Upload
hoangxuyen
View
219
Download
6
Embed Size (px)
Citation preview
[ Care and Use ManUal ]
Sep-Pak Cartridges and Plates 1
Contents
I. IntroduCtIon
II. usInG seP-PAK CArtIdGes And PLAtes
III. strAteGIes For soLId-PHAse eXtrACtIona. Retention-Cleanup-Elution Strategy
b. Pass-Through Cleanup Strategy
IV. stePs oF An sPe ProCedure
V. storAGe And dIsPosAL oF used CArtrIdGes
VI. AddItIonAL InForMAtIon
VII. LIterAture reFerenCes
seP-PAK CArtrIdGes And PLAtes
Waters Sep-Pak® cartridges and 96-well plates are convenient,
reproducible, disposable solid-phase extraction (SPE) devices
for sample preparation. Sep-Pak cartridges and plates are
manufactured in a Waters ISO 13485 and ISO 9002-registered
facility in compliance with cGMP guidelines for the U.S. Food and
Drug Administration for Class 1 Medical Devices.
[ Care and Use ManUal ]
Sep-Pak Cartridges and Plates 2
I. IntroduCtIon
Sep-Pak devices are available in many styles to accommodate
manual and automated solid-phase extraction. Devices are available
in convenient Plus, Light, Classic, Vac, Vac RC, and 96-well plate
formats.
Available Sorbent Chemistries:
Reversed Phase: C18, tC18, C8, tC2
Normal Phase: Silica, Alumina A, Alumina B, Alumina N, Florisil®
Ion Exchange and Mixed Mode: Aminopropyl, PSA, Accell™ Plus
QMA, Accell Plus CM, Cyanopropropyl, Diol
Speciality Products: Porapak® RDX, DNPH-Silica, XPoSure™, Ozone
Scrubber, Dry, PS2, AC2, Carbon Black/Aminopropyl
II. usInG seP-PAK CArtrIdGes And 96-WeLL PLAtes
Below is a five step guideline for using Sep-Pak cartridges. Each separation
is different and not all steps may be required for your application. Typical
strong and weak solvents used in the sample preparation steps are listed in
Table 1 below.
Table 1: Typical strong and weak solvents used in the
sample preparation.
Prepare sample: The sample can be applied to the Sep-Pak Cartridge in
either liquid or a gas phase. If the sample is a solid, it must be dissolved or
extracted prior to loading.
Condition/Equilibrate Cartridge: A conditioning step is required for
reversed-phase sorbents (e.g., C18). Use a strong solvent to wet the station-
ary phase. Conditioning is followed by an equilibration step using a weak
solvent.
Load Sample: The sample is loaded onto the cartridge.
Wash: The wash step removed interferences while retaining the analyte. If
weakly retained interferences are present in the sample matrix, wash with an
appropriate weak solvent.
Elute: Use an appropriate stronger solvent to elute the analyte while retain-
ing more strongly bound interferences.
III. strAteGIes For soLId-PHAse eXtrACtIon
Purification: Use as a “chemical filter” to retain the component of interest
while interferences elute, or to retain the interferences while the analyte
passes through unretained.
Trace Enrichment or Concentration: Use to concentrate an analyte from
a dilute solution. This is done when the analyte concentration is below the
detection limit of the analytical method.
Fractionation: Use with a step gradient of increasing solvent strength to
selectivity elute and isolate analytes based on differences in polarity.
Solvent Exchange: Use to adsorb the analyte and elute with a
desirable solvent. This is helpful of the sample is dissolved in a solvent
that is incompatible with the analytical method.
Separation Mode Typical Weak Solvents
Typical Strong Solvents
Reversed Phase Water or bufer. May contain low concentrations of organics, such as methanol.
Acetonitrile, methanol, or aqueous/organic mixture
Normal Phase Hexane, toluene Methylene chloride, methanol
Weak Cation Exchange
Low ionic strength buffer, pH >4
High ionic strength buffer, or pH <2
Strong Anion Exchange
Low ionic strength buffer, pH <8
High ionic strength buffer, or pH >10
[ Care and Use ManUal ]
Sep-Pak Cartridges and Plates 3
IV. stePs oF An sPe ProCedure
The following section describes the steps involved in a complete solid-phase
extraction procedure:
1. Pretreatment of Sample
Solid samples (soil, tissue, etc.)
• Shake, sonicate or use soxhlet extraction.
- extract sample with polar organic solvent (methanol,
acetonitrile) for polar analytes.
- extract sample with organic solvent and drying agent (dichloromethane, acetone) for non-polar analytes and
multiresidue extraction.
Non Aqueous Liquid • If the sample is soluble in water, dilute it with water for
reverse-phase SPE.
• If the sample is soluble in hexane, dilute it with or exchange to
hexane for SPE.
Wastewater• Filter or centrifuge as necessary.
