Management of Design Changes through the Product Lifecycle
1 Introduction
The International Pharmaceutical Aerosol Consortium on Regulation and Science (IPAC-RS) is an
international association of innovator and generic companies that develop, manufacture or market
orally inhaled and nasal drug products for local and systemic treatment of a variety of debilitating
diseases such as asthma, chronic obstructive pulmonary disease and diabetes.
IPAC-RS member companies are primarily involved in the development of “Combination Products”,
so are very interested in medical device design guidance and the impact upon “Combination
Products”. We are committed to advancing consensus-based, scientifically driven standards and
regulations for these products, with the purpose of facilitating the availability of high-quality, safe,
and efficacious drug products to patients.
The current members of IPAC-RS are: 3M, AstraZeneca, Boehringer Ingelheim, Catalent, Chiesi,
GlaxoSmithKline, MannKind Corporation, MAP Pharmaceuticals, Merck & Co., Inc., Mylan, Novartis,
Pfizer, Sunovion, Teva, and Vectura Ltd. Aptar Pharma, Medspray, Rexam, SHL, and West are
supplier members.
IPAC-RS has several working groups. The objective of the IPAC-RS Device Working Group is to
understand and promote best practices for orally inhaled and nasal drug product device design. In
response to concerns from IPAC-RS member companies about the consistency of the approach
adopted for the management of device changes through the product lifecycle, IPAC-RS established a
Executive Summary
Situation
There is significant variability as to how companies manage device design changes. This is supported
by the survey conducted by IPAC-RS.
Target
Regulators, member companies and therefore patients would benefit significantly if industry
adopted a common framework for the management of device changes.
Proposal
It is proposed to adopt a common framework for the management of device changes. This
framework is based upon a risk based approach to product development as defined in ICH Q8, Q9
and Q10. It is evident that given the scope of potential changes for an undefined number of devices
that it is impossible to develop definitive guidance. The purpose of the guidance defined is to be
indicative of what sponsoring companies should consider.
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
2
sub-group to review the consistency across industry of the management of device changes
throughout the product lifecycle.
The group determined that the most effective method to establish and validate the consistency, or
lack thereof, of management approaches across industry was to conduct a survey to establish a
baseline of current practices for a range of typical scenarios that could be encountered throughout
the lifecycle of a “Combination Product”.
The objective of the Device Survey was to assess current attitudes toward device changes in order
to:
• Establish a view on the ‘as is’ situation in relation to device changes,
• Facilitate a move towards consensus on the appropriateness of in-vivo and in-vitro testing,
• Highlight areas where regulatory requirements may differ from what is perceived to be
technically required.
A full description of the methodology adopted and the results of the survey are presented in
Appendix 1.
It is evident from the results that the hypothesis as to the lack of consistency in how members
companies both manage the change process and how they notify regulatory authorities of such
changes is validated. It was further concluded that Risk Management, as defined by ICH, does not
appear to be informing decision making. The survey supports the premise that the development of
guidance utilising a risk based management approach to evaluate and manage device changes for
OINDP could be of significant value to all stakeholders (i.e. patients, regulators & industry).
The remainder of this document describes an approach to realise the previously stated objective.
2 Scope
All “Combination Products” post start of Pivotal Clinical Studies, pre and post market authorisation,
are within the scope of this proposal.
There is no distinction made between Innovator and Generic Drug Products. For both types of
products, developers must demonstrate that as the product evolves through the product lifecycle,
the link between the “Combination Product” tested in the clinical setting, for which market
authorisation was granted, or is intended, must assure that the safety, quality and efficacy of the
drug product is not changed.
Formulation, Primary Pack and process changes, other than those related to the drug delivery
system, are out of scope (see Section 4.5).
It is important to note that the proposed guidance relates to product lifecycle management of
device changes. The development of generic substitutable or product line extensions is out of scope.
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
3
3 A Risk Based Approach to Design Changes through the Product
Lifecycle
The risk based approach proposed for the management of device changes through the product
lifecycle is based upon ICH Q8, Q9 & Q10.
ICH Q9 states “Fundamental to this process is the concept of quality risk management.
It is important to understand that product quality should be maintained throughout the product
lifecycle such that the attributes that are important to the quality of the drug (medicinal) product
remain consistent with those used in the clinical studies. An effective quality risk management
approach can further ensure the high quality of the drug (medicinal) product to the patient by
providing a proactive means to identify and control potential quality issues during development and
manufacturing. Additionally, use of quality risk management can improve the decision making if a
quality problem arises. Effective quality risk management can facilitate better and more informed
decisions, can provide regulators with greater assurance of a company’s ability to deal with potential
risks and can beneficially affect the extent and level of direct regulatory oversight.
