Graft Material Micro-porous ePTFE tubing (102 ± 25µm)

Stent Material (Composition)

CoCr (L605)

Stent Graft Design Single Stent

Strut Dimensions (Strut Width x Strut Thick-ness)

0.135 x 0.145 mm (SV)0.145 x 0.145 mm (MV) 0.165 x 0.145 mm (LV)

Introducer Sheath Compatibility

6F up to Ø 6 x 58 mm7F

Guide Wire 0.035”

Shaft Size 5Fr

Balloon Marker Material

Platinum / Iridium

Nominal Pressure9 bar Ø 5.0 - 7.0 mm8 bar Ø 8.0 - 10.0 mm

Rated Burst Pressure13 bar Ø 5.0 - 7.0 mm12 bar Ø 8.0 - 10.0 mm

Catheter Shaft Length 75 and 120 cm

Nominal Stent Diameters5.0 , 6.0 mm (SV)7.0 , 8.0 mm (MV) 9.0 , 10.0 mm (LV)

Nominal Stent Lengths18, 22, 28, 38, 58 (SV)18, 23, 27, 37, 57 (MV) 27, 37, 57 (LV)

Shelf Life 3 years

PERIPHERAL

PeripheralStentGraftSystem

PeripheralS

tentGraftS

ystem

TheStentComplianceoftheBeGraftPeripheral

ADULT

PeripheralInterventions

Yourbenefits

smart through a patented ‘one layer’ stent graft solution competitive we are proud to have the lowest profile on the market * flexible 6F Introducer sheath compatibility up to ø 6 x 58 mm efficient we have proven excellency of crossability and trackability * compatible biocompatible CoCr alloy and micro-porous ePTFE covered expansion range from Ø 5.0 mm to Ø 10.0 mm, stent lengths from 18 mm to 58 mm

Yourspecifications

* Data on file at Bentley InnoMed

Indications

The BeGraft Peripheral Stent Graft System is indicated for intraluminal chronic placement in iliac and renal arteries for:

Restoring and improving the patency Treating aneurysms, acute perforations, acute ruptures and fistulas

YournewBe-formationBentley InnoMed is proud to introduce a new arrival for minimally invasive procedures.

Inflation Pressure

[bar]

Stent Outer Diameter [mm]

Ø 5.0 Ø 6.0 Ø 7.0 Ø 8.0 Ø 9.0 Ø 10.0

8 8.0 9.0 10.0

9 5.0 6.0 7.0 8.3 9.2 10.2

10 5.2 6.2 7.2 8.5 9.4 10.4

11 5.3 6.4 7.3 8.7 9.6 10.6

12 5.4 6.5 7.5 8.8 9.7 10.7

13 5.5 6.6 7.6

NP Nominal Pressure

Rated Burst PressureRBP

Dear Customer, Bentley InnoMed guarantees a unique know-

ledge in the development of minimally invasive

products based on many years of hard work

and collaboration with our customers and clients.

As a result of requests from the market,

we combine all our expertise to develop

new products of high quality like the BeGraft

Peripheral Stent System.

This folder has now been enriched with Layar,

which allows you to view digital content with your

smartphone. Dowload the free Layar app, hold

the phone above the images and ‘Tap to View.’

Thank you for your trust,

Miko Obradovic and team

Join innovation by scanning the new BeGraft Peripheral product with the Layar app.

Recommended