View
248
Download
0
Category
Preview:
Citation preview
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
1/90
User Guide
Site MasterCable and Antenna Analyzer
with Spectrum Analyzer
S331E, 2 MHz to 4 GHz
S332E, 2 MHz to 4 GHz, Spectrum Analyzer, 100 kHz to 4 GHzS361E, 2 MHz to 6 GHz
S362E, 2 MHz to 6 GHz, Spectrum Analyzer, 100 kHz to 6 GHz
Anritsu Company490 Jarvis DriveMorgan Hill, CA 95037-2809USA
Part Number: 10580-00252Revision: B
Published: January 2010Copyright 2009 Anritsu Company
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
2/90
WARRANTYThe Anritsu product(s) listed on the title page is (are) warranted against defects in materials andworkma nsh ip for one year from the da te of shipmen t.
Anritsus obligation covers repairing or replacing products which prove to be defective during thewarranty period. Buyers shall prepay transportation charges for equipment returned to Anritsu forwar ra nt y repairs. Obligat ion is limited to the original pu rchaser . Anr itsu is not liable for consequent ialdamages.
LIMITATION OF WARRANTYThe foregoing war ra nt y does not app ly to Anritsu conn ectors th at ha ve failed due to norma l wear. Also,the warranty does not apply to defects resulting from improper or inadequate maintenance by theBuyer, u na ut horized modificat ion or misuse, or operation outside of the en vironm enta l specifications ofthe product. No other warranty is expressed or implied, and the remedies provided herein are theBuyers sole and exclusive remedies.
DISCLAIMER OF WARRANTYDISCLAIMER OF WARRANTIES. TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLELAW, ANRITSU COMPANY AND ITS SUPPLIERS DISCLAIM ALL WARRANTIES, EITHEREXPRESS OR IMPLIED, INCLUDING, BUT NOT LIMITED TO, IMPLIED WARRANTIES OFMERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, WITH REGARD TO THESOFTWARE PRODUCT. THE USER ASSUMES THE ENTIRE RISK OF USING THE PROGRAM.ANY LIABILITY OF PROVIDER OR MANUFACTURER WILL BE LIMITED EXCLUSIVELY TOPRODUCT REPLACEMENT.
NO LIABILITY FOR CONS EQU EN TIAL DAMAGES. TO TH E MAXIMUM EXTENT PE RMITTED BYAPPLICABLE LAW, IN NO EVENT SHALL ANRITSU COMPANY OR ITS SUPPLIERS BE LIABLEFOR ANY SPECIAL, INCIDENTAL, INDIRECT, OR CONSEQUENTIAL DAMAGES WHATSOEVER(INCLUDING, WITHOUT LIMITATION, DAMAGES FOR LOSS OF BUSINESS PROFITS,BUSINESS INTERRUPTION, LOSS OF BUSIN ESS IN FORMATION, OR ANY OTHER PE CUNIARYLOSS) ARISING OUT OF THE USE OF OR INABILITY TO USE THE SOFTWARE PRODUCTS,EVEN IF ANRITSU COMPANY HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGES.BECAUSE SOME STATES AND JURISDICTIONS DO NOT ALLOW THE EXCLUSION ORLIMITATION OF LIABILITY FOR CONSEQUENTIAL OR INCIDENTAL DAMAGES, THE ABOVELIMITATION MAY NOT APPLY TO YOU.
TRADEMARK ACKNOWLEDGMENTS
VxWorks is a registered trademark, and WindML is a trademark of Wind River Systems, Inc.BTS Maste r i s a t ra demark of Anri tsu Company.
NOTICEAnritsu Compan y has pr epared this ma nua l for u se by Anritsu Compan y personnel and customers as aguide for the proper installation, operation and maintenance of Anritsu Company equipment andcompu ter pr ogra ms. The dr awings, specificat ions, an d inform at ion cont ained h erein ar e the pr operty ofAnritsu Company, and any unauthorized use or disclosure of these drawings, specifications, andinform at ion is pr ohibited; they sh all not be repr oduced, copied, or u sed in wh ole or in par t a s th e basisfor manufacture or sale of the equipment or software programs without the prior written consent ofAnritsu Company.
UPDATES
Upda tes, if an y, can be downloaded from th e Document s ar ea of th e Anr itsu Website at :http://www.us.anritsu.com
For th e latest ser vice and sa les contact informa tion in your ar ea, please visit:http://www.anritsu.com/contact.asp
http://www.us.anritsu.com/http://www.us.anritsu.com/http://www.anritsu.com/Contact.asphttp://www.anritsu.com/Contact.asphttp://www.anritsu.com/Contact.asphttp://www.us.anritsu.com/7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
3/90
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
4/90
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
5/90
CE Conformity Marking
Anr itsu affixes t he CE Conform ity ma rk ing ont o its conform ing products in accorda nce with
Coun cil Directives of The Council Of The Eu ropean Commu nities in order to indicat e t ha t
th ese products conform to th e EMC a nd LVD directive of the E ur opean Un ion (EU).
C-tick Conformity Marking
Anr itsu affixes t he C-tick m ar king onto its conform ing products in a ccorda nce with t he
electr oma gnetic compliance regulat ions of Aust ra lia an d New Zealan d in order to indicat e
th at th ese products conform to the E MC regulat ions of Aust ra lia and N ew Zealan d.
Notes On Export Management
This product an d its man ua ls may require an Export License or a pproval by th e governm ent
of th e pr oduct count ry of origin for re-export from your coun tr y.
Before you export th is product or a ny of its m an ua ls, please cont act Anrits u Compan y to
confirm wheth er or n ot th ese items a re export -cont rolled.
When disposing of export -cont rolled items , the products an d m an ua ls need t o be broken or
shr edded to such a degree th at th ey can not be unlawfully used for m ilita ry pur poses.
Mercury Notification
This product us es an LCD backlight la mp t ha t cont ains m ercury. Disposal ma y be regulated
due t o environm ent al considera tions. Please cont act your local au th orities or, with in th e
Un ited St at es, th e Electronics Indu st ries Allian ce (www.eiae.org) for disposal or r ecycling
information.
VxWorks Runtime WindML Target License
NI Device License2000-1421
License 2000-1189
3-2000-1486WindRiver USB Runtime License
2000-1372
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
6/90
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
7/90
Site Master User Guide PN: 10580-00252 Rev. B Safety-1
Safety Symbols
To prevent th e risk of personal injury or loss relat ed to equipment ma lfun ction, Anr itsu
Compan y uses t he following symbols to indicat e safety-relat ed inform at ion. For your own
safety, please r ead t he inform at ion carefully before operating th e equipment .
Symbols Used in Manuals
Safety Symbols Used on Equipment and in ManualsThe following safety symbols ar e us ed inside or on th e equipment near opera tion locat ions to
provide inform at ion a bout safety items an d opera tion precau tions. Ensu re t ha t you clear ly
underst and th e meanings of the symbols and tak e th e necessary precaut ions before operating
th e equipm ent . Some or a ll of the following five symbols ma y or m ay n ot be used on a ll
Anritsu equipment. In addition, th ere ma y be other labels att ached to products t hat ar e not
shown in th e diagram s in this manu al.
This indicat es a pr ohibited opera tion. The prohibited operat ion is indicated
symbolically in or n ear th e bar red circle.
This indicates a compu lsory sa fety pr ecau tion. The requir ed opera tion is indicat edsymbolically in or nea r t he circle.
This indicates a war ning or caut ion. The cont ent s a re in dicat ed symbolically in or
near t he triangle.
This indicates a note. The cont ent s ar e described in t he box.
These indicat e tha t th e mar ked par t should be recycled.
Danger
This indicates a very dangerous procedure that could result in serious injury
or death, or loss related to equipment malfunction, if not performed properly.
Warning This indicates a hazardous procedure that could result in light-to-severe
injury or loss related to equipment malfunction, if proper precautions are not
taken.
Caution
This indicates a hazardous procedure that could result in loss related to
equipment malfunction if proper precautions are not taken.
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
8/90
Safety-2 PN: 10580-00252 Rev. B Site Master User Guide
For Safety
Warning Always refer to the operation manual when working near locations at
which the alert mark, shown on the left, is attached. If the operation,
etc., is performed without heeding the advice in the operationmanual, there is a risk of personal injury. In addition, the equipment
performance may be reduced. Moreover, this alert mark is sometimes
used with other marks and descriptions indicating other dangers.
Warning
When supplying power to this equipment, connect the accessory
3-pin power cord to a 3-pin grounded power outlet. If a grounded
3-pin outlet is not available, use a conversion adapter and ground thegreen wire, or connect the frame ground on the rear panel of the
equipment to ground. If power is supplied without grounding the
equipment, there is a risk of receiving a severe or fatal electric shock.
Warning
This equipment can not be repaired by the operator. Do not attempt to
remove the equipment covers or to disassemble internal
components. Only qualified service technicians with a knowledge of
electrical fire and shock hazards should service this equipment.
There are high-voltage parts in this equipment presenting a risk of
severe injury or fatal electric shock to untrained personnel. In
addition, there is a risk of damage to precision components.
