36m SANS BATAN (SMARTer) – INDONESIA
“Improvement The Neutron Intensity By Focusing Optics”
Edy Giri Rachman Putra*, Epung Saepul Bahrum, Bharoto, Eddy Santoso
Neutron Scattering Laboratory – BATAN, Kawasan Puspiptek Serpong, Tangerang 15314, Indonesia
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
Layout of the presentation
• Introduction• The instrument• The performance• The future programs
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
Neutron NewsVol. 17, issue 4October/November/December 2006Neutron Scattering Facilities in Serpong – Part 1
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
Neutron NewsVol. 18, issue 1January/February/March 2007Neutron Scattering Facilities in Serpong – Part 2
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
The Reactor – GA Siwabessy
Descriptive parameters of RSG – GAS
Power : 30MW (Thermal)Peak Thermal Neutron Flux : 2.5 x 1014 cm-2 s-1
CoreActive core volume (dm3) : 180Active core height (cm) : 60Loading (Kg 235U) : 8.675Number of fuel elements : 40Number of control elements : 8Fuel type : U3SiAl - MTR235U enriched (%) : 19.75235U density (g cm-3) : 2.96Moderator / coolant : H2O
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
Reactor Operation Schedule 2006
Reactor Operation Schedule 2007
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
~ 170 days operation/year
Neutron Scattering Laboratory - BATAN
Neutron beam instruments
(DN1-M) RSM : Diffractometer for Residual Stress Measurement(DN2) FCD/TD : Four Circle Diffractometer/Texture Diffractometer(SN1) TAS : Triple Axis Spectrometer(RN1) NRF : Neutron Radiography Facility(DN3) HRPD : High Resolution Powder Diffractometer(SN2) SMARTer : Small-Angle Neutron Scattering Spectrometer(SN3) HRSANS : High-Resolution Small-Angle Neutron Scattering Spectrometer
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
SMARTer
Low Angle Diffractometer LADI
Macromolecules Neutron DiffractometerMaNDi
Low Q Diffractometer LQD
AUstralia SANS AUSANS
A new 40m Australia SANS QUOKKA
MalaYsia SANS MYSANS
A Liquid And Disordered Materials Diffractometer ALADDIN
Small Angle Neutron Diffractometer SANDI
SANS-University of Tokyo SANS-U
SANS-JAEA SANS-J
A new 40m HANARO SANS ?36 meter Small-Angle Neutron Scattering Spectrometer - BATAN
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
1992 : Commissioned1992 – 1996 : Testing, setting up & calibration > wavelength, resolution*1996 – 1997 : Pre-amp & 2D-detector problems (Prof. Mortensen – RISØ)1997 : IAEA Expert Mission Workshop > Porasil sample
Dr. P.S. Goyal (BARC) Dr. A. Wiendenmann (HMI)
1998 – 2003 : Calibration > AgBE, Porasil, Polymers Setting up > pinhole, collimator, etc.
2003 – 2005 : Off > computer problem & TDC
Start from here by Edy Giri Rachman Putra
2005 – : Optimizing, calibration, alignment > AgBE, porasil, polymers, etc.
