MEDICAL OXYGEN GENERATING PLANTS M-DGS2

Oxygen Plant model

Code

PRIMARY and SECONDARY SOURCESRESERVESOURCE

8 bar

5 bar

5 bar

5 bar

7-9

bar

200 bar 200 bar

8 bar

5 bar

l Oxygen concentration produced by oxygen plant in percentage 94±2%;

l Oxygen values: taken from European Pharmocopoein;

l Oxygen supply pressure to the consumer, 4,5÷5,0 bar;

3 l Oxygen efficiency ≤ 1,4 kW/m ;

l 2 Screw type compressors;

l 2 Refrigerating Air Dryers;

l 2 Air Receivers;

l 2 Cole Towers with filters;

l 2 Oxygen concentrators;

l Oxygen control filtration unit;

l Reserve oxygen cylinder manifold.

MEDICAL OXYGEN PLANTS CONSIST OF:

Products comply with ISO 7396-1:2016 requirements

O 2

capacity l/min

730220730230730240730250730260730270730280730290730300730310730320730330730340730350

MDGS2/40MDGS2/60MDGS2/80MDGS2/110MDGS2/145MDGS2/210MDGS2/265MDGS2/315MDGS2/380MDGS2/475MDGS2/600MDGS2/760MDGS2/950MDGS2/1210

2x402x602x802x1102x1452x2102x2652x3152x3802x4752x6002x7602x9502x1210

PRIMARY and SECONDARY SOURCES RESERVESOURCE

www.medtech-lt.ru

MEDICAL OXYGEN GENERATING PLANTS M-DGS1

Oxygen Plant model

Code

l Oxygen concentration produced by oxygen plant in percentage 94±2%;

l Oxygen values: taken from European Pharmocopoein;

l Oxygen supply pressure to the consumer, 4,5÷5,0 bar;

3 l Oxygen efficiency ≤ 1,4 kW/m ;

l 1 Screw type compressor;

l 1 Refrigerating Air Dryer;

l 2 Air Receivers;

l 1 Cole Towers with filters;

l 1 Oxygen concentrator;

l 1 high pressure oxygen compressor;

l Oxygen control filtration unit;

l Reserve oxygen cylinder manifold.

MEDICAL OXYGEN PLANTS CONSIST OF:

O 2

capacity l/min

740220740230740240740250740260740270740280740290740300740310740320740330740340740350

MDGS1/40MDGS1/60MDGS1/80MDGS1/110MDGS1/145MDGS1/210MDGS1/265MDGS1/315MDGS1/380MDGS1/475MDGS1/600MDGS1/760MDGS1/950MDGS1/1210

4060801101452102653153804756007609501210

Products comply with ISO 7396-1:2016 requirements

Recommended