iSite 3.5: System Preferences
CONFIDENTIAL 2TRN-0035-05 Rev 04/19/06
Learning Objectives
• General Preferences
• Window Window/Center
• Image Processing
• Filters
• Annotations
• DICOM Sources
• Screen Overlay
• Paper Printing
Understand the purpose and configuration of:
CONFIDENTIAL 3TRN-0035-05 Rev 04/19/06
Learning Objectives (cont.)
• iExport
• iQuery
• Organization Dictionary
• Location Dictionary
• Plug Ins
• DICOM Print
• Body Part Dictionary (iSite Radiology only)
• Hanging Protocols (iSite Radiology only)
CONFIDENTIAL 4TRN-0035-05 Rev 04/19/06
Table of Contents
• General Preferences
• Window Window/Center
• Image Processing
• Filters
• Annotations
• DICOM Sources
• Screen Overlay
• Paper Printing
CONFIDENTIAL 5TRN-0035-05 Rev 04/19/06
Table of Contents (cont.)
• iExport
• iQuery
• Organization Dictionary
• Location Dictionary
• Plug Ins
• DICOM Print
• Body Part Dictionary (iSite Radiology only)
• Hanging Protocols (iSite Radiology only)
CONFIDENTIAL 6TRN-0035-05 Rev 04/19/06
Preferences Dialog
CONFIDENTIAL 7TRN-0035-05 Rev 04/19/06
System Preferences
• System Preferences configure system-wide settings for every user
• System and Machine Preferences are associated with the Security Codes ISTUSRPREF and ISTSUPPORT
• System Preferences are typically only accessible by administrators and are not visible to standard users
CONFIDENTIAL 8TRN-0035-05 Rev 04/19/06
Importing and Exporting Preferences
• Importing and Exporting Preferences
• Many preference settings can be imported and exported to facilitate setting up User and System Preferences
• The User Preferences that can be imported and exported are Window Width/Center and Filters
• The System Preferences that can be imported and exported are Window Width/Center, Filters, Annotation, DICOM Sources, Screen Overlays, Paper Printing, Organization Dictionary, Location Dictionary and Body Part Dictionary
• Please note that importing preferences will override any preferences in the system
CONFIDENTIAL 9TRN-0035-05 Rev 04/19/06
Preferences Dialog
• Preferences can be accessed by selecting the ‘P’ icon in the upper-right corner of the iSite Control Strip
CONFIDENTIAL 10TRN-0035-05 Rev 04/19/06
General Preferences
CONFIDENTIAL 11TRN-0035-05 Rev 04/19/06
Logon Welcome Message
• The ‘Logon welcome message’ may be configured with help desk information or may be used to notify users of scheduled maintenance of the system
CONFIDENTIAL 12TRN-0035-05 Rev 04/19/06
Logon Welcome Message
• The ‘Logon welcome message’ shown at left is displayed on the log-in page
CONFIDENTIAL 13TRN-0035-05 Rev 04/19/06
Image/Series Linking
• iSite System Administrators can set the number of degrees to allow for linking variance
• This determines if a series can be linked with another series and this variance is configurable at a site level
CONFIDENTIAL 14TRN-0035-05 Rev 04/19/06
Allow DICOM Query From Exam Timeline
• The ‘Allow DICOM Query From Exam Timeline’ option permits users to perform a DICOM query from the Exam Timeline if the user has the Security Code ISTQUERY
CONFIDENTIAL 15TRN-0035-05 Rev 04/19/06
Allow Exams to be Marked Read More Than Once
• The ‘Allow Exams to be Marked Read More Than Once’ option permits users to “Mark Read” exams which have already been marked as read
• The ‘Allow Exams to be Marked Read More Than Once’ is useful for over reads and peer review
CONFIDENTIAL 16TRN-0035-05 Rev 04/19/06
Image Orientation
• The ‘Image Orientation’ option allows iSite System Administrators to allow users to set their own flipping preferences for Axial CTs or always display Axial and Oblique CTs with the anatomical right on the left side of the image window
CONFIDENTIAL 17TRN-0035-05 Rev 04/19/06
Image Orientation
• Please note that the visibility of this option is dependent on the System Preferences set by your institution's iSite System Administrator