Caution
Electrostatic Discharge (ESD) can damage the highly sensitive
circuits in the instrument. ESD is most likely to occur as test devices
are being connected to, or disconnected from, the instruments front
and rear panel ports and connectors. You can protect the instrument
and test devices by wearing a static-discharge wristband.
Alternatively, you can ground yourself to discharge any static chargeby touching the outer chassis of the grounded instrument before
touching the instruments front and rear panel ports and connectors.
Avoid touching the test port center conductors unless you are
properly grounded and have eliminated the possibility of static
discharge.
Repair of damage that is found to be caused by electrostaticdischarge is not covered under warranty.
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
9/90
Site Master User Guide PN: 10580-00252 Rev. B Contents-1
Table of Contents
Chapter 1General Information
1-1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11-2 Chapter Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1-3 Available Models. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1-4 Available Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1-5 Standard Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1-6 Optional Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1-7 Additional Documents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1-8 General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1-9 Site Master Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
1-10 Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
1-11 Calibration Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
1-12 Annual Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
1-13 ESD Caution. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
1-14 Battery Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
1-15 Soft Carrying Case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
1-16 Tilt Bail Stand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
1-17 Secure Environment Workplace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Site Master Memory Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Erase All User Files in Internal Memory . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Recommended Usage in a Secure Environment . . . . . . . . . . . . . . . . 1-10
Chapter 2Instrument Overview
2-1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2-2 Chapter Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2-3 Turning On the Site Master. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12-4 Front Panel Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Front Panel Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Touch Screen Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Keypad Menu Keys (1 to 9). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
LED Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
2-5 Display Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
2-6 Test Panel Connector Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
10/90
Contents-2 PN: 10580-00252 Rev. B Site Master User Guide
2-7 Symbols and Indicators. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Calibration Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Battery Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Additional Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
2-8 Data Entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
Numeric Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
Parameter Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
Text Entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
2-9 Mode Selector Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Chapter 3Quick Start Guide
3-1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3-2 Measurement Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3-3 Cable & Antenna Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Select the Measurement Type. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Set the Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Set the Amplitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Turn on Markers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Single Limit Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
DTF Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Calibrate with OSL Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
3-4 Spectrum Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Set Start and Stop Frequencies. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Enter the Center Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Select a Signal Standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Set the Measurement Frequency Bandwidth. . . . . . . . . . . . . . . . . . . . . 3-7
Set the Amplitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Power Offset Set Up for Compensating External Loss . . . . . . . . . . . . . 3-8
Set the Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Single Limit Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Segmented Limit Lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Create a Limit Envelope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10Setting Up Markers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Select a Smart Measurement Type . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
11/90
Site Master User Guide PN: 10580-00252 Rev. B Contents-3
3-5 Saving Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
3-6 Useful MST Utilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Converting Files to .DAT File Format . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Group Edit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Print All to PDF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
Chapter 4File Management
4-1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4-2 Managing Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Save Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Save Dialog Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Quick Name Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Recall Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Recall Dialog Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3Copying Files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Deleting Files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Delete Dialog Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4-3 File Menu Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
4-4 File Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Save Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Save Location Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Save On Event Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
Recall Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11Copy Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Delete Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
Chapter 5System Operations
5-1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
5-2 System Menu Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5-3 System Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
System Options Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Display Settings Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5Reset Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
5-4 Preset Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
5-5 Self Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
5-6 Updating Site Master Firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
5-7 Site Master Firmware Emergency Repair. . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
12/90
Contents-4 PN: 10580-00252 Rev. B Site Master User Guide
Chapter 6GPS (Option 31)
6-1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6-2 Chapter Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6-3 Activating the GPS Feature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16-4 Saving and Recalling Traces with GPS Information . . . . . . . . . . . . . . . . . . 6-3
Saving Traces with GPS Information. . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Recalling GPS Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
6-5 GPS Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Chapter 7Bias Tee (Option 10)
7-1 Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Chapter 8Master Software Tools
8-1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
8-2 MST Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
8-3 Feature Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
8-4 Installing MST. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
8-5 Connecting to the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
8-6 Updating Site Master Firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Appendix AMeasurement Guides
A-1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
Index
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
13/90
Site Master User Guide PN: 10580-00252 Rev. B 1-1
Chapter 1 General Information
1-1 Introduction
This chapt er pr ovides inform at ion a bout frequen cy ra nge, available options, additiona l
docum ent s, genera l overview, preventive main ten an ce, an d an nu al verificat ion r equiremen ts
for t he Anritsu Ha ndh eld S331E, S361E, S332E, and S362E Site Mast er models. Thr oughout
this m anu al, the t erm Site Master will refer to th e S331E, S361E, S332E, an d S362E.
1-2 Chapter Overview
Available Models on page 1-1
Available Options on page 1-2
Sta nda rd Accessories on page 1-2
Optiona l Accessories on page 1-3
Additiona l Docum ent s on page 1-3
Genera l Descript ion on page 1-3
Site Mast er Specificat ions on page 1-4
Pr eventive Maintena nce on page 1-4
Calibrat ion Requirements on page 1-4
Ann ua l Verificat ion on page 1-4
ESD Caution on page 1-5
Batt ery Replacement on page 1-5
Soft Carr ying Case on page 1-7
Tilt Bail Stan d on page 1-8
Secure En vironment Workplace on page 1-9
1-3 Available Models
Table 1-1 lists th e Site Mast er models and frequency ra nges described in th is User Guide.
Table 1-1. Site Master Models
Model Frequency Range
S331E Cable & Antenna Analyzer, 2 MHz to 4 GHz
S361E Cable & Antenna Analyzer, 2 MHz to 6 GHz
S332E Cable & Antenna Analyzer, 2 MHz to 4 GHz, Spectrum Analyzer, 100 kHz to 4 GHz
S362E Cable & Antenna Analyzer, 2 MHz to 6 GHz, Spectrum Analyzer, 100 kHz to 6 GHz
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
14/90
1-4 Available Options General Information
1-2 PN: 10580-00252 Rev. B Site Master User Guide
1-4 Available Options
Available options for th e Site Ma ster models a re sh own in Table 1-2.
1-5 Standard AccessoriesThe Anr itsu Sit e Master in cludes a one year war ra nt y which includes: batt ery, firmwa re,
softwa re, a nd Certificat e of Calibra tion a nd Conform an ce. The following items ar e su pplied
with t he product.
Table 1-2. Available Options
S331E S332E S361E S362E Description
S331E-0021 S332E-0021 S361E-0021 S362E-0021 2-Port Transmission Measurement
S331E-0010 S332E-0010 S361E-0010 S362E-0010Bias-Tee (Requires Option 0021 on
S331E and S361E)
S331E-0031 S332E-0031 S361E-0031 S362E-0031GPS Receiver (Requires Antenna
P/N 2000-1528-R)
S331E-0019 S332E-0019 S361E-0019 S362E-0019 High-Accuracy Power Meter
S332E-0029 S362E-0029 Power Meter
S332E-0025 S362E-0025 Interference Analyzer
S332E-0027 S362E-0027 Channel ScannerS332E-0090 S362E-0090 Gated Sweep
S332E-0028 S362E-0028C/W Signal Generator (Requires
CW Signal Generator Kit, P/N 69793)
S332E-0509 S362E-0509 AM/FM/PM Analyzer
S331E-0098 S332E-0098 S361E-0098 S362E-0098 Standard Calibration to Z540
S331E-0099 S332E-0099 S361E-0099 S362E-0099Premium Calibration to Z540 plus test
data
Table 1-3. Standard Accessories for Site Master Models
Part Number Description
10580-00252 Site Master User Guide
3-68736 Soft Carrying Case
2300-498
MST CD: Master Software Tools, User/Measurement Guides,
Programming Manual, Troubleshooting Guides, Application Notes,
Data Sheet
633-44 Rechargeable Li-Ion Battery
40-168-R AC-DC Adapter
806-141-R Automotive Cigarette Lighter 12 VDC Adapter
3-2000-1498 USB A/5-pin mini-B Cable, 10 feet/305 cm
11410-00484 Site Master S331E, S332E, S361E, S362E Technical Data Sheet
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
15/90
General Information 1-6 Optional Accessories
Site Master User Guide PN: 10580-00252 Rev. B 1-3
1-6 Optional Accessories
The Site Ma ster Techn ical Dat a S heet (P/N 11410-00484) cont ains a list a nd description of
available optional accessories. The da ta sheet is available on t he Ma ster Software Tools
CD-ROM provided with th e inst ru ment or t he Anr itsu web site: http://us.anritsu.com.
1-7 Additional Documents
This user guide is specific to the Sit e Mast er a nd includes a genera l description about t he
Site Master. For inform ation a bout Cable & Antenn a Measur ement, Spectr um Analysis,Int erference Ana lysis, 2-port Tr an smission Measu rem ent s, Power Meter, and Mast er
Softwa re Tools, refer to the individua l Measu remen t Guides listed in Appendix A,
Measurement Guides.
1-8 General Description
The Site Master S331E/S361E is a h and held cable and an tenna ana lyzer designed to mak e
Retur n Loss, VSWR, Cable Loss, an d Distan ce-To-Fa ult (DTF) measu rem ent s in t he field.
The cable and an tenna ana lyzer a lso includes 1-port phase an d smith char t m easuremen ts.
