CHROMATIC CONFOCAL MICROSCOPY USING LIQUID CRYSTAL PANELS
by
Qi Cui
____________________________ Copyright © Qi Cui 2019
A Report Submitted to the Faculty of the
COLLEGE OF OPTICAL SCIENCES
In Partial Fulfillment of the Requirements
For the Degree of
MASTER OF SCIENCE
In the Graduate College
THE UNIVERSITY OF ARIZONA
2019
2
3
ACKNOWELDGEMENT
First and foremost, I would like to thank my mentor, professor Rongguang Liang. During my three years at U of Arizona, I’ve learnt much from professor Rongguang Liang, not only the knowledge of optical systems, but also the right way to do research. The invaluable experience in Applied Optics Lab has prepared me well to start my PhD, and I believe it would also be precious in the future. Second, I would like to thank professor Dongkyun Kang and professor Leilei Peng for being my Master’s oral exam committee member. I also thank professor Kang for his suggestions and help on my student career. Third, I would like to thank my colleagues in Applied Optics Lab. I thank Xiaobo Tian, Zhihan Hong, Wei-chen Liao, Bofan Song, Wenjun Kang, Shaobai Li, for their help and beneficial talks. They helped me on my research and my courses. Fourth, I would like to thank my friends at U of A. I thank Aoxue Han, Hwang-Jye Yang, Chloë Castle, Tyler Joseph, Jieun Ryu, Moonseob Jin, and other friends. I thank them for their invaluable friendship. Fifth, I would like to thank my professors and TAs at U of A. I especially thank professor Amit Ashok and professor Tom Milster. I thank them for their recommendation letters for my Fall 2018 and Fall 2019 PhD applications. In the end, I thank my families for their endless love and supports in my life.
4
DEDICATION
To my parents, and Tong Shen
5
TABLE OF CONTENTS LIST OF FIGURES .......................................................................................................... 6 ABSTRACT ...................................................................................................................... 7 CHAPTER 1: BACKGROUND ....................................................................................... 8 CHAPTER 2: CHROMATIC CONFOCAL SYSTEM .................................................... 10 CHAPTER 3: SYSTEM CALIBRATION AND CHARACTERIZATION .................... 12 3.1 Source spectrum normalization........................................................................ 12 3.2 Dispersion prism description and evaluation line ............................................ 13 3.3 LCD scanning method ..................................................................................... 15 3.4 Evaluation of line response and axial resolution ............................................. 16 CHAPTER 4: EXPERIMENTAL RESULTS .................................................................. 18 4.1 50 μm step standard ......................................................................................... 18 4.2 Onion epidermis 3D imaging ........................................................................... 19 CHAPTER 5: CONCLUSION ......................................................................................... 20 REFERENCES ................................................................................................................. 21
6
LIST OF FIGURES
Chapter 2 Fig. 1. Schematic diagram of confocal system ················································ 11
Chapter 3 Fig. 2. Spectrum of plasma lamp light after passing through two LCDs ·················· 12 Fig. 3. Zemax simulation on Amici prism ····················································· 14 Fig. 4. Point array pattern and evaluation lines ··············································· 16 Fig. 5. Evaluation line response curve ························································· 17
Chapter 4 Fig. 6. 50 μm step standard measurement ····················································· 18 Fig. 7. Onion epidermis 3D imaging ··························································· 19
7
Abstract:
A chromatic confocal microscopy system in combination with two liquid crystal display (LCD)
panels is proposed and demonstrated for surface profiling. The major advantage of this system is
no mechanical translation is needed for three-dimensional (3D) imaging. The axial scanning is
realized thanks to the chromatic aberration in the objective, whereas the lateral scanning is
achieved by turning on different pixels on LCDs. Chromatic aberration of objective lens is used
to provide wavelength-to-depth coding and decoding is realized by using a dispersion prism.
System performance is validated with a 50 μm step standard and the capability of 3D imaging is
demonstrated with an onion epidermis.
8
Chapter 1: Background
Confocal microscopy [1] is a powerful tool to measure object surface and three-dimensional
volume structures, owing to its optical sectioning ability [2-4]. A traditional confocal system
requires point by point scanning in both lateral (x,y) and axial (z) directions, therefore, the
scanning speed is relatively slow. To improve the scanning speed, slit confocal microscopy
system [5] and spinning-disk confocal microscopy system [6] were developed, but mechanical
scanning is still required for lateral and axial scanning.
