, % time pacing)

– A function of useful battery capacity

– A function of

• Static current drain

• Circuit efficiency

• Output Impedance

• The lower the programmed sensitivity the MORE sensitive the device

– Lead integrity also affects sensing

Page 33: 1 Applying Electrical Concepts to Pacemakers Module 4

33

Status Check

• Determine the threshold amplitude

Capture threshold = lowest value with consist capture

This is at 1.25 V

Click for Answer

1.25 V 0.05 V0.75 V1.00 V

Page 34: 1 Applying Electrical Concepts to Pacemakers Module 4

34

Status Check

OR

Click for Answer

Pacemaker A is able to “see” signals as small as 0.5 mV. Thus, it is more sensitive.

Programmed Sensitivity 0.5 mV

Pacemaker A

Programmed Sensitivity 2.5 mV

Pacemaker B

• Which of these pacemakers is more sensitive?

Page 35: 1 Applying Electrical Concepts to Pacemakers Module 4

35

Status Check

• A pacemaker lead must flex and move as the heart beats. On average, how many times does a heart beat in 1 year?

Click for Answer

35 MILLION times. It is not a simple task to design a lead that is small, reliable, and lasts a lifetime.

Page 36: 1 Applying Electrical Concepts to Pacemakers Module 4

36

Status Check

Click for Answer

Lead Fracture:

• High Impedance

• Possible failure to capture myocardium

Do you notice anything on this x-ray?

Page 37: 1 Applying Electrical Concepts to Pacemakers Module 4

37

Status CheckWhat would you expect?

• Which value is out of range?

• What could have caused this?

Pacemaker Interrogation Report

Mode: DDDR

Lower: Rate 60 ppm

UTR: 130 ppm

USR: 130 ppm

Atrial Lead Impedance: 475 Ohms

Ventricular Lead Impedance: 195 Ohms

Insulation failure

Click for Answer

Page 38: 1 Applying Electrical Concepts to Pacemakers Module 4

38

Brief Statements

Indications

• Implantable Pulse Generators (IPGs) are indicated for rate adaptive pacing in patients who ay benefit from increased pacing rates concurrent with increases in activity and increases in activity and/or minute ventilation. Pacemakers are also indicated for dual chamber and atrial tracking modes in patients who may benefit from maintenance of AV synchrony. Dual chamber modes are specifically indicated for treatment of conduction disorders that require restoration of both rate and AV synchrony, which include various degrees of AV block to maintain the atrial contribution to cardiac output and VVI intolerance (e.g. pacemaker syndrome) in the presence of persistent sinus rhythm.

• Implantable cardioverter defibrillators (ICDs) are indicated for ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias.

• Cardiac Resynchronization Therapy (CRT) ICDs are indicated for ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias and for the reduction of the symptoms of moderate to severe heart failure (NYHA Functional Class III or IV) in those patients who remain symptomatic despite stable, optimal medical therapy and have a left ventricular ejection fraction less than or equal to 35% and a QRS duration of ≥130 ms.

• CRT IPGs are indicated for the reduction of the symptoms of moderate to severe heart failure (NYHA Functional Class III or IV) in those patients who remain symptomatic despite stable, optimal medical therapy, and have a left ventricular ejection fraction less than or equal to 35% and a QRS duration of ≥130 ms.

Contraindications

• IPGs and CRT IPGs are contraindicated for dual chamber atrial pacing in patients with chronic refractory atrial tachyarrhythmias; asynchronous pacing in the presence (or likelihood) of competitive paced and intrinsic rhythms; unipolar pacing for patients with an implanted cardioverter defibrillator because it may cause unwanted delivery or inhibition of ICD therapy; and certain IPGs are contraindicated for use with epicardial leads and with abdominal implantation.

• ICDs and CRT ICDs are contraindicated in patients whose ventricular tachyarrhythmias may have transient or reversible causes, patients with incessant VT or VF, and for patients who have a unipolar pacemaker. ICDs are also contraindicated for patients whose primary disorder is bradyarrhythmia.

