Share Report Abuse Next Blog» Create Blog Sign In cGMP and Regulatory Updates This Blog is created for the updation of knowledge on current Good Manufacturing Practices (cGMP) and current Updates from all regulatory authorities. Also one can archive Standard Operating Procedures (SOPs) for any Pharma plant. FRIDAY, OCTOBER 31, 2008 Procedure for Leak testing of strips / blisters 1.0 OBJECTIVE To lay down a procedure for leak testing of strips / blisters. 2.0 SCOPE All packing areas. 3.0 RESPONSIBILITY Production officer. 4.0 ACCOUNTABILITY Production executive 5.0 PROCEDURE 5.1 Take the required number of strips / blisters as mentioned in the Annexure – I. Check the quality of strips/ blisters for any damages. 5.2 Tie the collected strips / blisters with a rubber band. 5.3 Ensure that all strips / blisters are dipped in water and close the lid. 5.4 Connect opening of the desiccator to the vacuum pump. 5.5 Apply a vacuum of 300 mm of Hg and close the knob of the desiccator. 5.6 Keep the vacuum for 30sec. 5.7 Release the vacuum by opening the knob of the desiccator slowly and open the lid remove the strips / blisters. 5.8 Wipe out the traces of water from the strips / blisters by using lint free duster . 5.9 Open the pocket of strips by using scissors to remove the tablets / capsules. 5.10 Open the blisters manually. Subscribe via email Enter your email address: Subscribe Delivered by FeedBurner FeedBurner FeedCount Blog Archive 2010 (18) 2009 (165) 2008 (376) December (43) Search Ads by Google Soil Testing Gas Leak Test Glucose Test Strips Leak Testing Hose Testing Calibration Testing METTLER TOLEDO calibrations meet ISO9001, ISO17025 & other standards www.mt.com/calibrationservice Blister Packaging Blister Machines Continuous Blister Packaging www.ulmapackaging.com Leak Test made easy Quick Connectors for high pressure leak testing of auto components www.weh.in/quick-coupling.html Great Lakes Vacuum We rebuild & repair vacuum pumps, blowers & also stock oil and parts! www.greatlakesvacuum.com Portable Vacuum Pump Economical,Cool & Quite, Long Life Versatile, Tough, Light weight ColeParmer.in/Vacuum_Pumps Water Sensor Switch Detect water leaks from 1/16 inch. Alarm or control based on detection www.antx.com 5/3/2010 cGMP and Regulatory Updates: Procedu… …vinvarun.biz/…/procedure-for-leak-tes… 1/7

_ Procedure for Leak Testing of Strips _ Blisters

Embed Size (px)

Citation preview

Page 1: _ Procedure for Leak Testing of Strips _ Blisters

Share Report Abuse Next Blog» Create Blog Sign In

cGMP and Regulatory UpdatesThis Blog is created for the updation of knowledge on current Good Manufacturing Practices (cGMP) and currentUpdates from all regulatory authorities. Also one can archive Standard Operating Procedures (SOPs) for anyPharma plant.

FRIDAY, OCTOBER 31, 2008

Procedure for Leak testing of strips /

blisters

1.0 OBJECTIVE

To lay down a procedure for leak testing of strips / blisters.

2.0 SCOPE

All packing areas.

3.0 RESPONSIBILITY

Production officer.

4.0 ACCOUNTABILITY

Production executive

5.0 PROCEDURE

5.1 Take the required number of strips / blisters as mentioned

in the Annexure – I. Check the quality of strips/ blisters for any

damages.

5.2 Tie the collected strips / blisters with a rubber band.

5.3 Ensure that all strips / blisters are dipped in water and close

the lid.

5.4 Connect opening of the desiccator to the vacuum pump.

5.5 Apply a vacuum of 300 mm of Hg and close the knob of the

desiccator.

5.6 Keep the vacuum for 30sec.

5.7 Release the vacuum by opening the knob of the desiccator

slowly and open the lid remove the strips / blisters.

5.8 Wipe out the traces of water from the strips / blisters by

using lint free duster .

5.9 Open the pocket of strips by using scissors to remove the

tablets / capsules.

5.10 Open the blisters manually.

Subscribe via email

Enter your email address:

Subscribe

Delivered by FeedBurner

FeedBurner FeedCount

Blog Archive

► 2010 (18)

► 2009 (165)

▼ 2008 (376)

► December (43)

Search

Ads by Google Soil Testing Gas Leak Test Glucose Test Strips Leak Testing Hose Testing

Calibration TestingMETTLER TOLEDO calibrationsmeet ISO9001, ISO17025 & otherstandardswww.mt.com/calibrationservice

Blister PackagingBlister Machines Continuous BlisterPackagingwww.ulmapackaging.com

Leak Test made easyQuick Connectors for high pressureleak testing of auto componentswww.weh.in/quick-coupling.html

Great Lakes Vacuum

We rebuild & repair vacuum

pumps, blowers & also stock oil

and parts!www.greatlakesvacuum.com

Portable Vacuum Pump

Economical,Cool & Quite, Long

Life Versatile, Tough, Light

weightColeParmer.in/Vacuum_Pumps

Water Sensor Switch

Detect water leaks from 1/16

inch. Alarm or control based on

detectionwww.antx.com

5/3/2010 cGMP and Regulatory Updates: Procedu…

…vinvarun.biz/…/procedure-for-leak-tes… 1/7

Page 2: _ Procedure for Leak Testing of Strips _ Blisters

5.11 Check for any traces of water inside the strips / blisters.

5.12 Number of tablets or capsules, which have become wet,

should not be more than 1%. If the leakage is more than the

above percentage repeat the leak test.