The 2-port tr an smission mea sur ement option includes two power levels and access to a
built-in 32 volt bias t ee (Option 10).
The S332E/S362E is an int egrat ed multi-functional t est instr ument tha t eliminates th e need
to carr y and learn m ultiple test sets. In addition t o th e cable & ant enna measur ements, the
S332E/S362E can be configur ed t o include a Spectr um Ana lyzer, 2-port Tra nsm ission
Measur ement , Inter ference Ana lyzer, Chann el Scan ner , CW Signa l Generat or, AM/FM/PM
Ana lyzer, Power Meter , and High Accur acy Power Met er. A GPS r eceiver can be added t o
both t he S331E/S361E an d th e S332E/S362E Site Master m odels.
The br ight 8.4" TFT color display pr ovides easy viewing in a va riet y of light ing cond itions a nd
th e combina tion of a t ouch screen a nd keypad ena bles user s to navigat e menu s with t he touch
screen and ent er num bers with t he keypad. All Site Master models ar e equipped with a
Li-Ion ba tt ery delivering m ore t ha n four hours of bat ter y life for th e S331E/S361E a nd m ore
th an t hr ee hour s of bat ter y life for t he S332E/S362E Site Mast ers.
The inter na l memory is large enough to store appr oxima tely 2,000 tra ces or set ups.
Measurement s an d setups can a lso be stored in a USB flash drive or t ran sferred to a P C using
th e included U SB cable.
Mast er Softwa re Tools (MST), a P C based softwa re pr ogram , can be used t o crea te r eport s,
view and organize data, an alyze historical dat a, add m ark ers a nd limit lines, rename t ra ces
and tr ace analysis. Please see C ha pter 8 for a brief overview of Mas ter Softwar e Tools an d th e
Mas ter Softwa re Tools User s Guide (.pdf file is available on th e MST CD-ROM) for a dditiona l
inform at ion. (.DAT files can be opened wit h t he legacy Ha ndh eld Softwa re Tools (HH ST)
application.)
Caution
When using the Automotive Cigarette Lighter 12 VDC Adapter, Anritsu PartNumber 806-141-R, always verify that the supply is rated for a minimum of
60 Watts at 12 VDC, and that the socket is clear of any dirt or debris. If the adapter
plug becomes hot to the touch during operation, then discontinue use immediately.
http://us.anritsu.com/http://us.anritsu.com/7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
16/90
1-9 Site Master Specifications General Information
1-4 PN: 10580-00252 Rev. B Site Master User Guide
1-9 Site Master Specifications
Refer to the Site Master Technical Data Sheet (P/N 11410-00484) for general specifications,
deta iled mea sur ement specificat ions for a ll available measu rem ent modes, order ing
inform at ion, power sen sors, an d available accessories. The dat a s heet is in cluded with th e
instr um ent an d also available on t he Ma ster Softwa re Tools CD-ROM. It is also available on
the Anritsu Website: http://us.anritsu.com.
1-10 Preventive Maintenance
Site Mast er prevent ive main ten an ce consists of cleaning th e unit a nd inspecting and cleanin g
th e RF conn ectors on th e instr um ent a nd all accessories. Clean th e Site Master with a soft,
lint-free cloth dam pened with wat er or wat er an d a m ild cleaning solut ion.
Clean the RF connectors a nd cent er pins with a cotton swab dam pened with denatu red
alcohol. Visua lly ins pect t he conn ectors. Th e finger s of th e N(f) connectors a nd t he pin s of t he
N(m) conn ectors sh ould be unbr oken a nd u niform in appea ra nce. If you a re u nsu re whet her
th e conn ectors ar e unda ma ged, gauge th e conn ectors to confirm th at th e dimensions a re
correct.
Visually inspect th e test port cable(s). The t est port cable should be uniform in appea ra nce,
an d not str etched, kinked, dented, or broken.
1-11 Calibration Requirements
Anr itsu recomm ends a nn ua l calibrat ion a nd per form an ce verificat ion by local Anritsuservice cent ers. The Cable and Antenn a Analyzer mode requir es calibrat ion sta nda rds for
OPE N, SHORT, an d LOAD (OSL) or I nst aCa l module, which a re sold separa tely.
1-12 Annual Verification
Anr itsu r ecomm ends a n a nn ua l calibrat ion a nd perform an ce verificat ion of th e Site Mast er
an d th e OSL calibrat ion component s an d Inst aCa l module by local Anritsu service cent ers.
The Site Ma ster is self-calibrat ing an d th ere a re n o field-adjust able component s. The OSL
calibrat ion component s ar e cru cial to the int egrity of the calibra tion. As a result , they mu st be
verified per iodically to ensu re per form an ce conformit y. This is especially import an t if theOSL calibration components have been accidentally dropped or over-torqued.
Cont act inform at ion for Anrit su S ervice Center s is available at:
http://www.anritsu.com/Contact.asp
Caution To avoid damaging the display or case, do not use solvents or abrasive cleaners.
http://us.anritsu.com/http://www.anritsu.com/Contact.asphttp://www.anritsu.com/Contact.asphttp://us.anritsu.com/7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
17/90
General Information 1-13 ESD Caution
Site Master User Guide PN: 10580-00252 Rev. B 1-5
1-13 ESD Caution
The Site Mast er, like oth er high perform an ce instr um ent s, is susceptible to electr osta tic
discha rge ( ES D) dam age. Coaxial cables an d an ten na s often build up a st at ic cha rge, which
(if allowed to discha rge by conn ecting directly to the Site Ma ster without discha rging th e
stat ic char ge) may dama ge the Site Master input circuitry. Site Master operat ors m ust be
awa re of th e poten tial for E SD dama ge an d ta ke all necessar y precau tions.
Operat ors should exercise practices out lined within indu str y sta nda rds su ch as J EDE C-625
(EIA-625), MIL-HDBK-263, and MI L-STD-1686, which per ta in t o ES D an d ESDS devices,
equipment , an d practices. Becau se these apply to th e Site Master, it is recomm ended th at an y
sta tic cha rges th at ma y be present be dissipat ed before conn ecting coaxial cables or a nt enna s
to the Site Mas ter . This may be as simple as tem pora rily att aching a short or load device to
the cable or a nt enna prior t o atta ching to the Site Master. It is importa nt to remember tha t
th e opera tor ma y also car ry a st at ic cha rge th at can cau se dama ge. Following the pra ctices
out lined in t he above sta nda rds will ensu re a sa fe environm ent for both personnel an d
equipment.
1-14 Battery Replacement
The bat ter y can be replaced without t he u se of tools. The bat ter y compa rt ment is locat ed on
th e lower left side of th e instr um ent (when you a re facing t he mea sur ement display). To
remove th e batt ery:
1. Slide th e cat ch t owar d th e bott om of th e inst ru men t
2. Pu ll the t op of the door a way from th e unit
3. Lift out t he bat ter y door.
4. Remove the batt ery pack from t he instr ument by grabbing th e batter y lanyar d an dpulling out .
Replacemen t is t he opposite of rem oval. The ba tt ery key side (slot below th e conta cts) sh ould
be facing the front on t he u nit a nd slide in first .
Note
When inserting the battery the battery label should face the back of the instrument
and the guide slot on the battery should be below the contacts. If the battery door
does not latch closed, the battery may be inserted incorrectly.
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
18/90
1-14 Battery Replacement General Information
1-6 PN: 10580-00252 Rev. B Site Master User Guide
The batt ery tha t is supplied with t he Site Master ma y need cha rging before use. The batt ery
can be cha rged while it is inst alled in th e Site Mast er by using either th e AC-DC Adapt er
(40-168-R) or th e 12-Volt DC a dap ter (806-141-R), or out side th e Site Ma st er wit h t he
optional Dua l Bat ter y Cha rger (2000-1374). Refer t o Batt ery Symbols on page 2-11for a
description of battery symbols.
Figure 1-1. Battery Compartment
Note
Use only Anritsu Company approved batteries, adapters, and chargers with this
instrument.
Anritsu Company recommends removing the battery for long-term storage of the
instrument.
Caution
When using the Automotive Cigarette Lighter 12 VDC Adapter, Anritsu Part
Number 806-141-R, always verify that the supply is rated for a minimum of 60
Watts @ 12 VDC, and that the socket is clear of any dirt or debris. If the adapter
plug becomes hot to the touch during operation, then discontinue use immediately.
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
19/90
General Information 1-15 Soft Carrying Case
Site Master User Guide PN: 10580-00252 Rev. B 1-7
1-15 Soft Carrying Case
The Site Ma ster can be opera ted wh ile in th e soft car rying case. On t he back of the case is a
large storage pouch for accessories and supplies.
To insta ll th e inst ru men t int o th e soft carr ying case:
1. The front pa nel of th e cas e is secur ed with h ook-an d-loop fast ener s. Fu lly close the frontpan el of the case. When closed, the front pan el supports th e sha pe of th e case while you
are inserting the Site Master.
2. Place th e soft car rying case face down on a sta ble surface, with t he front pa nel fully
closed a nd la ying flat .