Chromatic confocal microscopy [7-9] was proposed to avoid axial mechanical scanning, taking
advantages of wavelength-to-depth coding ability of chromatic aberration. Using the objective
with uncorrected axial chromatic aberration, different wavelengths are focused at different
depths. Depth information can be easily collected without any axial mechanical scanning. The
maximum depth scan range is determined by bandwidth of light source and the dispersion ability
of objective lens. A broader bandwidth light source and strong dispersion objective lens provide
longer depth scan range. Xenon lamp or a supercontinuum laser is often used to achieve large
depth scan range [10,11].
Slit scan method and spinning disk method in combination with chromatic confocal principle
were developed to further increase the scanning speed [12,13]. However, lateral mechanical
scanning is still necessary. In 2000, a non-translational confocal system using digital micromirror
device (DMD) was proposed to avoid any mechanical scanning. Lateral scanning was achieved
by switching DMD mirror states, and a tunable laser was used to achieve wavelength scanning
[14]. Botvinick el al [15] and Wang et al [16] also used DMD to avoid lateral mechanical
scanning, but no axial scanning process was mentioned in their work.
9
Recently, Hillenbrand et al incorporated a dispersion prism into chromatic confocal microscopy
system [17]. The dispersion prism is used to analyze the spectrum of all lateral channels at the
same time, therefore any axial scanning is avoided. However, a movable pinhole array is still
needed to cover the entire field of view.
In this paper, we propose a simple chromatic confocal system without moving component using
two liquid crystal display (LCD) panels to create a virtual movable illumination pinhole array
and the corresponding detection pinhole array. The confocal system would be described in
Section 2, and system characteristics are evaluated in Section 3. In Section 4, experimental
measurements of a 50 μm step standard and an onion epidermis are presented.
10
Chapter 2: Chromatic confocal system
The schematic diagram of the chromatic confocal system is shown in Figure 1. A plasma lamp
with a 450-650 nm bandpass filter (Semrock, FF01-550/200-25) is used as the light source. Light
from the light source is collimated by an achromatic doublet L1 (Thorlabs, AC254-100-A) and
passes through a linear polarizer P1, illuminates the illumination LCD with 1024 × 768 pixels
and then passes through a linear analyzer A1. A point array pattern is displayed on the
illumination LCD. By passing only the light at locations of displayed point array, the
illumination LCD works as the point source array. Light from the illumination LCD is collimated
by a camera lens L2 (f = 105 mm) and focused by a microscope objective L3 (10X, NA = 0.1)
onto the object. The f-number of L2 has been set as 5.6. Owing to remaining chromatic
aberration in the objective, different wavelengths are focused onto different depths onto the
object. Light reflected from the object is collimated by the objective L3 and focused by the
doublet L2 onto the detection LCD, which is conjugate with the illumination LCD. A linear
polarizer P2 is placed before and a linear analyzer A2 after the detection LCD. The same point
array pattern is displayed on the detection LCD, and detection LCD works as detection pinhole
array. For a single wavelength, out-of-focus light is rejected by those pinholes. Light passing
through the detection LCD is collimated by an achromatic doublet L4 (Thorlabs, AC254-050-A-
ML). A custom-made dispersion prism is used to disperse different wavelengths into different
propagation directions. For each detection pinhole, different wavelengths are focused by an
achromatic doublet L5 (Thorlabs, AC254-050-A-ML) onto a gray-scale detector (pco.edge 5.5,
2560 × 2160 pixel resolution, 6.5 μm pixel pitch) at different locations along a line determined
by the dispersion prism. This “evaluation line” represents spectrum of the light reflected from the
object [17]. By analyzing all evaluation lines formed on the detector, we can get depth
11
information of corresponding points on the object. By changing point array pattern displayed on
two LCDs sequentially, lateral scan is achieved without any mechanical translation. Note that
two LCDs are connected to the same PC, therefore point array pattern displayed on two LCDs
are exactly the same and can be changed synchronously.
Fig. 1. Schematic diagram of confocal system.
