WP214 Imaging Spectral Colorimeter Viewing Angle Measurement Report
Backlit LCD Display
Issued: April 23, 2014
Westboro Photonics Page 2
Table of Contents
1.0 WP214 IMAGING SPECTRAL COLORIMETER ........................................................................................................... 3
2.0 UNIQUE CAPABILITIES .................................................................................................................................................. 3
3.0 MEASUREMENT METHOD ............................................................................................................................................. 4
4.0 RESULTS .......................................................................................................................................................................... 5
4.1 DISPLAY WHITE .................................................................................................................................................................. 5 4.1.1 Screenshot .................................................................................................................................................................... 5 4.1.2 Tabular data ................................................................................................................................................................. 5 4.1.3 Distribution of axial spots to 50 degrees in CIE 1976 space ................................................................................. 6
4.2 DISPLAY RED ...................................................................................................................................................................... 7 4.2.1 Screenshot .................................................................................................................................................................... 7 4.2.2 Tabular data ................................................................................................................................................................. 7 4.2.3 Distribution of axial spots to 50 degrees in CIE 1976 space ................................................................................. 8
4.3 DISPLAY GREEN ................................................................................................................................................................. 9 4.3.1 Screenshot .................................................................................................................................................................... 9 4.3.2 Tabular Data ................................................................................................................................................................. 9 4.3.3 Distribution of axial spots to 50 degrees in CIE 1976 space ............................................................................... 10
4.4 DISPLAY BLUE .................................................................................................................................................................. 11 4.4.1 Screenshot .................................................................................................................................................................. 11 4.4.2 Tabular Data ............................................................................................................................................................... 11 4.4.3 Distribution of axial spots to 50 degrees in CIE 1976 space ............................................................................... 12
4.5 SPECTRAL DATA FOR RED, GREEN, BLUE, WHITE AND BLACK DISPLAY ........................................................................ 13 4.6 CIE 1976 WITH ALL MEASUREMENTS TO 50 DEGREES .................................................................................................... 16 4.7 CONTRAST ........................................................................................................................................................................ 17
Westboro Photonics Page 3
1.0 WP214 Imaging Spectral Colorimeter
The WP214 was developed to offer increased productivity to manufacturers and OEM customers who measure flat-panel displays, LED clusters, and multifunction information displays for luminance and chromaticity.
The system features a high dynamic range, cooled CCD imager with 1MP resolution and an integrated spectroradiometer which serves as the reference to correct the imager results. An automated filter wheel contains four tristimulus filters and five optional positions for scotopic or radiometric filters.
The function of the spectroradiometer is to provide high accuracy tristimulus measurements at one measurement spot, and by comparing the imaging colorimeter tristimulus data in that same location, a vector correction can be determined for the imaging colorimeter tristimulus values. This correction is applied to the imaging colorimeter data for display uniformity and viewing angle measurements.
2.0 Unique Capabilities
Choosing a measurement solution for display characterization is an important decision. Westboro Photonics is a leader in display measurement, with more than 20 years’ experience working with LED and display metrology. When evaluating technical alternatives, we invite you to consider the following important differentiators of the WP214 system:
Integrated spectroradiometer ensures superior color accuracy
Substantial time savings with simultaneous recording of an entire image vs. spot measurement systems (e.g. a complete control panel or display)
Integrated image processing functions permit new methods of analysis (e.g. automated target and background identification)
Short measurement cycles and simple setup of intuitive pass/fail criteria for use in production environments.
Compact, portable design and minimum training requirement facilitate a wide range of applications.
Built-in scripting capability to enable rapid capturing, analysis and reporting sequences.
Software development kit available for custom applications and automation
Wide assortment of Interchangeable lens options o Viewing Angle to +/- 80° when equipped with CONOMETER® lens
Westboro Photonics Page 4
3.0 Measurement Method
Sample Details:
Sample No. Description Date Analyzed
1 Mobile Display
October 21, 2013
Hardware configuration and setup:
A WP214 (S/N 2140113002, pre-production system) with CONOMETER® lens was mounted to a stage on a manual rail system.