CONFIDENTIAL 18TRN-0035-05 Rev 04/19/06
Image Orientation
• Users who wish to automatically display axial and oblique CTs with the anatomical right displayed on the left, regardless of the initial orientation of the image, may select the “Axial/Oblique CT:…" box
CONFIDENTIAL 19TRN-0035-05 Rev 04/19/06
Window Width/Center
CONFIDENTIAL 20TRN-0035-05 Rev 04/19/06
Window Width/Center
• Users may select the Window Width and Center dialog to add, modify or delete the default preset window width and center values display
CONFIDENTIAL 21TRN-0035-05 Rev 04/19/06
Window Width/Center
• The Add Window Width/Center may be used to create new Window Width/Center presets by establishing a name, indicating the Modality and setting the Width and Center
CONFIDENTIAL 22TRN-0035-05 Rev 04/19/06
Window Width/Center
• To edit a window width and center setting, “double click” a listing or highlight a listing and click the ‘Properties’ button
• When you have finished making changes, click ‘Apply’ and then ‘OK’ to save and exit the Dialog box
CONFIDENTIAL 23TRN-0035-05 Rev 04/19/06
Window Width/Center
• Window Width/Center settings are mapped to the number keys in the order displayed here and in the Context Menu, so that users may apply a Window Width/Center setting easily while viewing images
CONFIDENTIAL 24TRN-0035-05 Rev 04/19/06
Image Processing
CONFIDENTIAL 25TRN-0035-05 Rev 04/19/06
Image Processing
• Edge Enhance: Uses a Laplacian method to increase the visibility of edges in the image
• Edge Detection: Provides an alternative to the Edge Enhance tool, using a Sobel operator to increase the visibility of edges
• Median Filtering: Uses a common filtering method to reduce noise in an image
• CLAHE (Contrast Limited Adaptive Histogram Equalization): Uses an adaptive form of histogram equalization that enhances the contrast adaptively across the image
CONFIDENTIAL 26TRN-0035-05 Rev 04/19/06
Image Processing
• Image Processing Filters have several configurable parameters which may be configured by each individual user
CONFIDENTIAL 27TRN-0035-05 Rev 04/19/06
Image Processing
• Image Processing Filters can be applied manually (via a right mouse click option)
• Any time an Image Processing Filter is applied to an image, iSite displays a “P ” (Processed) Icon in the meta data area of that image
CONFIDENTIAL 28TRN-0035-05 Rev 04/19/06
Image Processing – Edge Enhance
• ‘Edge Enhance’ is performed with a Laplacian operator to discriminate the edges; the result is scaled and added back into the original image data
• A 5 point or 9 point operator may be selected
CONFIDENTIAL 29TRN-0035-05 Rev 04/19/06
Image Processing – Edge Enhance
• The intensity of ‘Edge Enhance’ is controlled by the “degree” parameter which can be “Light”, “Medium” or “Strong”
• Edge Enhance is applicable to any monochrome image
CONFIDENTIAL 30TRN-0035-05 Rev 04/19/06
Image Processing – Edge Enhance
The image on the right side has the ‘Edge Enhance’ filter applied
CONFIDENTIAL 31TRN-0035-05 Rev 04/19/06
Image Processing – Edge Detect
• This class of algorithm accentuates the rising and falling components of edges
• This is done with a Sobel operator; the result is scaled and added back into the original image data
CONFIDENTIAL 32TRN-0035-05 Rev 04/19/06
Image Processing – Edge Detect
• The intensity of ‘Edge Detect’ is controlled by the “degree” parameter which can be “Light”, “Medium” or “Strong”
CONFIDENTIAL 33TRN-0035-05 Rev 04/19/06
Image Processing – Edge Detect
The image on the right side has the ‘Edge Detect’ filter applied
CONFIDENTIAL 34TRN-0035-05 Rev 04/19/06
Image Processing – Median Filter
• ‘Median Filter’ is used as a noise reduction filter
