MEDICAL CYLINDER VALVES
13485QUALITYASSURANCE
ISO0029
TABLE OF CONTENTS
3
INTRODUCTION P.004
- Ceodeux meditec solutions P.004
- Quality first... P.004
- The main medical gases P.005
- Product range overview P.006
- Technology overview P.008
- How to select valves P.012
VALVES INTEGRATED WITH PRESSURE REGULATOR P.014
- M707 – Lifeball P.014
- M706/M713 | LIFE+ P.016
- M741 | LIFE-s P.018
- M800 – Emergency Back Pack P.020
- Protective Guard P.021
CYLINDER VALVE WITH RESIDUAL PRESSURE VALVE P.020
- M510/M515/M501/M506/M520/M525 Standard RPV Cylinder Valve P.022
- M540/M545/M530/M535 | Standard RPV off line P.023
- M511/M516/M508/M521/M526 | Standard RPV in line P.024
- M978/M979/C878 | Standard RPV in line P.025
- M100/M105/M040/M045/M200/M205 | Standard cylinder valve P.026
STANDARD CYLINDER VALVE P.027
- M972/M973/M974/M975 | standard cylinder valve small cone P.027
- D113 – Packed Valve P.028
- D413 – Tied Diaphragm P.029
SPECIAL NITRIC OXIDE CYLINDER VALVE P.030
- M966 – Pin index Cylinder Valve P.030
- M650 – Pin index RPV Cylinder Valve P.031
All Rotarex regulators are produced in Europe in accordance with international standards (ISO; CGA....) and are guaran-teed to provide safe and reliable performance in operation. All locations are ISO 13485.
CEODEUX MEDITEC SOLUTIONS
QUALITY FIRST...
4 INTRODUCTION
ALL RIGHTS OF CHANGE RESERVED
Patient therapy and safety is our main focus in the design, installation, commissioning and operation of medical gas pipeline systems
QUANTITY OF SUPPLYValves are designed in such a way sufficient medical gas capacity can be ensured in order to provide the required flows of gases for the intended number of patients to be treated at any time.
IDENTITY OF SUPPLYAll our products are designed, produced and controlled for a specific medical gas. All points to which the user or technical staff can connect are gas specified. There is no risk of misconnection.
QUALITY OF SUPPLYOur valves maintain quality of supplied medical gases thanks to appropri-ate selection of material, components cleanliness and internal filtration equipment.Products designs ensure safe supply by minimizing ignition risks, by avoid-ing patient poisoning risks and by guaranteeing very good performances for accurate therapies.
CEODEUX MEDITEC products are designed, validated, produced and con-trolled during their whole lifetime according to a very stringent quality system which exceeds requirements of ISO 13485 standard.
DESIGNAll our products comply with local and international standards. Some reference standards:
ISO 10524-3 / CGA E-18 VIPR
ISO 10297 / CGA V1 Standard Cylinder Valves
ISO 15996 RPV Valves
ISO 11117 Protection Guard
Product design is done through a risk assessment process and is combining proven technologies and innovation to reach the best performance levels.
VALIDATIONProduct validation process guarantees that products are developed and produced to withstand the complete lifecycle (installation, commission-ing, normal service and maintenance)
PRODUCTIONAll products are produced in a controlled environment in order to maintain cleanlinessAn individual traceability is ensured with individual serial numberProducts are individually controlled and data are recorded for a perfect product management.All products are cleaned according to ISO 15001
THE MAIN MEDICAL GASES
5INTRODUCTION
ALL RIGHTS OF CHANGE RESERVED
Symbol Name Application Source Type
O2 OxygenOxygen is one of the most extensively used gases for respiratory therapy and life-support and is additionally used in anesthetic procedures.
- Compressed gas cylinder - Liquid tank - Concentrator
Air AirMedical air is mainly used in respiratory therapy as a power source for patient ventilators, and for blending with oxygen. It is also used as the driving gas for nebulized drugs and chemotherapy agents.
- Compressor - Compressed gas cylinder
N2O Nitrous OxideNitrous oxide is used for anesthetic and analgesic purposes, being mixed with air, oxygen, and nebulized agents.
- Liquefied gas cylinder
CO2 Carbon DioxideCarbon dioxide is used less commonly now as a respiratory stimulant, and for insufflation during surgery.
- Liquefied gas cylinder
Air Surgical AirSurgical air (of medical air quality) is also used, at a higher pressure, to power a variety of surgical tools and other devices such as tourniquets.
- Compressor - Compressed gas cylinder
N2 NitrogenAs an alternative, nitrogen can be used for the same purpose as surgical air.
- Compressed gas cylinder
ALL RIGHTS OF CHANGE RESERVED
STANDARD CYLINDER VALVE WITH RESIDUAL PRESSURE VALVE OFF LINE
M510/M515 P. 22 M501/M506 P. 22 M520/M525 P. 22 M540/M545 P.23 M530/M535 P. 23
Gas type O2 CO2 Various gases O2 Various gases
Material Chrome plated brass Chrome plated brass Chrome plated brass Chrome plated brass Chrome plated brass
Inlet pressure 200 bar/2900 psi 200 bar/2900 psi 200 bar/2900 psi 300 bar/4350 psi 300 bar/4350 psi
STANDARD CYLINDER VALVE WITH RESIDUAL PRESSURE VALVE IN LINE
M511/M516 P. 24 M508 P. 24 M521/M526 P. 24 M978/M979 P. 25
SMALL CONEC878 P. 25
INTEGRATED GAUGEGas type O2 CO2 Various gases O2, various gases o
Material Chrome plated brass Chrome plated brass Chrome plated brass Chrome plated brass Chrome plated brass
Inlet pressure 200 bar/2900 psi 200 bar/2900 psi 200 bar/2900 psi 200 bar/2900 psi 200 bar/2900 psi
6
VIPR
LIFEBALL P. 14 LIFE + P. 16 LIFE-s P. 18
Gas type O2 O2 O2
Material Chrome plated brass Chrome plated brass Chrome plated brass
Inlet pressure 200 bar/300 bar
2000 psi/3000 psi
M706: 200 bar/2900 psi
M709: 300 bar/4350 psi
200 bar/2900 psi
M800 EMERGENCY OXYGEN BACK PACK P. 20 M706/M713 PROTECTIVE GUARD P. 21
Gas type O2 Ø
Material Chrome plated brass Plastic PA6.6
Inlet pressure 200 bar/2900 psi Ø
PRODUCT RANGE OVERVIEW
INTRODUCTION
ALL RIGHTS OF CHANGE RESERVED
PIN INDEX VALVE
M966 PIN INDEX CYLINDER VALVE P. 30 M650 P. 31
PIN INDEX RPV CYLINDER VALVE
Gas type O2 O2
Material Chrome plated brass Chrome plated or natural brass
Inlet pressure 232 bar/3364 psi 232 bar/3364 psi
SPECIAL NITRIC OXIDE MIXTURES CYLINDER VALVE
D113 PACKED VALVE P. 28 D413 TIED DIAPHRAGM P. 29
Gas type No No
Material Stainless steel Stainless steel
Inlet pressure 200 bar/2900 psi 200 bar/2900 psi
7
PRODUCT RANGE OVERVIEW
INTRODUCTION
STANDARD CYLINDER VALVE
M100/M105 P. 26 C040/C045 P. 26 M200/M205 P. 26 M972/M973 P. 27
SMALL CONEM974/M975 P. 27
SMALL CONEGas type O2 CO2 Various gases O2 Various gases
Material Chrome plated brass Chrome plated brass Chrome plated brass Chrome plated brass Chrome plated brass
Inlet pressure 200 bar / 2900 psi 200 bar / 2900 psi 200 bar / 2900 psi 200 bar/2900 psi 200 bar/2900 psi
ALL RIGHTS OF CHANGE RESERVED
TECHNOLOGY OVERVIEW
INTRODUCTION8
Ceodeux Meditec uses 4 main technologies to achieve a proper and reliable gas use from the cylinder:
VIPR - Valve with integrated pressure regulator Design to save your time: Easy and Ready to use medical valve with in-built regulator, flow meter, contains gauge and outlet pressure connector.
