Care and Maintenance of the Liko: Mobile Lifts Ceiling Lifts Sit to Stand Lifts Lift Stretchers
English7EN180818-02
2013-01-23
This instruction applies to Liko mobile lifts, ceiling lifts, sit to stand lifts, and lift stretchers, yet in this text we refer only to lifts.
To avoid possible transmission of infectious disease, it is important to clean the lift if it has become soiled or if it is to be used by another patient. Also refer to your facility’s cleaning and disinfection policies for any additional guidelines on frequency.
It is important to follow the recommendations in this manual to maintain the warranty and expected life time.
This manual gives detailed cleaning and disinfecting procedures for Liko lifts. The instructions in this manual do not replace your facility’s cleaning and disinfection policies.
This manual is intended for use by facility-approved persons only. To ignore this restriction could cause severe injury to people and serious damage to equipment.
Liko/Hill-Rom recommends that the lift be disinfected prior to first use.
Liko/Hill-Rom recommends to disinfect the lift prior to disposal in accordance with local environmental regulations.
Liko’s products undergo continuous development, which is why we reserve the right to make product changes without prior notice. The only warranty Liko/Hill-Rom makes is the express written warranty extended on the sale or rental of its products.
The information contained in this manual is subject to change without notice. Visit our website www.liko.com for updated versions of this manual or contact your local Liko/Hill-Rom representative for advice or more information.
Safety recommendations• Wear protective equipment according to manufacturer’s instruction and per facility protocol throughout the
cleaning operations, such as: rubber gloves, goggles, apron, face mask and shoe covers.• Unplug mains (AC power source) before cleaning and disinfection.• Never clean the lift by pouring water on it, steam cleaning it, or by using a high-pressure jet. • Refer to the recommendations made by the cleaning and disinfecting product manufacturer.
Inadequate cleaning may increase risk of cross contamination!
27EN180818-02 www.liko.com
App
licat
ion
of c
omm
only
use
d C
lean
ing
/ Dis
infe
ctan
ts o
n Li
ko p
rodu
cts
Che
mic
al c
lass
Act
ive
ingr
edie
ntpH
Cle
aner
s /
Dis
infe
ctan
t *)
Man
ufac
ture
r *)
May
not
be
used
on
the
follo
win
g ite
ms:
Qua
tern
ary
amm
oniu
m c
hlor
ide
Did
ecyl
dim
ethy
l am
mon
ium
chl
orid
e =
8.70
4%
Alk
yl d
imet
hyl b
enzy
l am
mon
ium
chl
orid
e =
8.19
%9.
0 –
10.0
in u
seVi
rex
II (2
56)
John
son/
Div
erse
yFo
ot re
st fo
r Sab
ina™
an
d R
oll-O
n™
Qua
tern
ary
amm
oniu
m c
hlor
ide
Alk
yl d
imet
hyl b
enzy
l am
mon
ium
chl
orid
e =
13.2
38%
A
lkyl
dim
ethy
l eth
ylbe
nzyl
am
mon
ium
chl
orid
e =
13.2
38%
9.5
in u
seH
B Q
uat 2
5L3M
Acc
eler
ated
Hyd
roge
n P
erox
ide
Hyd
roge
n P
erox
ide
0.1
-1.5
%B
enzy
lAlc
ohol
: 1-5
%H
ydro
gen
Per
oxid
e 0.
1 -1
.5%
Ben
zylA
lcoh
ol: 1
-5%
3O
xivi
r Tb
John
son/
Div
erse
yTh
e lif
t str
aps
for
Gol
vo™
and
cei
ling
lifts
Phe
nolic
Orth
o-P
heny
lphe
nol =
3.4
0%
Orth
o-B
enzy
l-par
a-C
hlor
ophe
nol =
3.0
33.
1 +/
- 0.4
in u
seW
exci
deW
exfo
rd L
abs
Ble
ach
Sod
ium
hyp
ochl
orite
12.2
Dis
patc
hC
alte
chTh
e lif
t str
aps
for
Gol
vo™
and
cei
ling
lifts
Alc
ohol
Isop
ropy
l alc
ohol
= 7
0%5.