• Neighborhoods around each pixel are sampled and the central pixel is replaced with the median value of the neighborhood
• A 3x3 or 5x5 filter may be selected
CONFIDENTIAL 35TRN-0035-05 Rev 04/19/06
Image Processing – Median Filter
The image on the right side has the ‘Median Filter’ applied
CONFIDENTIAL 36TRN-0035-05 Rev 04/19/06
Image Processing - CLAHE
• Contrast Limited Adaptive Histogram Equalization (CLAHE) is a variant of histogram equalization
• The drawback of the general histogram equalization method is that the computation is performed across the entire image
CONFIDENTIAL 37TRN-0035-05 Rev 04/19/06
Image Processing - CLAHE
• In medical images, the local behavior of the histogram can be radically different than either the global, or non-local behavior
• This class of method performs the histogram equalization on a parameterized region-by-region basis and stitches the results to prevent any boundary edges
CONFIDENTIAL 38TRN-0035-05 Rev 04/19/06
Image Processing - CLAHE
• The following parameters can be set via the “Preferences”
• Contextual Region Dimension
• Number Of Bins
• Clip Limit
• Map Level
CONFIDENTIAL 39TRN-0035-05 Rev 04/19/06
Image Processing – CLAHE
The image on the right side has ‘CLAHE’ applied
CONFIDENTIAL 40TRN-0035-05 Rev 04/19/06
Filters
CONFIDENTIAL 41TRN-0035-05 Rev 04/19/06
Filters
• Filters allow you to create custom search criteria for Worklists
CONFIDENTIAL 42TRN-0035-05 Rev 04/19/06
Filters
• Filters can be used to create Reading Worklists to meet specific needs
• Exercise care in creating filters so exams are not accidentally excluded
• To create a filter click ‘Add’
CONFIDENTIAL 43TRN-0035-05 Rev 04/19/06
Filters• This screen allows the
user to generate a filter
• Users are not required to fill in all the fields because the default for an empty field is ‘all’
• When you have finished making changes, click ‘Apply’ and then ‘OK’ to save and exit the Dialog box
CONFIDENTIAL 44TRN-0035-05 Rev 04/19/06
Filters• The Filter shown will
provide a worklist which includes:
–“CARDIAC” subspecialty (as defined in the Exam Dictionary) –Unread–All Exam Statuses
CONFIDENTIAL 45TRN-0035-05 Rev 04/19/06
Filters• The Filter shown will
provide a worklist which includes:
–“CR” and “DX” (DR) images for patients in –“Critical Care” Units–For Body Part “Chest”–“Stat” priority–“Unread Only”–“Completed” Exam Status
CONFIDENTIAL 46TRN-0035-05 Rev 04/19/06
Filters• The Filter shown will
provide a worklist which includes:
–“CR” and “DX” (DR) images for patients in –“Outpatient” Units–For Subspecialty “BONE”–“Non-Stat” priority–All Exam Statuses, except “In Progress”–For 30 Days–Ordered by Dr.s Smith, Jones, or Davis
CONFIDENTIAL 47TRN-0035-05 Rev 04/19/06
Annotations
CONFIDENTIAL 48TRN-0035-05 Rev 04/19/06
Annotations
• Image annotations may be added to an exam to identify key images or regions of interest
• The ‘Annotations’ option allows the iSite System Administrator to create, edit and delete preset annotations
CONFIDENTIAL 49TRN-0035-05 Rev 04/19/06
Annotations
• Select the ‘Add’ button to add new image annotation presets
CONFIDENTIAL 50TRN-0035-05 Rev 04/19/06
Annotations
• Provide a name, which should be the annotation
• Font styles and sizes may be selected
• Arrows may be added for emphasis
CONFIDENTIAL 51TRN-0035-05 Rev 04/19/06
Annotations
• The name of the annotation is displayed as an annotation option on the Context Menu
CONFIDENTIAL 52TRN-0035-05 Rev 04/19/06
Annotations• The annotation may
be resized or moved within the image window
CONFIDENTIAL 53TRN-0035-05 Rev 04/19/06
DICOM Sources
CONFIDENTIAL 54TRN-0035-05 Rev 04/19/06
DICOM Sources
• DICOM Sources enable the configuration of default display settings for Window Width/Level, Series Splitting and Sorting criteria for a modality or a group of modalities with a common manufacturer or model name, etc.