A new controlling & acquisition system A new instrument setting
Improving the performance
The time line of SMARTer
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
Sample environmentMagnetic permanentLow and high temperature furnaceStretching/drawing frame**Shear cell*Stop-flow cell*
Upgrading & improving the performance
Control systemWindows menu
Live on data collecting
• A new controlling & data collecting system• Pinhole & collimation alignment• A new sample changer system• A small furnace/heater for liquid sample• A new beam stop system• A new sample to detector distance• Wavelength calibration• Q-calibration• High temperature experiment < 100 C• Magnetic field experiment• Low & high temperature experiment*
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
Specifications
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
Instrumental parameters of SMARTer (36m SANS BATAN)
Beam port and neutron guide : S5 (radial), NG1Incident wavelength : 0.3 – 0.6 nm (VS = 3500 – 7000 rpm)Wavelength resolution : / ~ 10% - 20%Q-range : 0.02 < Q < 6 nm-1
Max. flux at sample position : 4 x 106 neutron/cm s(at = 0.32 and 22.5 MW)Beam size : min. 5mm in diameterSample to detector distance : 1.3m L 18mSample environment : open air atmosphere
small furnace/heater up to 150 °CData acquisition : OS LinuxControlling system : Windows 98
Speed (rpm)
Experimental Wavelength
Theoretical Wavelength
FWHM(Q)
Q/Q(%)
3500 5.66 Å 5.83 Å 0.114 10.8
4000 4.92 Å 5.10 Å 0.127 11.9
4500 4.35 Å 4.53 Å 0.133 12.4
5000 3.90 Å 4.08 Å 0.145 13.5
5500 3.52 Å 3.71 Å 0.155 14.3
6000 3.22 Å 3.40 Å 0.172 15.9
6500 3.02 Å 3.14 Å 0.186 17.5
7000 2.74 Å 2.92 Å 0.203 18.9
Performance
2D pattern from AgBE at = 3.22 Å
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
J. Appl. Cryst. 2007, 40, s447 - s452
Inter-laboratory comparison
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
E. Giri Rachman Putra, et al., J. Appl. Cryst. 2007, 40, s447 - s452
Inter-laboratory comparison
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
Sample-to-detector distance : 18mCollimation length : 18mWavelength (resolution) : 5.66 Å (11%)Pinhole settings : 666555Source aperture : 30 mm (dia)Sample aperture : 10 mm (dia)Beam stop : 50 mm (dia)Sample thickness : 1 mmSample transmission : 93%Collecting time : 2hrs
Inter-laboratory comparison
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
E. Giri Rachman Putra, et al., J. Appl. Cryst. 2007, 40, s447 - s452
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
Experiment on Solid State Polymer Sample
Block copolymer PS-PEP
E. Giri Rachman Putra, et al., J. Appl. Cryst. 2007, 40, s447 - s452
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
Experiment on Porous Materials
The best resolution
L1 = 18mL2 = 4m, 8m, 18mPinhole setting = 666555 (30, 30, 30, 20, 20, 20 mm in diameter)Collecting time = 1hr
Experiment on CuNiFe Magnetic Sample
0 kOe 420 kOe 840 kOe 10,000 kOe
H
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
Polymer solutions (F88 Pluronics)Tem
pera
ture
Temperature dependence
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
0.1
2
4
68
1
2
4
68
10
2
4
68
100
Inte
nsi
ty (
rela
tive
)
0.012 3 4 5 6 7 8 9
0.12 3
Q (1/Å)
KCl 1.5M KCl 1.2M KCl 1.0M KCl 0.8M KCl 0.6M KCl 0.4M KCl 0.2M KCl 0.1M KCl 0.05M theoretical
Salt
con
cen
trati
on
Polymer solutions (F88 Pluronics)
Salt concentration dependence
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
Porous silicaQ-calibration
1997Neutron Scattering LaboratoryNational Nuclear Energy Agency (BATAN)
SANS BATAN HMI
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
3
2
1
0
-1
-2
log
I
-2.5 -2.0 -1.5 -1.0 -0.5 0.0log Q
2005
HMI
Neutron Scattering LaboratoryNational Nuclear Energy Agency (BATAN)
SANS BATAN
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
Porous silicaQ-calibration
3
2
1
0
-1
-2
log
I
-2.5 -2.0 -1.5 -1.0 -0.5
log Q
Porasil CPG-10-75Data was taken usingSMARTer
-3.8
2006
HMISANS BATAN
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
Porous silicaQ-calibration
(late 2006)
HMI
Neutron Scattering LaboratoryNational Nuclear Energy Agency (BATAN)
3
2
1
0
-1
-2
log
I
-2.5 -2.0 -1.5 -1.0 -0.5log Q
Porasil CPG-10-75Data was taken using SMARTer11 December 2006
SANS BATAN
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
Porous silicaQ-calibration
Problems at SMARTer
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
• Reactor run at 15MW• Neutron flux > getting lower• No preset count measurement• Intensity in relative scale• Low flux at low q-range
• Modify the beam stopper & rubber cover.• Test a new monitor detector & pre-amp.• Build a new system for preset count exp. (hard & software)!• Make a new sample aperture.• Modify the live on data acquisition program.• Modify the data reduction program.• Make a new system for 6 sample changers with temp. control (software).• Build a new sample exchanger.• Do some experiments to optimize the SMARTer!• Flux measurements!• Local, regional & international network (CRP, TC, etc)
Dr. Kohlbrecher’s (IAEA expert mission – December 2006) recommendation;
Neutron Scattering Laboratory
National Nuclear Energy Agency of Indonesia (BATAN)
Things to do (future works)