  • Pearl GTL Project

    LINDE ENGINEERING

    Shell Project No: HP-3000-QAT Contractor Project No : 5887

    Project Document Number: T- 4.250.915 Project Rev.: B Page 2 of 14

    Purchase Order Number:

    2MH127

    Tag/Item Number(s): N/A

    Unit(s): ASU

    PCWBS: Z

  • QSGTL PEARL PLANT C2 AIR SEPARATION UNITS CONSTRUCTION

    E B 09.11.08 Ready for Construction FB EO AG RFC

    D A 09.10.08 Issue for Review FB EO AG IFR

    C O 02.12.07 Issued for Company Comments BR UA MC /ZG ICC

    B Issued for Review MS MC ZG IFR

    A Issued for Internal Review MS MC ZG IIR

    GAMA Rev.

    Project Rev

    Date Description Prepared

    By Checked

    By

    Approved By

    Status

    No. No. ORIGINATOR

    LINDE ENGINEERING AG

    Document Title:

    PROCEDURE FOR CALIBRATION

    Contractor

    Gama Qatar Co. W. L. L.

    Document No. LOC470

    GAM PRO QAC GTL 028 E

    Org.

    Code

    Doc.

    Type

    Disc. Code

    Geogr. Area

    Seq.

    No

    Rev. No

    Vendor Code

    Project Document Number: T-4.250.915

    Rev No: B

  • Doc. Title : GAM-PRO-QAC-GTL-028 Rev. E PROCEDURE FOR CALIBRATION

    Date : 09.11.2008

    T-4.250.915 Rev. B Page 4 of 14

    QQQ AAA TTT AAA RRR

    TABLE OF CONTENTS

    1. SCOPE ________________________________________________________________________ 5

    2. PURPOSE _____________________________________________________________________ 5

    3. REFERENCES _________________________________________________________________ 5

    4. DEFINITIONS _________________________________________________________________ 5

    5. RESPONSIBILITIES ____________________________________________________________ 6

    6. PROCEDURE __________________________________________________________________ 6

    6.1. GENERAL REQUIREMENTS _______________________________________________________6

    6.2. IDENTIFICATION OF INSTRUMENTS ______________________________________________6

    6.3. CALIBRATION OF INSTRUMENTS _________________________________________________7 6.3.1. Periodic Calibration ____________________________________________________________________ 7 6.3.2. Re-calibration _________________________________________________________________________ 7

    6.4. PRECISION OF CALIBRATION ____________________________________________________7

    6.5. METHOD OF CALIBRATION ______________________________________________________7

    6.6. DOCUMENTATION OF CALIBRATION _____________________________________________8

    7. ATTACHMENTS ________________________________________________________________ 9

  • Doc. Title : GAM-PRO-QAC-GTL-028 Rev. E PROCEDURE FOR CALIBRATION

    Date : 09.11.2008

    T-4.250.915 Rev. B Page 5 of 14

    QQQ AAA TTT AAA RRR

    1. SCOPE

    This procedure describes the criteria and methods that shall be applied by GAMA in ensuring the accuracy of the measuring and test instruments that shall be used in the erection activities of the QSGTL Pearl Plant C2 Air Separation Units.

    2. PURPOSE

    The purpose of this Quality Control Procedure is to set a guideline of the testing requirements as well as the responsibilities of implementing this procedure.

    3. REFERENCES

    T.4.096.674 : Calibration of Testing and Measuring Tools and Equipment PQP-08.F01 : List of Measuring and Test Devices Subject to Calibration PQP-08.F02 : Calibration Label

    Relevant Manufacturers Manual for equipments. Also all relevant ITPs and QCRs.

    4. DEFINITIONS

    Company : QSGTL and/or its PMC

    Contractor : Linde Engineering AG

    Sub Contractor : Gama Qatar

    Project : QSGTL C2 ASU PEARL Construction Works

    ITP : Inspection Test Plan

    QCP : Quality Control Procedure

    QA/QC : Quality Assurance / Quality Control Department

    ITL : Independent Testing Laboratory

    MRR : Material Receiving Report

    HSE : Health and Safety and Environment

  • Doc. Title : GAM-PRO-QAC-GTL-028 Rev. E PROCEDURE FOR CALIBRATION

    Date : 09.11.2008

    T-4.250.915 Rev. B Page 6 of 14

    QQQ AAA TTT AAA RRR

    5. RESPONSIBILITIES

    The QA/QC Manager is responsible for monitoring and implementing the following activities:

    Establishing the calibration schedule for all equipment included in the system.

    Keeping a record of all calibrated instruments.

    Ensuring that instruments are calibrated at all times during the work.

    Storage of calibration instruments, which are defined as secondary "MASTER".

    Master instruments are generally stored in a locked-up cabinet under the control of the QA/QC Manager. Particularly, over-sized instruments shall be kept at adequate areas and/or in proper containers.

    The above mentioned activities shall be carried on by the QA/QC Department together with the department that uses the instrument.

    6. PROCEDURE

    6.1. GENERAL REQUIREMENTS

    Instruments and/or devices having appropriate calibration certificates can be utilised within the expiry period. After the expiry dates, the calibration of these instruments and/or devices should be calibrated.

    The present procedure is applicable to instruments that require to be calibrated periodically, such as pressure gages, hardness testers, paint thickness gages, optical measurement devices, etc.

    Calibration intervals for these instruments and/or devices are given in Attachment 1 according to T.4.096.674 Calibration of Testing and Measuring Tools and Equipment.

