8/13/2019 Infrared Spectroscopy 03
1/70
Infrared (IR) Spectroscopy
IR deals with the interaction of infrared radiation with matter.The IR spectrum of a compound can provide importantinformation about its chemical nature and molecular structure.
Most commonly, the spectrum is obtained by measuring theabsorptionof IR radiation, although infrared emission andreflection are also used.
Widely applied in the analysis of organic materials, also usefulfor polyatomic inorganic molecules and for organometalliccompounds.
8/13/2019 Infrared Spectroscopy 03
2/70
Overview
1. Electromagnetic radiation
2. Vibrations
3. Principle of IR experiment
4. IR spectrum
5. Types of vibration
6. CGF/Fingerprint regions
7. IR activity of vibrations
8. Interpretation of IR spectra
9. Instrumentation
10. Sample preparation
8/13/2019 Infrared Spectroscopy 03
3/70
Electromagnetic Radiation
The propagation of electromagneticradiation in a vacuum is constant for allregions of the spectrum (= velocity oflight):
c =
1 = 10 10m 1 nm = 10 9m 1 m = 10 6m
Another unit commonly used is the wavenumber, which is linear with energy:
Work by Einstein, Planck and Bohr indicated that electromagnetic radiation can be
regarded as a stream of particles or quanta, for which the energy is given by the
Bohr equation:
8/13/2019 Infrared Spectroscopy 03
4/70
The Electromagnetic Spectrum
8/13/2019 Infrared Spectroscopy 03
5/70
LIMIT OF RED LIGHT: 800 nm, 0.8 m, 12500 cm-1
NEAR INFRARED: 0.8 -2.5 m, 12500 - 4000 cm-1
MID INFRARED: 2.5 - 50 m, 4000 - 200 cm-1
FAR INFRARED:50 - 1000 m, 200 - 10 cm-1
Divisions arise because of different optical materials and
instrumentation.
Infrared region
8/13/2019 Infrared Spectroscopy 03
6/70
Molecular spectra
There are three basic types of optical
spectra that we can observe formolecules:1.Electronic or vibronic spectra (UV-visible-near IR)(transitions between a specific vibrational and rotational level
of one electronic state and a vibrational and rotational level ofanother electronic state)
2.Vibrational or vibrational-rotational
spectra (IR region)(transitions from the rotational levels of one vibrational levelto the rotational levels of another vibrational level in the sameelectronic state)
3.Rotational spectra (microwave region)(transitions between rotational levels of the same vibrationallevel of the same electronic state)
8/13/2019 Infrared Spectroscopy 03
7/70
Infrared radiation in the range from 10,000 100 cm 1isabsorbed and converted by an organic molecule into energy ofmolecular vibration
> this absorption is quantized:
Vibrational spectra (I): Harmonic oscillator model
A simple harmonic oscillator is a mechanical system consisting ofa point mass connected to a massless spring. The mass is underaction of a restoring force proportional to the displacement of
particle from its equilibrium position and the force constantf(alsokin followings) of the spring.
8/13/2019 Infrared Spectroscopy 03
8/70
The vibrational frequency is increasing with:
increasing force constant f = increasing bond strength
decreasing atomic mass
Example:f cc>f c=c>f c-c
The vibrational energy V(r) can be calculated using the (classical) model of the
harmonic oscillator:
Using this potential energy function in the Schrdinger equation, the vibrationalfrequency can be calculated:
8/13/2019 Infrared Spectroscopy 03
9/70
Vibrational spectra (II): Anharmonic oscillator model
The actual potential energyof vibrations fits the
parabolic function fairlywell only near theequilibrium internuclear
distance. The Morsepotential function moreclosely resembles the
potential energy ofvibrations in a molecule for
all internuclear distances-anharmonic oscillatormodel.
Fig. 12-1
8/13/2019 Infrared Spectroscopy 03
10/70
The energy difference between
the transition from n to n+1corresponds to the energy of theabsorbed light quantum
The difference between two
adjacent energy levels gets
smaller with increasing n untildissociation of the moleculeoccurs (Dissociation energy ED)
EVI B ( En+1En) =h osc
Note:
Weaker transitions called overtonesare sometimes observed. These correspond to=2 or 3, and their frequencies are less than two or three times the fundamentalfrequency (=1) because of anharmonicity.
Typical energy spacings for vibrational levels are on the order of 10-20J. from theBolzmann distribution, it can be shown that at room temperature typically 1% or less ofthe molecules are in excited states in the absence of external radiation. Thus most
absorption transitions observed at room temperature are from the =0 to the =1 level.