5.13 Record it in the respective batch packing record, with

signature and date.

ANNEXURE -I

Note:

1. Perform leak test at the starting of blistering / strip sealing,

operation.

2. If there is machine break down more than 1/2 hour perform

leak test immediately after starting the operation.

3. Whenever leakage observed during leak test, ensure that

rectification is done.

4. If leak test failure is occurring repeatedly, bring to the notice

of Executive / Sr. Executive for corrective action.at 8:05 PM

1 comments:

Post a Comment

► December (43)

► November (84)

▼ October (105)

Procedure for Cleaning of

Production areas after B...

Maintenance of dies and

punches.

Procedure for cleaning of

HDPE pallets.

Cleaning procedure for

Inprocess container

Correction of wrong entry in

BMR / BPR

Procedure for Leak testing of

strips / blisters

Recording and monitoring of

differential air press...

Procedure for defoiling of

blisters / strips

Procedure for handling of

finished goods

Returning of unused Raw and

Packing materials

Destruction of old, torn

rejected FBD and FEM

bags...

Transferring of Materials

from Dispensing to Modul...

Cleaning procedure for New

Equipment

Transferring of machine

under major break downs

Procedure for Inventory of

Sieves

Maintenance of Dies and

Punches stock card.

Returning of on-line rejected

packing material

Packing line start up and

closing procedure

Handling of New Machines

Procedure for Indenting of

Stereos

Returning of on-line rejected

raw Material

Procedure for Disposal of

Materials

Exit Procedure - Annexure Y

Entry Procedure - Annexure Y

vicky said...

But mention about the amount of water to be

taken,also.

November 3, 2008 11:55 AM

RTD Calibration GuideHow to Calibrate a PT100 or RTD FlukeApplication Note Downloadwww.HartScientific.com

Torque testerquality control devices online & laboratorywww.at2e.com

5/3/2010 cGMP and Regulatory Updates: Procedu…

…vinvarun.biz/…/procedure-for-leak-tes… 2/7

Page 3: _ Procedure for Leak Testing of Strips _ Blisters

Newer Post Older PostHome

Subscribe to: Post Comments (Atom)

Procedure for cleaning of

Production Aids

Procedure for Maintaining

Environmental Conditions...

Gowning room Exit

Procedure

Gowning room Entry

Procedure

Exit Procedure - Annexure X

HVAC SYSTEM - BASIS OF

DESIGN PARAMETERS &

EQUIPME...

Qualification Protocol of

Heating, Ventilation and...

SOP - PAINTING OF THE

FACILITY

INSTRUMENT OPERATION

PROCEDURE DISSOLUTION

APPARAT...

INSTRUMENT CALIBRATION

DISINTEGRATION TEST

APPARA...

INSTRUMENT OPERATION

PROCEDURE

DISINTEGRATION TEST...

INSTRUMENT OPERATING

PROCEDURE UV-

SPECTROPHOTOMETE...

OPERATION OF HIGH

PRESSURE LIQUID

CHROMATOGRAPHY

INSTRUMENT CALIBRATION

PROCEDURE GAS

CHROMATOGRAPH...

CALIBRATION OF FLUID BED

DRIER TEMP GUAGE

CALIBRATION OF 6- STATION

DISSOLUTION APPARATUS

CALIBRATION OF DRIER

TEMPERATURE GUAGE

CALIBRATION OF DRY AND

WET BULB THERMOMETER

CALIBRATION OF FLAME

PHOTOMETER

CALIBRATION OF

FLUORIMETER

CALIBRATION OF HPLC

CALIBRATION OF INFRA RED

SPECTROPHOTOMETER

CALIBRATION OF MIXING

Oil Field TubularsPup Joints, X & R Nipples TubularAccessories,Flow Couplingwww.petroquip.com

Vacuum TechnologyIn manufacturing equipment at lowcost.Welcome:086 024-23820832!www.ulvac-usy.com

5/3/2010 cGMP and Regulatory Updates: Procedu…

…vinvarun.biz/…/procedure-for-leak-tes… 3/7

Page 4: _ Procedure for Leak Testing of Strips _ Blisters

TANK (GRADUATED)

CALIBRATION OF pH METER

INSTRUMENT OPERATION

PROCEDURE - KARL FISHER

TITRA...