0 –
7.0
Vira
guar
dVe
ridie
nH
and
cont
rols
for a
ll lif
tsQ
uate
rnar
y am
mon
ium
n-A
lkyl
dim
ethy
l ben
zyl a
mm
oniu
m c
hlor
ides
= 0
.105
%n-
Alk
yl d
imet
hyl e
thyl
benz
yl a
mm
oniu
m c
hlor
ides
= 0
.105
%11
.5 -
12.5
CS
IC
entra
l Sol
utio
ns In
c.Vi
king
™, L
iko
M22
0™,
Liko
M23
0™, U
no™
, Sa
bina
™, G
olvo
™,
Liko
Ligh
t™, R
oll-O
n™,
Liko
rall™
, Mul
tiral
l™B
enzy
l-C12
-18-
alky
ldim
ethy
lam
mon
ium
, ch
lorid
esB
enzy
l-C12
-18-
alky
ldim
ethy
lam
mon
ium
, chl
orid
es (2
2 %
)2-
Phe
noxy
etha
nol (
20 %
)Tr
idec
ylpo
lyet
hyle
ngly
cole
ther
(15
%)
Pro
pan-
2-ol
(8 %
)
appr
ox 8
.6 in
use
Terr
alin
Pro
tect
Shü
lke
Foot
rest
for S
abin
a™
and
Rol
l-On™
Org
anic
per
oxid
e (ty
pe E
, sol
id)
Mag
nesi
um m
onop
erox
ypht
alat
e he
xahy
drat
e (5
0-10
0%)
Ani
onic
sur
fact
ant (
5-10
%)
Non
ioni
c su
rfact
ant (
1-5%
)
5.3
in u
seD
ism
ozon
Pur
Bod
eTh
e lif
t str
aps
for
Gol
vo™
and
cei
ling
lifts
Eth
anol
Hyd
roge
n pe
roxy
de (2
.5-1
0 %
)La
uryl
dim
ethy
lam
ine
oxid
(0-2
.5 %
)E
than
ol (2
.5-1
0 %
)
7A
niox
y-S
pray
WS
Ani
osC
ontr
ol b
ox fo
r all
mob
ile li
fts
Troc
lose
ne s
odiu
mA
dipi
c ac
id 1
0-30
%A
mor
phou
s si
lica
< 1%
Sod
ium
Tol
uene
sul
phon
ate
5-10
%Tr
oclo
sene
sod
ium
10-
30 %
4-6
in u
seC
hlor
-Cle
anG
uest
Med
ical
Ltd
The
lift s
trap
s fo
r G
olvo
™ a
nd c
eilin
g lif
ts
*) O
r equ
ival
ent
37EN180818-02 www.liko.com
Cleaning and Disinfection RecommendationsAll parts must be kept clean at all times in order to avoid the risk of infection. All necessary precautions must be taken to remove all visible stains and soiling.The cleaning and disinfecting method described below applies specifically to this device and all its accessories and is designed to save time and to help combat nosocomial infections more effectively.
Equipment: • Protective equipment (such as: rubber gloves, goggles, apron, face mask and shoe covers) as recommended
by the facility protocol and manufacturers instructions• Clean buckets• Cloths for washing and drying• Soft brush • Warm water• To find Cleaning / Disifectants compatible or not compatible for use on Liko’s products, follow the ”Application
of commonly used Cleaning / Disinfectants on Liko products” at page 2 in this document.
47EN180818-02 www.liko.com
Viking™ liftCleaning Instruction
1. Unplug mains (AC power source) before cleaning and disinfection. 2. Clean the lift with a moist cloth, using warm water and a facility approved neutral detergent. A soft brush can be
used to remove stains and resistant soil.3. Wipe down the entire lift starting from the top down. Do not use a cloth that is dripping wet. To have access to all
areas run the lift into the highest and lowest positions and extend the base width adjustment entirely in and out. Remove the Battery to have access behind the battery.
NOTE! Do not clean the piston rod! 4. Pay special attention to the following areas: • Sling bar (different designs) • Mechanical emergency lowering (different designs) • Handles • Hand control (different designs) • Emergency stop • Operation panel/display (where applicable) • Lever for base width adjustment (where applicable) • Pedal for base width adjustment (where applicable) • Locking handles • Wheels
Wheels
Sling bar Handles
Locking handle
Emergency stop
Mechanical emergency
lowering
Lever for base width adjustment
DisinfectRepeat the cleaning instructions after reading the ”Application of commonly used Cleaning / Disifectants on Liko products” at page 2 in this document. Follow the manufacturer’s instructions for proper use of the disinfectant.After you disinfect the lift, wipe the lift with a moist cloth using clean water to remove all disinfectant residue.
The lift may not be cleaned with CSI or equivalent. The hand control may not be cleaned with Viraguard or equivalent. The control box may not be cleaned with Anioxy Spray or equivalent.
