How do I troubleshoot Start-Up Check errors on a CIRRUS HD-OCT? ZEISS Quick Help: CIRRUS™ HD-OCT
This guide reviews remedies for three standard error messages that may occur during the Start-Up Check process: Scanning module error, preventative maintenance error and Archive volume error.
Scanning module error
Message: "Errors were detected in the CIRRUS HD-OCT scanning module. You cannot acquire or review scans until the problem is corrected. Please contact Carl Zeiss Meditec Customer Service for assistance.” (Figure 1)
Error detail: Click Details to open the Details Dialog box. Scroll to the bottom of the window to view the error detail (Figure 2).
Figure 1
Figure 2
Error Message
Error Detail
How do I troubleshoot Start-Up Check errors on a CIRRUS HD-OCT? | ZEISS Quick Help: CIRRUS™ HD-OCT
Page 2 of 5
Troubleshoot: Disconnected controller
1. Click the Windows Logo in the left corner of the menu bar, select Control Panel and then select Network and Sharing Center. From the left-hand navigation column of the Network and Sharing Center window, click Change adapter settings (Figure 3) to reveal all existing Networking Connections — External, Internal and Wireless (Figure 4).
2. Locate the Internal Network icon (Figure 5a) and confirm that it indicates Enabled. If it indicates Disabled as shown below, right-click on the icon and select Enable (Figure 5b).
Figure 3
Figure 4
Figure 5b
Click here to see the list of Network adapters.
Figure 5a
3. Return to the Internal Network icon, right-click on the icon and select Properties from the drop-down menu (Figure 6a). From the list of connection items under the Networking tab, select Internet Protocol Version 4 (TCP/IPv4), then click Properties (Figure 6b) and verify the following Internet Protocol Properties: (Figure 7)
• "Use the following IP address" radio button should be selected – IP address is configured to 192.168.52.100 – Subnet mask is configured to 255.255.255.0
• "Use the following DNS server address" radio button should be selected – Preferred DNS server field is empty – Alternate DNS server field is empty.
Note: If the error message persists after you verify that the Internet Protocol Properties are correct, contact Carl Zeiss Meditec Technical Support at 1-800-341-6968.
How do I troubleshoot Start-Up Check errors on a CIRRUS HD-OCT? | ZEISS Quick Help: CIRRUS™ HD-OCT
Figure 6b
Page 3 of 5
Figure 7
Figure 6a
How do I troubleshoot Start-Up Check errors on a CIRRUS HD-OCT? | ZEISS Quick Help: CIRRUS™ HD-OCT
Preventive maintenance errorMessage: The instrument is prompting a notice for preventative maintenance (Figure 1).
Error Message
Troubleshoot: Preventive maintenance
This message does not indicate a problem with the CIRRUS HD-OCT instrument but rather serves as a reminder to schedule a preventative maintenance appointment for the instrument. ZEISS recommends regular preventative maintenance service. To schedule preventative maintenance service, contact Carl Zeiss Meditec Technical Support at 1-800-341-6968.
Note: Only ZEISS certified, factory-trained Carl Zeiss Meditec personnel may perform a preventative maintenance.
Figure 1
Page 4 of 5
How do I troubleshoot Start-Up Check errors on a CIRRUS HD-OCT? | ZEISS Quick Help: CIRRUS™ HD-OCT
Archive volume errorMessage: “Archive volume is not available” or “Archive volume has not been created” (Figure 1).
Note: This error is associated with the yellow Status message located near the bottom left corner of the application screen (Figure 2).
Troubleshoot: Archive volume
This message indicates that the instrument does not have a current archive, either one is not available or one has not been created.
Archive volume is not available: This message indicates that the instrument has an archive that is not set as current, or its archive location is not available. If the archive location on the network is not available, consult your Network Administrator.
Archive volume has not yet been created: This message indicates that an archive has not been created for the instrument. For assistance with the creation of an archive volume, refer to ZEISS Quick Help document How do I create a new (non-DICOM) archive on my CIRRUS HD-OCT?
Carl Zeiss Meditec AG Goeschwitzer Strasse 51–5207745 JenaGermanywww.zeiss.com/med
Carl Zeiss Meditec, Inc.5160 Hacienda DriveDublin, CA 94568USAwww.zeiss.com/med
Refer to the CIRRUS HD-OCT User Manual Instructions for Use for safe and effective operation of the instrument.
Page 5 of 5
SER.
9124
Pr
inte
d in
the
Unite
d St
ates
. CZ
-VIII
/201
7 US
A Ed
ition
.Th
e co
nten
ts o
f the
refe
renc
e gu
ide
may
diff
er fr
om th
e cu
rrent
stat
us o
f app
rova
l of t
he p
rodu
ct o
r ser
vice
offer
ing
in y
our c
ount
ry. P
leas
e co
ntac
t our
regi
onal
repr
esen
tativ
e fo
r mor
e in
form
atio
n. S
ubje
ct to
cha
nge
in d
esig
n an
d sc
ope
of d
elive
ry
and
as a
resu
lt of
ong
oing
tech
nica
l dev
elop
men
t. CI
RRUS
is a
eith
er tr
adem
ark
or re
gist
ered
trad
emar
k of
Car
l Zei
ss M
edite
c AG
or o
ther
com
pani
es o
f the
ZEI
SS G
roup
in G
erm
any
and/
or o
ther
cou
ntrie
s. ©
Carl
Zeiss
Med
itec,
Inc.
, 201
7. A
ll rig
hts r
eser
ved.
Figure 2
Figure 1
Recommended