CALIBRATION OF FLAME

PHOTOMETER

CALIBRATION OF UV-VISIBLE

SPECTROPHOTOMETER

CALIBRATION OF

VISCOMETER

WATER SAMPLING &

ANALYSIS FOR VALIDATION

VISUAL INSPECTION OF

CONTROL SAMPLES

PROCEDURE FOR VIABLE

PARTICLE COUNT BY AIR

SAMPLIN...

AUTOCLAVE VALIDATION

PROTOCOL

B12 Assay Method

PROCEDURE FOR ASSAY OF

ANTIBIOTICS BY DIFFUSION

ME...

BIOLOAD IN PRIMARY

PACKING MATERIALS

FUNCTIONS OF

MICROBIOLOGY

DEPARTMENT

Growth Promotion Test

IDENTIFICATION TESTS FOR

PURE CULTURES

PROCEDURE FOR

FUMIGATION and DE-

FUMIGATION

RECEIPT AND MAINTAINANCE

OF CONTROL SAMPLE

ART WORK

ANALYSIS AT PUBLIC

LABORATORY

Rejection of non-confirming

product

Personnel Validation

Out of Specification Results

LABELLING PROCEDURE AND

CONTROL

LABELING PROCEDURE FOR

REAGENTS AND SOLVENTS

Corrective and Preventive

5/3/2010 cGMP and Regulatory Updates: Procedu…

…vinvarun.biz/…/procedure-for-leak-tes… 4/7

Page 5: _ Procedure for Leak Testing of Strips _ Blisters

action SOP

BIOLOAD IN STEAM

PROCEDURE FOR SAMPLING

& ANALYSIS OF

COMPRESSED AI...

PROCEDURE FOR

MONITORING THE

BACTERIALOGICAL QUALI...

INSTRUMENT OPERATION

B.O.D. INCUBATOR

PROCEDURE FOR

CALIBRATION OF B.O.D.

INCUBATOR

PROCEDURE FOR

DESTRUCTION OF USED

MEDIA

DISINFECTANT EFFICACY

TEST (BY SWABING

METHOD)

PROCEDURE FOR

MONITORING

BACTERIOLOGICAL

QUALITY O...

METHOD FOR THE

ANTIMICROBIAL EFFICACY

TEST

PROCEDURE FOR

ENVIRONMENTAL

MONITORING BY SETTLE

P...

PROCEDURE FOR FINGER

DAB TEST - SOP

PROCEDURE FOR

CALIBRATION OF

AUTOCLAVE -SOP

INSTRUMENT OPERATION

REFRIGERATOR

INSTRUMENT

OPERATION,CLEANING &

CALIBRATION OF MIC...

Instrument Operation Of

Digital Colony Counter

PROCEDURE FOR OPERATION

AND CLEANING OF

AUTOCLAVE

Instrument Operation of

Antibiotic Zone reader

BATCH RECORD REVIEW SOP

ISSUE, ENTRY, REVIEW AND

CONTROL OF BATCH

RECORDS

5/3/2010 cGMP and Regulatory Updates: Procedu…

…vinvarun.biz/…/procedure-for-leak-tes… 5/7

Page 6: _ Procedure for Leak Testing of Strips _ Blisters

DOCUMENT CONTROL SOP

STANDARD OPERATING

PROCEDURE FOR SOP

ROLE OF QUALITY

ASSURANCE MANAGER

ROLE OF QUALITY

ASSURANCE

WATER SYSTEM VALIDATION

INSTRUMENT CALIBRATION -

THERMOMETER

INSTRUMENT CALIBRATION

UV LIGHT VIEW CHAMBER

INSTRUMENT OPERATION UV

LIGHT VIEW CHAMBER -

SOP

INSTRUMENT OPERATION

VACUUM OVEN -SOP

INSTRUMENT CALIBRATION

BROOKFEILD VISCOMETER

INSTRUMENT OPERATION

BROOKFEILD VISCOMETER

CLEANING OF SAMPLING

EQUIPMENT

HANDLING OF HAZARDOUS

CHEMICALS - SOP

CALIBRATION OF

VOLUMETRIC GLASS

WARES

► September (81)

► August (38)

► July (25)

Sponsored Link

Subscribe Now: Feed Icon

Subscribe in a reader

Favorite Links

Insurance-analyzer

About Water

Pharma-Career-Jobs

5/3/2010 cGMP and Regulatory Updates: Procedu…

…vinvarun.biz/…/procedure-for-leak-tes… 6/7

Page 7: _ Procedure for Leak Testing of Strips _ Blisters

Standard

Operating

Procedure

Christopher

Bradle...

New $30.49

Best $16.98

160th AVN GP

Multiple Authors

Best $50.00

Taxi To the Dark

Side

Alex Gibney,

Brian...

New $26.99

Best $6.49

Standard

Operating

Procedure

New $14.99

Best $14.99

Ricerca gets

citations for failure

t...

Washington

Informa...

New $5.95

Best $5.95

Privacy Information>

Search

Subscribe via email

Subscribe to cGMP & Regulatory

Updates by Email

5/3/2010 cGMP and Regulatory Updates: Procedu…

…vinvarun.biz/…/procedure-for-leak-tes… 7/7