Page 39: 1 Applying Electrical Concepts to Pacemakers Module 4

39

Brief Statements (continued)

Warnings/Precautions

• Changes in a patient’s disease and/or medications may alter the efficacy of the device’s programmed parameters. Patients should avoid sources of magnetic and electromagnetic radiation to avoid possible underdetection, inappropriate sensing and/or therapy delivery, tissue damage, induction of an arrhythmia, device electrical reset or device damage. Do not place transthoracic defibrillation paddles directly over the device. Additionally, for CRT ICDs and CRT IPGs, certain programming and device operations may not provide cardiac resynchronization. Also for CRT IPGs, Elective Replacement Indicator (ERI) results in the device switching to VVI pacing at 65 ppm. In this mode, patients may experience loss of cardiac resynchronization therapy and / or loss of AV synchrony. For this reason, the device should be replaced prior to ERI being set.

Potential complications

• Potential complications include, but are not limited to, rejection phenomena, erosion through the skin, muscle or nerve stimulation, oversensing, failure to detect and/or terminate arrhythmia episodes, and surgical complications such as hematoma, infection, inflammation, and thrombosis. An additional complication for ICDs and CRT ICDs is the acceleration of ventricular tachycardia.

• See the device manual for detailed information regarding the implant procedure, indications, contraindications, warnings, precautions, and potential complications/adverse events. For further information, please call Medtronic at 1-800-328-2518 and/or consult Medtronic’s website at www.medtronic.com.

Caution: Federal law (USA) restricts these devices to sale by or on the order of a physician.

Page 40: 1 Applying Electrical Concepts to Pacemakers Module 4

40

Brief Statement: Medtronic Leads

Indications

• Medtronic leads are used as part of a cardiac rhythm disease management system. Leads are intended for pacing and sensing and/or defibrillation. Defibrillation leads have application for patients for whom implantable cardioverter defibrillation is indicated

Contraindications

• Medtronic leads are contraindicated for the following:

• ventricular use in patients with tricuspid valvular disease or a tricuspid mechanical heart valve.

• patients for whom a single dose of 1.0 mg of dexamethasone sodium phosphate or dexamethasone acetate may be contraindicated. (includes all leads which contain these steroids)

• Epicardial leads should not be used on patients with a heavily infracted or fibrotic myocardium.

• The SelectSecure Model 3830 Lead is also contraindicated for the following:

• patients for whom a single dose of 40.µg of beclomethasone dipropionate may be contraindicated.

• patients with obstructed or inadequate vasculature for intravenous catheterization.

Page 41: 1 Applying Electrical Concepts to Pacemakers Module 4

41

Brief Statement: Medtronic Leads (continued)

Warnings/Precautions

• People with metal implants such as pacemakers, implantable cardioverter defibrillators (ICDs), and accompanying leads should not receive diathermy treatment. The interaction between the implant and diathermy can cause tissue damage, fibrillation, or damage to the device components, which could result in serious injury, loss of therapy, or the need to reprogram or replace the device.

• For the SelectSecure Model 3830 lead, total patient exposure to beclomethasone 17,21-dipropionate should be considered when implanting multiple leads. No drug interactions with inhaled beclomethasone 17,21-dipropionate have been described. Drug interactions of beclomethasone 17,21-dipropionate with the Model 3830 lead have not been studied.

Potential Complications

• Potential complications include, but are not limited to, valve damage, fibrillation and other arrhythmias, thrombosis, thrombotic and air embolism, cardiac perforation, heart wall rupture, cardiac tamponade, muscle or nerve stimulation, pericardial rub, infection, myocardial irritability, and pneumothorax. Other potential complications related to the lead may include lead dislodgement, lead conductor fracture, insulation failure, threshold elevation or exit block.

• See specific device manual for detailed information regarding the implant procedure, indications, contraindications, warnings, precautions, and potential complications/adverse events. For further information, please call Medtronic at 1-800-328-2518 and/or consult Medtronic’s website at www.medtronic.com.

Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.

Page 42: 1 Applying Electrical Concepts to Pacemakers Module 4

42

Disclosure

NOTE:

This presentation is provided for general educational purposes only and should not be considered the exclusive source for this

type of information. At all times, it is the professional responsibility of the practitioner to exercise independent

clinical judgment in a particular situation.