37
ADAP-GC: Deconvolution of Co-Eluting Metabolites from GC/TOF-MS Data for Metabolomics Studies Xiuxia Du Department of Bioinformatics and Genomics University of North Carolina at Charlotte

Pittcon 3 5-2014

Embed Size (px)

Citation preview

ADAP-GC: Deconvolution of Co-Eluting Metabolites from GC/TOF-MS

Data for Metabolomics Studies

Xiuxia Du

Department of Bioinformatics and Genomics University of North Carolina at Charlotte

Outline

•  Background

•  Why is deconvolution necessary?

•  How is deconvolution done in ADAP-GC?

•  ADAP-GC software

•  Next step

2

GC-MS vs. LC-MS M E T A B O L O M I C SO N C H N S C O H P C N S H C O

2

Selection Guide

Alkylsilyl derivativesEicosanoidsEssential oilsEstersPerfumesTerpenesWaxesVolatilesCaratenoidsFlavenoidsLipids

AlcoholsAlkaloidsAmino acidsCatecholaminesFatty acidsPhenolicsPolar organics ProstaglandinsSteroids

Organic AcidsOrganic AminesNucleosidesIonic SpeciesNucleotidesPolyamines

Less Polar More Polar

GC/MSGC/MS LC/MS

overlap

Figure 1. Classes of chemicals and the analytical techniques with which they aremost compatible.

A simple identification method uses onlythe background-subtracted EI spectraand a search of a general-purpose EIlibrary such as the NIST library. A morepowerful identification method involvessearching both chromatographicretention time and mass spectra of ananalyte against an application-specificlibrary containing expected retentiontimes and EI spectra for compounds.Currently, there are no publicly availableEI libraries devoted exclusively toendogenous metabolites.

With EI, the molecular ion is often lost.If the metabolite spectrum is not in thelibrary being searched, the absence ofmass information for the molecular ionmakes it difficult to limit the number ofchemical possibilities. Complementarychemical ionization (CI) can be used topreserve the molecular ion, but it comeswith the loss of the structural informationthat EI fragmentation provides. Therefore,GC/MS is best suited for targetingknown or anticipated metabolites.

The cost of GC/MS systems issubstantially less than that of LC/MSsystems.

LC/MS analysis

Liquid chromatography can separatemetabolites that are not volatile andhave not been derivatized. As a result,LC/MS can analyze a much wider rangeof chemical species than GC/MS.Samples commonly analyzed by LC/MSinclude amino acids (18 out of 20 aminoacids can be derivatized, but theremaining two can not) and sugarslarger than trimers.

Electrospray ionization (ESI) andatmospheric pressure chemicalionization (APCI) are the two ionizationtechniques most commonly used inLC/MS. Unlike EI for GC/MS, ionsuppression can occur with both ESI andAPCI so co-eluting compounds may beunderestimated or not detected at all.Therefore, for complex samples, greaterseparation is necessary for the reliableLC/MS results. Unlike GC/MS, LC/MSalmost always produces a molecular ionthat can be used to limit the possibleidentities of a given analyte.

There are no spectral libraries forLC/MS identification. However, becausethe molecular ion is usually present inLC/MS analyses, it's mass can be usedto search a database of metabolitessuch as the METLIN database. Inaddition, the development of accurate-mass time-of-flight mass spectrometershas enabled the calculation of anempirical formula from the molecularion. To truly address unknownmetabolites, MS/MS fragmentation on a Q-TOF with manual de novointerpretation is the next step in theidentification process.

LC/MS is best suited for a discovery-based approach when researchingunknown metabolites, or when many ofthe targeted metabolites are not readilyamenable to GC/MS analysis due tovolatility issues.

The cost of an LC/MS system issubstantially more than that of a GC/MS system.

3

Ionization

•  Electron ionization (EI) • Hard method

•  Small molecules, 1-1000 Da

•  Electrospray ionization (ESI) •  Soft method

•  Small molecules, peptides, proteins, up to 200,00 Da

4

EI

M + e− →M +• + 2e−

EI fragmentation of CH3OHCH3OH→CH3OH

+

CH 3OH→CH2O = H + +HCH3OH→+ CH3 +OHCH2O = H + →CHO ≡ H + +H

5

EI breaks up molecules …

Molecular ion

in predictable ways.

