  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    1/29

    BONE REMODELING FOLLOWNG HIP FEMORAL ENDOPROSTHESIS

    BYProfessor Hazem Abdel Azeem

    Cairo University

    March 2001

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    2/29

    Introduction

    Factors affecting survivorship of femoral componentStress shielding and bone loss.

    Bone mineral changes in the femora of implanted prosthesis.Inability of conventional radiography to adequatelyevaluate these change in its early stage.

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    3/29

    (A). The patient was revised with a 19.5-mm extensively porous coated cobalt-chromium alloy stem (B). At the 20-month follow-up, there was extensive thinning of the proximal cortical bone, but the patient was asymptomatic (C). Adapted from Sumner DR, Turner TM. Total hip revision surgery. New York: Raven Press, 1995.

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    4/29

    Contact radiographs of retrieved femora from a dog that had anextensively porous-coated titanium alloy stem implanted in theleft femur for 24 months, showing marked cortical bone loss in

    the proximal medial region.

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    5/29

    Literature review of different

    densitometry in THAAUTHOR PROSTHESIS CHANGE Follow-up

    McCarthy et al Cemented (unspecified composition) 0 1 to 14 yr Cohen and Rushton Cemented (CoCr) -7% BMDb (Gruen zone 7)-8% BMDb (Gruen zone 2) 1 yr

    Korovessis et al Cemented (CoCr) -16% BMD(ROI overlapping lesser trochanter) 1 mo-10 yr (mean, 2 yK ilgus et al Cement les s CoCr-ex tens ively porous -c oat ed -40% B MDc (m edial mean)-24% B MDc (lat eral mean) 2-5 yr K ilgus et al Cement les s CoCr-prox im ally porous -c oat ed -19% BM Dc (m edial m ean)-11% BM Dc (lat eral m ean) 2-6 y r

    Sychterz and Engh Cementless CoCr-extens ively porous-coated -42% BMC (medial+lateral mean) 1.5-13.5 yr Hughes et al Cementless CoCr-proximal 5/8 porous-coated -34% BMD (proximal of Gruen zone 7) 45 mo

    K oroves si s et al C em ent les s Ti 6A l7N b-g rit -blas ted s urfac e -24% B M D (R OI overlap pi ng les ser tro cha nt er) 1 m o-1 0 y r(m ea n, 2 y r K irat li et al Cem ent les s Ti6A l4V-prox im ally porous -c oat ed -24% BM D (RO Is enc om pas sing les ser t roc hant er) 2 y r Nakamura Cementless Ti6Al4V -23% BMDd (Gruen zone 7) 2 yr Nakamura Cementless Ti6Al4V -20% BMD (Gruen zone 7) 12-24 moNakamura Cementless Ti6Al4V -27% BMD (Gruen zone 7) 25-48 mosNakamura Cementless Ti6Al4V -45% BMD (Gruen zone 7) 49-84 moBryan et al Cementless Ti6Al4V-extens ively porous-c oated on AP surfac es -34% BMC (Gruen zone 7 9-10 yr

    Caniggia et al Cementless Ti alloy with hydroxyapatite coating -23% BMD (proximal medial cortex) 4 moHughes et al Cementless Ti-proximal 3/8porous-coated -15% BMD (proximal of Gruen zone 7) 38 mo

    Niinimki and Jalovara Cementless polyacetal resin with a steel core -40% BMD (Gruen zone 7) 7.5-9.5 yr

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    6/29

    AIM OF THE WORK

    A trial to find out a method for assessmentof the effect of stress shielding acquired by

    presence of the prosthesis in the upper endof the femur Objective study based on the deterioration

    of the quality of the bone around the stem inrelation to time

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    7/29

    AIM OF THE WORK (Contd.)

    Should this have impact ontechnical details of prosthesisdesign, cemented or cementlessand technique of cementing ??