Two primary principles of quality risk management are:
• The evaluation of the risk to quality should be based on scientific knowledge and ultimately
link to the protection of the patient.
• The level of effort, formality and documentation of the quality risk management process
should be commensurate with the level of risk.
3.1 Definitions
Combination Product – The combination of a device and a drug product.
Critical Quality Attribute – “A physical, chemical, biological or microbiological property or
characteristic that should be within an appropriate limit, range, or distribution to ensure the desired
product quality.”
• Drug Product Critical Quality Attribute (DP CQA) – An attribute of the drug product that has
a consequence for the safety and/or efficacy of the drug product e.g. emitted dose, fine
particle mass, assay, impurities. These attributes typically define the release specification of
a Drug Product.
• Device Critical Quality Attribute (Device CQA) – An attribute of the device that directly
impacts a Drug Product Quality Critical Attribute (e.g. metering of the formulated drug
product, presentation to the airflow path in a reproducible and reliable manner will directly
impact emitted dose) – please note that this is not a formal ICH definition but is drafted to
align with the principles of the ICH definition with respect to safety and efficacy.
Critical Process Parameter (CPP) – “A process parameter whose variability has an impact on a
critical quality attribute and therefore should be monitored or controlled to ensure the process
produces the desired quality.”
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
4
• Drug Product Critical Process Parameter (DP CPP) - A unit manufacturing operation (e.g.
filling) or a single manufacturing step/process parameter of a unit manufacturing operation
(e.g. sealing dwell temperature) that has the potential to directly impact a Drug Product
Quality Critical Attribute
• Device Quality Critical Process Parameter (Device QCPP) - A unit manufacturing operation
(e.g. injection moulding) or a single manufacturing step/process parameter of a unit
manufacturing operation (e.g. ultrasonic weld energy) that has the potential to directly
impact a Device Critical Quality Attribute.
3.2 Drug Product Critical Quality Attributes of Common OINDPs
The next section defines the typical Drug Product Quality Critical Attributes for an orally inhaled and
nasal drug product irrespective of whether the medicament is for topical or systemic drug delivery.
The oral inhaler and nasal spray were selected as they are the two primary combination products
that IPAC-RS design, develop and manufacture.
As per the definitions (see Section 4.1) these are the attributes of the drug product that determine
the safety and/or efficacy of the medicament.
Table 1 Typical Drug Product Critical Quality Attributes for OINDPs
Critical Quality Attribute
Oral Inhaler Nasal Spray
Assay � �
Impurities � �
Emitted Dose � �
Emitted Dose Uniformity � �
Aerodynamic Particle Size Distribution
(APSD) or Droplet Size Distribution
� ×
Droplet Size Distribution or Particle Size
Distribution
× �
Airflow Resistance/Resitivity � ×
Plume Geometry � �
Extractables � �
Leachables � �
Foreign Particulate Matter � ×
Microbiological Quality � �
It is the responsibility of a sponsoring company to define and justify the Drug Product CQAs for their
Combination Product.
It is important to consider whether demonstration of the equivalence of the Drug Product CQAs
following a design change is sufficient to demonstrate that the drug product is equivalent in respect
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
5
of clinical safety and efficacy. The design change must also be considered in terms of whether the
change will impact upon;
• How the target patient population may interact with the drug product to assure delivery of a
Drug Product CQA (e.g. airflow resistance of the oral inhaler is increased, patient inspiratory
effort is increased such that a subset of the target patient population can no longer
aerosolise a passive oral inhaler).
• Whether the design change could elicit as physiological change that would not be detected
by standard in-vitro test methods (i.e. Quality Control Test Method) as described by the
Drug Product CQAs described in Table 1 (e.g. a change to the method of actuation of a nasal
spray product causes the soft palate to close, a change to the external mouthpiece form of a
oral inhaler changes the patients oropharyngealeal geometry).
In addition, to the standard in vitro Quality Control methods it may be appropriate to utilise more
discriminating methodologies associated with lung cast and nasal cast models in association with
inhalation manoeuvres that are more representative of the target patient populations. These
approaches should be considered in the context of the design change proposed and their potential
to provide more discriminating data.
4 Device Changes Decision Tree
The diagram provided below describes the decision process flow that defines what, if any, testing
may be conducted to verify any change through the drug product lifecycle – see Scope Section 2.
It is evident that given the scope of potential changes and the range of different devices that it is
impossible to develop definitive guidance. The purpose of the guidance is to be indicative of what
sponsoring companies should consider.
