SURGICAL PROCEDURE GUIDE FOR IR SYSTEM USING THE UNIVERSAL KIT

Preview:

DESCRIPTION

SURGICAL PROCEDURE GUIDE FOR IR SYSTEM USING THE UNIVERSAL KIT. SURGICAL KIT – UNIVERSAL. 2. 1 – Guide Drill 2 – Pilot Drills 3 – Twist Drills 4 – Cortical Drills 5 – Drill Extension 6 – Fixture Driver (Manual) 7 – Fixture Driver (Machine) 8 – Torque Wrench 9 – 1.2mm Hex Driver - PowerPoint PPT Presentation

Citation preview

SURGICAL PROCEDURE GUIDEFOR IR SYSTEM USING THE UNIVERSAL KIT

SURGICAL KIT – UNIVERSAL

4

9 8

7 6

3

510

11

11 – Guide Drill2 – Pilot Drills 3 – Twist Drills4 – Cortical Drills5 – Drill Extension6 – Fixture Driver (Manual)7 – Fixture Driver (Machine)8 – Torque Wrench9 – 1.2mm Hex Driver10 – Parallel Pin11 – 1.2mm Hex Screw

2

GUIDE DRILL

BIOTEM Guide Drill• No slipping• Precise Positioning

Round Drill• Risk of slipping may lead

to: • Less precise

positioning• Damaging bone

tissue

Recommended rotational speed1200 rpm

Round Drill vs. Point Drill

Item Code

PD2029

PILOT DRILLS

Recommended rotational speed 700-800 rpm

Marker signals depth reached

Groove disperses cooling liquid

Diameter (Ø)

2.0mm 2.9mm

Item Code BTD2015 BTD2915

TWIST DRILLS

Recommended rotational speed 700-800 rpm

Marker signals depth reached

Groove disperses cooling liquid

Diameter (Ø)

3.3mm 3.6mm 4.0mm 4.6mm 5.6mm

Item Code BTD3315 BTD3615 BTD4015 BTD4615 BTD5615

CORTICAL DRILLS

Recommended rotational speed 700-800 rpm

Diameter (Ø)

3.4mm 3.7mm 4.0mm 5.0mm 6.0mm

Item Code CDA3411S CDA3711S CDA4011S CDA5011S CDA6011S

Groove Line sets the limit of drilling

DRILL EXTENSION

Robust connective capsule design with easy assembly / disassembly

Common connective groove designs create a risk for bending

Item Code

ETB4526

PARALLEL PIN

Item Code

DGA1620

Insert this side into Ø 2.0mm osteotomy (after using the straight drill)

Insert this side into the implant screw hole

HAND DRIVERS

Hex Driver (for torque wrench)

Hex Driver (Manual)

Hex (Ø) 1.2mm 1.2mm

Length 8mm (Short)

15mm (Long)

Item Code

HDT1208 HDT1215

Hex (Ø) 1.2mm 1.2mm

Length 8mm (Short)

15mm (Long)

Item Code

HDB1208 HDB1215

FIXTURE DRIVERS

Neck Size Mini Regular / Wide

Mini Regular / Wide

Color Code

White Blue White Blue

Short FDIM3108C FDIM2908C FDIH3108 FDIH2908

Long FDIM3112C FDIM2912C FDIH3115 FDIH2915

Contra-Angle Torque Wrench

Joining the driver with the implant – Simple and Safe

1. Insert driver into the appropriate connective space while applying pressure to the top of the implant holder.

2. Apply pressure at the bottom of the holder in order to release the implant-driver assembly.

DRILLING SEQUENCEIR FIXTURE DRILLING SEQUENCE

Guide Drill BTD2015 BTD2915 BTD3315 BTD3615 BTD4015 BTD4615

- Ø 2.0mm Ø 2.9mm Ø 3.3mm Ø 3.6mm Ø 4.0mm Ø 4.6mm

CDA3413S

Mini Ø 3.5mm

CDA4013S

Regular Ø 4.1mm

CDA4513S Regular / Wide Ø 4.8mm

CDA5013S

Wide Ø 5.5mm

1

1

2

2

3

4

3 4

*NOTE: Use cortical drill previous to inserting implant in hard bone.