8/13/2019 Infrared Spectroscopy 03
11/70
8/13/2019 Infrared Spectroscopy 03
12/70
The vibrational spectra appear as bands rather than lines. When vibrational
spectra of gaseous diatomic molecules are observed under high-resolutionconditions, each band can be found to contain a large number of closelyspaced componentsband spectra. The structure observed is due to thata single vibrational energy change is accompanied by a number ofrotational energy changes. The form of such a vibration-rotation spectrumcan be predicted from the energy levels of a vibrating-rotating molecule.
> vibrational-rotational bands
Vibrational spectra (III): Rotation-vibration transitions
A vibrational absorption transition from to +1 gives rise to three sets
of lines called branches:
Lower-frequency P branch: =1, J=-1;
Higher-frequency R branch: =1, J=+1;
Q branch: branch: =1, J=0.
8/13/2019 Infrared Spectroscopy 03
13/70
Spectrum of the Rotating Oscillator
The selection rules allow only transitions with = +1 and J = 1
(the transition with J = 0 is normally not allowed except those with
an odd number of electrons (e.g. NO)).
P R
8/13/2019 Infrared Spectroscopy 03
14/70
The IR absorption spectrum can be obtained with gas-phase orwith condensed-phase molecules. For gas-phase moleculesvibration-rotation spectra are observed, while in condensed
phases, the rotational structure is lost.
For most routine analytical applications of infraredspectrometry, spectra are obtained with condensed-phasesamples. Hence, the discuss here centers around the vibrationaltransitions observed with molecules present as pure liquid, as
solutions, or in the solid state.
8/13/2019 Infrared Spectroscopy 03
15/70
Molecular vibrations How many vibrations are possible (=fundamental vibrations)?
A molecule has as many degrees of freedom as the total degree offreedom of its individual atoms. Each atom has three degrees of freedom(corresponding to the Cartesian coordinates), thus in an N-atommolecule there will be 3N degree of freedom.
In molecules, movements of the atoms are constrained by interactionsthrough chemical bonds.
Translation- the movement of the entire molecule while the positionsof the atoms relative to each other remain fixed: 3 degrees of
translational freedom.Rotational transitionsinteratomic distances remain constant but theentire molecule rotates with respect to three mutually perpendicularaxes: 3 rotational freedom (nonlinear), 2 rotational freedom (linear).
8/13/2019 Infrared Spectroscopy 03
16/70
Fundamental Vibrations
Vibrationsrelative positions of the atoms change while the averageposition and orientation of the molecule remain fixed.
8/13/2019 Infrared Spectroscopy 03
17/70
There are two different types of vibrational modes:
Vibrations can either involve a change in bond length
(stretching) or bond angle (bending)
Vibration Types
8/13/2019 Infrared Spectroscopy 03
18/70
8/13/2019 Infrared Spectroscopy 03
19/70
Principle of IR experiments
E-vector in electromagnetic radiationhas frequency Molecular vibrations involving change in dipole moment set upfluctuating electric field
Vibrational energies: fundamental (= one quantum)
Energy transferred to molecule by resonance when vibrationfrequency is the same as that of the electromagnetic radiation
IR SAMPLE SAMPLE*
(MOLECULE, GS) (VIB.)
8/13/2019 Infrared Spectroscopy 03
20/70
Vibrations which do not change the dipole moment areInfrared Inactive
(homonuclear diatomics).
Selection Rules
The energy associated with a quantum of light may be transferred to themolecule if work can be performed on the molecule in the form ofdisplacement of charge.
Selection rule:
A molecule wil l absorb infrared radiation i f the change in vibrational
states is associated with a change in the dipole moment () of themolecule.
= qrq: electrical charge, r: directed distance of that charge from somedefined origin of coordinates from the molecule.
Dipole moment is greater when electronegativity difference between theatoms in a bond is greater. Some electronegativity values are:H 2.2; C 2.55; N 3.04; O 3.44; F 3.98; P 2.19; S 2.58; Cl 3.16
8/13/2019 Infrared Spectroscopy 03
21/70
The theoretical number of fundamental vibrations (absorption
frequencies) will seldom be observed
> overtones (multiples of a given frequency), combination (sum oftwo other vibrations) or difference (the difference of two othervibrations) tones increase the number of bands
> the following effects will reduce the number of theoreticalbands: frequencies which fall outside the measured spectral region (400-
4000 cm 1) bands which are too weak bands are too close and coalesce occurrence of a degenerate band from several absorptions of the
same frequency lack of change in molecular dipole
Why not 3N-6/3N-5 bands in IR spectrum?