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    8/29

    LIMITATIONS OF THE STUDY

    Presence of pre-existing evidentosteoporosisElapse of two years as osteolysis will takeshare in process of bone lossPathological lesions in the upper end of thefemur

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    9/29

    Inclusion Criteria

    Cases of femoral endoprosthesis in patients before age 50 (possibly not osteoporotic)Active patient (was back to work)Cemented or cementlessStandard stem sizes

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    10/29

    Material

    39 patientsunilateral total hip arthroplastyfemoral component: 24 cemented and15 cementless.Cementless: extensive coating with HA

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    11/29

    Material (contd.)

    Age distribution: range: 36 - 50 years.

    Sex distribution: 18 males. 21 females.

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    12/29

    METHODS

    Time of DEXA: 3, 6 and 12 months postoperative

    Equipment use: a human DPZ-IQMeasurements made in the 7 zones of Gruen.Base line is the findings of the first

    periprosthetic DEXA evaluation (3 months postoperative)

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    13/29

    METHODS (Contd.)

    Percentage of loss is calculated from thefirst measurement spaced three monthly for two years (Osteolysis takes share after twoyears and may mix up with the process of local osteoporosis)

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    14/29

    Technical details include pic.

    A base line densitometry study is done onthe contralateral un-operated side

    The software allows positioning of the prosthesis simulator to deduce the value of the prosthesis on that side and gives a

    standard value in the seven areas

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    15/29

    Technical details (contd.)include

    pic.The software measures the bone densityaround the prosthesis on the affected side

    Comparison between both sides gives thedeterioration rate caused by stress shieldingin every of the seven areas

    In the second study, further loss isestimated the same way and so on

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    16/29

    The 7 Zones of Gruen

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    17/29

    The 7 Zones of Gruen

    (Further Analysis)

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    18/29

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    19/29

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    20/29

    FOLLOW UP PERIOD

    2 YEARS

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    21/29

    Results

    The third postoperative month measurement was usedas the baseline.

    There was bone loss from the third month onwards(16 percent per year in zone 7)Bone loss leveled between the 12th month and 2years.

    Bone loss is not uniform, greatest decrease was seenin Gruens zone 7 (16%)Least in Zone 4 (4%)

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    22/29

    Example of Densitometry

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    23/29

    Covariate Factors of Bone LossSex: Women have significantly higher bone loss (more

    pronounced in zone 7)

    Implant-related difference: Implant size

    More bone loss with bigger sizes

    Cementing

    Less bone loss with cementingFemoral fill Less loss with femoral fill

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    24/29

    Effect of Femoral Fill

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    25/29

    Discussion

    Early bone loss cannot be identified by standard x-rays; hence value of densitometry.

    Rapid bone loss occurs up to two years: 16% per year in area 7

    Bone loss slows down after two years (Lit.)Resorption is greater with bulkier and stiffer devices, both in the metaphysis and diaphysis

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    26/29

    Discussion (Contd.)

    Stress shielding is more effective on the proximal part of the femur especially the

    medial sideStress shielding is less with cemented

    prosthesis (wider surface contact)

    Stress was higher in non-cemented prosthesis (the contact at maximum is 20%)

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    27/29

    Discussion (Contd.)

    Stress shielding is taken off suddenly inzone 4 in cementless prosthesis. This may

    allow fracture in this zone.Stress shielding may be graduated bycementing (transitional zone)

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    28/29

    Conlusion

    DEXA gives better insight into boneremodeling early post-implantation

    DEXA may help optimization of femoraldesign and better techniqueLong stem devices may be less beneficial with

    more stress shielding.Cemented stems may be associated with less

    bone loss and stress shielding.

  • 8/14/2019 1-18 DEXA and Periprosthetic Loosening

    29/29

    Conclusion (contd.)

    No correlation as yet has been found between observed bone resorption and

    clinical result.Method used permits a quantitativeapproach to bone remodeling to

    differentiate between physiological changesand pathological bone loss