CHEM 213 Instrumental Analysis Lab Lecture – Fe and Extraction

Preview:

Citation preview

CHEM 213Instrumental Analysis

Lab Lecture – Fe and Extraction

2

Chem 102 – used Crystal field theory to explain colour of coordination complexes

Chem 213 – use intensity of colour to determine concentration of metal ion

3 + Fe2+ Fe(bipy)32+

Bipyridyl(bipy)

red

Determination of Trace Iron (LM, p14-22)

Spectrophotometry – any method using light to measure chemical composition

3

P/P0 %T A 1 100 0 0.1 10 10.01 1 2

Fe: Concentration to Colour

Intensity of colour = Absorbance (A)

Measure absorbance of soln to determine the iron concentration.

Transmitted

Transmittance (T) = P/Po

Absorbance A = -log(T) = abc

a = absorptivity

b = pathlength

c = concentration

4

Spectrophotometer (Text p410, LM p19-21)

Spec 20D+pg 21

On/Off Zero control

Readout Mode

Sample comp.

Readout On/Off

Genesys 20pg 21

5

Spectrophotometer (Text p410, LM p19-21)

Sample cell matching LM p 19 (1 hour)- Need 8 test tubes with the same pathlength (b)- prepare dilute amaranth dye soln

- Set wavelength to 510 nm

- Insert test tube in sample compartment

- Adjust trans/absorb control to 90%

- Insert 2nd tube in spec

- Put on alignment line

- Rotate tube to get T=90%

- Repeat until you have 8 matched tubes

Record spectrometer number

- Mark tube with line

6

• Reduce any Fe(III) to Fe(II) so all iron is in the Fe(II) state, because only Fe(II) forms coloured complex

Determination of Trace Iron (LM, p14-22)

3 + Fe2+ Fe(bipy)32+

Bipyridyl(bipy)

red

1. Fe3+acidic

2. xs + Fe2+ Fe(bipy)32+

3. Adjust pH to ~ 5 by adding sodium acetate

+ xs NH2OH Fe2+

hydroxylamine reducing agent

7

Determination of Trace Iron(LM, p14-22)

Measure absorption spectrum of the complex to determine the wavelength of max. absorbance.

Absorbance spectrum – graph of absorbance of light vs. wavelength

A = abc

Measure A for most conc standard from 640 to 400 nm at 20 nm intervals.

The 0 %T (20D Only) and the 100 %T (A = 0) settings must be reset for each new wavelength.

8

Why is 100 %T (A = 0) reset at each wavelength?

Determination of Trace Iron (LM, p14-22)

Light intensity (Po) varies with wavelength)

A is maxBest sensitivity

9

Standards = pipet 5, 10, 15 & 20 mL Fe3+ into flasks

Blank = soln with all of the reagents but no analyte

[Fe3+]=0

Zero (set A =0) instrument using blank

Measure absorbance of all standards and samples

Re-zero

Take second absorbance reading for all stds and samples

Prepare Calibration Curve

10

Prepare Calibration Curve

Blank

standards

Asample

Csample

11

Calibration Curves LM, Appendix A

12

Calibration Curves

Calibration Curve – graph of instrument response vs. conc of analyte

- when response for sample is measured, its conc is determined by interpolating from the graph

- blank value should be part of the calibration curve, i.e. signal for [C] = 0.0

-standards must cover the complete range of expected unknowns

- plot signal on the Y axis and conc on the X axis.

- Normally response is linear

or with math transformation can be made linear

Signal = m(conc) + b

m = slope b = intercept

13

Drawing a Calibration Curve

Manually on graph paper

14

0 1 2 3 4 50

0.1

0.2

0.3

0.4

0.5

0.6

Concentration, ppm

Inst

Res

po

nse

Useful for spotting:

Manual Graphing of Calibration Curve

15

0 1 2 3 4 50

0.1

0.2

0.3

0.4

0.5

0.6

Concentration, ppm

Inst

Res

po

nse

Useful for spotting: outliers

Manual Graphing of Calibration Curve

16

0 1 2 3 4 50

0.1

0.2

0.3

0.4

0.5

0.6

Concentration, ppm

Inst

Res

po

nse

Manual Graphing of Calibration Curve

Useful for spotting: outliers curvature

17

Mathematically calculate line using least squares

fits data to y = mx + b ?

determines m and b

determines error in m and b?

allows accurate calculation of sample conc

Manually on graph paper

- best for spotting outliers, curvature, etc.

- disgard outliers

Manual Graphing of Calibration Curve

Calculate least squares line

18

Solvent extraction

Dextromethorphancough suppressant

Expectorant Glyceryl guaiacolate

Antihistamines Chloropheniramine maleate Diphenyhydramine HCl

Decongestants Phenylephrine HCl

Analgesics Acetaminophen

Preservatives

Flavouring

Colouring

Determination of Dextromethorphan in Cough Syrup

toluenewater

19Dextromethorphan

DMH+

Bromothymol blueBTB-

+ (DMH+BTB-)

ion pair

(DMH+ BTB-)org

yellow

B. Solvent Extraction System

Ion-association Extraction

water

toluene

Colour = Absorbance @ 400 nm [BTB-]org

xs

[DMH+]aq

Calibration curve: (absorbance-blank) vs. [DMH+]aq

20

First week solution preparation (pg 50)

Do NOT add BTBDo NOT prep sample

21

b) Mohr pipet – use like a buret

d) Fluted filter paper – allows liquid to pass through quickly

Recommended