BIOPUREWater purification systems for healthcare applications
Page154
Cleaning and sterilizing reusable medical equipment is becoming increasingly regulated by industry guidelines and international standards, as concern grows over infection control in hospitals and the spread of hepatitis, CJD, MRSA and other resistant pathogens. In addition to protecting patients and staff by preventing pathogen cross-contamination, the equipment itself must also be protected from damage by the sterilization or disinfection process.
BIOPURE
“ Water is all we do. We know water.”
With this in mind we applied our expertise gained from 50 years of innovation in water purification technologies, to design our unrivalled BIOPURE range. Our reliable water purification systems employ the latest technologies to ensure a broad removal of impurities, continual bacterial destruction and endotoxin removal. A compact design with an in-built wrap-around reservoir ensures a minimal amount of precious hospital or surgery space is occupied. Our BIOPURE range is easy and cost-effective to operate and a self-disinfection program contributes towards straightforward maintenance. Peace of mind is provided by a unique E-key, which restricts access to the most critical functions.
Water is vital to many
Biopure water minimizes the contamination risk to patients
What is biopurity?
Our BIOPURE range purifies tap water to the water purity essential to meet the demands of the latest medical standards:
EN• 15883, HTM2030/2031, NHS MESc32, MDA/HIS, ISO 15883 Part 4, EN285
Typical biopure water quality specifications are:
Bacteria Total Viable Count of • <10 CFU/100ml
Endotoxin levels of <• 0.25 EU/ml
Conductivity of <• 30 μS/cm
Water quality is maintained by:
Regularly sanitizing and validating • rinsewater systems to ensure they continue to meet the water specification
Routinely taking water samples to • demonstrate compliance
These water quality guidelines and standards serve to minimize the risk of cross infection to patients from a range of infectious pathogens.
Page155
1-2BIO
PURE
BIOPURE product finder
* For more information on applications and scientific water standards see our education guidebook “The Pure LabWater Guide”. A free copy is available at www.elgalabwater.com
120
200
600
60
30
15
7
BIO
PURE
7/15
BIO
PURE
60/
120
BIO
PURE
200
/300
/600
Matching systems to applications supporting you all the wayThis brochure outlines the key features of all the water purification products in our BIOPURE range. This product family is designed to deliver compliant purified (biopure) water at the point of use for a wide range of washer disinfectors, endoscope reprocessors and autoclaves. Whether you require a single bench top or wall mounted unit, or you need to feed a suite of multiple washers, we have a model that meets your needs.
To help you choose the system that is most suited to your requirements we have developed the BIOPURE product finder. Our teams of technical and application specialists are always available to answer any questions as well as provide friendly, expert advice in choosing the most suitable water purification systems.
Maximum water rate liters/hr
Water quality* Biopure Biopure Biopure
Water quality Conductivity µS/cm <1 <5 <30
Water quality CFU/10ml Bacteria <1 Bacteria <1 Bacteria <1
Water quality EU/ml Endotoxin <0.25 Endotoxin <0.25 Endotoxin <0.25
Typical applications
Small bench-top washer disinfectors and autoclaves
Small washer disinfectors and endoscope reprocessors
Large or multiple automated endoscope reprocessors and/or washer disinfectors
Page156
Every BIOPURE comes complete with one extra feature – first class service and support wherever you are in the world. ELGA has installed many thousands of systems globally, and our service engineers will apply their expertise to the installation, validation and maintenance of your water purification systems in compliance with all the relevant codes. Visit www.elgalabwater.com to find your nearest contact.
Service excellence
We aim to provide first-time-fix service•
Preventative maintenance contracts are • structured to match your precise needs, thereby minimizing interruptions to your workflow and sustaining system reliability
All our calibration equipment is maintained • to traceable standards and operated by fully trained service technicians, ensuring your water quality is consistently maintained at the required standard
Training
“Hands on” operation training is arranged • on or off site to ensure optimal system performance and minimize the risk of an interrupted workflow
For total confidence in your pure water systems
Worldwide service and support
Page157
3-4Service and Support
Technical assistance
Our specialized local team will ensure you • find the perfect product to match both your budget and applications
Our dedicated local help desk is always • available to provide advice, troubleshooting and parts identification
Online facility
We offer a user-friendly “Ask the Expert” • service and downloadable application information and maintenance planners
Download
Application information, including water • purity recommendations and system design advice
Consumable and maintenance planners • to ensure optimum cost efficiency and performance
Quality assured
Our commitment to the highest quality control processes guarantees reliability and compliance with both international, environmental and user organization standards.