- Compact and lightweight : ideal for small cylinder - Fully protected by an ergonomic plastic guard - Filtration system ensures safety against ignition. - Flow meter with orifice plate precision design increase high flow precision and low flow precision for pediatric patients
RESIDUAL PRESSURE VALVE - Open/close fonctionnalities - Improved safety thanks to RPV technology included (avoid all contamination risks)
- Safe, long, trouble free life under all service conditions - Exist in offset and inline version
STANDARD CYLINDER VALVE - Standard open/close fonctionnalities - High life cycle thanks to smart choice of materials - Safe, long, trouble free life under all service conditions
PIN INDEX VALVE - Standard open/close fonctionnalities - Yoke outlet connection (PIN INDEX System) - Key operated or handwheel manipulation - High life cycle thanks to smart choice of materials - Safe, long, trouble free life under all service conditions
CEODEUX MEDITEC VIPR
M707 P. 014
M706 P. 016
M713 P. 016
M741 P. 018
M800 P. 020
PRODUCT FINDER
CEODEUX MEDITEC RPV valve
M510/M515/M501/M506/M520/M525/M540/M545M530/M535/ M511/M516/M508/M521/M526 P. 024
M978/M979/C878 P. 025
PRODUCT FINDER
CEODEUX MEDITEC Standard valve
M100/M105/M040/M045/M200/M205/ M972/M973/M974/M975 P. 026
D113 P. 028
D413 P. 029
PRODUCT FINDER
CEODEUX MEDITEC Pin Index valve
M966 P. 030
M650 P. 031
PRODUCT FINDER
ALL RIGHTS OF CHANGE RESERVED
TECHNOLOGY OVERVIEW
9INTRODUCTION
VALVE INLET - Anti dust tube according to the last amendment of EN ISO 10524-3
· Better protection of the valve against contamination and eliminates the risk of ignition in case accumulation of particles in a filter
- Excess flow device · Additional safety to avoid excess flow risks and optimize normal flow in use
FILLING PORT - Approved for 300 bar (tested at 360 bar) - Active and non-active - Spring in Copper Beryllium material highly compatible with oxidizing gases
- O-rings in EPDM - High leak integrity due to seat O-ring and 2-parts design, including degasing hole
- High performance of the filling port against ignition with protruding pin design
- Filling port fully integrated in the valve body - Endurance performance is twice the requirment of the ISO standard (2000 cycles)
PRESSURE GAUGE - NEW Electronic system available (please refer to page 9 for further informations)
- Design fully in accordance with the EN ISO 10524-3 including a flow restrictor to prevent part ejection in case of failure
- Direct drive technology used for the Bourdon tube to reduce consequences of vibration or shocks
- Gauge display tight to water contamination - Safety Relief valve with 3 holes on the glass and tightness evidence sticker - Active pressure gauge for permanent content visibility
OUTLET PRESSURE: MEDICAL CONNECTOR - Same outlet as wall outlet used in hospitals with design according national standard
- Make connection to medical equipment possible to transport patients - Minimum outlet pressure guaranteed when high flow is required by downstream equipment
SHUT-OFF VALVE - 5000 cycles according CGA V9 (under 360 Bar) - None rotating lower spindle to avoid particles generation - Very low torque design to operate the shut-off valve up to 300 bar - ISO 10297
DETAILED FUNCTIONNALITIES OF OUR INTEGRATED VALVES
REGULATOR SAFETY RELIEF DEVICE - Integrated into the regulator design
· Quicker reaction - Large contact surface to increase sensitivity and reactivity to overpressure
- No risk to damage the device - High flow through the device due to the large diameter and large venting holes
FLOWMETER AND HOSEBARB - 12 positions - Continuous flow even in between setting positions during handwheel operation
- No possible positioning of the handwheel in-between settings - High precision of the flows - 10 year life time without maintenance - Easy replacement of the handwheel - High leak integrity performance - Flow range: Pediatric applications 0 -> 2 LpmHomecare 0 -> 6 LpmHospital 0 -> 15/ 0 -> 25 Lpm Possibility to be customed
PRESSURE REGULATOR - Piston-type regulator - Improved safety and reliability - Direct acting valve (decrease pressure curve) according to ISO 10524-3 and CGA-18
- High compatible materials with Oxygen and Medical applications
RESIDUAL PRESSURE VALVE - Protection of the cylinder against external contamination - High endurance performance in terms of leak integrity and functionality
· According standard ISO15996 – 100.000 cycles · Specific seat design to protect sealing O-ring during operation
- 3 bar residual pressure guaranteed at maximum flow setting
TECHNOLOGY OVERVIEW
DETAILED FUNCTIONNALITIES OF OUR INTEGRATED VALVES
10 INTRODUCTION
ALL RIGHTS OF CHANGE RESERVED
GROUNDBREAKING INNOVATION TO PERFECTLY REACH OXYGEN THERAPY REQUIREMENTS:
• Reactivity : Quick and accurate calculation of remaining time • Safety : Perfectly suitable for high pressure pure Oxygen environnement• Patient & Medical staff Services : Improved cylinders fl eet management.
- Quicker operations in emergency situations - Bluetooth for wireless communication
• Optimization : Longer cylinder use thanks to accurate cylinder content management
TECHNOLOGY OVERVIEW
11INTRODUCTION
CEODEUX MEDITEC INTELLIGENT ELECTRONIC SYSTEM CHANGE THE WAY YOU USE GAS
ALL RIGHTS OF CHANGE RESERVED
Get in touch with excellence
NEW CONTROL GAS USE THANKS TO INFORMATION TRANSPARENCY
NEW GAS THERAPY OPTIMIZATION THANKS TO ANTICIPATION
COLOR CODED CYLINDER CONTENT LEVEL INDICATIONS2
1 CYLINDER CONTENT BAR GRAPH (%)
3 PRECISE CYLINDER PRESSURE
5 CALCULATION SYMBOL FLOW
6 AVAILABLE WIRELESS BLUETOOTH COMMUNICATION
4 REMAINING TIME FOR USE AT SELECTED FLOW RATE
2
1
3
45
Cylinder content* bar graph (%). Visual indications of real quantity of gas into cylinder whatever pressure and temperature parameters are.