    6.2. IDENTIFICATION OF INSTRUMENTS

    Each instrument that is subject to calibration shall be identified by a serial number.

    a) This serial number will be placed on the instrument by such means that shall provide durable identification. All inspection, measuring and test instruments with missing or illegible serial numbers shall be re-calibrated prior to their use.

    b) All inspection, measuring and test instruments that are employed for the inspection or acceptance of the construction work shall be marked with utilising stickers or tags, which shall enable immediate visual confirmation of its calibration status.

  • Doc. Title : GAM-PRO-QAC-GTL-028 Rev. E PROCEDURE FOR CALIBRATION

    Date : 09.11.2008

    T-4.250.915 Rev. B Page 7 of 14

    QQQ AAA TTT AAA RRR

    6.3. CALIBRATION OF INSTRUMENTS

    Instruments subject to calibration shall be calibrated on the following occasions:

    6.3.1. Periodic Calibration

    This calibration will be carried out before the defined expiry dates by the QA/QC personnel or by a calibration service or agency under the control of the QA/QC Department.

    6.3.2. Re-calibration

    The instruments may be re-calibrated at any time before the expiry date of the calibration if there are doubts as to the accuracy of the equipment.

    6.4. PRECISION OF CALIBRATION

    The precision of the calibration shall be defined by the QA/QC Department in accordance with the class of precision stated by the manufacturer of the instrument.

    6.5. METHOD OF CALIBRATION

    The calibration of the devices shall be carried out according to the manufacturer instructions. In case of the calibration procedure and/or instruction of the device manufacturer is not available, then one of the following procedures shall apply:

    a) Calibration against MASTER devices,

    Master devices shall possess the manufacturers certificates.

    All primary "MASTER" instruments (2nd level) shall undergo calibration at predetermined periods depending on their characteristics and tendency to lose calibration. Secondary instruments (3rd level) shall be calibrated by comparison with the Masters.

    b) Calibration Control by direct measurement,

    Where and when applicable, the measuring devices shall be calibrated by direct measurement with appropriate and acceptable measuring methods. Measurements shall be carried out with certified devices.

    During the controls, if the device has calibration means, then it will be calibrated, supported with an appropriate measurement record.

    c) Calibration by an authorised institute,

    Calibration of the sophisticated devices, including optical measuring devices shall only be carried out by an authorised institute.

  • Doc. Title : GAM-PRO-QAC-GTL-028 Rev. E PROCEDURE FOR CALIBRATION

    Date : 09.11.2008

    T-4.250.915 Rev. B Page 8 of 14

    QQQ AAA TTT AAA RRR

    6.6. DOCUMENTATION OF CALIBRATION

    A "control list" will be compiled for the measuring and test equipment and shall give the relevant data.

    The calibration label (Attachment 2) of the calibrated equipment shall indicate the acceptance of the calibration. When this labelling is neither practical nor considered essential for the control purposes, other procedures shall be established to ensure the conformity with this requirement. Any other procedure shall be in compliance with the requirements of related specifications and subject to approval.

    Prior to use of any inspection, measuring and test instrument for the construction work, the responsible engineer shall assure that the effective calibration period of the device has not yet expired and the instrument has not been damaged to a visible degree.

  • Doc. Title : GAM-PRO-QAC-GTL-028 Rev. E PROCEDURE FOR CALIBRATION

    Date : 09.11.2008

    T-4.250.915 Rev. B Page 9 of 14

    QQQ AAA TTT AAA RRR

    7. ATTACHMENTS

    Attachment 1 Calibration Intervals

    Attachment 2 Calibration Labels

  • Doc. Title : GAM-PRO-QAC-GTL-028 Rev. E PROCEDURE FOR CALIBRATION

    Date : 09.11.2008

    T-4.250.915 Rev. B Page 10 of 14

    QQQ AAA TTT AAA RRR

  • Doc. Title : GAM-PRO-QAC-GTL-028 Rev. E PROCEDURE FOR CALIBRATION

    Date : 09.11.2008

    T-4.250.915 Rev. B Page 11 of 14

    QQQ AAA TTT AAA RRR

  • Doc. Title : GAM-PRO-QAC-GTL-028 Rev. E PROCEDURE FOR CALIBRATION

    Date : 09.11.2008

    T-4.250.915 Rev. B Page 12 of 14

    QQQ AAA TTT AAA RRR

    Equipment

    Calibration/

    Functional

    Test

    Calibration

    Period

    (Months)

    Remarks

    Calibration of survey

    instruments CAL 6 months

    Calibration of Relief valve CAL 1 week

    Calibration of MT yoke CAL 1 month

    Calibration of optical

    measuring devices CAL

    12 months and

    each testing

    Calibration of ultrasonic

    thickness measuring device CAL 12 month

    Calibration of surface

    thermometer CAL 2 years

    Calibration of PMI machine CAL 12months

    Calibration of Hygro meter CAL 12 months

    Calibration of Elco meter CAL 12 months

    Calibration of Holiday tester CAL 12 months

  • Doc. Title : GAM-PRO-QAC-GTL-028 Rev. E PROCEDURE FOR CALIBRATION

    Date : 09.11.2008

    T-4.250.915 Rev. B Page 13 of 14

    QQQ AAA TTT AAA RRR

    ATTACHMENT 2

    CALIBRATION LABELS

    INS

    S.NO

    __________ Date : ___/___/___

    Due __________ Sign __________

    CALIBRATION" LABEL COLOUR: WHITE

    CALIBRATED

  • Doc. Title : GAM-PRO-QAC-GTL-028 Rev. E PROCEDURE FOR CALIBRATION

    Date : 09.11.2008

    T-4.250.915 Rev. B Page 14 of 14

    QQQ AAA TTT AAA RRR

    "DO NOT USE" LABEL COLOUR: RED

    H O L D