BatteryHand control
Pedal for base width adjustment
Do not clean the piston rod!
Control box
Operation panel/ display
57EN180818-02 www.liko.com
Liko M220™/Liko M230™ liftsCleaning Instruction
1. Unplug mains (AC power source) before cleaning and disinfection. 2. Clean the lift with a moist cloth, using warm water and a facility approved neutral detergent. A soft brush can be
used to remove stains and resistant soil.3. Wipe down the entire lift starting from the top down. Do not use a cloth that is dripping wet. To have access to all
areas run the lift into the highest and lowest positions and extend the base width adjustment entirely in and out. Remove the Battery to have access behind the battery.
NOTE! Do not clean the piston rod! 4. Pay special attention to the following areas: • Sling bar • Mechanical emergency lowering • Handles • Hand control • Emergency stop • Pedal for base width adjustment (where applicable) • Wheels
DisinfectRepeat the cleaning instructions after reading the ”Application of commonly used Cleaning / Disifectants on Liko products” at page 2 in this document. Follow the manufacturer’s instructions for proper use of the disinfectant.After you disinfect the lift, wipe the lift with a moist cloth using clean water to remove all disinfectant residue.
The lift may not be cleaned with CSI or equivalent. The hand control may not be cleaned with Viraguard or equivalent. The control box may not be cleaned with Anioxy Spray or equivalent.
Pedal for base width adjustment
Hand control
Do not clean the piston rod!
Sling bar
Mechanical emergency lowering Handles
Emergency stop
Wheels
Battery
Control box
67EN180818-02 www.liko.com
Uno™ liftCleaning Instruction
1. Unplug mains (AC power source) before cleaning and disinfection. 2. Clean the lift with a moist cloth, using warm water and a facility approved neutral detergent. A soft brush can be
used to remove stains and resistant soil.3. Wipe down the entire lift starting from the top down. Do not use a cloth that is dripping wet. To have access to all
areas run the lift into the highest and lowest positions and extend the base width adjustment entirely in and out. Remove the Battery to have access behind the battery.
NOTE! Do not clean the piston rod! 4. Pay special attention to the following areas: • Sling bar (different designs) • Mechanical emergency lowering • Handles • Hand control • Emergency stop • Lever for base width adjustment (where applicable) • Locking handles • Wheels
Sling barHandles
Locking handles
Mechanical emergency lowering
Emergency stop
Wheels
Hand control
Lever for base width adjustment
DisinfectRepeat the cleaning instructions after reading the ”Application of commonly used Cleaning / Disifectants on Liko products” at page 2 in this document. Follow the manufacturer’s instructions for proper use of the disinfectant. After you disinfect the lift, wipe the lift with a moist cloth using clean water to remove all disinfectant residue.
The lift may not be cleaned with CSI or equivalent. The hand control may not be cleaned with Viraguard or equivalent. The control box may not be cleaned with Anioxy Spray or equivalent.
BatteryControl box
Do not clean the piston rod!
77EN180818-02 www.liko.com
Sabina™ liftCleaning Instruction
1. Unplug mains (AC power source) before cleaning and disinfection. 2. Clean the lift with a moist cloth, using warm water and a facility approved neutral detergent. A soft brush can be
used to remove stains and resistant soil.3. Wipe down the entire lift starting from the top down. Do not use a cloth that is dripping wet. To have access to all
areas run the lift into the highest and lowest positions and extend the base width adjustment entirely in and out. Remove the Battery to have access behind the battery. The calf strap may be machine washed 60-80°C (140-176°F), see ”Care and Maintenance of Liko Slings” for more information.
NOTE! Do not clean the piston rod! 4. Pay special attention to the following areas: • Sling bar (different designs) • Side support (where applicable) • Handles • Mechanical emergency lowering (different designs) • Hand control • Emergency stop • Lower-leg support • Foot rest • Locking handles • Wheels • Lever for base width adjustment (where applicable)
Lever for base width adjustment
DisinfectRepeat the cleaning instructions after reading the ”Application of commonly used Cleaning / Disifectants on Liko products” at page 2 in this document. Follow the manufacturer’s instructions for proper use of the disinfectant.After you disinfect the lift, wipe the lift with a moist cloth using clean water to remove all disinfectant residue.
