Modern business drivers are continually pushing to reduce the time it takes to get a product or service to market, reduce the risk and cost associated with that, and to improve quality In laboratories, delivering an analytical result that’s ‘right first time’ (RFT) is the answer - No reprocessing data or re-running injections - No out of specification (OOS) results or reporting/calculation errors Using chromatography tools for RFT analysis automatically gives: - High quality of results and confidence in results - Lower cost of analysis - Improved lab efficiency - Faster release to market and return on investment (ROI)
Citation preview
1
The world leader in serving science
Darren Barrington-Light
Product Specialist, Chromatography Software,
Thermo Fisher Scientific
May 21, 2014
Getting It “Right First Time” Seminar Series – Part 3
PP71148_E 04/14S
2
Agenda
• Introduction• How to get it right the first time• Thermo Scientific™ Dionex™ Chromeleon™
Chromatography Data System (CDS) tools for RFT analysis• Sequence Ready Check• Smart Startup• Summary
3
Introduction
• Modern business drivers are continually pushing to reduce the time it takes to get a product or service to market, reduce the risk and cost associated with that, and to improve quality
• In laboratories, delivering an analytical result that’s ‘right first time’ (RFT) is the answer• No reprocessing data or re-running injections
• No out of specification (OOS) results or reporting/calculation errors
• Getting RFT analysis automatically gives: High quality of results and confidence in results Lower cost of analysis Improved lab efficiency Faster release to market and return on investment (ROI)
4
How to Get It Right the First Time?
5
How to Get It Right the First Time?
• Robust and reliable instrumentation• Qualification automation
• Monitor instrument status
•Robust, reproducible, and reliable methodology• Standard operating procedures
• Method validation
• Automated sequence creation
• Run-time checks
• In-run reactions to actual chromatographic results
• Automated result calculation and export
• Operators• Make learning curve as shallow as possible
• Provide tools to help inexperienced users with more complex tasks
6
Chromeleon CDS Tools for RFT Analysis
7
Chromeleon CDS Tools for RFT Analysis
• Instrumentation• Predictive performance – instrument qualification, service and wear part
monitors
• Methodology• ICH Method Validation Templates
• eWorkflows™
• Sequence Ready Check
• Smart Startup
• System Suitability Tests with Intelligent Run Control
• Spreadsheet-based Report Designer with integral Report Check
• Operators• All the above combine to ensure ANY operator can achieve RFT
analysis
8
Sequence Ready Check
9
Sequence Ready Check
• How do you know that the sequence you have just created will run without any problems? What if you add injections during the run or modify the Instrument Method? If there are issues, how do you identify them?
• Sequence Ready Check verifies that:• Sequence list contains valid vial positions and methods
• Instrument is ready for operation
• Instrument Methods used to run each injection of every sequence in the Queue are error free
• Any Injection Custom Variables used in Instrument Method are valid
• Instrument Method Check verifies that all Instrument Method entries are valid
10
Sequence Ready Check
• Runs automatically• When Queue is started
• Before each injection
• If sequence is added to running Queue
• Run manually at any time
• Results only shown if issues found• Abort – must be corrected
• Error – must be corrected
• Warning – should be corrected
• Information – info only• e.g. eluent volume required
11
Instrument Method Check
• Automatic checks inside Instrument Method Editor ensure only valid methods can be saved
• Incorrect entries are clearly highlighted and described
• Method tabs cannot be exited with incorrect entries
12
Right First Time Analysis
• Sequence Ready Check• Identifies problems with sequence before run starts and during run
• Works for all sequences and instruments
• Instrument Method Editor Check• Identifies and highlights issues with Instrument Method
• Works for all methods and instruments
• Ensure only valid sequences and methods can be run or added to Queue
• More right first time analyses
13
Smart Startup
14
Smart Startup
• How do you know your instrument is ready to perform the analysis? Which settings do you need? How long should you wait for it to equilibrate? Ever left your system equilibrating for ages?• Usually left to analyst to decide - subjective. Often equilibrate too long or
add extra ‘sacrificial’ injections wasting solvent, sample, vials and time.
• Smart Startup initializes and equilibrates your instrument with correct chromatographic conditions before first injection.
Start your sequence with ‘Start with Smart Startup’
15
Smart Startup
• Available for Thermo Scientific™ Dionex™ UltiMate™ 3000 HPLC and modular Thermo Scientific™ Dionex™ ICS systems
• Dedicated ePanel provided for Smart Startup• Set Smart Startup per sequence in Queue tab
• Can simplify automation of method switching
• Can set delayed start
16
Smart Startup
• First runs instrument wake up routine• Ensures modules are active and connected
• Next prepares instrument• Purges pump, runs gradient wash, flushes system
• Washes autosampler
• Switches column oven valves
• Heats or cools temperature-controlled modules
• Switches lamps on
• Finally sets initial conditions for all modules and equilibrates
• After 10 min warm up, key parameters are monitored• Pump pressure ripple
• Sampler and oven temperature
• Detector noise and drift
17
Smart Startup
• Parameter status is monitored and updated in real time
18
Smart Startup
19
Smart Startup
20
Smart Startup
21
Smart Startup
22
Smart Startup
23
Smart Startup
24
Smart Startup
25
Smart Startup
26
Smart Startup
• Once everything is within specification sequence starts
27
Smart Startup – Customization
• Smart Startup conditions are automatically extracted from Instrument Method
• Easily accessible in Instrument Method Editor via dedicated tab
28
Smart Startup – Customization
• Can select whether to startup specific modules and perform certain procedures• Purge pump
• Flush system
• Set temperatures
• Lamps/cells on
• Electrolytic devices on
• Can access advanced settings for finer control
29
Smart Startup – Customization
• Can edit all settings including acceptance criteria• Pump pressure ripple limit
• Noise & drift limits
• Temperature limits
• Maximum equilibration time limits time for system to reach equilibration before stopping Smart Startup
30
Smart Standby and Shutdown
• Smart Shutdown• Safely shuts system down
• Can flush system to remove buffers
• Switches off • Flow
• Detector lamps/cells
• Electrolytic devices
• Ovens
• Smart Standby• Automatically puts system into
standby mode
• Low flow, lamps off, lower temps
• Set in the Queue
31
Live Demo
• Let’s have a look at the Ready Check and Smart Startup in action …
32
Summary
33
Summary
• Sequence Ready Check and Instrument Method Check• Checks that every sequence will run correctly and discovers any issues
before and during the run
• Identifies and highlights issues with Instrument Method
• Works for all sequences, methods and instruments
• Ensures only valid sequences and methods will be run giving more right first time analyses
• Smart Startup:• Automates your instrument initialization and equilibration
• Ensures first injection is always good
• Removes subjectivity from system readiness and frees up analyst time
• Eliminates wasted solvent, sample and run time
• Aids method switching increasing instrument utilization
34
Next time…
• Discover even more Chromeleon CDS tools to ensure you get “right first time, every time” analysis