Double lumen remodeling balloon: New technique for treatment of MCA bifurcation aneurysm

Preview:

DESCRIPTION

Double lumen remodeling balloon: New technique for treatment of MCA bifurcation aneurysm. Kadziolka K, Leautaud A., Estrade L., W. Mustafa, Pierot L. CHU REMIS Interventional Neuroradiology Department France. New technology. - PowerPoint PPT Presentation

Citation preview

Double lumen remodeling balloon:New technique for treatment of MCA

bifurcation aneurysm

Kadziolka K, Leautaud A., Estrade L., W. Mustafa,Pierot L. CHU REMIS Interventional Neuroradiology Department France

New technology

As medical technology advances such as stenting, remodeling,

an increasing number of intracranial aneurysms are being addressed for

EVT.

Aneurysm morphology

MCA large, bifurcation aneurysms are

technically challenging due to the complex anatomy.

Bifurcation aneurysms are often wide necked and incorporate the origin of arterial branches in the aneurysm broad base.

Endovascular therapeutic options

Remodeling with use of hyper compliant balloon.

Double balloon technique ( kissing balloon)

Distal round balloon technique

Stent with or without balloon

„Y” stenting ( dual antiplatelet premedicatio )

Double lumen distal remodeling balloon?

Distal round balloon technique

v

Dual lumen balloon microcatheter

Primary experience with use of novel double lumen balloon

ASCENT 4 x 7  mm Micrus Endovascular Corp, San Jose, California,

USA

Balloon Catheter Technical Specification

Guidewire compatibility ≤.014” Tip length distal to balloon 3 mm Inner lumen diameter .0170” Outer diameter 2.9F Marker band 3 cm from distal tip allows delivery

of embolics (Ascent 4x7 mm) Compatible with DMSO Guiding catheter compatibility ,050” minimum ID

Patients and Methods

A 60-year-old male presented with unruptured left MCA bifurcation aneurysm

measuring 11×7 mm with neck 4 mm. Dome to neck ratio 2.5.

Both parents suffered from ruptured aneurysm subarachnoid hemorrhage.

Periprocedural Medication

Systemic heparinisation: 50 IU/kg bolus infusion followed by 1500 IU/hour.

ASA 250 mg IV.

Systemic heparinisation prolonged for 24 hours.

Endovasculare procedure

Endovasculare procedure

Results

Results

Clinical outcome at discharge was unchanged.

Anatomic angiographic outcome appears as residual neck.

Follow-up Examination

Follow up MRI after 3 months. DSA follow-up at 6 months after the treatment

Follow-up Examination

Follow-up Examination

Anatomic angiographic outcome appears as complete occlusion.

Embolisation of the aneurysm with dome to neck diameter ≥ 6mm /3 mm distal microcatheter tip/

Rigidity of the double lumen catheter

Cigar ellipsoidal shape instade of round more spherical

Technical difficulties and limitations

Our recomendations for use of Ascent double lumen balloon

New compliant guide catheters 6F or coaxial guiding platform.

Microwire .014" rather then smaller

Conclusion

Double lumen balloon provides a means to acheive reconstruction of complex mca bifurfation aneurysm and can be an alternative to double balloon(kissing)technique or “Y” stenting for some selective bifurcation aneurysms.

Due to development of new endovascular devices treatment of mca bifurcation becomes safer, feasible and durable.