Designed and manufactured under an ISO9001:2000 total quality system. Tested to comply with CE, EMC, EN 61010, ETL (UL CSA), PIRA, WEEE Directive and other standards as appropriate.
Page158
Designed to meet the latest medical standards
When high purity water is required for healthcare applications, the BIOPURE range offers a choice of systems with the flexibility to meet your needs.
General applications:
From small bench-top washer disinfectors and autoclaves and single endoscope reprocessor installations, to large or multiple automated endoscope reprocessors and washer disinfector installations
The BIOPURE solution for overcoming your water purification challenges
Challenge BIOPURE Solution
Your application demands biopure water at point-of-use
Integrated water purification technologies ensure broad removal of impurities, continual bacterial destruction and endotoxin removal. Ongoing control minimizes potential bacterial growth through simple and complete sanitization of all wetted parts. Water is continuously re-circulated through the purification technologies ensuring that microbial purity is optimized and biopurity maintained.
You need simple control and operation that includes easy maintenance
Easy-to-read display enables water quality, temperature and consumables to be viewed at a glance. An easy-to-use self-disinfection program ensures biopurity is effortlessly maintained.
Space is at a premium A compact design with unique wrap-around reservoir ensures a small footprint and maximizes the space available.
You do not want unauthorized users to access the control functions on your BIOPURE systems
A unique E-key means complete control over who has access to the most critical functions
Easy to monitor, operate and maintain
BIOPURE
“On demand delivery of biopure water”
“Designed for biopurity”
Page159
5-6BIO
PURE Range
BIOPURE 7/15
Provides 7 or 15 liters/hr of biopure water at a price that is comparable to purified bottled water without the hassle of continual ordering and storage. The BIOPURE 7/15 is cost-effective and is more environmentally friendly than traditional distillation units. The unit’s compact size allows it to be positioned on or under a bench, wall mounted or fitted into a cabinet. A manual fill option is available via a dispensing gun for filling autoclaves, washer disinfectors and manual washing.
BIOPURE 60/120
Cost-effectively provides 60 or 120 liters/hr of biopure water. The BIOPURE 60/120 ensures a continual supply of purified water for installations such as single washer disinfectors and endoscope reprocessors.
BIOPURE 200/300/600
For large or multiple automated endoscope reprocessors and washer disinfector installations the BIOPURE 200/300/600 cost-effectively provides biopure water at rates of 200 to 600 liters/hr.
High efficiency RO for broad removal of impurities
4
5
Optipure purification cartridge
Improves inorganic quality and • reduces endotoxin levels
6
6
UV sterilization
For continual bacterial • destruction
Molecular Ultrafilter provides final microbial and endotoxin removal
Full recirculation of water within the system and Docking Vessel optimizes microbial purity
8
7
9
9
8
7
Vent filter prevents contamination of reservoir
Easy-to-use monthly disinfection program
Designed for ease-of-use • by practice staff
3
5
3
2
Unique system validation and protection
Automated audio- • visual alarms
Communication
Interfaces via RS• 232 for communications and monitoring flexibility
1
2
1
4
“Tap water supply to biopure – a system sized to your needs”
Page160
LITR
387
70-0
3
www.elgalabwater.com
Commitment to global
citizenship
The Veolia Environment Foundation supports worldwide projects contributing to sustainable development, with a special focus on outreach, workforce development and the environment.
Since 2004, with an annual budget of €5 million, the Foundation has supported more than 450 projects.
(see www.fondation.veolia.com for details)
Contact us:
ELGA offices and distributors are located in more than 60 countries and are fully trained in all ELGA systems.
To find your nearest ELGA representative, go to www.elgalabwater.com and select your country for contact details.
ELGA Global Operations Centre
Tel: +44 1494 887 500 Fax: +44 1494 887 505
Email: info@elgalabwater.com Web: www.elgalabwater.com
ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions. © VWS (UK) Ltd. 2010 – All rights reserved. ELGA® and BIOPURE® are registered trademarks of VWS (UK) Ltd.
Your local ELGA representative
The LabWater Specialists Our commitment to providing water purification systems for washer disinfectors, endoscope reprocessors and autoclaves minimizes the contamination risk to patients by complying with the very latest medical standards.