Red: Warning area : < 10% cylinder content Yellow : Pre-Warning area : 10% < cylinder content < 30% Green : Confortable area : >30% cylinder content
Precise cylinder pressure Cylinder pressure measure-ment at real conditions Bar – Psi - kPa units available.
At changing conditions calculation process before displaying remaining time.
Remaining time for use at selected fl ow. Accurate and quick calculation of the remaining time according to fl ow selection and cylinder content. Precise displayed unit: hh:mm
Indicate when a sensor default was detected into the system
Indicate when an electronic failure was detectead into the system
Indicate when battery level of the system doesn’t guarantee anymore the accurate func-tionning of the valve
Indicate when environment is not conformed with valve functioning (e.g. high or low temperature)
Indicate when gas fl ow is critical (Less of 10 minutes of use or 10 bar remain-ing in the cylinder)
* Quantity of gas into the cylinder
6
Intelligent and complete alarm system to clearly indicate when user face a critical situation. Audible and visible alerts for enhanced patient safety
SYSTEMS FAILURES NORMAL USE ALARMS
ALL RIGHTS OF CHANGE RESERVED
HOW TO SELECT VALVES
BODY MATERIALS
Most Ceodeux Meditec valves are available in brass or chrome plated brass.
Chrome plated or Raw brass: The most commonly used material for high velocity oxygen applications due to its cost effectiveness, good strength, resistence and low-friction flow properties.
Need more information? One of our material engineers would be happy to discuss the pros and cons of each option to help you choose the best solution.
INLET PRESSURE
OUTLET PRESSURE
Different models are designed for different inlet pressure to fit with your applications need (working pressure). For high pressure (200bar/3000psi) the Ceodeux Meditec valves range covers any requirements.
The available options are clearly indicated on each product page.Please specify required inlet pressures when ordering. We can also accommodate special requests.
Especially for VIPR: different models are designed for different outlet pres-sure to fit with your applications need (working pressure). For low pressure needed in use (4,5bar/50psi) and for accurate flow (from 0-2 to 0-25 Lpm) the Ceodeux Meditec VIPR range covers any requirements.
The available options are clearly indicated on each product page.Please specify required inlet pressures when ordering. We can also accommodate special requests.
INLET CONNECTIONS
Specific cylinder valve connections are required for each gas type. The most common connection types are ¾’’-16 UNF and 25E.Other standards and dimensions are available on request.
OUTLET CONNECTIONS
VIPR: Specific outlet connections are available (medical connector types). According to your national standards the Ceodeux Meditec valves range covers any requirements.
Specific flow outlet connections are available (hose barb) according to EN 13544-2 and supplied from flow selector.
Valves:According to your national standards and according to upstream device to be connected (pressure regulator, flex hose,...) the Ceodeux Meditec valves range covers any requirements.
Gas Compatibility: make sure the body material is compatible with the gas type you will be using.
INTRODUCTION12
ALL RIGHTS OF CHANGE RESERVED
HOW TO SELECT VALVES
13INTRODUCTION
8 9
ALL RIGHTS OF CHANGE RESERVED ALL RIGHTS OF CHANGE RESERVED
SPECIFICATIONS
ALL FUNCTIONS IN ONE DEVICE
- Active Pressure Measurement - High Pressure Shut Off Valve - Filling Port with Check Valve - Residual Pressure Valve - Pressure Regulator - Safety Relief Device - Flow Selector - Medical Outlet Connection (Optional)
REGULATORY
- Medical CE certifi ed according to MDD 93/42/EEC amended 2007/47/EC
- π certifi ed according TPED Directive 2010/35/EU - Product complies with EN ISO 10524-3 and CGA E-18
- Medical integrated valve for Oxygen and others medical gases
- For all hospital and homecare applications
- Complies with the most stringent standards and requirements
- Approved by the major medical authorities worldwide
- Available in many diff erent confi gurations
M706 | ALPINOX INTEGRATED VALVE
KEY ADVANTAGES
Inlet Connection 17E / 25E / M18 / ¼ ”- 16 UNF
Filling Port ISO 5145 / CGA V1 with check valve
Flow Outlet Hose Barb
Pressure Outlet Acc. to national standards
Inlet pressure 200 bar / 2900 psi
Temperature range
-20°C to +60°C
Outlet pressure 4.5 bar (EU)3.5 bar (USA)
Flow range 12 pos.
0-6 Lpm / 0-15 Lpm / 0-25 Lpm
Body Chrome plated brass
Seats PA 6.6
O-ring EPDM
Filters Sintered Bronze 20 μm
OPTIONS
Protection Guard CSN / MIF / CBNE with bed hanger Inlet Device Filter / Anti-dust tube / Excess Flow device
PRODUCT REFERENCES
Gas Working pressure Inlet Connection Filling port Outlet Pressure Flow Range Medical connector
M70600153 O2 200 bar 25E ISO 5145 4.5 bar 0-25 Lpm DIN 13260-2
M70600152 O2 200 bar 25E ISO 5145 4.5 bar 0-25 Lpm NFS 90-116
M70600130 O2 2000 psi ¼ ” - 16 UNF CGA 540 50 psi 0-25 Lpm DISS 1240
M70600109 O2 200 bar 17E ISO 5145 4.5 bar 0-15 Lpm None
M70600121 N2 O / O2 135 bar 25E ISO 5145 4.5 bar 0-15 Lpm NFS 90-116
Others versions and possibilities available upon request
Simple Use - Ready-to-use Design - Simple, quick and intuitive function to deliver gas - Easy access to all functions
High Performances - Proven and Reliable Single stage piston regulator - High Pressure and Flow Stability and Accuracy outstanding fl ow outlet pressure stability. Exceptionally accurate.
- Wide fl ow ranges for many applications
Safety First - No high pressure manipulation - Shut off valve to isolate high pressure if required - Preset integrated pressure relief device - Flow selector design reducing likelihood of being set between two settings
- Continuous fl ow between settings - Active pressure gauge for a permanent indication - RPV function in order to avoid any potential contamination - Choice of material ensures patient health and safety
M706 M706 with plastic guard
CYLINDER VALVES INTEGRATED WITH PRESSURE REGULATORCYLINDER VALVES INTEGRATED WITH PRESSURE REGULATOR
TECHNICAL PARAMETER
Working pressure
Outlet pressure (only for VIPR)
Gas type or medical mixture
Material: raw brass or chrome plated
Inlet connection: 17E, 25E, 3⁄14”-16 UNF
Outlet connection valve- according to standards
Filling port (only for VIPR): acc ISO 5145 or CGA V1.,...