12
Chapter 3: System calibration and characterization
3.1 Source spectrum normalization
Due to the light filter nature of pinholes, a multi-point scanning confocal system like scanning-
disk confocal system usually has a low light efficiency and suffers from low signal-to-noise
ratio. Using a plasma lamp is an effective way to decrease the exposure time and increase the
signal-to-noise ratio. The spectrum of plasma lamp light passed through two LCDs is measured
by a high resolution spectrometer (Ocean optics, HR2000+) and used to normalize the intensity
distribution on the evaluation lines. Note that spectrometer should be placed after two LCDs,
because LCD has different transmissions over different wavelengths. The measured spectrum is
shown in Figure 2. In our experiment, we only used the wavelengths between 505 nm and 650
nm. The reason is that the light intensity at 500 nm is relatively low. Normalizing with small
intensities would dramatically amplify the noises. This step guarantees that the signal-to-noise
ratio of the normalized intensity of the evaluation line is high enough to extract depth
information accurately.
450 500 550 600 650
wavelength in nm
0
2000
4000
6000
8000
10000
12000
14000
Inte
nsity
spectrum
Fig. 2. Spectrum of plasma lamp light after passing through two LCDs.
13
3.2 Dispersion prism description and evaluation line
In order to analyze spectrum of all lateral channels at the same time, a dispersion prism is used to
disperse different wavelengths to form evaluation lines on the detector. We used a double Amici
prism (DAP) designed by Mu et al [18] in our experiment. The main reason of using this design
is that this prism is free of distortion aberration. Distortion aberration introduced by a normal
dispersion prism would decrease quality of evaluation lines and hence confocal images of the
object. The configuration of double Amici prism is shown in Fig. 3(a). This prism consists of
three wedge elements. The first and third elements are made of H-LAK52 glass and second
element is made of H-ZF88 glass. The optimal angles of each elements are given in the diagram.
To evaluate the prism performance, we input the parameters into ZEMAX in an optical system
with image space F/# = 5 (corresponding image space NA = 0.1). A 50 mm focal length paraxial
lens is added after the prism to focus light onto the image plane without introducing any
aberration. Two fields are examined: the first field is on the optical axis (x = 0˚ and y = 0˚), the
second field is 70 percent of full field of a 50 mm focal lens with a diameter = 25.4 mm (x = 0˚
and y = 10.1˚). Layout of the system, ray aberrations and distortion in percent are shown in
figure 3(b), 3(c) and 3(d). Figure 3(b) shows the system layout. The blue lines represent light
rays at 10.1˚ field, and green lines represent light rays at 0˚ field. Figure 3(c) indicates the lateral
color of the system. The length of evaluation line is determined by the focus position difference
between 450 nm and 650 nm. If the prism is used with a 50 mm focal length imaging lens, the
length of evaluation line is 545 μm. This length would be used to design point array pattern
displayed on LCDs. Figure 3(d) shows the distortion aberration of the prism. The maximum
distortion is 0.3 percent and it is negligible.
14
In general, the length of evaluation line is determined by the spectrum of light source, the
dispersion prism, and the focal length of imaging lens. For a given prism and source bandwidth,
a longer focal length imaging lens would increase the length of the evaluation line. Note that
there is a trade-off between spatial and spectral resolution due to the fixed pixel number on the
detector. A longer evaluation line on the detector means more pixels are used to represent
spectrum of an object point, hence the spectral resolution is increased, but decreasing the spatial
resolution because smaller number of object points could be sampled. The axial resolution of the
chromatic confocal imaging system is not determined by the full width half maximum (FWHM)
of the longitudinal point spread function (PSF). Instead, it is determined by the length of the
evaluation line and the chromatic aberration in the objective.
Fig. 3. (a) Configuration of double Amici prism, (b) the layout of DAP in F/5 optical system, (c) positions of the rays with different wavelengths for the field of 10.1˚, and (d) distortion aberration in percent.
15
3.3 LCD scanning method
The LCD consists of 1024 × 768 pixels, the pixel size is 12 μm × 12 μm. The light can pass
through LCD only if some pixels are turned on. After passing through LCD, the polarization
direction of the incident beam rotates by 90 degrees. By turning on some pixels while turning
off the others, a point array pattern is generated on two LCDs, which can be used as multi-point
light source and detection pinhole array, respectively.