ELGA is an integral part of Veolia Water Solutions and Technologies, the world’s leading water service company. Veolia Water Solutions and Technologies has a worldwide team of over 9,500 people and is renowned for its capabilities in providing water solutions to customers throughout the entire water cycle.
The ELGA team focuses exclusively on water and its purification. It continually contributes to the unique technical and scientific applications expertise developed during the past 50 years. We are experienced in meeting the challenges that arise during the development, installation and servicing of single point-of-use water purification systems as well as large projects involving consultation with architects, consultants and clients.
Page161
BIOPUREHEALTHCARE
Cost effective – provides the convenience of on-demand pure water at a price comparable with purified • bottled water
Space saving – the BIOPURE’S compact size means that it can be positioned under a bench, wall mounted • or fitted into a cabinet
Guaranteed bacterial specification, provided by integrated purification technologies combined with • easy system disinfection
Water specification is maintained in the reservoir by regular recirculation through the purification technologies•
The BIOPURE 7/15 has been designed to deliver compliant purified water at the point of use for a wide range of washer disinfectors and autoclaves to minimize contamination risk to patients.
BIOPURE 7/15
FeedwaterInlet
Boost Pump(optional)
Sanitisation Port
Pre-Treatment
ReverseOsmosis
ReverseOsmosis
Drain
Water Quality Sensor
Recirculation Pump
Outlet toDocking Vessel
or Reservoir
Feed fromDocking Vessel
or Reservoir
DispenseTap
UV Lamp
Drain
Outlet to application
UltrafiltrationMembrane
Optipure Purification
Cartridge
Water Quality Sensor
BIOPURE 7/15
Page162
BIOPUREHEALTHCARE
Treated Water SpecificationsMODEL BIOPURE 7 BIOPURE 15Make-up rate @15°C 7 l/hr 15 l/hrDelivery flow rate – maximum 2 l/min @ 15psi (1bar) 2 l/min @ 15psi (1bar)Inorganics (conductivity @25°C) <1 µS/cm <1 µS/cmBacteria - typical <1 CFU/10ml1 <1 CFU/10ml1
Endotoxin <0.25 EU/ml <0.25 EU/ml1System to be installed and regularly disinfected following ELGA LabWater installation design guidelines. Sample to be taken from hygienic sample valve and sampled in accordance with ELGA LabWater guidelines.
Dimensions and weights
Dimensions Height 460mm (18.1in), Width 550mm (21.7in), Depth 270mm (10.6in)Operational weight 20Kg 21KgInstallation Bench/Wall2 Bench/Wall2
2Install in a properly ventilated area.
Feedwater Requirement
Source Tap water as detailed belowConductivity <1400 µS/cm <1400 µS/cmContaminantHardness <250 ppm as CaCO3 <250 ppm as CaCO3
Free chlorine 0.5 ppm 0.5 ppmChloramine <0.2 ppm <0.2 ppmSilica <30 SiO2 ppm <30 SiO2 ppmFouling index <10 FI <10 FIIron/manganese <0.2 ppm <0.2 ppmOrganics N/A N/ATemperature 4 – 40°C (Recommended 15 - 25°C)Flowrate (requirement at 15°C) 1.3 l/min 1.4 l/minDrain requirements (gravity fall with air gap) 1.2 l/min 1.2 l/min
Feedwater Pressure
Maximum without internal boost pump 90psi (6.0bar) 90psi (6.0bar)Minimum without internal boost pump 60psi (4.0bar) 60psi (4.0bar)Maximum with internal boost pump 30psi (2.0bar) 30psi (2.0bar)Maximum without internal boost pump Flooded Suction Flooded Suction
Electrical Requirements
Mains Input 100-240V ac, 50-60HzSystem control voltage (not including pumps and UV) 24V dcPower consumption with boost pump 80VAPower consumption without boost pump 50VAElectrical protection rating 13 amp supply with suitable ELBNoise level <45 dBA
ELGA LabWaterTel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505Email: info@elgalabwater.com Website: www.elgalabwater.comELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions. © VWS (UK) Ltd. 2010 – All rights reserved. ELGA® is a registered trademark of VWS (UK) Ltd.