Residual pressure: if requested or no
Gauge: if specifi c request
Temperature range: -20°C/+65°C or specifi c
Description of the fi nal application: hospital homecare, emergency, or other
Guard: select a model and confi rm type of cylinder and cylinder weight
Guard color: white or green / Other colors on request + marking
Bed hanger (only for VIPR): if required or not
Environment: if special
Product applicable standards: EN ISO 10524-3 (VIPR) EN ISO 10297, CGA E-18
Certifi cation / Approval / RegistrationCE marked - 93/42/EEC European Medical Device Directive, FDA 510(k) – USA, TGA (Therapeutic Goods Administration) - Australia, Health Canada – Singapore, ANVISA – Brasil
To choose the right valve for your application and get the best results, you should identify the following technical parameters:
Each product page is designed to provide you the essential technical information at a glance:
Standard manometer available
14
ALL RIGHTS OF CHANGE RESERVED
VIPR
ALL FUNCTIONS IN ONE DEVICE
- Active Pressure Measurement - Filling Port with Check Valve - Residual Pressure Valve - Pressure Regulator - Safety Relief Device - Flow Selector - Medical Outlet Connection (Optional)
REGULATORY
- Medical CE certifi ed according to MDD 93/42/EEC amended 2007/47/EC
- π certifi ed according TPED Directive 2010/35/EU - Product complies with EN ISO 10524-3 and CGA E-18
M707 | LIFEBALL
- Ultra compact medical integrated valve for O2 and other medical gases
- For all hospital and homecare applications
- Centralized control unit on top of the value
- High legibility of fl ow settings with large and phosphorescent chraracters
- Very compact and light design
- High presicion thanks to compensated regulator
- Unique intellingent electronic system to display and manage yuour gas use (for more info please refer to page 9)
1515
ALL RIGHTS OF CHANGE RESERVED
VIPR
SPECIFICATIONS
Inlet Connection 25E/17E/¾”-16UNF
Filling Port CGA 540/CGA 577/ISO W24x2/ ISO W30x2/W22x14R
Flow Outlet Hose Barb acc. to Iso 13544-2
Pressure Outlet Acc. to national standards
Inlet pressure 200 bar/2000 psi 300 bar/3000 psi
Temperature range
-20°C to +60°C
Outlet pressure 3.5 bar/43 psi 4.5 bar/58 psi
Flow range 12 pos.
0-2 Lpm / 0-6 Lpm / 0-15 Lpm
Body Chrome plated brass
Seats PA 6.6
O-ring EPDM
Filters Sintered Bronze 60 μm
MAIN FUNCTIONNALITIES
To better understand roles and benefits of each of these functions please refer to our technological review section on page 9
1 VALVE INLET
2 ANTIDUST TUBE
3 FILLING PORT
4 GAUGE
4
7
5
8
6
5 RPV
6 HOSE BARB
7 FLOWMETER
8 REGULATOR & SAFETY RELIEF DEVICE
9 BED HANGER
10 OUTLET PRESSURE: MEDICAL CONNECTOR
1
3
9
2
10
PRODUCT CONFIGURATOR
GasInlet
ConnectionWorking Pressure
Filling Port
HP SafetyOutlet
PressureFlow Range
Pressure outlet
Flow outlet Inlet optionsGuard Color
Bed Hanger
M707 O2 A 200 ISO N 3.5 2 N HB6 EFD W BH
O2 O2 17E A 200bar 200 ISO 5145
ISO None N 3.5bar 3.5 0-2Lpm 2 None N Hose Barb Ø6mm
HB6 Excess flow device
EFD White W Yes BH
Air AIR 25E B 300bar 300 CGA V1
CGA CGA S1.1
CGA 4.5bar 4.5 0-6Lpm 6 AFNOR NF Anti dust tube
ADT Green G
N2O/O2 N2O 3/4"-16 UNF
CCS 2000Psi 2000 UNI UNI JIS JIS 0-15Lpm 15 DIN DIN
3000Psi 3000 JIS JIS 0-25Lpm 25 BS BS
w14,5MPa 14.5 SS SS
JIS JIS
AS AS
DISS DISS
16
ALL RIGHTS OF CHANGE RESERVED
ALL FUNCTIONS IN ONE DEVICE
- Active Pressure Measurement - High Pressure Shut Off Valve - Filling Port with Check Valve - Residual Pressure Valve - Pressure Regulator - Safety Relief Device - Flow Selector - Medical Outlet Connection (Optional)
REGULATORY
- Medical CE certifi ed according to MDD 93/42/EEC amended 2007/47/EC
- π certifi ed according TPED Directive 2010/35/EU - Product complies with EN ISO 10524-3 and CGA E-18
- Medical integrated valve for Oxygen and others medical gases
- For all hospital and homecare applications
- Complies with the most stringent standards and requirements
- Approved by the major medical authorities worldwide
- Available in many diff erent confi gurations
M706/M713 | LIFE+
Other combination available
Confi g. 1
Confi g. 2
VIPR
+
Filling PortShut Off Valve
Flow Outlet+ Pressure Gauge
MedicalOutlet
Connector
Filling Port
Shut Off Valve
Flow Outlet+ Pressure Gauge
MedicalOutlet
Connector
17
ALL RIGHTS OF CHANGE RESERVED
Confi g 1
SPECIFICATIONS
Inlet Connection 17E / 25E / M18 / ¾ ”- 16 UNF
Filling Port ISO 5145 / CGA V1 with check valve
Flow Outlet Hose Barb
Pressure Outlet Acc. to national standards
Inlet pressure M706:200 bar / 2900 psiM709:300bar
Temperature range
-20°C to +60°C
Outlet pressure 4.5 bar (EU)3.5 bar (USA)
Flow range 12 pos.