* *

**

SURGICAL PROCEDURE MANUALFOR IR SYSTEM USING THE EXTENDED KIT

SURGICAL KIT - EXTENDED

21

4 65

7

108

9

1113

1 – Guide Drill2 – Spare Hole3 – Straight Drills4 – Tapered Drills5 – Marking Drills6 – Cortical Drills7 – Parallel Pins8 – Hand Drivers9 – Torque Wrench10 – Drill Extension11 – Spare Holes12 – Implant Drivers (2 Stage)13 – Implant Drivers (1 Stage)

12

3

Item Code

KIT0700

GUIDE DRILL

BIOTEM Guide Drill• No slipping• Precise Positioning

Round Drill• Risk of slipping may lead

to: • Less precise

positioning• Damaging bone

tissue

Recommended rotational speed1200 rpm

Round Drill vs. Point Drill

Item Code

PD2029

STRAIGHT DRILLS

Recommended rotational speed 1200 rpmHarvest bone at low rotational speed (30-50 rpm)

Length 8.5mm 10mm 11.5mm 13mm 15mm

Item Code BSD2008 BSD2010 BSD2011 BSD2013 BSD2015

Ø 2.0mm

Irrigation groove for easy penetration of cooling media

Stopper for increased drilling precision (prevents over-drilling)

Straight drill design

TAPERED DRILLS (1/2)

Recommended rotational speed 1200 rpmHarvest bone at low rotational speed (30-50 rpm)

Irrigation groove for easy penetration of cooling media

Stopper for increased drilling precision (prevents over-drilling)

Taper angle of 15°

TAPERED DRILLS (2/2)

Length 8.5mm 10mm 11.5mm 13mm 15mm

Item Code BTD3008 BTD3010 BTD3011 BTD3013 BTD3015

Ø 3.0mm

Length 8.5mm 10mm 11.5mm 13mm 15mm

Item Code BTD3308 BTD3310 BTD3311 BTD3313 BTD3315

Length 8.5mm 10mm 11.5mm 13mm 15mm

Item Code BTD3608 BTD3610 BTD3611 BTD3613 BTD3615

Length 8.5mm 10mm 11.5mm 13mm 15mm

Item Code BTD3808 BTD3810 BTD3811 BTD3813 BTD3815

Length 8.5mm 10mm 11.5mm 13mm 15mm

Item Code BTD4208 BTD4210 BTD4211 BTD4213 BTD4215

Ø 3.6mm

Ø 3.3mm

Ø 3.8mm

Ø 4.2mm

MARKING DRILLS

Recommended rotational speed 700-800 rpmHarvest bone at low rotational speed (30-50 rpm)

Diameter (Ø)

2.7mm 4.6mm 5.0mm 5.6mm

Item Code MDR2713 MDR4613 MDR5013 MDR5613

Irrigation groove for easy penetration of cooling media

Stopper for increased drilling precision (prevents over-drilling)

Lateral microgrooves enhance cooling media dispersion

13 11.5

10 8.5 7

CORTICAL DRILLS

Recommended rotational speed 700-800 rpm

Diameter (Ø)

3.4mm 3.7mm 4.0mm 4.5mm 5.0mm 6.0mm

Item Code

CDA3411 CDA3711 CDA4011 CDA4511 CDA5011 CDA6011

Marker signals Limit of groove line

Groove Line sets the limit of drilling

DRILL EXTENSION

Robust connective capsule design with easy assembly / disassembly

Common connective groove designs create a risk for bending

Item Code

ETB4526

PARALLEL PIN

Item Code

DGA1620

Insert this side into Ø 2.0mm osteotomy (after using the straight drill)

Insert this side into the implant screw hole

HAND DRIVERS

Hex Driver (for torque wrench)

Hex Driver (Manual)

Hex (Ø) 1.2mm 1.2mm

Length 8mm (Short)

15mm (Long)

Item Code

HDT1208 HDT1215

Hex (Ø) 1.2mm 1.2mm

Length 8mm (Short)

15mm (Long)

Item Code

HDB1208 HDB1215

FIXTURE DRIVERS

Neck Size Mini Regular / Wide

Mini Regular / Wide

Color Code

White Blue White Blue

Short FDIM3108C FDIM2908C FDIH3108 FDIH2908

Long FDIM3112C FDIM2912C FDIH3115 FDIH2915

Contra-Angle Torque Wrench

Joining the Driver with the Implant Fixture

1. Apply pressure on the top part of the fixture mount then insert driver into the appropriate connective space

2. Apply pressure on the bottom part of the fixture mount in order to release the implant-driver assembly.

Ergonomic mount design allows for a simple assembly with the driver as well as a safe manipulation.

DRILLING SEQUENCE

IR FIXTURE DRILLING SEQUENCE (1/2)

Point Drill Straight Drill Tapered Drill Implant Tapered Drill Tapered Drill Implant

- Ø 2.0mm Ø 3.0mm Ø 3.5mm Ø 3.3mm Ø 3.6mm Ø 4.1mm

DRILLING SEQUENCE

IR FIXTURE DRILLING SEQUENCE (2/2)

Tapered Drill Implant Marking Drill Marking Drill Implant

Ø 4.2mm Ø 4.8mm Ø 4.6mm Ø 5.0mm Ø 5.5mm

Recommended