The lift may not be cleaned with CSI or equivalent. The hand control may not be cleaned with Viraguard or equivalent. The control box may not be cleaned with Anioxy Spray or equivalent. The foot rest may not be cleaned with Terralin Protect, Virex II or equivalent. The leg support on Sabina II, S/N up to 460899 and leg support on Sabina 200, S/N up to 3501599 may only be cleaned with warm water and a facility approved neutral detergent. The leg support on Sabina II S/N from 460900 and leg support on Sabina 200, S/N from 3501600 may be cleaned with the recommended disinfectants.
Side support
Sling bar
Lower-leg support
Foot restLocking handles
Locking handles
Wheels
Handles
Emergency stop
Hand control
Battery
Control box
Mechanical emergency
lowering
Calf strap
Product label with serial no. (S/N)
Do not clean the piston rod!
87EN180818-02 www.liko.com
Lift strap
Mechanical emergency lowering
Arm rest
Golvo™ liftCleaning Instruction
1. Unplug mains (AC power source) before cleaning and disinfection. 2. Clean the lift with a moist cloth, using warm water and a facility approved neutral detergent. A soft brush can be
used to remove stains and resistant soil.3. Wipe down the entire lift starting from the top down. Do not use a cloth that is dripping wet. To have access to
all areas run the lift into the highest and lowest positions and extend the base width adjustment entirely out. Remove the Battery to have access behind the battery. Lower the sling bar to get access to the whole lift strap.Fully extend the lift strap by using the emergency lowering. After cleaning the lift strap, and before you raise the sling bar, make sure the lift strap has dried. After the emergency lowering device is used, the lifting level needs to be restored; see the Instruction Guide for the Golvo lift.
4. Pay special attention to the following areas: • Lift strap • Sling bar (different designs) • Arm rest • Mechanical emergency lowering • Hand control (different designs) • Handles • Operation panel (where applicable) • Emergency stop • Wheels
Sling bar
Handles
Wheels
Emergency stop
Operation panel
DisinfectRepeat the cleaning instructions after reading the ”Application of commonly used Cleaning / Disifectants on Liko products” at page 2 in this document. Follow the manufacturer’s instructions for proper use of the disinfectant.After you disinfect the lift, wipe the lift with a moist cloth using clean water to remove all disinfectant residue.
The lift may not be cleaned with CSI or equivalent. The hand control may not be cleaned with Viraguard or equivalent. The control box may not be cleaned with Anioxy Spray or equivalent. The lift strap may not be cleaned with Oxivir Tb, Dispatch, Chlor-Clean, Dismozon Pur or equivalent.
Hand control
Control box
Battery
97EN180818-02 www.liko.com
LikoLight™ liftCleaning Instruction
1. Unplug mains (AC power source) before cleaning and disinfection. 2. Clean the lift with a moist cloth, using warm water and a facility approved neutral detergent. A soft brush can be
used to remove stains and resistant soil.3. Wipe down the entire lift starting from the top down. Do not use a cloth that is dripping wet. To have access to all
areas run the lift into the highest and lowest positions and extend the base width adjustment entirely in and out. Remove the Battery to have access behind the battery.
NOTE! Do not clean the piston rod! 4. Pay special attention to the following areas: • Sling bar (different designs) • Mechanical emergency lowering • Handles • Hand control • Emergency stop • Pedal for base width adjustment • Lever for base width adjustment (where applicable) • Locking handles • Wheels
Sling bar
Handles
Locking handles
Mechanical emergency lowering
Emergency stop
Wheels
Hand control
DisinfectRepeat the cleaning instructions after reading the ”Application of commonly used Cleaning / Disifectants on Liko products” at page 2 in this document. Follow the manufacturer’s instructions for proper use of the disinfectant.After you disinfect the lift, wipe the lift with a moist cloth using clean water to remove all disinfectant residue.
The lift may not be cleaned with CSI or equivalent. The hand control may not be cleaned with Viraguard or equivalent. The control box may not be cleaned with Anioxy Spray or equivalent.
Pedal for base width adjustment
Lever for base width adjustment
Battery
Control box
Do not clean the piston rod!
107EN180818-02 www.liko.com
Wheels
Torso cushion
Rear supports
HandlesSafety lockSeat frame
Lower-leg support
Foot rest
Locking handle
Locking handles
RollOn™ liftCleaning Instruction
1. Clean the lift with a moist cloth, using warm water and a facility approved neutral detergent. A soft brush can be used to remove stains and resistant soil.
2. Wipe down the entire lift starting from the top down. Do not use a cloth that is dripping wet.3. Pay special attention to the following areas: • Torso cushion • Safety lock • Handles • Seat frame • Raising handles • Rear supports • Locking handles • Lower-leg support • Foot rest • Wheels
DisinfectRepeat the cleaning instructions after reading the ”Application of commonly used Cleaning / Disifectants on Liko products” at page 2 in this document. Follow the manufacturer’s instructions for proper use of the disinfectant.After you disinfect the lift, wipe the lift with a moist cloth using clean water to remove all disinfectant residue.