8/13/2019 Infrared Spectroscopy 03
22/70
Infrared Spectrum
of Carbon Dioxide
8/13/2019 Infrared Spectroscopy 03
23/70
Vibrational Modes for a CH2Group
8/13/2019 Infrared Spectroscopy 03
24/70
Absorption Regions
8/13/2019 Infrared Spectroscopy 03
25/70
8/13/2019 Infrared Spectroscopy 03
26/70
8/13/2019 Infrared Spectroscopy 03
27/70
Fingerprint region
In the region from 1300 to 400 cm-1, vibrational frequencies are
affected by the entire molecule, as the broader ranges for groupabsorptions in the figure belowfingerprint region.
Absorption in this fingerprint region is characteristic of the moleculeas a whole. This region finds widespread use for identification purpose
by comparison with library spectra.
8/13/2019 Infrared Spectroscopy 03
28/70
8/13/2019 Infrared Spectroscopy 03
29/70
When two bond oscillators share a common atom, they seldom behave
as individual oscillators (unless the individual oscillation frequencies are
widely different).
The frequency of the asymmetric stretching vibration in CO2is at a
shorter wavelength (higher frequency) than for a carbonyl group in
aliphatic ketones (around 1715 cm1).
> there must be strong mechanical coupling or interaction!
Example: CO stretching band in
Methanol: 1034 cm 1
Ethanol: 1053 cm 1
not an isolated stretching vibration, but rather a coupled symmetric
stretching invloving CCO stretching
Coupled Interactions
8/13/2019 Infrared Spectroscopy 03
30/70
The vibrations must be of the same symmetry
The interaction is greatest, when the coupled groups absorb
(individually) near the same frequency--- the same energies of isolated
vibrations.
Strong coupling between stretching vibrations requires a common atombetween the two groups
Coupling between bending and stretching vibrations can occur if the
stretching bond forms one side of the changing angle.
A common bond is required for coupling of bending vibrations. Coupling is negligible when groups are separated by one or more carbon
atoms and the vibrations are mutually perpendicular.
Requirements for Coupled Interactions
8/13/2019 Infrared Spectroscopy 03
31/70
8/13/2019 Infrared Spectroscopy 03
32/70
Hydrogen bonding alters the force constant of both groups:
the XH stretching bands move to lower frequency
the stretching frequency of the acceptor group (B) is also reduced,
but to a lesser degree
The XH bending vibration usually shifts to a shorter wavelength
Effect of Hydrogen Bonding
8/13/2019 Infrared Spectroscopy 03
33/70
y axis is %T or A
x axis is wavenumber (or wavelength)
Iosample I
T = I/Io %T = 100 I/Io
T transmission / transmittance
A = -log T
A absorbance(no units)
(Note A (but not T) concentration)
IR spectrum
8/13/2019 Infrared Spectroscopy 03
34/70
Absorbance spectrum of polystyrene
> generally used for quantitative
work
Ordinate Scaling
Transmittance spectrum of polystyrene
> traditionally used for spectral
interpretation
8/13/2019 Infrared Spectroscopy 03
35/70
8/13/2019 Infrared Spectroscopy 03
36/70
8/13/2019 Infrared Spectroscopy 03
37/70
Dispersive instruments: with a monochromator to be used inthe mid-IR region for spectral scanning and quantitativeanalysis.
Fourier transform IR (FTIR) systems: widely applied andquite popular in the far-IR and mid-IR spectrometry.
Nondispersive instruments: use filters for wavelengthselection or an infrared-absorbing gas in the detection systemfor the analysis of gas at specific wavelength.
8/13/2019 Infrared Spectroscopy 03
38/70
Dispersive IR spectrophotometers
Simplified diagram of a double beam infrared spectrometer
Modern dispersive IR spectrophotometers are invariably double-beam
instruments, but many allow single-beam operationvia a front-panelswitch.
8/13/2019 Infrared Spectroscopy 03
39/70
Double-beam operation compensates for atmospheric absorption, for thewavelength dependence of the source spectra radiance, the opticalefficiency of the mirrors and grating, and the detector instability, whichare serious in the IR region.single-beam instruments not practical.
Double-beam operation allows a stable 100% T baseline in the spectra.
