Upload
rmsanchep
View
214
Download
0
Embed Size (px)
Citation preview
8/13/2019 05-Thermocouples
1/4
23
5. MEASURING WITH THERMOCOUPLE
Objective
1. Check the function of two-wire transducer XTR101 for measuring temperature with
thermocouples.
2. Check the use of measuring module Janascard AD232 with isothermal terminal board and
serial interface.
Introduction:
In the setup, there are thermocouples of K-type. It is thermocouple NiCr-Ni (nickel/chromium
- nickel), also known as Chromel-Alumel (Ch-A). There are three K-thermocouples in the
laboratory miniature furnace. Their sensitivity is approximately =&
42 V/K.
The first thermocouple is connected directly to the module Janascard AD232.
The second thermocouple is connected to transducer XTR101. The reference ends are placed
in a bulk aluminum block together with a bipolar transistor whose PN junction is used by
XTR101 for measuring the temperature of reference (cold) ends. To evaluate the quality of
reference ends temperature compensation by XTR101, the aluminum block is heated up by
current (approx. 1 A) passing through a power transistor mounted on front side. The
temperature of the aluminum block can be inspected by a digital room thermometer with
external probe. The probe is also inserted into the aluminum block. The lower of the two
displays shows the temperature of aluminum block.
Note: the third thermocouple is connected to so called electronic ice point (battery powered
device for reference ends temperature compensation).
Procedure
5.1. Two wire transducer XTR 101
5.1.1. Study the manual of XTR101
5.1.2. Influence of reference (cold) end temperatureSimply stated, the generated thermoelectric voltage is proportional to the difference of
temperatures between the measuring (hot) junction and the reference (cold) ends. Thus,
warming up of the reference ends will influence the generated voltage.
While the furnace is off (stable temperature of junction) the reference ends in aluminum block
are heated up by current in the attached power transistor. Warm up the block several degrees
above the ambient temperature and observe the change of generated voltage by a voltmeter.
The same thermoelectric voltage is connected also to the input of XTR101 circuit. The
temperature of reference ends is sensed by PN junction of an auxiliary transistor. If the
resistor network of XTR101 is properly designed and the PN junction is connected, the circuit
8/13/2019 05-Thermocouples
2/4
24
compensates any changes of reference ends temperature and the resulting output voltage is
not influenced.
The output signal of XTR101 is measured by another voltmeter. This voltage should be stable
in spite of heating of the cold ends of thermocouple and thus variations of the thermoelectric
voltage in the input. (The XTR101 is able to compensate the heating up of reference ends.)
5.1.3. Influence of the wire resistance
While the temperature of the hot junction and the cold ends is stable, examine the influence of
wire resistance. The chip XTR101 output works in current loop, i.e. the output signal is the
current in circuit between the voltage supply and the XTR101. (In our case, the measured
signal is voltage drop on a sensing resistor RL). The current should be proportional only to the
measured quantity (here - temperature). However, when the resistance of supply wires is too
high, the chip is not able to regulate the current properly and signal is distorted.
Increase the resistance of the supply circuit by setting the decade resistor box from zero to
higher and higher value. Observe what is the maximum resistance before signal is distorted.
5.2. Using measuring module Janascard AD232
Apply line voltage (100-220V) to the furnace and let it heat up to 150-200 C. Measure the
generated thermoelectric voltage and also the output voltage of XTR101 (10 values). At the
same time, monitor the temperature in furnace by module Janascard AD232. The controlling
software in PC can display the voltage and recalculated temperature according to setup.
Record also the output voltage of third thermocouple compensated by electronic ice point.
Fig. 5.1 The furnace and thermocouples
Furnace
XTR101 circuit
Aluminum block
(thermostat)
8/13/2019 05-Thermocouples
3/4
25
Fig. 5.2 Block diagram of XTR 101
Fig. 5.3 Cold-end temperature compensation
Fig. 5.4 Setting up the offset value (4-20mA)
Thermocouple
sensing PNunction fortemperaturecompensation
8/13/2019 05-Thermocouples
4/4
26
Compensation box (electronic ice point) Omega CJ
Precission of compensation: 0,25 C at 25 C
0,5 C for 15 to 35 C
0,75 C for 10 C to 50 CSupply: lithium pill 3,6 V
Note: Turn off after measurement !!!