LITR 38764-02
Page163
ProtekPretreatment
Cartridge
Drain
FeedwaterInlet
UV Lamp
Air Gap
BoostPump
ReverseOsmosisModules
Optional E-Cartridge
VF
50 LitreReservoir
Delivery/ Recirculation
Pump
Optional Optipure
PurificationCartridge
UltrafiltrationModule
To Distribution
Loop
Return From Loop
Water Quality Sensor
Water Quality Sensor
BIOPUREHEALTHCARE
Reliable water quality – unique disinfection design for purification of RO membranes as well as the local • distribution loop
Guaranteed bacterial specification, provided by integrated purification technologies combined with easy • system disinfection
Space saving – “all in one” design allows easy under-bench installation •Economic design – unique “wrap around” reservoir •Unique system operation protection using automated alarms and access to important control systems •
limited by “E-key”
The BIOPURE 60/120 has been specifically designed to deliver compliant purified water at the point of use for a wide range of washer disinfectors and endoscope reprocessors to minimize contamination risk to patients.
BIOPURE 60/120
BIOPURE 60/120
Page164
BIOPUREHEALTHCARE
Treated Water SpecificationsMODEL BIOPURE 60 BIOPURE 120Make-up rate @15°C 60 l/hr 120 l/hrDelivery flow rate – maximum Up to 10 l/min @ 22.5psi (1.5bar) Up to 10 l/min @ 22.5psi (1.5bar)Inorganics (conductivity @25°C) <30 µS/cm1 <30 µS/cm1
Inorganics (conductivity @25°C) (Optipure Purification Cartridge installed) <5 µS/cm1 <5 µS/cm1
Bacteria - typical <1 CFU/10ml2 <1 CFU/10ml2
Endotoxin <0.25 EU/ml <0.25 EU/ml1Subject to a maximum feed water TDS of 500 ppm. For applications outside of this range, please contact Technical Support. 2System to be installed and regularly disinfected following ELGA LabWater installation design guidelines. Sample to be taken from hygienic sample valve and sampled in accordance with ELGA LabWater Guidelines.
Dimensions and weights
Dimensions Height 820mm (32.8in), 834mm (33.4in) including castors, Width 790mm (31.1in), Depth 480mm (18.9in)
Supply weight 58kg (127.7lb) 58kg (127.7lb)Operational weight 115kg (253.5lb) 115kg (253.5lb)Installation Floor mounted3 Floor mounted3
3Install in a properly ventilated area.
Feedwater RequirementSource Tap water as detailed belowConductivity <1400 µS/cmContaminant Hardness <250 ppm as CaCO3
Free Chlorine <1 ppm Cl2
Chloramine <0.2 ppm Cl2
Silica <30 ppm SiO2
Fouling Index <10 FIIron/manganese <0.05Fe/Mn ppmOrganics <3 ppm TOCTemperature 4 – 40°C (Recommended 15 - 25°C)Flowrate (requirement at 15°C) 10 l/min 10 l/minDrain requirements (gravity fall with air gap) 20 l/min 20 l/minFeedwater pressure 90psi (6bar) maximum, 30psi (2bar) minimum
Failure to comply with the feed water pretreatment recommendations will affect the life and performance of key components and may invalidate the warranty.
Feedwater requirements are specified for systems fitted with a Protek pretreatment cartridge.
If feedwater purity is variable or values are outside any of the above ranges, additional pretreatment is recommended to be installed in the feedwater supply to the BIOPURE system.
If in doubt seek advice from Technical Support at ELGA LabWater.
Electrical RequirementsMains Input 230V ac, 50HzSystem control voltage (not including pumps and UV) 24V dcPower consumption (peak demand) 800VAElectrical protection rating 13 amp supply with suitable ELBNoise level <60 dBA
ELGA LabWaterTel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505Email: info@elgalabwater.com Website: www.elgalabwater.comELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions. © VWS (UK) Ltd. 2010 – All rights reserved. ELGA® is a registered trademark of VWS (UK) Ltd.
LITR 38765-02
Page165
BIOPUREHEALTHCARE
Compact footprint to save space and facilitate easy installation•Guaranteed bacterial specification, provided by onboard technology combined with easy system disinfection •Reliable water quality – unique sanitization design for purification of RO membranes as well as the local •
distribution loop Unique system operation protection using automated alarms and access to important control systems •
limited by “E key”
The BIOPURE 200/300/600 features a unique, award winning design incorporating a very high output water purification system with a large 350 liter storage reservoir. The system is designed to feed large or multiple automated endoscope reprocessors and washer disinfector installations. The BIOPURE 200/300/600 cost effectively delivers compliant purified water at rates of 200, 300 and 600 liters/hr.