2. Condition StepFor reversed-phase sorbents, preconditioning of the sorbent with an
organic solvent, such as methanol, acetonitrile, isopropanol, or tetrahy-
drofuran, is usually necessary to obtain reproducible results. Without
this step, a highly aqueous solvent cannot penetrate the hydrophobic
surface and wet the sorbent. Thus, only a small fraction of the sorbent
surface area would be available for interaction with the analyte. For the
same reason, it is important not to let silica-based SPE cartridges dry out
between the solvation step and the addition of the sample. A complete
preconditioning of a reversed-phase cartridge includes the solvation step
and an equilibration with a low-strength solvent such as water or buffer.
a. Retention-Cleanup-Elution StrategyAs the sample is loaded onto the cartridge, the analytes of interest are
retained by the sorbent. If needed, an optimized series of washes are
used to remove matrix interference from the cartridge. A strong solvent
is used to elute the analytes from the cartridge. Sample enrichment
results when the final elution volume is smaller than the load volume.
Figure 1: Retention-Cleanup-Elution
b. Pass-Through Cleanup StrategyPass-through cleanup methods optimize matrix retention while the
analytes of interest pass-through the cartridge unretained. No sample
enrichment occurs during the solid-phase extraction (SPE) step.
1. Sample is passed through sorbent and collected
• No sample enrichment
2. Matrix interferences are retained on sorbent
Figure 2: Pass-Through
Load Sample(Black)
StepElute 1
StepElute 2
One cartridge can separate all three dyes
StepElute 3
NOTE: Different strength solvents can be used to separate the dyes.
Stationary Phase
Particles
[ Care and Use ManUal ]
Sep-Pak Cartridges and Plates 4
5. Elution StepOnce the interferences are washed off the cartridge, a strong
solvent is introduced to elute the analytes of interest. The volume
and flow rate of the eluting solvents should be precisely controlled as
in the load step to ensure reproducible results.
Table 2. Guidelines on the various types of separation mechanisms
V. storAGe And dIsPosAL oF used CArtrIdGes
Cartridges stored in their original sealed pouch remain stable for long
periods. To store unused cartridges in opened pouches, squeeze the air out
of the pouch, fold over the open end of the pouch twice, seal with tape, and
store in a desiccator.
Dispose of used cartridges safely in accordance with applicable government
or local regulations.
3. Load StepWhen the analytes of interest are not retained by the sorbent, this is
called analyte breakthrough. For some methods, such as pass-through
cleanup, analyte breakthrough is desirable and is maximized for
those specific methods. However, in all other cases, analyte break-
through is unwanted and contributes to poor recovery and method
reproducibility. Breakthrough occurs when:
• There is too high an organic concentration in the load
solution for very polar analytes. Dilute sample at least 1:1
with water or buffer prior to loading.
• The analytes are bound to proteins, they may pass through
the sorbent. Ensure that analytes are not bound to proteins
by acidifying or basifying the sample.
• Sorbent is overloaded by the matrix component.
Therefore, it is important to choose the correct sorbent
mass (see Table 1).
• The flow rate of the load step is too fast. There is not
enough contact time between the analytes and the sorbent.
Look at the drops and adjust the vacuum so that you see
discrete droplets, not a stream of liquid.
Table 1. Choice of cartridges based on sample size
* LP=large particules (60 µm)
4. Wash StepThe wash steps are designed to remove unwanted matrix components
that remain from the loading step. The ideal wash solvent removes
only the matrix while keeping the analytes bound to the sorbent. For
complex samples this is impossible, so the wash steps are optimized
using pH, solvent strength, and solvent polarity to remove as much
matrix as possible while maintaining acceptable analyte recovery.
Sample Size Cartridge
1 - 10 mL 1 cc/30 mg or 3 cc/60 mg
10 - 100 mL 3 cc/60 mg or 6 cc/200 mg
100 - 500 mL 6 cc/200 mg or 6 cc/500 mg (LP*)
500 - 1000 mL 6 cc/500 mg (LP) or 12 cc/1 g (LP*)
Reversed Phase Normal Phase Ion Exchange
Analyte Low to moderate polarity/hydrophobic
Moderate to highly polar/uncharged
Charged or ionized
Matrix Aqueous Non-polar organic solvent
Aqueous/low ionic strength
Condition/
Equilibrate
1. Solvate polar organic 2. Water
Non-polar organic
Low ionic strength buffer
Wash Aqueous/buffer Non-polar Low ionic strength buffer
Elution
Steps
Increase polar organic content in steps
Increase moderate to high polarity organic content in steps
Stronger buffers–ionic strength or pH to neutralize the charge
[ Care and Use ManUal ]
Sep-Pak Cartridges and Plates 5
Table 4: Accessories and Spare Parts for the Waters Extraction Manifolds
Description Qty. Part No.