3. Open t he zippered back of th e case.
4. Insert the inst rum ent face down into the case, tak e care t hat th e connectors a re
properly situ at ed in th e case t op opening. You m ay find it ea sier to insert th e
conn ectors first , then pull th e corn ers over t he bott om of th e Site Mast er.
5. Close the back panel an d secur e with t he zipper t o secur e the Site Mast er.
Note
The soft case has two zippers near the back. The zipper closer to the front of the
case opens to install and remove the instrument. The zipper closer to the back of
the case opens an adjustable support panel that can be used to provide support for
improved stability and air flow while the instrument is in the case. This support
panel also contains the storage pouch.
Figure 1-2. Instrument Inserted into the Soft Carrying Case
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
20/90
1-16 Tilt Bail Stand General Information
1-8 PN: 10580-00252 Rev. B Site Master User Guide
The soft car rying case includes a det achable shoulder st ra p, which can be conn ected t o the
D-rings of th e case.
1-16 Tilt Bail Stand
A Tilt Bail is at ta ched t o the back of th e Site Ma ster for desk top opera tion. The tilt ba il
pr ovides two sett ings of backwa rd t ilt for im pr oved st ability. To deploy th e tilt ba il, pu ll the
bott om of th e tilt bail away from th e back of th e instr um ent. To store th e tilt bail, push th e
bott om of the bail towar ds th e back of the inst rum ent unt il it a tt aches to the Site Master.
CautionThe soft case has panel openings for the fan inlet and exhaust ports. Do not block
the air flow through the panels when the unit is operating.
NoteDo not use the tilt bail while the instrument is in the soft case. The soft case has anadjustable support panel in the back zipper.
Figure 1-3. Tilt Bail Extended
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
21/90
General Information 1-17 Secure Environment Workplace
Site Master User Guide PN: 10580-00252 Rev. B 1-9
1-17 Secure Environment Workplace
This section det ails th e types of mem ory in t he Site Ma ster , how to delete stored user files in
inter na l memory, an d recomm ended usa ge in a secur e environmen t workplace.
Site Master Memory Types
The ins tr um ent cont ains non-volatile disk-on-a-chip m emory, EEP ROM, and volatile DRAM
memory. The instr ument is also supplied with an externa l USB flash drive. The inst rum ent
does not h ave a ha rd disk dr ive or an y oth er type of volatile or non-volatile mem ory.
Disk-On-A-Chip (DOC)
DOC is used for storage of instr um ent firmwa re, factory calibrat ion in form at ion, us er
mea sur ement s, setups, an d .jpg screen ima ges. User inform at ion st ored on the DOC is erased
by the ma ster reset process described below.
EEPROM
This memory stores th e model nu mber, serial num ber, and calibrat ion dat a for th einstru ment . Also stored here ar e th e user-set operating par amet ers su ch as frequency ran ge.
During the ma ster r eset process all operating para meter st ored in the EE PROM are set to
sta nda rd factory defau lt values.
RAM Memory
This is volat ile mem ory used t o store par am eters needed for t he n orm al opera tion of th e
instru ment along with curr ent m easurement s. This memory is reset whenever t he
instrument is restart ed.
External USB Flash Drive
This memory may be selected as the destination for saved measuremen ts a nd setu ps for theinstr um ent . The user can also copy the cont ent s of the int ern al disk-on-chip mem ory to th e
externa l flash m emory for storage or data tr ansfer. The extern al F lash USB can be
reform at ted or san itized usin g softwa re on a P C.
Refer t o the Chapter 4, File Management for additional information on saving and copying
files t o the U SB flash drive.
Erase All User Files in Internal Memory
Perform a Mast er Reset:
1. Turn th e instrum ent on.2. Press the Shift button then th e System (8) button.
3. Press the System Options submenu key.
4. Press the Reset key, then th e Master Reset key.
5. A dialog box will be displayed on t he screen war ning t ha t a ll settings will be ret ur ned t o
factory default values a nd all user files will be deleted. This deletion is a sta nda rd file
delete and does not involve overwriting exiting information.
6. Press the ENTER butt on t o complete the m aster reset.
7. The instr um ent will reboot a nd th e reset is complete.
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
22/90
1-17 Secure Environment Workplace General Information
1-10 PN: 10580-00252 Rev. B Site Master User Guide
Recommended Usage in a Secure Environment
Set t he Site Mast er to save files to the externa l USB Flash dr ive:
1. Atta ch th e external Flash drive and tu rn t he instrum ent on.
2. Press the Shift button th en the File (7) button.
3. Press the Save submenu k ey.
4. Press the Change Save Location submenu key, th en select t he USB drive with th e
rotary knob, Up/Down ar row keys, or t he t ouchscreen.
5. Press the Set Location submenu key.
The exter na l USB dr ive is n ow th e defau lt locat ion for saving files.
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
23/90
Site Master User Guide PN: 10580-00252 Rev. B 2-1
Chapter 2 Instrument Overview
2-1 Introduction
This chapt er pr ovides a br ief overview of th e Anr itsu S ite Mast er. The int ent of this chapt er is
to acquaint the u ser with th e instrum ent. For detailed measur ement inform ation, refer to a
specific mea sur ement guide listed in Appendix A, Measurement Guides.
2-2 Chapter Overview
Turn ing On the Site Master on page 2-1
Front P anel Overview on page 2-2
Display Overview on page 2-6
Test Pa nel Conn ector Overview on page 2-9
Symbols and Indicat ors on page 2-11
Data E ntr y on page 2-13
Mode Selector Menu on page 2-14
2-3 Turning On the Site Master
The Anritsu Site Mast er S331E/S361E m odels are capable of approximat ely four hours a nd
th e S332E/S362E models a re capa ble of approximat ely thr ee hours of cont inuous operat ion
from a fully cha rged, field-replaceable bat ter y (see Section 1-14 Batt ery Replacementon p age 1-5).
The Sit e Mast er can also be opera ted from a 12 Vdc sour ce (which will also simulta neously
cha rge th e batt ery). This can be achieved with either th e Anr itsu AC-DC Adapt er (Anr itsu
par t n um ber 40-168-R) or 12 Vdc Aut omotive Cigaret te Lighter Adapt er (Anr itsu par t
nu mber 806-141-R). Both items a re included with t he Site Mas ter (Table 1-3).
Caution
When using the Automotive Cigarette Lighter 12 VDC Adapter, Anritsu
Part Number 806-141-R, always verify that the supply is rated for a minimum of60 Watts @ 12 VDC, and that the socket is clear of any dirt or debris. If the
adapter plug becomes hot to the touch during operation, discontinue useimmediately.
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
24/90
2-4 Front Panel Overview Instrument Overview
2-2 PN: 10580-00252 Rev. B Site Master User Guide
To tur n on the Site Master , press the green On/Off button on the front panel (Figure 2-1)
The Site Ma ster ta kes appr oxima tely sixty seconds t o complete power wa rm -up a nd t o load
th e applicat ion softwa re. At th e completion of this pr ocess, the inst ru men t is r eady for u se.
2-4 Front Panel Overview
The Site Mast er men u-driven int erface is easy to use an d requires little tr aining. The
Site Master u ses a touch screen and keypad for da ta input. The five bott om m enu keys and
eight subm enu keys on th e right side are touch screen keys. The men u an d submen u keys will
vary depending upon the selected mode of operation, see Mode Selector Menu on page 2-14.
Nu mer ic keys 1 thr ough 9 ar e dual pur pose, depending upon th e cur rent mode of opera tion.
The dua l-pur pose keys ar e labeled with a nu mber on the key itself an d th e altern at e fun ction
is print ed in blue above each of the keys. Use th e blue Shift key to access th e functions pr inted
on t he pan el. The Escape key, used for a bort ing dat a en tr y, is t he oval but ton locat ed above
nu mer ic key 9. The rota ry kn ob, the four ar row keys, an d th e keypad can be used to cha ngeth e value of an a ctive param eter.
The Menu key provides gra phical icons of all th e insta lled mea sur ement modes and u ser
defined short-cuts (see Menu Key on page 2-3). The locat ions of th e keys ar e shown in
Figure 2-1.
Figure 2-1. Site Master Overview
NoteKeep the fan inlet and exhaust ports clear of obstructions at all times for proper
ventilation and cooling of the instrument.
Touch Screen
Main Menu KeysOn/Off Button
Shift Key
Numeric Keypad/
Shift Menu Keys (1 to 9)
Printed in Blue
Touch Screen
Submenu Keys
Fan Inlet Port
Fan Inlet PortFan Exhaust
Port
Calibration
Status and Type
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
25/90
Instrument Overview 2-4 Front Panel Overview
Site Master User Guide PN: 10580-00252 Rev. B 2-3
Front Panel Keys
Menu Key
Pr ess th is key to display a grid of short cut icons for inst alled measu rem ent m odes an d user
selected m enu s an d setu p files.
Figure 2-2 shows th e Menu key screen with sh ort cut icons for t he inst alled measur ementmodes. Touch one of the icons in th e top two rows to cha nge m odes. These icons a re
preinst alled and can not be moved or deleted.
Figure 2-2. Menu Key Screen, Icons for Installed Measurements
NoteThe display of the Menu screen will vary depending on Site Master model and
installed options.
