Biopotentials and Electrophysiology

Measurement

Page 3: بنام خداوند عليم

What are biopotentials

Biopotential: An electric potential that is measured between points in living cells, tissues, and organisms, and which accompanies all biochemical processes.

• Also describes the transfer of information between and within cells

Page 4: بنام خداوند عليم

Mechanism behind biopotentials

• Concentration of potassium (K+) ions is 30-50 times higher inside as compared to outside

• Sodium ion (Na+) concentration is 10 times higher outside the membrane than inside

• In resting state the member is permeable only for potassium ions

Potassium flows outwards leaving an equal number of negative ions inside

Electrostatic attraction pulls potassium and chloride ions close to the membrane

Electric field directed inward formsElectrostatic force vs. diffusional force

• Nernst equation:

• Goldman-Hodgkin-Katz equation:

mVVm 100...70

ko

ki

kk c

c

Fz

RTV

,

,ln

CliClNaiNaKiK

CliClNaiNaKiK

km cPcPcP

cPcPcP

Fz

RTV

,,,

,,,ln

restmV ,mVVm 100...70

Page 5: بنام خداوند عليم

Mechanism behind biopotentials • When membrane stimulation exceeds a

threshold level of about 20 mV, so called action potential occurs:1. Sodium and potassium ionic permeabilities

of the membrane change2. Sodium ion permeability increases very

rapidly at first, allowing sodium ions to flow from outside to inside, making the inside more positive

3. The more slowly increasing potassium ion permeability allows potassium ions to flow from inside to outside, thus returning membrane potential to its resting value

4. While at rest, the Na-K pump restores the ion concentrations to their original values

• The number of ions flowing through an open channel >106/sec

• Body is an inhomogeneous volume conductor and these ion fluxes create measurable potentials on body surface

Page 6: بنام خداوند عليم

0 mV

Page 7: بنام خداوند عليم

-80 mV

Page 8: بنام خداوند عليم

-80 mV

+++ ++

+

+

--

-

--

-

- +

Page 9: بنام خداوند عليم

-80 mV

[K+] = 135[Na+] = 7[Cl-] = 11A-

[K+] = 2.5[Na+] = 125

[Cl-] = 130A-

+++ ++

+

+

--

-

--

-

- +

Page 10: بنام خداوند عليم

Example: Na+ and K+ channels

Page 11: بنام خداوند عليم

The status of channels

The opening ofNa+ channel

Inactive of Na+ channel;Opening K+ channel

Closing of Na+ &K+ channels

The opening ofleakage channel

Page 12: بنام خداوند عليم

Electrocardiography (ECG)

• Measures galvanically the electric activity of the heart• Well known and traditional, first measurements by

Augustus Waller using capillary electrometer (year 1887)• Very widely used method in clinical environment• Very high diagnostic value

Page 13: بنام خداوند عليم
Page 14: بنام خداوند عليم

12-Lead ECG measurement

• Most widely used ECG measurement setup in clinical environment• Signal is measured non-invasively with 9 electrodes• Lots of measurement data and international reference databases• Well-known measurement and diagnosis practices• This particular method was adopted due to historical reasons, now it is

already rather obsolete

Einthoven leads: I, II & III Goldberger augmented leads: VR, VL & VF Precordial leads: V1-V6

Page 15: بنام خداوند عليم
Page 16: بنام خداوند عليم
Page 17: بنام خداوند عليم
Page 18: بنام خداوند عليم
Page 19: بنام خداوند عليم
Page 20: بنام خداوند عليم
Page 21: بنام خداوند عليم

1. Atrial depolarization

2. Ventriculardepolarization

3. Ventricular repolarization

Page 22: بنام خداوند عليم
Page 23: بنام خداوند عليم

ECG basics

• Amplitude: 1-5 mV• Bandwidth: 0.05-100 Hz

• Largest measurement error sources:• Motion artifacts• 50/60 Hz powerline interference

• Typical applications:• Diagnosis of ischemia• Arrhythmia• Conduction defects

Page 24: بنام خداوند عليم

Electroencephalography (EEG)

• Measures the brain’s electric activity from the scalp

• Measured signal results from the activity of billions of neurons

• Amplitude: 0.001-0.01 mV

• Bandwidth: 0.5-40 Hz

• Errors:• Thermal RF noise• 50/60 Hz power lines• Blink artifacts and similar

• Typical applications:• Sleep studies• Seizure detection• Cortical mapping

Page 25: بنام خداوند عليم
Page 26: بنام خداوند عليم

EEG measurement setup• 10-20 Lead system is most

widely clinically accepted• Certain physiological

featuresare used as reference points

• Allow localization of diagnostic features in the vicinity of the electrode

• Often a readily available wire or rubber mesh is used

• Brain research utilizes even 256 or 512 channel EEG hats

Page 27: بنام خداوند عليم

Electromyography (EMG)

• Measures the electric activity of active muscle fibers• Electrodes are always connected very close to the muscle

group being measured• Rectified and integrated EMG signal gives rough indication

of the muscle activity• Needle electrodes can be used to measure individual muscle

fibers

• Amplitude: 1-10 mV• Bandwidth: 20-2000 Hz

• Main sources of errors are 50/60 Hz and RF interference

• Applications: muscle function, neuromuscular disease, prosthesis

Page 28: بنام خداوند عليم

Electrooculography (EOG)

• Electric potentials are created as a result of the movement of the eyeballs

• Potential varies in proportion to the amplitude of the movement• In many ways a challenging measurement with some clinical

value

• Amplitude: 0.01-0.1 mV• Bandwidth: DC-10 Hz

• Primary sources of error include skin potential and motion

• Applications: eye position, sleep state, vestibulo-ocular reflex

Page 29: بنام خداوند عليم
Page 30: بنام خداوند عليم

FREQUENCIES OF BIOPOTENTIALS

Signal Frequency range (Hz) Amplitude range(mV)

ECG 0.01 – 100 0.05 – 3

EEG 0.1 – 80 0.001 – 1

EOG 0.01 – 10 0.001 – 0.3

EMG 50 – 3000 0.01 – 100

Page 31: بنام خداوند عليم