The lift may not be cleaned with CSI or equivalent. The foot rest may not be cleaned with Terralin Protect or Virex II or equivalent.
Raising handles
117EN180818-02 www.liko.com
Likorall™ liftCleaning Instruction
1. Unplug mains (AC power source) before cleaning and disinfection. 2. Clean the lift with a moist cloth, using warm water and a facility approved neutral detergent. A soft brush can be
used to remove stains and resistant soil.3. Wipe down the entire lift starting from the top down. Do not use a cloth that is dripping wet. Lower the sling bar to
get access to the whole lift strap. After cleaning the lift strap and before you raise the sling bar, make sure the lift strap has dried.
The Likorall lift with mechanical emergency lowering: fully extend the lift strap by using the mechanical emergency lowering. After cleaning the lift strap and before you raise the sling bar, make sure the lift strap has dried. After the mechanical emergency lowering device is used, the lifting level needs to be restored; see the Instruction Guide for the Likorall lift.
4. Pay special attention to the following areas: • Lift strap • Electrical emergency lowering/raising • Emergency stop • Emergency stop cord • Mechanical emergency lowering (where applicable) • Sling bar (different designs) • Hand control (different designs) • Electrical emergency lowering device (where applicable)
DisinfectRepeat the cleaning instructions after reading the ”Application of commonly used Cleaning / Disifectants on Liko products” at page 2 in this document. Follow the manufacturer’s instructions for proper use of the disinfectant.After you disinfect the lift, wipe the lift with a moist cloth using clean water to remove all disinfectant residue.
The lift may not be cleaned with CSI or equivalent. The hand control may not be cleaned with Viraguard or equivalent. The lift strap may not be cleaned with Oxivir Tb, Dispatch, Chlor-Clean, Dismozon Pur or equivalent.
Lift strap
Mecanical emergency lowering
Sling bar
Emergency stop cord
Emergency stop
Electrical emergency lowering device
Electrical emergency lowering/raising
Hand control
127EN180818-02 www.liko.com
DisinfectRepeat the cleaning instructions after reading the ”Application of commonly used Cleaning / Disifectants on Liko products” at page 2 in this document. Follow the manufacturer’s instructions for proper use of the disinfectant.After you disinfect the lift, wipe the lift with a moist cloth using clean water to remove all disinfectant residue.
The lift may not be cleaned with CSI or equivalent. The hand control may not be cleaned with Viraguard or equivalent. The lift strap may not be cleaned with Oxivir Tb, Dispatch, Chlor-Clean, Dismozon Pur or equivalent.
Multirall™ liftCleaning Instruction
1. Unplug mains (AC power source) before cleaning and disinfection. 2. Clean the lift with a moist cloth, using warm water and a facility approved neutral detergent. A soft brush can be
used to remove stains and resistant soil.3. Wipe down the entire lift starting from the top down. Do not use a cloth that is dripping wet. Lower the sling bar to
get access to the whole lift strap. After cleaning the lift strap and before you raise the sling bar, make sure the lift strap has dried.
4. Pay special attention to the following areas: • Electrical emergency lowering/raising • Mechanical emergency lowering • Emergency stop • Handles • Lift strap • Hand control • Sling bar (different designs)
Electrical emergency lowering/raising
Lift strap
Sling bar
Emergency stop
Hand control
Handles
Mechanical emergency lowering
137EN180818-02 www.liko.com
Lift StretchersCleaning Instruction
1. Clean the stretcher with a moist cloth, using warm water and a facility approved neutral detergent. A soft brush can be used to remove stains and resistant soil.
2. Wipe down the entire stretcher starting from the top down. Do not use a cloth that is dripping wet.
DisinfectRepeat the cleaning instructions after reading the ”Application of commonly used Cleaning / Disifectants on Liko products” at page 2 in this document. Follow the manufacturer’s instructions for proper use of the disinfectant.After you disinfect the stretcher, wipe the stretcher with a moist cloth using clean water to remove all disinfectant residue.
The stretcher may not be cleaned with CSI or equivalent.
Liko FlexoStretch™/Liko UltraStretch™
LikoStretch 1800™
LikoStretch Mod 600 IC™
LikoStretch 1900™
Liko OctoStretch™
www.l iko.com