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
26/90
2-4 Front Panel Overview Instrument Overview
2-4 PN: 10580-00252 Rev. B Site Master User Guide
Figure 2-3 shows th e Menu key screen with sh ort cut icons for t he inst alled measur ement
modes and four rows of user-defined sh ort cut s to men us a nd set up files.
Pr ess an d hold down a ny key for a few seconds t o add a sh ort cut to th is screen.To add
shortcut setu p files (.stp), open t he r ecall menu an d hold down on t he file na me for severa l
seconds. Th en select th e locat ion for th e short cut .
User-defined short cut s will sta y in memory un til deleted. To delete or m ove a short cut but ton,
press the Menu key then press a nd h old the sh ort cut for a pproxima tely 3 seconds. The
Customize Button dialog box will open t o allow a bu tt on to be deleted or m oved. Pr ess Esc to
exit th e Menu sh ort cut display.
Help for t he Menu shortcut screen is available by pressing the icon in th e lower-right corn erof th e display.
Esc Key
Pr ess this k ey to cancel any setting th at is curr ently being made.
Enter Key
Pr ess th is key to finalize data in put or select a h ighlighted item from a list.
Arrow Keys
The four ar row keys (ar oun d th e Enter key) are u sed t o scroll up , down, left, or r ight. Th e
ar row keys can often be used t o cha nge a valu e or t o cha nge a selection from a list. Thisfun ction is similar t o the function of th e rotar y knob. The a rr ow keys ar e also used to move
markers.
Figure 2-3. Menu Key Screen
NoteThe Factory Default reset will delete all user created shortcut icons from the Menu
screen. Refer to the Reset Menu on page 5-6 for additional information.
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
27/90
Instrument Overview 2-4 Front Panel Overview
Site Master User Guide PN: 10580-00252 Rev. B 2-5
Shift Key
Pr essing the Shift key and th en a num ber key execut es the function t hat is indicated in blue
text above the n umber k ey. When t he Shift key is active, its icon is displa yed at t he t op-right
of the measu rement display area by the bat tery charge indicat or.
Number Keypad
The Nu mber k eypad has t wo fun ctions: The prim ar y fun ction is number ent ry. The seconda ry
fun ction of th e num ber keypad is to list var ious m enu s. See Keypad Menu Keys (1 to 9)
on p age 2-5.
Rotary Knob
Tur ning t he rota ry kn ob cha nges nu mer ical values, scrolls th rough selecta ble item s from alist, and m oves mar kers. Values or items ma y be with in a dialog box or a n edit window.
Touch Screen Keys
Main Menu Touch Screen Keys
These five main men u k eys are h orizont ally arr an ged along t he lower edge of the t ouch
screen. The main m enu k ey fun ctions cha nge to ma tch specific instr um ent Mode sett ings. The
ma in menu keys genera te function-specific subm enus . The various measu rem ent m odes are
selected by pr essing th e Shift key and th en the Mode (9) key. Descriptions of the var iousmea sur ement modes can be foun d in the applicable Measur ement Guides listed in
Appendix A, Measurement Guides.
Submenu Touch Screen Keys
These submen u keys are ar ra nged along the r ight -ha nd edge of th e touch screen. The
submenu labels chan ge as instr ument measur ement sett ings chan ge. The curr ent submenu
title is shown at th e top of the subm enu key block.
Keypad Menu Keys (1 to 9)
Pr essing the Shift key and th en a n umber key selects t he menu function t hat is printed in
blue cha ra cter s above the n um ber key. See Figure 2-1 on page 2-2.
Not all Seconda ry Fu nction Menu s ar e active in various mea sur ement modes. If any one of
th ese menu s is available in a specific inst ru men t m ode of opera tion, then it can be called from
the n umber keypad. It m ay also be available from a m ain m enu key or a submenu key.
The Pr eset Menu (1) and System Menu (8) are described in Chapter 5, System Operations .
The Sweep Menu (3), Measure Menu (4), Tra ce Menu (5), and Limit Menu (6) vary depending
on m easurement mode, see th e Measurement Guides listed in Appendix A for information.The F ile Menu (7) is described in Chapter 4, File Management. The Mode Menu (9) isdescribed in Mode Selector Menu on page 2-14.
Figure 2-4. Shift Key Icon
NoteAvailable measurement modes are based on model and options purchased. Refer
to Table 1-1 and Table 1-2 for additional information.
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
28/90
2-5 Display Overview Instrument Overview
2-6 PN: 10580-00252 Rev. B Site Master User Guide
LED Indicators
Power LED
The P ower LE D is locat ed to th e left of the On/Off key. The LED is solid green when t he u nit
is on a nd slowly blinks when th e unit is off but ha s extern al power.
Charge LEDThe Cha rge LED is locat ed to the r ight of th e On/Off key. The LE D slowly blinks when th e
bat ter y is cha rging and is solid green wh en th e batt ery is fully cha rged.
2-5 Display Overview
Figure 2-5 and Figure 2-6 illust ra te some of th e key inform at ion a rea s of th e Site Mast er in
Cable and Ant enna mode and Spectrum Analyzer mode. For deta iled informa tion on either
mode, refer to the Measu remen t Gu ides listed in Appendix A, Measurement Guides.
Figure 2-5. Cable and Antenna Analyzer Return Loss Measurement Display
Date/Time
GPS Icon
GPS Location
SaveBattery
ChargeSubmenu
Touch Screen
Keys
Measurement
Settings
Summary
Calibration
Status, Type
Main Menu
Touch Screen Keys
Marker Table
Frequency
Standard
Trace
Measurement
Title
http://-/?-http://-/?-7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
29/90
Instrument Overview 2-5 Display Overview
Site Master User Guide PN: 10580-00252 Rev. B 2-7
Figure 2-6. Spectrum Analyzer Display (S332E and S362E only)
Date/Time
GPS Icon
GPS Location
Save
Battery
ChargeSubmenu
Touch Screen
Keys
Measurement
Settings
Summary
Main Menu
Touch Screen Keys
Marker Table
Frequency
Standard
Trace
Measurement
Title
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
30/90
2-5 Display Overview Instrument Overview
2-8 PN: 10580-00252 Rev. B Site Master User Guide
In addition t o the defau lt color display, Site Mast er offers th e following display sett ings:
Black & White for print ing an d viewing in broad da ylight conditions
Night Vision optimized for night-time viewing
High Contrast for oth er cha llenging viewing cond itions
Figure 2-7. Site Master Display Settings
Default Colors Black & White
High ContrastNight Vision
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
31/90
Instrument Overview 2-6 Test Panel Connector Overview
Site Master User Guide PN: 10580-00252 Rev. B 2-9
2-6 Test Panel Connector Overview
Test pan el conn ectors for t he Site Mast er S332E ar e shown in Figure 2-8.
External Power
The extern al power conn ector is used to power th e unit an d for bat ter y cha rging. Inpu t is
12 VDC to 15 VDC at up t o 5.0 A. The green flashing P ower LED n ear th e power switch
indicat es that th e instr ument has externa l power.
USB Interface Type A
The Site Ma ster ha s t wo Type A USB conn ectors t ha t a ccept U SB F lash Memory devices for
storing measur ements, setups dat a, and screen images.
USB Interface Mini-B
The US B 2.0 Mini-B conn ector can be u sed to conn ect th e Site Ma ster directly to a P C. The
first tim e th e Site Mast er is conn ected t o a PC, th e norma l USB device detection by th e
compu ter opera ting system will ta ke place. The CD-ROM tha t is shipped with th e instr um ent
cont ains a driver for Windows XP t ha t is inst alled when Mast er Softwa re Tools is inst alled.
Drivers ar e not a vailable for ea rlier versions of th e Windows opera ting system . During t he
driver inst allat ion pr ocess, place the CD-ROM in th e compu ter drive and sp ecify tha t t he
insta llation wizard s hould search th e CD-ROM for t he dr iver.
Figure 2-8. S332E Test Panel Connector
Warning
When using the AC-DC Adapter, always use a three-wire power cable that is
connected to a three-wire power line outlet. If power is supplied without grounding
the equipment in this manner, then the user is at risk of receiving a severe or fatal
electric shock.
NoteFor proper detection, Master Software Tools should be installed on the PC prior toconnecting the Site Master to the USB port.
RF Out USB Type A
External
Power
GPS
USB Mini-B
RF InExternal
ReferenceExternal
Trigger
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
32/90
2-6 Test Panel Connector Overview Instrument Overview
2-10 PN: 10580-00252 Rev. B Site Master User Guide
Headset Jack
The h eadset jack pr ovides au dio out put from t he bu ilt-in AM/FM/SSB demodulator for t esting
an d t roubleshooting wireless comm un icat ion s ystems. The jack a ccepts a 2.5 mm 3-wire
miniat ur e phone plug such as t hose comm only used with cellular telephones.
Ext Trigger In (S332E, S362E Models)
A TTL signa l th at is applied to th e Exter na l Trigger female BNC input conn ector causes a
single sweep to occur . In t he Spectru m Ana lyzer mode, it is us ed in zero span, a nd t riggering
occur s on th e rising edge of the signal. After t he sweep is complete, th e resu ltan t tr ace is
displayed un til the next t rigger signal ar rives.
