PA Intermodulation Distortion Measurements Using the PNA

Preview:

Citation preview

PA Intermodulation Distortion Measurements Using the PNA

María Jesús Cañavate Sánchez

Index

Galashiels

Orkney

1. Introduction: characterization of a power amplifier

2. PNA N5225A

3. S-Parameters

4. AM/AM

5. AM/PM

6. Two-tone test

1. Introduction: characterization of a power amplifier

Galashiels

Orkney

S-Parameters

DUT: Amplifier

1. Introduction: characterization of a power amplifier

Galashiels

Orkney

Nonlinear characteristic

AM/AM and AM/PM compression

of a typical Traveling Wave Tube

(TWT) amplifier.

1. Introduction: characterization of a power amplifier

Galashiels

Orkney

Intermodulation distortion

DUT: Amplifier

Frequency spectrum of the amplifier

input signal: 5 unmodulated carriers.

AM/AM and AM/PM compression

of a typical Traveling Wave Tube

(TWT) amplifier.

Frequency spectrum of the amplifier

output signal: 5 unmodulated

carriers and IMP3.

2. PNA N5225A

Galashiels

Orkney

- Number of ports: 4

- Frequency range: 10MHz – 50GHz

- Max. port power, 20GHz: +13dBm

- Receiver damage level: +15dBm

2. PNA N5225A

Galashiels

Orkney

Damage: +15dBm

Damage: +30dBm

Damage: +43dBm

Coupling factor: 15dB

3. S-Parameters

Galashiels

Orkney

Power flow analysis (for all types of PA measurements)

3. S-Parameters

Galashiels

Orkney

Before calibration: make sure settings are at desired levels

3. S-Parameters

Galashiels

Orkney

Select power and internal attenuators levels

- Output power: -10 dBm

- Gain: 26 dB

- External attenuation: 6 dB

- Internal attenuation: 15 dB

- Measured S21 drops by the attenuation

- After calibration, attenuator losses will

be removed

- Before calibration: Gain = 5 dB

- After calibration: Gain = 26 dB

- Internal attenuation still not available

3. S-Parameters

Galashiels

Orkney

Calibration performance (for all types of PA measurements)

Source–power calibration at the input point of the amplifier

To ensure a constant and known power

Material needed: power sensor

Two-port SOLT calibration

To remove effects of the external and internal attenuators

Material needed: mechanical calibration kit

SOLT: short-open-load-through

3. S-Parameters

Galashiels

Orkney

Calibration performance

Source–power calibration Two-port calibration

3. S-Parameters

Galashiels

Orkney

Measurements

4. AM/AM

Galashiels

Orkney

S21 measurement under a power sweep condition

- Attenuator settings are fixed

- Source-power calibration at

source and receiver ports

4. AM/AM

Galashiels

Orkney

S21 measurement example

Gain

AM/AM

5. AM/PM

Galashiels

Orkney

Procedure:

S21 measurement under a power sweep condition

(AM/AM).

Change the S21 measurement from log magnitude format

to phase format.

No new calibration is required.

6. Two-tone test

Galashiels

Orkney

2 different configurations (option 087):

Using internal PNA components

Only extra cables are required

Using external components

Combiner

Coupler

Extra cables

6. Two-tone test

Galashiels

Orkney

Using internal PNA components:

6. Two-tone test

Galashiels

Orkney

Using external components:

6. Two-tone test

Galashiels

Orkney

6. Two-tone test

Galashiels

Orkney

Frequency tab:

6. Two-tone test

Galashiels

Orkney

Power tab:

Thank you!

Galashiels

Orkney

PhD student: María Jesús Cañavate Sánchez // mjc31@hw.ac.uk