6

GC-EI-MS

EI MS

7

Raw data

8

Deconvolution

3

As a review, let's look at the deconvolution process. AMDISconsiders the peak shapes of all extracted ions and their apexretention times (RT). In this example, only some of theextracted ion chromatograms (EICs) are overlaid for claritywith the apex spectrum (Figure 1A).

Figure 1A-1C. Simplified deconvolution process (continued).

Figure 1A

50

170280

31075

185

160

Extracted IonChromatograms(EIC)

After de-skewing

50

170

280

75 late retention time

185 shape & early retention time

310 early retention time

160 shape

Same shape and sameretention time

50

170280

31075

185

160

Extracted IonChromatograms(EIC)

Figure 1B

50

170

280

Only the ions in blackhave the same shapeand retention time asshown by 50, 170, 280-plus others

Figure 1B shows the EICs after the different peak shapes or RTs are eliminated from Figure 1A. Ions 50, 170, 280 and a few others remain.

Ion 160 EIC has the same RT as ions 50, 170 and 280, but hasa different peak shape. Ion 185 has a different peak shape andan earlier RT. Ions 75 and 310 have similar peak shapes butthey have different RTs.

www.agilent.com

9

Deconvolution

3

As a review, let's look at the deconvolution process. AMDISconsiders the peak shapes of all extracted ions and their apexretention times (RT). In this example, only some of theextracted ion chromatograms (EICs) are overlaid for claritywith the apex spectrum (Figure 1A).

Figure 1A-1C. Simplified deconvolution process (continued).

Figure 1A

50

170280

31075

185

160

Extracted IonChromatograms(EIC)

After de-skewing

50

170

280

75 late retention time

185 shape & early retention time

310 early retention time

160 shape

Same shape and sameretention time

50

170280

31075

185

160

Extracted IonChromatograms(EIC)

Figure 1B

50

170

280

Only the ions in blackhave the same shapeand retention time asshown by 50, 170, 280-plus others

Figure 1B shows the EICs after the different peak shapes or RTs are eliminated from Figure 1A. Ions 50, 170, 280 and a few others remain.

Ion 160 EIC has the same RT as ions 50, 170 and 280, but hasa different peak shape. Ion 185 has a different peak shape andan earlier RT. Ions 75 and 310 have similar peak shapes butthey have different RTs.

www.agilent.com

10

Deconvolution

4

Deconvolution finds the components from a complex TIC.Each component is searched against a retention time locking(RTL) library in AMDIS format. In addition to spectral match-ing, the locked RT can also be used as a criterion for hits.Depending on the match factor from the search, target com-pounds can be identified or flagged in a complex TIC. Thepower of deconvolution is appreciated while comparing thetop two spectra in Figure 2. The raw scan or original nonde-convoluted scan is shown on top. The clean scan, that is the

deconvoluted component, is shown in the middle. The bottomscan is the identified compound in the AMDIS library.Without deconvolution, the analyst would visually comparethe background subtracted raw scan and library scans forconfirmation. It would be very difficult, if not impossible, tosay that Fenbuconazole, the target compound in this example,is present using that type of comparison.

50

170280

Extracted IonChromatograms(EIC)

Figure 1C

Thesedeconvoluted ions are groupedtogether as a component

50

170

280

Figure 1C shows all of the ions in black that have similar peak shapes and RTs, within the criteria set earlier by the analyst. These aregrouped together and referred to as a component by AMDIS.

Figure 1A-1C. Simplified deconvolution process (continued). www.agilent.com

11

GC-EI-MS data processing workflow

peak detection

deconvolution alignment denoising

baseline correction

library search

EIC extraction

raw MS data

12

•  For low-resolution mass measurement: relatively easy

•  For high-resolution mass measurement: more involved

EIC extraction

13

Peak picking

•  Each EIC chromatographic peak is characterized by its apex elution time, left and right boundary, peak height, and peak shape.