It is acceptable to conduct testing to validate multiple changes (see Section 4.3 & 4.4). However, it
must be recognised that such an approach increases the risk that non-equivalence between design
revisions will be determined and that the source of the non-equivalence will not be immediately
evident.
It is also possible, but less likely, that two changes may offset their impact upon a Drug Product
Critical Quality Attribute and further change at a later date may result in non-equivalence and the
source of the issue will be harder to define.
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
6
Figure 1: Managing Device Change: A Decision Tree
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
7
It is the responsibility of the sponsoring company to determine the design and type of clinicval
bridging study required.
4.1 In Vitro Test Protocol
When considering device changes the key objective must be to demonstrate equivalence of the drug
product post change with reference to those DPCQAs that have the potential to be impacted (i.e.
emitted dose, emitted dose uniformity, APSD Profile and fine particle mass uniformity).
Similarly, changes to Device Critical Quality Attributes, Drug Product Quality Critical Process
Parameters and the Device Critical Quality Process Parameters that have the potential to impact a
Drug Product Critical Quality Attribute may be assessed using the same test methodology.
Key consideration must be given to whether the methods being utilised to assess the potential
impact upon a Drug Product CQA are sufficiently differentiating cognisant of:
• how the target patient population may interact with the drug product to assure delivery of a
Drug Product CQA ; and
• whether the design change could elicit a physiological change that would not be detected by
standard in-vitro test methods (i.e. Quality Control Test Methods).
The test protocol should use a risk based assessment to determine:
a) The Drug Product Critical Quality Attribute(s) that has potential to be impacted
b) The Drug Product Critical Quality Attribute(s) that are not impacted
c) The rationale for (a) and (b). Consideration must be given to any changes in Drug Product
Critical Process Parameters, Device Critical Quality Attributes and/or Device Critical Process
Parameters and their potential to impact a Drug Product Critical Quality Attribute.
d) What, if any, supporting evidence exists that can be used to understand potential for the
proposed change to impact the Drug Product Critical Quality Attribute(s). This supporting
evidence may be data generated by the sponsoring company or may be information in the
public domain (i.e. Scientific Literature, IPAC-RS Publications et.).
e) A pre-defined test protocol, based upon the potential risk and acceptance criteria for each
of the impacted Drug Product Critical Quality Attributes, that provide assurance that the
safety and efficacy of the Drug Product has not been compromised. These protocols should
be sufficiently powered to evaluate the impact of any change using an appropriate
statistical method.
4.1.1 Stability Assessment Requirements
Stability testing, including an assessment of assay and impurities, should only be considered
necessary if:
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
8
• the primary pack is an integral part of the drug delivery system e.g. reservoir Dry Powder
Inhaler, Nasal Spray Pump;
• the change proposed has the potential to alter how the formulation is protected from water,
light, oxygen etc.; and/or
• the component(s) to be revised are in intimate contact with the formulation for more than
the transient time to deliver a dose.
For changes where the primary pack is integral to the drug delivery system it is recommended that
long term stability evaluation should be conducted. Changes could be implemented on the basis of
shorter term predictive accelerated stability studies but a commitment to longer term studies would
be required.
For all other instances it may be appropriate to conduct accelerated predictive stability studies that
assess impact upon specific Drug Product Critical Quality Attributes.
A protocol defining the tests to be performed, a rationale for including the tests defined (the
rationale should consider how the change impacts upon the ability of the drug delivery system to
protect, meter and dispense the medicament) and acceptance criteria, that assure the
safety/efficacy of the drug product has not been compromised, should be pre-approved by the
sponsoring company before adoption of the design revision.
4.1.2 Mechanical Function & Mechanical Safety Assessment
The Mechanical Function and Mechanical Safety Assessment should evaluate attributes of the
design, based upon a risk assessment, that could be compromised by the inclusion of the change
proposed. For example, replacement of a type of polymer for a component under load, may cause
the part to fail on long term storage or at extremes of the design as described by the variance in the
detailed engineering drawings.
This assessment should also be cross-referenced with the Design FMEA to ensure that the integrity
of the individual components that may be subject to change and the overall product assembly are
not compromised (the Risk Priority Number is not higher than the originally acceptable limits).
A protocol defining the tests to be performed, a rationale for including the tests defined and
acceptance criteria, that assures that the safety and efficacy of the drug product has not been
compromised, should be pre-approved before adoption of the design revision.
4.2 Patient Handling Studies
Any Patient Handling/Human Factors Study must be designed to consider:
• What is the design change(s)?
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
9
• What, if any subset, of the patient populations ability to use the design maybe
compromised?