The sample was placed on a level stage and the measurement area was aligned using the image preview feature of the Photometrica software. The precise position of the AOI’s on the surface of the sample was not tracked or tightly controlled.
All measurements were performed in a dark room.
Westboro Photonics Page 5
4.0 Results
4.1 Display White
4.1.1 Screenshot
Figure 1: Pseudo color luminance plot with display set to white (log scale)
4.1.2 Tabular data
Name Luminance cd/m²
u' v' X y CCT (K)
CENTER 234.3 0.1957 0.4667 0.3086 0.327 6750
10 E 213.9 0.1974 0.4688 0.3125 0.3299 6507
10 N 211.1 0.1975 0.4707 0.3144 0.3329 6386
10 S 192.8 0.1948 0.4671 0.3078 0.3281 6781
10 W 203.1 0.1955 0.4674 0.309 0.3282 6717
20 E 158.1 0.2001 0.4658 0.3133 0.3241 6506
20 N 155.4 0.1949 0.4748 0.3147 0.3407 6322
20 S 129.7 0.1964 0.4643 0.3074 0.323 6858
20 W 140 0.1933 0.468 0.3067 0.3301 6822
Westboro Photonics Page 6
30 E 103.7 0.1969 0.4699 0.313 0.3319 6467
30 N 94.53 0.1992 0.4739 0.3194 0.3377 6108
30 S 78.01 0.1981 0.4608 0.3066 0.317 6977
30 W 83.84 0.1968 0.4603 0.3045 0.3166 7118
40 E 55.07 0.1969 0.4678 0.3111 0.3286 6593
40 N 59.56 0.1942 0.4758 0.3147 0.3427 6312
40 S 50.61 0.1941 0.4619 0.3025 0.3199 7194
40 W 40.78 0.1978 0.4558 0.3021 0.3094 7395
50 E 29.86 0.1972 0.4732 0.3163 0.3373 6264
50 N 40.66 0.197 0.4723 0.3152 0.3358 6327
50 S 37.91 0.1906 0.4719 0.3067 0.3374 6756
50 W 24.08 0.194 0.459 0.3001 0.3155 7424
4.1.3 Distribution of axial spots to 50 degrees in CIE 1976 space
Figure 2: All AOI’s plotted in the CIE1976 color space with the display set to white
Westboro Photonics Page 7
4.2 Display Red
4.2.1 Screenshot
Figure 2: Pseudo color luminance plot with display set to red (log scale)
4.2.2 Tabular data
Name Luminance (cd/m²)
u' v' x Y Dominant Wavelength (nm)
Purity
CENTER 43.8 0.433 0.5192 0.6195 0.3302 611 0.85
10 E 41.05 0.4314 0.5193 0.6183 0.3308 611 0.85
10 N 40.08 0.4374 0.5198 0.624 0.3296 612 0.86
10 S 35.85 0.4358 0.5186 0.6208 0.3283 612 0.85
10 W 37.89 0.4354 0.5203 0.6233 0.331 611 0.86
20 E 31.73 0.4252 0.5198 0.6139 0.3336 610 0.84
20 N 29.72 0.4275 0.5216 0.6187 0.3355 609 0.86
20 S 25.49 0.4303 0.5189 0.6168 0.3306 611 0.84
20 W 25.36 0.4346 0.5213 0.6242 0.3328 610 0.87
30 E 20.52 0.425 0.5202 0.6143 0.3341 610 0.85
30 N 18.02 0.431 0.5208 0.6203 0.3332 610 0.86
30 S 15.45 0.4322 0.5172 0.6157 0.3274 613 0.83
Westboro Photonics Page 8
30 W 15.37 0.4374 0.5211 0.6261 0.3315 611 0.87
40 E 10.74 0.4262 0.5191 0.6136 0.3321 611 0.84