8/13/2019 Infrared Spectroscopy 03
40/70
Dispersive spectrophotometers Designs
Null type instrument
8/13/2019 Infrared Spectroscopy 03
41/70
Components of dispersive spectrophotometers
Nernst Glower heated rare earth oxide rod(~1500 K)
1-50 m
(mid- to far-IR)
Globar heated SiC rod (~1500 K) 1-50 m(mid- to far-IR)
W filament lamp 1100 K 0.78-2.5 m
(Near-IR)
Hg arc lamp plasma 50 - 300 m(far-IR)
CO2 laser stimulated emission lines 9-11 m
1. IR source
8/13/2019 Infrared Spectroscopy 03
42/70
8/13/2019 Infrared Spectroscopy 03
43/70
3. Optical system
8/13/2019 Infrared Spectroscopy 03
44/70
Reflection gratings( made from various plastics): the groovespacing is greater (e.g. 120 grooves mm-1). To reduce the effect ofoverlapping orders and stray radiation, filters or a preceding prism
are usually employed. Two or more gratings are often used withseveral filters to scan a wide region.
Mirrorsbut not lenses are used to focus and collimate the IRradiation. Generally made from Pyrex or another material with lowcoefficient of thermal expansion. Front surfaces coated with avacuum-deposited thin metal film of Al, Ag, or Au.
8/13/2019 Infrared Spectroscopy 03
45/70
Windowsare used for sample cells and to permit various compartmentto be isolated from the environment.
transparent to IR over the wavelength regioninert to the various chemicals analyzedcapable of being shaped, ground, and polished to the desiredoptical quality
Fourier Transform Infrared Spectrometer (FTIR)
8/13/2019 Infrared Spectroscopy 03
46/70
The Fourier transform method provides an alternatives to the useof monochromators based on dispersion.
In conversional dispersive spectroscopy, frequencies are separatedand only a small portion is detected at any particular instant, whilethe remainder is discarded. The immediate result is afrequency-domain spectrum.
Fourier transform infrared spectroscopy generates time-domainspectraas the immediately available data, in which the intensity isobtained as a function of time.
Direct observation of a time-domain spectrum is not immediatelyuseful because it is not possible to deduce, by inspection,frequency-domain spectra from the corresponding time-domainwaveform (Fourier transformis thus introduced).
Fourier Transform Infrared Spectrometer (FTIR)
Si l b FTIR S t t
8/13/2019 Infrared Spectroscopy 03
47/70
In one arm of the interferometer, the IR source radiation travels through the beamsplitter to the fixed mirror back to the beam splitter through the sample and to thedetector. In the other arm, the IR source radiation travels to the beam splitter to themovable mirror, back through the beam splitter to the sample and to the detector. Thedifference in pathlengths of the two beams is the retardation . An He-NE laser is usedas a monochromatic reference source. The laser beam is sent through the interferometer
in the opposite direction to that of the IR beam.
Single-beam FTIR Spectrometer
8/13/2019 Infrared Spectroscopy 03
48/70
Double-beam FTIR
Spectrometer
Interferometer
8/13/2019 Infrared Spectroscopy 03
49/70
Interferometer
Michelson interferometer
If moving mirror moves 1/4 (1/2 round-trip) waves are out of phase at beam-splitting mirror - no signal
If moving mirror moves 1/2 (1 round-trip) waves are in phase at beam-splittingmirrorsignal
...
8/13/2019 Infrared Spectroscopy 03
50/70
Interferograms
8/13/2019 Infrared Spectroscopy 03
51/70
Difference in pathlength called retardation Plot vs. signal - cosine wave with frequency proportional to lightfrequency butsignal varies at much lower frequency
One full cycle when mirror moves distance /2 (round-trip = )
Frequency of signal:
Substituting =c/
If mirror velocity is 1.5 cm/s
Bolometer, pyroelectric, photoconducting IR detectors can "see changes
on 10-4s time scale!
MMMM VVf 2
2/==
c
Vf MM
2=
VMMvelocity of moving mirror
10
10 10/103
/3
== scm
scm
f
8/13/2019 Infrared Spectroscopy 03
52/70
Computer needed to turn complex interferograms into spectra.
8/13/2019 Infrared Spectroscopy 03
53/70
Advantages of FTIR
8/13/2019 Infrared Spectroscopy 03
54/70
very high resolution (< 0.1 cm 1)Two closely spaced lines only separated if one complete "beat" is recorded. As
lines get closer together, must increase.(cm1) = 1/Mirror motion is 1/2 Resolution governed by distance movable mirror travels
very high sensitivity (nanogram quantity)can be coupled with GC analysis (> measure IR spectra in gas-phase)
High S/N ratios - high throughputFew optics, no slits mean high intensity of light
Rapid (
8/13/2019 Infrared Spectroscopy 03
55/70
Usually to improve resolution
decrease slit width but less lightmakes spectrum "noisier" - signalto noise ratio (S/N)
n # scans
S/N improves with more scans
(noise is random, signal is not!)
nN
S
SS
S
nN
S
i==
2)(
To improve S/N ratio
8/13/2019 Infrared Spectroscopy 03
56/70
For routine instrument calibration, run the spectrum ofpolystyrene film (or indene) at resolution 2 cm-1. Band
positions are available in the literature.