RF In
50 Type-N female conn ector. Maximum input is +26 dBm a t 50 VDC.
RF Out/Reflection In
RF outpu t, 50 impeda nce, for r eflection measu rem ent s. Maximu m inpu t is +23 dBm a t
50 VDC.
GPS Antenna Connector
The GPS an ten na conn ection on t he Site Ma ster is type SMA-fema le. GPS function is
described in Chapter 6, GPS (Option 31) .
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
33/90
Instrument Overview 2-7 Symbols and Indicators
Site Master User Guide PN: 10580-00252 Rev. B 2-11
2-7 Symbols and Indicators
The following symbols an d indicat ors ind icat e th e instr um ent sta tu s or condit ion on th e
display.
Calibration Symbols
The cur ren t calibra tion sta tu s an d type is displayed in the u pper-left of th e screen wh en in
Cable & Ant enna Ana lyzer m ode. See Figure 2-5 on page 2-6. The five sta tus messages are
described n ext.
Cal Status: ON, Flex
The Site Mast er ha s been calibrat ed with discret e Open, Short , an d Load component s. This is
a F lexCal calibra tion indicat ing it is possible to cha nge th e frequen cy ra nge after calibrat ion.
Cal Status: ON, Standard
The Site Mast er ha s been calibrat ed with discret e Open, Short , an d Load component s. This is
a St an dar d calibrat ion indicating it is not possible to cha nge th e frequen cy ran ge aftercalibrat ion with out perform ing another calibrat ion.
Cal Status: ON, Flex, Insta
The Site Mast er ha s been calibra ted with t he Inst aCa l module. This is a FlexCal calibrat ion
indicat ing it is possible to change t he frequency ran ge after calibrat ion.
Cal Status: ON, Standard, Insta
The Site Master has been calibrat ed with the In staCal m odule. The Site Master has been
calibrat ed with discret e Open, Short , and Load component s. This is a St an dar d calibrat ion
indicat ing it is not possible to cha nge th e frequen cy ra nge after calibrat ion with out
perform ing another calibrat ion.
Cal Status Off:
The Site Master has not been calibrated.
For calibrat ion pr ocedur es refer to the Cable & Ant enna Measur ement Guide
(PN: 10580-00241) listed in Appendix A.
Battery Symbols
The bat ter y symbol above the display indicates th e cha rge rema ining in t he bat ter y. The
colored s ection in side th e symbol cha nges size an d color wit h t he cha rge level.
G r e e n : Bat ter y is 30% to 100% cha rged
Yellow: Bat ter y is 10% to 30% cha rged
R e d : Bat ter y 0% to 10% cha rged
Lig h tn in g B o lt : Bat ter y is being cha rged (any color sym bol)
Detailed batt ery inform at ion is also available in th e Sta tu s dialog box (System> Status).
Figure 2-9. Battery Status
http://-/?-http://-/?-7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
34/90
2-7 Symbols and Indicators Instrument Overview
2-12 PN: 10580-00252 Rev. B Site Master User Guide
When eith er t he AC-DC Adapt er (40-168-R) or th e 12 Volt DC ada pter (806-141-R) is
conn ected, the ba tt ery au tomat ically receives a char ge, an d th e batt ery symbol with th e
light ning bolt is displayed (Figure 2-10).
The green Charge LED flashes when t he batt ery is char ging, and rema ins on st eady when the
bat ter y is fully cha rged.
When opera ting from extern al power without a ba tt ery insta lled, the bat ter y symbol is
rep laced by a red plu g body (Figure 2-11).
Additional Symbols
Single Sweep
Single Sweep is selected. Pres s Continuous in the Sweep men u t o resu me cont inuous sweeping.
Floppy Icon
Shortcut t o the Save subm enu. Touch th e icon to open t he t ouch screen k eyboar d for saving
measurements, setups, or screen displays.
Figure 2-10. Battery Charging Icon
Caution Use only Anritsu-approved batteries, adapters, and chargers with this instrument.
Figure 2-11. Battery Not Installed
http://-/?-http://-/?-7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
35/90
Instrument Overview 2-8 Data Entry
Site Master User Guide PN: 10580-00252 Rev. B 2-13
2-8 Data Entry
Numeric Values
Nu mer ic values ar e cha nged using th e rotar y knob, ar row keys, or t he keypad. Pr essing one
of the ma in men u keys will display a list of submenu s on the r ight side of the touch screen.
When t he value on a su bmenu key is displayed in red, it is ready for chan ging. When u singthe r ota ry knob or a rrow keys th e chan ging value is shown on the submenu and in red on th e
grat icule. When usin g the keypad, the new value is shown in red on t he grat icule an d the
subm enu chan ges to Un its. Selecting a u nit for t he new value completes t he ent ry.
Parameter Setting
Pop-up list boxes or edit boxes ar e used t o pr ovide selection list s an d selection edit ors. Scroll
th rough a list of items or para met ers with t he ar row keys, the rotar y knob, or t he touch
screen. These list boxes an d edit boxes frequen tly display a r an ge of possible valu es or limit s
for possible values.
Finalize the input by pressing th e Enter key. At a ny tim e before finalizing th e input , press t he
escape (Esc) key to abort t he change an d ret ain t he previously existing sett ing.
Some para met ers (such as for a nt enna s or couplers) can be added to list boxes by crea ting
them and import ing them using Master Softwar e Tools.
Text Entry
When ent ering text, as when sa ving a mea sur ement , the touch screen keyboar d is displayed
(Figure 2-12). Cha ra cter s ar e enter ed directly with th e touch screen keyboar d. The keypad
can be used for n um eric ent ry. The left a nd r ight ar row keys will scroll th e cur sor t hr ough t he
filename. See Save Menu on page 4-8 for additional inform at ion.
Figure 2-12. Touch Screen Keyboard
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
36/90
2-14 PN: 10580-00252 Rev. B Site Master User Guide
2-9 Mode Selector Menu
To access th e fun ctions un der t he Mode men u, select t he Shift key, then th e Mode (9) key. Us e
th e directional ar row keys, the r ota ry kn ob, or t he t ouch screen to highlight th e selection, and
press the Enter key to select. The list of modes tha t a ppear in th is menu will vary depending
upon t he options th at are insta lled and activat ed in th e instr ument . Figure 2-13 is an
example of th e Mode menu . Your instr um ent m ay not show th e same list. The cur ren t m ode isdisplayed below t he bat ter y symbol.
Th e Menu key is another option t o quickly cha nge measu rem ent m odes. Press th e Menu key
th en select one of th e Measur ement icons in t he t op two rows (Figure 2-2 on page 2-3).
Figure 2-13. Mode Selector Menu
http://-/?-http://-/?-7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
37/90
Site Master User Guide PN: 10580-00252 Rev. B 3-1
Chapter 3 Quick Start Guide
3-1 Introduction
This cha pter provides a brief overview of basic mea sur ement setu ps. For deta iled
mea sur ement inform at ion, refer to a specific mea sur ement guide listed in
Appendix A, Measurement Guides. This chapt er provides quick st art measur ement
inform at ion for th e following mea sur ement modes:
Section 3-3 Cable & Antenn a Analyzer on page 3-2
Section 3-4 Spectr um Analyzer on page 3-7
3-2 Measurement Mode Selection
Press the Menu key and use t he touch screen t o select t he appr opriat e measu rem ent icon.
Figure 3-1. Menu Screen with Icons for Installed Measurement Modes
Note The display of the Menu screen will vary depending on installed options.
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
38/90
3-3 Cable & Antenna Analyzer Quick Start Guide
3-2 PN: 10580-00252 Rev. B Site Master User Guide
3-3 Cable & Antenna Analyzer
Set t he inst ru ment to Cable & Ant enn a Analyzer m ode as described in th e previous section.
Select the Measurement Type
Press the Measurement main m enu key and select th e appropriate measu rement.
Set the Frequency
1. Press the Freq/Dist main menu key.
2. Press the Start Freq submenu key and use t he keypad, rotary knob, or t he ar row keys to
enter t he star t frequency.
3. Press the Stop Freq submenu key and use t he keypad, rota ry knob, or t he ar row keys to
ent er t he stop frequen cy.
Set the Amplitude
1. Press the Amplitude main menu key.
2. Press the Top subm enu k ey an d use th e keypad, rota ry knob, or t he ar row keys to edit
th e top scale value. Press Enter to set.
3. Press the Bottom submenu key and use t he keypad, rotar y knob, or t he ar row keys to
edit th e bott om scale value. Press Enter to set.
Figure 3-2. Menu Screen with Icons for Installed Measurement Modes
Note For Amplitude in Smith Chart measurements, refer to Smith Chart on page 2-23of the Cable & Antenna Measurement Guide listed in Appendix A.
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
39/90
Quick Start Guide 3-3 Cable & Antenna Analyzer
Site Master User Guide PN: 10580-00252 Rev. B 3-3
Turn on Markers
1. Press the Marker ma in menu key.
2. Press the Marker 1 2 3 4 5 6 submenu key and select th e mark er num ber 1 butt on using
the t ouch screen. The underlined number on t he Marker submenu key indicat es the
active mar ker.