40 N 11.45 0.4322 0.5202 0.6204 0.3319 611 0.86
40 S 9.782 0.4391 0.5146 0.6174 0.3216 615 0.82
40 W 7.8 0.4443 0.5205 0.631 0.3286 612 0.88
50 E 6.046 0.4335 0.5186 0.619 0.3292 612 0.85
50 N 8.249 0.4311 0.5193 0.618 0.3308 611 0.85
50 S 7.241 0.4317 0.5155 0.6126 0.3252 614 0.81
50 W 4.286 0.4473 0.5189 0.6308 0.3252 613 0.87
4.2.3 Distribution of axial spots to 50 degrees in CIE 1976 space
Figure 4: All AOI’s plotted in the CIE1976 color space with the display set to red
Westboro Photonics Page 9
4.3 Display Green
4.3.1 Screenshot
Figure 5: Pseudo color luminance plot with display set to green (log scale)
4.3.2 Tabular Data
Name Luminance (cd/m²) u' v' x Y
Dominant Wavelength (nm) Purity
CENTER 169 0.1249 0.5638 0.3015 0.6047 548 0.73
10 E 153.4 0.1259 0.5642 0.304 0.6052 549 0.74
10 N 153.7 0.1275 0.564 0.3067 0.6029 549 0.74
10 S 138.5 0.1258 0.5634 0.3027 0.6027 549 0.73
10 W 147 0.1253 0.5636 0.3021 0.6037 549 0.73
20 E 113.3 0.1281 0.5637 0.3074 0.6014 550 0.74
20 N 112.5 0.1252 0.5649 0.3035 0.6086 549 0.75
20 S 95.08 0.1249 0.5634 0.301 0.6034 548 0.73
20 W 101.7 0.1247 0.5639 0.3012 0.6056 548 0.74
30 E 74.8 0.1246 0.5646 0.3019 0.6081 549 0.75
30 N 69.53 0.1247 0.5652 0.3029 0.6104 549 0.76
30 S 56.82 0.1244 0.5634 0.3 0.6037 548 0.73
Westboro Photonics Page 10
30 W 61.66 0.126 0.564 0.3039 0.6047 549 0.74
40 E 40.08 0.1258 0.5658 0.3058 0.6112 549 0.77
40 N 44.26 0.1236 0.5659 0.3018 0.614 549 0.76
40 S 37.61 0.1215 0.5644 0.2957 0.6103 547 0.74
40 W 30.2 0.1218 0.564 0.2957 0.6086 547 0.73
50 E 21.15 0.1233 0.5663 0.3015 0.6156 549 0.77
50 N 31.13 0.1206 0.567 0.2973 0.6212 548 0.77
50 S 29.09 0.1191 0.5667 0.2939 0.6215 547 0.76
50 W 17.63 0.1217 0.565 0.2968 0.6123 548 0.74
4.3.3 Distribution of axial spots to 50 degrees in CIE 1976 space
Figure 6: All AOI’s plotted in the CIE1976 color space with the display set to green
Westboro Photonics Page 11
4.4 Display Blue
4.4.1 Screenshot
Figure 7: Pseudo color luminance plot with display set to blue (log scale)
4.4.2 Tabular Data
Name Luminance
cd/m² u' v' x Y
Dominant Wavelength (nm)
Purity
CENTER 18.84 0.1729 0.1623 0.1491 0.0622 465.4 0.9276
10 E 16.85 0.1744 0.1628 0.1504 0.06236 465.1 0.9241
10 N 16.74 0.1733 0.1646 0.1499 0.06326 465.4 0.9237
10 S 15.56 0.1735 0.1622 0.1495 0.06212 465.3 0.9267
10 W 16.43 0.1722 0.1625 0.1486 0.06232 465.6 0.9286
20 E 12.59 0.1743 0.1581 0.1492 0.06015 465 0.9309
20 N 11.81 0.1688 0.1625 0.1459 0.0624 466.3 0.9352
20 S 10.56 0.1728 0.1572 0.1478 0.05975 465.3 0.935
20 W 11.23 0.1701 0.1599 0.1464 0.06115 466 0.9361