0-6 Lpm / 0-15 Lpm / 0-25 Lpm
Body Chrome plated brass
Seats PA 6.6
O-ring EPDM
Filters Sintered Bronze 20 μm
OPTIONS
Protection Guard CSN2 (see main picture) / MIF (see on page 19) with bed hanger
Inlet Device Filter / Anti-dust tube / Excess Flow device
VIPR
MAIN FUNCTIONNALITIES
To better understand roles and benefi ts of each of these functions please refer to our technological review section on page 9
1 VALVE INLET
2 ANTIDUST TUBE
3 FILLING PORT
4 GAUGE
5 RPV
6 HOSE BARB
6
7 FLOWMETER
8 REGULATOR & SAFETY RELIEF DEVICE
9 SHUT OFF VALVE
10 OUTLET PRESSURE: MEDICAL CONNECTOR
4
7
5
8
1
3
9
2
10
PRODUCT CONFIGURATOR
GasInlet
ConnectionWorking Pressure
Filling Port
HP SafetyOutlet
PressureFlow Range
Pressure outlet
Flow outletInlet
optionsGuard type
Guard Color
Confi gu-ration
M706/M713 O2 A 200 ISO N 3.5 2 N HB6 EFD CSN W
O2 O2 17E A 200bar 200 ISO 5145
ISO None N 3.5bar 3.5 0-2Lpm 2 None N Hose Barb Ø6mm
HB6 Excess fl ow device
EFD CSN CSN White W Confi g 1
Air AIR 25E B 300bar 300 CGA V1
CGA CGA S1.1
CGA 4.5bar 4.5 0-6Lpm 6 AFNOR NF Anti dust tube
ADT MIF MIF Green G Confi g 2
N2O/O2
N2O 3/4"-16 UNF
CCS 2000Psi 2000 UNI UNI JIS JIS 0-15Lpm 15 DIN DIN
3000Psi 3000 JIS JIS 0-25Lpm 25 BS BS
w14,5MPa
14.5 SS SS
JIS JIS
AS AS
DISS DISS
18
ALL RIGHTS OF CHANGE RESERVED
ALL FUNCTIONS IN ONE DEVICE
- Active Pressure Measurement - Cylinder Safety Device - Filling Port with Check Valve - Pressure Regulator - Safety Relief Device - Flow Selector - Medical Outlet Connection (Optional)
REGULATORY
- Product complies with CGA E-18
- Compact Medical Integrated Valve for Oxygen and others medical gases
- For all hospital and homecare applications
- For situations where performance, compactness and light weight are required
- Available in many different configurations
- Provided with a plastic integrated hanger to carry and protect the valve
M741 | LIFE-s
VIPR
PRODUCT CONFIGURATOR
GasInlet
ConnectionWorking Pressure
Filling Port HP SafetyOutlet
PressureFlow Range
Pressure outlet
Flow outlet Inlet options Guard Color
M741 O2 A 200 ISO N 3.5 2 N HB6 EFD
O2 O2 17E A 200bar 200 ISO 5145 ISO None N 3.5bar 3.5 0-2Lpm 2 None N Hose Barb Ø6mm
HB6 Excess flow device
EFD White W
Air AIR 25E B 300bar 300 CGA V1 CGA CGA S1.1
CGA 4.5bar 4.5 0-6Lpm 6 AFNOR NF Anti dust tube
ADT Green G
N2O/O2 N2O 3/4"-16 UNF
CCS 2000Psi 2000 UNI UNI JIS JIS 0-15Lpm 15 DIN DIN
3000Psi 3000 JIS JIS 0-25Lpm 25 BS BS
w14,5MPa 14.5 SS SS
JIS JIS
AS AS
DISS DISS
19
ALL RIGHTS OF CHANGE RESERVED
SPECIFICATIONS
Inlet Connection ¾” - 16 UNF
Filling Port CGA 540 with check valve
Flow Outlet Hose Barb
Pressure Outlet DISS 12409/16”-18UNF
Inlet pressure 134 bar2000 psi / bar
Temperature range
-20°C to +65°C
Outlet pressure 3.5 bar/50 psi
Flow range 12 pos.
0-6 Lpm / 0-15 Lpm / 0-25 Lpm
Body Chrome plated brass
Seats PA 6.6
O-ring EPDM
Filters Sintered Bronze
OPTIONS
Inlet Device Filter / Anti-dust tube / Excess Flow device
VIPR
MAIN FUNCTIONNALITIES
To better understand roles and benefits of each of these functions please refer to our technological review section on page 9
1 VALVE INLET
2 FILLING PORT
3 GAUGE
4 HOSE BARB
5 FLOWMETER
6 REGULATOR & SAFETY RELIEF DEVICE
7 OUTLET PRESSURE: MEDICAL CONNECTOR3
7
5
1
6
2
4
20
ALL RIGHTS OF CHANGE RESERVED
20
M800 | EMERGENCY BACK PACK
KIT CONTENTS
- 1 back pack: made of polyester · PVC reinforecement and inner foam construction · Large storage for additional items · Shoulder straps for hand-free intervention
- 1 cylinder · made of aluminium alloy · 2L oxygen at 200 bar
- 1 integrated valve · complies with EN ISO 10524-3 · 0-15 Lpm fl ow range · Residual pressure device · Pressure gauge 0-315 bar · Pressure-relief valve · Excess fl ow device
- 2 Nasal cannulas · Made of lightweight fl exible plastic tubing · Adult and children sizes
KEY FEATURES
- The bag is reinforced, waterproof and secures a 2 liters oxygen cylinder equipped with an integrated valve and 2 sizes of nasal cannulas.
- A complete safe contained medical oxygen breathing kit for mobile oxygen therapy
SPECIFICATIONS
Gas Medical Oxygen
Cylinder content 2L
Cylinder Pressure 200 bar/2900 psi
Flow range 0-15 Lpm
Gas Autonomy 26 min (at 15 Lpm)
Total weight 5,8 Kg
CE marking π / CE0029
VIPR
21
ALL RIGHTS OF CHANGE RESERVED
21
ALL FUNCTIONS IN ONE DEVICE
- Available with an optional bed hanger for an easier handling and support
REGULATORY
- Designed and tested according to EN ISO 11117 standard
- Guards and Caps to protect the valve during service and transportation while giving the product an easy access and a good looking design
- Designed to protect all products depending on customers specifi c requirements
SPECIFICATIONS
Connection On valve body / thread W80 x 1/11”
Permitted mass Up to 60 Kg
Material PA 6.6
Temperature range
-20°C to +65°C
Dimensions Ø143 mm max.