It must be recognised that these studies are primarily qualitative and as such are not statistically
powered to elucidate whether the revision of a design issue will impact usability of the device. It is
highly recommended that sponsoring companies continuously monitor patient feedback from all
clinical studies.
4.3 In Vivo Test Protocol Definition
For orally inhaled and/or nasal drug products for delivery of medicaments by systemic absorption it
may be appropriate and clinically relevant to demonstrate equivalence by means of a
pharmacokinetic study.
Systemic pharmacokinetic data have a well-established role in the bioequivalence evaluation of
orally administered systemically acting drugs but there is doubt about its applicability to inhaled
topically acting drugs. Whereas pharmacokinetic data clearly have a role is establishing
bioequivalence in terms of systemic safety, there is a need to better understand the relevance to
efficacy. Therefore, it may only be possible to demonstrate that the inhaled topical medicaments
have the same safety/efficacy by conducting both a pharmacokinetic and pharmacodynamic study.
Consideration should be given to power the study appropriately to demonstrate the study
objectives.
5 Regulatory Notification
It is evident from the survey that sponsoring companies are unclear of how to inform the regulatory
authorities of design changes through the product lifecycle. The proposal outlined in Section 4.3
defines a series of key decision points that drives sponsoring companies to determine whether they
need to undertake programmes of work to support the device changes that they would propose.
The proposal for the mechanism whereby sponsoring companies notify the regulatory authorities
uses the same framework – see Table 2
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS). All Rights Reserved.
10
Table 2 Risk Based
Framework for Notifying Regulators
Potential to impact
how the target
population and/or
patient/HCP may
interact with the
device
Potential to elicit a
change in
physiological response
that would not be
detected by standard
in-vitro methods
Potential to impact a
Drug Product CQA,
Device CQA etc.
Design Phase
Post Phase IIb - Approval
Post Approval
x x x Update IND/CTA, if Registered Detail is impacted,
when transitions from Phase IIB to Phase III
Recorded in NDA/ IMPD
US: Annual update
EU: Type IAIN change
x
x
�
Update IND/CTA, if Registered Detail is impacted,
when the new design is studied in the clinic
Recorded in NDA/ CTA
US: CBE 30
EU: Type IA or IB change
x
x
�
� �
�
�
�
x
�
x
�
x
�
x
x
�
�
Updated IND/CTA, if the Registered Detail is
impacted, when the new design is studied in the
clinic
Recorded in NDA
US: PAS
EU: Type II EU change
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
11
6 Conclusion
The proposal presented has been developed in the context of a Device Survey conducted by IPAC-RS
that clearly demonstrates a lack of uniformity in approach. It is evident that all parties involved in
the design, development and product lifecycle of “Combination Products” (i.e. patients, sponsoring
companies and regulators) would benefit significantly from the development of guidance such that
common standards for the management of change through the product lifecycle can realised.
The approach outlined in this proposal has been developed cognisant of the concept of a Risk-Based
Approach to Product Development as described in ICH Q8, Q9 and Q10. The proposal defines a risk
based approach that enables device changes to be implemented post start of pivotal clinical through
the product lifecycle on the basis of a supporting data package that is targeted at assuring that the
safety/efficacy of the drug product is not compromised.
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
12
Appendix 1:
IPAC RS – Product Lifecycle Device Change Management Survey
Methodology
The IPAC-RS Product Lifecycle Device Change Management Survey presented 15 scenarios that
related to a series of common types of changes. These included material of construction changes,
introduction/replacement of a supplier, scale up through the industrialisation phase of a programme
and design revisions that member companies may determine to implement for a variety of reasons.
Respondents were provided with information about each of these scenarios in turn and asked to
indicate the following for each scenario:
• what activities they would undertake in support of the change proposed cognisant of the
phase of the development process. These activities are classified as either being non clinical
or clinical.
• whether their decision to conduct these activities would be driven from a technical and/or a
regulatory perspective.
• how they would notify the agency of any change.
Non-clinical activities that could be selected by respondents included:
• mechanical verification,
• device robustness,
• physiochemical parameters,
• dimensional measurements,
• filling line trials,
• emitted dose,
• dose content uniformity,
• Aerodynamic Particle Size Distribution – ACI or NGI,
• Aerodynamic Particle Size Distribution – Lung Cast Model,
• Pack Integrity,
• Spray Pattern/Plume Geometry,
• extractable/leachables,
• other.
Clinical testing activities that could be selected included:
• user handling study,
• design validation,
• pharmacokinetic bioequivalence study,
• pharmacodynamic bioequivalence study,
• clinical efficacy,
• clinical safety study,
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
13
• flow profile measurement,
• other.