3. Use th e arr ow keys, th e keypad, or t he rotar y knob to move the ma rk er. The cur ren t
value for t he selected m ar ker is sh own a bove the u pper-left corn er of th e graph . It is
also possible to drag th e mar ker u sing the t ouch screen.
4. Delta Ma rk ers a re a vailable for ea ch of th e six reference mar kers. F or t he selected
ma rker , Toggle the Delta On/Off submenu key to tur n on the Delta m ar ker.
Peak/Valley Auto Markers
When ma king Retur n Loss an d VSWR measur ement s, the Pea k/Valley Aut o feat ur e can be
used t o au tomat ically tur n on Mar ker 1 t o peak, Mar ker 2 t o valley, an d display M1 and M2in th e Mark er Table. This featu re is n ot available for DTF measu rements.
1. Pr ess the Marker main menu key.
2. Pr ess the Peak/Valley Auto key.
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
40/90
3-3 Cable & Antenna Analyzer Quick Start Guide
3-4 PN: 10580-00252 Rev. B Site Master User Guide
Single Limit Line
1. Press Shift and then Limit (6) to ent er th e Limit m enu.
2. Press the Limit On/Off key to tur n on the Limit.
3. Press Single Limit and then use th e num eric keypad, the a rr ow keys, or t he rotary kn ob
to chan ge the limit value and t hen press Enter.
4. Press the Limit Alarm key to tur n on or off the Limit Alarm .
NoteRefer to the Cable & Antenna Measurement Guide listed in Appendix A for
creating multi-segment limit lines.
Figure 3-3. Single Limit Lines
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
41/90
Quick Start Guide 3-3 Cable & Antenna Analyzer
Site Master User Guide PN: 10580-00252 Rev. B 3-5
DTF Setup
1. Press the Measurements main menu key and select DTF Return Loss or DTF VSWR.
2. Press the Freq/Dist main menu key.
3. Press the Units subm enu k ey an d select m to display dista nce in m eter s or ft to display
dista nce in feet.4. Press DTF Aid an d use th e touch screen, or a rr ow keys to na vigate t hr ough a ll th e DTF
parameters.
a . Set Start Distance and Stop Distance. Stop Distan ce needs to be sma ller t ha n
Dmax.
b . Enter the Start and Stop frequencies.
c . Press Cable, select t he a ppropriat e cable from th e cable list a nd pr ess Enter.
d . Press Continue.
5. Press Shift then Calibrate (2) to calibrat e the inst ru men t. Refer to Calibrat e with OSL
Calibrat ion on page 3-6 for additional in form at ion.
6. Press the Marker ma in menu key and set th e appropriate mar kers.
7. Press Shift and Limit (6) to enter a nd set t he appr opriat e limit lines.
8. Press Shift and File (7) to save the mea sur ement . See th e User Guide for det ails.
Figure 3-4. DTF Aid
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
42/90
3-3 Cable & Antenna Analyzer Quick Start Guide
3-6 PN: 10580-00252 Rev. B Site Master User Guide
Calibrate with OSL Calibration
1. Press the Freq/Dist main m enu key and enter the appr opriate frequency ra nge
2. Press Shift then Calibrate (2) key.
3. Select Standard or FlexCal.
4. Press Start Cal and follow instructions on screen.
5. Conn ect Open t o RF Out and press the Enterkey.
6. Conn ect Short t o RF Out and press the Enterkey.
7. Conn ect Load to RF Out and press the Enterkey.
8. Verify th at th e calibrat ion h as been properly perform ed by checking that th e Cal Stat us
messa ge is now displaying ON, Standard or ON, FlexCal.
NoteRefer to the Cable & Antenna Measurement Guide listed in Appendix A for
calibration details.
Figure 3-5. Calibration Setup with OSL Cal
Power Charge
+/-.0
3Sweep
2
Calibrate
1
Preset
6
Limit
5
Trace
4
Measure
9
Mode
8
System
7
File
ShiftEsc
Enter
Menu
SiteMasterS331E
OPEN
LOAD
SHORT CALIBRATION
RFOUT/REFLECTIONTEST PORT
TEST PORT CABLE (OPTIONAL)
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
43/90
Quick Start Guide 3-4 Spectrum Analyzer
Site Master User Guide PN: 10580-00252 Rev. B 3-7
3-4 Spectrum Analyzer
Set t he inst ru men t t o Spectr um Ana lyzer mode as described in Section 3-2 Measurement
Mode Selection on page 3-1.
Set Start and Stop Frequencies
1. Press the Freq main menu key.
2. Press the Start Freq submenu key.
3. Ent er t he desired start frequency using the keypad, the ar row keys, or t he r ota ry knob.
When ent ering a frequen cy using the keypad, the submen u key labels cha nge to GHz,
MHz, kHz, and Hz. Press the a ppropriate un it key. Pr essing the Enter key has th e same
affect a s pressing t he MHz submenu key.
4. Press the Stop Freq submenu k ey.
5. En ter t he desired stop frequency.
Enter the Center Frequency
1. Press the Freq main menu key.
2. Press the Center Freq submenu key.
3. En ter t he desired cent er frequency using th e keypad, th e arr ow keys, or t he rotar y knob.
When ent ering a frequen cy using the keypad, the submen u key labels cha nge to GHz,
MHz, kHz, and Hz. Press the a ppropriate un it key. Pr essing the Enter key has th e same
affect a s pressing t he MHz submenu key.
The cent er frequency an d span is sh own at th e bott om of th e screen.
Select a Signal Standard
1. Press the Freq main menu key.
2. Press the Signal Standard subm enu key. The Signal St an dar ds dialog box opens.
3. Highlight a signal sta ndar d and pr ess Enter to select.
4. Press the Channel submenu key to chan ge the channel value in t he Cha nnel Editor.
The signal sta nda rd is sh own in yellow at th e top of the screen.
Set the Measurement Frequency Bandwidth1. Press the BW main m enu key to display the BW menu.
Press the RBW and/or t he VBW submenu key to manua lly chan ge the values.
Set RBW and VBW aut oma tically by pressing th e Auto RBW submenu key or t heAuto VBW submenu key.
2. Press the RBW/VBW subm enu k ey to cha nge the r esolution bandwidth an d video
bandwidth ra tio.
3. Press the Span/RBW submen u key to cha nge the span width t o resolut ion ban dwidth
ratio.
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
44/90
3-4 Spectrum Analyzer Quick Start Guide
3-8 PN: 10580-00252 Rev. B Site Master User Guide
Set the Amplitude
Press the Amplitude main menu key to display th e Amplitude menu.
Set Amplitude Reference Level and Scale
1. Press the Reference Level submenu key and use t he ar row keys, rotary knob, or t hekeypad t o cha nge th e reference level. Press Enter to set t he r eference level value.
2. Press the Scale submenu key and use t he ar row keys, rotary knob, or t he keypad toenter the desired scale. Press Enter to set th e scale value.
Set Amplitude Range and Scale
1. Press the Auto Atten subm enu k ey to set a n optima l reference level based on th e
measur ed signal.
2. Press the Scale submenu key.
3. En ter t he desired scale unit s by using th e keypad, th e arr ow keys, or t he rotar y knob.Press Enter to set. The y-axis scale is au tomat ically renu mbered.
Power Offset Set Up for Compensating External Loss
To obta in accur at e result s, compensat e for a ny extern al at ten ua tion by using power offset.
In power offset m ode, th e compen sat ion factor is in dB. (Exter na l att enua tion can be creat ed
by using an externa l cable or an extern al high power a tt enuat or.)
Press the RL Offset submenu key and use t he keypad, the ar row keys, or t he rotar y knob to
ent er t he desired offset value. When using th e rotar y knob, the value cha nges in increm ent s
of 0.1 dB. Using t he Left/Right ar row keys changes t he value in 10% increment s of th e valueshown on th e Scale submenu key. When u sing the Up/Down ar row keys, the value chan ges in
the increment sh own on the Scale submenu key. When u sing the keypad, enter t he new value
then press Enter or th e dB subm enu key to set th e value. The power offset is displayed in th e
instru ment settings summ ary column on t he left side of the measur ement display.
Set the Span
1. Press the Span main m enu key or t he Freq ma in m enu key followed by the
Span submenu k ey.
2. To select full span , press t he Full Span submen u k ey. Selecting full span overr ides any
previously set Sta rt an d Stop frequen cies.
3. For a single frequency measur ement, press t he Zero Span submenu key.
NoteTo quickly move the span value up or down, press the Span Up 1-2-5 or SpanDown 1-2-5 submenu keys These keys facilitate a zoom-in, zoom-out feature in a
1-2-5 sequence.
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
45/90
Quick Start Guide 3-4 Spectrum Analyzer
Site Master User Guide PN: 10580-00252 Rev. B 3-9
Single Limit Line
Press the Limit menu key to display the Limit menu .
1. Press the Limit (Upper / Lower) subm enu k ey to select t he desired limit line,
Upper or Lower.