Colors White / Black / Blue / Green
OPTIONS
Bed Hanger Assembled on the guard or cap (nif)
Label Customized labeling and/or marking
M706/M713 PROTECTIVE GUARD
MIF CSN 2
VIPR
ALL RIGHTS OF CHANGE RESERVED
22
M510/M515/M501/M506/M520/M525 | STANDARD RPV OFF LINE
KEY FEATURES
- Low operating torque - MRI compatible - Safe, long, trouble free life under all service conditions
- Designed and tested according EN ISO 10297 and EN ISO 15996
OPTIONS
- Inlets: 17 E – 25 E – M 18 – US threads - Filter - Burst disc safety device - Thread for dip tube connection - Chrome plating
- High pressure brass cylinder valve for medical gases, O-ring seal type with residual pressure
RPV VALVE
SPECIFICATIONS
Working pressure 200 bar/ 2900 psi
Temperature range -25°C to +70°C
Helium leak rate at pmax
10-3 mbar l/sec
Seat orifi ce size 4 mm
Handwheel material aluminium
Seat material polyamid
O-ring material EPDM
Body material Chrome plated brass
PRODUCT CONFIGURATOR
Gas Inlet Connection Working Pressure Outlet connection HP Safety
O2O2 17E A 200bar 200 ISO 5145 ISO None N
N2O N2O 25E B 300bar 300 CGA V1 CGA CGA S1.1 CGA
Various gases (*) VG * ¾ "-16 UNF CCS 2000Psi 2000 UNI UNI JIS JIS
CO2 CO2 ¾ " NGT OBS 3000Psi 3000 DIN DIN Eu CO2/N2O EU
AFNOR AFNOR
BS BS
SS SS
AS AS
Can be connected with Ceodeux Meditec regulators
Ref catalog: MPR-EN-1115
PRODUCT SERIE FEATURES
GAS SAFETY
O2 CO2 VARIOUS GASES WITH WITHOUT
M510
M515
M501
M506
M520
M525
* gas or mixture must be confi rmed
ALL RIGHTS OF CHANGE RESERVED
Ref catalog: MPR-EN-1115
23
M540/M545/M530/M535 | STANDARD RPV OFF LINE
KEY FEATURES
- Low operating torque - MRI compatible - Safe, long, trouble free life under all service conditions
- Designed and tested according EN ISO 10297 and EN ISO 15996
OPTIONS
- Inlets: 17 E – 25 E – M 18 – US threads - Filter - Burst disc safety device - Thread for dip tube connection - Chrome plating
- High pressure brass cylinder valve for medical gases, O-ring seal type with residual pressure
RPV VALVE
SPECIFICATIONS
Working pressure 300 bar/ 4350 psi
Temperature range -25°C to +70°C
Helium leak rate at pmax
10-3 mbar l/sec
Seat orifi ce size 4 mm
handwheel material aluminium
Seat material polyamid
O-ring material EPDM
Body material Chrome plated brass
Can be connected with Ceodeux Meditec regulators
PRODUCT SERIE FEATURES
GAS SAFETY
O2 VARIOUS GASES WITH WITHOUT
M540
M545
M530
M535
PRODUCT CONFIGURATOR
Gas Inlet Connection Working Pressure Outlet connection HP Safety
O2O2 17E A 300bar 3000 ISO 5145 ISO None N
N2O N2O 25E B CGA V1 CGA CGA S1.1 CGA
Various gases (*) VG * ¾ "-16 UNF CCS UNI UNI JIS JIS
¾ " NGT OBS DIN DIN Eu CO2/N2O EU
AFNOR AFNOR
BS BS
SS SS
AS AS
* gas or mixture must be confi rmed
ALL RIGHTS OF CHANGE RESERVED
24
M511/M516/M508/M521/M526 | STANDARD RPV IN LINE
KEY FEATURES
- Low operating torque - MRI compatible - Safe, long, trouble free life under all service conditions
- Designed and tested according EN ISO 10297 and EN ISO 15996
OPTIONS
- Inlets: 17 E – 25 E – M 18 – US threads - Filter - Burst disc safety device - Thread for dip tube connection - Chrome plating
- High pressure brass cylinder valve for medical gases, O-ring seal type with residual pressure
RPV VALVE
SPECIFICATIONS
Working pressure 200 bar/ 2900 psi
Temperature range -25°C to +70°C
Helium leak rate at pmax
10-3 mbar l/sec
Seat orifi ce size 4 mm
handwheel material aluminium
Seat material polyamid
O-ring material EPDM
Body material Chrome plated brass
Can be connected with Ceodeux Meditec regulators
Ref catalog: MPR-EN-1115 PRODUCT SERIE FEATURES
GAS SAFETY
O2 CO2 VARIOUS GASES WITH WITHOUT
M511
M516
M508
M521
M526
PRODUCT CONFIGURATOR
Gas Inlet Connection Working Pressure Outlet connection HP Safety
O2O2 17E A 200bar 200 ISO 5145 None
N2O N2O 25E B 300bar 300 CGA V1 CGA S1.1
Various gases (*) VG * ¾ "-16 UNF CCS 2000Psi 2000 UNI JIS
CO2 CO2 ¾ " NGT OBS 3000Psi 3000 JIS
* gas or mixture must be confi rmed
ALL RIGHTS OF CHANGE RESERVED
Ref catalog: MPR-EN-1115
25
M978/M979/C878 | STANDARD RPV IN LINE
KEY FEATURES
- Low operating torque - MRI compatible - Safe, long, trouble free life under all service conditions
- Designed and tested according EN ISO 10297 and EN ISO 15996
OPTIONS
- Inlets: 17 E – 25 E – M 18 – US threads - Filter - Burst disc safety device - Thread for dip tube connection - Chrome plating
- High pressure brass cylinder valve for medical gases, O-ring seal type with residual pressure
RPV VALVE
SPECIFICATIONS
Working pressure 200 bar/ 2900 psi
Temperature range -25°C to +70°C
Helium leak rate at pmax
10-3 mbar l/sec
Seat orifi ce size 4 mm
handwheel material aluminium
Seat material polyamid
O-ring material EPDM
Body material Chrome plated brass
Specifi c M978: O2
M9769: various gasessC878: O2 + integrated valve
Can be connected with Ceodeux Meditec regulators
PRODUCT SERIE FEATURES
GAS SAFETY
O2 VARIOUS GASES WITH WITHOUT
M978
M979
* gas or mixture must be confi rmed
PRODUCT CONFIGURATOR
Gas Inlet Connection Working Pressure Outlet connection HP Safety
O2O2 17E A 200bar 200 DIN DIN None N
N2O N2O M18X1,5 RCS Eu CO2/N2O EU
Various gases (*) VG *
CO2 CO2
ALL RIGHTS OF CHANGE RESERVED
RPV VALVE26
SPECIFICATIONS
Working pressure 232 bar
Temperature range -25°C to +70°C
Helium leak rate at pmax
10-3 mbar l/sec
Seat orifi ce size M100/M105/M200/M205 : 4mmM040/M045: 8mm
Magnetic compatibility
yes
handwheel material polyamid
Seat material polyamid
O-ring material EPDM
Body material Chrome plated brass
KEY FEATURES
- Low operating torque - MRI compatible - Safe, long, trouble free life under all service conditions
- Approved valve design - Testing according ISO 10297 and CGA U9 Standard
OPTIONS
- Inlets: 17 E – 25 E – M 18 – US threads - Filter - Burst disc safety device - Thread for dip tube connection - Chrome plating
- High pressure brass cylinder valve for medical oxygen
- O-ring seal type
M100/M105/M040/M045/M200/M205 | STANDARD CYLINDER VALVE
Can be connected with Ceodeux Meditec regulators
Ref catalog: MPR-EN-1115
- High pressure brass cylinder valve for medical oxygen
- O-ring seal type
PRODUCT SERIE FEATURES
GAS SAFETY
O2 CO2 VARIOUS GASES WITH WITHOUT
M100
M105
M040
M045
M200
PRODUCT CONFIGURATOR
Gas Inlet Connection Working Pressure Outlet connection HP Safety
O2O2 17E A 300bar 3000 ISO 5145 ISO None N
N2O N2O 25E B CGA V1 CGA CGA S1.1 CGA
Various gases (*) VG * ¾ "-16 UNF CCS UNI UNI JIS JIS
¾ " NGT OBS DIN DIN Eu CO2/N2O EU
AFNOR AFNOR
BS BS
SS SS
AS AS
* gas or mixture must be confi rmed
ALL RIGHTS OF CHANGE RESERVED
M972/M973/M974/M975 | STANDARD CYLINDER VALVE SMALL CONE
SPECIFICATIONS
Working pressure 200 bar
Temperature range -20°C to +70°C
Helium leak rate at pmax
10-3 mbar l/sec
Seat orifi ce size 2.5 mm
Residual pressure 1-3 bar
handwheel material aluminium
Seat material PA 6.6
O-ring material EPDM
Body material Chrome plated brass
Filter material brass sintered
KEY FEATURES
- Low operating torque under up to 200 bar service pressure
- MRI compatible - Designed and tested acc. EN ISO 10297, EN ISO 15996 - Residual pressure setting 1-3 bar - Manufactured in cleanroom - Inlet with fi lter - Possibility of integrated gauge