CONFIDENTIAL 55TRN-0035-05 Rev 04/19/06
DICOM Sources
• DICOM Sources also allows you to stack RF or XA images
CONFIDENTIAL 56TRN-0035-05 Rev 04/19/06
DICOM Sources
• When adding a DICOM source, the iSite System Administrator should be as specific as possible to prevent “overlapping” DICOM Sources
CONFIDENTIAL 57TRN-0035-05 Rev 04/19/06
DICOM Sources
• A warning message is provided in the event of duplicate DICOM Sources to prevent “overlapping” DICOM Sources names or “overlapping” DICOM Sources configurations
CONFIDENTIAL 58TRN-0035-05 Rev 04/19/06
DICOM Sources
• If two or more DICOM sources are configured with an equal number of matching identifiers, then the DICOM source containing an Exam Code will take precedence
CONFIDENTIAL 59TRN-0035-05 Rev 04/19/06
DICOM Sources
• In the Window/Center tab, choose which values to use for this source
CONFIDENTIAL 60TRN-0035-05 Rev 04/19/06
DICOM Sources
• Histogram Calculation
– This is the default selection in the ‘Add DICOM Source’ dialog
– The histogram is a graph showing the total tonal distribution in the image
– The histogram maps Luminance, which is defined from the way human eye perceives the brightness of different colors
– Every pixel in the color or gray image computes to a luminance value between 0 and 255. The histogram uses bins to graph the pixel count of every possible value of luminance or brightness. The total tonal range of a pixel’s 8 bit tone value is 0 to 255, where the 0 is the blackest at the left end and 255 is the whitest at the right end
CONFIDENTIAL 61TRN-0035-05 Rev 04/19/06
DICOM Sources
• DICOM Values
– DICOM Values are the Window Center and Window Width values provided by the modality
– DICOM Sources utilizing DICOM Values should be configured for a specific AE Title to ensure that images from other devices are not impacted
– The DICOM modality values are stored in Window Center (0028,1050) and Window Width (0028,1051) DICOM attributes
CONFIDENTIAL 62TRN-0035-05 Rev 04/19/06
DICOM Sources
• The Series Splitting option enables iSite System Administrators to specify various rules for splitting series
• For example creating separate image windows for monochrome images versus color images
CONFIDENTIAL 63TRN-0035-05 Rev 04/19/06
DICOM Sources
• The Sorting tab allows the user to specify the sorting criteria for series images
• After selecting the details for the DICOM source, click ‘OK’ and then ‘Apply’ or ‘OK’ when you return to the Preferences Dialog in order for the changes to be preserved
CONFIDENTIAL 64TRN-0035-05 Rev 04/19/06
Screen Overlays
CONFIDENTIAL 65TRN-0035-05 Rev 04/19/06
Screen Overlays
• iSite System Administrators can configure the information overlays for various modalities
• There are options to specify where various patient and exam information will be displayed on the image
CONFIDENTIAL 66TRN-0035-05 Rev 04/19/06
Screen Overlays
• To add an item, click, drag and drop the item into the screen at left
• Items can be removed by dragging them out of the overlay window
CONFIDENTIAL 67TRN-0035-05 Rev 04/19/06
Screen Overlays
• When entering a name for the new overlay, note that this name must correspond to a Modality Type
CONFIDENTIAL 68TRN-0035-05 Rev 04/19/06
Screen Overlays
• Image Number and Series Number may be added utilizing the method demonstrated in the accompanying image
CONFIDENTIAL 69TRN-0035-05 Rev 04/19/06
Screen Overlays