2. Activate th e selected limit line by pressing th e On Off submenu k ey so tha t On isunderlined.
3. Press the Limit Move subm enu k ey to display th e Limit Move menu . Press th e first
Move Limit subm enu key an d use th e arr ows keys, rotar y knob, or keypad to cha nge the
dBm level of the limit line.
4. Press the Back submenu key to retu rn t o th e Limit menu .
5. If necessary, press th e Set Default Limit subm enu k ey to redr aw t he limit line in view.
Segmented Limit Lines
The following procedur e crea tes limit lines for a Retur n Loss Measur ement . Limits ar e set t o12 dB between 1000 MHz an d 1400 MHz and to 15 dB between 1600 MHz a nd 2000 MHz.
1. Press the Limit (Upper / Lower) subm enu k ey to select t he desired limit line, Upper or
Lower.
2. Activate th e selected limit line by pressing th e On Off submenu k ey so tha t On is
underlined.
3. Press the Limit Move submen u key to display th e Limit Move men u.
Create a Sample Segmented Limit Line
4. Press the Move Limit submenu k ey and use th e num ber keys to enter a limit value of
12 dB. The limit value can also be moved by using th e arr ow keys or t he rota ry knob.Press the Back submenu key to retur n to the Limit menu .
5. Press the Limit Edit submenu key to display the E dit menu.
6. Press the Add Point submenu key.
7. Press the Frequency submenu key and enter 1400 MHz.
8. Press the Add Point submenu key
9. Press the Frequency submenu key and enter 1600 MHz.
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
46/90
3-4 Spectrum Analyzer Quick Start Guide
3-10 PN: 10580-00252 Rev. B Site Master User Guide
Create a Limit Envelope
1. Press Shift then Limit (6) to open t he Limit men u.
2. Select Limit Envelope.
3. Press the Create Envelope key.
Figure 3-6. Limit Envelope
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
47/90
Quick Start Guide 3-4 Spectrum Analyzer
Site Master User Guide PN: 10580-00252 Rev. B 3-11
Setting Up Markers
Press the Marker main m enu key to display the Mar ker menu .
Selecting, Activating, and Placing a Marker
1. Press the Marker 1 2 3 4 5 6 submenu key and t hen select th e desired marker u sing the
touch screen ma rker butt ons. The selected m ark er is un derlined on th e Markersubmenu key.
2. Press the On Off submenu k ey so th at On is under lined. The selected ma rker isdisplayed in r ed an d rea dy to be moved.
3. Use th e rotar y knob to place the m ar ker on th e desired frequen cy.
4. Repeat steps 1 t hrough 3 to activat e an d move additional m ar kers.
Selecting, Activating, and Placing a Delta Marker:
1. Press the Marker 1 2 3 4 5 6 submenu key and select t he desired delta m ark er. The
selected marker is underlined.
2. Press the Delta On Off submenu key so th at On is under lined. The selected ma rk er is
displayed in r ed an d rea dy to be moved.
3. Use th e rotar y knob to place the delta ma rker on t he desired frequen cy.
4. Repeat steps 1 t hrough 3 to activat e an d move additional m ar kers.
Viewing Marker Data in a Table Format
1. Press the More submenu key.
2. Press the Marker Table On Off submenu k ey so tha t On is underlined. All marker and
delta m ar ker data a re displayed in a ta ble under the measur ement graph.
Figure 3-7. Marker Table
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
48/90
3-4 Spectrum Analyzer Quick Start Guide
3-12 PN: 10580-00252 Rev. B Site Master User Guide
Select a Smart Measurement Type
In Spectru m Analyzer mode, press Shift then Measure (4) and select a smar t m easurement
using the submenu k eys.
Figure 3-8. Spectrum Analyzer Measure Menu
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
49/90
Quick Start Guide 3-5 Saving Measurements
Site Master User Guide PN: 10580-00252 Rev. B 3-13
3-5 Saving Measurements
Measu rem ent files can be st ored in th e following form at s:
.VNA or .DAT for Ca ble and Antenn a a na lyzer mea sur ement s
.SPA for Spectru m Analyzer meas ur ement s
Sa ving files in .VNA or .SP A is recomm ended a s it en ables u ser s to edit, view, and a na lyzetr aces with Mast er Softwa re Tools (MST).
Procedure for saving files:
1. Press Shift then File (#7).
2. Press Save Measurement.
3. Press Change Save Location an d set t he curr ent locat ion t o be the US B flash dr ive or
interna l memory, and then pr ess Set Location.
4. Press Change Type (Setup/JPG/) an d select Measurement .VNA or Measurement .DAT
or Measurement (when in Spectru m Ana lyzer mode).
5. Ent er th e file nam e using the keyboard and pr ess Enter.
Refer t o Chapter 4, File Management for more deta ils about work ing with files.
Note
Anritsu recommends saving files to the internal memory and then transferring the
files to an external USB memory device if needed. Refer to Chapter 4, Copying
Files for more details.
The .DAT file format is only recommended for users who need to work with thisformat or prefer using Handheld Software Tools (HHST). Traces saved as .DAT
can be viewed, edited, and analyzed with Handheld Software Tools. If the DUAL
measurement display is turned on, files will be saved as Filename_1 and
Filename_2.
.DAT is only supported for Return Loss, VSWR, Cable Loss, DTF RL, DTF VSWR
and only supports 137, 275, 551 data points. 1102 and 2204 data points are not
supported in the .DAT file format. Use the .VNA file format if these resolutions are
required.
.DAT files cannot be recalled to the instrument for viewing. If this is required, use
the .VNA file format.
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
50/90
3-6 Useful MST Utilities Quick Start Guide
3-14 PN: 10580-00252 Rev. B Site Master User Guide
3-6 Useful MST Utilities
Converting Files to .DAT File Format
1. Est ablish a conn ection with MST.
2. Download measur ements:a . Go to Sync | Download all measurements.
b . Select th e folder on t he compu ter or select Local, an d t hen set t he locat ion.
c . ClickDevice an d dra g all of th e tr aces int o the mea sur ement window. All of th e
tr aces will au toma tically be saved in th e selected Local locat ion.
3. To convert all of the files in a folder, select File | Save Folder as .DAT, and th en selectthe folder used above. To convert files one-by-one, select File | Save As, and then
cha nge th e extension t o .DAT to convert th e tr ace.
4. Files conver ted t o .DAT file forma t can be opened in HH ST.
Figure 3-9. MST Dialog
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
51/90
Quick Start Guide 3-6 Useful MST Utilities
Site Master User Guide PN: 10580-00252 Rev. B 3-15
Group Edit
The Group E dit featu re a llows ma rker s an d limit lines to be copied from one tra ce to all of the
tr aces in a folder. In addition, the t itle and s ubt itle can be quickly rena med for a ll of th e
tr aces in a folder. For example, to add a cell site nu mber on th e title.
To cha nge th e tit le to be the cell site nu mber for a ll tra ces in a folder:
1. Select Tools | Group Edit.
2. Set th e applicat ion type to VNA.
3. Select th e locat ion of the folder.
4. Ent er t he cell site nu mber a nd checkPlot Title.
5. ClickApply to ren am e all of th e plot tit les in t he selected folder.
Figure 3-10. MST Measurement Group Edit Dialog
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
52/90
3-6 Useful MST Utilities Quick Start Guide
3-16 PN: 10580-00252 Rev. B Site Master User Guide
Print All to PDF
If Adobe Acrobat is in sta lled on th e compu ter with MST, tr aces can be convert ed to PDF using
Print All an d selecting Print to PDF. This crea tes a compa ct a nd port able PDF report of all of
th e tr aces in a folder with just one click.
Figure 3-11. MST Print All Measurements Dialog
7/30/2019 Anritsu S361E Site Master and S3xxE User Guide
53/90
Site Master User Guide PN: 10580-00252 Rev. B 4-1
Chapter 4 File Management
4-1 Introduction
This cha pter will review th e file ma na gement featu res of th e Site Master a nd deta il th e
File men u. The su bmenu s un der t his men u a llow th e user t o save, recall, copy, an d delete files
in inter na l memory or an exter na l USB flash dr ive.
4-2 Managing Files
Press the Shift key then th e File (7) key on t he n um eric keypad t o list t he File men u. Follow
th e additional st eps below.
Save Files
Set the Save Location
Press Save then the Change Save Location subm enu keys an d select th e locat ion t o save files.
You can save files to the int erna l memory or t o an extern al US B flash drive. You can also
creat e new folder s. If an exter na l USB flash dr ive is connected or disconn ected , press Refresh
Directories to updat e th e locat ion t ree. Press t he Set Location key to st ore t he s ave locat ion.
Save Measurement As
Th e Save Measurement As key is used t o quickly save mea sur ement s with a specific file
nam e. The Site Master saves the measur ement with th e latest file nam e that was used to
save a measurement and with a n umber th at is au tomatically increment ed and appended to
the end of the file nam e. For inst ance, if the last measur ement was saved with th e nam e
System Retur n Loss, Save Measurement As saves the next measurement as System Return
Loss_#1, System Retur n Loss_#2 etc. The file nam e used can be cha nged usin g th e Save
dia log box (Figure 4-1).
Save a Measurement
Press the Save Measurement key and enter the n ame for t he measu rement file. Themea su rem
Recommended