The survey was open for completion from May 2010 through January 2011. The preliminary results
are based on responses obtained by September 2010; at that point, 125 responses had been
obtained.
Figure 1 presents a summary of the demographics of the respondents to the survey
Results
The results showing how a sponsoring company would respond to each of the scenarios are
provided in the following format;
• Two graphs are presented for each scenario, one graph summarizes results for non-clinical
activities and one summarizes results for clinical activities.
• Each bar represents the response to a particular test. The abbreviated test name is given
below the bar.
• The height of the bar gives the numbers of respondents that would do the test as a
percentage of the total answering the question.
• The blue, orange and red portions represent the proportion that would perform the test for
technical only, regulatory only or technical & regulatory reasons respectively.
Abbreviations used are as follows:
Non-Clinical Test Abbreviations Clinical Test Abbreviations
MechVer Mechanical variation UserH User Handling Study
DevRob Device robustness DevVal Device Validation (Function) Study
Phy&Dim Physiochemical parameters and
dimensional measurements
PK BE Pharmacokinetic Bioequivalence Study
Filling Filling Line Trials PD BE Pharmacodynamic Bioequivalence
Study
EmMass Emitted Mass (Shot Weight) ClinEff Clinical Efficacy Study
DCU Dose Content Uniformity ClinSaf Clinical Safety Study
APSD Aerodynamic Particle Size Distribution
– ACI or NGI
FlowProf Flow Profile Measurement
LungCast Aerodynamic Particle Size Distribution Other Other – to be specified
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
14
– Lung Cast Model
PackInt Package integrity related tests
Spray Spray Pattern/ Plume Geometry (MDIs)
EXs Extractables characterization
LEs Leachables characterization
Other Other – to be specified
Complete Results of IPAC-RS
Product Lifecycle Device Change Management Survey
Scenario 1 DPI operating button change
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
15
Scenario 2 Addition of a mdi dose counter
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
16
Scenario 3 DPI mouthpiece change
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
17
Scenario 4 Capsule to multi-dose DPI
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
18
Scenario 5 DPI components change for automisation
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
19
Scenario 6 Base elastomer manufacturing change
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
20
Scenario 7 Polymer additive source/supplier change
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
21
Scenario 8 Mouthpiece material change
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
22
Scenario 9 Ink change
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
23
Scenario 10 Primary Pack manufacturing plant relocation
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
24
Scenario 11 MDI valve & filling process change
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
25
Scenario 12 New material conditioning process
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
26
Scenario 13 Moulding process change
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
27
Scenario 14 Injection mould scale-up
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
28
Scenario 15 Ultrasonic welding process change
Copyright © 2013 International Pharmaceutical Aerosol Consortium on Regulation & Science (IPAC-RS).
All Rights Reserved.
29
Summary of Results
The following general observations can be made from a review of the responses to all scenarios:
• Clinical testing was almost as common as in-vitro testing for design changes, while for
material & process changes, in-vitro testing was much more common.
• Several respondents would use long-term stability testing (>6 months) at 40 C/ 75% RH to
support changes. It was unclear whether companies would conduct short term accelerated
stability studies to implement a change and commit to longer stability studies in parallel.
• Mechanical testing, device robustness, and dimensions were seen by many as key elements
for most changes.
• Shot weight, DDU, ACI and spray pattern were often done regardless of whether there was a
perceived technical need.
• The lung cast model and flow profile measurement were not used often.
• Extractable and Leachable testing were primarily conducted for material and process
changes, but frequently also conducted for pure design changes.
• Device validation and user handling studies were the most common clinical studies.
• About half of respondents would do clinical work purely for regulatory reasons
The survey revealed considerable variance regarding: (1) whether to test in the context of a device
change; and (2) if so, what types of tests should be conducted; as well as (3) the rationale for
testing. There was also uncertainty about the regulatory pathway that should be taken to obtain
approval of a proposed change.
It was concluded from the survey that there was a lack of consistency in the approach being
adopted. It was evident that Risk Management, as defined by ICH, does not appear to be informing
decision making. It was unclear whether the driver of the inconsistencies noted was that
respondents answered based on what they would do or what they believed should be done.
Additional work could be conducted to understand the inconsistency in approach (e.g. a sub-
analysis to determine whether inconsistency was driven by region, background or experience of the
respondents). However, it was concluded that given the variability observed that this would be
unlikely to affect the overall conclusion.
The survey suggests that the development of an agreed framework for guidance utilizing a risk-
management based approach to evaluate and manage device changes for OINDP could be of a
significant value to all stakeholders (i.e. patients regulators & industry).
Recommended