Higher resolution calibrations may be made from gas-
phase spectra (e.g. HCl gas).
Spectrum calibration
S l ti t h i
8/13/2019 Infrared Spectroscopy 03
57/70
Sample preparation techniques
The preparation of samples for infrared spectrometry is often the mostchallenging task in obtaining an IR spectrum. Since almost all substances absorbIR radiation at some wave length, and solvents must be carefully chosen for the
wavelength region and the sample of interest.
Infrared spectra may be obtained for gases, liquids or
solids (neat or in solution)
8/13/2019 Infrared Spectroscopy 03
58/70
8/13/2019 Infrared Spectroscopy 03
59/70
8/13/2019 Infrared Spectroscopy 03
60/70
Regions of transparency for common infrared solvents.
The horizontal lines indicate regions where solvent transmits at least25% of the incident radiation in a 1-mm cell.
8/13/2019 Infrared Spectroscopy 03
61/70
Solid samples
Spectra of solids are obtained as alkali halide discs (KBr), mulls
(e.g. Nujol, a highly refined mixture of saturated hydrocarbons) and
films (solvent or melt casting)
Alkali halide discs:
1. A milligram or less of the fine ground sample mixed with about
100 mg of dry KBr powder in a mortar or ball mill.
2. The mixture compressed in a die to form transparent disc.
Mulls
1. Grinding a few milligrams of the powdered sample with a mortar
or with pulverizing equipment. A few drops of the mineral oiladded (grinding continued to form a smooth paste).
2. The IR of the paste can be obtained as the liquid sample.
8/13/2019 Infrared Spectroscopy 03
62/70
1.Fundamental chemistry
Determination of molecular structure/geometry.
e.g. Determination of bond lengths, bond angles of
gaseous molecules
2. Qualitative analysissimple, fast, nondestructive
Monitoring trace gases: NDIR.Rapid, simultaneous
analysis of GC, moisture, N in soil. Analysis of fragments
left at the scene of a crime
Quantitative determination of hydrocarbons on filters, in
air, or in water
Main uses of IR spectroscopy:
8/13/2019 Infrared Spectroscopy 03
63/70
Near-infrared and Far-infrared absorption
The techniques and applications of near-infrared (NIR) and
far-infrared (FIR) spectrometry are quite different from those
discussed above for conventional, mid-IR spectrometry.
Near-infrared: 0.8 -2.5 m, 12500 - 4000 cm
-1
Mid-infrared: 2.5 - 50 m, 4000 - 200 cm-1
Far-infrared:50 - 1000 m, 200 - 10 cm-1
8/13/2019 Infrared Spectroscopy 03
64/70
Near-infrared spectrometry
NIR shows some similarities to UV-visible spectrophotometry and
some to mid-IR spectrometry. Indeed the spectrophotometers
used in this region are often combined UV-visible-NIR ones.
The majority of the absorption bands observed are due to
overtones (or combination) of fundamental bands that occur in
the region 3 to 6 m, usually hydrogen-stretching vibrations.
NIR is most widely used for quantitative organic functional-group
analysis. The NIR region has also been used for qualitative
analyses and studies of hydrogen bonding, solute-solvent
interactions, organometallic compounds, and inorganiccompounds.
8/13/2019 Infrared Spectroscopy 03
65/70
Far-infrared spectrometry
Almost all FIR studies are now carried out with FTIRspectrometers.
The far-IR region can provide unique information.
i) The fundamental vibrations of many organometallic and
inorganic molecules fall in this region due to the heavy atomsand weak bonds in these molecules.
ii) Lattice vibrations of crystalline materials occur in this region,
iii) Electron valence/conduction band transition in
semiconductors often correspond to far-IR wavelengths.
8/13/2019 Infrared Spectroscopy 03
66/70
Infrared Spectrum of CCl
8/13/2019 Infrared Spectroscopy 03
67/70
Infrared Spectrum of CCl4
8/13/2019 Infrared Spectroscopy 03
68/70
8/13/2019 Infrared Spectroscopy 03
69/70
8/13/2019 Infrared Spectroscopy 03
70/70