30 E 8.303 0.1719 0.1638 0.1486 0.06292 465.7 0.9274
30 N 7.123 0.1702 0.1641 0.1473 0.06314 466.1 0.9302
Westboro Photonics Page 12
30 S 6.265 0.1745 0.1513 0.1478 0.05697 464.8 0.9399
30 W 6.792 0.1737 0.1509 0.1471 0.05681 465 0.9419
40 E 4.278 0.1773 0.1595 0.1518 0.0607 464.3 0.9237
40 N 4.477 0.174 0.1651 0.1506 0.0635 465.2 0.9216
40 S 4.046 0.1746 0.1528 0.1482 0.05764 464.8 0.9378
40 W 3.418 0.176 0.1496 0.1485 0.0561 464.4 0.9397
50 E 2.404 0.1763 0.1705 0.1536 0.06601 464.9 0.9098
50 N 2.874 0.1753 0.1495 0.148 0.05608 464.5 0.9411
50 S 2.752 0.1778 0.1521 0.1505 0.0572 464 0.9331
50 W 1.921 0.1832 0.1453 0.153 0.05395 462.4 0.9333
4.4.3 Distribution of axial spots to 50 degrees in CIE 1976 space
Figure 8: All AOI’s plotted in the CIE1976 color space with the display set to blue
Westboro Photonics Page 13
4.5 Display Black
4.5.1 Screenshot
Figure 7: Pseudo color luminance plot with display set to blue (log scale)
4.5.2 Tabular Data
Name Luminance
(cd/m²) u' v' x y
CENTER 0.1591 0.184 0.4362 0.2703 0.2849
10 E 0.1899 0.1884 0.449 0.2852 0.3021
10 N 0.1522 0.1838 0.4384 0.2717 0.2881
10 S 0.1418 0.1871 0.4359 0.2738 0.2835
10 W 0.1469 0.1843 0.4417 0.2747 0.2926
20 E 0.2119 0.192 0.4624 0.3004 0.3215
20 N 0.1323 0.1855 0.4392 0.2743 0.2887
20 S 0.1189 0.1932 0.4336 0.2795 0.2788
20 W 0.1512 0.1876 0.4604 0.2932 0.3198
30 E 0.2082 0.1919 0.4777 0.3135 0.3468
Westboro Photonics Page 14
30 N 0.1077 0.1893 0.4395 0.2791 0.2881
30 S 0.09967 0.2014 0.4332 0.2887 0.276
30 W 0.1418 0.1997 0.4671 0.314 0.3264
40 E 0.1716 0.197 0.4814 0.3235 0.3513
40 N 0.09004 0.1944 0.4401 0.2856 0.2874
40 S 0.08958 0.2042 0.4368 0.2947 0.2802
40 W 0.1083 0.2118 0.4626 0.3248 0.3153
50 E 0.1362 0.2079 0.4841 0.3401 0.352
50 N 0.07753 0.2071 0.4326 0.2948 0.2737
50 S 0.0875 0.205 0.4473 0.3038 0.2946
50 W 0.08667 0.2171 0.4573 0.3264 0.3056
Westboro Photonics Page 15
4.6 Spectral data for Red, Green, Blue, White and Black Display
Figure 9: Spectral distribution plots for all colors (log scale)
Westboro Photonics Page 16
4.7 CIE 1976 with all measurements to 50 degrees
Figure 10: CIE 1976 chromaticity plot for all colors
Westboro Photonics Page 17
4.8 Contrast
Figure 11: Contrast computation (White / Black)
Name CENTER 10 E 10 N 10 S 10 W 20 E 20 N
Contrast 1473 1127 1386 1360 1383 747.2 1175
Name 20 S 20 W 30 E 30 N 30 S 30 W 40 E
Contrast 1090 925.9 498.1 877.2 782.4 590.6 320.5
Name 40 N 40 S 40 W 50 E 50 N 50 S 50 W
Contrast 661 565.2 375.9 219.5 524.7 434.3 277.7