In our experiment, a single pixel is used as a pinhole. The structure of point array pattern is
designed to achieve zero cross-talk and maximum utilization of detector pixels. The spacing
between adjacent points on LCDs in x direction is 48 μm, and in y direction is 516 μm. A
schematic diagram of point array pattern is shown in Fig. 4(a). The spacing in y direction is
much larger than that in x direction, because a point light source passing through the prism
would spread a 504 μm evaluation line in y direction, therefore the spacing in y direction needs
to be large enough to avoid cross-talk. Note the measured length of evaluation line is smaller
than our simulation result, which is 545 μm. The reason is that intensity of wavelengths between
450-460 nm is very small, this part on evaluation line is overwhelmed by noises and hard to
discriminate. The 48 μm spacing in x direction guarantees that no cross-talk in x directions and it
provides some tolerances for system misalignment. An evaluation line pattern is shown in Fig.
4(b), each evaluation line is generated by a point source in Fig. 4(a) and its orientation is
determined by the dispersion prism.
To achieve lateral scanning, a point array pattern on illumination LCD is shifted in raster
scanning manner to cover the entire field of the objective. By shifting the corresponding pinhole
array in detection LCD, the confocal images can be taken without any mechanical translation.
16
The frame rate of a commercial LCD is 60 Hz, which limits the imaging speed of this confocal
system.
Fig. 4. (a) Point array pattern. (b) Evaluation lines.
3.4 Evaluation of line response and axial resolution
Two important characteristics of a chromatic confocal system are chromatic focus shift and axial
resolution. The chromatic focus shift is characterized as a function of focus position versus
wavelength, it is used to find surface profile of the object. A common method to characterize
chromatic focus shift is realized by using a tunable laser light source and a mirror object [14]. A
single pixel is opened on the optical axis and used as a point light source. For each tuned
wavelength, the mirror is scanned along z axis to find a depth which results in maximum
detected intensity. By finding all wavelength-to-depth mappings, chromatic focus shift
relationship can be determined.
17
To extract the depth information from the normalized evaluation line, we first established the
relationship between the imaging depth and the pixel location of the maximum normalized
intensity. We used a flat mirror as the object and attached it to a z-direction motorized translation
stage (Thorlabs, MT1-Z8). The mirror was scanned in a range of 100 μm, and at every registered
depth the location of maximum normalized intensity on the evaluation line was recorded. The
plot is shown in Fig. 5. The fitted function is shown on the top left of the figure. In our
experiment, the length of evaluation line is 77 pixels on the detector. We fitted the measured data
with a third order polynomial model.
The axial resolution is another important characteristic of the confocal system. As we mentioned
in Section 3.2, the axial resolution is determined by the length of evaluation line and the
chromatic focal shift. With a longer evaluation line, more pixels can be used to image the
spectrum, and therefore more precise the wavelength can be used to find the corresponding axial
depth. In our experiment, the scanning depth is 100 μm, the length of evaluation line is 77 pixels,
therefore the axial resolution is 1.3 μm.
Fig. 5. Depth versus the location of normalized maximum intensity on the evaluation line.
18
Chapter 4: Experimental results
4.1 50 μm step standard
To verify the surface profiling capability of our system, we first measured a 50 μm step standard
(VLSI, SHS-50.0QC). Wide-field image of this step standard is shown in Fig. 6(a). This image
was taken when all LCD pixels were turned on and the dispersion prism was removed. To obtain
confocal images, a point array pattern described in Section 3.3 was displayed and shifted across
all areas on the illumination LCD for lateral scanning. The scanning step is 1 pixel and the total
scanning step is 172. During the scanning process, the evaluation lines of all object points were
recorded. Depth of each object point was deduced from the evaluation lines using the method
discussed in Section 3.4. The measured surface profile is shown in Fig. 6(b). The measured
height is 50.86 μm and root mean square (RMS) error is 1.43 μm. The measurement error is less
than 1.3 μm, which is the theoretical axial resolution calculated in Section 3.4. RMS error is
slightly greater than 1.3 μm, this may be due to the measurement error at some object point
locations. This experiment demonstrates that the proposed chromatic confocal system is capable
to do surface profiling based on evaluation-line processing.
Fig. 6. (a) Wide field image of a 50 μm step standard. (b) Reconstructed surface profile.
19
4.2 Onion epidermis 3D imaging
To demonstrate three-dimensional imaging ability of the proposed system, we measured an
onion epidermis. We used the same lateral scanning procedure described in Section 4.1. The
evaluation lines of all object points were recorded and used to reconstruct 3D image of the onion
epidermis by finding maximum intensity on each evaluation line. Fig. 7(a) shows a reconstructed
image of the onion epidermis. In this figure, every object point is in focus. Fig. 7(b) provides
surface profile of this onion epidermis. The experiment further demonstrates the 3D imaging
capability of the proposed confocal imaging system without any moving component.