- OPTIONS
- Inlet: 17E – M18 – M 14 - Without fi lter
Can be connected with Ceodeux Meditec regulators
Ref catalog: MPR-EN-1115
STANDARD CYLINDER VALVE 27
- High pressure brass cylinder valve for medical oxygen
- O-ring seal type
- High pressure brass cylinder valve for medical oxygen
- O-ring seal type - Small cone inlet size
PRODUCT SERIE FEATURES
GAS SAFETY
O2 VARIOUS GASES WITH WITHOUT
M972
M973
M974
M975
PRODUCT CONFIGURATOR
Gas Inlet Connection Working Pressure Outlet connection HP Safety
O2O2 17E A 200bar 200 DIN DIN None N
N2O N2O M18X1,5 RCS Eu CO2/N2O EU
Various gases (*) VG *
CO2 CO2
* gas or mixture must be confi rmed
ALL RIGHTS OF CHANGE RESERVED
STANDARD CYLINDER VALVE28
- High pressure cylinder valve with RPV, packed needle stem design, for NO mixtures
D113 | PACKED VALVE
SPECIFICATIONS
Working pressure 200 bar/3000psi
Temperature range -40°C to +70°C
Helium leak rate at pmax
10-6 mbar l/sec10-5 mbar l/sec10-8 mbar l/sec
Flow coeffi cient Cv 0.3
Seat orifi ce size 3,3 mm
Body material Stainless steel
Residual pressure 1 to 5 bar
RPV seat material Viton®
KEY FEATURES
- Freedom from seizure when used with corrosive gases - gas does not come into contact with the valve operating mechanism
- Positive and eff ective operation - valve spindle is mechanically linked to the operating handwheel, which only requires low closing torque
- Long life - valve spindle and seat are replaceable - Adjustable gland packing - in the unlikely event of a leakage occuring past the valve spindle, when opening the valve, rectifi cation is readily achieved by retightening the gland packing nut
- Low internal volume - this is particularly benefi cial, when using the valve for hygroscopic, corrosive or high purity gases, here inert gas purging may be necessary before opening or after closing the valve
- Left hand threaded counter screw to secure sealing mechanism even under excessive torque
12
3~
6 3~
3 9 , 5~7 , 5~
PRODUCT CONFIGURATOR
Gas Inlet Connection Working Pressure Outlet connection HP Safety
"NO mixture up to 1000 ppm” ¾ " NGT OBS 200bar 200 ISO 5145 ISO None N
25E B CGA V1 CGA CGA S1.1 CGA
Can be connected with Ceodeux Meditec regulators
Ref catalog: MPR-EN-1115
ALL RIGHTS OF CHANGE RESERVED
SPECIAL NITRIC OXIDE CYLINDER VALVE 29
- High fl ow cylinder valves with RPV, tied diaphragm seal type, for NO mixtures
D413 | TIED DIAPHRAGM
KEY FEATURES
- Reduced dead spaces and gas wetted volumes allow easy purging process
- Freedom of internal threads and springs insures minimum particulate generation
- Very low internal volume with minimum dead spaces. This is particularly benefi cial, when using the valve for hygroscopic, corrosive, pyrophoric or high purity gases where inert gas purging is necessary before opening or after closing the valve
- Freedom from seizure where used with corrosive gases.Gas does not come into contact with the valve operating mechanism
- Positive and eff ective operation, valve lower spindle is mechanically linked to the upper spindle and the operating handwheel
- Backed up welded diaphragms provide a permanent and durable gas tight seal
- O-Rings in the gland and on spindle guarantee a secondary sealing
- Threaded vent hole
SPECIFICATIONS
Working pressure 200 bar/3000 psi
Temperature range -40°C to +70°C
Helium leak rate at pmax
10-8 mbar l/sec
Flow coeffi cient Cv 0,35
Seat orifi ce size 4 mm
Gas wetted materials PCTFE
Body material 316L Stainless steel
Residual pressure 1 to 5 bar
RPV seat material Viton®
7 , 5~ 5 6~
12
0~
5 4~
PRODUCT CONFIGURATOR
Gas Inlet Connection Working Pressure Outlet connection HP Safety
"NO mixture up to 1000 ppm” ¾ " NGT OBS 200bar 200 ISO 5145 ISO None N
25E B CGA V1 CGA CGA S1.1 CGA
Can be connected with Ceodeux Meditec regulators
Ref catalog: MPR-EN-1115
ALL RIGHTS OF CHANGE RESERVED
SPECIAL NITRIC OXIDE CYLINDER VALVE30
SPECIFICATIONS
Working pressure 232 bar
Temperature range -25°C to +70°C
Helium leak rate at pmax
10-3 mbar l/sec
Seat orifi ce size 2.5 mm
Magnetic compatibility
yes
Key – operated chrome plated
Handwheel material plastic switch blade
Seat material PA 6.6
O-ring material EPDM
Body material Chrome plated brass
Filter material brass sintered
KEY FEATURES
- Low operating torque - MRI compatible - Key-operated or handwheel operating system - Safe, long, trouble-free life under all service conditions
- Testing according EN ISO 10297 and CGA U9 Standard
OPTIONS
- Inlets: 17E – 19E – 19T – LT – T – M18 – US threads - Natural brass - Burst disc safety device - Fusible metal safety device - Other fi lter (lang)
- High pressure brass cylinder valve for medical gases and oxygen,
- with yoke outlet connection (PIN INDEX System)
- O-ring-seal
M966 | PIN INDEX CYLINDER VALVE
Can be connected with Ceodeux Meditec regulators
I.E: Alpinox Pin indexRef catalog: MPR-EN-1115
10
3~
3 2~
* gas or mixture must be confi rmed
PRODUCT CONFIGURATOR
Gas Inlet Connection Working Pressure Outlet connection HP Safety
O2O2 17E A 200bar 200 ISO 407 ISO None N
N2O N2O M18X1,5 RCS CGA CGA CGA S1.1 CGA
Various gases (*) VG * 25E B
CO2 CO2
ALL RIGHTS OF CHANGE RESERVED
31PIN INDEX VALVE
- PIN INDEX valve for medical gases with active residual pressure function
M650 | PIN INDEX RPV CYLINDER VALVE
SPECIFICATIONS
Working pressure 232 bar
Temperature range -25°C to +70°C
Helium leak rate at pmax
10-3 mbar l/sec
Seat orifi ce size 2.5 mm
Cv 0.08
Magnetic compatibility
yes
Key – operated chrome plated
handwheel material removable plastic switch blade
Seat material PA 6.6
O-ring material EPDM
Body material chrome plated / natural brass
Filter material brass sintered
KEY FEATURES
- High tensile brass (CZ115) - Low operating torque - MRI compatible - Key-operated or handwheel operating system - Design and tested acc. EN849, EN ISO407, EN ISO15996 and CGA V9
- Residual pressure - Manufacturing in clean room
OPTIONS
- Inlets: 17E – 19E – 19T – LT – T – M18 – US threads - Natural brass - Burst disc safety device - Fusible metal safety device - Other fi lter (lang)Can be connected with
Ceodeux Meditec regulators
I.E: Alpinox Pin indexRef catalog: MPR-EN-1115
PRODUCT CONFIGURATOR
Gas Inlet Connection Working Pressure Outlet connection HP Safety
O2O2 17E A 200bar 200 ISO 407 ISO None N
N2O N2O M18X1,5 RCS CGA CGA CGA S1.1 CGA
Various gases (*) VG * 25E B
CO2 CO2
* gas or mixture must be confi rmed
32
ALL RIGHTS OF CHANGE RESERVED
NOTES
3333
ALL RIGHTS OF CHANGE RESERVED
NOTES
A WORLD OF GAS CONTROL SOLUTIONSROTAREX is helping engineers worldwide to get better gas results: from ultra high purity production and medical care facilities to industrial and LPG applications, as well as alternative energy vehicles, fi re suppression, diving, aerospace, cryogenics, laboratory, petro-chemical and welding. ROTAREX applies over 90 years of know-how and experience to custom design, develop and manufacture the high performance valves, regulators and fi ttings to suit your needs, all in one hand. Discover the diff erence ROTAREX can make in your world.