BIOPURE 200/300/600
ReverseOsmosis
SanitisationPort
BoostPump
Air GapPre-Filter
FeedwaterInlet
Drain
350 LitreReservoir
VF
Variable SpeedRecirculation
Pump
From Loop
To Distribution Loop
Hygienic Sample Valve
Ultra Filter
UV lamp
Water Quality Sensor
Water Quality Sensor
BIOPURE 200/300/600
Page166
BIOPUREHEALTHCARE
Treated Water SpecificationsMODEL BIOPURE 200 BIOPURE 300 BIOPURE 600Make-up rate @15°C 200 l/hr 300 l/hr 600 l/hrDelivery flow rate – maximum Up to 30 l/min @ 52psi (3.5bar) Up to 30 l/min @ 52psi (3.5bar) Up to 30 l/min @ 52psi (3.5bar)Inorganics (conductivity @25°C) <30µS/cm <30µS/cm <30µS/cmBacteria - typical <1CFU/10ml2 <1CFU/10ml2 <1CFU/10ml2
Endotoxin <0.25 EU/ml <0.25 EU/ml <0.25 EU/ml1Subject to a maximum feed water TDS of 700 ppm. For applications outside of this range, please contact Technical Support. 2System to be installed and regularly disinfected following ELGA LabWater installation design guidelines. Sample to be taken from hygienic sample valve and sampled in accordance with ELGA LabWater Guidelines.
Dimensions and weightsDimensions Height 1820mm (71.7in)
Width 730mm (28.75in) Depth 890mm (35in)
Height 1820mm (71.7in) Width 730mm (28.75in)
Depth 890mm (35in)
Height 1820mm (71.7in) Width 756mm (28.75in) Depth 1100mm (35in)
Supply weight 180kg (396lb) 180kg (396lb) 210kg (463lb)Operational weight 530kg (1168lb) 530kg (1168lb) 560kg (1235lb)Installation Floor Mounted3 Floor Mounted3 Floor Mounted3
3Install in a properly ventilated area. All connections – Inlet, sanitization drain, reservoir drain, ringmain outlet and return are 3/4” BSPP (female).
Feedwater RequirementSource Tap water as detailed belowConductivity Conductivity <1400 µS/cmContaminant Measure Range Pretreatment
Calcium Ca ppm as CaCO3<250 None4
<250 Softner or use very low RO recovery4
Free chlorine Cl2 ppm<0.1 None
0.1 – 0.5 20 inch carbon block>0.5 Cylinder of carbon sized correctly to obtain <0.1ppm
Chloramine Cl2 ppm<0.1 None
0.1 – 0.5 20 inch high efficiency carbon5
>0.5 Cylinder of carbon sized correctly to obtain <0.1ppm
Silica SiO2 ppm <30 None>30 20 inch cartridge depth filter or use very low RO recovery5
Fouling index FI<10 None
10 to 20 20 inch cartridge depth filter>20 Back washable media filter with a minimum flow rate of 20 l/min
Iron/manganese Fe/Mn ppm<0.05 None>0.05 20 inch cartridge depth filter6
>0.1 Back-washable Fe filter7
Organics TOC ppm C<2 None
2 to 3 20 inch carbon block5
>3 Cylinder of carbon sized correctly for TOC demand5
Temperature 4 - 40°C (Recommended 15 - 25°C)Flowrate (requirement at 15°C) 20 l/minDrain requirements (gravity fall with air gap) 45 l/min
Feedwater pressure 150psi (10bar) maximum, 30psi (2bar) minimum4Check LSI, increase frequency of acid cleaning. 5Increase frequency of alkaline cleaning. 6Please use ELGA Part Numbers LA683 (Filter Housing) and LC172 (Carbon Filter). 7Increase frequency of acid cleaning.
Electrical RequirementsMains Input 230V ac, 50Hz/60HzSystem control voltage (not including pumps and UV) 24V dc
Power consumption (peak demand) 2200VAElectrical protection rating 20 ampsNoise level during recirculation <70dBA
ELGA LabWaterTel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505Email: info@elgalabwater.com Website: www.elgalabwater.comELGA® is the global laboratory water brand name of Veolia Water. VWS (UK) Ltd. Registered in England & Wales No. 327847 ©Copyright 2009 ELGA LabWater/VWS (UK) Ltd. All rights reserved. As part of our policy of continual improvement we reserve the right to alter the specifications given in this datasheet.