Needle valves(required when using silica-based SPE cartridges)
20/pk WAT200685
Needle 20/pk WAT200691
Cover, 20 position without gauge assembly 1 WAT200686
Gauge assembly, Vacuum 1 WAT200687
Reservoir, glass with outlet valve 1 WAT200688
Outlet valve kit 1 WAT200689
Gasket for cover 1 WAT200690
Ejector tool 1 WAT058839
Luer plugs 25/pk WAT058851
Rubber ball ring (for vacuum gauge assembly) 1 WAT058840
Reversible vial rack for 1 mL or 4 mL Autosampler vials
1 WAT058871
13 x 75 mm test tube rack 1 WAT200678
13 x 100 mm test tube rack 1 WAT200679
16 x 75 mm test tube rack 1 WAT200680
16 x 100 mm test tube rack 1 WAT200681
Reservoir, 30 cc (for Sep-Pak Plus, Light, Vac & Classic cartridges)
48/pk WAT011390
Reservoir, 60 cc (for Sep-Pak Plus, Light & Vac cartridges)
12/pk WAT024659
Adapter, male-male Luer (for Sep-Pak Classic cartridges)
100/pk WAT024310
Adapter (to attach reservoir to 1, 3 & 6 cc Sep-Pak Vac cartridges)
12/pk WAT054260
Adapter (to attach reservoir to 12, 20 & 35 cc Sep-Pak Vac cartridges)
10/pk WAT048160
Vacuum pump (110 V, 60 Hz) 1 WAT085114
Vacuum pump (220 V, 50 Hz) 1 WAT085115
VI. AddItIonAL InForMAtIon
Sep-Pak accessories are available to help with your solid-phase extraction.
Ordering information for these products is given below.
For more detailed information, consult the Chromatography Columns
and Supplies Catalog by visiting www.waters.com (literature code:
720002419EN), or contact your Waters representative.
If you require technical assistance, or have any questions regarding sample
preparation, visit www.waters.com or call us at 1-800-252-4752.
Table 2: Sep-Pak Cartridge Accessories
Table 3: Extraction Manifolds (Includes 20 needle tips, 25 plugs and
ejector tool)
Description Qty. Part No.
30 cc Reservoir 48/pk WAT011390
60 cc Reservoir 12/pk WAT024659
Reservoir adapters for 1, 3 and 6 cc VAC 12/pk WAT054260
Reservoir adapters for 12, 20 and 35 cc VAC 10/pk WAT048160
Male-male adapter 100/pk WAT024310
Male luer plugs 100/pk WAT044395
Female luer plugs 100/pk WAT044385
Sep-Pak connections kit 1 WAT011400
Description Part No.
20 position extraction manifold without rack WAT200677
20 position extraction manifold with 13 x 75 mm test tube rack
WAT200606
20 position extraction manifold with 13 x 100 mm test tube
WAT200607
20 position extraction manifold with 16 x 75 mm test tube rack
WAT200608
20 position extraction manifold with 16 x 100 mm test tube
WAT200609
[ Care and Use ManUal ]
Sep-Pak Cartridges and Plates 6
VII. reLAted LIterAture
Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com
©2008 Waters Corporation, Waters. The Science of What’s Possible, Sep-Pak, Porapak, Accell, and XPoSure are trademarks of Waters Corporation. Florisil is a trademark of U.S. Silica.Sep-Pak Cartridges are covered by U.S. Patents # 4,211,658 and/or # 4,250,035 and several foreign patents.
June 2008 WAT011188 Rev 8 KK-PDF
Sales Offices:
Austria and European Export
(Central South Eastern Europe,
CIS and Middle East) 431 877 18 07
Australia 2 9933 1777
Belgium 32 2 726 1000
Brazil 55 11 5094 3788
Canada 800 252 4752
China 8621 6495 6999
CIS/Russia +7 495 3367000
Czech Republic 42 02 617 11384
Denmark 45 46 59 8080
Finland +358 9 5659 6288
France (33) 1 30 48 72 00
Germany 49 6196 400600
Hong Kong 852 29 64 1800
Hungary 36 1 350 5086
India and India Subcontinent
91 80 2 837 1900
Ireland 353 1 448 1500
Italy 39 02 274 211
Japan (81) 3 3471 7191
Korea (82) 2 820 2700
Mexico 5255 5200 1860
The Netherlands +31 (0)76-50 87 200
Norway 47 63 84 60 50
Poland (48) 22 833 4400
Puerto Rico 787 747 8445
Singapore 65 6273 1221
Spain 34 93 600 93 00
Sweden 46 8 555 11500
Switzerland 41 62 889 2030
Taiwan 886 2 2543 1898
United Kingdom 44 208 238 6100
Document Literature Reference
A Sample Preparation Primer and Guide to Solid-Phase Extraction Methods Development
WA20300
Sample Preparation Solutions Brochure 20000848EN
96-Well Collection Plate Options for the Waters Extraction Plate Manifold
720001263EN
Waters Sep-Pak Sample Extraction Products Brochure 720000860EN
Waters 96 and 384-Well Collection Plate Specifications
WA41941
Sorbent Selection Guide for SPE Wall Chart 720002007EN
Environmental Analysis Catalog 720002639EN
Environmental Chromatography Methods Guide 720002543EN
Food Safety Application Notebook 720002565EN