COMPLETE SOLUTIONS FROM SOURCE TO PROCESS.
MEDITECCYLINDER VALVES EQUIPMENT FIRETEC AUTOMOTIVE LPG/SRG
AUSTRALIA / BRASIL / CHINA / FRANCE / INDIA / ITALY / JAPAN / LUXEMBOURG / MIDDLE EAST / RUSSIA / S INGAPORE / SPAIN / THAILAND / USA
INDUSTRIALCYLINDER VALVES
UHPCYLINDER VALVES
MEDICAL VALVES& REGULATORS
FIRE SUPPRESSIONLPG VALVES
REGULATORS & GAUGES
SPECIALTY GAS VALVES REGULATORS & FITTINGS CRYOGENIC VALVES AUTOMOTIVE
WATERTEC
www.rotarex.com
WORLDWIDE HEADQUARTERS
Ceodeux Meditec Customer Service
24 rue de Diekirch, B.P. 19
7505 Lintgen, Luxembourg
Tel.: +352 32 78 32 1
Fax: +352 32 78 32 854
E-mail: info@meditec.rotarex.com
© 2012 - ROTAREX S.A.
CEODEUX MEDITEC - CEODEUX S.A.
MCV-EN-1115
REGIONAL / COUNTRY HEAD QUARTERS
NORTH AMERICA
USARotarex North America
221 Westec Drive
Westmoreland Technology Park I
Mt. Pleasant, Pennsylvania 15666 USA
Tel.: +1 724-696-43 40
Fax: +1 724-696-43 64
E-mail: info@northamerica.rotarex.com
SOUTH AMERICA
BRASILRotarex Brazil Ltda
Cond. Ind. Portal da Anhanguera Estr.
Municipal Gov. Mário Covas, S/N
13279-411 Bairro Macuco -Valinhos
São Paulo Brazil
Tel.: +55 19 3518 0800
Fax: +55 19 3869-1503
E-mail: info@brazil.rotarex.com
EUROPE
FRANCE, BENELUX, GERMANY24, rue de Diekirch, BP 19
L-7505 Lintgen, Luxembourg
Tel.: +352 32 78 32-1
Fax: +352 32 78 32 317
E-mail: salesequipment@rotarex.com
ITALYRotarex Italia S.r.l.
46 Via Giacomo Matteotti
I-25080, Ciliverghe di Mazzano (BS) Italy
Tel.: +39 030 212 05 50
Fax: +39 030 212 23 62
E-mail: info@italia.rotarex.com
SPAINRotarex Spain
C/Alcala, 4° Oficina 404
28014 Madrid
Tel.: +34 (91) 570 89 05
Mobile: + 34 (650) 908 856
E-mail: info@spain.rotarex.com
RUSSIARotarex Rus
Tverskaya street, 20/1 bldg. 1 of 506
125009, Moscow Russia
Tel.: +7 985 125 7776
Skype: rotarexrus
E-mail: info-russia@rotarex.com
POLANDRotarex Polska
13 Gröbli
PL 49-300 Brzeg, Poland
Tel.: +48 77 416 40 16
Fax: +48 77 416 20 99
E-mail: info@poland.rotarex.com
ASIA
MIDDLE EASTRotarex Middle East
Jebel Ali Free Zone - Lob 17, Office 415
P.O. Box 261749
Dubai, U.A.E.
Tel.: +971 (0) 4 887 6701
Fax: +971 (0) 4 887 6702
E-mail: info@middle-east.rotarex.com
CHINARotarex Star
60 Yuan Zhong Road
Shanghai Nanhui Industrial Zone
201300, Shanghai China
Tel.: +86-21 5800 4000
Fax: +86-21 5800 3226
E-mail: info@china.rotarex.com
SINGAPORERotarex Fareast Pte Ltd
10 Ubi Crescent
Ubi Techpark, Lobby C, #06-55
408564 Singapore
Tel.: +65 64 72 37 27
Fax: +65 64 72 45 28
E-mail: info@singapore.rotarex.com
JAPANRotarex Japan Ltd
6F, Yawaragi Bldg., 1-27-11 Taito
Taito-ku, Tokyo
110-0016 JAPAN
Tel.: +81 3 5817 8108
Mob.: +81 80 3755 2539
Fax: +81 3 5817 8168
E-mail: info@japan.rotarex.com
INDIARotarex ENGG. PVT. LTD.
C-105, Mahindra Apartments
Vikaspuri, New Delhi 110018
mob.: +91-9873467666
E-mail: atul.nigam@rotarex.in
THAILANDRotarex (Thailand) Co Ltd.
297, Wanglee Tower
3rd floor, Unit C, Surawong Road
Suriyawong, Bangrak,
Bangkok 10500 Thailand
Tel.: +66 2635 71 79
E-mail: info@thailand.rotarex.com
SOUTH KOREARotarex Luz Tech Co.
305-500, 547 Yongsan-dong
Yuseong-gu
Taejon, Korea
Tel.: +82 42 825 8911
Fax: +82 42 825 8913
E-mail: info@rotarex.co.kr
OCEANIA
AUSTRALIARotarex Australia/New Zealand
Hornsby Heights,
Sydney, NSW, 2077
Tel.: +61 477 477 481
E-mail: info@australia.rotarex.com
Recommended