!14

Background ADAP-GC ADAP-LC ADAP-Stats ADAP-CAGT

An automated data analysis pipeline for GC-TOF-MS metabonomics studies. Journal of proteome research 2010, 9 (11), 5974-81. !

ADAP-GC 1.0 Deconvolution

ex = a1,a2,,an{ }ey = b1,b2,,bn{ }

Let the abundance values of two EICs be

Then, the similarity between the two EICs can be measured by

r =ex •ey

ex •ey

15

Why ADAP-GC 2.0?

ADAP-GC 2.0: Deconvolution of Coeluting Metabolites from GC/TOF-MS Data for Metabolomics Studies. Analytical chemistry 2012, 84 (15), 6619-29.

•  43, 73, and 117: shared •  217: unique to uridine •  132: unique to n-eicosanoic acid

810 881

909 948

16

ADAP-GC 2.0

17

ADAP-GC 2.0

•  An EIC peak could result from the elution of a single or multiple co-eluting components.

•  Chromatographic Peak Features (CPF) is defined.

•  Simple CPF and composite CPF are identified.

•  Deconvolution is performed.

determination of deconvolution windows

selection of model CPFs

construction of spectrum for each component

correction of splitting issues

decon procedure

18

Selection of Model CPFs

•  Step 1: select good candidates

•  Step 2: determine the number of components by hierarchical clustering of the good candidates

•  Step 3: determine the model CPF for each component

sharpness = Ii − Ii−1

Ii−1i=2

p

∑ +Ii − Ii+1

Ii+1i=p

n−1

total score = c1( ) mass( )+ c2( ) gaussian similarity( )+ c3( ) apex intensity( )+ c4( ) SNR( )

19

Construction of Spectrum

•  Each composite CPF is a linear summation of model CPFs.

•  Weights are determined by constrained optimization.

•  The weights that correspond to the same model CPF yield the spectrum of a component.

E = X i[ ]− akMk i[ ]k=1

K

∑#

$%

&

'(

i=1

n

∑2

20

ADAP-GC 2.0 Deconvolution

Background ADAP-GC ADAP-LC ADAP-Stats ADAP-CAGT

•  Putting the pieces together …

ADAP-GC 2.0: Deconvolution of Coeluting Metabolites from GC/TOF-MS Data for Metabolomics Studies. Analytical chemistry 2012, 84 (15), 6619-29.

21

•  compute pairwise spectrum similarity

•  keep the best model CPF

•  do a second deconvolution

Resolving Splitting Issues

22

Degree of Co-elution

23

Alignment

•  Component-based: the same component across samples are identified based on spectrum and retention time similarity

scoretotal si, sj( ) = 0.9scorespec si, sj( )+ 0.1scoreRT si, sj( )scoreRT =1− ΔRT w

An automated data analysis pipeline for GC-TOF-MS metabonomics studies. Journal of proteome research 2010, 9 (11), 5974-81.

24

Alignment

•  For each component, the best representative spectrum across all of the samples are determined

RT: 21.2955 21.2938 21.2938 21.2947 21.3097

21.2955 21.2972 21.2963 21.3038 21.3030

25

Alignment

!

26

Export

•  Identity and quantity in .csv files

•  Spectra in .msp format that can be read by NIST MS Search software and other library search tools

27

Software

28

Software

29

Software Background ADAP-GC ADAP-LC ADAP-Stats ADAP-CAGT

30

Software Background ADAP-GC ADAP-LC ADAP-Stats ADAP-CAGT

31

Software

32

Software

33

Next step

•  Many parameters must be pre-specified in the current data processing.

•  How to reduce reliance on parameter settings?

•  Is an adaptive workflow possible?

34

Acknowledgement

•  Du lab §  Wenxin Jiang

§  Yan Ni

§  Peter Pham

§  Kyle Suttlemyre

§  Fei Xu

§  Wenchao Zhang

35

Acknowledgement

•  Dr. Wei Jia’s group @ UNC-Greensboro §  Yunping Qiu

§  Guoxiang Xie

§  Xiaojiao Zheng

•  Dr. Steve Zeisel’s group @ UNC-Chapel Hill

•  Mingming Su @ DHMRI

36

Thank you!

37