Fig. 7. (a) Reconstructed 3D image of an onion epidermis and (b) surface profile.
20
Chapter 5: Conclusion
Chromatic confocal microscopy is a promising tool to measure object surface and three-
dimensional volume structures. In this paper, we present a chromatic confocal system without
moving components for surface profiling. Two LCD panels are incorporated into the system and
used as multi-point light source and detection pinhole array. By changing point array pattern
displayed on two LCDs, lateral mechanical scan can be avoided. Axial scan is avoided by using
an objective lens with uncorrected chromatic aberration and a dispersion prism. The speed of this
system is limited by LCD frame rate. The axial resolution of our system is 1.3 μm and can be
further increased by using a longer focal length imaging lens. However, the trade-off between
lateral and axial resolution always exists and must be carefully balanced. The future research
would focus on how to improve lateral and spatial resolution at the same time and how to reduce
the number of necessary images to reconstruct the confocal image faster.
21
References:
1. M. Minsky, “microscopy apparatus,” U.S. Patent 3,013,467 (19 Dec. 1961). 2. D. K. Hamilton and T. Wilson, “Three-dimensional surface measurement using the confocal
scanning microscope,” Appl. Phys. B 27, 211–213 (1982). 3. K. Carlsson and N. Aslund, "Confocal imaging for 3-D digital microscopy," Appl. Opt. 26,
3232-3238 (1987). 4. D. K. Hamilton, T. Wilson, and C. J. R. Sheppard, “Experimental observations of depth-
discrimination properties of scanning microscopes,” Opt. Lett. 6, 625–626 (1981). 5. C. J. R. Sheppard and X. Q. Mao, “Confocal microscopes with slit apertures,” J. Mod. Opt.
35, 1169–1185 (1988). 6. M. Petran, M. Hadravsky, M. D. Egger, and Galambos, “Tandem-scanning reflected-light
microscope,” J. Opt. Soc. Am. 58, 661-664 (1968). 7. G. Molesini, G. Pedrini, P. Poggi and F. Quercioli, “Focus-wavelength encoded optical
profilimeter,” Opt. Commun. 49, 229-233 (1984). 8. M. A. Browne, O. Akinyemi, and A. Boyde, “Confocal surface profiling using chromatic
aberration,” Scanning 14, 145–153 (1992). 9. M. Maly and A. Boyde, “Real-time stereoscopic confocal reflection microscopy using
objective lens with linear longitudinal chromatic dispersion,” Scanning 16, 187–192 (1994). 10. K. Shi, P. Li, S. Yin, and Z. Liu, "Chromatic confocal microscopy using supercontinuum
light," Opt. Express 12, 2096-2101 (2004). 11. R. J. Garzón, J. Menese, G. Tribillon, T. Gharbi, and A. Plata, “Chromatic confocal
microscopy by means of continuum light generated through a standard single-mode fibre,” J. Opt. A: Pure Appl. Opt. 6, 544–548 (2004).
12. P. C. Lin, P. Sun, L. Zhu and Y. Fainman , “Single-shot depth-section imaging through chromatic slit-scan confocal microscopy,” Appl. Opt. 37, 6764-6770 (1998).
13. H. Tiziani and H. Uhde, "Three-dimensional image sensing by chromatic confocal microscopy," Appl. Opt. 33, 1838-1843 (1994).
14. S. Cha, P. C. Lin, L. Zhu, P. Sun and Y. Fainman, “Nontranslational three-dimensional profilometry by chromatic confocal microscopy with dynamically configurable micromirror scanning,” Appl. Opt. 39, 2605-2613 (2000).
15. E. L. Botvinick, F. Li, S. Cha, D. A. Gough, Y. Fainman, and J. H. Price, “In-vivo confocal microscopy based on the Texas Instruments digital micromirror device,” in Optical Diagnostics of Living Cells III (2000).
16. Q. Wang, J. Zheng, K. Wang, K. Gui, H. Guo, and S. Zhuang, "Parallel detection experiment of fluorescence confocal microscopy using DMD," Scanning 38, 234 (2016).
17. M. Hillenbrand, R. Weiss, C. Endrödy, A. Grewe, M. Hoffmann, and S. Sinzinger, "Chromatic confocal matrix sensor with actuated pinhole arrays," Appl. Opt. 54, 4927-4936 (2015).