• Window Width and Window Level may be displayed together using the method demonstrated in the accompanying image
CONFIDENTIAL 70TRN-0035-05 Rev 04/19/06
Paper Printing
CONFIDENTIAL 71TRN-0035-05 Rev 04/19/06
Paper Printing
• iSite allows you to configure the way in which exams and images are printed
• iSite System Administrators can create print configurations that users can select to specify various options
CONFIDENTIAL 72TRN-0035-05 Rev 04/19/06
Paper Printing
• For assistance in configuring Paper Printing, please refer to the user manual or contact Customer Care
CONFIDENTIAL 73TRN-0035-05 Rev 04/19/06
iExport
CONFIDENTIAL 74TRN-0035-05 Rev 04/19/06
iExport
• iExport is the mechanism by which iSite exports exams in DICOM format to another DICOM device, such as a mass CD Burner or 3D imaging system
CONFIDENTIAL 75TRN-0035-05 Rev 04/19/06
iExport
• From this page, DICOM forwarding addresses can be added and/or edited
• iExport destinations will be visible in the iExport dialog, which can be accessed via the popup menu in the worklists or from the exam timelines
CONFIDENTIAL 76TRN-0035-05 Rev 04/19/06
iExport
• The AE Title, Host IP and Port of the remote device must be obtained from the device vendor
• The Query/Retrieve service must be supported by the device and supported values must be obtained from the device vendor
CONFIDENTIAL 77TRN-0035-05 Rev 04/19/06
iQuery
CONFIDENTIAL 78TRN-0035-05 Rev 04/19/06
iQuery
• iQuery is the iSite mechanism that allows studies to be imported into the iSite Server for viewing with iSite
• From this page, archives from which to pull studies can be added, edited or deleted
CONFIDENTIAL 79TRN-0035-05 Rev 04/19/06
iQuery
• The AE Title, Host IP and Port of the remote device must be obtained from the device vendor
• The Query/Retrieve service must be supported by the device and supported values must be obtained from the device vendor
CONFIDENTIAL 80TRN-0035-05 Rev 04/19/06
Organization Dictionary
CONFIDENTIAL 81TRN-0035-05 Rev 04/19/06
Organization Dictionary
• The Organization Dictionary lets you specify the various organizations present in your institution, if you use ‘organization’ in your exam records
CONFIDENTIAL 82TRN-0035-05 Rev 04/19/06
Organization Dictionary
• The listings are then used for the purposes of exam searches
• The names in this dictionary must match the organizations listed in iSuite
CONFIDENTIAL 83TRN-0035-05 Rev 04/19/06
Location Dictionary
CONFIDENTIAL 84TRN-0035-05 Rev 04/19/06
Location Dictionary
• The Location Dictionary allows you to specify the various locations in the institution, if ‘location’ is a mapped field from external systems
• Locations must match locations mapped from external systems
CONFIDENTIAL 85TRN-0035-05 Rev 04/19/06
Location Dictionary
• The Patient Location Menu may then be utilized as search criteria on the Patient Directory
CONFIDENTIAL 86TRN-0035-05 Rev 04/19/06
Plug Ins
CONFIDENTIAL 87TRN-0035-05 Rev 04/19/06
Plug Ins
• Plug Ins (Applications Program Interface, API) provide enhanced feature sets such as advanced workflow and visualization to users
CONFIDENTIAL 88TRN-0035-05 Rev 04/19/06
Plug In Dialog
• Choose a display name for the Plug Ins folder and then set whether it will be displayed in the folder tree
• Plug Ins visible in the Folder List are visible to all users
CONFIDENTIAL 89TRN-0035-05 Rev 04/19/06
Plug In Dialog
• Enter the URL of the Plug In
• Select enable in either iSite Radiology, iSite Enterprise or both
• Please Note: Do not deselect ‘Disable API’ prior to contacting Philips iSite Care Center
CONFIDENTIAL 90TRN-0035-05 Rev 04/19/06
Plug In Dialog
• Users may require specific Security Codes to utilize the Plug In if the ‘Requires Security Code’ option is enabled
CONFIDENTIAL 91TRN-0035-05 Rev 04/19/06
DICOM Print
CONFIDENTIAL 92TRN-0035-05 Rev 04/19/06
DICOM Print
• The DICOM Print option controls the printers to which exams and images will be printed
• You will need to obtain information and assistance from the printer vendor to properly configure and validate the print settings
CONFIDENTIAL 93TRN-0035-05 Rev 04/19/06
DICOM Print
• Image Display Format, FilmSize ID and Medium Type must be supported by the printer device and supported values must be obtained from the printer vendor
• Min Density, Max Density and Printer Configuration values must be obtained from the printer vendor to ensure proper contrast resolution
CONFIDENTIAL 94TRN-0035-05 Rev 04/19/06
DICOM Print Warning: Considerations for Life-Size Printing
• For LifeSize Printing, the correct Image Box Height and Width settings must be obtained from the printer vendor
• Philips strongly recommends validating that the LifeSize print option is indeed printing life-sized images after completing the configuration
CONFIDENTIAL 95TRN-0035-05 Rev 04/19/06
DICOM Print Warning: Considerations for Life-Size Printing
• When using iSite for measurement or life-size printing, Philips recommends the use of an appropriately placed object of known size for determining the magnification factor applied to the image
CONFIDENTIAL 96TRN-0035-05 Rev 04/19/06
DICOM Print Warning: Considerations for Life-Size Printing
• For assistance in configuring iSite for DICOM Printing, please refer to the user manual or contact Customer Care
• Philips strongly recommends validating that the LifeSize print option is indeed printing life-sized images after completing the configuration
CONFIDENTIAL 97TRN-0035-05 Rev 04/19/06
Body Part Dictionary
CONFIDENTIAL 98TRN-0035-05 Rev 04/19/06
Body Part Dictionary
• iSite allows you to configure the body part dictionaries that your institution will be using in conjunction with various parts of the program
• The body part definitions can be edited and removed
CONFIDENTIAL 99TRN-0035-05 Rev 04/19/06
Body Part Dictionary
• Select ‘Body Part Dictionary’ from the ‘System Preferences’ list
• Select ‘Add’ to create a new entry in the dictionary; or
• Select ‘Properties’ to edit an existing entry
CONFIDENTIAL 100TRN-0035-05 Rev 04/19/06
• As illustrated at left, the body part “head” is equivalent to “brain,” “iac,” “orbits,” “sinus,” “skull”
• Additionally, the body parts listed in the ‘Equivalents’ frame, are equivalent to the body part “Name” as well as all other equivalents
• For example: “head” is not only
equivalent to “brain” but also equivalent to “iac”
Edit Body Part Dictionary Dialog
CONFIDENTIAL 101TRN-0035-05 Rev 04/19/06
Hanging Protocols
CONFIDENTIAL 102TRN-0035-05 Rev 04/19/06
Hanging Protocols
• The Hanging Protocols preference area allows the user to view the iSite Radiology Hanging Protocols and to delete them if necessary
CONFIDENTIAL 103TRN-0035-05 Rev 04/19/06
Hanging Protocols
• All of the Hanging Protocol property windows are ‘Read Only’ and cannot be edited from the Preferences page
• Hanging Protocols must be edited from within the Radiology canvas page
CONFIDENTIAL 104TRN-0035-05 Rev 04/19/06